Part No. 2296694-01, Rev. 03 Quick Start Guide MC-III TM Flow Analyzer A Quick Reference on Mounting, Wiring, & Configuring the MC-III EXP or WP Flow Analyzer For complete instructions, see MC-III EXP User Manual, Part No. 9A-50165003 MC-III WP User Manual, Part No. 9A-50165009 in the MC-III software Help menu

description

Mc III Quickstart Guide (1)

Transcript of Mc III Quickstart Guide (1)

Page 1: Mc III Quickstart Guide (1)

Part No. 2296694-01, Rev. 03

Quick Start Guide MC-III TM Flow Analyzer

A Quick Reference on Mounting, Wiring, & Configuring

the MC-III EXP or WP Flow Analyzer

For complete instructions, see MC-III EXP User Manual, Part No. 9A-50165003

MC-III WP User Manual, Part No. 9A-50165009in the MC-III software Help menu

Page 2: Mc III Quickstart Guide (1)

2

Pipe plugs

Explosion-proofenclosure

3/4-in. to 1-in.pipe reducer*

1-in. pipe union*

Cableassembly

Magnetic pickup

Turbine flowmeter

MC-III EXP1. PositiontheMC-IIIEXPabovetheflowmeterpickupadapter.2. PlugtheconnectoroftheMC-IIIEXPcableintothemagneticpickupand

hand-tightentheknurlednutontheconnector.3. MounttheMC-IIIEXPontheflowmeterpickupadapterwiththedisplay

facingthedesireddirection,andtightenallsectionsofthepipeunion.BeforemountinganATEX-approveddevice,see“MC-IIIEXP(ATEX)”below.

Caution: Do not use Teflon tape on threads of the union, adapter, or pipe plugs.

Mounting Your MC-III

ATEX-approvedstandoff tubeadapter

MC-III EXP (ATEX)ATEX-approvedMC-IIIEXPsarefittedwithastandofftubeadapterwhenpur-chasedwithanATEX-approvedBarton7000SeriesorNuFloturbinemeter.

Toavoiddamagingthesignalcableduringinstallation,makesurethesignalcableisdisconnectedfromtheMC-IIIcircuitboardbeforemountingtheMC-IIItoaflowmeter.

*Pipereducerandunionareusedin CSA-approvedMC-IIIdevicesonly.

Page 3: Mc III Quickstart Guide (1)

3

MC-III WP1. PositiontheMC-IIIWPabovetheflowmeter.2. PlugtheMC-IIIWPcableconnectorintothemagneticpickupandhand-

tightentheknurlednutontheconnector.3. ScrewtheMC-IIIWPmountontotheflowmeterthreadssurroundingthe

magneticpickup.4. Withthedisplayfacing

thedesireddirection,tightenthetwoscrewsoneithersideofthemounttoprevent horizontalshifting.

5. Withthedisplay orientedverticallyinthedesireddirection,tightenthetwohex-headboltsoneithersideofthetopsectionofthemounttopreventverticalshifting.

Page 4: Mc III Quickstart Guide (1)

4

CAUTION—All field wiring must conform to the National Electrical Code, NFPA 70, Article 501-4(b) for installations within the United States or the Canadian Electric Code for installations within Canada. Local wiring ordinances may also apply. All field wiring must have a wire range of 22 to 14 AWG and terminal block screws must be tightened to a minimum torque of 5 to 7 in-lb to secure the wiring within the ter-minal block. Only personnel who are experienced with field wiring should perform these procedures.

Wiring Your MC-III

Ground screw

Pipe plugs

Ground screws

MC-III EXP

FieldwiringenterstheMC-IIIEXPenclosurethrougheitheroftwothreadedconduitopeningsinthetopoftheenclosure.Whennotinuse,theopeningsaresealedwithpipeplugs.

MC-III WP

FieldwiringenterstheMC-IIIWP enclosurethroughconduithubsinthebottomoftheenclosure.Onehubissuppliedandtwoadditionalentrancesarepluggedbutavail-ableforusewithcustomer-suppliedhubs).Theflowmetercableisroutedthroughacordconnectorinthebackofthebottomenclosurepanel.

ForDiv.2installations,aground wiremustbeconnectedtothe internalgroundscrewinsidethe enclosure.

ForDC-poweredinstallations,route thegroundconductorintothe enclosurewiththeincomingpower conductors.

Page 5: Mc III Quickstart Guide (1)

5

6-30 VDC

TB1

PULS

EIN

PUT

RESE

TIN

PUT

TFM

A&SGND

TB3

EXT POWER

TURBINEMAGNETIC PICKUPVMAX = 3.9 V

J1

J2

RESETSWITCH

BATT

ERY

A

B

BATTERY CONNECTION

Caution: Never use external power with a 4-20 mA output.

