CL560 MULTIFUNCTION CALIBRATOR User Guide Applies for Main Firmware version 1.90 Dear user, We have made every effort to ensure the accuracy of the contents of this manual. Should any errors be detected, we would greatly appreciate to receive suggestions to improve the quality of the contents of this manual. The above not withstanding, we can assume no responsibility for any errors in this manual or their eventual consequences. We reserve rights to make modifications to this manual without any further notice. (8822000) / CL560 / 012512

Transcript of MC5 User Manual

Page 1: MC5 User Manual

CL560MULTIFUNCTION CALIBRATOR

User GuideApplies for Main Firmware version 1.90

Dear user,

We have made every effort to ensure the accuracy of the contents of this manual.Should any errors be detected, we would greatly appreciate to receive suggestionsto improve the quality of the contents of this manual.

The above not withstanding, we can assume no responsibility for any errors in thismanual or their eventual consequences.

We reserve rights to make modifications to this manual without any further notice.

(8822000) / CL560 / 012512

Page 2: MC5 User Manual

Contents

Trademarks

HART® is a registered trademark of the HART Communication Foundation.Any use of the word “HART” hereafter inthis document implies the registered trademark.

Other trademarks are property of their respective owners.

Page 3: MC5 User Manual

Contents

Contents

Part A, General

CL560’s Modularity and Options 24Hardware modules/options ..................24

Other Connectable devices ............25Firmware options ..................................25

Safety 26Certifications and Compliances(EC Declaration of Conformity) ............26Safety Precautions and Warnings ........27

General Warnings ..........................27Warnings Concerning the useof Electrical Modules (E and ET) ....28General Warnings ConcerningPressure Measurement ..................28Warnings ConcerningHigh Pressure .................................29

Service 30Firmware Update ..................................30Recalibrating CL560 ............................30The Battery Charger .............................30Cleaning CL560 ...................................31

Cleaning the Contacts of theInternal Reference JunctionModule ............................................31

Introduction 2About This Manual .................................2

Typographical Conventions ..............3Unpacking and Inspection ......................3

CL560 Hardware 4Operational Sections and Connections ..4

The Upper Panel ..............................5The Connectors on theLeft Side of CL560 ............................6The Front Panel ................................7

Memory ................................................10Batteries ............................................... 11

About the Charger andthe Charging Procedure .................12Removing/Replacing theBattery Pack ...................................13

Support for Table Top Use ....................14The Wrist Strap andthe Neck Support Strap ........................14The Optional Carrying Case .................14

CL560 Firmware 15General Description .............................15

Startup Procedure ...........................15Basic Mode .....................................16Maintenance ...................................16Calibration Mode ............................16Help Function .................................17Printing ...........................................17

The User Interface ...............................18The Status Bar ................................18The Function Key Bar .....................19Menus .............................................19The Display Area ............................20

Page 4: MC5 User Manual

Contents

Part B, Startup and Basic Operation

Starting CL560 34Startup Procedure ................................34Basic Mode, Defined ............................35

Measuring 37Pressure Measurement ........................39

Using Internal Modules ...................39Using External Modules ..................39Zeroing a Pressure Module ............40

Current Measurement ..........................41External supply ...............................41Internal Supply ...............................41

Voltage Measurement ..........................42Measuring Low Voltages ................42Measuring Voltages up to ±50 V .....43

Resistance Measurement ....................44Switch State Sensing ...........................45Limit Switch Test ..................................46

Performing the Limit Switch Test ....46Frequency Measurement .....................48Pulse Counting .....................................49RTD Measurement (Temperature) .......50Thermocouple Measurement(Temperature) .......................................51

Internal Reference Junction ...........51External Reference Junction ..........51

Special Measurements 53Mathematical Special Measurements ..54

Minimum value ...............................54Maximum value ..............................54Min/Max value .................................54Rate of Change ..............................54Special Filtering and Resolution .....55Deviation Measurement ..................55

Special Measurements UsingTwo Ports Simultaneously ....................56

Difference Measurement ................56Redundant Measurement ...............57

Generating/Simulating 58General ................................................58

Changing the Value of theGenerated/Simulated Signal ..........59

Current Generation ..............................60Using the ET module’soutput terminals ..............................60Using the E module’soutput terminals ..............................60

Voltage Generation ..............................62Generating Voltages up to ±12 V ....62Low Voltage Generation .................62

Frequency Generation .........................64Pulse Generation .................................65RTD and Resistance Simulation ..........66Thermocouple Simulation ....................67

Internal Reference Junction ...........67External Reference Junction ..........67

Special Generations 69Opening the Step or RampConfiguration Window ..........................69Stepping ...............................................70Ramping ...............................................71

Alarm Limit Settings 73

Page 5: MC5 User Manual

Contents

Configuring the Calibrator 76Settings ................................................76Setting Time and Date ..........................78

Advanced Utilities 79Measuring the EnvironmentTemperature with the ENV Sensor .......79Display Mode Settings .........................80

Scaling ............................................81Displaying Values in Percentage ....82Displaying Error Values ..................82

Transmitter/Switch Simulation ..............83Transmitter Simulation ....................84Switch Simulation ...........................85

Data Logging ........................................86General ...........................................86Configuring .....................................86Starting the Data Log ......................87Viewing the Results ........................88Transferring the Results toa Personal Computer ......................88

Generating Signals UsingExternal Devices ..................................89

Controlling the External Devicein Basic Mode .................................90Pressure Controller Settings ...........91Temperature Controller Settings .....92

Printing .................................................93

Part C, Advanced Operation and Configurations

Additional Information 94Things to Consider whenMeasuring Pressure .............................95

General ...........................................95Pressure Type ................................95Pressure Modules and theirNaming Conventions ......................96Square Rooting ..............................96

Thermocouple Measurement/Simulation,Connections and Troubleshooting .......97

Internal Reference Junction ...........97External Reference Junction ..........98Error situations .............................100

Resistance and RTD Measurement,Connections .......................................101

4-wire System ...............................1013-wire System ...............................101Using a Compensation Loop ........1022-wire System ...............................102

Current Measurement Parallel toa Test Diode, Connections .................103Parallel Functions in CL560 ...............104

Page 6: MC5 User Manual

Contents

Part D, Calibration

General 106Phases of Instrument Calibration .......107

As Found Calibration ....................108Adjustment ...................................108As Left Calibration ........................109

Required Modules for DifferentInput/Output Signal Combinations ..... 110

Calibrating an Instrument 112Selecting the Instrumentto Be Calibrated ................................. 112The Instrument Window ..................... 113A Calibration Procedure Using CL560 114

About Automatic Calibration ......... 116About Manual Calibration ............. 116Autocapture Feature ..................... 117

Examples of Instrument Calibration ... 117Pressure Transmittersand Sensors ................................. 118Temperature Sensors ...................120Temperature Indicatorsand Recorders ..............................122Electrical Limit Switches ...............124

Using External Devices inCalibration Mode ................................126CL560’s Support for InstrumentAdjustment .........................................127

Maintaining CL560’s InstrumentDatabase 128

Adding New Instruments ....................128Editing Instrument Data ......................129

General Data Page .......................129Instrument Input Page ..................130Instrument Output Page ...............130Calibration Settings Page .............131Calibration Instructions Page .......132

Deleting Instruments ..........................132

Viewing Calibration Results 133Calibration Result Windows ...............134

How to Choose Which CalibrationRun is Viewed ...............................134

Deleting Calibration Results ...............134

Appendix 1,User Guide forMC5 CL560 Option 136

Appendix 2,Technical Data 150

Appendix 3,Quick Guide for theCL560 Datalog Viewer 156

Appendix 4,Index 160

Page 7: MC5 User Manual

�������

������������������� �����

� ������������������� ���������� ����� �� ����� ������� �������������������� ������

� ������� ������������������������ ��� ���

� ������� �������������������������� � ���

� ���� ��� ���!� ����������� ��������

� " ���!���� ������ ���� �������� #�����!� $��� ���� ��� ���%��������

Page 8: MC5 User Manual

�������

�����������

��������� ������������ ��������� �������� ���� �����

������ ����������� ����������������� ����� ����� ������ ������ ����� ����!�� � �"���!�������#����!�������#������� �����������$����"� ����%������&�������� �"���� ������� '����������� �%�������$� �������� �������� ������������#��� ������ ��� �"���"����������"���� ����� � ����������#��%%���� �����!�����������(��������� ���������������������������!��) ���"���"� ��������!�������������%

*���+��������� ��������� ��������������� ��)��"����"������%�*����������+� ����!� ����� �!��"�� ������������� ��� ������������������� ��� �!��"������������������� ����"�������!� ����"%

������!������������� �� ��� ���� ���� !�������#������� ���� ������!��������!������ ����������%� ��� ����!������������ ��� � ����,���������) ������������!�����������������%����������������� ������ ���-�.*�� ���� ����������%

��������!���������������������������� ���������"#����!��� � �"����������������� � �"%

����������� ���*� ��/������ ��� ��� ) ���� ������!����0��#�1#�������2%

� 3������� ����������������������%�*����� �����������!�������������"%

� 3����1������ ����������� ���������������������������� ����������� ��� ����%

� 3����������������� ���� ����)�������#�����!� ������ ������ � � ������������������������ � ���� ������ ������� ��!�������������������#�.*2�����*(��������������(� ����� �%

� 3����2������������������ ���� �������������������������� ���� ��� +������� �� ��������������%

*����)���!����������� �!��"������� ������������� )��!���%�*�����!����������� �!��"�������� �����4����5-��� ����)���67%*��������������������!������� �� ������������� )��!��������� � ��������4������! ������5 ���3����1���� )�7%�/������� ������ ��!��) ���� �������������������$� �+�� ��� ���������� �������!��� ����������4���%

Page 9: MC5 User Manual

8

������������ ���������

�����,��!����������� ��������� ��,�������!� ������� ��������������#��%%

9��������!��0�� �����������

���������!�������,���5� ,�����,��������&���)��7�����!� ������� ������������ #��%%

������������+�������������

:���� �����������+�"������������������������ ����������+�"����� ���� ��� ������ � ���� ���������!�� �� ��,�� 5����� �7�� ��!��"���������������� ���������������#��%%

:���� ��+�"���(����

� ���� ��� ��� ����� ������������"�������� �����!���������������� ��!��� �%������������ ������� ����!������� �������������� �� !�!��� !���� �������!������ !�%�*������� )��������#�� �)��#� ��!���������� �������"�������������"���)������������� ������� �%��������������� �����) ��������� ���������#�!��+��������������� ���!����#������ �����������������!����������� �����!�� � ��� ��#�������������!������ ���������� ������������!�� ���%�*������������������ ������������� �0�

� ��� ���� ������ � ����� *� ��/������ ��� ;������"������ 1�����"�3��+#�< �-� ��������������1�����"�3��+� ��!���������� ��� �������� �����"� ���������������� ����������!������0

��!���������������� �������&� ����� ��!������0�* ������������������ !�� �������&�� ����� ��!������0�:������ � ������������������

� ��� !�

:��������� !� �����)� ���������� ����������� ����!� ��#������������'��� ���!� ���(���������!����=%

���"����)��������������� ���������������������"������"������#��������� ����!��+ �� ����)���!�� ���%��������������� ���������� !� ���������������������������%

) �����*

���� �����!����!���� ������������������������� ��$� ������� ����������� +� ������������,��$ �� ��,-.��/,0������ �1�.23������!������������������� ��$��� 4 �����

�������� �

Page 10: MC5 User Manual

=

�������

�� ����������

����������������0

� �3��� ����(�����!��������5><����?71�����"�!��+��38�%

� ���������� �!����!�������� 1����� � �������!����������+���!!�������!� ����!!�������� ��������� ������������������ �!���� �����!�������0��6��@�A���B��56=C@C6��CB:7%

A6��@�A=��B��5��C@C6�=CB:7� �������� ������������ ��%� 9��������!�������0�����@�A���B��5�=C@C6=�CB:7%

<��0�*����� �+������������������ �����"������������� ������ �������!�� ��� ���,��������� � ��%

� -�� � �"0���@�D��E.-

������!������ )���!�� � ��� ��������)� ������ ���!!��� ,��%

������� �������� �� ���� ����

� � � � � � � �� � � � �� � � �

� � � � � � � �� � � � �� � � �

� � � � � � � � � � �� � � � � � � � � �

� � � � � � � � � � � � � � �

� � � � � � � �� � � � � � � �� � � � � � � � �

� � � � � � � �� � � � � � �

� � � � � � � � �

� � � � � � � � � � � � � �� � � � � � � � � � � � � � � �� � � � � � � � � �

� � � � � � � � � � � � � � � � � � � � � �� � � � � � � �

� � � � � � � � � �� � � �� � � � � � � � �

� ! � �" � � � �� � � � �

� ! � �# � � � �� � �

� ! � �$ � � � �� � � � �

���� ���� ������������� �������!��������� ������ �� �� ������,�!���%�

<��%

F��!� ��� ��� ����� ����!��) ���! ������ 5��� �����������! �������������� ���� ��������7���������,��!������ ���� ����������%*������ ���� ���� "�����������"� )��"� � � � �����"� ���� ������ ������! �����%

Page 11: MC5 User Manual

���������������

*����!!���!�����������!������ ��� ���� ��� ��������(���������0

&+���� ����������'�����

�����������������������>,�������3���������������5>G*�7%�*���������� ���������������� ��������� ����������!!���!�������� �����+��� ����5-� ������ �+����������!!���!����%�

6����� ����������'�����

/!������������������3�����������������"���� ��������� �������%�������������"������� ��������������� �������%�����������1������ �������������)��������!������������������� ��������!��������������� ������������������������ �������!!���!����#�������!���������)��������������������"�����������#��%%����������������>,�������3�������������%�

*������������������������3���������������������������������������������� ���� ����%�*���!�� ����1������ �������� ����� �"����������������������� ������� ���������������������� �!���������������������� �� ���������+�!����������%�<�����"��� ������� �� ��������� �� ����������� �� !�������������&��������%

���������!����������������������+��� ���36�@�38%�*���1������� �����������������)����������) �� �%�*��� ��������� � � ����������������%

*��������������!����������� ���������� ��������!���������������� ��������� �%�������������� )��� $� �����"�!� ����"�������� ��������� ����������� �����������C���(8��C!� ������%��) ��! �� �� � $� ���������� ���������� �(� ������� ��!�������������(����!��������������%

*��) ������ ��������� �����#����������� ���� ����"� ��������� ������!�������������������������5��,%���$�����<�#�!!�,%�8%�����C��7%������������������ ����$� ������������������������ ��5�"! ����"�!��������������� �����!�������������� ��������������7#��!!�"���������������� ������!�����������������������"&����,�����!���%

*���)��!��������!����� �������� ��������!���������������)�������������+���������� �����%�.����������������� ��� ���� �+� �� ���!������������!��������������%�9����������!�����" ���!��!����������" ���!���� ������ ���) ���������!����H%

������-��� ���

Page 12: MC5 User Manual

�������

���� ���������������������������� ����

*��������� �����������5�����) � 7�������������������������� �0

&�%���� ������ ��� ����6����� �

*� ���������� �� ������!� ���� ���!��������������������������%�*� �������� ����"������������) ����������!��������������������%������������������������������������������ ������� ����� �����!�������� ����������%�

�� ����7��������6����� �

*������3(3.*�����������"��������� ���������� ���������� ���!��� ����3�%�*���3����"���)������� ���� ����� ������!����������� ��� �� ����������#��%%������ ���������������������� ���������� �������������������������������3�%�

*���������������������������� ���������� ���������������!� ����!�������!� ����%�*���!� �������"������������!� ������������ �����������������5�%%���� ���� ����������7%

) �����*

�������!� $�������%�����$!�#� �+����������������������� ������� ���������

�+��� �!�6���� ����6����� �

*����/G��������� ������� ���������� ��!�����������������#���!�������������(��"����+�����%��������%�

/� ����, � ��"� ����������� ������� ���� ��3���������� ��������%

Page 13: MC5 User Manual

H

���������������

*��������!�����������)��������� ��%�9��������������! ������� ������������ ������! ���������!���� ����9��� �������������� ��#�������������������5�%%�� �!��"�����+�"����7%�*������ ����������������� ��������� ���� ��������� ��!�����!��%

&����� ��'�����/&� ����3

*���>�������������������������� ��$���� � ��0�%��� ��#��8������������9��!%�������������������� �������� ��������������� ���������������������%���� � ����"������� ����!�� � � �"������� ���������������!!�"���� ���������� ������=CI2��! ����!!�"%�

*���>����������� ������������!� ����-�.*�����%�*� ����� ������ ��� �� ������ ���������� ���-�.*���!�� � � ��%�;���-�.*� ������� ����4���� �� ������&��! �����!!�"#���� ��������H�� ��� ��� �������������-�.*������� ��� �� �������� �����"� �������%

��� � ���� ������ ������ ���� ��-�.*� ������������ �� ���!!���� ,�6%

&����� �� ����� ��� ����'�����/&�� ����3

*���>*������� ���!�� ���"���� ���� ��� ���!�������� ������������ ���� �������%���� ������ �)�������� ���������"����!������� ���������������������� ������������������%��� ��#������#����89��!����������%�

������ ����!�� � � ��0

� � �I���������������������*(���������������� ��� ���������� ������������������4���� ��������� �I��������������%

� .�� �����������.*2������������%

������� �(� ����� �� ��!�� � � ��0

� *(��� ����� ���� ��� ��������� ������������������4���� ������� � � I������������%

� .�� ������� ���.*2�� ����� �%

� I����#� �������#���$����"� ���!����������� �

� � � � � � � � � � � � � � �

� � � � � �

� � � � � � � � � �

� � � �� � � � � � � � � � �

� � � �

� � � � �

� � � � � � � � � � � � �

� � � � � � ! " �# � � ! $ � � � %

� � � � & � ' (

� � � � �

� � � � � � �� � � � �� � � � � � � � � � � � � � � � � � � � � � � � � �

! � !

) � & � � � * + ' ,

- . � � � � � . � � �

� � � � � � � � � � � � � � �

� � � � � �

� � � � � � � � � �

� � � �� � � � � � � � � � �

� � � �

� � � � �

� � � � � � � � � � � � �

� � � � � � ! " �# � � ! $ � � � %

� � � � & � ' (

� � � � �

� � � � � � �� � � � �� � � � � � � � � � � � � � � � � � � � � � � � � �

! � !

) � & � � � * + ' ,

- . � � � � � . � � �

������-��� ���

Page 14: MC5 User Manual

D

�������

:��������;������'����

*���*(�������������(� ����� �� ������������������4���� �� ����!� ������� � ��������>*������%���� ���!�� ���"���� �������������������������������� �����������������4���� ����!����� �� ������ ���� �� �������!������ �������������������� ���� ��������!��%�

*���.��������� J����� �������� ���� ���������� ���������*(��!������� ��� !!��� ���%�!��� ���� � , �� ���� �� ���� ����� � ��� �������������������� ������ ���(!���������.���������J���� �������%���+������������������� ���(!����������!��� �"� �� �� ������������.��������J���� �������%�.����������� ��������������� ������ � ������� ���(!���������������%�-����� ���� �� �����$����%�2���!����������� ���(!��� ������ ����!�� ������� , ������ %������� ���"��� ������������������������������������������%�

*(������������������� ����� ����"������������ ����� ��� �������� ���������� 4���� ���"��� �� ����� �I����� ���� ����� �����>*������%�*������������������4���� ����������������)������������� ����������������������������������4���� ������ ����� ���������������������4���� �����!����������)������� �������������%

3����1��� �� ��������� ����� ���� ������ �� ���� + ����� ���������4���� ������ ��������)� �����%

<���� !

���������������+� �� ������������ )��� �!��"%�*��� ������ ���� ���� �!��"� ���=��,�8���! ,���%

*�$� �+�"������������������������� �!��"0

� 3������������������ ��������� �� �%� /��������!� ������ �� ���� �+�"���������������������%

*�$� �+�"�������������+� ����� ������0

� 3������������������ ��������� �� �%� /������������ ������ ��� ���� �+�"�����������������+� ��

�� ������%

*�������������� ������������� ����"���)������������������ ��%

� � � � � � � � � � � � � � �

� � � � � �

� � � � � � � � � �

� � � �� � � � � � � � � � �

� � � �

� � � � �

� � � � � � � � � � � � �

� � � � � � ! " �# � � ! $ � � � %

� � � � & � ' (

� � � � �

� � � � � � �� � � � �� � � � � � � � � � � � � � � � � � � � � � � � � �

! � !

) � & � � � * + ' ,

- . � � � � � . � � �

Page 15: MC5 User Manual

?

=�!$� ��

*���+�"�������������&��+�"�����������!�������� ������� ������ �������� �0

����������=�!�� �������&�����=�!

*���������+�"����������>�����+�"������������������������!!�������������������� �!��"%��*���������+�"����������� �����) �����������������������%�*��"�������)����)������!�� �������� �� ������� ��� ���� ��#��%%� ������� ���������������� ����� �!��"%*���>�����+�"�� � �������������� ����)�����%

����=�!���$�%������<���� !

*���� ���+�"�������������+�� ���������� �!��"���������%����� ���������� �������� ������������������������ �������������� ��!��"�5�������!����<���� !���!���D7����� �����!� �� ������������5�������!���������������!���6H7%

*���-��!�+�"�� �!��"����������� � )�����!%

*�����(����+�"�� ��������������������%�3�����������(����+�"������������������������ �����(��%�*� ������"�������� ��!���)�������� ��������(���� ��� ���������%

������-��� ���

Page 16: MC5 User Manual

6�

�������

3���� ������A(��+�"����������� ��������������������� ��)����%<��%�*���A(��+�"� ���!!� ��������"� ������� ��� ����%

*���2�� ����+�"������������� ����! �������������� ��)��������� ����������"��� ���%

����2������=�!�

*���:���� ��F�"���������������� ������ �!��"%�*������� ��������:���� ��F�"�)�� �����!��� ���� ����� ���� �%�*��� � ��!����������� �!��"� �� ������ ��������:���� ��F�"����������������������%

����> ����=�!�

*���<���� ��+�"�����������"������ ��������� ���������0

� F�"��6���H��������������������������+�"�%� F�"��������D��������������������������)�����!����������

!� ��%�*��"���"������������� ������ � ��������!�� ��� ����!!��!�� ��%

� F�"�?������������� ��������!� ���������� ���� ����� � ��� �������������"%�*�������� ��� �"�������?�+�"� ��������� � ��������� �+�"#��,��!��������� ���� �0�;��������� ����������#� ����?�+�"�!������������������?%�*�� � �������� ���������#� "�� ��� ��)�� ������ ���� � +�"��� ���� �����(��:���� ��F�"� �����)� �����%

�������������������"��� ������"������� ���"����%�*� ������������������ ��������!�� � �������������"������)�����#��%%� �������������%��������������"���"��������������"�� ���������$� �����������"%�*���������� �����,����� � ������������������ ��������������������������� ��� �� �� �������"%����������!�������� ���������"� ������������"����������%

Page 17: MC5 User Manual

66

���������������!!������������������������������������ ���������+��� ��������� ��%�;������ ����+�� ��������� ��#�"���������2�"�1������"������ ��%������������� ����"�������������������"��"!�%

*�����+�� ��������� ����������������0��������I����0 6%��I�������0 ���*"!�0 ��

*������������������������������� ���!������� ��������� ���) �������0

��I����0 6��C@C�=��I��#��:��$����"0 ��(���-'

*������� ��������� ��� �� ������1�����"�3��+%�*������������������� �����"����������������������1�����"�3��+� ��� ������������������&��1����/� �%����"����)��� ��������������������������� ���"����"������������ ����������1�����"�3��+� � ��������������� ������������� ����������������1�����"�3��+%

*�����, ����!���� ��� ��� ����������� ��)�� �����!��� ������������������ ����������� ���� ����� �!��"� � ��%�������������������!�������������������������������=I������� �������!�!�"��������������, ����!���� ��� ��%�>)��� �������������, ����� ���#� ������������� ������������������ ��� ������ ����� ��� �����%������)�����!���� ��� ��� ��6������%

�����+�� ��������� ������� �����#� ������, ����!���� ��� ��������!�����������$��� �"������������� ��%�����)�����!���� ��� �� ���!!�, �����"�=�����%

*����!!���������������������&��� �!��"��� ����! �����������������"%�*��� � ��������! ������ �#� �������������� ��������������������� ��5������ ����������+�� ��������� ��7%

<���%

�����&������"� ���� ���� �������� ���+(��������� ����� �� �������������! ����������������� ������ ��� ��������%�.�������������+� ������!�� �"��� ���������� ��� ����� ��� �� � ��������������� ����� �����%

2�������)�������� �������1�����"�3��+�����2�"�1�����"������� ������������ ��%���������"����� ������� ��� �� �� ������� ��������!!���)������������,�������!�� �%

�������������

�������������

������-��� ���

Page 18: MC5 User Manual

6�

�������

���������� ��������������� ���������������

*���������� ��������������������������������������������������%�*������� ��������� ��� ������"��������!��������������� ��!�������� ����������!��������������������� ��%�

;���������� ������������#��������� ��������� ���� ��������+�������������)��������������� ��%������ ������#���� ��� ��) � ���� ������������������� ��%

;���������������������� ��� �����#��������� �� ��� ���������� �5�� �+ ������� ��7��� ��!������5������������� ��7%���������"����������� �� ���� ������ �� !����%�>�!�"� ������ ��� ���� ����"������� ���!!�,%��K�����%

;���������������������� ��� �������#����������� �������������%����� ��������������� ��������� ���!�) �������!!���)���������!��)��������������� �������� ����� �

) ������*

�"&�(>�?��@&��@�:�&:��:(A6<&<�)6�@��@&����6#:��(:�

����� ����� ����������%��� ������� �-������,B�CA���

����� ��������������!�$�������������� ��������� ��� ��������������+����B��.��/-�B�.23�

� ! � � % � � � � � � � � � � � �

� ! � � % � � � � � � � � % ! �

Page 19: MC5 User Manual

68

�!����" �����������#���������$

*����)������!���������1�����"�3��+#�!������������� ��!��������0

-� ,�

6% *�����������!� ���� ��5����� �!��"���� �������������!7����� ���������!!��%

�% 3���� ���� ��)��� ����� ��� ����������� ������!!��%�*���1�����"3��+�!!�������� ��"����!���� ����%

*���!���������1�����"�3��+#�� �!�"��� ��� ���� ���!����%�;����"���������� �+#�����1�����"�3��+� ���������� �� ���!����%

<��%

�����������1����/� �� ���3���!�������#�����1�����"�3��+� ����%*���1�����"�3��+���������� ������� �� )�����������������������%��) ���,!� ������1�����"�3��+���� $� ��%

������-��� ���

Page 20: MC5 User Manual

6=

�������

������������������������*�����!!��� )���"������) � ������� ���������� ��!��������� ������ �!%� �� ��� ������!!������ �������+�������������!�����������������������!������ �� ������! �����%

��������������� ������ ����������������������������� � �������!�������������������� ���������� ������� ���������%�*������+��!!�������!����!��"����� �� � ������� ����� �0�3� � �������� ���������� �������� ����� �� ���� � �!��"� ���� �+ �%���������� )��"0�-������������#��%%���)��)������������������� �!��"� ���������������)������"���"��%�*����"������������� ����� ��� �+ � ������������� ������%

�������� �������� ����������&����������" ������� ��!���� ���� �����) ������������ ���������%�*��� ����" �� ����� ���������������� ��� �����!�� ��� � � ��#�� +�0

� *��������#������������������ !�� >,�������!��������������� ��!��������!��!� *��!��������������� ����������� ��������� /������ ���5�� ���+7

*�������" �� ����� �� �� ���� ��� ���� �������� ������ ��� ��) �������%

Page 21: MC5 User Manual

6�

�� ���������

�����&��� �� ���� ����)��� ��:��9-�����"%�*�������� �� �������� )��"����"� ���!����� ���� � �� ���� ����)������� �)��� �� ����������!�� � � ��� ����������%�9���2�� � ������ �����!���8�������� ������ �����!��� ������� �� ���� ��"��������%�

!� �����"������ *��� ��� ��! ������ �����"� ����� ���� ���� ����� ����� ���� � ��� ���%������� ������� ����������+��� ���������+������%�>������ ������ ��������)����� ���+�� � �������� �!��"��������������,�� �����������+������%

*������ �����!������� ���"������ ���������� ������� �%

����������������

>)��"�� ��������� ��������������9�����!�3������������+������������ ��� �"���������) ����"�!����� �������������%

����������������� ��!�����������������"#�������� ����� ����������� �� �!��"��%

������ ����������������� ����"� !������� ��1�� �����%��������!������ )������� !� ���� ����9�����!�3�������� �� ��������� �� ����3����1����� ��������%

� � � � � " � �� � & � � ' " � �

� � � ( � � )( � ( � � �

� * � � � % � � ! � � � � � � + � � � � �

� � � � � � � % � � � � � � � � � � %

� � � � � % � � � % � � ! � � � � � � + � � � � � �

� � � � � � % � � � � � � � � � ' � � �

� � � � � � � � � ' � � � + � � � � � � � � � � � � � �

� � � � � � � � � � * � � � � �

, � � - � � % � � ! � � � � � �

� � � � � � � � � � � � � + � � � � � �

� � � � � � � � � � � � � � � � � � � �

' � � � � # � % % � � %

� � � � � � � � � � . � � � � � � � � � � � � � � � � � � � �

/ � � � � �� & ' �

� � # � / � � � � & (� & ' �

� � � � � " � �� � & � � ' " � �

� � � ( � � )( � ( � � �

� � ( " � � � � " � � " � � �

������: �� ���

Page 22: MC5 User Manual

6�

�������

#���%���

���1�� ������"�������������������������(� ������� � ����%*��������� � � ��!������"� ��� ������� �� ���)� �����%�1�� ����� ������������������� �������� �������������� ����������������� ���� ��!������������� ���������%

