ME 322: Instrumentation Lecture 16

15
ME 322: Instrumentation Lecture 16 February 25, 2015 Professor Miles Greiner Lab 6 calculations (Excel demo)

description

ME 322: Instrumentation Lecture 16. February 26, 2014 Professor Miles Greiner. Announcements/Reminders. HW 6 due Friday No lab this week Career Fair, Tomorrow , February 27, 2014 Internships Prepare for permanent employment next year www.unr.edu/engineering/careerfair - PowerPoint PPT Presentation

Transcript of ME 322: Instrumentation Lecture 16

Page 1: ME 322: Instrumentation Lecture 16

ME 322: InstrumentationLecture 16

February 25, 2015Professor Miles Greiner

Lab 6 calculations (Excel demo)

Page 2: ME 322: Instrumentation Lecture 16

Announcements/Reminders

• HW 6 due Friday• Career Fair, Tomorrow, February 26, 2015

– Internships– Prepare for permanent employment next year– www.unr.edu/engineering/careerfair– Differential Fees pay for this

Page 3: ME 322: Instrumentation Lecture 16

Lab 6 Air Volume Flow Rate and Centerline Speed in a Wind Tunnel

• Plexiglas Tube and Schedule-40 Pipe have different diameters• Control flow rate using a variable speed blower

– Cover blower exit for very low speeds• For a range of flow rates, measure

– Volume flow Q rate using a Presso Venturi Tube (in pipe)– Centerline speed VC using a Pitot-Static Tube (in Plexiglas tube)

• For both measure pressures difference using calibrated transmitters/digital multimeters

• Both VC and Q increase with blower flow rate– Is VS < VC < VP?

Page 4: ME 322: Instrumentation Lecture 16

Instrument Schematic

• Measure atmospheric conditions PATM and TATM – Using hand-held digital-barometer – = 0.5 kPa, = 1°C (95%?)

• Need PStat to find pressure– Is PStat < = or > than PATM?– Use 40-in-WC transmitter to find Gage Pressure PG = PATM – PS – PS = PATM - PG

• To measure Pitot-Static tube pressure difference PP – Use 3-in-WC transmitter

• To measure Venturi tube pressure difference PV – Use 40-in-WC transmitter

DTubeDPipe

PP PG

PV Static

Total+ +--IV

IP IG

+-

Variable SpeedBlower Pipe

PlexiglasTube

Pitot-Static Probe VC

Venturi TubeQ

Atm40 in WC

3 in WC 40 in WC

BarometerPATM TATM

Page 5: ME 322: Instrumentation Lecture 16

Summary• Before Experiment• Measure tube diameter• Calculate transmitter uncertainties• Use hand held barometer to measure

– PATM

– TATM • °C

Tatm Patm Dpipe Dtube Apipe Atube K W40-inch W3-inch

[°C] [mbar] [inch] [inch] [m2] [m2] [-] [Pa] [Pa]22 873 2.07 2.25 0.002165 0.002565 0.381 25 1.9

Page 6: ME 322: Instrumentation Lecture 16

During Experiment• For each blower setting measure currents and find the values &

uncertainties – Pressure Measurements

• P = rgh = rg(FS)(I – 4mA)/16 mA• WP = 0.0025rg(FS)

– Static Pressure, PStat = WOB (Work on Board)• WOB• WOB

– Air density WOB• WOB

– Volume flow rate WOB• WOB

– Centerline speed WOB• WOB

– Check Pipe Reynolds numbers, = • (300 K)• Venturi calibration (KPresso) is within 2% for 54,000 < < 137,000

Blower Condition

IV

[mA]IP

[mA]IG

[mA]Blower off 4.01 4.02 4

1 8.9 14.9 52 8.59 14.34 4.993 8.33 13.17 4.884 7.74 12.05 4.785 7.15 11.4 4.726 6.74 10.87 4.687 6.39 10.4 4.628 6.09 8.38 4.439 5.59 7.96 4.410 5.09 6.18 4.21

Blower off 4.02 4.02 4.01

Page 7: ME 322: Instrumentation Lecture 16

Consistency Check

• For a given volume flow rate Q– VS = Q/A– VP = 2VS

• What area should we use?– APipe or ATube

Page 8: ME 322: Instrumentation Lecture 16

Measured Results

• Determine speed and flow rate uncertainty for a range of blower speeds

Page 10: ME 322: Instrumentation Lecture 16
Page 11: ME 322: Instrumentation Lecture 16

Pressure Transmitter Uncertainty• Pressure

• = 998.7 kg/m3, g = 9.81 m/s2

• FS = (3 or 40 inch)

• Manufacturer stated uncertainty: 0.25% Full Scale– (95%?)– For FS = 3 inch WC

• PFS = rWghFS = (998.7 kg/m3)(9.81 m/s2) (3 inch) = 746.6 Pa• wP = 0.0025 PFS = 1.9 Pa

– For FS = 40 inch WC• PFS = rWghFS = (998.7 kg/m3)(9.81 m/s2) (40 inch) = 9954 Pa• wP = 0.0025 PFS = 25 Pa

Page 12: ME 322: Instrumentation Lecture 16

Static Pressure• PStat = PATM – PG

– Use for , RAir = 0.2870 kPa-m3/kg-K– So want PStat in [kPa]

• Inputs– PATM

• Measure using barometer• = 500 Pa = 0.5 kPa (95%)

– PGAGE • Measure using 40 inch WC gage• = 25 Pa = 0.025 kPa (95%)

Page 13: ME 322: Instrumentation Lecture 16

Static Pressure Uncertainty

• PStat = PATM – PG (Linear Sum?)

Page 14: ME 322: Instrumentation Lecture 16

Transmitter Pressures

3” or 40” WC

FS Wρ

Pitot 3” WC 1.9 PaVenturi

and Gage

40” WC 25 Pa

Page 15: ME 322: Instrumentation Lecture 16

Gas Pressure and Density