1. Ifthebattery/batterypackisnotalreadyconnectedtothecircuitboard, attachthebatterycabletoconnectorJ1.

2. ConnecttheflowmetersignalcabletoterminalblockTB1.

3. Ifanexternalpowersupplyistobeused(thebatterywillprovidebackuppower),wirethepowersupplytoterminalblockTB3.

ForMC-IIIWPDiv.2installations,connectaprotectiveovercurrentdeviceratedat0.5Amaximum(circuitbreakerorfuse)tothepositivesupplylineoftheDCpowersupplyinthesafearea.Adisconnectswitchforthepowersupplymustalsobeinstalledinthesafearea.

AZenerdiode(PartNo.1.5KE33CA)mustbeinstalledforCEapproval.

TB3

TB1

PULS

EIN

PUT

RESE

TIN

PUT

TFM

A&SGND

EXT POWER

TB2

RS485SLAVE

4-20OUT

PULSEOUT

6-30 V

DC

POWER SUPPLY6 to 30 VDC

J1

J2

RESET SWITCH

BATT

ERY

GROUND SCREWREQUIRED FORDIV. 2 INSTALLATIONS

TB3 MARKING:GND = SIGNAL GROUND

Page 6: Mc III Quickstart Guide (1)

6

TB1

PULS

EIN

PUT

RESE

TIN

PUT

TFM

A&SGND

TB3

EXT POWER

TB2

RS485SLAVE

4-20OUT

PULSEOUT

6-30 VDC

J1

J2

RESET SWITCH

BATT

ERY

Leave this end of shield disconnected.

POWER SUPPLY5 to 50 VDC

*

Resistor may be included in pulse readout device. Size the resistor to limit the current to 60 mA.

*

PULSE READOUTDEVICE

TB1

PULS

EIN

PUT

RESE

TIN

PUT

TFM

A&SGND

TB3

EXT POWER

TB2

RS485SLAVE

4-20OUT

PULSEOUT

6-30 V

DC

J1

J2

RESET SWITCH

BATT

ERY

POWER SUPPLY8 to 30 VDC

LOAD

Resistor may be included in readout device.

4-20 mA and flowmeter frequency (amp & square) cannot be used simultaneously.

*

*

4. Ifapulseoutputistobeused,wirethepowersupplyandthepulsereadoutdevicetoterminalblockTB2.Themaximumcurrentratingofthepulseoutputcircuitis60mAat50VDC.

AZenerdiode(PartNo.1.5KE56CA)mustbeinstalledforCEapproval.

5. Ifa4-20mArateoutputistobeused,wirethepowersupplytoterminalblockTB2.Thepowersupplyvoltagerequiredtopowerthecurrentloopdependsontheloopresistance.SeeanMC-IIIusermanualfordetails.

Caution: Never use external power with a 4-20 mA output.

Wiring Your MC-III

Page 7: Mc III Quickstart Guide (1)

7

6. Ifaflowmeterfrequencyoutputistobeusedtoprovideflowrateand/ortotalinformationtoperipheralequipment,wirethepowersupplyandthefrequencyreadoutdevicetoterminalblockTB3.Thereadoutdevicemustoperatewith50mAorless.

TB1

PULS

EIN

PUT

RESE

TIN

PUT

TFM

A&SGND

TB3

EXT POWER

TB2

RS485SLAVE

4-20OUT

PULSEOUT

6-30 VDC

J1

J2

RESETSWITCH

BATT

ERY

Leave this end of shield disconnected.

POWER SUPPLY5 to 30 VDC

FREQUENCYREADOUT DEVICE

*

Resistor may be included in frequencyreadout device. Size the resistor to limit the current to 50 mA.

4-20 mA and flowmeter frequency(amp & square) cannot be usedsimultaneously.

*

TB3 MARKINGS:GND = SIGNAL GROUNDA&S = AMP & SQUARE

Caution: When using the flowmeter frequency output and powering the device from an external power supply, make sure both power supplies share a common negative (-) terminal or are totally isolated from each other.

7. ToconfiguretheMC-IIIwithsoftware,connecttheMC-IIItoacomputerbywiringaconvertercabletotheRS-485terminalsofTB2.Seepage12forwiringdiagrams.

Page 8: Mc III Quickstart Guide (1)

8

Configuring with the KeypadMostroutinefunctionscanbeconfiguredfromthesix-buttonkeypad.

Thisguideprovidesinstructionsforenteringacalibrationfactor,andselectingtheunitsforthetotalandflowratedisplays.Forinstructionsonconfiguringallotherparameters,seetheMC-IIIusermanuals.