9��!! ������.��! ������������������� �(� ����� ��� ���������)��"� ���� ��%

������ �� ����� �����1�� ��������������� ���� ��!����1��� �� �������%

3������ ��������������1�� ��9����&�� � ���� ��)��� ����� ��������� � ���� ������ �%

%��������

*� ���� ������� ������������� ��������� ���� ������ ��%

��� � ����"������� ������!�� � � �"�������� ������������5��$� �����!��� ��7%

�� �����������)������4����������������� ��3���������� ��������%

���������%���

�����&���� �����"� ����� ���� �� ����������%�*��������)��"��!��� ��������� �� ���� ����������� ��������� ��������� ��������� ������%���������"���������������������������� ������ %�%����� ����������� ��������� ��� ���� ��� ���"� ����� �� ��)��� �������&�� �����"%��!� ����"���������������� ������ ������ ���� ������ ���%

��������!!�������������� ������� ����,��%�*��"����!���������� �� �����!������������� ���� ������������"����!�� ���%�L����"��������������+ ������ ������� ����,��0�"� ����������#���D��������������2�������������%���� � ����"#���� ���� ���������������������������������� ���� ��!�������%

������� ���� �� �������� ������ �� ���)� ������ ��3����2��� �� �������%

� ��$� �����:�����

*�����!� ������!�������� ����� ������������ ���������������� ����� �� ����������"����) � ��� �������� 5����!� ����� �� ����!� ���!� ����� ���)� �����7%�*�������� �� ���� �������� ������� ���� ������ ������+��� ��!�� ������) � ������������� ��3����) ������%

���� ������ �������� ����� ���� ���������� ��!��������� ��3���2����� ��������%

Page 23: MC5 User Manual

6H

&����������

*���-��!������ �� ���������� �����! �������� ���������������������������� �� ���������!������M������N�������� � �"��)� ������ ����������"� ���� �%� � ��� "�� ����� ���!#� 4���!����� ���� � +�"%��� �� � �������������!�!!���!%

������

L������!� ���������������������"�� ���� ��� ���������"�!����� ������ �+�"��������� �+�"�� ����������"%

��� � ���� ������ ����!� �� �� ������� ��!���������� ��������%

) �����*

(��!������������������ ����� % �� $��� �� ���������������� �!��������������� !�� ����������������������������%���$��������� �

0 0 1 2 3 1 0 2 2 2 � � 4 0 5 4 6 �

� � � � � , � � � � % �� � � � � � � � � � � � � � � � � � � � � � � �

/ � � � � � � � � � � � ! � � � � � � � � � � � � � � � � �� � 6 7 � � � � � � � � � � � 1 �� � � � � � � � � � � � � � � � � � � � � � ' � � � � �� � � � � & � ! � � 1

� � � � � � � � 8 � % � � � � � � � � � � � � � � � � � 8 � � � � � � � 9 � � � � � � � � � � � � � � � � � � � � �: � � � - � 4 � � � � � � � �: � � � - � 0 � � � � � � � � � 1

" � � � . � � 5� � � � � / � � � � � � � %� � � � � / � . � � � � � � � � % � � � � � � � � � %

; � # �

� � � �

������: �� ���

Page 24: MC5 User Manual

6D

�������

��������� �������*����� �����������������/���� �� ������������� ��������� �! �����0�

����!�� �������������������� �������� ������!��) ���! �����#�������� �!��������������� �������� ��������� �����!����%

�����������#��

*���9������1�����������!������� �!��"� ��) � ������������ ��%���� ��� ) ���� �������� ������ ��%

*���� ����5�������7����� ��� �!��"��������������)���������������"%*���������"��"���� ����!�������"���!����"����5 7� ��"������� �� ����������"� ������� ������� ��� ����!� ����2�"�1�����"����� ��%������ �����������������"������������������������"��� � �����%�*������������� ��� �!��"������� �����������%�*����� �������� ��� �!��"���������!����������������� �������!� ������) �������������#� ������������ ������������������%

*������������� ��5� �����7����� ��� �!��"����� � ���� ������� �� �������������"����#�� +�0

� ����������� ���������� �� �+ ��������� ���������+��� ��%

� ��$���� �����+� �������������������%� ���"���� �� ��� ������� ��� �� �������,���������) ��#

�%%���-�.*� �����������������������5� ��� �7%

<���������������������� �� ����!�"��������������� ��%�*����"����������) � ������"� ����������%

� � � � � � � �

� � �

� � � � � � � � � � � �

� � � � � � � � � � � � � � � � � �� � � � � � � � � � � � �

� � � ! " �

� � � �: � � � - � 4 : � � � - � 0� � ( "� � �

& � ! � � � � � � � � � �

2 1 < 6 0 4 � + � �� % � % � �

� � � � � � =

# � � � � $ % � � " � & � ' � ( ( ( ( ( ( � � ) � �

# * " � + � ) � �

� � � � � � � � , � � � � � � � � � � - � � � � � � � ) � �

2 1 6

2 1 4 4 0

0 � � � � � � � �

+ � �

� � � � �

� � � " � � � � � � � � � � � � � � � �

9 � � � � � �

> � � � � � ?

> � 4 5 � � ( � 0 � ?

> � � % 1 � " � � � ?

" � � �

> + � � ?

$ � � � � � � � �

' � � � � � � � � �

@ � �� �+ � �� �� � ; 0 & �� � ; %

( - ( . � ( ( � � � � � . ( �

4 �

; � � �

� � � �> % � % � ?

� � � � �� � � �

� � � � �

A � � �

$ � � � � � � � � � � � � �

� � � � � � � , � � � � � � � � ( ( ( ( (

� � � � � � � � / � � 0 � � � � � � �

� � � � � � � � � � � � � � � %

' � � � � � � � � � � � � � � � �� � � � � : � � � - � 4 � � � � 0 �

� � �

� � � ) � � � � �

� � � � � � � � � � � � � � �

� � � � � � � � �� � � � � � � � � � � �� � � � � � % �

� � � � � / � � �

$ � � � � � � � @ � � � + � �

/ � � � � � � 7 � � ! � � % � � � � � � � � � � � � � � � � �� � � � � � � � � � �

� � � � � � � � �� � � � � � � � � � � �� � � � � � % �

� � � � � � � � � � � � � � � - 5� � � � � � � � � � � 1

� � � � � � � � � � , � � �

� � � � � � � � � � � � � � � � � �

� � � � � � $ � �

Page 25: MC5 User Manual

6?

�����������'���#��

*���:���� ��F�"�1���������������������� �!��"� ��) � ������������ ��%�*������� ��������:���� ��F�"��)�� �����!��� �������� ���� �%�����"���:���� ��+�"���,���������������������� �� ��� ��������������������%

� + � �� % � % � �

� � � � � � =

2 1 6 + � �

� � � � � � � � � �

� � � � � � %

� � � �� � ( "� � �

& � ! � � : � � � - � 4 : � � � - � 0� � � � � � � �

� � � � � � � � � � �

> ( � � � ?

� + � �� % � % � �

� � � � � � =

2 1 6 + � �

� � � � � � �� � � � � � � � �> ( � � � ?

> ) ) � ) ) � ?

$ � � � � � � � � � �

� � � � � � � � �

� � � � �

� � � � �

� � � � � � � � � �

� � � � � � � � �

� � � �� � ( "� � �

& � ! � � : � � � - � 4 : � � � - � 0� � � � � � � �

> � 0 5 � � B � 4 2 2 ?

� � � � � � � � � �

� � � �� � ( "� � �

& � ! � �

2 1 C 2 D E < � + � �� % � % � �

� � � � � � =

# � � � � $ % � � " � & � ' � ( ( ( ( ( ( � � ) � �

2 1 6

2 1 < E C <

0 � � � � � � � �

+ � �

� � � � �

� � � " � � � � � � � � � � � � � � � �

9 � � � � � �

> � � � � � ?

> � 4 5 � � ( � 0 � ?

> � � % 1 � " � � � ?

" � � �

> + � � ?

$ � � � � � � � �

' � � � � � � � � �

( - ( . � ( ( � � � � � . 2 �

4 �

; � � �

� � � � �� � � �> % � % � ?A � � � �� � � � �� � � �

: � � � - � 4 : � � � - � 0� � � � � � � �

� � � �� � � � + � � � � � �$ � � � � � ( "� � �� � �

� � � �� � � � � � %

%����

*��� :���� �� F�"� ��!�� �� ��������� �� ��� �"���(����%�*�������+�"� ������� ������� ���������%����������� ����������� ��� ���� ����������4���#� ���� ������:����� ��F�"� �������������������%�

��� ����!���������������)����� !���#� �������&���!!���������(�� ������ ���� ���������� �!(� �� �� ����%� �����������#��������� ����� �+�"������ ��������������)� ����������!���%���������!� �� ����������� ������������ ��+�"���� %�9����� ���������!� ���������� �������������� ���)����0

6% ��� ���� ������� ����� ��������������������������� �����"#��%%� ���������� ��������������������������!� � ������! ��������)�%

�% ��!!��!�� ���!������������� ������������)� ������!� ��%*����������������� �� ��� �!��"��� �� �������+���� ����������%/������� ����� �+�"�������� ����� �+�"���������� ���!!��!�� ��%�*�����������!� �� ������!!��!�� ��#������ ��������

�+�"������� �+�"%�*����������!!��!������ ������������ ����"�� �#�!��������� �+�"��������(�����:���� ��F�"%

8% ������������ ����� �!� �����!����������!��) �������%9��� ��������:���� ��F�"����������!��������������%�������!��) ���! �����#�����; �� �6�����!������ ��!����%��� �� �������:���� ��F�"��( ����!�"�#���������:���� �F�"��(��$�������������������!�� ������������%

=% ���� � �� �!������#��%%�) � ����� � ���� ������ ��������� �� ������������������+%

������: �� ���

Page 26: MC5 User Manual

��

�������

����(����������

*�����"���������� �!��"������)�� �������� ���������������������� )�����+�(���� ��%�*������ �! ������� )�����)��) � ����"! ����������������� ��� ��������� �!��"�������"���%

# ���'� ��� ���7����� �����

*���� �!��"������ ��� ) ���� ���� �� �� ��� ������ )����,����������� �������������(������ �)�����%

�� ������ ������� �� �� ����� �)����� �� ������ ����� ���� � ���� ��� �����%���� �#��%%��������� ��� ����������� �������� ��)�����%

�����)������� ������� ���������) � ���#������������ )��� ���� �������������� +�"���� �����(%���� :���� �F�"%

� ��$� �����

*��� � �!��"� ����� �� � ) ���� �������� �� ����� ������� ���� �%

*���� ���� �� �� �!��"���������������������� ����������� �!���� ����%� *��� ������ �����!�� ����������������������!���� ���%

*����� ��� �� �� �!��"��������������!�%�*����������!�� ����������������� ���� �������������%

������� �����)������

*��������� !����"� �� ��� ���� � �� �� �������%�*���! ���������� ��� ��������� ���� �� �� ���.��! ������ ��%

*���������� ������������ ����� �� �� �� ����������"������)����� ����� �!��"����������������� ���� ��� ����%

/��� ����������+�"�� ���)������ ����� ����%

0 0 1 2 3 1 0 2 2 2 � � D 5 2 F

� � � � � $ � � G � � � �

� � � � � � � � � � �

� � � � 3 � � 4 � � � 1 � 0 �

� � � " � � � � � � � � � � � � � � � �

5 � � � � � � 6 � � 7 � � 8

� � 1 � G � � �@ ; H

. ( (

4 1 2 2 2 2 2

4 4 1 3 D E C

� � �� � � � + � � � � � � � � ( "

� �

$ � � �

0 0 1 2 3 1 0 2 2 2 � � D 5 4 < �

$ � � �

� � � � �2 1 < 2

* � � � � �

� � 4 2 2 � � E D 6I � � � � � � 3 2 �

4 4 1 3 3 0 6

) 2 1 2 6 � J � � � � � �

6 2 1 2 2

� � �� � � � � �$ � � � �

K

2

)2 J 6 2 J 4 2 2 J

� � ' � � � � � � � � � � � � > � � 5 � � � � � 1 ?

� � � � � � � > � 5 � � � � 1 ?

� �

0 0 1 2 3 1 0 2 2 2 � � D 5 E C �

9 5 � # $ % 0

� � � � � � � � � � �

: � � � � � 1

# � � �

; � � � � � � ( <

9 � � � � � � � �

; � � � � � � � ( ( <

3 � � � � � � � �

9 � � � � � �

( � = � � � � � � � � �

9 � / � ( � <

� ( ( � <

� � � � � �� 5 � � � % � � �

0400

4

D 1 2 2 2 24 F 1 2 2 2 2

� 5

Page 27: MC5 User Manual

�6

� $����

*��������������#��%%� ����) � � �� ��� ���� �� �������� �� ����� ������%�*������ �����)�� ���� ���� �!��"�����%�*���������������������������������� �!��"%�������������������������������� ������������������� ������%�*��"� �� ����� �������� ������ ����"����������"�� ��������� �+�"�%

*��� ����� � +�"�� ������ ���� � ������ ��������� ��%�*��� �����

+�"�������� ���� � ��� ���!�������� ���5 ���!!� �����7%

- ��%

�������������������������������������������#�������������� ��+�"��� $� �+�"� 4��!� �� �����!�� ������%

@�����������

*������!� �� � �����!�� ��� ��� %� ��� � �!��"�����!� ��,�� ����� ����������������� ������ �+�"%

<���� !���� �&�� ������� �� ������������&�������< �

*��������� ����� �������� � ����(����������������������������� � ������� �� ����� �!��"�����%�/�������(%����:���� ��F�"����)����� ������ ������� ����� ��1�� �����%������ ���� �� �� �#���������������+�"�%

> ����2�����

*���������� � �"������������ � �������� ��� ���0

6% 3������������ ��+�"#� ���+�"%�*����������������)��������!�������������)����%

�% 3��������� �+�"��������( ����:���� ��F�"��)� ������ �������� ���� �� �� �%�*����"�������� ���������)����%�<� � ����!!������������������������)����%

����!�������� �)������"�!���� ������ �+�"%�*�� ������������ ���)����#����������(������:���� ��F�"%�9�������3����1�����!�� ����������������� ������� ��� ����� ���������� ����� ���%

0 0 1 2 3 1 0 2 2 2 � � D 5 < C �

� � � � / � � @� � % �( � � � � � ( "

� � ( � � ( ( ( � � ( � � > � ' � 5 � � 3 � � 6 � ' � # � � � � 6

� � � �

> , ?

) 2 1 2 2 2 2 0

2 1 3 3 3 3 F

4 1 3 3 3 3 D

E 1 2 2 2 2

< 1 2 2 2 2

6 1 2 2 2 2

F 1 2 2 2 2

C 1 2 2 2 2

D 1 2 2 2 2

3 1 2 2 2 2

3 1 3 3 3 3

3 1 2 2 2 2

D 1 2 2 2 2

& � � �

> , ?

) 2 1 2 2 2 2 6

2 1 3 3 3 3 F

0 1 2 2 2 2

E 1 2 2 2 4

< 1 2 2 2 4

6 1 2 2 2 0

F 1 2 2 2 4

C 1 2 2 2 4

D 1 2 2 2 4

3 1 2 2 2 0

4 2 1 2 2 2 4

3 1 2 2 2 0

D 1 2 2 2 0

� � � � �

> J ?

2 1 2 2 E

2 1 2 2 2

2 1 2 2 0

2 1 2 2 4

2 1 2 2 4

2 1 2 2 0

2 1 2 2 4

2 1 2 2 4

2 1 2 2 4

2 1 2 2 0

2 1 2 2 0

2 1 2 2 0

2 1 2 2 0� �

9 � / � ( � <

� ( ( � <

< 1 2 2 2 24 F 1 2 2 2 2 � 5

0 0 1 2 3 1 0 2 2 2 � � 4 0 5 4 6 �

� � � � � , � � � � % �� � � � � � � � � � � � � � � � � � � � � � � �

/ � � � � � � � � � � � ! � � � � � � � � � � � � � � � � �� � 6 7 � � � � � � � � � � � 1 �� � � � � � � � � � � � � � � � � � � � � � ' � � � � �� � � � � & � ! � � 1

� � � � � � � � 8 � % � � � � � � � � � � � � � � � � � 8 � � � � � � � 9 � � � � � � � � � � � � � � � � � � � � �: � � � - � 4 � � � � � � � �: � � � - � 0 � � � � � � � � � 1

" � � � . � � 5� � � � � / � � � � � � � %� � � � � / � . � � � � � � � � % � � � � � � � � � %

; � # �

� � � �

9 � / � ( � <

� ( ( � <

< 1 2 2 2 20 2 L � 5

������: �� ���

Page 28: MC5 User Manual

��

�������

<���%

L�������� ��������� ��� �� ���� �������� ����������� �� ��� ����, ���� � � �%�/��� �����(�&�����:���� ��F�"� �� ���)����� ������� ���� ������������������������� �� ��%

*������������� ��������+�"�0� #� ����� � ������)� ������ �������� ��� ���%�*���+�"����"���!��������������%

��+��������

3�������"��� ��������� ��+�"��������( ����:���� ��F�"��)� ����� ��������� ���� �� �� �� ��������� � ���� ��,�� � ���%�*���� �������� ��������)� ����������������!���� ��� ������ �%� �/��� ���������� �� +�"�� 56� �� H7� �� ������� ������������%�/��� ����������+�"�� ��)�� ���� ������ �� ���� ��,�� � ���%9�������������������� ������� ������ � +�"%�/��� �����(�&����:���� ��F�"������)���� ���������������%� ��� ���� ���������� "� ���������� ���������� ������� �����)� ����������������#���"����� ��

���� �+�"������������������� )��%

����!�� ������ ���,�� ��� �����(�������:���� ��F�"%�*�� �����5������7������� ������,�#����������(������:���� ��F�"%

# � � � � � � � $ � � � 4 2 F 1 4

4 F 1 4 2 1 0 2 2 2 � � 4 6 5 0 4 �

J � � � � � � �

2 1 6 2

2 1 E 2

2 1 4 2

2 1 4 2

� � � � � � � � � ' � � � � � � � � � � �

' � � � $ �

. � ; � � �

� M � # �

� � ( � & �

� � 9 � � �

� � � � " �

, � : � B �

N � A � O �

P � Q � & �

R � S � T �

U � V � W �

� � � � X � � Y �

Z � [ � \ �

� � / � � �

� � , � � � � $ �

� � , � � � � % � � �

� � � � � � " � � � � � � � ? � 6

9 � @ � � � � � �

5 6 @ � � � � � �

� � � � � � 5 6 @ � � � � � �

5 6 @ � � � � � �

=

=

]

]

# � � ( - �

Page 29: MC5 User Manual

�8

<����<���������

2�!�2 �� � ���� ���� ����� ��������� ���� � � ������������!�����)�����%�L����)��������������������)� ������!� ��%�*��� � ��� �� �)� �������!� ��� ��� �!��"���� ��������� � �� ��)�� ����2�!�2 �� � ��� ���%

��2�!�2 ��� ���!���� ����"��!��������� �+�"�����"����������� ��+�"���������( ����:���� ��F�"��)� ������ ��������� ����� �� �� �%�9�������� �� �������!!���� ������(��� ���� �������� �� ���������������� ��� ����������������) � ����!���%

/���� ���������������+�"�� ����� ������� ����� �+�"��������������� �����)� ������!� ��%�9����������� ����!� ��� ��� ���

�+�"������� �+�"%

���8�������

3!��!�� ���� ���� � � ���� ��2�!�2 ��� ���� �,��!�� �����3!��!� �����!!���� ����4���� �� ��������%�*����&����! ����������3!��!�� ��� �����!�������������6����� ����!���6D%

9���� ����3!��!�� �������������� ��������� � ���� �"�5���!�����2�!�2 ��� ���70�>���� � ���"��!���� ��������� ������+�"������!���������3!��!�� ��#�������������)������������!%

"�������������

9����� ��� ������������� ����"���)����������������)�����!�� ��%�9����� ��� �����������������#����� ������ �����) �� ���� ��� �� %

9����� �� � ����������� ����� �������� ) � ���� !���%�;���� ���� �����5���� � � ��� ���� �)������ ��,�7� ����������������"��!���������+�"#������ ��������������� �!��"������!� ��%

*��� ����� �+�"������������� ������� ��������� ��%�*��� �����

+�"��������������� ������!��������� ���5 ���!!� �����7%

9���������������!� ��� ��������(#�����:���� ��F�"���� �������� �+�"������� �+�"%

� � � � � $ � � � � � � � ? � 6

� � � � �

M � � � � � � � � � � � �

� � � � � $ � � � � � � � ? � 6

4 F 1 4 2 1 0 2 2 2 � � 4 6 5 0 4 �

� & � � � � & ( � ' � , � � � � � '

4 2 4 ) B # ) 2 2 4 1 44 4 0 ) � � ) 2 2 E 1 4

� � -� � 4 2 F 1 4� � 4 4 0 1 4 0� � 4 4 0 1 4 6 ) 4� � 4 4 0 1 4 6 ) 0� � 4 4 0 1 4 F� �

4 4 0 ) � � ) 2 2 C 1 4

, ,

������: �� ���

Page 30: MC5 User Manual

�=

�������

�� ������������������������

������ ����������)�����!� ��������������� ������ ��������� ��� ���%�*� ����+��� ��!�� ���� ����"������ ������ �����!�� � � ������� �� ���������� ��$� �������%� ������ � ������������ ��� ������#������������������"�������������"�� �����)������������� ��������������%�

#���$����������%����

���������&�������� �������������1����/� ��51/7%��������������������!� ���#�����������������������!� �������� ��������������������)� ������ ����������������(�������(� ���������� ���%�� *�������������� � ��������������������������� ��������(��������%�9������!��������������� ������!����������� �� ���� ������ �������� ���� ��������%

�� >������&��-�.*������ ��� ����$� ����������������!��� )�� �� ����!� �� �� ��������%

MODULE DESCRIPTION

Base Unit (BU) Required module. Includes the case,display, keyboard, battery pack, commonelectronics and the ENV, AUX andCOMP/PRT connectors as well as aconnector for an external pressure module(PX1).

Internal PressureModules(1

(P1, P2 and P3)

Modules with positive and compoundgauge pressure measurement capabilityand a barometric module enabling alsoabsolute pressure measurement togetherwith other modules.

External PressureModule connector(1

(PX1)

Connection for external modules capableof measuring high pressures up to 1000bar (approx. 14500 psi).

Electrical Module(E)

Voltage, low voltage, current and frequencymeasurement. Also pulse counting, switchtesting, HART communication(2, currentgeneration, and 24 V loop supply.

Electrical andTemperatureModule (ET)

Resistance, RTD and thermocouplemeasurement/simulation. Low voltagemeasurement/generation. Voltage, current,frequency and pulse generation.

Internal ReferenceJunction Module(RJ)

Internal reference junction compensationfor thermocouple measurement/simulation.Cannot be used without the ET Module.

Page 31: MC5 User Manual

��

)����� ���������������

*����� ����� ������ �������������) �������������������������������%�*������ ��� ��� ���������������������) �������������� ������������"�������������� ������ ������������ �� �����!�������5)�� �� ������ ��������� ���!� ����70

� >,�������3���������������5>G*7� >�) ����������!��������������5><I7� 1�����"�!�������!�������!� �����5���3(3.*7� 3�����������������0�3��=#

2���+�23��6������2���+�23���6��5�/G7� *��!��������1����(2�"�1��+�0�������#

*�+F� #������+(J��������-�.*�9� ��� � ��5�/G7%

��� � ���� ������ ��������������"�!�������!�������!� ����#�!����������������������� ���!�������������(��"����+�� ���)� ������ �!���C������ ��������%

&��$�������� *������������� �� ������ !!��� ��������� ����!�������!������ �����������������������#������� �(� ����� ��������� ���� ����+�%�*���!� �������� )��"����� � ������������������������������&������� ��� �"%�

*������ ��� �� ����!� �������� �����������"��)� �������� ����� ������������ �� �����!������5)�� �� ������ ��������� ��!� ����70

� 9!�� ������!��������������� 2� )���� ���>,�������2�) ���� ������������������ �� �����/G

��������%� -�.*������ ��� ��5.�$� ��������>������7� ���� ������������� �

�����&�������� �"������!� ��

Page 32: MC5 User Manual

��

�������

������

�����&�� ����� �� ����(����� !��� 5�3��7%�*���������"� !��+����� �)�����)����������������!�!���)��� ��� �������������������%9������������ ���� �+ �� �� ������ � ��%�

*�������� �����������&������� �������������� ������ ������ �� ��%������������������+�� ��� �������!��� ������ ��� �� �!���!�������

���������!��������������� ����������� ���������������5?��!� 7�����������)��!��������!�������%��������������������!�����������!���������������,���������������&����, ����!��������)����#����)��!��������!�������)������,�����!�������������������� ��������������������%

��������� �� �������� ���'(��"�������� ������ ������)�������������������>���� ���� )��D?(88�(>>���������������"����� �"� ���������� ������� '������������0

><����D6�6 >� �� �#><����D6�6 ����� �"#><��6����8�� -���� ����������#><��6����8�8 I������������� ��#

��������� �)������ ���� )��H8(�8(>>���������������"������ �" ���������� ������� '�����������0

><���?�� � �I����%

Page 33: MC5 User Manual

�H

�������*������� �� ����� ���������� ������ ����!��� � ����� ���� ������������������������"��+ �����!�!��%�;�+ �� ��������� �)�)��������������!��������#� ���!�����������(�������� ���� ����������%�1������� ��+� � ��� �+� ��������� ���������������� ��������"�������(� ���������!���������������� ������������� ����������������� !�#����%

/�����������"� ��"����������� ��������� �������������������"%�9��������������� ���������!�� ���� �������������������� ���������������0

� ;��������������������� ���) �����"�������� ;���������� ���������� � ������,!������ ������!������������ ������)��������� � ��� ��������� ������������ �������!��

9��� ���� �� ����������"���������!���������� �������� )��� � �� �+ �� ���������� �����%�*�!��)������� ���� ����������������"������� ����$����"� �����������#�+��!�������� �����5���������6������7����������� ��������������� ��� ����������� ���� �� � ������� �%

*�������+������

�������!� $�������%�����$!�#� �+���������������'����� ������� ���������

��������������$ ����!�� �������� ����8� 4 �������������� 8��������!� ������!����������� ��$� ����

(��!������������������ ����� % �� $��� �� ���������������� �!��������������� !�� ����������������������������%���$��������� �

���������� �E �����$ ����������� ���� ��� $���# ����!�� E��������������$ ����!��!���� ������������� ��� 4 ������ ����<�����������$ ���������������!� ������������ ������ ������

�%����������������������$ ����������������������������� !� ���$�������!�F�� ������ ������%�������%��������>����F� �� �� �������� ���� � ����$ �������� ��� �������� ��� ������� ���

:�� ��� $���$ �������� !�%����� ��� ��������� ���������� ���������%������� �� �+���� !�$�������!��+�����%�F�$���� ��!������������� ����!���������� � ������������ %����� �8���F�������!����������� ��� ����� �����%������ ������� �� �8��������� �����

����� ��������������!�$�������������� ��������� ��� ��������������+����B��.��/-�B�.23�

9����"

Page 34: MC5 User Manual

�D

�������

+������� �������������������,�������%�������-,�����,�.

���� � ��� ���� �������� �������� �� ������������ ������8������ � ������%���%��� ��� ����%��������� ��� �� ������ �$���������$���������� ������������ � !������������ �������������F��� ��!�� �������� �%��� ����������A<�7,�$������� �!���� �� ����������� �����������%����@���%��F������+��������������E�������������� !� ��������� � !�

����������������� �� �% ������� �����$���������������&�� ��&� �����F���� ������������� �������������!������ +�����A����������� �������$������������� ����������

' +� ������%��� ������ ������������ �� ������$�����G�CA�6��!�F�����%��F����������������%��� ������� �����&�� ���&��������������!���������+���� ��%��� �������'��F����������8����%��� ��� !�$����������������$��� 4 �����

*�������+������� �����������������%������!���

���� �����!����!���� ��������������������� �������������������� ���������� ����������� +� ������������,�C$ � ��,-.��/,0�C���� ��1�.23������!������������������� ��$��� 48 �����

��� %����� ����� ���� ��$� ���F� ���� ��� ��������������!����� ��������� ���� �������� � ��� ���� ������ / +����9����> 3��6�����������������������9�������������������8�������/�!�� ��!��������� ����������� ���������� ������,��$ ����� ���3F� ���!�������������������� ��� �������������������$��!�����+ ��� ��� ���

��� !�����������4�������!��� �$������������������������� �!������������������������������������������% �%�������%���8���������!��� ��&������� �� ������������� ��� ���������! ����� ����������� ���������������� ������ ��

��� !����������������� ��� ��� ����������� ����������E������������� $����������� ��� � !�������!� ����������� ���������������� �� ����������E�������� ���� �������� �����������&+���� �� ������� %���������E���������� ������������

>�%����+�������� +� ������������ ��������� ����F�$���������� ������+���� ��������������� ������� +� �����8��������� ����������� ����������E�������� +� ������������+���� �� ��������� ���� ����������������6����������� �8������ ��������%���������������+���� �� �����

>�%������� �����������!���� ��������������� ��������������� �� ����� !�� ������� � ��� E ������� ��$$$����������$������� ����� �� ����� ���

Page 35: MC5 User Manual

�?

