Mass Spectrometry - Department of Chemistrymslab.chem.umn.edu/class/lecture/1120_Hyphenated.pdf ·...
Transcript of Mass Spectrometry - Department of Chemistrymslab.chem.umn.edu/class/lecture/1120_Hyphenated.pdf ·...
Reasons for Using Chromatography with MS
• Mixture analysis by MS alone is difficult• Fragmentation from ionization (EI or CI)
• Fragments from each component in one spectrum yields intractable interpretation
• Suppression effects in soft techniques (ESI)• Analytes that are less abundant or ionize poorly
may not be detected• Quantitation of mixtures is less accurate and less
sensitive than quantitation of components that have been chromatographically separated
Chromatography -MS Data
• Rather than averaging all data into one spectrum, each full scan is saved separately to create a chromatogram– TIC - Total Ion Chromatogram– EIC - Extracted Ion Chromatogram
• Scan speed is critical for resolution of close-eluting components– peaks could be missed or averaged together– spectral tilting
• Intensities of low/high mass ions skewed
Chromatography -MS Data
• Identification of Components– Background subtraction can be useful in
cleaning up spectra• Can remove background ions• Can remove ions from co-eluting tailing compound
– EIC can support/contradict that two MS peaks are from the same/different compounds• Comparing whether time profiles overlap
– EIC can show if ions come from the peak or are in the background
Offline LCMS
• LC may also be coupled with MS by fraction collection (ion exchange, size exclusion chromatography, TLC, Flash Column)
• Fractions may be analyzed by any compatible MS technique
• Offline Automation– LC eluent can be spotted onto a MALDI target
using a robot for MS analysis
LC in line with MS
• Liquid chromatography may be directly coupled to mass spectrometry using an APCI or ESI source
• Typical flow rates:– 100 µl/min-2.5ml/min: APCI source– > 1 ml/min: ESI source, usually need splitter (≥ 4.6 mm
column)– 1 µl/min - 1 ml/min: ESI source (200 µm- 4.6 mm
column)– < 1 µl/min: nanospray source (≤ 150 µm column)
Types of LC coupled inline to MS
• Most common= reversed phase (i.e. C18, C8, C4)
• Ion exchange• Normal phase• Affinity• 2D Chromatography (Ion Exchange
followed by reversed phase)
Restrictions
• Non-volatile salts should be avoided as signal suppression will result
• Strong ion-pairing agents such as trifluoroacetate should similarly be avoided
• Need polar solvents for ESI– Normal phase not compatible with ESI (Use APCI)
• Additives must be compatible with ionization mode (i.e. pH of separation must match desired ions)– Or use post-column addition
GCMS
• With the advent of capillary columns, GC columns can be inserted directly into the ionization source
• GC is typically coupled to EI, CI, and FI• GC-MS is applicable to a narrower range of
compounds due to volatility requirements• When GC/MS is applicable, it is a fairly
trouble free technique
Tandem Mass Spectrometry (MS/MS or MSn)
• Soft ionization techniques such as ESI only result in parent ions, revealing no structural information
• Hard ionization techniques yield fragments, but specific precursor-product relationships cannot be obtained
• Isolate and fragment precursor ion and measure m/z of product ions to obtain structural information
• One mass analyzer isolates, another scans for products
Common Methods of Fragmentation
• Threshold Dissociation (Weakest bonds are broken)– Collision Induced Dissociation (CID)
• Apply an electric field to accelerate precursor ions and induce bond cleavage via collision with neutral gas molecules (Ar, N2)
– Infrared Multiphoton Dissociation (IRMPD) • Use infrared laser to break bonds
• Electron Capture Dissociation (ECD) (Random)– Multiply-charged cations capture electrons, inducing odd-
electron fragmentation
CID
AcceleratedPrecursor Ion
Fragmentationthrough one ormore collisions
Product Ions
(And Neutrals)
Ar
+
+
+
+
IRMPD
IsolatedPrecursor Ion
Fragmentationby one ormore photons
Product Ions
(And Neutrals)
hν
+
+
+
+
ECD/ETD
IsolatedPrecursor Ion
Capture of oneor more electrons
Product Ions
(And Neutrals)
e-
+
++
•
or M-
Instruments for MS/MS
• Instruments where one or more analyzers are placed in series– Triple Quadrupole (CID at eV levels)
– Four Sector (CID at keV levels)
– Quadrupole-Time-of-Flight (Q-TOF) (CID eV)
– TOF-TOF• Trapping Instruments
– Quadrupole Ion Trap (CID eV, IRMPD, ECD)
– FT-ICR (CID, SORI-CID, IRMPD, ECD)
Quadrupole Ion Trap
M+H
•MS/MS is done in one analyzer- Precursor Ion is Isolated by selective ejection- Precursor ion is given kinetic energy- Fragments are formed and trapped- Trap is scanned to detect products- Products can be isolated and fragmented etc.
Tandem Mass Spectrometry in an Ion Trap1) Measure Full Spectrum 2) Ion Isolation:
e.g., Precursor m/z 426.7 ± 1.5
300 900 1500 2000m/z
50
100
Rel
ativ
e A
bund
ance
426.7
852.3
1138.7700.2
426.7
3) Add Collision Energy
Fragmentation of 426.7
4) Measure m/z of Fragment Ions
SLNVALSLNVASLNVSLNSLS
RLR
ALRVALR
NVALRLNVALR
[For Peptide SLNVALR]
Courtesy of the MSCLS at the Univ. of Minn.
