Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf ·...

7
Measuring kinetic rate constants of multiple-component reactions with optical biosensors David A. Edwards a , Ryan M. Evans b, * , Wenbin Li a a Department of Mathematical Sciences, University of Delaware, Newark, DE 19716, USA b Applied and Computational Mathematics Division, Information and Technology Lab, National Institute for Standards and Technology, Gaithersburg, MD 20899, USA article info Article history: Received 20 March 2017 Received in revised form 14 June 2017 Accepted 16 June 2017 Available online 22 June 2017 Keywords: Optical biosensors Kinetic rate constants Mathematical modeling Parameter identication abstract One may measure the kinetic rate constants associated with biochemical reactions using an optical biosensor: an instrument in which ligand molecules are convected through a ow cell over a surface to which receptors are immobilized. If there are multiple reactants, one is faced with the problem of tting multiple kinetic rate constants to one signal, since data from all of the reacting species is lumped together. Even in the presence of ambiguous data, one may use a series of experiments to accurately determine the rate constants. Moreover, the true set of rate constants may be identied by either postprocessing the signals or adjusting the ligand inow concentrations. Published by Elsevier Inc. Introduction Many biochemical reactions in nature involve a stream of ligand molecules owing through a uid-lled volume over a surface to which receptors are conned. Such surface-volume reactions occur during platelet adhesion [1], drug absorption [2], and antigen- antibody interactions [3]. Fundamental to understanding these reactions is obtaining accurate quantitative measurements of the underlying kinetic rate constants. As seen in Fig. 1 , scientists use optical biosensors for measuring rate constants associated with surface-volume reactions. In a typical bisosensor experiment, ligand molecules are con- vected through a ow cell over a surface to which receptors are immobilized. Mass changes on the surface due to ligand binding are averaged over a portion of the channel oor to produce sensogram readings of the form S ðtÞ¼ sBðtÞ; (1.1) where BðtÞ denotes the concentration of bound ligand, and s is proportional to the molecular weight of the ligand. Recent advances in biosensor technology have enabled re- searchers to measure kinetic rate constants associated with multiple simultaneous independent reactions. Examples of such technology includes array-based [4,5] and multiple-channel [6e8] biosensors, and while these devices typically yield multiple inde- pendent measurements for each array-spot or channel, scientists are currently attempting to infer multiple kinetic rate constants asso- ciated with coupled reactions from a single signal. In particular, chemists are now using biosensor experiments to study how cells cope with DNA damage. Harmful DNA lesions can impair a cell's ability to replicate DNA. One way a cell may respond to such a lesion is through DNA translesion synthesis. For a com- plete description of this process we refer the interested reader to [9e11]; however, for our purposes it is sufcient to know that DNA translesion synthesis involves multiple interacting components: a Proliferating Cell Nuclear Antigen (PCNA) molecule, polymerase h, and polymerase d. Further, in order for a successful DNA translesion synthesis event to occur, polymerase h must bind with a PCNA molecule. A central question surrounding DNA translesion syn- thesis is whether polymerase h directly binds with the PCNA, or whether the polymerase h-PCNA complex forms as a result of a catalysis-type ligand switching process [12]. The former scenario is depicted in Fig. 2, which shows poly- merase h directly binding with a PCNA molecule, i.e. the reaction P 1 : E þ L 2 ) * 2 ka 2 k d EL 2 : (1.2a) * Corresponding author. E-mail address: [email protected] (R.M. Evans). Contents lists available at ScienceDirect Analytical Biochemistry journal homepage: www.elsevier.com/locate/yabio http://dx.doi.org/10.1016/j.ab.2017.06.009 0003-2697/Published by Elsevier Inc. Analytical Biochemistry 533 (2017) 41e47

Transcript of Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf ·...