+������� ��������&�����������

@��������������� �� !��� ��������(��!���������������������+�������� ���E���������������������������9��F� ��� �����8����������� ������ ��� �������������E���������� ������:� � �����!� �������������������� ����� ��� ���!�������������������������������� ������$�������� �������������

)���������� �F������!��� � ���������� ��� �!���9��F�����8� ��!����!���������E�����������!� !��� ������������������������ �� ��������������������� ������ ��� ���������8���� ��� ����9����������� ��� �������$��������������������� ����������� ���������� ����������$ ��/G��������3��� ����

6��!�������������F� ��� �4�������� E�������� � �������� ��� E�� ���������������%����� ����������������% �%��������������8����!������F�����������!��� �������� ������� 6��� ��� ������������ ���������������������� $����� ��� !� ������8��������� ����� ����������� �������:� �� �����!� ���� ���!������������������������� ��� E�������� �������������������������� ���� �� ��� � ����������� �����

���������9����������� ��� ������� ������������������ � ��� ���� ������ ������������������ ���������� ��������� $����!�� �����������E��� ������������9���������� �����8��%�� � ����� ���� ���� ������� ��� ���� �� ��������� �$����)�����������9��F� ��� �4������ ������� �����������!��� �"��!�� �� ��� �4�� ���� ��������������� ��9��� ��� ������� E ���

<�������������� ���$���������������������9�������� ����

���E��� �������� ������ ������� !� $������������� ���������������4���%���������������� ��������� ��!����������������� ����������!��� �����������������������������8���� ���%�� ���+����� ���� ��� �� ��������%�� ����������!��� ��������������������� ������������������

@������������� ������ �������$� ������ ��$�� E�������8� ����� ���������!�������������� !� �����D�!�������� ����� E����� �� !�$��� �������$� ������������%�����!���������� E8����� ��D����� $���������� �������������E����6�� �� ��$$$����������������$�������� ����F���� �� ����� ���������� E�D������ ���� ��!������� ��%�F� ��� �� �� ��9��� ���� ��������� ���� ��

9����"

Page 36: MC5 User Manual

8�

�������

������

���"�$��� � ������) ���!����������"�!������� ������)����� ������������������%�>�%������������ ���������� %���+��������8������������ �( �� ���� ��� ����������� ��%���

*��������#�� �)������� ��� ����������"����� ����������"��%

&��$����������*���$� �+���� �"������� ������ �� �� ����)��� �� ���)� ������ �����+ ���������&�� ���� ��%���������M2 �����N�!���������� ���� ����"�������������� �� ����)��� �������� �����%

����"������� ����3���������!���������������!���������� ���� ��������������������������������������� ���!���� ��"��3�%

.�������������+�!��������� ��������������� �������#��� ���%%����� ���� ��������������� ���%���������+����!�� ��������������������!��" �������!������� ��%

+�������� ����,�������������� "��� ����� ��!�������� )�� ��� ������ �� ������� ����������� ���� ���������%�

�������������������*���������� ����� ����������������) ���%�;������������� ����������� ��� �"������ ��������� ������ �!���������� ��%�

Page 37: MC5 User Manual

86

���� ����,�������������������� �#������������+��� ������ ������� �������� ����!�5! �����!7%�;� �������� �� ���������������� ������ ��� ������ ������� ���!���� ����%�<�)��� ������"� ���������������%�

����������� ��������������/�������� ��������0������%�����

*������������� ���� ���������.���������J���� ��1��+���"���������� �� ����� ����� � ��%�*��� � ���!�� ��)�� �����!��� ���������) ������������� ������� �%

��������"�!��������)������������������.���������J���� ��1��+�"��� �������� �� )�������� �����%�< �"���������������������%.��)������ �!�� � �������!��������+������)��%�*����)��� ����������� ����"����������� �+%

9��) ��

Page 38: MC5 User Manual

8�

�������

5>�!�"7

Page 39: MC5 User Manual

Startup and BasicOperation

Things discussed in Part B:

• What happens during thestartup procedure.

• Measuring signals and doingsome special measurements.

• Generating/simulating signals.• Step and Ramp functions.• Alarm limits.

Page 40: MC5 User Manual

34

Startup and Basic Operation

Starting CL560

Startup ProcedureWhen CL560 is started, a startup picture appears. After a self test,some basic information of the calibrator at hand appears in the lowerpart of the screen. If you want to view the calibrator information fora longer period, press the D/Wait function key. Then the calibratorinformation is visible until you press the D/Continue function key.

25.11.2002 13:52

Wait

Serial number

E module version

Calibration due date

Main version

ET module version

23512365

1.50

20.03.2003

1.90

1.70

www.beamex.com

If a module’s version number cannot be seen, the module is notincluded in the CL560 at hand.

The calibration due date that is listed in the startup window is theearliest calibration due date for all connected modules. If the cali-brator requires recalibration, CL560 stops at the calibrator informa-tion window and the text “Calibrate Soon” appears below the cali-bration due date row.

Page 41: MC5 User Manual

35

Basic Mode, DefinedEvery time CL560 is switched on, the startup procedure ends inBasic Mode.

All non-calibration related measurements and generations are per-formed in the Basic Mode. Briefly: in Basic Mode CL560 works likea high quality multimeter. When returning from CL560’s higher leveloperations (calibration, viewing of calibration results, calibrator anduser configurations), you always return to the Basic Mode.

Where should I go today...

Basic Mode: Measurement Generation...

In Basic Mode, the two available measurement/ generation/simulation windows have default settings based either on factorysettings or settings defined when CL560 was previously used.

The first time the D/Menu key is pressed, the Window 1 Setupmenu is available. Other possible menus can be selected from thefunction keys: B/Window 2 Setup and C/Others. The latter functionkey opens a menu with some special functions and also includesthe possibility to go to higher level operations.

Starting CL560

Page 42: MC5 User Manual

36

Startup and Basic Operation

Example of a Basic Mode screenwith pressure measurement config-ured in Window 1 and current mea-surement configured in Window 2:

What can be done in Basic Mode

• Measure signals (*• Generate signals (*• Simulate signals (*• Start special measurement (min/max value etc.)• Perform a Limit Switch Test• Set alarm limits• Use the ramping function• Use the stepping function

*) Available options depend on the installed modules.

Next…

Measuring on page 37Generating/Simulating on page 58Special Measurements on page 53Alarm Limit Settings on page 73Special Generations on page 69.

22.05.2000 8:33 1 Pressure

2 Current

P2:INT20C/-1.0000...20.6840 bar g

E: Current Measurement

gaugebar2.6475

12.4731

ModeCalibration MENU

mA

Page 43: MC5 User Manual

37

Measuring

All measurements in Basic Mode require that you first select theWindow to be used (Commands: Start with D/Menu and continueeither with A/Window 1 Setup or B/Window 2 Setup). Each mea-surement has its own unique 1/Quantity and 2/Function/Port set-tings in their window’s menu. The other window menu settings, e.g.measuring unit, refine the measurement characteristics.

When presenting measurements in this manual, the first paragraphtells the module (or modules) that is/are required for the measure-ment. Because of CL560’s modularity you may or may not have therequired module. If the module is not included in your CL560, the1/Quantity and 2/Function/Port settings needed for the measure-ment are not available as choices in the pop-up lists.

Each measurement also has at least one picture with a circle aroundsome of CL560’s terminals, like the one below.

The circle indicates active terminals for each 1/Quantity and2/Function/Port setting in the window menu.

If the picture has more than two terminals circled, then the lighterpart is somehow optional. In the following picture, the HART termi-nal is optional during current measurement.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,T/C, Low V

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V I meas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,+24V I

2-w xmtr

Measuring

Page 44: MC5 User Manual

38

Startup and Basic Operation

Warning!

Do not apply voltage higher than 50 V (max 2 A) between anyterminals.

Page 45: MC5 User Manual

39

Pressure MeasurementSee chapter Things to Consider when Measuring Pressure on page 95 for more infor-mation on pressure measurement and internal/external pressure modules.

Required settings Options/description

Quantity PressurePressure Type g gauge pressure or

abs absolute pressure.

The available pressure types may be restricted because of the selected pressure port /pressure module. For more information concerning pressure types, see chapter Pres-sure Type on page 95 .

Using Internal Modules

Select an internal pressure module portwith a suitable measuring range and withan ability to measure the required pressuretype:

Port P1: INTxxxx,P2: INTyyyy orP3: INTzzzz.

Using External Modules

Select an external pressure module portwith a connected pressure module and asuitable measuring range and with an abil-ity to measure the required pressure type,e.g.:

Port PX1: EXTxxxx

T / C , L o w V4 - w m e a s

R , R T D

3 & 4 - w m e a s

V , I , + 2 4 V I M E A S / S I N K

V , H A R T ®

L o w VM E A S & IO U T P U T G E NS E N S O R M E A S U R E & S I M U L A T E

E T E

T / C I N T . R J

T / C O R E X TW I R E S O N L Y

2 - w x m t r

T / C , L o w V4 - w m e a s

R , R T D

3 & 4 - w m e a s

V , I , + 2 4 V I M E A S / S I N K

V , H A R T ®

L o w VM E A S & IO U T P U T G E NS E N S O R M E A S U R E & S I M U L A T E

E T E

T / C IN T . R J

T / C O R E X TW I R E S O N L Y

2 - w x m t r

Connecting and Removing External Pressure Modules

An external pressure module may be connected and removed atany time. If a removed module was part of an active measurement,CL560 automatically changes the measurement to a suitable inter-nal pressure module. CL560 also emits a “beep” to inform you ofthe fact that the external pressure module used for pressure mea-surement was disconnected.

Measuring

Page 46: MC5 User Manual

40

Startup and Basic Operation

Configuring a Supply Voltage or Current to the Pressure Sensor

To configure the pressuresensor supply setting, Openthe appropriate windowsetup menu (select D/Menu,A/Window 1 Setup orB/Window 2 Setup). Con-tinue to the third submenuby pressing the numericalkey twice. The adjacentpicture displays all thesubmenus available whenthe selected Quantity isPressure.

Select 2/Sensor Supply and choose either Current or Voltage sup-ply. A field for entering the supply Voltage/Current value appears inthe pressure measurement window.

Zeroing a Pressure Module

If the selected pressure module does not display zero gauge pres-sure when the applied pressure is zero, the module has to be ze-roed.

Open the appropriate window setup menu (D/Menu, A/Window 1Setup or B/Window 2 Setup) and select menu option7/Zero Pressure Module. If a secondary pressure module is ac-tive in the selected window, a pop-up menu will appear for choosingeither to zero the primary or the secondary pressure module.

NOTE!

Zeroing a pressure module is especially important when theoperating position of CL560 is changed or the location of CL560is changed in the vertical direction. Both of the above men-tioned factors affect notably on the pressure measurement mod-ules. Measuring pressure below 100 mbar (approx. 40 iwc)should be done with a firmly mounted CL560 (e.g. placed on atable top).

Next…

Special Measurements on page 53Alarm Limit Settings on page 73Special Generations on page 69Calibration, see Part D.

mbar(gauge)

Alarm <

0.5 bar

ControllerSettings

CloseMENUMode

OthersWindow 1 Window 2Setup Setup

Sensor Supply[None]

mbar(gauge)

Alarm <

0.5 bar

SpecialMeasurement[None]

[--(--)]

Function and Port Info

Alarm

Second

Extra Info

Port Info

CloseMENUMode

OthersWindow 1 Window 2Setup Setup

[P2: EXT100]

Second port

CloseMENUMode

Others

0.70834 mbar(gauge)

Alarm <P1: INT2C / -1.000000...2 bar

0.5

0.4374

2 Current

bar

Pressure

E: Current Measurement

Quantity[Pressure]

[P1: INT2C]

[Eng. Unit]

Unit[bar]

Funct/Port

Display Mode

06.05.2002 15:32 1

HART

PressureType[gauge]Zero PressureModule

Window 1 Window 2Setup Setup

Page 47: MC5 User Manual

41

Current MeasurementThe current measurement terminals are located in the E module.

Required settings Options/description

Quantity CurrentFunct/Port E: I(meas)

External supply

When the measuring circuit includes anexternal power supply, use the terminalsshown in the picture below.

Internal Supply

When you want to use CL560’s 24 V termi-nal to supply the measuring circuit, use theterminals shown in the picture below.

Based on the terminals in use, CL560 automatically acts as either apure current measuring unit or as a current measuring unit while atthe same time supplying the measuring circuit.

Notes.

Check the polarity of your connections. The arrows in the previouspictures describe the correct flow of current.

If you are using a transmitter with HART capability, make also aconnection to CL560’s HART terminal. For more information con-cerning HART connections, see Appendix 1, chapter ConnectingCL560 and a HART Instrument on page 138. Information con-cerning current measurement parallel to a test diode can be foundin part C, chapter Current Measurement Parallel to a Test Diode,Connections on page 103.

Next…

Current Generation on page 60Special Measurements on page 53.Alarm Limit Settings on page 73.Calibration, see Part D.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I, meas/sink

2-w xmtr

I T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,+24V I

2-w xmtr

Measuring

Page 48: MC5 User Manual

42

Startup and Basic Operation

Voltage MeasurementThe E module has terminals for low voltage measurement within the range ±1 V andterminals for voltage measurement within ±50 V range. The ET module also has low volt-age measurement terminals with a range of ±500 mV. The ET module terminals are alsoused when measuring/simulating thermocouples using an external Reference Junction.

Required settings Options/description

Quantity VoltageFunct/Port ET: LowV(mea), E: LowV(mea) or E: V(meas)

CL560 displays the measured low voltage in the selected window.

Hint!

Low voltage measurement can be used for non-standard thermo-couple measurement. You will see the measured temperature inmillivolts and need a table to convert the measured millivolt value tocorresponding temperature values. In this case, use copper exten-sion cords to connect the non-standard thermocouple to CL560’sterminals.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,T/C, Low V

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET E

Max input:60 VDC/30 VAC

V, I,

Measuring Low Voltages

Select either Function/port ET: LowV(mea)or E: LowV(mea) and choose a suitableunit. The following pictures display the ac-tive terminals for both available ports.

ET: LowV(mea), ±500 mV:

E: LowV(mea), ±1 V:

Page 49: MC5 User Manual

43

Measuring Voltages up to ±50 V

Select Function/port E: V(meas) and choose a suitable unit.

CL560 displays the measured voltage in the selected window.

Warning!

Do not apply voltages higher than 50 V between any of CL560’sterminals.

Next…

Voltage Generation on page 62Thermocouple Measurement (Temperature) on page 51Special Measurements on page 53.Alarm Limit Settings on page 73.Calibration, see Part D.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,

Measuring

Page 50: MC5 User Manual

44

Startup and Basic Operation

Resistance MeasurementResistance measurement terminals are located in the ET module.

Required settings Options/description

Quantity ResistanceFunct/Port ET: R(meas)

The following picture displays the activeterminals:

The two leftmost terminals are used in 2-wire systems. CL560 automatically checksthe connection and displays the found wir-ing system (2-wire, 3-wire or 4-wire) in themeasuring window. For more informationconcerning wiring options, see Resistanceand RTD Measurement, Connections onpage 101.

Note.

If the measured resistance value is infinite or very high (> 4000 ohm),the text “+OVER” is displayed in the measuring window. This meansthat the circuit is broken or the connection is wrong. Wrong connec-tion may also cause erroneous reading, typically too low. If neces-sary, use the 2-wire ohm measurement to check the wiring beforefinal connection.

Next…

RTD and Resistance Simulation on page 66RTD Measurement (Temperature) on page 50Special Measurements on page 53.Alarm Limit Settings on page 73.Calibration, see Part D.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,4-w meas

R, RTD

Page 51: MC5 User Manual

45

Switch State SensingThe switch state detection terminals are located in the E module.

Required settings Options/description

Quantity SwitchFunct/Port E: Switch (selected automatically when the

corresponding Quantity setting is activated )

Also check the Sound setting. The optionChange means that CL560 beeps everytime the switch changes its state. Whenusing options Open or Closed the soundis continuously on when the switch isopen respectively closed.

The contact should be free of external po-tential. If this is not possible, use DC volt-age within the range -10 V to +30 V. CL560

shows voltages above approx. +1.5 V asopen contact and voltages below approx.+1.5 V as closed contact.

Hint!

Switch state detection may also be used for binary signal detection.

Next…

Limit Switch Test on page 46Calibration, see Part D.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,

Measuring

Page 52: MC5 User Manual

46

Startup and Basic Operation

Limit Switch TestA limit switch test displays the opening and closing point of a limitswitch. CL560 supports limit switch testing of any type of limitswitches as long as CL560 is able to either measure or generate/simulate the switch’s input signal and is also capable of detectingthe switch state.

Notes.

Basic Mode’s limit switch test is a restricted version compared tothe limit switch test available in Calibration Mode. This test givesyou the approximates of the latest opening and closing points andalso the current switch status. The limit switch test in CalibrationMode offers more accurate results and additionally, statistical dataof repeated switch tests. More of the limit switch test in CalibrationMode in Part D of this manual.

In Basic Mode CL560 does not support limit switch testing whensimulating resistance or an RTD sensor. The limit switch test inCalibration Mode also supports resistance and RTD sensor input.

Performing the Limit Switch Test

Configure, e.g. Window 1 to either mea-sure or generate/simulate the switch’s in-put signal (Function keys D/Menu andA/Window 1 Setup in Basic Mode) andWindow 2 for switch state detection (Func-tion keys D/Menu and B/Window 2 Setupin Basic Mode). It doesn’t matter which win-dow is assigned for which function as longas both the switch’s input signal and theswitch state are available for CL560.

Slowly change the switch’s input signal andwatch how the “Opened @” and “Closed@” data is updated as the switch changesits state.

07.01.2002 15:15 1 Voltage

2 Switch

ET: Voltage Generation

E: Switch Sense

Measurement V

V1.87000

Closed

ModeCalibration

RampingStop MENU

Opened @ VClosed @ V

1.86712

0.985491.29990

Field

Page 53: MC5 User Manual

47

Note.

The accuracy of the “Opened @” and “Closed @” values dependgreatly on the change rate of the switch’s input signal.

Hint.

If you use CL560 to generate/simulate the switch’s input signal,you can use ramping to create changing input signals for the switchunder test.

Where to find more information when measuring a switch’s input signal…

Pressure Measurement on page 39Using External Modules on page 39Voltage Measurement on page 42Current Measurement on page 41Thermocouple Measurement (Temperature) on page 51RTD Measurement (Temperature) on page 50

Where to find more information when generating a switch’s input signal…

Voltage Generation on page 62Current Generation on page 60Thermocouple Simulation on page 67Ramping on page 71

Measuring

Page 54: MC5 User Manual

48

Startup and Basic Operation

Frequency MeasurementThe frequency measurement terminals are located in the E module.

Required settings Options/description

Quantity Freq.Funct/Port E: f(meas)

Also check the Unit setting.

CL560 compares the external potentialagainst an adjustable reference voltage(Trigger Level, range: -1 … +15 V) whenmeasuring the frequency.

Next…

Switch State Sensing on page 45Pulse Counting on page 49Frequency Generation on page 64Calibration, see Part D.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,

Page 55: MC5 User Manual

49

Pulse CountingThe pulse counter terminals are located in the E module.

Required settings Options/description

Quantity PulsesFunct/Port E: Pls(count)

Also check the Trigg.Edge setting.

CL560 compares the external potentialagainst an adjustable reference voltage(Trigger Level, range: -1 … +15 V) whencounting pulses.

The counter may be cleared (zeroed) by selecting D/Menu, 6/Clearcounter.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,

Next…

Frequency Measurement on page 48Switch State Sensing on page 45Pulse Generation on page 65Calibration, see Part D.

Measuring

Page 56: MC5 User Manual

50

Startup and Basic Operation

RTD Measurement (Temperature)RTD-measurement terminals are located in the ET module.

Required settings Options/description

Quantity RTD-Temp.Funct/Port ET: RTD(mea)Sensor Type Available RTD sensors

The following picture displays the activeterminals:

The two leftmost terminals are used in 2-wire systems. CL560 automatically checksthe connection and displays the found wir-ing system (2-wire, 3-wire or 4-wire) in themeasuring window. For more informationconcerning wiring options, see Resistanceand RTD Measurement, Connections onpage 101.

Note.

If the measured resistance value is infinite or very high (> 4000ohm), the text “+OVER” is displayed in the measuring window. Thismeans that the circuit is broken or the connection is wrong. Wrongconnection may also cause erroneous reading, typically too low. Ifnecessary, use the 2-wire ohm measurement to check the wiringbefore final connection.

Next…

RTD and Resistance Simulation on page 66Thermocouple Measurement (Temperature) on page 51Resistance Measurement on page 44Special Measurements on page 53.Alarm Limit Settings on page 73.Calibration, see Part D.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,4-w meas

R, RTD

Page 57: MC5 User Manual

51

Thermocouple Measurement (Temperature)Thermocouple measurement terminals are located in the ET module.

Required settings Options/description

Quantity T/C-TempFunct/Port ET: TCi(mea) (for the internal Reference Junction), or

ET: TCx(mea) (for other RJ compensation methods)Sensor Type Available thermocouplesReference Junction Depends on the Function/Port setting.

See subsequent chapters.Second port Available only if RTD sensor is not used as the Reference

Junction compensation method.

Internal Reference Junction

CL560’s Internal Reference Junction Mod-ule is an optional module. To use the Inter-nal Reference Junction, select Function/Port ET: TCi(mea). The Reference Junc-tion compensation method Internal is au-tomatically selected.

External Reference Junction

To use an External Reference Junction,select Function/Port ET: TCx(mea) andchoose one of the available ReferenceJunction compensation methods: Entered,0°C or available RTD sensors.

Additional information is found in chapterInternal Reference Junction on page 97.

Additional information is found in chapterExternal Reference Junction on page 98.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,

T/C OR EXTWIRES ONLY

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,T/C, Low V

Measuring

Page 58: MC5 User Manual

52

Startup and Basic Operation

Note.

CL560 has a comprehensive set of predefined sensor types. Addi-tional sensor types are available as options.

Warning!

If you connect an RTD sensor to the ET module’s R, RTD con-nectors, there is no galvanic isolation between the thermo-couple and the RTD sensor.

Next…

Thermocouple Simulation on page 67RTD Measurement (Temperature) on page 50Voltage Measurement on page 42Special Measurements on page 53.Alarm Limit Settings on page 73.Calibration, see Part D.

Problems with thermovoltage measurement? See Error situationson page 100.

Page 59: MC5 User Manual

53

Special Measurements

Special measurements are extra utilities that can be activated toperform a special function alongside the normal measurement.

Only one of the special measurements may be activated at a giventime for each window/quantity. Activating another special measure-ment deactivates the earlier special measurement in the same win-dow (or the earlier special measurement using the same quantity).

All special measurements are started similarly:

• Select D/Menu followed by B/Window 2 Setup if needed.

• As the special measurement setting command is located inthe window setup’s submenu, select either the numerical key

or the cursor key to open the submenu.

• Then select 1/Special Measurements and the type of spe-cial measurement from the provided pop-up list (or deactivatea special measurement by selecting the option “None”).

Some of the special measurements do not require any additionalinformation. They start immediately after they are activated. Rateof Change Measurement, Special Filtering and Resolution,Redundant Measurement and Deviation Measurement do how-ever need some additional information. Therefore when you acti-vate them you will have to enter the required additional informationbefore the special measurement is started.

Notes.

Special measurements are only available in Basic Mode.

If you change any of the main settings of the measurement (Quan-tity, Funct/Port), the defined special measurement is deactivated.

To zero/restart a special measurement without altering any mainmeasurement settings, do the same menu commands you used forstarting the special measurement.

Special Measurements

Page 60: MC5 User Manual

54

Startup and Basic Operation

Mathematical Special MeasurementsMathematical special measurements may be assigned to all measurements.

Minimum value

Start the minimum value measurement by selecting:

D/Menu and B/Window 2 Setup, if needed, or  ,

1/Special Measurements,Minimum Reading.

Maximum value

Start the maximum value measurement by selecting:

D/Menu and B/Window 2 Setup, if needed, or  ,

1/Special Measurements,Maximum Reading.

Min/Max value

Start the minimum/maximum value measurement by selecting:

D/Menu and B/Window 2 Setup, if needed, or  ,

1/Special Measurements,Min/Max Readings.

Rate of Change

Start the rate of change mea-surement by selecting:

D/Menuand B/Window 2 Setup, ifneeded,

 or  ,1/Special Measurements,Rate of Change….

Before the special measurementis started CL560 prompts for therate of change unit which can beselected from the following op-tions: 1/s, 1/min and 1/h.

22.09.2000 13:22 RATE OF CHANGE MEASUREMENT

QuantityPortUnit

Rate of Change Unit

T/C-temperatureET: TCi(meas)°C

1/s

StartCancel Edit

Page 61: MC5 User Manual

55

Special Filtering and Resolution

Start using special filtering andresolution by selecting:

D/Menuand B/Window 2 Setup, ifneeded,

 or  ,1/Special Measurements,Filter and Resolution….

Select the Time Constant settingand how many decimals should becut off from the actual resolution ofthe current measurement. The fil-ter is a 1st order digital filter.

None means “No additional filter-ing” and “Use the modules own resolution” respectively.

Note.

You cannot “worsen” the resolution to less than one significant digit.

Deviation Measurement

In deviation measurement the read-ing of the main port is compared toa reference value. The referencevalue is subtracted from the read-ing of the main port.

Start the deviation measurement byselecting:

D/Menuand B/Window 2 Setup, ifneeded,

 or  ,1/Special Measurements,Deviation Meas….

You can enter the reference valuewhen you start the special measurement mode.

Note.

Keep in mind that when the displayed deviation reading is smallcompared to the actual reading, a significant part of the deviationvalue may be measurement error. See the specifications for mea-surement errors at the actual measurement level.

19.05.2000 13:55 SPECIAL FILTERING AND RESOLUTION

QuantityPortUnit

Time Constant

T/C-temperatureET: TCi(meas)°C

Decimals -2

The setting -1, for instance, means oneless decimals.

5 s

StartCancel Edit

22.09.2000 13:32DEVIATION MEASUREMENT

QuantityPort

Measure deviation from the value

T/C-temperatureET: TCi(mea)

22.0000 °C

StartCancel Edit

Special Measurements

Page 62: MC5 User Manual

56

Startup and Basic Operation

Special Measurements Using Two Ports SimultaneouslyThe following special measurements only apply when two ports aresimultaneously used in the same window. The Second Port set-ting in the window’s setup menu allows you to select another portfor the same window. The Second Port setting is enabled depend-ing on selected quantity and available modules.

Difference Measurement

In difference measurement the difference between the reading ofthe main port and the second port is calculated. The reading of thesecond port is subtracted from the reading of the main port.

Start the difference measurement by selecting:

D/Menu and B/Window 2 Setup, if needed, or  ,

1/Special Measurement,Difference Meas….

Notes.

Difference measurement is not available as an option in the list ofspecial measurements, unless two measurement ports are activatedin the same window.

No setting window appears, but the calculated difference is dis-played in the special measurement row.

The measurement value of the second port may be assigned to theExtra Info row at the bottom of the window.

Keep in mind that when the displayed difference reading is smallcompared to the actual reading, a significant part of the differ-ence value may be measurement error. See the specificationsfor measurement errors at the actual measurement level.

If the measuring spans of the selected ports are different, makesure you do not exceed the measurement range of either port.

Page 63: MC5 User Manual

57

Redundant Measurement

In redundant measurement themeasurements of the main port andthe second port are compared witheach other. If the readings differmore than the entered limit value,CL560 gives an audible alarm.

Start the redundant measurement byselecting:

D/Menuand B/Window 2 Setup, ifneeded,

 or  ,1/Special Measurement,Redundant Meas….

You can enter the allowed deviation between the two measurementswhen you start the special measurement mode.

Notes.

Redundant measurement is not available as an option in the list ofspecial measurements, unless two measurement ports are selectedfor use in the same window.

The measurement value of the second port is displayed on the spe-cial measurement row. You may assign the allowed deviation valueto be displayed on the Extra info row at the bottom of the window.

When setting the allowed difference, take the accuracies of themeasurements into account.

If the measuring spans of the selected ports are different, makesure you do not exceed the measurement range of either port.

22.09.2000 13:41REDUNDANT MEASUREMENT

Main MeasurementQuantity Pressure

0.00100 bar

Port P1

Second MeasurementQuantity PressurePort P2

While displaying the main measurement, compares it to the second and alarms if readings differ more than

StartCancel Edit

Special Measurements

Page 64: MC5 User Manual

58

Startup and Basic Operation

Generating/Simulating

GeneralThis chapter and the subchapters describe only how to generate/simulate signals using CL560’s own modules. Part C discusses theuse of external devices connected to the auxiliary interface.

CL560 is capable to perform the following generation/simulationfunctions:

• Voltage generation• Current generation• Frequency and pulse generation• Thermocouple simulation• RTD and resistance simulation

Generation/simulation in Basic Mode require that you first selectthe Window to be used (Commands: Start with D/Menu and con-tinue either with A/Window 1 Setup or B/Window 2 Setup). Eachgeneration/simulation has its own unique 1/Quantity and 2/Func-tion/Port settings in their window’s menu. The other window menusettings (generation/simulation unit etc.) refine the generation/simu-lation characteristics.

