MAR 300 Laboratory Exercise 1
-
Upload
joshua-bergeron -
Category
Documents
-
view
215 -
download
0
Transcript of MAR 300 Laboratory Exercise 1
-
7/27/2019 MAR 300 Laboratory Exercise 1
1/3
Bergeron 1
Laboratory Exercise 1: Spectrophotometry
Joshua Bergeron
9-5-2013
-
7/27/2019 MAR 300 Laboratory Exercise 1
2/3
Bergeron 2
Introduction: This lab exercised the use of a spectrophotometer in the determination of the
unknown concentration of a known solution. The Genesys 10 spectrophotometer has the
capability of determining a reference wavelength and generating a standard curve based on an
analysis of known concentrations of a solution.
This exercise is relevant to any accurate determination of the concentration of a solution.
The methodology in this lab is an efficient and accurate method for determining an unknownconcentration in a known solution.
Procedure:
Step 1) Determine a Reference Wavelength
a. Prepare a blank dH2O and a sample of 0.1M solutionb. Insert them into Genesys 10 Spectrophotometerc. Press [TEST]; scroll to Scanning & press [ENTER]d. Select stored tests; scroll to Scan(PAR); select Load Teste. Ensure it will scan from 400-700 nmf. Set to read the blank; select Collect Baselineg. Select Measure Sample; select Edit Graph: Math: Peaks: & Valleysh. Print graph; select tabular & print tablei. The reference wavelength corresponds to the largest absorbance value
Step 2) Determine a Standard Curve
a. Prepare 5 samples from 0.1 M to 0.005 M of known solutiona) Vs = Volume of stock solution required to reach target molarityb) Vs = (0.002 L (0.0075 M))/0.100 M = 1.5 mLc) Vs = (0.002 L (0.0050 M))/0.100 M = 1.0 mLd) Vs = (0.002 L (0.0025 M))/0.100 M = 500 Le) Vs = (0.002 L (0.0005 M))/0.100 M = 1oo L
b. Press [TEST]; scroll to Standard Curve; press [ENTER]a) Set appropriate test parameters
c. Select Run Standards & key in the concentrations of samples 1-5determined above
d. Select Measure Standards; print absorbance, slope, and interceptStep 3) Determine the unknown concentration from the known solution
a. press [ESC] to exit
b. Set the reference wavelength and measure the absorbance of the sample
c. Compare the absorbance to the standard curve in order to estimate the unknown
concentration
-
7/27/2019 MAR 300 Laboratory Exercise 1
3/3
Bergeron 3
Results:
Wavelength 400 produced the highest absorbance value of 0.289 and was used as the Reference
Wavelength.
A standard curve with a standard deviation of 0.012 was produced with a slope of 3.38
The unknown sample was measure at 400nm and yielded a measured absorbance of 0.187Based on the standard curve results, the unknown concentration is estimated to be 0.050 M
See Attached Lab Results
Discussion: My initial experiment resulted in errors within my standard curve. After repeating
the experiment, I realized I had failed to fill the blank cuvette with a blank sample. This
prevented a proper determination of absorbance within my standard curve. I find this procedure
to be highly efficient and accurate.
Once the error in my method was corrected, I was able to follow the procedure and
determine the unknown concentration.
Questions:
1) What was the point of blanking the spectrophotometer with dH2O? dH2O is free of anything
that would alter absorbance. Its important to use a blank that only lacks the element you are
trying to measure.
2) If you wanted to measure the absorbance of a sample of natural seawater, what would you
suggest using as your BLANK? What potential sources of error would be inherent in the method
you suggest? Be specific. I would attempt to produce a blank that only lacked the element I
was trying to measure. If it was a chlorophyll determination, I would either filter the seawater
or attempt to create artificial seawater. The difficulty arises in the recreation of naturalseawater- even artificial seawater doesnt accurately represent the content of any one
random natural seawater sample. Your method for producing an appropriate blank depends
entirely on which element within natural seawater you are measuring the absorbance of.
3) Take a closer look at Equation 1 as it relates to Beers Law. Which would have a more
significant impact on absorbance: molecular abundance (concentration) or the distance through
which the light beam must pass (path length)? Explain your rationale. IfA=lc then either a
change in l (pathlength) orc (concentration) would effect absorbance. They represent an
equal ratio in the equation.
4) When you were calculating the concentration of your unknown (using your standard curve),
why were you able to assume that the y-intercept was zero? The standard curve used for the
concentration determination was graphed from the origin. This allows you to assume that the
y-intercept is always zero.