Page 1: Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf · Measuring kinetic rate constants of multiple-component reactions with optical biosensors

lable at ScienceDirect

Analytical Biochemistry 533 (2017) 41e47

Contents lists avai

Analytical Biochemistry

journal homepage: www.elsevier .com/locate/yabio

Measuring kinetic rate constants of multiple-component reactionswith optical biosensors

David A. Edwards a, Ryan M. Evans b, *, Wenbin Li a

a Department of Mathematical Sciences, University of Delaware, Newark, DE 19716, USAb Applied and Computational Mathematics Division, Information and Technology Lab, National Institute for Standards and Technology, Gaithersburg, MD20899, USA

a r t i c l e i n f o

Article history:Received 20 March 2017Received in revised form14 June 2017Accepted 16 June 2017Available online 22 June 2017

Keywords:Optical biosensorsKinetic rate constantsMathematical modelingParameter identification

* Corresponding author.E-mail address: [email protected] (R.M. Evans).

http://dx.doi.org/10.1016/j.ab.2017.06.0090003-2697/Published by Elsevier Inc.

a b s t r a c t

One may measure the kinetic rate constants associated with biochemical reactions using an opticalbiosensor: an instrument in which ligand molecules are convected through a flow cell over a surface towhich receptors are immobilized. If there are multiple reactants, one is faced with the problem of fittingmultiple kinetic rate constants to one signal, since data from all of the reacting species is lumpedtogether. Even in the presence of ambiguous data, one may use a series of experiments to accuratelydetermine the rate constants. Moreover, the true set of rate constants may be identified by eitherpostprocessing the signals or adjusting the ligand inflow concentrations.

Published by Elsevier Inc.

Introduction

Many biochemical reactions in nature involve a stream of ligandmolecules flowing through a fluid-filled volume over a surface towhich receptors are confined. Such surface-volume reactions occurduring platelet adhesion [1], drug absorption [2], and antigen-antibody interactions [3]. Fundamental to understanding thesereactions is obtaining accurate quantitative measurements of theunderlying kinetic rate constants. As seen in Fig. 1, scientists useoptical biosensors for measuring rate constants associated withsurface-volume reactions.

In a typical bisosensor experiment, ligand molecules are con-vected through a flow cell over a surface to which receptors areimmobilized. Mass changes on the surface due to ligand binding areaveraged over a portion of the channel floor to produce sensogramreadings of the form

S ðtÞ ¼ sBðtÞ; (1.1)

where BðtÞ denotes the concentration of bound ligand, and s isproportional to the molecular weight of the ligand.

Recent advances in biosensor technology have enabled re-searchers to measure kinetic rate constants associated with

multiple simultaneous independent reactions. Examples of suchtechnology includes array-based [4,5] and multiple-channel [6e8]biosensors, and while these devices typically yield multiple inde-pendent measurements for each array-spot or channel, scientists arecurrently attempting to infer multiple kinetic rate constants asso-ciated with coupled reactions from a single signal.

In particular, chemists are now using biosensor experiments tostudy how cells cope with DNA damage. Harmful DNA lesions canimpair a cell's ability to replicate DNA. One way a cell may respondto such a lesion is through DNA translesion synthesis. For a com-plete description of this process we refer the interested reader to[9e11]; however, for our purposes it is sufficient to know that DNAtranslesion synthesis involves multiple interacting components: aProliferating Cell Nuclear Antigen (PCNA) molecule, polymerase h,and polymerase d. Further, in order for a successful DNA translesionsynthesis event to occur, polymerase h must bind with a PCNAmolecule. A central question surrounding DNA translesion syn-thesis is whether polymerase h directly binds with the PCNA, orwhether the polymerase h-PCNA complex forms as a result of acatalysis-type ligand switching process [12].

The former scenario is depicted in Fig. 2, which shows poly-merase h directly binding with a PCNA molecule, i.e. the reaction

P1 : E þ L2 ) *2ka

2kdEL2: (1.2a)

Page 2: Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf · Measuring kinetic rate constants of multiple-component reactions with optical biosensors

Fig. 1. Cross-sectional schematic of an optical biosensor experiment. The instrument has length l and height h. Ligand molecules are convected into instrument in a Poiseuille flowprofile and transported to the surface to bind with receptors. The receptors are limited to the reacting zone x ¼ xmin to x ¼ xmax.