The presentation of each generation/simulation begins with a para-graph defining the modules that include terminals required for thegeneration/simulation (Most generated/simulated signals are donewith the ET module but current generation may also be done withthe E module). Because of CL560’s modularity you may or may nothave the required module. If the module is not included in yourCL560, the 1/Quantity and 2/Function/Port settings needed forthe generation/simulation are not available as choices in the pop-up lists.

Each generation/simulation also has at least one picture with a circlearound some of CL560’s terminals, like the one below.

The circle indicates active terminals for each 1/Quantity and2/Function/Port setting in the window menu.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,V, I,

Page 65: MC5 User Manual

59

Warnings!

When selecting a generation/simulation function, CL560 alwaysstarts with zero output (not when simulating a resistance). Thisis done to prevent damages in the connected instrument.

Be careful when increasing the generated/simulated signal. If aninstrument that cannot withstand the generated signal is con-nected to CL560, the instrument may get seriously damaged.

Changing the Value of the Generated/Simulated Signal

Move the field indicator on the display until it surrounds the value ofgenerated/simulated signal (use the B/Field Function Key or thecursor keys).

Enter the new value using the numeric keys. You may cancel theediting by pressing the A/Cancel Function Key. The C/çDeleteFunction Key removes the rightmost digit.

Accept the new value by pressing either the key or the D/OKFunction Key.

Note.

The dual function of the keys: , and is not available in ageneration field. The keys only represent numbers.

Fine Tuning the Generated/Simulated Signal

Press either of the left or right arrowkeys ( or ) to start the Fine Tun-ing of the selected numeric field.Then one of the digits in the num-ber in underlined indicating which digit can be fine tuned. If needed,press the left or right arrow buttons again to select another digit forFine Tuning.

To change the value of the undelined digit, use the up and downarrow keys ( ). The changes take effect immediately..

To end Fine Tuning, press either the key or the D/Close Func-tion Key. The C/Edit Function Key ends Fine Tuning and allows youto enter a number using the numeric keys.

Notes.

You cannot exceed the minimum/maximum limit of the quantity withthe Fine Tuning utility.

The fine tuned value follows the resolution properties of the gener-ated/simulated quantity.

1000.00

Generating/Simulating

Page 66: MC5 User Manual

60

Startup and Basic Operation

Current GenerationCL560 can generate current both with the E and the ET module. Additionally the E modulecan be used both in source and sink mode.

Required settings Options/description

Quantity CurrentFunct/Port ET: I(gen) or E: I(control)

Using the ET module’s outputterminals

When generating current from the ETmodule’s output terminals, select Funct/Port option ET: I(gen). The following pic-ture displays the active terminals:

Notes.

If you are simultaneously using the ET module’s output terminalsand some of the ET module’s measuring terminals, there is no gal-vanic isolation between the measuring circuit and the current gen-eration.

The voltage between the output connectors of the ET module islimited to ±15 Volts. There is no sinking possibility in the ET mod-ule.

Using the E module’s outputterminals

When generating current from the Emodule’s output terminals, you can eitheruse CL560’s internal 24V supply (sourcemode) or use an external supply and letCL560 control the current (sink mode).

Active terminals in source mode (below):

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,V, I,

Based on the terminals in use, CL560 automatically selects either source or sink mode.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,+24V I

2-w xmtr

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I, meas/sink

2-w xmtr

I

Active terminals in sink mode (below):

Page 67: MC5 User Manual

61

Note.

The output current of the E module is a “pure” analog signal withoutany embedded HART signal.

Notes applying to both the ET: I(gen) and the E: I(control) ports:

Check the polarity of your connections. The arrows in the previouspictures describe the correct flow of current.

If the loop resistance of the external circuit is too high or infinite, themessage “O.LOAD” is displayed until the load is reasonable.

Warnings!

Make sure that you don’t exceed the maximum current allowedby the instrument under test.

If you open the mA generation loop, CL560 tries to maintain thecurrent by increasing the output voltage. If you then close theloop again, the current is first too high, but returns quickly tothe correct level. If this current peak could damage the compo-nents of the loop, make sure that the loop will not open or pro-tect it against overcurrent. For the same reason, always enter0 mA output before connecting the loop.

Next…

Current Measurement on page 41Special Generations on page 69Calibration, see Part D.

Generating/Simulating

Page 68: MC5 User Manual

62

Startup and Basic Operation

Voltage GenerationThe ET module has terminals for voltage generation within ±12 V range and terminals forlow voltage sensor simulation with the range ±500 mV. The low voltage terminals arealso used when measuring/simulating thermocouples using an external Reference Junc-tion.

Required settings Options/description

Quantity VoltageFunct/Port ET: V(gen) or ET: LowV(sim)

Generating Voltages up to ±12 V

Select Function/port ET: V(gen) andchoose a suitable unit.

CL560 displays both the set value and theinternally measured value for the generatedvoltage in the selected window. Max. out-put current depends on the voltage levelas follows: 10 mA if the Voltage is within±10 V or 3 mA if the Voltage is outside of±10 V.

Low Voltage Generation

Select Function/port ET: LowV(sim) andchoose a suitable unit. The max. outputcurrent is 5 mA.

CL560 displays both the set value and theinternally measured value for the generatedvoltage in the selected window.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,V, I, T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,T/C, Low V

Page 69: MC5 User Manual

63

Note.

If the resistance of the external circuit in voltage generation is verylow (obvious short circuit), the message “O.LOAD” is displayed untilthe load is reasonable.

Hint!

Low voltage generation can be used when calibrating instrumentswith millivolt input. You can use it also to simulate special thermo-couples not defined in CL560. If you simulate a thermocouple inmillivolt mode, you must convert from temperature to millivolts manu-ally.

Warning!

If you short circuit the voltage output, CL560 tries to maintainthe voltage by increasing the output current. If you then removethe short circuit, the voltage is first too high, but returns quicklyto the correct level. If this voltage peak could damage the com-ponents of the circuit, make sure that the circuit can not beshorted or protect the circuit against overvoltage. For the samereason, always enter 0 V output before connecting the circuit.

Next…

Voltage Measurement on page 42Thermocouple Simulation on page 52Special Generations on page 69Calibration, see Part D.

Generating/Simulating

Page 70: MC5 User Manual

64

Startup and Basic Operation

Frequency GenerationThe frequency generation terminals are located in the ET module.

Required settings Options/description

Quantity Freq.Funct/Port ET: f(gen)

Also check the Unit and Waveform set-tings.

CL560 generates sine or square wave withgiven amplitude and frequency. Minimumallowed frequency is 1 Hz, but the specifi-cation is valid for frequencies above 40 Hzonly. When generating a square wave, theoutput DC-level can be set to symmetric orall positive.

The amplitude (Vpp) setting range is: 0.01 Vto 12 V.

Next…

Pulse Generation on page 65Frequency Measurement on page 48Special Generations on page 69Calibration, see Part D.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,V, I,

Page 71: MC5 User Manual

65

Pulse GenerationThe pulse generation terminals are located in the ET module. Pulse generation is similarto square wave frequency output, but it generates a predefined count of all positive orsymmetric pulses.

Required settings Options/description

Quantity PulsesFunct/Port ET: Pls(gen)

Also check the Polarity setting.

Enter the amount of pulses CL560 shouldgenerate in the pulse generation window.Also enter the amplitude and frequencyto be used.

The pulse generation starts immediatelyafter the amount of pulses value is entered.

Note.

If you change the frequency after the pulse generation has started,the frequency is adjusted during the current cycle, but the frequencyis not correct until the next cycle.

Next…

Frequency Generation on page 64Pulse Counting on page 49Special Generations on page 69.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,V, I,

Generating/Simulating

Page 72: MC5 User Manual

66

Startup and Basic Operation

RTD and Resistance SimulationCL560 mimics the RTD or the resistor to be measured by the instrument under test. Theinstrument under test generates the current for the resistance measurement. CL560 con-trols the voltage across its terminals so that the resistance (voltage to current ratio) cor-responds to the simulated temperature or resistance. CL560 can simulate resistancebetween 1 to 4000 ohm.

RTD and resistance simulation terminals are located in the ET module.

Required settings Options/description

Quantity RTD-Temp. (for RTD simulation)Resistance (for resistance simulation)

Funct/Port ET: RTD(sim) (for RTD simulation)ET: R(sim) (for resistance simulation)

Sensor Type Available RTD sensors. Only neededwhen doing RTD simulation.

The correct resistance value is between theresistance simulation terminals of the cali-brator. Use of 2-, 3- or 4-wire connection isup to the receiver instrument. Use only thetwo leftmost R, RTD terminals with everywiring option. Connect the possible thirdand fourth wire according to the require-ments of the connected instrument, but useonly the two leftmost R, RTD terminals.

Notes.

In RTD and resistance simulation CL560monitors the resistance measurement cur-rent. If the current is too high, it cannotsimulate the right resistance value. In thatcase it shows the message “H.CURR”. Re-spectively, if the measurement current isso low that it may affect the accuracy, themessage “L.CURR” is displayed.

Accurate operation of the simulation elec-tronics requires that the current suppliedby the instrument under test does not varyrapidly. The simulation result is not accu-rate if the instrument under test uses ACcurrent. If the instrument under test usespulsed measurement current it should waita few milliseconds before starting the mea-surement after setting the current.

Next…

Thermocouple Measurement (Temperature) on page 51RTD Measurement (Temperature) on page 50Resistance Measurement on page 44Special Generations on page 69Calibration, see Part D.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

V, I,R, RTD

Page 73: MC5 User Manual

67

Thermocouple SimulationCL560 is able to mimic a thermocouple at given temperatures. This property is calledthermocouple simulation. It enables checking and calibration of temperature indicators,recorders, transmitters and other equipment related to temperature measurement with athermocouple probe.

In thermocouple simulation the original thermocouple is disconnected from the instru-ment under test and replaced with CL560. To the instrument under test CL560 appearsas a thermocouple at the given temperature.

Thermocouple simulation terminals are located in the ET module.

Required settings Options/description

Quantity T/C-TempFunct/Port ET: TCi(sim) (for the internal Reference Junction), or

ET: TCx(sim) (for other RJ compensation methods)Sensor Type Available thermocouple typesReference Junction Depends on the Function/Port setting.

See subsequent chapters.

Internal Reference Junction

CL560’s Internal Reference Junction Mod-ule is an optional module. To use the Inter-nal Reference Junction, select Function/Port ET: TCi(sim). The Reference Junctioncompensation method Internal is auto-matically selected.

External Reference Junction

To use an External Reference Junction,select Function/Port ET: TCx(sim) andchoose one of the available ReferenceJunction compensation methods: Entered,0°C or available RTD sensors.

Additional information is found in chapterInternal Reference Junction on page 97.

Additional information is found in chapterExternal Reference Junction on page 98.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,

T/C OR EXTWIRES ONLY

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

V, I,T/C, Low V

Generating/Simulating

Page 74: MC5 User Manual

68

Startup and Basic Operation

Notes.

If the resistance of the external circuit in thermocouple simulation isvery low (obvious short circuit), the message “O.LOAD” is displayeduntil the load is reasonable.

Check that the instrument under test is set for the same thermo-couple type as selected in CL560. If the type is not one of the avail-able types, you should simulate in millivolt mode and convert fromtemperature to millivolts manually. More information concerningmillivolt generation is in chapter Voltage Generation on page 62.

Wire-wound thermocouple temperature indicators and record-ers are calibrated with known loop resistance, commonly ten ortwenty ohms. The required resistance of the conductors is usuallymarked on the device. When calibrating such a device, tune theloop resistance to the nominal value with an additional resistor.

Use CL560’s 2-wire resistance measurement to tune the loop re-sistance of a wire wound device. Short circuit the measuring devicefor the time of resistance measurement. The resistance measure-ment may damage the device, if you do not install the short circuit.For more information on resistance measurement, see chapterResistance Measurement on page 44.

Warning!

If you short circuit the thermocouple simulation output, CL560tries to maintain the voltage by increasing the output current.If you then open the output again, the voltage is first too high,but returns quickly to the correct level. If this voltage peak coulddamage the components of the circuit, make sure that the cir-cuit can not be shorted or protect the circuit against overvolt-age. For the same reason, always enter 0°C output before con-necting the circuit.

If you connect an RTD sensor to the ET module’s R, RTD con-nectors, there is no galvanic isolation between the thermo-couple and the RTD sensor.

Next…

Thermocouple Measurement (Temperature) on page 51RTD and Resistance Simulation on page 66Special Generations on page 69Calibration, see Part D.

Problems with thermovoltage simulation? See Error situations on page 100.

Page 75: MC5 User Manual

69

Special Generations

Manual change of generated signals is sometimes a restriction.Therefore CL560 includes Stepping and Ramping as standard utili-ties for creating automatically alternating output signals (specialgenerations).

To be able to start the definition of a special generation, a suitablegeneration/simulation field has to be active in the Basic Mode Win-dow. If a generation/simulation is not started, select D/Menu andeither A/Window 1 Setup or B/Window 2 Setup to configure ageneration/simulation. For more information on how to configure ageneration/simulation, see chapter Generating/Simulating on page58 and its subchapters.

Notes.

Only one of the special generations may be active at a given time.

Special generations are available only in Basic Mode.

Opening the Step or Ramp Configuration WindowOpen the window setup menu for the generation/simulation quan-tity that you want to configure a Stepping function for. The menucommands are:

• D/Menu• and

B/Window 2 Setup,if needed,

• Press the numerickey to open thesecond menu page

• Choose3/Step, Ramp and

• Either 1/Step or2/Ramp in the nextmenu.

12.04.2002 12:52

ET: Voltage Generation

E: Switch Sense

V???????

Open

2 Switch

Measurement, V -------

1 Voltage

Opened @ VClosed @ V

--------------

12.04.2002 12:51

ET: Voltage Generation

E: Switch Sense

V???????

Open

2 Switch

Measurement, V -------

1 Voltage

Opened @ VClosed @ V

--------------

12.04.2002 12:51

ET: Voltage Generation

E: Switch Sense

V???????

Open

2 Switch

Measurement, V -------

1 Voltage

Opened @ VClosed @ V

--------------

Quantity[Voltage]

Function/Port[ET: V(gen)]

Unit[mA]

HART

SetupWindow 2

MENUClose

SetupWindow 1 Others

Display Mode[Eng. Unit]

Alarm[--(--)]

Function and Port Info

Extra Info

SetupWindow 2

MENUClose

SetupWindow 1 Others

Special Measurements[None]

Step, Ramp

Second Port Info

MENUClose

Step

Ramp

Back

Special Generations

Page 76: MC5 User Manual

70

Startup and Basic Operation

Stepping

Configure the Stepping settings accordingto your needs. The following pictures areexamples of stepping configurations. Thegraph below the configuration window dis-plays the result of the settings.

Note that because of the Starting pointsetting, the first repeat of the leftmost ex-ample is limited to a small starting step andonly two full steps. Generally: if the step-ping settings limit the first repeat to only afraction of a full repeat it is still calculatedas one repeat.

You may enter either the Step Size or theNumber of Steps. Because these valuesdepend on each other, the value of the otherfield is calculated whenever you make anychanges. Also, you can only enter an inte-ger as the Number of Steps value. If you

enter a step size that results in a non-inte-ger value for the Number of Steps field,CL560 displays a dashed line instead of avalue for the Number of Steps field. Thisindicates that the size of the step thatreaches the 100 % range limit is smallerthan the defined step size. See therightmost example.

To stop stepping press the function keyC/Stop Stepping. This applies for both acontinuous stepping (Repeats field set tozero) and a stepping with a predeterminedamount of repeats (max. value 65535).

22.09.2000 14:22

StopRamping

Field

STEPPING

QuantityPort

Stepping Mode

Step TimeRepeats

Starting Point

Step SizeNumber of StepsDivision of Steps

Range

CurrentET:I(gen)

Up

1.00

3From present

1.004Linear

8.000012.0000

s

mA

mA0 %

100 %

StartCancel

0 = continuous

Edit

STEPPING

QuantityPort

Stepping Mode

Step TimeRepeats

Starting Point

Step SizeNumber of StepsDivision of Steps

Range

CurrentET:I(gen)

Up/Down

0.50

0From 0 %

1.50-------Linear

8.000012.0000

s

mA

mA0 %

100 %

22.09.2000 14:39

0 = continuous

StopRamping

Field StartCancel Edit

Current, mA

Time, s

8

9

10

11

12

10-1 5 10 14

Start

Current, mA

Time, s

8

9

10

11

12

10-1 5 10 14

Start

The Stepping configuration windows shownbelow can are opened as described inchapter Opening the Step or Ramp Con-figuration Windowon page 69.

Page 77: MC5 User Manual

71

Warning!

Do not configure the range settings so that they exceed theallowed input range of the connected instrument. CL560 deter-mines the limits of the range settings based on the selectedquantity and port, not the connected instrument.

Ramping

The Ramping configuration windows shownbelow can are opened as described inchapter Opening the Step or Ramp Con-figuration Windowon page 69.

Configure the ramping settings accordingto your needs. The following pictures areexamples of ramping configurations. Thegraphs presented after the configurationwindows display the result of the settings.

zero) and a ramping with a predeterminedamount of repeats (max. value 65535).

To stop ramping press the function keyC/Stop Ramping. This applies for both acontinuous ramping (Repeats field set to

22.09.2000 10:52

CurrentET:I(gen)

500

0

8.000016.0000

s

mA0 %

100 %

0 s

ss

RAMPING

QuantityPort

Rise TimeWait in 100%Fall Time

Repeats0=continuous

Range

Wait in 0%

StopRamping

Field StartCancel Edit

RAMPING22.09.2000 10:43

QuantityPort

Rise TimeWait in 100%Fall Time

Repeats0=continuous

Range

CurrentET:I(gen)

112

2

8.000016.0000

s

mA0 %

100 %

2Wait in 0% s

ss

StopRamping

Field StartCancel Edit

Current, mA

Time, s

8

10

12

14

16

10-1 5 10 14

Start

Current, mA

Time, s10-1 5 10 14

Start

8

10

12

14

16

Special Generations

Page 78: MC5 User Manual

72

Startup and Basic Operation

Note.

CL560 actually makes the ramp in small steps. The steps are assmall as possible, slower ramps use smaller steps.

Warning!

Do not configure the range settings so that they exceed theallowed input range of the connected instrument. CL560 deter-mines the limits of the range settings based on the selectedquantity and port, not the connected instrument.

Page 79: MC5 User Manual

73

Alarm Limit Settings

Each main measurement in a window may have alarm limits set-tings. CL560 supports “higher than”, “lower than”, “high rate” and“low rate” alarms.

To set the alarm limits, open the window settings menu of the win-dow where the measurement is active (starting from the Basic Mode:D/Menu followed by A/Window 1 Setup or B/Window 2 Setup ifneeded). As the alarm setting command is located in the windowsetup’s submenu, select the numerical key to open the submenu.

Open the alarmsetting windowby selecting2/Alarm.

mbar(gauge)

Alarm <

0.5 bar

ControllerSettings

CloseMENUMode

OthersWindow 1 Window 2Setup Setup

Sensor Supply[None]

mbar(gauge)

Alarm <

0.5 bar

SpecialMeasurement[None]

[--(--)]

Function and Port Info

Alarm

Second

Extra Info

Port Info

CloseMENUMode

OthersWindow 1 Window 2Setup Setup

[P2: EXT100]

Second port

CloseMENUMode

Others

0.70834 mbar(gauge)

Alarm <P1: INT2C / -1.000000...2 bar

0.5

0.4374

2 Current

bar

Pressure

E: Current Measurement

Quantity[Pressure]

[P1: INT2C]

[Eng. Units]

Unit[bar]

Funct/Port

Display Mode

07.05.2002 12:53 1

HART

PressureType[gauge]Zero PressureModule

Window 1 Window 2Setup Setup

22.09.2000 15:21 ALARM SETTINGS

QuantityPort

Alarms activated

Alarm, when the Measurement

and/or

RTD-temperatureET: RTD(mea)

2419

<>

°C

Alarm, when the Rate of Change

and/or???????1

<>

1/h

Yes

CancelLimitClear OKEdit

Alarm Limit Settings

Page 80: MC5 User Manual

74

Startup and Basic Operation

The Alarms activated field makes it possible to deactivate thealarm limits without clearing the limits.

Set the limits as required. Individual alarm limits may be disabledby clearing the alarm limits using the B/Clear Limit Function Key.The “<“ (lower than) limit may have a higher value than the “>” (higherthan) limit. In that case CL560 emits an alarm when the measure-ment is inside the range specified by the alarm limits.

Note.

Remember to set/check also the rate alarm unit if you set any ratealarms.

Appearance of the alarm settings.

The alarm settings can be seen as symbols in the measuring win-dow and in the window setup menu as follows:

Symbol Definition

[ - - ( - - )] No limit settings[< - ( - - )] Low limit is set[ - > ( - - )] High limit is set[< > ( - - )] Low and high limits are set[> < ( - - )] Low and high limits are set. Low > High[ - - ( < - )] Low rate limit is set[ - - ( - > )] High rate limit is set[ - - ( < > )] Low and high rate limits are set[ - - ( > < )] Low and high rate limits are set. Low > High[< > ( < > )] All limits are set.

Acknowledging alarms

When an alarm limit is exceeded, CL560 emits an audible alarm.The alarm can be acknowledged from all other keys except thefollowing keys: , and or any key combinations includingone or more of the previously listed keys. The active alarm limit isunderlined in the measurement window as long as the alarm limit isexceeded.

High alarm limit >

Low alarm limit:

High alarm limit <

Low alarm limit:

ALARM

ALARM

No alarm

No alarm

No alarm

Low alarm limitHigh alarm limit

Low alarm limit

High alarm limit ALARM

Mea

sure

d va

lue

Mea

sure

d va

lue

Page 81: MC5 User Manual

Advanced Operationand Configurations

Things discussed in Part C:

• How to configure CL560 to suityour requirements.

• Advanced Utilities available inBasic Mode.

• Additional information: Usefulduring pressure measurement,thermocouple measurement/simulation and resistance/RTDmeasurement/simulation.

Page 82: MC5 User Manual

76

Advanced Operation and Configuration

Configuring the Calibrator

A group of settings are available in CL560’s Others menu’s Main-tenance option. The following chapters tell how you can changethem and what the default settings are.

SettingsThe settings window has two pages. To open the settings windowstarting from Basic Mode, press D/Menu, C/Others and 1/Set-tings. The leftmost picture is the settings page that opens first.

You can scroll between the pages using the B/Next Page FunctionKey.

Note.

All changes in settings become valid immediately.

25.09.2000 8:53 SETTINGS

Date FormatTime Format

Language

Auto-off Delays (0=never) [min]CalibratorBacklight

Sound VolumesKey ClickAlarmAttentionOutside SpanError

h:mm 24h

English

05

LowHighMediumMediumLow

Close

dd.mm.yyyy

PageNext Edit

25.09.2000 8:56

Net FrequencyBacklight Power Normal

50 Hz

ENV-sensor UsageTemperature UnitTemperature Scale

Both°CITS90

SETTINGS

ClosePageNext Edit

Page 83: MC5 User Manual

77

Date Format

A drop down list of available date formats:dd.mm.yyyy, yyyy.mm.dd and mm.dd.yyyy.

The default format is “dd.mm.yyyy”.

Time Format

A drop down list of available time formats.24 hour format: h:mm12 hour format: h:mm am/pm.

The default setting is the 24 hour format.

LanguageA drop down list of available User Interfacelanguages. The default language is En-glish.

Auto-Off Delay for the CalibratorWait time before auto power off is executed.

The default setting is “0” (never).

Auto-Off Delay for the Display Back-light

Wait time before the display light is auto-matically switched off.

The default setting is “5 min”.

Setting the time to zero disables displaybacklight auto-off.

Volume Settings

CL560 has individual volume settings forthe following sounds:

Sound Default value

*Key Click Medium*Alarm Medium*Attention Medium Outside Span Medium Error Medium

All settings have three different possiblevalues (Low, Medium, High). The onesstarting with an asterisk (*) may also be setto “Off”.

Line Frequency

Select the setting that suits the local linefrequency (50 or 60 Hz).

Note that wrong line frequency setting af-fects the accuracy of CL560.

The default setting is “50 Hz”.

Display Backlight Power

Select the backlight power from the avail-able options:

• Economy• Normal• Efficient

The default setting is “Economy”.

ENV Sensor UsageHow the Environment Sensor (ENV) shouldbe used during calibration. Select one ofthe available options:

• Not Used• Environment• Instrument• Both

The default setting is “Both”.

Temperature Unit

Select either °C (Centigrade) or °F (Fahr-enheit) as the temperature unit.

The default setting is “°C”.

Temperature ScaleCL560 supports the use of both interna-tional temperature scales: ITS90 and theolder IPTS68. Select which one of them isin use.

The default setting is “ITS90”.

Configuring the Calibrator

Page 84: MC5 User Manual

78

Advanced Operation and Configuration

Setting Time and DateTo change the time and date (starting from Basic Mode), pressD/Menu, C/Others, 2/Maintenance and 2/Time/Date.

Notes.

The date must always be given in dd.mm.yyyy format no matterwhat the configured date format is.

The time must always be given in the 24 hour format no matterwhat the configured time format is.

The date and time will be updated when you press the D/AcceptFunction Key.

13.06.2002 16:02 TIME/DATE SETTINGS

Date (dd.mm.yyyy)

Time (hh:mm:ss) 16

13

02

06

22

2002

Cancel AcceptEdit

Page 85: MC5 User Manual

79

Advanced Utilities

All the subsequent chapters:• Measuring the Environment Temperature with the ENV

Sensor on page 79,• Display Mode Settings on page 80,• Transmitter/Switch Simulation on page 83,• Data Logging on page 86,• Generating Signals Using External Devices on page 89

and• Printing on page 93

describe the use of additional features that are available in CL560.Some of the utilities are firmware options and/or they require someadditional hardware before they can be utilized.

Measuring the Environment Temperature with the ENV SensorThe ENV-sensor is a optional environment temperature probe that can be connected tothe ENV sensor interface on the left side of CL560.

Required settings Options/description

Quantity ENV-Temp.Funct/Port T-ENV(mea) (selected automatically when the

corresponding Quantity setting is activated )

Note!

The ENV-sensor is not accurate enough to be used as a refer-ence sensor. Use it as a tool for obtaining the environment tem-perature during calibration or as a temperature measurement toolwhen the accuracy of the measurement is not critical.

Advanced Utilities

Page 86: MC5 User Manual

80

Advanced Operation and Configuration

Display Mode SettingsBy default, CL560 displays all mea-surement, generation and simula-tion data in engineering units. InBasic Mode you can also selectsome special Display Modes. Tochange the Display Mode, selectthe Window setup menu (Com-mands: D/Menu and B/Window 2Setup, if needed) and choose3/Display Mode.

All of the available special DisplayModes are described in the subse-quent chapters.

Note.

All special Display Modes are reset to Engineering Units DisplayMode if the quantity or port of the window(s) assigned to the specialDisplay Mode are changed.

Warning!

Be careful not to mistakenly read a value shown in special Dis-play Mode as a value shown in Engineering Units. CL560 al-ways displays the true measurement in the window’s lower-most row (Extra Info row) when a special Display Mode is ac-tive.

14.06.2002 8:29

P2: INT20C/-1.0 ... 20.7 bar g

E: Current Measurement

gauge bar1.0357

8.3524 mA

2 Current

1 Pressure Quantity[Pressure]

Function/Port[P2: INT20C]

Unit[bar]

HART

SetupWindow 2

MENUClose

SetupWindow 1 Others

Display Mode[Eng. Units]

Pressure Type[gauge]Zero PressureModule

Eng. UnitsScalingPercentageError Value

Page 87: MC5 User Manual

81

Scaling

In scaling, the measured/generated/simulated value is displayed inanother, user defined quantity/unit. Start scaling as described inchapter Display Mode Settings on page 80.

When selecting Scaling Display Mode, the following configurationwindow appears:

SCALING26.04.2002 14:02

QuantityPort

PressureP1:INT2C

Measured Range

kPa

StopRamping

Field StartCancel Edit

0 %100 %

kPa0.000100.000

Scaled ModeScaled Unit mAScaled Range

0 %100 %

4.00020.000

Transfer Function Linear

Current

SCALING30.05.2002 10:09

QuantityPort

PressureP1:INT2C

Measured Range

kPa

StopRamping

Field StartCancel Edit

0 %100 %

kPa0.000100.000

Scaled Mode Custom UnitsScaled UnitScaled Range

0 %100 %

1.005.00

Transfer Function Linear

Gallons/min

Decimals 3

You can scale the quantity either to a quantity that already exists inCL560 or select Custom Units from the Scale Mode field. Thenyou can manually enter any unit to the Scale Unit field (see therightmost picture above).

When scaling is active in a BasicMode window, the window containsthe text “SCALED”. The true valueis displayed in Extra Info row. If theExtra Info row was assigned tosome other measurement beforescaling was started, the other mea-surement is automatically stopped.

Note.

Remember to set the amount ofdecimals to be displayed for aCustom Unit.The default value is 3.

30.05.2002 10:10

P2: INT20C/-1.0 ... 20.7 bar g

E: Current Measurement

Gallons/min20.088

8.3217 mA

2 Current

1 Pressure

Measurement bar 1.004

ModeCalibration MENU

SCALED

Advanced Utilities

Page 88: MC5 User Manual

82

Advanced Operation and Configuration

Displaying Values in Percentage

In Percentage Display Mode themeasured/generated/simulatedvalues are displayed as percent-age of a user defined range. Startthe Percentage Display Mode asdescribed in chapter DisplayMode Settings on page 80 andenter the Measured Range.

The true value is displayed in theExtra Info row.

