M2000 Hotplate

download M2000 Hotplate

of 43

Transcript of M2000 Hotplate

  • 8/4/2019 M2000 Hotplate

    1/43

    Page 1

    TABLE OF CONTENTS:

    SECTION 0: SYSTEM DESCRIPTION PAGE 3SPECIFICATIONS PAGE 3

    FEATURES PAGE 4

    SECTION 1: UNPACKING INSTRUCTIONS PAGE 6TO UNPACK THE UNIT PAGE 6

    SECTION 2: SYSTEM OPERATION PAGE 8KEYBOARD FUNCTIONS PAGE 8

    TEMPERATURE CONTROLLER PAGE 11TEMPERATURE CONTROLLER KEYBOARD PAGE 11

    SECTION 3: SYSTEM CONFIGURATION PAGE 13PROCEDURE PAGE 13

    SECTION 4: COLD PLATE OPERATIONS PAGE 18TEMPERATURE CONTROLLER PAGE 18

    SECTION 5: PROGRAMMING PAGE 20FLOW CHART SYSTEM PROGRAMMING PAGE 20

    PROCEDURE PAGE 21

    SECTION 6: RUNNING PROGRAMS PAGE 25PROCEDURE PAGE 25

    SECTION 7: PROCESS DESCRIPTION PAGE 30CONVENTIONAL OVENS PAGE 30

    HOTPLATE BAKING PAGE 31BAKING STYLES PAGE 32REFLOW SOLDERING PAGE 33SILICON WAFERS PAGE 33GALIUM ARSINIDE PAGE 33PHOTOMASKS AND DISPLAYS PAGE 34EXHAUST COVER PAGE 34BAKING PROCESS PAGE 35

  • 8/4/2019 M2000 Hotplate

    2/43

    Page 2

    TABLE OF CONTENTS:

    SECTION 8: OPTIONS PAGE 38THE OPTion KEY PAGE 38

    HARDWARE OPTIONS PAGE 40

    SECTION 9: ERRORS AND THEIR RECOVERY PAGE 41

  • 8/4/2019 M2000 Hotplate

    3/43

    Page 3

    Section 0

    DESCRIPTION:

    MODEL: Cee (Cost Effective Equipment) Series 2000

    Hotplate.

    TEMPERATURE CONTROL: PID ( Proportional Integral Derivative) with plati-num RTD sensor. Microprocessor controlled to a

    maximum temperature of 400 C with 0.1-C

    resolution.

    SUBSTRATE HANDLING: Substrates handled by a combination of Silicone

    belts and gas pillow.

    SUBSTRATE SIZE: Capacities are 3 to 6 inches depending on model

    and or retrofit.

    MAXIMUM THROUGHPUT: 18 substrates per minute with 0 bake time and 3-

    inch substrates.

    MAXIMUM BAKE TIME: 29 minutes 57 seconds (9 minutes 59 seconds per

    step).

    USER PROGRAM STORAGE: Lithium battery, life expectancy of approximately33 years.

    EXHAUST COVER Cee open-faced single-lip exhaust cover supplied

    as standard in all hotplates.

    POWER SUPPLY: 120 Volts AC (240 Volts AC, optional), 700

    Watts.

    UTILITIES: Nitrogen: 1/4" NPT Female 70 psi minimum. Vac-

    uum: 1/4" NPT Female 25" Hg

    Exhaust: 2" O.D. 20-100 CFM

    DIMENSIONS: 11" height x 24.1" length x 13.2" depth

    WEIGHT: 85 lb / 32.3 kg

  • 8/4/2019 M2000 Hotplate

    4/43

    Page 4

    Section 0

    DESCRIPTION:

    FEATURES

    The battery-backed memory stores up to ten user programs containing preset values for

    bake temperature, bake time, and bake method. A program lockout key preventsunauthorized changes to user programs.

    The Series 2000 Hotplate supports three bake methods: proximity, soft-contact, and

    hard-contact (For more information, see bake methods in Section 6). These bake methods

    provide maximum control of substrate warmup, solvent drying, and resin curing

    characteristics.

    A user program contains up to three consecutive bake methods.

    With the automatic substrate handling, the operator need only load unbaked substrates onto

    the entry track then remove baked substrates from the exit track. With the optional cassette

    elevators installed, even this procedure is handled automatically.

    A separate microprocessor allows greater control of temperature critical processes. The

    Series 2000 Hotplate realizes bake temperatures to maximum of 400 C with resolution of

    0.1 C.

    Easily installed retrofits facilitate changing substrate size from large photomasks to Silicon

    or GaAs wafers of less than 2 inches in diameter.

    A benchtop unit, the Series 2000 Hotplate covers less than two-square feet and requires

    only common utilities: 120 Volts AC; nitrogen gas; and vacuum.

    Built-in, diagnostic software aids in system installation and testing.

    Built-in communication facilitates interfacing with elevators.

    Single-step, operating mode allows substrate baking without prior programming setup for

    quick processing of one or two substrates.

  • 8/4/2019 M2000 Hotplate

    5/43

    Page 5

    Section 0

    DESCRIPTION:

    FEATURES continued

    The option key permits use of special or custom software to perform specific tasks not

    available in the standard hotplate. For example, modifications can include special bakemethods, temperature ramp programs, or statistical data storage of substrate temperature

    profiles during baking.