Fig. 2. Direct binding of polymerase h with a PCNA molecule. We have labeled poly-merase h and the PCNA molecule as L2 and E respectively.

Fig. 3. Schematic of the ligand switching process. First, polymerase d binds with thePCNA molecule; next, polymerase h binds with the polymerase h-PCNA complex;finally, polymerase d dissociates, leaving us with the desired polymerse h-PCNAcomplex. We have labeled polymerase d, polymerase h, and the PCNA molecule as L1,L2, and E respectively.

D.A. Edwards et al. / Analytical Biochemistry 533 (2017) 41e4742

Here, the PCNA and polymerase hmolecules are denoted as E and L2respectively. Additionally, 2ka denotes the rate at which L2 bindswith a PCNA molecule E, and 2kd denotes the rate at which L2dissociates from a PCNA molecule E. We will refer to this aspathway one, or simply P1 as in (1.2a).

The ligand switching process is shown in Fig. 3 and statedprecisely as

P2 : E þ L1 ) *1ka

1kdEL1; EL1

þ L2 ) *12ka

12kd

EL1L2 ) *21kd

21ka

EL2 þ L1: (1.2b)

First, polymerase d (denoted L1) binds with a PCNA molecule;next polymerase h associates with the polymerase d-PCNA com-plex; finally polymerase d switches out, thereby leaving the desiredpolymerase h-PCNA complex. Each one of the steps in this multiple-component process is reversible. In (1.2b) and Fig. 3, the rate con-stants 1ka and 1kd denote the rates at which polymerase d binds andunbinds with a PCNA molecule (respectively), jika denotes the rateat which ligand Li binds with the product ELj, and

jikd denotes the

rate at which Li dissociates from the product EL1L2. In the latter twoexpressions the indices i and j can equal one or two. We shall referto this as pathway two, or simply P2 as in (1.2b).

By measuring the rate constants associated with this multiple-component process, one could determine whether EL2 forms as aresult of ligand binding or the ligand switching process describedabove. To measure the kinetic rate constants, scientists firstimmobilize PCNA molecules on the surface of the biosensor. Afterreceptor immobilization, scientists inject L1 and L2 into the in-strument at concentrations C1 and C2. The two ligands are thentransported to the surface to bind with available receptor sites E,creating the three reacting species EL1, EL1L2, and EL2 at concen-trations B1ðtÞ;B12ðtÞ; and B2ðtÞ.

The presence of multiple reacting species changes the inter-pretation of the sensogram reading. Most biosensors, including theBIAcore, measure only mass changes at the surface due to ligandbinding; hence (1.1) becomes

S ðtÞ ¼ s1B1ðtÞ þ ðs1 þ s2ÞB12ðtÞ þ s2B2ðtÞ; (1.3)

where si is proportional to the molecular weight of Li, for i ¼ 1; 2.The lumped signal (1.3) complicates parameter estimation. Sinceone is faced with the challenge of fitting multiple rate constants toone signal, it may be difficult for standard algorithms to find aunique solution to the corresponding least-squares problem.

Indeed, the problem of inferring multiple kinetic rate constantsfrom one signal is an interesting problem which has received littleattention to date. This problem is addressed [13,14], although bothof these papers are on multiple receptor kinetics. Multiple-ligand

Page 3: Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf · Measuring kinetic rate constants of multiple-component reactions with optical biosensors

D.A. Edwards et al. / Analytical Biochemistry 533 (2017) 41e47 43

kinetics are discussed briefly in [15, pp.101e102]. Svitel et al. pro-pose amethod for inferring a distribution of kinetic rate constants ina model that could be used to describe multiple-ligand kinetics ormultiple receptor sites for the same ligand in Ref. [16]. However,their results differ from the present work in at least three respects.In Ref. [16], the authors consider a continuous distribution offunctionally distinct ligand molecules. By contrast, since we areconcerned with elucidating the role of polymerase h during DNAtranslesion synthesis, we are considering two distinct ligands.Additionally, their approach is based upon Tikhonov regularization,which introduces a bias into parameter estimation. Since this maylead to erroneous results, it is important that Tikhonov regulari-zation is combined with a priori knowledge about the system athand. Since conventional techniques such as flourescent micro-scopy may modify protein activity, a priori estimates of the rateconstants are unavailable, and methods not based on Tikhonovregularzation are necessary. Finally, in Ref. [16] the authorsconsider systems in which reactions associated with differentligand molecules proceed independently, which is clearly not thecase for the application that we have in mind. In fact, if the re-actions (1.2) decoupled then one could use standard techniques tomeasure the kinetic rate constants and the role of polymerase h inDNA translesion synthesis would be understood.