PERCENTAGE26.04.2002 14:21

QuantityPort

PressureP1:INT2C

Measured Range

kPa

StopRamping

Field StartCancel Edit

0 %100 %

kPa0.000100.000

ERROR DISPLAY26.04.2002 14:29

INPUTPort

PressureP1:INT2C

Measured Range

kPa

StopRamping

Field StartCancel Edit

0 %100 %

kPa0.000100.000

Transfer Function Linear

Error Calc. Method % of span

OUTPUTPort

CurrentE: I(meas)

Measured Range

mA0 %

100 %mA4.000

20.000

Window

Window

1

2

Displaying Error Values

The Error Display Mode utilizesboth of Basic Mode’s windows. Themeasured/generated/simulatedvalues of both windows are treatedas percentages of user definedranges. The input value is consid-ered as the “correct” value and theoutput value should follow the in-put value, while taking the Trans-fer Function into account.

The window from which Error Dis-play Mode is invoked is consideredas the instrument’s output signaland the other window as the inputsignal.

Start the Error Display Mode as described in chapter Display ModeSettings on page 80. Enter the Measured Range for both the in-put and the output signal, the Error Calculation Method and theTransfer Function (input/output correlation).

The true output value is displayed in the Extra Info row.

Note.

The Error Display Mode is only available for a Basic Mode windowset for measuring a quantity.

Page 89: MC5 User Manual

83

Transmitter/Switch SimulationCL560’s Basic Mode includes apossibility to simulate a transmit-ter and a switch. To start a Trans-mitter or as Switch simulation, con-figure Window 1 to measure a sig-nal (transmitter input) and and Win-dow 2 to either generate or simu-late a signal (transmitter output),then select D/Menu, C/Others and6/Transmitter/Switch Simula-tion. A pop-up list appears whereyou can select whether you wantto simulate a transmitter or aswitch.

Note.

To be able to start the Transmitter/Switch Simulation, both Ba-sic Mode windows need to be configured accordingly. If, e.g.window 2 is not assigned for generating/simulating a signal,CL560 can not start transmitter simulation.

18.04.2002 15:15

P2: INT20C/-100 ... 2068 kPa g

E: Current Source/Sink

gauge kPa0.02

mA

2 Current

1 Pressure

End MENU

Measurement mA 4.0000

Pause

4.0000

SetupWindow 2

MENUClose

SetupWindow 1 Others

Data Logging

Settings

Maintenance

HART

Settings

Transmitter/SwitchSimulation

Communication

Transmitter FunctionSwitch Function

Advanced Utilities

Page 90: MC5 User Manual

84

Advanced Operation and Configuration

Transmitter Simulation

Start the Transmitter Simulation as described in chapter Transmit-ter/Switch Simulation on page 83. If the settings of the Basic Modewindows suit Transmitter Simulation, a configuration window simi-lar to the lower left picture is shown.

TRANSMITTER SIMULATION26.04.2002 14:30

INPUTPort

PressureP1:INT2C

kPa

StopRamping

Field StartCancel Edit

100 %kPa0.000

100.000

Transfer Function

0 %

OUTPUTPort

CurrentE: I(control)

mA100 %mA4.0000

20.0000 %

Saturation Limit Output Values

mAMaximummA3.5000

23.000Minimum

Linear

30.05.2002 10:31

P2: INT20C/ (0.00 ... 100.00 kPa g)

E: I(control)/ (4.0000 ... 20.000 mA)

gauge kPa0.06

4.0000 mA

2 OUTPUT

1 INPUT

End MENU

Measurement mA 4.0000

Pause

TRANSMITTER SIMULATION

Enter the input and output spans, the saturation limits and also thetransfer function.

If the saturation limits differ from the range limits, CL560 extrapo-lates the output value based on the input value and the TransferFunction until the saturation limit is reached. Then if the input signaldrifts farther from the input range, the output value stays at the satu-ration limit.

The upper right picture shows the Basic Mode window while Trans-mitter Simulation is active. The second row of both windows displaythe active port and the input/output range of the simulated transmit-ter.

Notes.

To Zero a pressure module during Transmitter Simulation, pressD/MENU and 7/Zero Pressure Module.

When pausing the Transmitter Simulation, you can change the gen-eration/simulation signal as in Basic Mode.

Warning!

Remember to scale the output signal of the Transmitter Simu-lation so that the instrument connected to the output signalloop is not damaged.

Page 91: MC5 User Manual

85

Switch Simulation

Start the Switch Simulation as described in chapter Transmitter/Switch Simulation on page 83. If the settings of the Basic Modewindows suit Switch Simulation, a configuration window similar tothe lower left picture is shown.

SWITCH SIMULATION26.04.2002 14:38

StopRamping

Field StartCancel Edit

INPUTPort

PressureP1:INT2C

kPakPa50.000

40.000

OUTPUTPort

VoltageET: V(gen)

VV5.0000

0.5000DeactuatedActuated

DeactuatingActuating

Nominal Points

Switch Limits

30.05.2002 10:32

P2: INT20C/ (50.00/40.00 kPa g)

ET: V(gen)/ (5.000/0.500 V)

gauge kPa52.034

5.0010

2 OUTPUT

1 INPUT

End MENU

Measurement V 5.0011

Pause

SWITCH SIMULATION

Enter the actuating point and the deactuating point of the input sig-nal. Also enter the output signal levels for an actuated output and adeactuated output.

The upper right picture shows the Basic Mode window while SwitchSimulation is active. The second row of the upper window displaysthe active port and the actuating and deactuating points. The sec-ond row of the lower window displays the actuated and deactuatedoutput of the switch.

Note.

To Zero a pressure module during Switch Simulation, Press D/MENUand 7/Zero Pressure Module.

Warning!

Remember to scale the output signal of the Switch Simulationso that the instrument connected to the output signal loop isnot damaged.

Advanced Utilities

Page 92: MC5 User Manual

86

Advanced Operation and Configuration

Data Logging

General

The Data Logging utility is a firmware option that allows you to col-lect data using CL560. The collected data may then be viewed,transferred to a personal computer (PC) and printed using a utilityshipped together with CL560 if the Data Logging option is purchased.

The maximum amount of measuring channels can be up to seven(valid only if your CL560 includes a sufficient amount of measuringmodules/ports).

Data Logging may be configured to collect data using different trig-gers and saving methods as described further on.

Configuring

To start configuring the Data Log-ging (starting from Basic Mode),press

D/Menu,C/Others,5/Data Logging.

The upper part of the Data Loggingconfiguration window is reservedfor seven measurement channelsand the lower part for general set-tings. The following subchaptersdescribe how they are configured.

Assigning Measurement Ports to Channels

Move the cursor to one of the meas-urement channel rows in the DataLogging configuration window.Press C/Edit to open the followingchannel settings window: Select themeasurement port first. The othersettings depend on the port setting.

To clear a channel, select port Nonein the channel settings window. Toremove all channel settings, selectD/Menu, 1/Clear All Channels inthe Data Logging configuration win-dow.

20.08.2002 14:57 DATA LOGGING SETUP

1

Trigger

None2 None3 None4 None5 None6 None7 None

Save Method

Interval [s]Samples

Time [h min s]

PeriodicInstant Value

210

0

ModeBasic

LoggingData Edit MENU

0 20

20.08.2002 15:01 DATA LOGGING SETUP

Channel

PortPressure Type

Unit

1

Abs

bar

P3:B

Cancel OKEdit

Page 93: MC5 User Manual

87

General Data Logging Settings

Trigger

Available options: Periodic andKeyboard.When Keyboard is selected, thefields Interval, Samples and Timeare not needed and thus not shown.

Save Method

Options: Instant, Average, Mini-mum, Maximum, Min & Max andAll values.

Interval

Accepted values: 1 to 3600 sec-onds. Enter the value as an integer.

Samples

Maximum amount of samples: 70000.

Time

Automatically calculated based on the Interval and Samples set-tings. Alternatively, by changing the Time settings, the Samplesvalue is automatically changed to suit the new Time settings.

Note.

In addition to the previously mentioned limits, the maximum valuesof Samples and Time settings also depend on the available freememory. If you enter values that require more memory than is avail-able, CL560 automatically replaces the entered value with the cur-rently allowed maximum value.

When a data log is done, you canenter a name for the results.

Starting the Data Log

To start logging, select B/Data Log-ging in the Data Logging configu-ration window. Start the Logging byselecting C/Start Logging in theLogging window.

20.08.2002 15:02 DATA LOGGING SETUP

1

Trigger

2 ET:RTD(mea) °C Pt100 α3853 ET:TCi(mea) °C K NiCr/niAl4 None5 None6 None7 None

Save Method

Interval [s]Samples

Time [h min s]

PeriodicInstant Value

60720

12

ModeBasic

LoggingData Edit MENU

P3:B bar abs

0 0

Absbar1.0113

20.08.2002 15:32 DATA LOGGING

Pt100 α385°C24.17K NiCr/NiAl°C24.39Posit. kanttikHz-------Posit. kanttikHz-------Posit. kanttikHz-------Posit. kanttikHz-------

1

2

3

4

5

6

7

Count 647 38802 s

LoggingStop MENU Back

Advanced Utilities

Page 94: MC5 User Manual

88

Advanced Operation and Configuration

If the Trigger setting is Periodic, you can follow how the loggingadvances in the lower part of the window. Both the amount ofsamples to be logged and the remaining time is displayed.

If the Trigger setting is Keyboard, accept the values by pressingthe Function Key B/Accept Readings. .

Now the lower part of the window displays the amount samplesalready logged and the time since the logging was started

To stop/interrupt the logging, be it periodically triggered or triggeredfrom the keyboard, select C/Stop Logging.

Viewing the Results

Opening the Log Results windowcan be done both from the Configu-ration window and the Logging win-dow. Select D/MENU 2/ViewDatalog Results.

If you have logged using more thantwo channels, use the and cursor keys to scroll through all thecolumns. Alternatively: use the nu-meric keys to quickly jump to a col-umn.

Removing Datalog results

To remove a set of data loggingresults, select D/MENU 3/Remove Datalog Data.

To remove all datalog results, select D/MENU 4/Remove AllDatalog Data.

Transferring the Results to a Personal Computer

A 32-bit Windows® software called Datalog.exe is shipped togetherwith CL560 if you bought the Data Logging option. Start this soft-ware just as any other Windows® software.

All communication between the PC and CL560 is initiated fromDatalog.exe. More information of the software in Appendix 3.

20.08.2002 13:19

Back MENU

Just Testing - All values

Time

14:19:38

14:20:3814:21:3814:22:3814:23:3814:24:3814:25:3814:26:3814:27:3814:28:3814:29:3814:30:3814:31:3814:32:38

1: Instant

[bar]

1.01131.01131.01121.01121.01121.01111.01111.01111.01111.01101.01101.01101.0109

1: Average

[bar]

1.01131.01131.01131.01121.01121.01121.01111.01111.01111.01111.01101.01101.0109

1 2

04.09.2000 Pressure Pressure

Page 95: MC5 User Manual

89

Generating Signals Using External DevicesTo be able to utilize devices connected to the auxiliary instrumentinterface (AUX) you need to have the required firmware option. Thefollowing paragraph explains how to activate CL560 to use a con-nected external device. The instructions are also valid when theconnected device is changed.

To select an external device, open either Window 1 or Window 2setup menu (select D/Menu and either A/Window 1 Setup orB/Window 2 Setup in Basic Mode). Make sure the quantity settingof the window corresponds to the external device’s quantity require-ments. Continue to the third submenu by pressing the numericalkey twice. The following picture displays all the submenus avail-able when the selected Quantity is Pressure.

To be able to communicate with the external device, CL560 requiressome additional data which you can enter in the window that opensautomatically after you have selected the Controller Settings menuoption. This is presented on the next page.

mbar(gauge)

Alarm <

0.5 bar

ControllerSettings

CloseMENUMode

OthersWindow 1 Window 2Setup Setup

Sensor Supply[None]

mbar(gauge)

Alarm <

0.5 bar

SpecialMeasurement[None]

[--(--)]

Function and Port Info

Alarm

Second

Extra Info

Port Info

CloseMENUMode

OthersWindow 1 Window 2Setup Setup

[P2: EXT100]

Second port

CloseMENUMode

Others

0.70834 mbar(gauge)

Alarm <P1: INT2C / -1.000000...2 bar

0.5

0.4374

2 Current

bar

Pressure

E: Current Measurement

Quantity[Pressure]

[P1: INT2C]

[Eng. Unit]

Unit[bar]

Funct/Port

Display Mode

06.05.2002 15:32 1

HART

PressureType[gauge]Zero PressureModule

Window 1 Window 2Setup Setup

Advanced Utilities

Page 96: MC5 User Manual

90

Advanced Operation and Configuration

Select the Controller Type. Thelist of available controllers variesdepending on installed options andthe selected quantity. The othersettings vary depending on theController Type setting.

Note.

The Settings window for externaldevices may also be opened fromCalibration Mode’s Instrument win-dow. Additional information on howto use external devices in Calibra-tion Mode is in Part D chapter Us-ing External Devices in Calibration Mode on page 118.

Controlling the External Device in Basic Mode

When an external device is configured for use, the device name isdisplayed in the window it was assigned to, e.g. POC4 in the adja-cent picture.

Enter target values (setpoints) forthe connected external device inthe numerical field following thedevice name.

Note.

If a field indicator cannot be movedto the external device’s setpointfield (or the field indicator is notseen at all), make sure the connec-tions and the settings are in order.When CL560 is able to communi-cate with the connected device, acommunication symbol is shown inthe status bar’s rightmost end (seeadjacent picture).

25.09.2000 9:45 CONTROLLER SETTINGS

Controller Type

Cancel OK

POC4Control ModeDevice Address

Shut Off0

Serial Communication Settings

9600,N,8,1

Edit

25.09.2000 9:48 1 Pressure

2 Current

P2:INT20C/-1.0000...20.6840 bar g

E: Current Measurement

gaugebar0.0003

4.0073

ModeCalibration MENU

mA

POC4 0.00

Page 97: MC5 User Manual

91

Pressure Controller Settings

The following pressure controllers were supported when this UserGuide was written:

• Beamex POC4 and • Druck DPI510 equipped with the RS232 communication op-

tion (only gauge pressure type models). • Druck DPI 515

The following settings apply in general to all supported pressurecontrollers. If certain controllers have device specific data, it is men-tioned in the general description.

Control Mode

Possible settings: Shut Off and Continu-ous.

Shut Off means that the Pressure Control-ler shuts down after the desired pressurelevel is reached. When Continuous is se-lected, the Pressure Controller is active allthe time. Even after the setpoint value isreached.

The first method is to be preferred becauseit minimizes the amount of disturbances inthe system, provided there are no notablepressure leaks in the system.

Notes.

Although the pressure controller does measure the generated pres-sure, it is not passed on to CL560 in CL560’s Basic Mode. There-fore you should always connect the generated pressure also to apressure measurement module in CL560.

When a controller is used in Calibration Mode, Input Mode “Con-trol” is used when the instrument’s input signal is measured withCL560 and Input Mode “Ctrl/Mea” when the controller communi-cates the generated pressure to CL560.

Zero the pressure controller using its own user interface or pro-gram it to zero itself every time it generates zero pressure.

Device Address

Device address is needed if the PressureController is in addressed mode. Ad-dressed mode also allows daisy-chain con-necting several Pressure Controllers.

Serial communication settings

The communication settings shown in thewindow are the ones required by CL560.The settings vary depending on the con-nected pressure controller (usually the de-fault communication settings of the pres-sure controller). If the settings have beenchanged afterwards, refer to the pressurecontroller’s manual on how to change thecommunication settings.

Advanced Utilities

Page 98: MC5 User Manual

92

Advanced Operation and Configuration

Temperature Controller Settings

The following temperature controller manufacturers have modelsthat communicate with CL560 (the list was valid when this UserGuide was written):

• TekKnow, • Isotech,• Ametek/Jofra SE, ATC, CTC and ITC models and • HART Scientific, most Industrial Models.

When in doubt whether your controller type or model communi-cates with CL560, contact Omega’s support.

The following settings apply in general to all supported temperaturecontrollers. If certain controllers have device specific data, it is men-tioned in the general description.

Device AddressThe Device Address of TekKnow TC1200and Isotech is either 1 or 2, depending onthe device configuration.

The Device Address field is not visible fortemperature controllers that do not needthe setting.

Serial communication settingsThe communication settings shown in thewindow are the ones required by CL560.The settings vary depending on the con-nected temperature controller (usually thedefault communication settings of the tem-perature controller). If the settings havebeen changed afterwards, refer to the tem-perature controller’s manual on how tochange the communication settings.

Note.

Although the temperature controller does measure the tempera-ture of the bath/dry block it is not passed on to CL560 in CL560’sBasic Mode. Therefore it may be of use to also connect a tempera-ture sensor to a suitable port in CL560.

When a controller is used in Calibration Mode, Input Mode “Con-trol” is used when the instrument’s input signal is measured withCL560 and Input Mode “Ctrl/Mea” when the controller communi-cates the temperature to CL560.

Page 99: MC5 User Manual

93

PrintingWhen using the optional printer you may print out screenshots ofany user interface screen (e.g. calibrations results) in CL560.

You can print the screenshot by pressing the key and the key simultaneously. CL560 responds with a long beep to indicatethat the screen is dumped to the printer.

The printer can not be used at the same time as CL560 is con-nected to a computer because the printer and the computer usesthe same interface (the COMP/PRT connector).

Notes.

If you press the and keys and no printer is connected, theserial port is unavailable for PC communication for about a minute.

The printer is configured for communication with CL560 when de-livered. If you inadvertently alter the configuration, please contactOmega for instructions on how to restore the settings.

Warning!

Only use the printer provided by Omega. Using any other print-ers may damage the printer or CL560 or even both of them.

Use only cables provided by Omega when connecting CL560to a PC or a printer.

Advanced Utilities

Page 100: MC5 User Manual

94

Advanced Operation and Configuration

Additional Information

Most of the measurements, generations and simulations presentedin this manual are straight forward: Just make the required windowsettings and connect the instrument under test in the active termi-nals and that’s it!

Some cases require additional settings and things to check beforeyou can be certain that the measurement, generation or simulationworks as expected. A typical example of this is temperature mea-surement using a thermocouple. It is not enough to select the cor-rect quantity and port in CL560. The Sensor type and the Refer-ence Junction Mode has to be set accordingly, too. Wrong settingsgive erroneous results.

Whenever this additional information may be of use in a measure-ment, generation or simulation, text describing the function refersto one of the following chapters. An experienced user need not jumpto read this supplemental text, but for a beginner we highly recom-mend it.

The main subjects described here are:

Things to Consider when Measuring Pressure on page 95,Thermocouple Measurement/Simulation, Connections andTroubleshooting on page 97,Resistance and RTD Measurement, Connections on page 101Current Measurement Parallel to a Test Diode, Connectionson page 103 andParallel Functions in CL560 on page 104.

Page 101: MC5 User Manual

95

Things to Consider when Measuring Pressure

General

CL560 can measure pressure with its internal pressure modules, orwith external pressure modules. If a barometric module is present,the measurement results of other modules can be shown either asabsolute pressure or as gauge pressure.

Pressure Type

Every pressure measurement is in fact a pressure difference mea-surement: a given pressure value is compared against a referencepressure. In certain cases the reference pressure has a specialmeaning (like the atmospheric pressure) and the pressure mea-surements with reference to against these points are given a spe-cial name:

• Absolute pressure measurement uses vacuum as the refer-ence pressure.

• Barometric pressure (or atmospheric pressure) is the ab-solute pressure of the surrounding atmosphere. It is speciallytitled because of its importance.

• Gauge pressure measurement uses the atmospheric pres-sure as the reference. Gauge pressure may be zero, negativeor positive. Most of CL560’s pressure modules actually mea-sure gauge pressure although they may be used for measur-ing other pressure types too.

• Differential pressure measurement requires a special dif-ferential pressure module that has an input available for boththe reference pressure and the given pressure.

To be able to measure absolute pressure with CL560’s gauge pres-sure modules, you need a barometric pressure module. Certainhigh pressure modules may approximate absolute pressure mea-surement by adding an entered barometric pressure to the mea-sured gauge pressure.

CL560 may use two standard pressure modules to measure differ-ential pressure: One module measures the reference pressure andanother the given pressure. The pressure difference is calculatedby subtracting the reference pressure from the given pressure. Moreof this in Part B, chapter Special Measurements.

Always ensure that the pressure type setting of your pressuremeasurement is what you intended it to be. The pressure typesetting is crucial for getting meaningful pressure measurementresults.

Additional Information

Page 102: MC5 User Manual

96

Advanced Operation and Configuration

Pressure Modules and their Naming Conventions

CL560’s software refer to pressure modules as follows:Port: Module Name, e.g. P1: INT400m

which means that there is an internal pressure module in port P1with a 0 to 400 mbar (0 to 6 psi) pressure range.

The names of internal pressure modules start with the letters INT.The names also include numbers and possible additional letters asfollows:

1. The number defines the max. nominal pressure in the SI pres-sure unit bar.

2. If the number is followed by a small “m”, the given max. nomi-nal pressure is given in mbar.

3. If there is an additional letter C, the pressure module is a com-pound module with an ability to also measure negative gaugepressures.

Examples: Name Max. nominal pressure in psi

INT100m 100 mbar 1.5INT400mC ± 400 mbar ± 6INT2C -1 … 2 bar -14.5 … 30

Notes.

Compound modules with a max. nominal pressure above 1 bar (e.g.INT2C) are able to measure negative pressures down to -1 bar.Compound modules with a max. nominal pressure below 1 bar (e.g.INT400mC) are only able to measure as deep negative pressuresas positive pressures.

The name of the internal barometric pressure module is B.

Eternal pressure modules have a name that starts with the lettersEXT. The name includes numbers and possibly an additional lettersimilar to internal modules (does not apply to old XPMs).

Square Rooting

Pressure transmitters or converters used for flow measurement of-ten have square rooting character. To enable correct percentageerror calculation of square rooting transmitters, select “square root-ing” as the Transfer Function.

Square rooting is required when entering instrument data for a pres-sure instrument (see Part D, Calibration). Standard pressure mea-surement in Basic Mode does not require any Transfer Functioninformation.

Page 103: MC5 User Manual

97

Thermocouple Measurement/Simulation, Connections andTroubleshooting

To accurately measure the thermovoltage caused by the tempera-ture to be measured, the second thermovoltage caused by the Ref-erence Junction needs to be compensated. This is done using oneof the Reference Junction compensation methods described in thesubsequent chapters.

The Reference Junction compensation method has to be chosenboth when measuring and simulating thermocouples.

Internal Reference Junction

CL560’s Internal Reference Junction Mod-ule is an optional module. To be able to uti-lize Internal Reference Junction compen-sation this optional module has to be in-stalled into your CL560. If the measuring/simulating port is set to the Internal Refer-ence Junction Module (ET: TCi(mea) orET: TCi(sim)), CL560 automatically selectsthe Internal Reference Junction compen-sation method. No other Reference Junc-tion compensation methods are availableunless the measuring/simulating port ischanged.

Connection when measuring/simulatingthermovoltage using Internal ReferenceJunction Mode:

Note.

The range of the Internal Reference Junction’s temperature com-pensation is -10 … +50°C (14 … 122 °F).

See also…

External Reference Junction on page 98

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

T/C INT. RJ

T/C OR EXTWIRES ONLY

SENSOR MEASURE & SIMULATE

T/C sensor or a T/C signal receiver

T/C materials (T/C, extension or

compensation wires) T/C INT. RJ

T/C OR EXTWIRES ONLY

Additional Information

Page 104: MC5 User Manual

98

Advanced Operation and Configuration

External Reference Junction

When using an external Reference Junction, CL560 measures or simulates thethermovoltage using the “T/C Low V” terminals in the ET section. If the measuring/simu-lating port is set to the External Reference Junction (ET: TCx(mea) or ET: TCx(sim)),the following Reference Junction compensation methods are available:

0°C:To be used when:• The Reference Junction temperature is

fixed to 0°C (using, e.g. ice) and CL560is not used for measuring the ReferenceJunction temperature.

• The Reference Junction temperature iscontrolled with 0°C set point.

• A compensation box is used and theReference Junction temperature settingis 0°C.

• There is no Reference Junction compen-sation in the measuring/simulation cir-cuit but 0°C is a good enough approxi-mation.

Entered:To be used when:• The Reference Junction temperature is

fixed to a temperature that differs from0°C.

• The Reference Junction temperature iscontrolled and the controller’s set pointis not 0°C. Note: this method is of useonly if the accuracy of the controller isbetter than CL560’s own ReferenceJunction Module.

• A compensation box is used and theReference Junction temperature settingis other than 0°C.

• There is no Reference Junction compen-sation in the measuring/simulation cir-cuit, but you want to manually enter an-other Reference Junction temperaturethan 0°C.

Warning!

If you connect an RTD sensor to the ETmodule’s R, RTD connectors, there is nogalvanic isolation between the thermo-couple and the RTD sensor.

RTD sensor:To be used when:• The Reference Junction temperature is

measured using an RTD sensor con-nected to CL560’s RTD terminals.

Connection when measuring/simulatingthermovoltage using Reference Junc-tion compensation method RTD sensor:

T/C, Low V 4-w measR, RTD

3 & 4-w meas

T/C INT. RJ

T/C OR EXTWIRES ONLY

SENSOR MEASURE & SIMULATE

T/C sensor or a T/C signal receiver

T/C materials Copper

Tref

T/C, Low V

Page 105: MC5 User Manual

99

Connection when measuring/simulatingthermovoltage using a temperature con-troller and Reference Junction compen-sation method Entered (also RJ-mode0°C):

Connection when measuring/simulatingthermovoltage using a compensationbox and Reference Junction compensa-tion method Entered (also RJ-mode0°C):

See also…

Internal Reference Junction on page 97

T/C, Low V 4-w measR, RTD

3 & 4-w meas

T/C INT. RJ

T/C OR EXTWIRES ONLY

SENSOR MEASURE & SIMULATE

T/C materials Copper

Tref

T/C sensoror a T/C signal receiver

T/C, Low V

T/C, Low V 4-w measR, RTD

3 & 4-w meas

T/C INT. RJ

T/C OR EXTWIRES ONLY

SENSOR MEASURE & SIMULATE

t

T/C sensoror a T/C signal receiver

T/C materials

Copper

T/C, Low VT/C, Low V 4-w measR, RTD

3 & 4-w meas

T/C INT. RJ

T/C OR EXTWIRES ONLY

SENSOR MEASURE & SIMULATE

Copper

T/C sensoror a T/C signal receiver

Temperature Controller Tref

T/C materials

T/C, Low V

Connection when measuring/simulatingthermovoltage using a fixed temperatureand Reference Junction compensationmethod Entered (also RJ-mode 0°C):

Additional Information

Page 106: MC5 User Manual

100

Advanced Operation and Configuration

PROBLEM CAUSE

CL560 (or the instrumentunder test whensimulatingthermovoltages)measures thetemperature/millivoltsignal, but the displayedtemperature reading is allwrong.(The error may vary from0 to about ±50°Cdepending on the type ofthe error)

• The thermocouple type selected inCL560 does not correspond with theused thermocouple.

• The selected Reference Junctionmode is not in accordance with theused wiring.

• Extension or compensation cable typeor connections are incorrect.

• The polarity of the cables areincorrect.

CL560 (or the instrumentunder test) displaysrandom readings duringthermocouplemeasurement.

• Incorrect connections.• The wiring is broken.• Interference from a mobile phone or a

radio transmitter affects themeasurement.

CL560 displays unstablereadings duringthermocouple simulation.

• The instrument under calibration usesvoltage pulses to detect open sensor.When CL560 detects these pulses, ittries to compensate for them, whichcauses the unstable output. Preventthese pulses for the time ofcalibration. Refer to the servicemanual of the instrument undercalibration on how to prevent theseopen sensor detection pulses.

Error situations

The easiest way to avoid errors in thermocouple measurement andsimulation is to check carefully the used wiring and the ReferenceJunction mode. The following table describes the typical error situ-ations and possible causes/corrections when working with thermo-couples:

Page 107: MC5 User Manual

101

Resistance and RTD Measurement, ConnectionsThe two main problems in resistance and RTD measurement arethe effects of wiring resistances and thermovoltages in the resis-tance measurement circuit. If possible, use 4-wire connection toeliminate the effect of wiring resistance. The special resistance mea-surement sequence of CL560 eliminates the thermovoltages of theresistance measurement circuit.

CL560 continuously checks the connection type during resistanceand RTD measurement. The found connection type is displayed onthe resistance or RTD measurement window.

3-wire System

CL560 sources current through the resis-tor from the two left side terminals. CL560measures the voltage drop across the en-tire current loop and across the low sideconnection wire. If both left side connec-tion wires are identical, CL560 can com-pensate for the resistance of the connec-tion wires.

4-wire System

CL560 sources current through the resis-tor from the two left side terminals. CL560measures the voltage drop across the re-sistor from the two right side terminals. The4-wire method gives the resistance be-tween the terminals of the resistor, it is in-sensitive to the resistance of the connec-tion wires.

4-w measR, RTDT/C, Low V

4-w measR, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

4-w measR, RTDT/C, Low V

4-w measR, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

Additional Information

Page 108: MC5 User Manual

102

Advanced Operation and Configuration

Using a Compensation Loop

CL560 sources current through the resis-tor and the compensation loop from the twoleft side terminals. CL560 measures thevoltage drop across the entire current loopand across the compensation loop. If thecompensation loop and the connectionwires of the resistor are identical, CL560can compensate for the resistance of theconnection wires.

When the compensation loop wiring sys-tem is used, CL560 displays the text “3-wire”.

