Measurement Control and DAQ for Superconductors and Magnet labs

SCLabs 20-05-09 Adriaan Rijllart 1

Adriaan Rijllart, for the EN-ICE-MTA section

Page 2: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Outline

Status of control and DAQ systems for SC cable and magnet measurements

Maintenance plan (to keep present functionality) Development plan (to comply with new

requirements and assure long term functionality)

SCLabs 20-05-09 Adriaan Rijllart 2

Page 3: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Range of systems

Fixed test systems – SM18, Block 4, B163 Mobile systems – SM18, Cryolab (B165)

Not covered: systems for warm measurement such as for the Magnet Rescue Facility in B180, B181 and MQW in B867.

SCLabs 20-05-09 Adriaan Rijllart 3

Page 4: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Fixed measurement systems:

SCLabs 20-05-09 Adriaan Rijllart 4

Control Room

Test Bench

SM18

Page 5: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Fixed measurement systems:

SCLabs 20-05-09 Adriaan Rijllart 5

B163Block 4

Page 6: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Mobile measurement systems:

SCLabs 20-05-09 Adriaan Rijllart 6

Cryolab Mobile BenchSM18 High Voltage Test

Page 7: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Inventory of measurement systemsType Hardware OS Application Units

Quench DAQ Sun + VME Solaris HFDAQ, LFDAQ 7

Magnetic M Sun + VME Solaris MMP 16

High Voltage T Sun + GPIB Solaris HVT 6

Splice/Loss Sun + GPIB Solaris SRM/Loss 2

Snapback M PXI Pharlap B3B5 Snapback 1

HTS DAQ Mac + SCXI Mac OS HTS 2

Correctors DAQ Mac + SCXI Mac OS CQR 2

Diode DAQ PC + PCI Windows Diode 1

FRESCA Sun + GPIB Solaris FRESCA 1

M(H)/RRR/Rc Sun + GPIB Solaris Magnetisation/RRR/Rc 3

Critical current Sun + GPIB Solaris IC 4

Mobile Cryolab Mac + PCI Mac OS PMC 7

Mechanical M PC + GPIB Windows HBM ?

SCLabs 20-05-09 Adriaan Rijllart 7

Page 8: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Risk analysis of meas. systems

SCLabs 20-05-09 Adriaan Rijllart 8

Concern Problem Impact Prob Wght

LabVIEW on Solaris

Discontinued 5 years ago. But so far it keeps working.

High Low 3

Solaris No longer supported at CERN since 1 year. High Low 3

SBS34 Discontinued 4 years ago. Porting to another module requires difficult work.

Medium High 6

BAB740 Discontinued 4 years ago. BAB760 is produced and could replace the BAB740 with some work.

Medium High 6

VME 30 years old standard. Becoming obsolete. Produced by few companies. Expensive.

Low Medium 2

Page 9: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Measurement system Sun-VME

SCLabs 20-05-09 Adriaan Rijllart 9

EthernetEthernet

VME integratorsVME integrators

Power SupplyPower Supply

DVMsDVMsLossLoss M. M.

Operator ConsoleOperator Console

Quench DAQQuench DAQMagnetic MeasurementMagnetic Measurement

Sun

Sun Sun

VME ADCsVME ADCs

Page 10: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Upgrade to PXI

SCLabs 20-05-09 Adriaan Rijllart 10

EthernetEthernet

IntegratorsIntegratorsDVMsDVMs

Power SupplyPower Supply

Operator ConsoleOperator Console

Quench DAQQuench DAQMagnetic MeasurementMagnetic Measurement

ADCsADCs

Page 11: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Partial upgrade, keeping some VME

SCLabs 20-05-09 Adriaan Rijllart 11

EthernetEthernet

DVMsDVMs

Operator ConsoleOperator Console

Quench DAQQuench DAQMagnetic MeasurementMagnetic Measurement

ADCsADCs

VME IntegratorsVME Integrators

Page 12: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Reasons to adopt PXI Industry standard for instrumentation, based on PCI Supported by 60 vendors, commercialising >1000 modules PXI is 10 years old Smaller, cheaper than VME Key hardware features: triggering and synchronisation, shielding,

speed (PXI-express, follows PCI express development) Key software features by National Instruments: LabVIEW, LabVIEW

Real Time, integrated network communication, drivers Key OS features: Pharlap, VX-Works, Windows, Linux Connects to VME, which allows partial migration, keeping VME

investment where needed

SCLabs 20-05-09 Adriaan Rijllart 12

Page 13: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Maintaining present systems Step 1: Replace Sun with PXI, keep MXI-VME Advantages:

Up to date LabVIEW version Up to date OS (no Solaris)

Disadvantage: VME processors, modules and crates are kept (some

are obsolete)

SCLabs 20-05-09 Adriaan Rijllart 13

Page 14: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Upgrade plan Step 2: Replace VME with PXI

More compact, higher density modules 3 VME crates with 56 modules Replace 5 VME ADC with 1 PXI ADC

Higher performance ADCs Higher precision, up to 24 bit ADCs and 7 ½ digits DVMs Faster acquisition, if needed

Maintenance advantages: No SBS, BAB, VME, MXI

Disadvantage: higher cost

SCLabs 20-05-09 Adriaan Rijllart 14

Page 15: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Cost and resourcesMaintenance plan Unit cost

CHF# units Total cost CHF Manpower

(ManMonths)

SM18 + Block 4 replace Sun workstations

10 k 36 360 k 6

B163 PXI systems 10 k 8 80 k 1

SCLabs 20-05-09 Adriaan Rijllart 15

Upgrade plan Unit cost CHF

# units Total cost CHF Manpower (ManMonths)

SM18 DAQ 35 k 7 245 k 12

SM18 MM ? 9 ? ?

Block 4 DAQ 20 k 2 40 k 4

Block 4 MM ? 2 ? ?

B163 DVMs 3 k 16 48 k 4

Page 16: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Replacement of all Sun based systems

All Sun based measurement systems will need to be replaced (B180 Magnet Rescue Facility, B181, B867, …)

Not part of the SC laboratory review, but to be kept in mind

My inventory has 90 Sun workstations

SCLabs 20-05-09 Adriaan Rijllart 16

Page 17: Measurement Control and DAQ for Superconductors and Magnet labs SCLabs 20-05-09Adriaan Rijllart1 Adriaan Rijllart, for the EN-ICE-MTA section.

Recommendation Plan 1: maintenance.

Replace all Sun measurement stations with PXI processors, keeping the VME crates and modules.

Plan 2: upgrade. Develop one pilot DAQ system in PXI for validation. Replace all DAQ systems with PXI system.

B163 Plan 1: maintenance, replace Sun workstations. Plan 2: replace DVMs with PXI if needed for higher speed.

B165 Can be maintained as is (no Sun workstations) For new measurement systems PXI should be considered.

SCLabs 20-05-09 Adriaan Rijllart 17