We explore the difficulty of fitting the rate constants in (1.3) bydeveloping a mathematical model for the multiple-ligandbiosensor experiment described herein. This mathematical modelis used to show that there are two different sets of kinetic rateconstants which correspond to very similar sensogram signals.Moreover, by proposing and numerically simulating a set of fourexperiments, it is shown that one can recover the true rate con-stants associated with each of the signals. Furthermore, it is shownthat the sensogram readings can be disambiguated by post-processsing or varying the ligand inflow concentrations.

Mathematical model, numerical experiments, and results

The surface reactions (1.2) are modeled with a set of OrdinaryDifferential Equations (ODEs):

dB1dt

¼ 1kaðRt � BSÞC1 þ 12kdB12 � 1kdB1 � 1

2kdB1C2; (2.1a)

dB12dt

¼ 12kaB1C2 þ 2

1kaB2C1 � 12kdB12 � 2

1kdB12; (2.1b)

dB2dt

¼ 2kaðRt � BSÞC2 þ 21kdB12 � 2kdB2 � 2

1kaB2C1: (2.1c)

Here BS ¼ B1 þ B12 þ B2. The system (2.1) is subject to the initialconditions

B1ð0Þ ¼ B1;0;B12ð0Þ ¼ B12;0;B2ð0Þ ¼ B2;0: (2.2)

The system (2.1) and (2.2) is expressed more compactly in ma-trix vector form as

dBdt

¼ �ABþ f; Bð0Þ ¼ B0; (2.3)

BðtÞ ¼ ðB1ðtÞ;B12ðtÞ;B2ðtÞÞT :

Here, B0 ¼ ðB1;0;B12;0;B2;0ÞT is a vector containing the three initialconditions and

A¼0@ð1kaC1þ 1kdþ 1

2kaC2Þ 1kaC1�12kd 1kaC1

�12kaC2 ð12kdþ 2

1kdÞ �21kaC1

2kaC2 2kaC2�21kd ð2kaC2þ 2kdþ 2

1kaC1Þ

1A;

f ¼ ð1kaRtC1;0; 2kaRtC2ÞT :The ODE model (2.3) assumes that reaction kinetics are accu-

rately described by the well-stirred approximation. Since masstransport effects on ligand binding are typically minimal inexperimental regimes, they are not considered herein. For a dis-cussion of transport effects on multiple-component reactions inoptical biosensors one may see Refs. [13,17].

The gold standard for verifying the model is comparison withexperimental data. At press time, the PCNA experiments had notbeen completed. Therefore, we use synthetic data from numericalsimulations of the model (2.3) using an ODE solver in MATLAB. Thisdata was then used to generate the signal (1.3), which is writtenconcisely as

SðtÞ ¼ sTBðtÞ; (2.5)

s ¼ ðs1; s1 þ s2; s2ÞT :Even though the expression for B from (2.3) is in some sense

“exact”, the synthetic sensogram data S which results still lumpsthe solutions into a single signal from which the rate constantsmust be estimated, either with or without postprocessing fordisambiguation.

In the numerical experiments, ligands L1 and L2 were injectedsequentially. First, L1 was injected until (2.5) reached a steady state.The injection of L1 was stopped simultaneously with starting theinjection of L2, which continued until the signal sensogram signalreached a steady state again. Therewas no bound ligand at the startof the first injection (so B0 ¼ 0), while the initial condition for thesecond injection was the steady-state solution to (2.3) during thefirst injections. This numerical experiment was performed for twosets of parameters, which are labeled Signal 1 and Signal 2 in theexact column of Table 1. The corresponding signals are depicted inFigs. 4 and 5, and are labeled Signal 1 and Signal 2 in accordancewith the parameter values in Table 1.