2-wire System

Calibrator sources current through theresistor and measures the voltage dropacross the same terminals.

The result is acceptable, if the resistanceof the connection wires is low.

4-w measR, RTDT/C, Low V

4-w measR, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

4-w measR, RTDT/C, Low V 4-w measR, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

Page 109: MC5 User Manual

103

Current Measurement Parallel to a Test Diode, ConnectionsThe impedance of CL560’s milliampere input is low enough to enable current measure-ment parallel to a test diode in a 20 mA circuit. Connections:

Plus Side Test Diode Minus Side Test Diode

CL560 parallels the test diode of the transmitter and measures theexternally supplied output current of the transmitter. For more infor-mation concerning test diode connections of HART instruments,see Appendix 1, chapter Connecting CL560 and a HART Instru-ment on page 138.

Note.

In higher temperatures: The leakage of the diode in the instrumentmay affect the accuracy.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

+

I

Input signal X

2-w xmtr

+

-

Imeas/sinkT/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

+

I

Input signal X

2-w xmtr

+

-

Imeas/sink

Additional Information

Page 110: MC5 User Manual

104

Advanced Operation and Configuration

Parallel Functions in CL560This chapter explains what kind of simultaneous tasks you can dowith CL560.

Every module in CL560may simultaneouslyhave their own tasks.Also every availableoutput connector (ENV,AUX and COM/PRT)may have their own in-dependent tasks.

The ET module has adouble action ability: When one of the sensor measure & generatesection terminals (T/C INT RJ, T/C Low V, or R,RTD terminals) isused for measuring purposes. Then an electrical output signal maybe generated from the ET module’s output terminals. This feature isnot available when generating a signal from the sensor measure &generate section terminals.

The ET module is also capable of doing the following double tasks:

• Measuring Low Voltages or thermocouples (connected to anyterminal used when measuring thermocouples) and at the sametime either measuring or simulating an RTD with the RTD sen-sor terminals.

• Generating Low voltages or simulating a thermocouple (con-nected to any terminal used when simulating thermocouples)and at the same time measuring with the RTD sensor termi-nals.

Warning!

There is no galvanic isolation between the ET module’s mea-suring terminals and generating terminals.

When measuring in this section...

Electrical outputis possible in this section

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

Page 111: MC5 User Manual

Calibration

Things discussed in Part D:

• General presentation of whatcalibration is and the phasesof a typical calibration proce-dure.

• A description of a calibrationprocedure using CL560.

• Some application examples onhow to perform the calibrationof certain instruments.

• Creating, editing and deletingInstruments.

• Viewing calibration results.

Page 112: MC5 User Manual

106

Calibration

General

CL560 is a stand-alone calibrator as well as a calibrator communi-cating with calibration software.

CL560 supports stand-alone calibration and off-line calibration.The following list briefly describes these methods:

• In stand-alone calibration, all instrument and calibration resultdata is stored in CL560’s memory. No external calibration da-tabase is used.

• In off-line calibration (sometimes also called batch calibration),the instrument data is downloaded from a calibration software.The calibration is done just as in stand-alone calibration, butcalibration procedures are downloaded from a software andthe saved results are uploaded to the calibration software.

Page 113: MC5 User Manual

107

Phases of Instrument Calibration

The picture gives a general view of the phases in a calibrationprocedure. CL560 supports three separate guide texts (whiteblocks in the picture):

One is displayed when starting the calibration.Another when the adjustment utility is started.The third text is displayed when ending the calibration.

The guide texts may be entered in CL560 or they may be down-loaded from a PC with a calibration software.

Start

As Found test

Save As Found results

No. of repeatsdone?

Read Starting Guide Text

no yes

Adjustmentrequired?

Within limits?

As Left test

Adjustment

* If it is not possible to adjust the maximum error below the "Reject if" limit, consider replacing the instrument with a more accurate one.

Read Adjusting Guide Text

no

no* yes

yes

End

Save As Left results

No. of repeatsdone?

Read FinishingGuide Text

no yes

General

Page 114: MC5 User Manual

108

Calibration

As Found Calibration

The As Found calibration documents the state of the instrumentbefore carrying out any adjustments. The calibration states theamount of drift in the instrument during the calibration period.

CL560 supports multiple As Found tests. There is no fixed upperlimit for the number of repeats except the available free memory.

CL560 does not require you to save any As Found calibrations. Ifyou do not want to save information of the instrument’s state beforeadjustment, just do the adjustments needed and perform as manyAs Left calibrations as is needed.

Adjustment

CL560 calculates (among other figures) the maximum error foundduring the calibration. Depending on the value of the maximum er-ror, you decide whether you adjust the instrument or not. CL560supports four different error limit settings:

Reject if > acceptance limit for found maximum error,Adjust if > the instrument need to be adjusted if this

limit is exceeded,Do not adjust if < adjusting the instrument is not

necessary / of use andAdjust to < after adjusting, the maximum error should

not exceed this limit.

CL560 has a special adjustment utility tool that can be used whenthe input signal of the instrument is generated/simulated with CL560.See chapter CL560’s Support for Instrument Adjustment on page127. We recommend you use the utility whenever possible. In allother situations, use the error graph and its error value to have aview on how the adjustment affects the accuracy of the instrumentand how close the output signal is from the target value.

Use the instrument’s ZERO adjustment at lower end of the range.Adjust the higher end of the range with the instrument’s SPAN ad-justment. The two adjustment points do not necessarily need to bethe end points of the instrument range. You may want to use other

0

0 25 50 75 100%

Error

Page 115: MC5 User Manual

109

points to optimize the calibration at the most important part of therange. Note, however, that if the points are very close to each other,the error at other points of the range may be quite big (see As LeftCalibration on page 109).

As Left Calibration

Multiple As Left calibrations are allowed just as they are for As Foundcalibrations. The As Left calibrations document the state of the in-strument after the possible adjustment.

CL560 does not require you to save any As Left calibrations. If theAs Found calibration proves to be good and no adjustment isneeded, you may omit the As Left calibrations.

The pictures above display different As Left calibration graphs aftertwo different kinds of adjustments.

In the leftmost picture, the error is minimized at the end points of theinstrument range. This results in a relatively notable error in themiddle of the span. If the middle of the span is the area, where themeasurement is usually situated, then this method is not recom-mended.

The rightmost picture displays the error graph when the maximumerror is minimized by “lowering the graph”. As a result, the errorminimum is at approx. 30 % and 70 % of the span.

Note that these adjustment strategies are just examples. There isan unlimited amount of “correct” ways to adjust an instrument.

ZERO

SPAN0

0 25 50 75 100%

Error

0

0 25 50 75 100%

Error

0

0 25 50 75 100%

Error

General

Page 116: MC5 User Manual

110

Calibration

Required Modules for Different Input/Output Signal CombinationsConsider the type of your instrument before doing a calibration:

• What are the input and output quantities/ports?• Is the input signal measured, generated/simulated, controlled

or keyed (Input method)?• Is the output signal measured or keyed (Output method)?

Use the following tables find out the required modules for each sup-ported input/output quantity and method:

Table for choosing the input module Table for choosing the output module

QUANTITY METHOD REQ. MODULE QUANTITY METHOD REQ. MODULE

Voltage Measured E Voltage Measured E Generated ET Keyed none Keyed none HART E

Low Voltage E measured E Low Voltage E Measured E ET measured ET ET Measured ET Generated ET Keyed none

Keyed none HART E

Current Measured E Current Measured E E generated E Keyed none

ET generated ET HART E Keyed none Resistance Measured ET

Resistance Measured ET Keyed none Simulated ET HART E Keyed none Pressure Measured INT or EXT

Pressure Measured INT or EXT Keyed none Controlled INT or EXT HART E Controlled/Meas none Temp. Keyed none

Keyed none HART E

Temperature Keyed none TC temperature Measured ET Controlled/Meas none Keyed none

TC temperature Measured ET HART E Controlled ET RTD temperature Measured ET Controlled/Meas none Keyed none

Simulated ET HART E

RTD temperature Measured ET Frequency Measured E Controlled ET Keyed none Controlled/Meas none HART E Simulated ET Value Keyed none

Frequency Measured E Switch Measured E

Generated ET Keyed none

Value Keyed none

Page 117: MC5 User Manual

111

The general principle is: The same module may not be used forboth the input and the output signal. There are however some ex-ceptions to that rule:

• If the input signal is a current signal measured or generatedusing the E module, then any quantity with the output methodHART may also be connected to the E module.

• If when checking the required modules both the input and theoutput module is ET, but the cells are grayed, check the tablebelow. It presents all supported combinations where both theinput signal and the output signal is connected to the ET mod-ule.

Each row is a supported combination.

Supported input/output quantities and methods where both the input signal and the output signal is connected to the ET module:

INPUT QUANTITY INPUT METHOD OUTPUT QUANTITY OUTPUT METHOD

Voltage Generated Low Voltage ET Measured

Generated TC Temperature Measured

Generated RTD Temperature Measured

Low Voltage ET Measured RTD Temperature Measured

ET Generated RTD Temperature Measured

Current ET Generated Low Voltage ET Measured

ET Generated TC Temperature Measured

ET Generated RTD Temperature Measured

TC temperature Measured RTD Temperature Measured

Controlled RTD Temperature Measured

Simulated RTD Temperature Measured

RTD temperature Measured TC Temperature Measured

Controlled TC Temperature Measured

Simulated TC Temperature Measured

Frequency Generated Low Voltage ET Measured

Generated TC Temperature Measured

Generated RTD Temperature Measured

Generated Resistance Measured

General

Page 118: MC5 User Manual

112

Calibration

Calibrating an Instrument

Selecting the Instrument to Be CalibratedTo move from Basic Mode to Cali-bration Mode, press the A/Calibra-tion Mode Function Key.

A list of instruments available inCL560’s memory is presented. Ifyou do not have any instrumentdata in CL560’s memory, eitherdownload them from a calibrationsoftware or create instruments inCL560 as described in chapterAdding New Instruments on page128.

The Position ID / Device ID of analready calibrated instrument istick-marked ( ).

When you move the cursor (the inverted text), the additional datapresented in the lower part of the screen changes to reflect the dataof the current instrument.

Select one of the available instruments by pressing the , orthe C/Select key.

Note.

If there are a lot instrument data in CL560’s memory, displaying thelist may take a few seconds.

25.09.2000 10:16

POSITION/DEVICE ID

101-XI-001.1112-TT-003.1112-TT-007.1PT101.6PT112.12PT115.15-1PT115.15-2PT112.16TT112.08TT112.10TT112.12

Device ID

Position Name

Calibrated

TT112.07

Intake Temperature on Feed 1

Not Calibrated

ModeBasic MENUSelect

Page 119: MC5 User Manual

113

The Instrument WindowThe Instrument window gives youan overview of the selected instru-ment.

You can view detailed instrumentdata by pressing the D/MENUFunction Key and 1/View Instru-ment Details.

To edit the Instrument data, pressthe B/Edit Function Key. More in-formation on editing instrumentdata is in chapter Editing Instru-ment Data on page 129.

If your instrument is calibrated us-ing an external device that is com-municating with CL560, select D/MENU Function Key and 7/Con-troller Settings to select the device and start the communication.

To calibrate the selected instrument, simply press C/Calibrate.General calibration procedures are discussed in chapter A Cali-bration Procedure Using CL560 on page 114. Some calibrationexamples are available starting from chapter Examples of Instru-ment Calibration on page 117.

If a Starting Guide text is written, it appears before the CalibrationWindows are opened. To close the window displaying the note text,press D/OK.

25.09.2000 10:20 INSTRUMENT

112-TT-003.1

Edit Calibrate MENU

Intake Temperature on Feed 1

Cal. Points 5 ↑↓

Transfer Function Linear

Method Simulated

INPUT 0.00 ... 100.00 °C

Sensor Type Pt100 α385

Method Measured

OUTPUT 4.0000 ... 20.0000 mA

Back

Calibrating an Instrument

Page 120: MC5 User Manual

114

Calibration

A Calibration Procedure Using CL560The Calibration Windows:

A typical calibration procedure using CL560 is as follows:

We assume you have already selected the instrument to becalibrated and are viewing/editing the selected instrument’s datapages. See Selecting the Instrument to Be Calibrated on page112 for information on how to enter Calibration Mode and se-lecting instruments for calibration.

1. Make the required connections and test them by altering theinput signal in with the Instrument Adjustment utility (if youhaven’t done the connections and testing in Basic Mode).

2. Start the calibration run by pressing B/Start. The way CL560passes through all calibration points depends on the Calibra-tion Method setting:In Automatic calibration CL560 generates/simulates (or com-municates with an external device capable of generating/simu-lating) the input signal. See chapters

About Automatic Calibration on page 116 andAbout Manual Calibration on page 116.

Note.

You can pause a calibration run whenever you want bypressing the A/Stop Function Key. Then you are providedwith the following (self-explanatory) choices:

A/ContinueB/EndC/Skip Point.

25.09.2000 10:46 Input

Error1.00

Output

Pt100 α385°C (ITS90)

3.9762

-0.15 % of span

0.00

+

0

-0% 50% 100%

RTD Temperature [ET: Simul]

Current [E: Meas]

mA

Start MENU Back Adjust

Input Signal Window

ErrorGraph

Output Signal Window

Page 121: MC5 User Manual

115

3. When the calibration run isready, you can either Rejectthe results (A/Reject Calibra-tion) or Continue by pressingthe D/Saveè Function Key.

Note.

If the maximum error of thecalibration results exceed theAdjust if > error limit, a mes-sage window with the follow-ing text:

”The calibrated instrumentneed to be adjusted.”

4. Enter the environmental dataand save the results using oneof the options provided in theFunction Keys B to D.

Note.

Some of the temperature datamay be pre-entered by CL560,provided the data is availablefor CL560 (pressure modules,the E module, the ET moduleor the optional ENV sensorwas used during the calibra-tion).

5. Continue either with anothercalibration run or end the calibration procedure. In the lattercase you will see the possible Finishing Guide text beforeyou are back viewing the Instrument Data.

Note.

If you are calibrating instruments using off-line method then pleasekeep in mind this: When you upload the results from CL560 to thecalibration software, the results are by default also left in CL560’smemory. To free memory in CL560 by deleting the results, see De-leting Instruments on page 132.

See also:

CL560’s Support for Instrument Adjustment on page 127.Viewing Calibration Results on page 133.

25.09.2000 10:52 Input

Error1.00

Output

Pt100 α385°C (ITS90)

4.0293

0.18 % of span

0.00

+

0

-0% 50% 100%

RTD Temperature [ET: Simul]

Current [E: Meas]

mA

CalibrationReject Save  

25.09.2000 10:55 ENVIRONMENTAL DATA

Device TemperatureEnvironment Temp.Input Module Temp.Output Module Temp.

Humidity

Calibrated by:

Notes:

Andy Handyman

Just Testing

21.021.031.629.6

°C°C°C°C

30 %

As FoundSave

As LeftSave

As BothSave Back

Calibrating an Instrument

Page 122: MC5 User Manual

116

Calibration

About Automatic Calibration

Automatic calibration is possible when CL560 is able to generate/simulate the input signal. Another way to automate the calibrationis allowing CL560 to control a connected external device capableof generating/simulating the input signal, e.g. a POC4 Pressure Out-put Controller connected to CL560.

Automatic calibration is selected in the Calibration Method fieldon the Instrument Data pages. More information on editing instru-ment data is in chapter Editing Instrument Data on page 129.

In Automatic calibration CL560 generates/simulates (or commu-nicates with an external device capable of generating/simulating)the input signal. When the input signal is within the Maximum Al-lowed Calibration Point Deviation limits, CL560 waits for the out-put signal to stabilize (the Setpoint Delay setting). Then the cali-bration point is accepted automatically and CL560 continues withthe next calibration point.

If the input signal never stabilizes (the symbol is visible con-tinuously), but you want CL560 to save the results anyhow, pressthe B/Force Accept Function Key.

About Manual Calibration

In manual calibration, you will have to manually set the input signalto a level that the next calibration point requires. Manual calibrationis possible in almost all situations. Even for calibrations that couldbe done automatically.

If the input signal is measured, the input signal window displays a“Desired Input Value X.xxxx” text to help you remember what isthe next target value.

If the input signal is generated/simulated, CL560 automaticallychanges the input signal to the next target value, but does not con-tinue until you manually accept the point.

Use the B/Accept Readings Function Key to save the point dataand continue to the next point until all points are accepted.

Notes.

If the calibration points are defined for the output signal, set theinput signal to such a level that the instrument to be calibrated cre-ates the required output signal.

See also Autocapture Feature on next page

Page 123: MC5 User Manual

117

Autocapture Feature

The Autocapture Feature makes it possible to make a manual cali-bration that is semi-automatic when the input signal in measured.To be able to utilize the Autocapture Feature the instrument datahas to be set as follows:

• Input Method field set to “Measured”,• Calibration Method field set to “Automatic” and• Maximum Allowed Calibration Point Deviation field set to

greater than zero.

Then CL560 tracks the input signal and once the input signal isinside the Maximum Allowed Calibration Point Deviation limitfor a time period defined in the Setpoint Delay, the input and out-put values are automatically saved.

Activate the Autocapture Feature from the menu available whenviewing the calibration windows. When activated and a calibrationpoint is about to be captured (CL560 is waiting for the SetpointDelay time to pass while the measurements are stable enough andwithin the Maximum Allowed Calibration Point Deviation limit),an hourglass is shown in the input measurement window’s lowerright corner.

When autocapture “snatches” the input and output signal values,CL560 emits a beep and the next target value for the input is dis-played.

Examples of Instrument CalibrationThis User Guide has only a few examples of how to calibrate aninstrument. Many calibration procedures not presented here areeither very similar to the ones presented in the following chaptersor they can be “combined”, e.g. the calibration of a Pressure Indi-cator/Recorder is partly done as a Pressure Transmitter and partlyas a Temperature Indicator/Recorder calibration.

Use the information presented in Part B of this manual when con-necting/calibrating instruments that do not have a separate calibra-tion procedure presented in this part.

The following calibration examples are presented in this manual:

Pressure Transmitters on page 118,Temperature Sensors on page 120,Temperature Indicators and Recorders on page 122 andElectrical Limit Switches on page 124.

Calibrating an Instrument

Page 124: MC5 User Manual

118

Calibration

Pressure Transmitters and Sensors

This procedure suits all instruments with a pressure input and anykind of electrical output signal. Both the input and output signalrange should be measurable with CL560. To be able to perform anautomatic calibration, CL560 needs to have the firmware optionneeded to communicate with the pressure controller.

Required modules• An internal or external pressure module.• The E module for measuring the electrical output signal.

Preparations

1. Connect the input signal of the instru-ment to the pre-selected pressure mod-ule (INT, EXT, XPM). If you are calibrat-ing a pressure sensor, and have con-figured CL560 to provide the supplyvoltage or current, connect the ETmodule’s output terminals (marked witha circle in the picture) to the sensor’ssupply terminals. For more informationon sensor supply configuration, seechapter Instrument Input Page onpage 130.

2. Connect the pressure source both to theinstrument’s input and to the pre-se-lected pressure module (INT, EXT,XPM). If a pressure output controller isused, connect it to CL560’s AUX inter-face.

3. Connect the instrument’s output signalto the chosen electrical input port inCL560.

4. Test the connections in Basic Mode ifneeded. To quickly configure the BasicMode’s windows, go to CalibrationMode, select the instrument to be cali-brated and the Function Key C/Cali-brate but immediately return to BasicMode.

P

EPressureSource

0 6 . 1 0 . 2 0 0 0 1 1 : 4 9 I n p u t

E r r o r1 . 0 0

O u t p u t

G a u g eb a r

8 . 0 3 1 0

0 . 2 8 % o f s p a n

0 . 4 9 8 2

+

0

-0 % 5 0 % 1 0 0 %

P r e s s u r e [ P 1 : I N T 6 C ]

C u r r e n t [ E : M e a s ]

m A

P a u s eR e a d i n g s

A c c e p t

D e s i r e d I n p u t V a l u e 0 . 5 0 0 0

T / C , L o w V4 - w m e a s

R , R T D

3 & 4 - w m e a s

V , I ,+ 2 4 V I m e a s / s i n k

L o w V

H A R T ®

T / C I N T . R J

T / C O R E X TW I R E S O N L Y

2 - w x m t r

V , ,

?

M E A S U R EO U T P U TS E N S O R M E A S U R E & S I M U L A T EE T E

M a x i n p u t :6 0 V D C /3 0 V A C

C L 5 6 0M U L T I F U N C T I O N C A L I B R A T O R

Page 125: MC5 User Manual

119

Calibration

1. Move to Calibration Mode and select theinstrument to be calibrated. If appli-cable, activate the pressure controllercommunication as described in chap-ter The Instrument Window on page113.

2. Start the calibration as presented inchapter A Calibration Procedure Us-ing CL560 on page 114.

3. Set the pressure manually using, e.g.a pressure pump. If a pressure control-ler is connected to CL560, the pressurecontroller generates the pressure basedon CL560’s requests.

4. In manual calibration: Accept the pointsusing the B/Accept Readings FunctionKey. The calibration proceeds automati-cally if a pressure output controller isavailable and the Calibration Mode isset to Automatic. Additional CalibrationMethod specific data is in chaptersAbout Automatic Calibration andAbout Manual Calibration, both onpage 116.

5. Save or reject the results as is de-scribed in chapter A Calibration Pro-cedure Using CL560 on page 114.

6. Either do another calibration run or re-turn to the Instrument window.

Hint!

These instructions can be adapted for pneumatic pressure trans-mitters and converters, too. Instead of connecting the output sig-nal to a terminal in the E module, connect the pressure output sig-nal to a suitable pressure module.

Warning!

Working with pressure instruments can be hazardous. Onlyqualified personnel may use pressure instruments and pres-sure sources.

Next…

CL560’s Support for Instrument Adjustment on page 127.

Calibrating an Instrument

Page 126: MC5 User Manual

120

Calibration

Temperature Sensors

This procedure suits temperature sensors no matter if they are RTDsor thermocouples. The sensor’s output signal is measured with CL560or it is keyed into CL560. The reference temperature is either mea-sured with CL560 or it is entered or communicated (Input MethodControlled/Meas) to CL560. Automatic calibration requires the firm-ware option for communicating with the calibration bath / dry block.

If you measure the reference temperature using CL560, rememberthe following conditions:

• If the sensor to be calibrated is a thermocouple, the referencethermometer needs to be an RTD sensor.

• If the sensor to be calibrated is an RTD sensor, the referencethermometer needs to be a thermocouple.

Required modules• The ET module for measuring the sensor’s output signal and

possibly also for measuring the reference temperature.

Preparations

1. Connect the calibration bath / dry blockto CL560’s AUX connector (if appli-cable) using the communication cableprovided with CL560. Place the sensorof the instrument (and the sensor of thereference thermometer) into the calibra-tion bath/dry block. Connect it/them tosuitable terminals in CL560 (RTD orT/C). If the sensor is a thermocouple,select one of the available referencejunction methods:

• The internal Reference Junctionmodule is in use.

• The reference junction is fixed to0°C.

• The reference junction tempera-ture is entered.

• The reference junction tempera-ture is measured externally usingan RTD connected to the RTD ter-minals.

See Part C for additional informationconcerning the reference junction meth-ods.

Referencethermometer

Sensor

06.10.2000 11:06 Input

Error1.00

Output

°C (ITS90)

24.94

-0.06 % of span

25.01

+

0

-0% 50% 100%

TC Temperature [ET:Control]

RTD Temperature [ET: Meas]

°C (ITS90)

PauseReadings

Accept

Desired Input Value 25.00

Pt100 α385

K NiCr/NiAl

CL560M U L T I F U N C T I O N C A L I B R A T O R

T/C, Low V 4-w measR, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

?

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

2. Test the connections in Basic Mode ifneeded. To quickly configure the BasicMode’s windows, go to CalibrationMode, select the instrument to be cali-brated and the Function Key C/Cali-brate but immediately return to BasicMode.

Page 127: MC5 User Manual

121

Calibration

1. Move to Calibration Mode, select theinstrument to be calibrated and start thecalibration as presented in chapter ACalibration Procedure Using CL560on page 114.

2. Start the calibration as presented inchapter A Calibration Procedure Us-ing CL560 on page 114.

3. Set the temperature of the bath/dryblock to the required calibration point.If the temperature bath/dry block com-municates with CL560, CL560 controlsthe temperature of the bath/dry block.Wait until the temperature stabilizes.

4. If the reading of the reference tempera-ture sensor is entered manually, enterthe value and press B/Accept Read-ings Function Key. If the reference tem-perature sensor is connected to CL560and the Calibration Method is Manual,accept the readings using the B/AcceptReadings Function Key. In automaticcalibration the calibration proceeds au-tomatically. More information on Cali-bration Methods is in chapter AboutAutomatic Calibration and AboutManual Calibration, both on page 116.

5. Save or reject the results as is de-scribed in chapter A Calibration Pro-cedure Using CL560 on page 114.

6. Either do another calibration run or re-turn to the Instrument window.

Hint!

With small changes, this procedure should also suit all tempera-ture instruments with a fixed sensor. In that case, use a referencesensor that can be connected to CL560 to measure the input signaland either measure or enter the output signal of the instrument.

Use one of the following examples for the output signal:

• Temperature Transmitter with sensor. See the examplePressure Transmitters on page 118.

• Temperature Indicators/Recorders with sensor. See theexample Temperature Indicators and Recorders on page122.

• Temperature switches (thermostats) with sensor. See theexample Electrical Limit Switches on page 124.

Next…

CL560’s Support for Instrument Adjustment on page 127.

Calibrating an Instrument

Page 128: MC5 User Manual

122

Calibration

Temperature Indicators and Recorders

This procedure suits Temperature Indicators/Recorders no matterif the sensor is an RTD or a thermocouple. The instruments inputsignal is simulated using CL560 and the output reading is enteredinto CL560.

Required modules• The ET module for simulating the temperature sensor. The

Base Unit takes care of entered output signal.

Preparations

1. Connect suitable terminals in CL560 tothe input connectors of the indicator/recorder.

2. If the sensor to be simulated is a ther-mocouple, select one of the followingreference junction methods:

• The internal Reference Junctionmodule is in use.

• The reference junction is fixed to0°C.

• The reference junction tempera-ture is entered.

• The reference junction tempera-ture is measured externally usingan RTD connected to the RTD ter-minals.

See Part C for additional informationconcerning the reference junction meth-ods.

3. If the sensor to be simulated is an RTD,the used wiring system depends on theinstrument to be calibrated. Rememberto only use the two leftmost RTD termi-nals in CL560.

4. Test the connections in Basic Mode ifneeded. To quickly configure the BasicMode’s windows, go to CalibrationMode, select the instrument to be cali-brated and the Function Key C/Cali-brate but immediately return to BasicMode.

06.10.2000 10:46 Input

Error1.00

Output

°C (ITS90)

74.800

-0.20 % of span

75.00

+

0

-0% 50% 100%

RTD Temperature [ET: Simul.]

Value [Keyed]

°C

PauseReadings

Accept

Pt100 α385

Enter Output Reading

CL560M U L T I F U N C T I O N C A L I B R A T O R

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

?

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

Page 129: MC5 User Manual

123

Calibration

1. Move to Calibration Mode, select theinstrument to be calibrated and start thecalibration as presented in chapter ACalibration Procedure Using CL560on page 114.

2. CL560 simulates the output signal of thetemperature sensor for each calibrationpoint. The middlemost window displaysthe default value for the reading. Youhave two possibilities on how to con-tinue:

• Adjust the simulated signal untilthe reading of the indicator/re-corder is exactly the same as thedefault value displayed in CL560.This method is useful when cali-brating analog indicators/record-ers and the input signal sourcesupports fine adjustment.

• Select the field displaying the de-fault value and enter the actualreading. This method is usefulwhen calibrating indicators withdigital display or when the inputsignal cannot be fine adjusted.

3. Accept the points using the B/AcceptReadings Function Key.

4. Save or reject the results as is de-scribed in chapter A Calibration Pro-cedure Using CL560 on page 114.

5. Either do another calibration run or re-turn to the Instrument window.

Hint!

This procedure can be used as a reference when calibrating anykind of indicators/recorders. Adapt the input signal connections/set-tings according to the instruments input signal.

When calibrating a temperature instrument with a detachable sen-sor (the input signal is simulated with CL560), use this example asthe source for the input signal part.

Next…

CL560’s Support for Instrument Adjustment on page 127.

Calibrating an Instrument

Page 130: MC5 User Manual

124

Calibration

Electrical Limit Switches

This procedure suits limit switches with an electrical input. The in-put signal is generated with CL560.

Required modules• The ET module for generating the required electrical signal

(Voltage or Current).

• The E module for detecting the switch’s state.

Preparations

1. Connect the ET module’s terminalsmarked “OUTPUT” to the input of theswitch.

2. Connect the switch contact to the switchterminals in the E module.

3. Test the connections in Basic Mode ifneeded. To quickly configure the BasicMode’s windows, go to CalibrationMode, select the instrument to be cali-brated and the Function Key C/Cali-brate but immediately return to BasicMode.

Calibration

1. Move to Calibration Mode and select theinstrument to be calibrated.

2. Start the calibration as is presented inchapter A Calibration Procedure Us-ing CL560 on page 114. CL560 doesthe Prescan if it is enabled. During theprescan, CL560 searches for approxi-mate values for the actuating anddeactuating point. This speeds up thefinal test without sacrificing accuracy.The prescan test is done only once in acalibration.

• By default, the Prescan is set to on.If you do not want CL560 to performa prescan, disable it (commandsD/MENU and 2/Prescan). WhenPrescan is set to off, CL560 usesthe whole scan range also duringthe actual test.

