Manual of AWA6228 V).OCT & 1/3 OCT measurement

Hangzhou Aihua Instruments Co., Ltd

Page 2: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Boot up meter…

Press the start/reset key for at least 2 seconds to boot up the meter. And enter to the main menu.

Main menu

Page 3: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Calibration check before test

Press Esc to go back to main menu, and move the cursor to ‘Meas.’, press enter to go to interface of measurement.

Left / right keys to move the cursor, and up/down keys to change the setting mode. For calibration check, the setting should be ‘NOIS, ONE, LIST, INST’.

Page 4: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Put on the calibrator and turn it on.Notes: keep the meter and calibrator connected tightly and solidly.

Measurement interface

After you turn on the calibrator, the LAFp should be 93.8±0.2 dB. Otherwise, please do the calibration. Ref. to video I). Calibration.

Calibration check before test (Cont.)…

Page 5: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Start testing…

After finishing calibration, Press Esc key to return to main menu. Move the cursor to ‘Meas.’, then press enter.

Left / right keys to move cursor to 2nd box, up/down keys to changing setting to ‘OCT’ or ‘ 1/3’. Notes: the following will use 1/3 Oct as example.Measurement interface

Page 6: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Vary frequency…

Move cursor to ‘Page’, and press up/down keys to turn page to view more dBs in various frequencies.

Notes: in the OCT test, all the dB value response to vary frequency is Z-weighted.

Page 7: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Graphic view…

Move the cursor to 3rd box, and change setting to ‘Gra.’, then you can view it in graphic form.

Move cursor to here, can change the weighted setting from ‘A, C or Z’.

Page 8: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Big font view…

Change the setting in 3rd box to ‘BIG’, then you can focus on the one parameter that you care the most, and view it in bigger font.

Move the cursor here, to pick the parameter that you want to observe.

Page 9: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Start testing…

Move the cursor to forth box, and change the index to ‘Leq’, then it will show the equivalent continuous sound level

Press Run to start testing.

Page 10: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Recall data…When it is doing the test, the up-corner will display ‘RUN’ ; when you press pause , the word will change to be ‘PAUSE’; when the test finished, the word shown as ‘END’.

After the measurement finished, the data will be auto saved in the meter’s memory. Esc key go back to the main menu, and pick ‘Data’ to recall the saved data.

Page 11: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Recall Data(Cont.)…Left/right keys to move cursor and pick the data that you want to read. (1/3 means 1/3 OCT test.)

Page 12: Manual of AWA6228 V).OCT & 1/3 OCT measurement Hangzhou Aihua Instruments Co., Ltd.

Recall Data(Cont.)…

Press left/right keys to roll the screen horizontally.

Press enter to view the graph.

up/down keys to turn the page. Press Out to

print the data, while you have mini-printer attached

Page 2

Page 1Page 1Date Time Mode run time Range