First consider signal one, and notice the increase in the senso-gram reading upon the start of the second injection. Examining thebound state concentrations depicted in Fig. 4, it is evident that theincrease is due to a large influx of EL2. Since the rate constant 2ka isan order of magnitude larger than the others, L2 molecules almostimmediately bind with available receptor sites at the start of thesecond injection.

Now consider signal two; again, there is a sharp and suddenincrease in the sensogram reading at the start of the second in-jection. Since the rate constant 1

2ka is an order of magnitude largerthan the others, L2 molecules almost immediately bind with the EL1present upon the start of the second injection. This behavior isevident in the bound state concentrations depicted in Fig. 5: once L2is injected there is a sudden decrease in B1, and a sudden increasein B12.

Although the sensogram signals correspond to different sets ofrate constants, they appear quite similar to the eye. Indeed, Fig. 6indicates that the maximum relative difference between the tworeadings is approximately five percent. This is significant in light ofthe fact that true physical experimental data may be contaminatedby noise.

Page 4: Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf · Measuring kinetic rate constants of multiple-component reactions with optical biosensors

Table 1The values in the column labeled “Exact” were used to generate the signals 2.1 and 2.2. The notation C1 ¼ 10�11;0;mol=cm3 indicates that a value of C1 ¼10�11mol=cm3 was used during the first injection, and a value of C1 ¼ 0 mol=cm3 was used during the second; analogous notation is used for C2. The valuesin the column labeled “Fitted” are those recovered from our parameter estimation algorithm described on pages six and seven.

Parameter Signal 1 Signal 2 Signal 1 Signal 2 Reference

Exact Fitted

1kaðcm3=ðmol,sÞÞ 108 1:00� 108 [4,18,19]

1kdðs�1Þ 10�3 0.0010 [4,18,19]

1kaðcm3=ðmol,sÞÞ 3� 109 108 3:00� 109 1:00� 108 [4,18,19]

1kdðs�1Þ 10�3 10�5 0.0010 9:99� 10�6 [4,18,19]12kaðcm3=ðmol,sÞÞ 108 3� 109 1:00� 108 3:00� 109 [4,18,19]12kdðs�1Þ 10�3 9:99� 10�4 [4,18,19]21kaðcm3=ðmol,sÞÞ 108 [4,18,19]21kdðs�1Þ 10�3 9:99� 10�4 [4,18,19]

Rtðmol=cm2Þ 10�12 [4,19]

C1ðmol=cm3Þ 10�11 ; 0 [4]

C2ðmol=cm3Þ 0, 10�11 [4]

s1ðkDaÞ 220.2 [20e22][23]

s2ðkDaÞ 71.5 [23,24]

Fig. 4. Left: Sensogram signal 1. Ligand L1 was injected until the signal (2.5) reached steady state at t ¼ 5000. At this point we stopped injecting L1, and injected L2 until the signalreached steady state again at t ¼ 15000. The time t is measured in seconds. Right: corresponding bound ligand concentrations.

Fig. 5. Left: Sensogram signal 2, generated in the same manner as the reading in Fig. 4. Right: Corresponding bound ligand concentrations.

D.A. Edwards et al. / Analytical Biochemistry 533 (2017) 41e4744

Page 5: Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf · Measuring kinetic rate constants of multiple-component reactions with optical biosensors

Fig. 6. Relative difference between signal 1 and signal 2. Since the signals are identicalduring the first injection, we have depicted the relative difference during the secondinjection only; i.e., from t ¼ 5000 to t ¼ 15000.

D.A. Edwards et al. / Analytical Biochemistry 533 (2017) 41e47 45

Since the two parameter regimes identified in Table 1 lead tonearly identical sensogram signals, it may be difficult for standardalgorithms to identify the true set of rate constants. Therefore, a setof four numerical experiments has been performed to identify thetrue set of rate constants. Since true experimental data is stillforthcoming, the efficacy of this procedure is demonstrated bynumerically simulating these four experiments with our mathe-matical model.