3. The actual test is done automatically:CL560 slowly increases the input sig-nal until the switch actuates and con-tinues by decreasing the input signaluntil the switch deactuates. CL560’sscreen displays the obtained data asthe test advances.

06.10.2000 10:29 Input

Output

V

8.0310

5.023Voltage [ET: Gen.]

Switch [E: Meas.]

Actuate

Pause

Testing switch limits (4/5)

Deactuate-------7.000

5.500

4.500

3.000

CL560M U L T I F U N C T I O N C A L I B R A T O R

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

?

MEASUREOUTPUTSENSOR MEASURE & SIMULATE

ET EMax input:

60 VDC/30 VAC

Page 131: MC5 User Manual

125

• If for some reason the calibrationcannot be done automatically (be-cause the input signal is measured,not generated/simulated), you willhave to change the input signalmanually. The slower you raise theinput signal the more accurate theobserved actuation point is. Beequally careful when decreasing theinput signal in order to get thedeactuation point.

4. Save or reject the results as is de-scribed in chapter A Calibration Pro-cedure Using CL560 on page 114.

5. Either do another calibration run or re-turn to the Instrument window.

Notes.

Selecting Calibration Method Manual when the input signal is gen-erated/simulated with CL560 means that each pair of actuation/deactuation points have to be accepted manually.

If the prescan stops and displays an error message, try wideningthe scan range. It is also advisable to have the scan range sym-metrical in relation to the actuation/deactuation points.

Hint!

This procedure also suits other types of switches. How to calibrateother types of switches may be adapted by combining the informa-tion of this example with the input signal information from one of thefollowing examples:

• Pressure switches. See the example Pressure Transmit-ters on page 118.

• Temperature switches with a fixed sensor. See the exampleTemperature Sensors on page 120.

• Temperature switches with a detachable sensor. See theexample Temperature Indicators and Recorders on page122.

Next…

CL560’s Support for Instrument Adjustment on page 127.

Calibrating an Instrument

Page 132: MC5 User Manual

126

Calibration

Using External Devices in Calibration ModeIn order to utilize external devices when calibrating an instrument,the following things need to be set/available before the calibrationis started:

• The required firmware option has to be installed in CL560 (andwhen applicable, in the connected external device).

• The Instrument’s Input Method field has to be set either toControlled or Controlled/Meas. About the settings:

∗ Controlled: The input signal (pressure or temperature)is measured using one of CL560’s measurement inputports.

∗ Controlled/Meas: The input signal is measured by theexternal device and communicated to CL560.

In both cases: CL560 sends the set point values to the exter-nal device via the communication cable connected to the AUXconnector

• The Instrument’s Calibration Method field has to be set toAutomatic.

• In the Instrument Window:Select the D/Menu and 7/Con-troller Settings to select theController type and activatethe communication.

Notes.

The Input Method option Con-trolled is available for the follow-ing Input Quantities: Pressure,RTD temperature and TC tem-perature.

The Input Method option Con-trolled/Meas is available for thefollowing Input Quantities: Pressure, Temperature, RTD Tempera-ture and TC Temperature.

More information on how to create instruments into CL560’s data-base and how to edit the settings of an existing instrument is avail-able in chapter Maintaining CL560’s Instrument Database on page128 and its subchapters.

Chapter Generating Signals Using External Devices on page 81contains a description on how to utilize external devices in BasicMode and also information on controller settings.

8.11.2001 15:24 INSTRUMENT

112-TT-003.1

Edit Calibrate MENU

Intake Temperature on Feed 1

Cal. Points 5 ↑↓

Transfer Function Linear

Method Controlled

INPUT 0.00 ... 100.00 °C

Sensor Type Pt100 α385

Method Measured

OUTPUT 4.0000 ... 20.0000 mA

BackMENUClose

DeleteInstrument

ControllerSettings

ViewInstrumentDetails

ViewCalibrationResults

Page 133: MC5 User Manual

127

CL560’s Support for Instrument AdjustmentCL560 includes a special utility foradjusting instruments.

This tool is of use when CL560 gen-erates/simulates the input signal orcontrols an external device thatgenerates/simulates the input sig-nal (automatic calibration). Use thestandard calibration screen whenadjusting instruments that are cali-brated manually.

When viewing the three CalibrationWindows, open the adjustment util-ity by pressing the C/Adjust Func-tion Key.

Press the Function Keys B, C and D to quickly change the value ofthe input signal.

Additionally, the input signal field is editable during instrument ad-justment. This makes it possible for you to enter generation/simula-tion values other than the ones available via the Function Keys.

Hint.

For help on how to adjust/trim HART instruments, see Appendix 1.

25.09.2000 11:00 Input

Error1.00

Output

Pt100 α385°C (ITS90)

19.9918

-0.05 % of span

100.00

+

0

-0% 50% 100%

RTD Temperature [ET: Simul]

Current [E: Meas]

mA

0% 100% Back 50%

Calibrating an Instrument

Page 134: MC5 User Manual

128

Calibration

Maintaining CL560’s Instrument Database

This chapter and its subchapters describe how to create instruments,edit instrument data and delete instruments.

Adding instruments to CL560’s database by downloading the in-struments from an external calibration database software is not pre-sented here. When using such a software, consult the InstructionManual of the software.

CL560’s memory is dynamic. There are no fixed amount instrumentsand calibration data that you may keep in CL560’s memory. It alldepends on how many calibration repeats and how much historydata is kept in the memory.

Normally, you can safely store some hundreds of instruments andtheir calibration results without having “out of memory” messages.

If you want to delete instruments or calibration data from CL560’smemory, see chapters Deleting Instruments on page 132 andDeleting Calibration Results on page 134.

Adding New InstrumentsIf you go to Calibration Mode, butthere are no instruments in CL560’sdatabase, the Position ID list isempty except for the possibility tocreate a new instrument. Select the

or the key to create a newinstrument.

If the Position ID list is not empty,you can add new instruments byselecting D/MENU and 1/CreateNew Instrument.

03.10.2000 08:51

POSITION/DEVICE ID

Create New Instrument...

Device ID

Position Name

Calibrated

ModeBasic MENUSelect

Page 135: MC5 User Manual

129

Whichever way you start creating the instrument, you will first haveto select the Input Quantity and the Output Quantity. These selec-tions cannot be changed afterwards.

SELECT INSTRUMENT INPUT

VoltageLow VoltageCurrentResistancePressureTemperatureTC TemperatureRTD TemperatureFrequencyValue

SELECT INSTRUMENT OUTPUT

VoltageLow VoltageCurrentResistancePressureTemperatureTC TemperatureRTD TemperatureFrequencyValueSwitch

After the Quantity selections, there are five pages of instrumentdata that need to be completed. They are presented in chapterEditing Instrument Data.

Note.

There’s also a possibility to create a new instrument based on aninstrument already existing in CL560’s memory. Do this with theCopy Instrument option available in menu.

Editing Instrument DataThe instrument data pages may be re-edited whenever you want.

General Data Page

Out of the four text fields in the up-per part of this page, only PositionID or Device ID is required.

Change the Error calculation for-mula and the error limits accordingto your needs. Zero value in anyerror limit field means that the limitis not compared against the calcu-lated maximum error.

19.10.2000 10:04

Position ID

MENU

PT 101.6

Position Name Feed 3, pressure

Device ID

Device Name

Error Calc. Method % of span

Reject if 0.50

Adjust if 0.30

Do not Adjust if 0.10

Adjust to 0.10

>

>

<

<

SavepageNext Edit

INSTRUMENT GENERAL DATA

Maintaining CL560’s Instrument Database

Page 136: MC5 User Manual

130

Calibration

Instrument Input Page

The input quantity defines whichfields are required/visible. Commonfields for all quantities are:

Input Method,Unit andRange.

All other quantities except Valuealso have the Input Port field.

Some additional fields appear whenthe input quantity is Pressure, RTDTemperature or T/C Tempera-ture, e.g. Pressure Type in theadjacent picture.

The Sensor Supply fields at thebottom of the window are unique for pressure instruments. Theyallow configuring supply voltage or current to a pressure sensor.

Instrument Output Page

The output quantity includes thecorresponding common fields asthe input quantity.

The same additional fields are alsoavailable for the output quantity,with the following exceptions/addi-tions:

If the output quantity is Pressure,the pressure type is always“Gauge”.

Output quantity Frequency in-cludes a Trigger Level setting.

Remember to also check theTransfer Function setting at the bottom of the display. The defaultsetting is Linear.

27.09.2001 13:01INSTRUMENT INPUT

MENU

Input Method Controlled

Unit bar

Range 0.000000

SavePageNext

Pressure

Port 1 INT1C

Pressure Type Gauge

0 %1.000000100 %

Edit

Sensor Supply ET:V(gen)Level ??? V

03.10.2000 12:47INSTRUMENT OUTPUT

Output Method Measured

Range 4.000000

Current

Port E

Unit mA

0 %20.000000100 %

Transfer Function Linear

MENUSavepageNext Edit

Page 137: MC5 User Manual

131

Switches have an altogether differ-ent group of settings as shown inthe adjacent picture.

The actuation and deactuationpoint errors are calculated againstthe corresponding nominal values.

Error Calculation Directions de-fine error directions to be observedwhen calculating errors.

Scan Range values define therange where CL560 searches forthe approximate actuation anddeactuation point at the beginningof an automatic calibration of aswitch. The range limits should be symmetrical in relation to theactuation/deactuation points.

Calibration Settings Page

The Setpoint Delay setting is ofuse only when the CalibrationMethod is set to Automatic. Itdefines how long CL560 waits be-fore saving the input and outputvalues after input value waschanged to the next calibrationpoint.

Maximum Allowed CalibrationPoint Deviation is a limit that de-fines how far the input signal valueof the saved point can be from thetheoretical calibration point. Thissetting is used both in manual andautomatic calibration.

Again, the settings for a switch is somewhat different:

An additional Number of Repeats field is visible but no Calibra-tion Points or Maximum Allowed Calibration Point Deviationfields.

03.10.2000 13:05INSTRUMENT OUTPUT

Switch Type

Switch: Pressure [bar]

Nominal PointsActuating 0.700000Deactuating 0.600000

Error Calculation DirectionsActuating Up/DownDeactuating Up/Down

Normally Closed

Scan RangeStart Point 0.400000End Point 0.900000

MENUSavepageNext Edit

19.10.2000 10:06CALIBRATION SETTINGS

Calibration Method Automatic

As Found 0

Setpoint Delay 5.00

Maximum Allowed Calibration Point Deviation

Calibration Points 5

As Left 1

0.00<

s

%

MENUSavepageNext Edit

Cal. Period, Days 360<

Calibration Repeats

Maintaining CL560’s Instrument Database

Page 138: MC5 User Manual

132

Calibration

Calibration Instructions Page

In addition to the actual instrumentdata, there are three note fields.Starting Guide, Adjusting Guideand Finishing Guide. They can beused to provide the calibrating tech-nician with any kind of instructionsthat may be practical during thecalibration procedure. The notetexts may be written in CL560 orthey can be downloaded from acalibration software. On this page,the three first rows of the note textsare shown, but when viewed dur-ing calibration each note may beas long as a full sized text window.

The fields on the Instructions Page need not be utilized, but if theyinclude any text they will be shown at the appropriate stage of thecalibration.

Deleting InstrumentsIf you want to free some memory or otherwise do not need certaininstrument data anymore, you may remove instruments fromCL560’s memory as follows:

To remove all instruments (and their calibration data), open the Po-sition List and press D/Menu, 6/Delete All Instruments (see theleftmost picture below).

To remove a single instrument (and its calibration data), select itfrom the Position ID list. When viewing the instrument window, pressD/Menu and 5/Delete Instrument (see the rightmost pictureabove).

03.10.2000 13:52CALIBRATION INSTRUCTIONS

Starting Guide

Adjustment Guide

Finishing Guide

Required tools: Pressure pump or a Pressure Controller

Remember the "Do not Adjust if" limit...

Clean up after your work!

MENUSavepageNext Edit

25.09.2000 10:16

POSITION/DEVICE ID

101-XI-001.1112-TT-003.1112-TT-007.1PT101.6PT112.12PT115.15-1PT115.15-2PT112.16TT112.08TT112.10TT112.12

Device ID

Position Name

Calibrated

TT112.07

Intake Temperature on Feed 1

Not Calibrated

ModeBasic MENUSelect

MENUClose

Create New Instrument

Delete AllInstruments

14.06.2000 8:04 INSTRUMENT

112-TT-003.1

Edit Calibrate MENU

Intake Temperature on Feed 1

Cal. Points 5 ↑↓

Transfer Function Linear

Method Simulated

INPUT 0.00 ... 100.00 °C

Sensor Type Pt100 α385

Method Measured

OUTPUT 4.0000 ... 20.0000 mA

BackMENUClose

DeleteInstrument

Controller Settings

ViewInstrumentDetails

Make a Copy

ViewCalibrationResults

Page 139: MC5 User Manual

133

Viewing Calibration Results

You can open the calibration result windows from two places:

When viewing the instrument window, press D/Menu and 3/ViewCalibration Results (see the leftmost picture below).

When viewing the calibration windows, and you haven’t started acalibration run, press D/Menu and 2/View Calibration Results(see the rightmost picture above).

14.06.2000 8:06 INSTRUMENT

112-TT-003.1

Edit Calibrate MENU

Intake Temperature on Feed 1

Cal. Points 5 ↑↓

Transfer Function Linear

Method Simulated

INPUT 0.00 ... 100.00 °C

Sensor Type Pt100 α385

Method Measured

Output 4.0000 ... 20.0000 mA

BackMENUClose

DeleteInstrument

Controller Settings

ViewInstrumentDetails

Make a Copy

ViewCalibrationResults

29.05.2002 16:01 Input

Error1.00

Output

Pt100 α385°C (ITS90)

3.9762

-0.15 % of span

0.00

+

0

-0% 50% 100%

RTD Temperature [ET: Simul]

Current [E: Meas]

mA

Start MENU Back AdjustMENUClose

Prescan[On]

Start HARTAdjustment

ViewCalibrationResults

ZeroPressureModule

Viewing Calibration Results

Page 140: MC5 User Manual

134

Calibration

Calibration Result WindowsThe results are displayed in tableformat and as a graph. There arealso some additional data display-ing the statistics of the calibrationas well as environmental informa-tion that was automatically insertedduring the calibration or it wasmanually entered after the calibra-tion. If any calibration notes werewritten, they can also be seen onone of the calibration result pages.

Use the B/Next Page, to changefrom one result page to another.

The table with numeric results mayinclude more rows in the table thanwhat can be displayed. Use thevertical cursor keys to scrollthrough all rows.

How to Choose Which Calibration Run is Viewed

When viewing Calibration results, press D/MENU, 1/CalibrationHistory. A list of saved calibrations is displayed. Select the calibra-tion run to be viewed by moving the cursor (inverted text) to thedesired calibration run and pressing either the

or the key..

Deleting Calibration ResultsWhen viewing Calibration results, press D/MENU.

To remove the current calibration run, press 3/Remove Results.

To remove all calibration results (for the current instrument), press4/Remove All Results.

03.10.2000 15:22

BackPageNext MENU

25.09.2000 10:52 - As Left - Passed

Input

[°C]

0.0025.0050.0075.00100.0075.0050.0025.000.00

Output

[mA]

3.99307.989011.983415.977719.972915.987511.99257.99794.0032

Error

[%]

-0.044-0.069-0.104-0.139-0.170-0.078-0.047-0.0130.020

1 2

03.10.2000 15:22

PageNext MENU Back

0.40 Output Error [% of span]+

0

-0% 50% 100%

Maximum ErrorMaximum HysteresisMaximum UnlinearitySpan Error

-0.1700.0640.044

-0.126Zero Error -0.044

Page 141: MC5 User Manual

Appendixes

Appendix 1,User Guide forCL560 HART Option 136

Appendix 2,Technical Data 150

Appendix 3,Quick Guide for theDatalog Viewer 156

Appendix 4,Index 160

Page 142: MC5 User Manual

136

Appendix 1, User Guide for CL560 HART Option

Appendix 1,User Guide forCL560 HART Option

HART and CL560 ’s Basic Mode 141Connecting to a HART Instrument .....141

Disconnecting a HARTInstrument ....................................142Device Information .......................142Device Settings ............................143Trimming a HART Instrumentin Basic Mode ...............................143Adding a HART Instrument toCL560’s Instrument Database ......145

Calibrating a HART Instrument 146Selecting the Instrument tobe Calibrated ......................................146The Calibration Procedure .................147

Trimming a HART Instrument inCalibration Mode ..........................147

Contents

General 137CL560 and HART ...............................137

Connecting CL560 and a HARTInstrument 138

Supply From the Calibrator ................138External Supply ..................................138Using CL560 as a HARTCommunicator ....................................139Test Diode Connections .....................139

HART Communication Settings 140

Page 143: MC5 User Manual

137

General

HART (Highway Addressable Remote Transducer) is a digital com-munication protocol that uses sine waves superimposed on the stan-dard 4 - 20 mA current. Because the HART signal is small, and itsaverage value is zero it does not affect the accuracy of the analogcurrent signal.

HART is a master-slave communication protocol, which means thatduring normal operation, each slave (field device) communicationis initiated by a master communication device. Two masters (a pri-mary master and a secondary master) can connect to a HART loop.

The HART communication protocol is an open standard owned bythe member companies in the HART Communication Foundation(HCF). For additional information contact:HART Communication Foundation, 9390 Research Boulevard,Suite I-350, Austin, Texas, 8759 USA.Internet: http://www.hartcomm.org

CL560 and HARTCL560 Multifunction Calibrator’s HART communication is a firm-ware option that can be used provided an E module is included inthe CL560 at hand. The HART modem required for HART commu-nication is located in the E module and the terminals of the E mod-ule are used for connecting CL560 with the Hart instrument.

CL560 supports HART Revision 5. Earlier revisions are not sup-ported.

CL560 treats the Analog Output (AO) and the Digital Output (PV,Primary Variable) as separate instruments. To calibrate/trim bothoutput signals, an Analog Output instrument and a Digital Outputinstrument need to be created into CL560’s instrument database.This can be done manually or automatically by communicating withthe HART instrument to be calibrated/trimmed.

General

Page 144: MC5 User Manual

138

Appendix 1, User Guide for CL560 HART Option

Connecting CL560 and a HART Instrument

How the HART instrument is connected to CL560 depends on thesupply power connection and whether there is a test diode con-nected to the transmitter. The following pictures describe the con-nections for each case.

Supply From the CalibratorCL560 supplies thetransmitter and mea-sures the output cur-rent. The HART® termi-nal is used for digitalcommunication. CL560automatically includesthe required 270 ohmresistor.

External SupplyCL560 measures theexternally suppliedoutput current of thetransmitter. TheHART® terminal isused for digital com-munication. Remem-ber to make sure thatthe loop also in-cludes a resistor witha resistance between230 to 600 ohm orthat the impedanceof the loop itself is at least 230 ohms.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

+

I

Input signal X

I

2-w xmtr

+24V

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

+

I

Input signal X

2-w xmtr

Supplyetc.

250 ohm

Imeas/sink

Page 145: MC5 User Manual

139

Test Diode ConnectionsCL560 also supports HART® communication while measuring cur-rent parallel to a test diode in a 20 mA circuit. The external connec-tions vary depending on the diode connection of the transmitter:

Plus Side Test Diode Minus Side Test Diode

Using CL560 as a HART CommunicatorIf you want to useCL560 as a Hartcommunicator withoutmeasuring/controllingthe instrument’soutput current,connect CL560 asshown in the adjacentpicture. Again, makesure that there is aresistor (or enoughresistance) in theloop to enable Hartcommunication.

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

+

I

Input signal X

2-w xmtr

+

-

Imeas/sinkT/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

+

I

Input signal X

2-w xmtr

+

-

Imeas/sink

T/C, Low V4-w meas

R, RTD

3 & 4-w meas

V, I,+24V Imeas/sink

Low V

HART®

T/C INT. RJ

T/C OR EXTWIRES ONLY

2-w xmtr

V, ,

MEASUREOUTPUTSENSOR MEASURE & SIMULATEET E

Max input:60 VDC/30 VAC

+

I

Input signal X Supplyetc.

250 ohm

Connecting CL560 and a HART Instrument

Page 146: MC5 User Manual

140

Appendix 1, User Guide for CL560 HART Option

HART Communication Settings

Before attempting to communicatewith HART instruments it is advis-able to review CL560’s HART com-munication settings.

Open the settings window by press-ingD/Menu,C/Others and3/HART Communication Set-tings

Master Address

HART communication supports theuse of two masters, a Primary anda Secondary master. Select aunique master address for CL560.

Preambles used

Can be a value between 3 to 20.The smaller the value is, the fasterthe communication is, but instru-ments requiring a larger amount ofpreambles may not communicate.

Number of Retries

Can be a value between 1 to 10.Determines the amount of retrieswhen communication errors occur.

06.10.2000 15:30 1 Pressure

2 Current

P1: INT400MC/-41.400...41.400 kPa

E: Current Measurement

gaugekPa0.089

3.9845 mA

SetupWindow 1 Others

MENUCloseWindow 2

Setup

Settings

Maintenance

HARTCommunicationsettings

06.10.2000 15:32 HART COMMUNICATION SETTINGS

Master Address

Preambles usedNumber of Retries

53

Close

Secondary

Page 147: MC5 User Manual

141

HART and CL560’s Basic Mode

Connecting to a HART InstrumentTo communicate with a HART in-strument, press:D/Menu andB/Window 2 Setup, if needed5/HART.

Note.

The CL560 menu option is disabledif the other window already reservesthe E module. Change the setup ofthe other window to enable theCL560 menu option.

06.10.2000 14:37 1 PressureP1: INT400MC/-41.400...41.400 kPa

E: Current Measurement

gaugekPa0.089

3.9855 mA

2 Current

SetupWindow 1 Others

MENUCloseWindow 2

Setup

Quantity[Current]

Function/Port[E:I(Meas)]

Unit[mA]

Step, Ramp

HART

06.10.2000 15:41

0 - 3051 - PRESS TRANSM

1 - PT155 - NC TTR2 - PT311 - ARG JT

3 - PT413 - WBA4 NTB

4 - TT210 - PLGR NSWF

5 - TT921 - FPN ILL

6 - TT148 - TBB NE

Select Back

HART

Searching for HART Device at:

Devices found:

12

Restart

Refer to chapter ConnectingCL560 and a HART Instrument onpage 138 for information on how toconnect CL560 and a HART instru-ment.

CL560 searches for HART devicesconnected to the current loop andopens a window similar to the oneseen to the right.

Up to 16 instruments may be con-nected to the same current loop.Choose the appropriate instrumentby pressing 4/Select (optionallyeither the or the key).

Note.

Measuring the Analog Output (AO) of a HART instrument does not requireany special actions. Just connect the Analog Output to CL560’s currentmeasurement terminals in the E module.

HART and CL560’s Basic Mode

Page 148: MC5 User Manual

142

Appendix 1, User Guide for CL560 HART Option

The standard Window Setup menuis replaced by a HART menu. Thismenu is automatically opened af-ter the HART instrument is se-lected. All HART menu options aredescribed in the subsequent chap-ters. The measurement quantity ofthe window where the HART instru-ment was selected to is automati-cally changed to the quantity of theHART instrument’s Digital Output(PV).

Disconnecting a HART Instrument

The HART instrument is to be disconnected when you need tochange the Quantity of a window assigned for HART measure-ment.

To disconnect a HART instrument, selectD/Menu andB/Window 2 Setup, if needed1/Disconnect Device.

The HART communication is terminated and the HART menu isreplaced by the standard Window Setup menu.

Device Information

The adjacent picture displays thedata of the Device Info window.The following fields are editable:

Tag,Serial Number,Date,Descriptor andMessage.

06.10.2000 15:571 PressureP1: INT400MC/-41.400...41.400 kPa

E: HART/PV

gaugekPa0.087

0.084

tila Kalibrointi

kenttäValitse Valikko

kPa

2 Pressure

Disconnect Device

Device Information

Device Settings

Unit[kPa]

Trim Instrument

Add HARTInstrumentto Database

OthersMENUClose

SetupWindow 1 Window 2

Setup

06.10.2000 16:01 HART DEVICE INFORMATION

Tag

Device ID

Date

1109949

07.06.1999

Close

3051C

Descriptor NO SMOKE

Message JUST TESTING

Serial Number 7292301Manufacturer RosemountDevice Type 3051CSoftware Version 5.3.176Hardware Revision 1.0

Page 149: MC5 User Manual

143

Device Settings

The adjacent picture displays thedata of the Device Settings window.The following fields are editable:

Tag,Polling Address,PV/Range Unit,Lower Range,Upper Range,Damping,Transfer Function,Burst Mode andSensor Serial Number.

Trimming a HART Instrument in Basic Mode

Note that CL560’s Calibration Mode also includes a possibility totrim HART instruments. Trimming a HART Instrument in Basic Modeis only needed when you do not intend to calibrate and trim theHART instrument.

To start trimming the HARTinstrument, press:D/Menu andB/Window 2 Setup, if needed6/Trim Instrument.

From the pop-up list, select whichoutput, analog or digital, is to betrimmed. CL560 reads the instru-ment data while displaying the fol-lowing message:Reading HART instrument data.Please wait.

06.10.2000 16:09 DEVICE SETTINGS

Tag

Polling Address 0

Close

3051C

OutputPV/Range Unit mbarLower Range 0.00Upper Range 1000.00Damping (s) 0.40Transfer Function LinearBurst Mode Off

SensorSerial Number 1109949Lower Limit 0.00Upper Limit 2486.41Minimum Span 24.87

01.10.2000 16:141 PressureP1: INT400MC/-41.400...41.400 kPa

E: HART/PV

gaugekPa0.088

0.084

tila Kalibrointi

kenttäValitse Valikko

kPa

2 Pressure

Close

Disconnect Device

Device Information

Device Settings

Unit[kPa]

Trim Instrument

Add HARTInstrumentto Database

Analog Output (AO)Digital Output (PV)

HART and CL560’s Basic Mode

Page 150: MC5 User Manual

144

Appendix 1, User Guide for CL560 HART Option

For the Digital Output (PV), CL560 opens the Instrument Input Win-dow for checking/editing the settings. Use B/Next Page FunctionKey to check/edit the Instrument Output Window data.

Selecting the A/Trim Function Key opens the Calibration Windowswith the HART instrument’s trim menu opened. To change to Ana-log Output (AO), use the menu option 4/Select Output.

Warning!

Do not alter any settings unlessyou have the knowledge and areauthorized to trim the HART in-strument.

Note.

Some of the menu options may be dis-abled when trimming certain HART in-struments. Refer to the HARTinstrument’s manual for device depen-dent information concerning the trim-ming procedure.

06.10.2000 16:28INSTRUMENT INPUT

Input Method Measured

Unit mbar

Range 0.000000

Pressure

Port 1 INT1C

Pressure type Gauge

0 %1000.0000100 %

TrimpageNext Edit

19.04.2000 15:29INSTRUMENT OUTPUT

Output Method

Range 4.000000

Current

Port E

Unit mA

0 %20.000000100 %

Transfer Function Linear

Measured

TrimpageNext Edit

29.05.2002 15:42 Input

Error1.00

Output

kPa

4.3362

-0.04 % of span

0.0433

+

0

-0% 50% 100%

Pressure [P1:INT20C]

Current [E: Meas]

mA

gaugeTrim Lower0% Range

Trim Upper100% Range

Check Any Value

DeviceInformation

Device Settings

SelectOutput[AO]

End HARTAdjustment

Analog Output (AO)Digital Output (PV)

Page 151: MC5 User Manual

145

Adding a HART Instrument to CL560’s Instrument Database

CL560 has a semi-automatic utilityto add HART instruments toCL560’s instrument database. Se-lect 7/Add HART instrument toDatabase from the Basic Mode’sHART Menu. Continue by select-ing which output is to be added,analog or digital. The following in-strument data is acquired from theHART instrument:

The following instrument data isacquired from the HART instru-ment:

Input Quantity Input UnitInput Range Input MethodOutput Quantity Output UnitOutput Range Output MethodTransfer Function Device IDSetpoint Delay

When the instrument is created,CL560 automatically opens the In-strument Window, from where it iseasy to either check/edit instrumentdata or start calibrating the instru-ment. It is recommended to edit theinstrument data and carefully checkthe default values assigned for in-strument data fields that do nothave corresponding fields in HARTinstrument data. To edit the instru-ment data, press the 2/Edit button.

At least the following instrument data fields should be checked:

Error Calculation Method Reject if > Input PortCalibration Method Calibration Points Output Port

Keep in mind that changing the value of some of the above men-tioned fields may also require changes in related fields.

Note.

The HART instrument’s TAG is assigned to the Device ID field in CL560. NoPosition ID is automatically assigned to the instrument.

11.04.2002 9:571 PressureP1: INT400MC/-41.400...41.400 kPa

E: HART/PV

gaugekPa1.1

1.142 kPa

2 Pressure

Disconnect Device

Device Information

Device Settings

Unit[mbar]

Trim Instrument

Add HARTInstrumentto Database

Analog Output (AO)Digital Output (PV)

SetupWindow 2

MENUClose

SetupWindow 1 Others

06.10.2000 16:57 INSTRUMENT

112-TT-003.1

Edit Calibrate MENU

JUST TESTING

Cal. Points 3 ↑

Transfer Function Linear

Method Measured

INPUT 0.00 ... 1000.00 mbar

Pressure Type Gauge

Method Measured

OUTPUT 4.0000 ... 20.0000 mA

Back

HART and CL560’s Basic Mode

Page 152: MC5 User Manual

146

Appendix 1, User Guide for CL560 HART Option

Calibrating a HART Instrument

Selecting the Instrument to be CalibratedTo be able to calibrate the HARTinstrument, it has to be added toCL560’s instrument database. Theanalog output (AO) and the digitaloutput (PV) are treated as separateinstruments.