C+Ar Products
Product Ion Scan Mode in a Triple Quadrupole
Q1 Mass selection Q2 collision cell
Fragment Ion C
Q3 Full Scan
Ion C
Q1 only transmits
Ion C
Q3 Scans for products
A
BC
D
Ionsin
Source
Products
C+Ar C1+C2+C3
Multiple Reaction Monitoring (MRM)in a Triple Quadrupole
Q1 Mass selection Q2 collision cell
Fragment Ion C
Q3 Mass Selection
Ion C
Q1 only transmits
Ion C
Q3 only transmitsIon C2
A
BC
D
Ionsin
Source
Ion C2
Neutral Loss Scan Mode in a Triple Quadrupole
Q1 Pass A-H2O Q2 collision cell
Fragment ions one at a
time
Q3 Pass A
Ion A-H2OIon B
Ion CIon D
Ion EIon F
A-H2O A+ +H2O
Scans acrossmass range
Scans acrossmass range at18 amu lowerthan Q1
IonsA-F
A
Precursor Ion Scan Mode in a Triple Quadrupole
Q1 Mass selection Q2 collision cell
Fragment ions one at a
time
Q3 pass only 79
Ion A-PO4
Ion BIon C
Ion DIon E
Ion F
PO3- 79
Scans acrossmass range
Transmits only 79
IonsA-F
-/+ Polarity switch
Precursor ion of 79 (-)
Enhanced resolution (+)
MSMS (+)
-/+ Polarity switch
Charge state determination
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48 50 52Time, min
0.0
5.0e7
1.0e8
1.5e8
2.0e8
2.5e8
3.0e8
3.5e8
4.0e8
4.5e8
5.0e8
5.5e8
6.0e8
6.5e8
7.0e8
7.5e8
Inte
nsity
(cps
)
β-Casein Digest Total Ion Chromatogram
Courtesy of the MSCLS at the Univ. of Minn.
5 10 15 20 25 30 35 40 45 50 55 60 65Time, min
0.00
1.00e7
2.00e7
3.00e7
4.00e7
5.00e7
6.00e7
7.00e7
8.00e7
9.00e7
1.00e8
1.10e8
1.20e8
1.29e8
Inte
nsity
(cps
)Negative Precursor Ion Chromatogram
Courtesy of the MSCLS at the Univ. of Minn.
400 500 600 700 800 900 1000 1100 1200m/z, amu
100
300
500
700
900
1100
1300
1500
1700
1900
21001029.2
Inte
nsity
, cps
Precursor Ion Scan at 30 min
Courtesy of the MSCLS at the Univ. of Minn.
1031 1032 1033m/z, amu
1.00e4
2.00e4
3.00e4
4.00e4
5.00e4
6.00e4
7.00e4
8.00e4
9.00e4
1.00e5
1.10e5 1031.31031.7
1032.2
Inte
nsity
, cps
Positive Enhanced Resolution Scan
+2 ion, MW= 2060.6
Courtesy of the MSCLS at the Univ. of Minn.
100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu
2.0e5
6.0e5
1.0e6
1.4e6
1.8e6
2.2e6
2.6e6
3.0e6
3.4e6F Q pS E E Q Q Q T E D E
a1-NH3b1-NH3
I
b2
a2-NH3
b2-NH3
b3b3-NH3
b4b4-NH3 a5
b5
y12y12-NH3b6
y11y11-NH3
b7
y10y10-NH3
b8
y9
b8-NH3
y9-NH3
y8
b9-NH3y8-NH3
y7
y7-NH3
y6
y6-NH3
y5
y5-NH3
y4
y4-NH3
y3
y3-NH3
y2
y2-NH3
y1
y1-NH3
a1
Inte
nsity
, cps
MSMS on ion 1031.3 for sequence determination
Courtesy of the MSCLS at the Univ. of Minn.
Q-TOF Mass Analyzer
TOF
Sample Inlet
Q2(Collision cell)
Hexapole
Ion Guide
Q1
MCPDetector
Reflectron
Pusher
•Select 1 ion in Q1 (Product Scan)•Scan Q1 (Precursor Scan)•Cannot perform Neutral Loss Scan
FT-ICR-MS
• MS/MS is done in one analyzer– RF fields can selectively isolate one ion– RF fields (on or off resonance) can accelerate ions followed by
collisions and detection of products– IR laser or electron source can fragment isolated ions followed by
detection of products– Processes can be mixed, matched, repeated etc.
M+H
MS6 With FT-ICRCO2
O2C-
-CID CO2
-
•CID
•
-
•
- SO2•
-O
t-BuSH-
O
CID
O•
H
--
low energy
highenergy
t-BuSH
No Reaction
Phenol
ProtonTransfer
Reed et. al. JACS, 122, 10689-10696, (2000)
MS/MS Applications• Structural elucidation aid for unknown
compounds• Functional group confirmation• Positive Identification of known compounds
– environmental contaminants– biological metabolites– controlled substances (illegal drugs, steroids etc.)
• When Combined with Chromatography– Screening of mixtures for specific moieties– Structure-specific quantitation
• Thermochemical information (BDE etc.)• Chiral analysis