The first experiment consisted of injecting L1 only, and wassimulated by solving the ODE

dB1dt

¼ 1kaðRt � B1ÞC1 � 1kdB1: (2.6)

We used lsqcurvefit in MATLAB to fit the parameters 1ka and 1kdin (2.6) to data from the first numerical experiment. Then, anexperiment which consisted of injecting of L2 onlywas simulated inan analogous manner to the first experiment. The second experi-ment generated estimates of 2ka and 2kd. Next, a third experimentwas done by simulating the injection of L1 until the signal reached asteady-state, and then halting injection of L1 and immediatelystarting injection of L2 until the signal reached a steady-state again(analogous to the simulations in Figs. 4 and 5). The third experi-ment generated estimates of 12ka;

12kd, and

21kd. A fourth experiment

analogous to the third can be done to determine 12ka by switching

the roles of L1 and L2 in experiment three. Due to the similarity ofthis experiment to experiment three, this computation has beenomitted for simplicity.

Despite the closeness of sensogram signals 1 and 2, the columnlabeled “Fitted” in Table 1 reveals the results from our parameterestimation algorithm agree quite well with the original valuesrecorded in the “Exact” columnei.e., those used in our numericalexperiments depicted in Figs. 4 and 5. Our parameter estimationalgorithm is robust with respect to noise; however, the resultsincluded herein were computed without the addition of noise.

Although the proposed parameter estimation procedurecorrectly identifies the set of rate constants corresponding to eachof the signals, in practice it may be desirable to disambiguate apotentially unclear sensogram signal before parameter estimation.For the two signals considered herein, this can be done throughpostprocessing. To motivate how this is done, recall that the jumpin signal one results from L2 binding with available receptor sites.By excluding this from the signal (2.5), the resulting sensogramreading will not exhibit a sudden increase at the start of the secondinjection. Hence, the signal was postprocessed by subtracting the

time-dependent quantity s2~B2ðtÞ from (2.5) for t � 5000 (i.e., dur-ing the second injection). The function ~B2 satisfies

d~B2dt

¼ 2ka

�Rt

�1� 1kaC1

1kaC1 þ 1kd

�� ~B2

�C2 � 2kd~B2; (2.7)

~B2ð0Þ ¼ 0:

Note that in (2.7) the factor

Rt 1kaC1

1kaC1 þ 1kd(2.8)

has been subtracted from the total number of receptors Rt. Thefactor (2.8) is simply the steady state of (2.6), and hence corre-sponds to the number of receptors filled by B1 at the beginning ofthe second state. Hence the parenthetical quantity in (2.7) repre-sents the receptors available at the start of the second injection.

This subtraction excludes EL2 forming as a result of L2 bindingwith empty receptors at the start of the second injection. Thepostprocessed signals are depicted in Fig. 7. By excluding the influxof EL2 at the start of the second injection, signal one no longerexhibits an obvious increase at the start of the second injection.Conversely, even after postprocessing a sudden increase in senso-gram two is still evidentethis is due to the fact that the jump insignal two results from a large influx of EL1L2, not an influx EL2.

The two signals in Figs. 4 and 5 can also be disambiguated byadjusting the ligand inflowconcentrations. To demonstrate this, thesame numerical experiments as in Figs. 4 and 5 wereconducted, however the ligand concentrations were simulata-neously simultaneously reduced to C1 ¼ 4:5� 10�12 mol=cm3 andC2 ¼ 4:5� 10�12 mol=cm3. These values are 45 percent of thosegiven in Table 1, and all of the other values remained the same. Theresulting sensogram readings are depicted in Figs. 8 and 9.

First consider signal 1: notice that the increase seen at the startof the second injection is more pronounced than in signal 2. SinceC1 is less than half of its original value, there are fewer unbound L1molecules at the surface of the biosensor available for binding; inturn B1 reaches a lower steady-state value. This implies that thequantity Rt � B1 is larger at t ¼ 5000, and there are more emptyreceptor sites for L2 molecules to bind with at the start of thesecond injection. The latter results in the pronounced jump,because the rate constant 2ka is an order of magnitude larger thanthe other association constants and pathway one (1.2a) isprominent.