Adding the instruments can bedone using the utility described inchapter Adding a HART Instru-ment to CL560’s Instrument Da-tabase on page 145 or by sendingthe instrument data from a calibra-tion software. Additionally: you canalso manually enter the instrumentdata into CL560’s instrument data-base.

Select the instrument to be cali-brated from CL560’s list of availableinstruments.

If you are already connected to theHART instrument (HART commu-nication started in Basic Mode),CL560 continues directly from theInstrument Window to the Calibra-tion Windows.

If HART communication is notstarted and you are calibrating thedigital output (PV) of the HART in-strument, CL560 prompts you tostart the communication and selectthe instrument in a window similarto the adjacent picture.

CL560 accepts the HART instrument as the instrument to be cali-brated only if its instrument data matches with the data of the instru-ment selected in CL560.

06.10.2000 17:13 INSTRUMENT

112-TT-003.1

Edit Calibrate MENU

JUST TESTING

Transfer Function Linear

Method Measured

INPUT 0.00 ... 1000.00 mbar

Pressure Type Gauge

Method Measured

OUTPUT 4.0000 ... 20.0000 mA

Back

Cal. Points 5 ↑↓

06.10.2000 17:16

0 - 3051 - PRESS TRANSM

1 - PT155 - NC TTR

2 - PT311 - ARG JT

3 - PT413 - WBA4 NTB

4 - TT210 - PLGR NSWF

5 - TT921 - FPN ILL

6 - TT148 - TBB NE

Select Back

HART

Searching for HART Device at: 8

Devices found:

Restart

Page 153: MC5 User Manual

147

The Calibration ProcedureCalibrating a HART instrumentdoes not differ from the calibrationof a non-HART instrument withsimilar input quantity, input method,output quantity and output method.Refer to the examples in CL560User Guide’s Part D.

Note.

Calibrating the Analog Output (AO) ofa HART instrument does not necessar-ily require HART communication.CL560 treats it as a standard analoginstrument during the calibration proce-dure.

Trimming a HART Instrument in CalibrationMode

There is a special utility for trimminga HART instrument duringcalibration. To start the HARTadjustment, select D/MENU and1/Start HART Adjustment.

06.10.2000 17:21 Input

Error1.00

Output

mbar

510.24

-0.21 % of span

512.156

+

0

-0% 50% 100%

Pressure [P1:INT1C]

Pressure [E: HART]

mbar

Gauge

StopReadings

Accept

15.05.2002 13:22 Input

Error1.00

Output

kPa

4.3374

-0.06 % of span

0.0434

+

0

-0% 50% 100%

Pressure [P1:INT20C]

Current [E: Meas]

mA

gauge

MENUClose

Prescan[On]

Start HARTAdjustment

ViewCalibrationResults

ZeroPressureModule

Calibrating a HART Instrument

Page 154: MC5 User Manual

148

Appendix 1, User Guide for CL560 HART Option

How CL560 continues depends on the instrument previously se-lected for calibration:

• If the output signal of the previously selected instrument is thedigital output of a HART instrument, the instrument input datawindow, shown as the leftmost picture below is opened imme-diately.

• For all other output signals CL560 prompts you to start thecommunication and select the instrument to be calibrated (re-fer to the picture in chapter Selecting the Instrument to beCalibrated on page 146. Then the instrument input data win-dow shown below is opened.

Press B/Next Page Function Key to check/edit instrument outputdata and then A/Trim to start the adjustment.

CL560 opens the HART Adjustmentmenu with the Digital Output (PV)selected. To change to Analog Out-put (AO), use the menu option4/Select Output.

Warning!

Do not alter any settings unlessyou have the knowledge and areauthorized to trim the HART in-strument.

06.10.2000 16:28INSTRUMENT INPUT

Input Method Measured

Unit mbar

Range 0.000000

Pressure

Port 1 INT1C

Pressure type Gauge

0 %1000.0000100 %

TrimpageNext Edit

19.04.2000 15:29INSTRUMENT OUTPUT

Output Method

Range 4.000000

Current

Port E

Unit mA

0 %20.000000100 %

Transfer Function Linear

Measured

TrimpageNext Edit

29.05.2002 15:42 Input

Error1.00

Output

kPa

4.3362

-0.04 % of span

0.0433

+

0

-0% 50% 100%

Pressure [P1:INT20C]

Current [E: Meas]

mA

gaugeTrim Lower0% Range

Trim Upper100% Range

Check Any Value

DeviceInformation

Device Settings

SelectOutput[AO]

End HARTAdjustment

Analog Output (AO)Digital Output (PV)

Page 155: MC5 User Manual

149

When trimming the Digital Output(PV), the Input window displaysthe instrument’s input signal mea-sured by CL560. The Output win-dow displays the digital output ofthe instrument.

The lower part of the Output win-dow includes an additional trimfield. Use either the C/Fetch Func-tion Key to copy the value shownin the input window or manuallyenter a value the digital outputshould be trimmed to. Then usethe D/Send Function Key.

When trimming the Analog Output(AO), the Input window displaysthe current set point communicatedby CL560. The Output window dis-plays the current generated by theinstrument and measured byCL560.

Again, the lower part of the Out-put window includes an additionalfield. Use either the C/Fetch Func-tion Key to copy the value shownin the Output window or manuallyenter a value that is the correctoutput current. Then use theD/Send Function Key.

Note.

Some of the menu options may be disabled when trimming certain HARTinstruments. Refer to the HART instrument’s manual for device dependentinformation concerning the trimming procedure.

16.05.2002 15:58Input

Error1.00

Output

bar

1.021

-0.04 % of span

1.029

+

0

-0% 50% 100%

Pressure [P1:INT20C]

Pressure [E: HART \ PV]

bar

Trim 0% to:

TrimEnd Fetch MENU

Gauge

1.029

20.08.2002 15:44 Input

Error1.00

Output

mA

3.9982

N/A % of span

4.0000

+

0

-0% 50% 100%

Current [Keyed]

Current [E: Meas.]

mA

Enter measured 0%

TrimEnd Fetch MENU

3.9982

Calibrating a HART Instrument

Page 156: MC5 User Manual

150

Appendix 2, Technical Data

Appendix 2,Technical Data

ContentsCL560 General Specifications 151

Modules 152Pressure Modules (INT & EXT) ......... 152Temperature Electrical Module (ET) .. 153

RTD Measurementand Simulation ............................. 153Thermocouple Measurementand Simulation ............................. 154Reference JunctionModule (RJ) ................................. 154

Electrical Module (E) ......................... 155

Page 157: MC5 User Manual

151

GeneralDisplay 96 x 72 mm (3.78" x 2.83"), 320 x 240 pixels, back lit LCDWeight 1.7 - 2.3 kg (3.7 - 5.1 lbs)Dimensions 245 mm (9.6") x 192 mm (7.5") x 74 mm (2.9") (d/w/h)Case protection IP65 (dust and water proof)Keyboard Membrane protected individual keysBattery type Rechargeable NiMH, 4000 mAh, 7.2V DCBattery operation Average 10 hoursCharger supply 100...240 VAC, 50-60 HzOperating temperature -10...50°C (14...122°F)Specifications valid 15...35°C (59...95°F)Temperature coefficient < ±0.001% RDG / °C outside of 15...35°C (59...95°F)Storage temperature -20 to 60°C (-4 to 140°F)Humidity 0 to 80% R.H. non condensingMeasurement sample rate 2.5 / secondWarranty 3 years as standard, battery pack 1 year

CL560 General Specifications

CL560 General Specifications

Page 158: MC5 User Manual

152

Appendix 2, Technical Data

Modules

Pressure Modules (INT & EXT)

1) Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long term stability for men-tioned period. (k=2).

2) Every internal/external pressure module's range may be displayed also in absolute pressure if the Barometric Module (B) isinstalled.

Supports the following pressure units as standard: Pa, hPa, kPa, MPa, mbar, bar, lbf/ft2, psi, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, at,mmH2O, cmH2O, mH2O, iwc, ftH2O, mmHg, cmHg, mHg, inHg, mmHg(0°C), inHg(0°C), mmH2O(4°C), inH2O(4°C), ftH2O(4°C),inH2O(60°F), mmH2O(68°F), inH2O(68°F), ftH2O(68°F), torr, atm.

Pressure modules 20 bar and below, pressure connection G 1/8 (ISO 228/1) 60° internal cone.INT60, INT100, INT160 module pressure connection G 1/8 (ISO 228/1) female.EXT60, EXT100, EXT160, EXT250, EXT600, EXT1000 pressure module pressure connection G 1/4 (ISO 228/1) male.Wetted parts AISI316 stainless steel, Hastelloy, Nitrile rubber.

In the INT20C, EXT20C, INT60, EXT60, INT100, EXT100, INT160, EXT160 and EXT250 the maximum overpressure is twice therange.

The maximum overpressure for EXT600 is 900 bar and for EXT1000 1100 bar.

InternalModules

ExternalModules

Unit Range(2 Resolution 90 Days Uncertainty(1 (±) 1 Year Uncertainty(1 (±)

INT B EXT B kPa a mbar a psi a

80 to 120 800 to 1200 11.6 to 17.4

0.01 0.1 0.001

- - -

0.5 kPa0.6 0.5 mbar0.7 0.0073 psi

INT100m

EXT100m

kPambar iwc

0 to 100 to 100

0 to 40

0.00010.001 0.001

0.025% RDG + 0.025% FS

0.04% RDG + 0.025% FS

INT400mC

EXT400mC

kPambar iwc

±40±400 ±160

0.0010.01 0.001

0.025% RDG + 0.02% FS

0.04% RDG + 0.02% FS

INT1C

EXT1C

kPabar psi

±100±1

-14.5 to 15

0.0010.00001 0.0001

0.025% RDG + 0.01% FS

0.04% RDG + 0.01% FS

INT2C EXT2C

kPabar psi

-100 to 200-1 to 2

-14.5 to 30

0.0010.00001 0.0001

0.025% RDG + 0.01% FS

0.04% RDG + 0.01% FS

INT6C

EXT6C

kPabar psi

-100 to 600-1 to 6

-14.5 to 90

0.010.0001 0.001

0.025% RDG + 0.01% FS

0.04% RDG + 0.01% FS

INT20C

EXT20C

kPabar psi

-100 to 2000-1 to 20

-14.5 to 300

0.010.0001 0.001

0.025% RDG + 0.01% FS

0.04% RDG + 0.01% FS

INT60

EXT60

kPabar psi

0 to 60000 to 60

0 to 900

0.10.001 0.01

-- -

0.04% RDG + 0.01% FS

INT100

EXT100

MPabar psi

0 to 100 to 100

0 to 1500

0.00010.001 0.01

-- -

0.04% RDG + 0.01% FS

INT160

EXT160

MPabar psi

0 to 160 to 160

0 to 2400

0.00010.001 0.01

-- -

0.04% RDG + 0.013% FS

-

EXT250

MPabar psi

0 to 250 to 250

0 to 3700

0.0010.01 0.1

-- -

0.04% RDG + 0.015% FS

-

EXT600

MPabar psi

0 to 600 to 600

0 to 9000

0.0010.01 0.1

-- -

0.04% RDG + 0.015% FS

-

EXT1000

MPabar psi

0 to 1000 to 1000

0 to15000

0.0010.01 0.1

-- -

0.04% RDG + 0.015% FS

Page 159: MC5 User Manual

153

Temperature Electrical Module (ET)

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability formentioned period. (k=2)

2) Load effect < 5 µV/mA. Maximum output current 5 mA.3) Load effect < 100 µV/mA. Maximum output current 10 mA (0 … 10 V), 3 mA (10 … 12 V).4) Maximum load impedance 400 ohm.5) Amplitude range 0 … 12 Vpp. Amplitude setting accuracy up to 5 kHz ±(200 mV + 5% of set value). Waveforms: Square wave

(positive / symmetric) and sinewave (above 40 Hz).6) Pulse generation frequency range 0.1 ... 1000 Hz. Amplitude setting 0 ... 12 Vpp.7) Valid with measurement current 0.2 ... 5 mA (1 ... 1000 ohm), 0.1 .. 1mA (1 ... 4 kohm). Ohm/RTD simulation speed 1 ms.8) Whichever is greater.9) Specification valid with 4 wire connection. In 3 wire connection add 10 mohm.10) Bias current < 10 nA.

RTD Measurement and Simulation

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for men-tioned period. (k=2).

RTD types available as standard:

RTD/ohm simulation excitation current 0.2 … 5 mA (1 … 1000 ohm), 0.1 … 1 mA (1 ... 4 kohm).Also other RTD types available as option

• Pt50 (385) • Pt500 (385) • Pt100 (391)• Ni100 (618) • Pt100 (385) • Pt1000 (385)• Pt100 (3923) • Ni120 (672) • Pt200 (385)• Pt100 (375) • Pt100 (3926) • Cu10 (427)• Pt400 (385) • Pt100 (389)

Function Range Resolution 1 Year Uncertainty(1

mV generation(2 ± 500 mV 0.001 - 0.01 mV 0.02 % RDG + 4 µV V generation(3 ± 12 V 0.00001 - 0.0001 V 0.02 % RDG + 0.1 mV mA generation(4 ± 25 mA 0.0001 mA 0.02 % RDG + 1 µA Hz generation(5 0.00028 to 50000 Hz 0.000001 - 0.1 Hz 0.01 % RDG Pulse generation(6 0 to 9 999 999 pulssia 1 pulssi N/A Ohm simulation(7 1 to 4000 ohm 0.01 - 0.1 ohm 0.04 % RDG or 30 mohm(8

Ohm measurement(9 0 to 4000 ohm 0.001 - 0.1 ohm 0.02 % RDG + 3.5 mohm mV measurement(10 ± 500 mV 0.001 - 0.01 mV 0.02 % RDG + 4 µV

Function Range (°C) Range (°C) Measurement 1 Year Uncertainty(1 (±)

Simulation 1 Year Uncertainty(1 (±)

Pt-sensors -200 to 850°C -200 to 0°C 0.06°C 0.1°C 0 to 850°C 0.025% RDG + 0.06°C 0.025% RDG + 0.1°C

Modules

Page 160: MC5 User Manual

154

Appendix 2, Technical Data

Thermocouple Measurement and Simulation

Resolution 0.01°C.With internal reference junction (module RJ) add 0.1°C uncertainty.

Thermocouple types C 3) (ASTM E 988 - 96), G 3) (ASTM E 1751 - 95e1) and D 3) (ASTM E 988 - 96) also available as standard.Also other thermocouple types available as option.1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for

mentioned period. (k=2)2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.13) ±(0.02 % RDG + 4) µV4) DIN 43710

Reference Junction Module (RJ)

1) Uncertainty includes reference standard uncertainty, hyster-esis, nonlinearity, repeatability and typical long term stabilityfor mentioned period. (k=2)

Type Range (°C) Range (°C) 1 Year Uncertainty(1 (±) B(2 0 ... 1820 0 ... 200

200 ... 500 500 ... 800 800 ... 1820

(3

2.0 °C 0.6 °C 0.8 °C

R(2 -50 ... 1768 -50 ... 0 0 ... 150 150 ... 1400 1400 ... 1768

1.0 °C 0.7 °C 0.5 °C 0.5 °C

S(2 -50 ... 1768 -50 ... 0 0 ... 50 50 ... 1500 1500 ... 1768

1.0 °C 0.7 °C 0.6 °C 0.7 °C

E(2 -270 ... 1000 270 ... -200 -200 ... 0 0 ... 600 600 ... 1000

(3

0.08 % RDG + 0.07°C 0.015 % RDG + 0.07°C 0.026 % RDG

J(2 -210 ... 1200 -210 ... -200 -200 ... 0 0 ... 1200

(3

0.07 % RDG + 0.08°C 0.02 % RDG + 0.08°C

K(2 -270 ... 1372 -270 ... -200 -200 ... 0 0 ... 1000 1000 ... 1372

(3

0.1 % RDG + 0.1 °C 0.02 % RDG + 0.1 °C 0.03 % RDG

N(2 -270 ... 1300 -270 ... -200 -200 ... -100 -100 ... 0 0 ... 750 750 ... 1300

(3

0.2 % RDG 0.05 % RDG + 0.15°C 0.01 % RDG + 0.15°C 0.03 % RDG

T(2 -270 ... 400 -270 ... -250 -250 ... -200 -200 ... 0 0 ... 400

(3

0.7 °C 0.1 % RDG + 0.1°C 0.01 % RDG + 0.1°C

U(4 -200 ... 600 -200 ... 0 0 ... 600

0.1 % RDG + 0.15°C 0.01 % RDG + 0.15°C

L(4 -200 ... 900 -200 ... 0 0 ... 900

0.07 % RDG + 0.13°C 0.02 % RDG + 0.13°C

Range (°C) 1 Year Uncertainty(1 (±) -10 … 50 °C 0.1 °C

Page 161: MC5 User Manual

155

Electrical Module (E)

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability formentioned period. (k=2)

2) Bias current <10 nA3) Impedance >1 Mohm4) Impedance < 7.5 ohm5) Impedance > 1 Mohm. Frequency measurement minimum amplitude 0.5 Vpp (< 5 kHz),

1 Vpp (5...50 kHz). Pulse counting minimum amplitude 0.5 Vpp (pulse length > 100 µs),1 Vpp (pulse length 100 µs...10 µs).Trigger level range -1…+15 V.

6) Maximum load impedance 800 ohm

All specifications are subject to change without prior notice

Function Range Resolution 1 Year Uncertainty(1

mV measurement(2 ±1000 mV 0.001 - 0.01 mV 0.02 % RDG + 5 µVV measurement(3 ±50 V 0.00001 - 0.001 V 0.02 % RDG + 0.25 mVmA measurement(4 ±100 mA 0.0001 - 0.001 mA 0.02 % RDG + 1.5 µAHz measurement(5 0.0028 to 50000 Hz 0.000001 - 0.1 Hz 0.01 % RDGPulse counting(5 0 to 9 999 999 pulses 1 pulse N/AmA generation(6 0 to 25 mA 0.0001 mA 0.02 % RDG + 1.5 µA

Modules

Page 162: MC5 User Manual

156

Appendix 3, Quick Guide for the CL560 Datalog Viewer

Appendix 3,Quick Guide for theCL560 Datalog Viewer

ContentsCL560 Datalog Viewer 157General ............................................. 157Installing CL560 Datalog Viewer ....... 157

Communication Settings ............. 157File Management .............................. 157

Transferring data from CL560 ...... 157Saving Data on Hard Disk ........... 158Opening Existing Data Files ........ 158Importing and Exporting .............. 158

Other Utilities ..................................... 158Chart Properties .......................... 158Channel Properties ...................... 159Zooming ...................................... 159Printing the Graph ....................... 159

Page 163: MC5 User Manual

157

CL560 Datalog Viewer

GeneralCL560 Datalog Viewer is a utility for transferring Data Logging re-sults from CL560 to a personal computer (PC). The results can beviewed in CL560 Datalog Viewer and saved in the utility’s own for-mat or exported for use in other applications, e.g. Microsoft Excel®.

CL560 Datalog Viewer has two windows. One for displaying theresults in a table and another that displays a graph based on theresults.

The graph and the table have a linked cursor. Double-click any-where inside the graph area and you will see a vertical bar. Thecorresponding data row in the table is highlighted. Similarly, if youmove the highlighted row in the table (by clicking on another row orusing the arrow keys), the vertical bar in the graph window movesto the corresponding location in the graph.

Installing CL560 Datalog ViewerCL560 Datalog Viewer does not require a special installation pro-cedure. Just copy DataLog.exe to a folder that suits your needsand start the software from that location.

Communication Settings

Before connecting to CL560 Multifunction Calibrator, be sure to se-lect the correct Serial Port and Port Speed in the settings menu.

Always start using the maximum Port Speed. If communication prob-lems arise, decrease the port speed.

File Management

Transferring data from CL560

To upload the results, select the Transfer Data button in the toolbar.In the following dialog: Select the data to be downloaded and pressStart.

The data is uploaded to the RAM memory of the PC and is avail-able for viewing in CL560 Datalog Viewer.

CL560 Datalog Viewer

Page 164: MC5 User Manual

158

Appendix 3, Quick Guide for the CL560 Datalog Viewer

Tab Property Description

General Series Visibility Each data channel can be madevisible/invisible. Making a data channelinvisible is also possible through theChannel Property settings, but this is theonly place to make a channel visible againafter it has been set to “invisible”.

Animated Zoom Adds/removes animation effects whenzooming in. More information in chapterZooming.

Axis Axis First select the Axis to be defined.

- Autoscaling When selected, the axis minimum andmaximum is automatically determined.

- Min Axis minimum value. Editable only whenautoscaling is not selected.

- Max Axis maximum value. Editable only whenautoscaling is not selected.

- Axis Number Format The number format used for the selectedaxis. Supported formats are listed at theend of Datalog Viewer’s help file.

- Title Enter a title for the selected axis.

Saving Data on Hard Disk

When saving using the Save or Save As commands, CL560 DatalogViewer saves the results in its own format (*.lg5). This file formatpreserves all the changes in the property settings described in chap-ters Chart Properties and Channel Properties.

Opening Existing Data Files

Selecting the Open icon in the toolbar (or Open in the File menu)opens files in software’s own format (*.lg5). To read data saved inother supported formats, choose the Import option in the File menu.

Importing and Exporting

CL560 Datalog Viewer supports importing and exporting ofWorksheet files (*.csv) and text Files (*.txt). Both file formats arealso supported by several spreadsheet software, e.g. MicrosoftExcel®. When exporting, CL560 Datalog Viewer gives you the pos-sibility to select the CSV file separator (comma or a list separatorbased on Window’s settings).

Other UtilitiesAll of the property settings described in the subsequent chaptersare preserved only when saving data in CL560 Datalog Viewer’sown file format (*.lg5). Other file formats (*.csv and *.txt) save onlythe data, but no property settings.

Chart Properties

Select Chart Properties in the toolbar to edit the following settings:

Page 165: MC5 User Manual

159

Channel Properties

The property setting window for each channel can be opened click-ing on the line in the graph legend.

The following settings are available:

Zooming

To zoom in, drag you mouse from the upper left corner to the lowerright corner of the area to be zoomed in to. The important thing isthat your mouse moves from left to right.

You can scroll the zoomed area by selecting the secondary buttonand dragging desired direction.

To zoom back to the original size, drag your mouse from right toleft with a slight upward or downward movement.

Printing the Graph

To print the graph, select the Print Chart option from the toolbarbutton or use the option in the File menu. A standard print dialogappears for choosing the printer.

Tab Property Description

Paging Paging ON/OFF If paging is selected, the graph is dividedinto several horizontal “pages”.

- Points per Page How many points is displayed per page.

- Current Page A possibility to jump to a certain page.When paging is selected, you can alsoscroll from page to page as well as to thefirst and last page using the arrow iconsdisplayed in the lower left corner of thegraph window.

Title Title Visible Whether a title is visible or not. - Title text The title text to be shown in the graph.

Property Description

Axis control The channel can be assigned either to the left or theright axis. By default all channels are assigned to the leftaxis.

Line color Here you can change the color of the line if the defaultcolor does not suit your needs.

Line width A possibility to change the line width. The default widthis 2.

Series name Give the Channel a more descriptive name.

CL560 Datalog Viewer

Page 166: MC5 User Manual

160

Appendix 4, Index

Appendix 4,Index

22-wire system 102

33-wire system 101

44-wire system 101

AAcknowledging alarms 74Adding Instruments 128Adjustment 108, 127Advanced Utilities

Environment temperaturesensor 79

AlarmsAcknowledging 74Volume setting 77

As Found calibration 108As Left calibration 109Autocapture Feature 117Automatic calibration 116Auto-Off Delay 77

BBag 14Basic Mode, defined 35Batteries 11Battery Pack 13Binary Signal Detection 45Brightness 8

CCalibration

Adjustment 108As Found 108As Left 109Autocapture Feature 117Automatic 116Deleting the results 134Examples 117Generally 106Instrument

adjustment 127Instrument Window 113Maintaining the Instrument

Database 128Manual 116Phases 107Procedure, defined 114Required modules 110Saving the results 115Selecting the

Instrument 112Calibration Method

field 116, 126, 131Calibration Results

Deleting 134Viewing 133

Calibrator configuration,see Settings 76

Carrying case 14Case 14Change Rate 54Charger 12, 30Cleaning CL560 31Compensation loop 102Computer Interface 6

Configuring the Calibrator,see Settings 76

Connectable devices 25Connections 4

AUX Interface 6COMP/PRT Interface 6ENV Interface 6External pressure

modules 5Internal pressure

modules 5Contrast 8Copying an

instrument 129Creating Instruments 128Current generation 60Current measurement 41

DData Logging

Configuring 86Starting 87Transferring data to

a PC 88Viewing Results 88

Deleting Instruments 132Deviation measurement 55Difference

measurement 56Display

Brightness 8Contrast 8

Display Mode 80Error Display 82Percentage 82Scaling 81

Page 167: MC5 User Manual

161

Index

EEditing fields 21Electrical and Temperature

module (ET) 7Electrical limit switch

calibration 124Electrical module (E) 7Environment temperature

sensor 6, 79Error Display 82Error situations in

thermocouplemeasurement 100

Examples of InstrumentCalibration 117

ExternalDevices 6, 25, 89, 126Ametech/Jofra 92Calibration Mode 118DPI510 91HART Scientific 92Isotech 92POC4 91TekKnow 92Use in Basic Mode 90

External pressuremodules 5

External ReferenceJunction 98

FFiltering 55Fine Tuning 59Firmware

Basic Mode 16Editing fields 21Function Keys 19General description 15Help function 17Menus 19Status bar 18Updating 15User Interface 18

Frequency generation 64Frequency

measurement 48Function Keys 19

GGenerating (see also

Simulating)Changing the generated

signal 59Current 60Fine Tuning 59Frequencies 64General description 58Pulses 65Using the E module 60Using the ET

module 60, 62, 64, 65Voltages 62

HHelp function 17

IInput Method field 126,

130Instrument adjustment

support 127Instrument Database 128

CalibrationInstructions Page 132

CalibrationSettings Page 131

Deleting Instruments 132General Data Page 129Instrument

Input Page 130Instrument

Output Page 130Instrument Window 113Internal Pressure

Modules 5Internal Reference

Junction 97

KKeyboard, defined 9

LLimit Switch test 46Line Frequency 77

MMaintenance, see Settings

76Manual calibration 116Maximum value 54CL560 recalibration 30Measuring

Current 41Frequencies 48General description 37Limit Switch test 46Pressure 39Pulses 49Resistance 44Switch State Detection 45Temperatures (RTD

sensor) 50Temperatures (T/C) 51Using the E

module 41, 42, 48, 49Using the ET

module 42, 44, 50, 51Voltages 42

Menus 19Min/Max value 54Minimum value 54Modules 24

NNeck support strap 14

OOperational Sections 4Options 24, 25

Page 168: MC5 User Manual

162

Appendix 4, Index

PParallel functions

in CL560 104Percentage 82Pressure

Module names 96Pressure Types 95Sensor

supply 40, 118, 130Square rooting

instruments 96Pressure

Controllers 91, 126Pressure measurement 39Pressure module,

zeroing 40, 84, 85Pressure sensor/transmit-

ter calibration 118Printer Interface 6Printing 17, 93Pulse generation 65Pulse measurement 49

RRamping 71Recalibration 30Redundant

measurement 57Reference Junction

External 51, 67, 98Internal 8, 51, 67, 97Modes 98Module 8

Required modules whencalibrating aninstrument 110

ResistanceMeasurement 44

Resistance simulation 66Resolution 55RTD sensor simulation 66

SSafety 26Saving calibration

results 115Scaling 81Selecting the instrument to

be calibrated 112

Sensor supply 40, 118, 130Service 30Settings

Auto-Off Delay 77Date Format 77Display Backlight

Brightness 77Language 77Line Frequency 77POC4 91Temperature Scale 77Temperature Unit 77Time and Date 78Time Format 77Use of Environment

Sensor 77Volume settings 77

Simulatinga Switch 85a Transmitter 84

Simulating (see alsoGenerating)Resistance 66RTD sensors 66Thermocouples 67

Simultaneous measure-ment/generation/simulation 104

Software 15Special Display Mode 80Special Generations 69

Ramping 71Stepping 69

Special measurements 53Change Rate 54Deviation

measurement 55Difference

measurement 56Maximum value 54Min/Max value 54Minimum value 54Redundant

measurement 57Special Filtering 55Special Resolution 55

Stand 14Starting CL560 15, 34Status bar 18Stepping 69Support 14

Support for instrumentadjustment 127

Switch Simulation 83Switch State Sensing 45

TTemperature

Controllers 92, 126Temperature indicator

calibration 122Temperature Measurement

RTD sensor 50Thermocouple 51

Temperature recordercalibration 122

Temperature sensorcalibration 120

Test Diode, measurementparallel to a 103

Thermocouplesimulation 67

Transmitter Simulation 83Troubleshooting

thermocouplemeasurement 100

UUnpacking 3Updating the firmware 15Use of Environment

Sensor 77User Interface 18

WWarnings 27Viewing 8Viewing Calibration

Results 133Voltage generation 62Voltage measurement 42Wrist strap 14

ZZeroing a pressure

module 40, 84, 85