MDS Alignment All 3 modes have their own alignment.

10
MDS Alignment All 3 modes have their own alignment

Transcript of MDS Alignment All 3 modes have their own alignment.

Page 1: MDS Alignment All 3 modes have their own alignment.

MDS Alignment

All 3 modes have their own alignment

Page 2: MDS Alignment All 3 modes have their own alignment.

Photo Mode**Press MDS PHOTO button**

Set up all imaging conditions:

• Spot Size (1 or 2)

• Alpha Angle (2 or 3)

• Magnification (40-60k)

• Dose (15-20 e-/Å2/sec)

After you set up dose, record your C3 lens value so that you can find the correct dose without having to recalculate it

Page 3: MDS Alignment All 3 modes have their own alignment.

How to calculate specimen and image dose and dose rate

Example:

Io = 15 e/Å2/sec

Mi = 40,000x

If M s = 1.25 for screen

1 piAmp/cm2 = 625 x 10-12 e/Å2/sec

Is = 23.4 PiAmp/cm2

Page 4: MDS Alignment All 3 modes have their own alignment.

Relationship of dose or dose rate between object and detector

Object intensity = I0 = Ii Mi2

Ii = I0/ Mi2 Ii = Is/ Ms

2

Screen intensity = Is = I0 Ms2/ Mi

2

Page 5: MDS Alignment All 3 modes have their own alignment.
Page 6: MDS Alignment All 3 modes have their own alignment.

Dose Calculation (e-/Å2/sec)

Go to a completely open area (no ice or carbon, just vacuum)

Go to 50,000x Magnification

*Your dose at 50k in pA/cm2 is equal to

the dose in e-/Å2/sec

Adjust the following settings to obtain the

desired dose for your sample:

• Size of your beam (using Brightness knob)

• Spot Size

• Exposure time

Page 7: MDS Alignment All 3 modes have their own alignment.

MDS Focus

**Press MDS FOCUS button**

Should be set up the same way as MDS PHOTO

• Be in MAG 1

• Select desired spot size, alpha angle and magnification

*generally these settings should all the be same as the settings in PHOTO mode

Page 8: MDS Alignment All 3 modes have their own alignment.

MDS Search

**Press MDS SEARCH button**• Go to MAG 1

• Repeat Voltage Axis Alignment

• Go to Spot Size of 5 (you may also have to do the spot size alignment at this point if they are misaligned)

• Spread beam clockwise (using Brightness)

until you hear a beep

• Press SA DIFF

• Adjust DIFF FOCUS knob until the beam

is the desired size

Page 9: MDS Alignment All 3 modes have their own alignment.

Aligning MDS Photo and MDS Search

• After setting up MDS SEARCH mode, go back to MDS PHOTO

• Find an area of interest such as ice contamination

• Center that area in Photo mode using the trackball

• Press MDS FOCUS

• Press MDS SEARCH

• If the area of interest is not in the center than press PLA and use your DEF/STIG X and Y knobs to bring the area to the center

• Turn off PLA

Page 10: MDS Alignment All 3 modes have their own alignment.

Alignment of Search Mode

Photo Mode Misaligned Search Mode:Use PLA & def/stig x y to center area of interest

Aligned Search Mode