In sensogram two, there was no obvious increase at the start ofthe second injection. A comparison with Fig. 4 shows that this isbecause less EL1L2 forms. Since less EL1 is present for L2 moleculesto bind with, the jump at the start of the second injection is not aspronounced. This is due to the fact the rate constant 1

2ka is an orderof magnitude larger than the other association constants, sopathway two (1.2b) is prominent.

Conclusions

Scientists are currently attempting to determine the pathway ofpolyermase h during DNA translesion synthesis using opticalbiosensor experiments. The presence of multiple reacting specieson the sensor surface complicates the associated parameter esti-mation problem. Using a mathematical model, two parametersregimes that lead to nearly identical sensogram signals have beenidentified. By proposing and numerically simulating a set of fourexperiments, it has been shown that it is possible to recover thetrue set of rate constants.

Furthermore, it has been shown that the signals may be

Page 6: Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf · Measuring kinetic rate constants of multiple-component reactions with optical biosensors

Fig. 7. The signals depicted in Figs. 4 and 5 after postprocessing. In both figures the postprocessed signal is depicted by the dotted line, and the original signal (shown in Figs. 4 and5) is depicted by the dashed line.

Fig. 8. Left: Signal 1 after adjusting ligand inflow concentrations. Right: Corresponding bound state concentrations.

Fig. 9. Left: Signal 2 after adjusting ligand inflow concentrations. Right: Corresponding bound state concentrations.

D.A. Edwards et al. / Analytical Biochemistry 533 (2017) 41e4746

disambiguated in two ways. An estimate for binding from onepathway can be subtracted from the data, revealing the bindingfrom the other pathway. Alternatively, one may use differing valuesof the inflow concentrations to isolate binding in each pathway.

These results are motivated by the study of the role of polyermase hduring DNA translesion synthesis; however, they may are genericenough to be applied to other multiple-component biosensorexperiments.

Page 7: Measuring kinetic rate constants of multiple-component ...edwards/download/pubdir/pubj51.pdf · Measuring kinetic rate constants of multiple-component reactions with optical biosensors

D.A. Edwards et al. / Analytical Biochemistry 533 (2017) 41e47 47

Future work will include incorporating mass-transport effectsinto our parameter estimation algorithm. Moreover, wewill seek todevelop MATLAB software to fit the rate constants in (1.2) to trueexperimental data, and use this software to verify the post-processing strategy discussed herein.

Funding

The authors were supported by the National Science Founda-tion, under award number NSF-DMS 1312529. In addition, thesecond author was partially supported by the National ResearchCouncil, in the form of a postdoctoral fellowship.

Disclaimer

This paper is an official contribution of the National Institute ofStandards and Technology and is not subject to copyright in theUnited States. Commercial products are identified in order toadequately specify certain procedures. In no case does such iden-tification imply recommendation or endorsement by the NationalInstitute of Standards and Technology, nor does it imply that theidentified products are necessarily the best available for thepurpose.

References

[1] S. Austin, Haemostasis, Medicine 37 (3) (2009) 133e136.[2] C. Bertucci, A. Piccoli, M. Pistolozzi, Optical biosensors as a tool for early

determination of absorption and distribution parameters of lead candidatesand drugs, Comb. Chem. High Throughput Screen. 10 (6) (2007) 433e440.

[3] M. Raghaven, M.Y. Chen, L.N. Gastinel, P.J. Bjorkman, Investigation of theinteraction between class I MHC-related F_c receptor and its immunoglobulinG ligand, Immunity 1 (4) (1994) 303e315.

[4] R.L. Rich, M.J. Cannon, J. Jenkins, P. Pandian, S. Sundaram, R. Magyar,J. Brockman, J. Lambert, D.G. Myszka, Extracting kinetic rate constants fromsurface plasmon resonance array systems, Anal. Biochem. 373 (1) (2008)112e120.