Page 9: Mc III Quickstart Guide (1)

9

Parameterchangestypicallyrequirefourbasicsteps:

1. Selectamenu(K-FACTOR,OUTPUT,ORDISPLAY).2. PressUPARROWtochangetheparameterorvaluedisplayed.3. PressLEFTARROWtoadvancetothenextconfigurable

parameterorvalue.4. PressENTERtosavethenewconfigurationsetting.

Enter a Calibration Factor

00000000M3

BBLGALLIT

CFCF

X1000X1000

Select a K-Factor Unit.PressK-FACTORMENU.

K-FACTORMENU

PressUPARROWuntilthecorrectunitisdisplayed.

TEST

Select a Decimal Point Position.PressLEFTARROWtoselectthedecimalpoint.

LOG

PressUPARROWtochangethepositionofthedecimalpoint.

TEST

PressLEFTARROWtosavethedecimalpointselectionand proceedwithenteringacalibra-tionfactor.

LOG

Enter a Calibration Factor.PressUPARROWuntilthelastdigitofthecalibrationfactorisdisplayed.

TEST

ThenpressLEFTARROWtoselectthenextdigittotheleft.RepeatusingUPandLEFTarrowstoenterremainingdigits.

LOG

PressENTERtoadvancetotheInputSensitivitymenu.

SAVE

ENTER

Enter an Input Sensitivity Setting.PressUPARROWtoselectlow,medium,orhigh(foraturbinemeterinputsignal)or“pulseinput”(forapreamplifierinputsignal).

TEST

PressENTERtosaveyourselections.

SAVE

ENTER

Page 10: Mc III Quickstart Guide (1)

10

Select a Unit of Measure for Volume

00000000000000

M3

BBL

GAL

LIT

M3

BBL

GAL

LIT

/SEC

/MIN

/HR

/DAY

CF

CF

X1000

X1000STANDARD

Select a Volume Unit.PressDISPLAYMENU. DISPLAY

MENU

PressUPARROWuntilthecorrectunitisdisplayed.Note—Ifacalculateddivisorwasentered,selectuser-defined(nounitsvisible).

TEST

Toreadthevolumeintermsofthousandsofunits(ex.1.0=1,000bbl),continuepressingUPARROWuntilboththeunitofchoiceandtheX1000optionaredisplayed.

TEST

Select a Decimal Point Position.PressLEFTARROWuntilthedecimalpointintheTotaldisplaybeginsblinking. LOG

PressUPARROWtochangethepositionofthedecimalpoint.TEST

PressENTERtosaveyourselections.SAVE

ENTER

Configuring with the Keypad

Page 11: Mc III Quickstart Guide (1)

11

Select a Unit of Measure for Rate

00000000000000

M3

BBL

GAL

LIT

M3

BBL

GAL

LIT

/SEC

/MIN

/HR

/DAY

CF

CF

X1000

X1000STANDARD

00000000000000

M3

BBL

GAL

LIT

M3

BBL

GAL

LIT

/SEC

/MIN

/HR

/DAY

CF

CF

X1000

X1000STANDARD

Select the Volume Portion of the Flow Rate Unit.PressDISPLAYMENU. DISPLAY

MENU

PressLEFTARROWuntilavolumeunitofmeasureisdisplayedintheratedisplayportionoftheLCDwindow.TheseunitsarelocatedtotheleftorjustabovetheRatedisplay.

LOG

PressUPARROWuntilthedesiredunitofmeasureisdisplayed.Note—Ifacalculateddivisorwasentered,selectuser-defined (nounitsvisible).

TEST

Select the Time Portion of the Flow Rate Unit.PressLEFTARROWtoselectthetimeportionoftheRateunitofmeasure. LOG

PressUPARROWuntilthedesiredunitofmeasureisdisplayed.TEST

Select a Decimal Point Position.PressLEFTARROWuntilthedecimalpointintheRatedisplaybeginsblinking. LOG

PressUPARROWtochangethepositionofthedecimalpoint.TEST

PressENTERtosaveyourselections.SAVE

ENTER

Page 12: Mc III Quickstart Guide (1)

12

Step 1: Install the MC-III SoftwareFollowtheinstructionsonthecoveroftheCDpackettoloadthesoftware.AnMC-IIIiconwillappearonthe computerdesktopwheninstallationiscomplete.

Step 2: Connect the Computer to the MC-III

Note—An RS-485 converter is required to connect a computer to the MC-III flow analyzer. The instructions below are for an RS-485-to USB converter (Part No. 2296650-01). An RS-485-to-RS-232 9-pin converter (Part No. 9A-101283116) is also available.

1. RemoveoropenthecoveroftheMC-IIIenclosuretoaccessthecircuitboard.

2. VerifythatthebatteryisconnectedtotheJ1connector(page5).Ifusinganexternalpowersupply,verifythatitiswiredproperly(page5).