[5] M. Zumbrum, D.A. Edwards, Multiple surface reactions in arrays with appli-cations to optical biosensors, Bull. Math. Biol. 76 (7) (2014) 1783e1808.

[6] C. Boozer, J. Ladd, S. Chen, S. Jiang, DNA-directed protein immobilization forsimultaneous detection of multiple analytes by surface plasmon resonance

biosensor, Anal. Chem. 78 (5) (2006) 1515e1519.[7] J. Dost�alek, H. Vaisocherov�a, J. Homola, Multichannel surface plasmon reso-

nance biosensor with wavelength division multiplexing, Sensors Actuators BChem. 108 (1) (2005) 758e764.

[8] J. Homola, H. Vaisocherov�a, J. Dost�alek, M. Piliarik, Multi-analyte surfaceplasmon resonance biosensing, Methods 37 (1) (2005) 26e36.

[9] E.C. Friedberg, Suffering in silence: the tolerance of DNA damage, Nat. Rev.Mol. Cell Biol. 6 (12) (2005) 943e953.

[10] A.R. Lehmann, A. Niimi, T. Ogi, S. Brown, S. Sabbioneda, J.F. Wing,P.L. Kannouche, C.M. Green, Translesion synthesis: Y-family polymerases andthe polymerase switch, DNA repair 6 (7) (2007) 891e899.

[11] B.S. Plosky, R. Woodgate, Switching from high-fidelity replicases to low-fidelity lesion-bypass polymerases, Curr. Opin. Genet. Dev. 14 (2) (2004)113e119.

[12] Z. Zhuang, R.E. Johnson, L. Haracska, L. Prakash, S. Prakash, S.J. Benkovic,Regulation of polymerase exchange between Polh and Pold by mono-ubiquitination of PCNA and the movement of DNA polymerase holoenzyme,Proc. Natl. Acad. Sci. 105 (14) (2008) 5361e5366.

[13] R. M. Evans, D. A. Edwards, Receptor heterogeneity in optical biosensors,J. Math. Biol., Accepted.

[14] J. Svitel, H. Boukari, D. Van Ryk, R.C. Willson, P. Schuck, Probing the functionalheterogeneity of surface binding sites by analysis of experimental bindingtraces and the effect of mass transport limitation, Biophysical J. 92 (5) (2007)1742e1758.

[15] R.L. Rich, D.G. Myszka, Extracting kinetic rate constants from binding re-sponses, in: M.A. Cooper (Ed.), Label-free Bioesnsors, Cambridge UniversityPress, 2009.

[16] J. Svitel, A. Balbo, R.A. Mariuzza, N.R. Gonzales, P. Schuck, Combined affinityand rate constant distributions of ligand populations from experimentalsurface binding kinetics and equilibria, Biophysical J. 84 (6) (2003)4062e4077.

[17] R. M. Evans, D. A. Edwards, Transport Effects On Multiple-Component Re-actions in Optical Biosensors, Submitted.

[18] General Electric Life Sciences, GE Healthcare Bio-sciences AB, Bj€orkgatan, vol.30, April 2013, 751 84 Uppsala, Sweden, BIAcore T200 data file.

[19] M.L. Yarmush, D.B. Patankar, D.M. Yarmush, An analysis of transport re-sistances in the operation of BIAcore; implications for kinetic studies of bio-specific interactions, Mol. Immunol. 33 (15) (1996) 1203e1214.

[20] UniProt, DNA Polymerase Delta Catalytic Subunit. URL http://www.uniprot.org/uniprot/P15436.

[21] UniProt, Dna Polymerase Delta Small Subunit. URL http://www.uniprot.org/uniprot/P46957.

[22] UniProt, Dna Polymerase Delta Subunit 3. URL http://www.uniprot.org/uniprot/P47110.

[23] Protein Information Resource, Composition/molecular Weight Calculation.URL http://pir.georgetown.edu/pirwww/search/comp_mw.shtml.

[24] UniProt, DNA Polymerase Eta. URL http://www.uniprot.org/uniprot/Q04049.