3. ConnecttheRS-485toUSBconvertercabletotheRS-485terminalsonTB2oftheMC-IIIcircuitboardasshown.

4. UseauniversalA/BUSBcabletoconnectyourcomputertotheconverter.

TD ( B )

TD ( A )

GND

UNIVERSALA/B USB CABLE

TX +

TX -

TB2

4-20 O

UT

RS

485 S

LAVE

P

ULS

E

OU

T

Part No. 2296650-01

Communication Adapters (RS-485 or USB)Ifanoptionalcommunicationadapterisinstalled,theMC-IIIEXPmaybeconnectedtoacomputerwithoutopeningtheenclosureorinstallingaddition-alfieldwiring.CommunicationadaptersareavailableforusewithRS-232orUSBcomputerports.IfusingaUSBport,auser-supplieduniversalserialbusUSBA/Bcableisrequired.SeetheMC-IIIEXPUserManualfordetails.

Important—The RS-485 to RS-232 adapter is approved for use with CSA-certified and ATEX-certified MC-III EXP analyzers. The RS-485 to USB adapter is approved only for use with a CSA-certified MC-III EXP analyzer.

Connecting to the MC-III

Page 13: Mc III Quickstart Guide (1)

13

Step 3: Run the Program1. Double-clicktheMC-IIIicononthedesktoporselectStart-Programs-

NuFlo-MC-III-MC-IIItolaunchthesoftware. AWelcomescreenwillappear,followedbyaDeviceAutorunOptions

screen.2. SelectLaunch the Configuration Wizardandfollowtheinstructions

onthescreen. Foraccesstoallconfigurableparameters,selectGo to MC-III Main (seetheMC-IIIusermanualsforinstructions).

Toautomateafunctiontoruneachtimethesoftwareconnectswiththeinstrument,clickonthefunctionontheDeviceAutorunOptionsscreenandchecktheAlways do...checkboxatthebottomofthescreen.

Page 14: Mc III Quickstart Guide (1)

14

Configuring with Software

Gas Volume Correction

Gasturbinemetersmeasuregasinactualcubicfeet(ACF).Tomeasure gasintermsofstandardcubicfeet(SCF),flowinggaspressuresand temperaturesmustbereferencedbacktostandardconditions,basedon fixed(average)parameters.These parameters are accessible only through MC-III interface software.

Toconfiguretheinstrumenttomeasuregasflowinstandardcubicfeet, performthefollowingsteps:

1. SelectGo to MC-III MainfromtheDeviceAutorunOptionsscreen.2. ClicktheK-Factor Entryiconinthescrollbaratleft.3. IntheGasVolumeCorrectionsectionoftheK-FactorEntryscreen,

checktheEnable volume correctioncheckbox(seescreenonpage15).TheSTANDARDannunciatorwillappearontheLCD.

4. Adjustatmosphericpressure,ifdesired.5. Enterthebasepressureandunitandthebasetemperatureandunit.6. Entertheworkingpressure,workingtemperature,andcorresponding

units.7. Enteraknowncompressibilityfactor,orpressCalculate compressibility

from gas comp.toenterthegascomposition.ClickCalculate.8. PressApply or OKtosavethenewcompensationsettings.

Page 15: Mc III Quickstart Guide (1)

15

Multipoint Linearization

TheMC-IIIcanbecalibratedwithupto12calibrationpointsusing MC-III interface software.(Onlysingle-pointcalibrationissupportedthroughthe ConfigurationWizard.)

Toconfiguretheinstrumentwiththeseparameters,performthefollowingsteps:

1. SelectGo to MC-III MainfromtheDeviceAutorunOptionsscreen.2. ClicktheK-FactorEntryiconinthescrollbaratleft(seescreenonpage

14).3. Selectthefactorunits.4. Selectthemultipointcheckbox.5. Enterthenumberofcalibrationpointsdesiredusingthe“plus”and

“minus”buttons(upto12pointsmaximum).6. Clickoneachactivatedcalibrationpointfield,andentertheappropriate

frequency(Hz)andcalibrationfactor.7. Repeatforallcalibrationpoints.8. PressApply or OKtosavethenewcalibrationsettings.

Page 16: Mc III Quickstart Guide (1)

Copyright 2013 Cameron International Corporation

HOUSTON HEAD OFFICE 281.582.9500

NORTH AMERICA 1.800.654.3760

[email protected]

ASIAPACIFIC +603.5569.0501

[email protected]

EUROPE,MIDDLE EAST& AFRICA

[email protected]

M E A S U R E M E N T S Y S T E M S

U S A • C A N A D A • U K • C H I N A • U A E • A L G E R I A • M A L A Y S I A • I N D I A • R U S S I A

w w w . c - a - m . c o m / f l o