    Information on various options will be distributed as they become available.

  • 8/4/2019 M2000 Hotplate

    6/43

    Page 6

    Section 1.

    UNPACKING INSTRUCTIONS

    The Series 2000 Hotplates are shipped in plyboard crates and packed with urethane foam during

    shipment to reduce the possibility of damage to the unit.

    TO UNPACK THE UNIT

    1. Remove all loose packing material from the carton.

    2. Remove the plastic bag containing the keys.

    3. Lift the unit out of the carton by grasping the ends of the hotplate only . Do not lift the

    unit by any of the top covers or protrusions. Do not flip the unit on its side or ends.

    4. Remove the plastic wrap from the unit.

    5. Open the bag containing the keys: insert one into the PROG/RUN switch on the front of

    the unit, the other should be stored in a safe place.

    6. On the back of the unit are two, 1/4" NPT pipe fittings: GAS and VAC. Connect these to

    nitrogen gas at 60 psi or greater and vacuum at 25" Hg, respectively.

    POWER OFFRUN PROG.

    ON

  • 8/4/2019 M2000 Hotplate

    7/43

    Page 7

    Section 1.

    UNPACKING INSTRUCTIONS

    TO UNPACK THE UNIT continued

    7. Plug in the unit; verify operation by turning on the power switch. The hotplate display

    should read "CEE -----SERIES 2000."

    8. Press the DIAG key to enter the diagnostic mode. Press the "3" key to turn on the gas

    jets in the chuck; this allows precise leveling of the unit.

    9. Place a wafer of correct size on the chuck. Adjust the leveling legs on the base of the

    unit until the wafer does not tend appreciably toward the front or back of the chuck--but

    does tend toward the right side of the chuck. Allow a 1/2" gap to the bench surface to

    allow adequate ventilation.

    10. Press the reset key to turn off the nitrogen flow.

    * * * * *

    This completes the setup procedure for the Series 2000 Hotplate.

  • 8/4/2019 M2000 Hotplate

    8/43

    Page 8

    Section 2.

    SYSTEM OPERATION:

    KEYBOARD FUNCTIONS

    The Series 2000 Hotplate has two independent microprocessors: system and temperature

    controllers.

    The keyboard shown below guides the system controller. All programming and functioning of

    the machine operate through here.

    The second keyboard regulates the temperature-control microprocessor. However, the systemcontroller automatically and directly instructs the temperature controller which rarely

    necessitates use of the temperature-control microprocessor.

    KEYBOARD FUNCTIONS

    0-9 number Selects a program to modify or run, as well as, encodes

    numerical data.

    PROGram Places the system in program mode. Using this key while in

    run mode identifies the current bake program or most recent

    bake program.

    RUN Places the system in run mode. Allows selection of desired

    program to execute.

    1 2 3 4 5

    6 7 8 9 0

    OPT.

    PROG RUN DIAG

    START STOP RESET

    ENTER CLEAR CONF

    PROX

    CONT

    VAC

    CEE ------ SERIES 2000

  • 8/4/2019 M2000 Hotplate

    9/43

    Page 9

    Section 2.

    SYSTEM OPERATION:

    KEYBOARD FUNCTIONS continued

    DIAGnostics Starts diagnostic routines. Primarily used for system checkout,

    DIAGnostics allows testing of all sensors in addition to the

    manual operation of motors.

    START Executes a user, bake program. In the event of an error,

    START instructs the system to continue from the point at which

    the error occurred.

    STOP Stops a user program when the current cycle ends and immedi-

    ately stops the bake cycle during single-step baking.

    RESET Instructs the controller to perform a complete system reset. All

    motors, solenoids, and elevators halt; the system returns to the

    series number display. Note: after a reset, the controller will

    not remember a substrate on the chuck--manually remove it.

    ENTER Signifies acceptance of the currently displayed choice during

    data entry and program selection. This key and the clear key

    function similarily to a calculator keyboard.

    1 2 3 4 5

    6 7 8 9 0

    OPT.

    PROG RUN DIAG

    START STOP RESET

    ENTER CLEAR CONF

    PROX

    CONT

    VAC

    CEE ------ SERIES 2000

  • 8/4/2019 M2000 Hotplate

    10/43

    Page 10

    Section 2.

    SYSTEM OPERATION:

    KEYBOARD FUNCTIONS continued

    CLEAR Removes unwanted or invalid information during data entry.

    CLEAR also selects the nobake option when configuring a user

    program.

    CONFiguration Defines system. Allows selection of a usable, bake-temperature

    range in percent.

    OPTions Accesses optional software to perform specific or custom tasks.

    PROXimity Selects proximity bake during system programming.

    CONTact Selects soft-contact bake during system programming.

    VACuum Selects hard-contact bake during system programming.

    1 2 3 4 5

    6 7 8 9 0

    OPT.

    PROG RUN DIAG

    START STOP RESET

    ENTER CLEAR CONF

    PROX

    CONT

    VAC

    CEE ------ SERIES 2000

  • 8/4/2019 M2000 Hotplate

    11/43

    Page 11

    Section 2.

    SYSTEM OPERATION:

    TEMPERATURE CONTROLLER

    Except under specified circumstances, the user does not manipulate the temperature-control

    keyboard located on the left-hand side of the Series 2000 Hotplate. The system controller

    instructs the temperature controller during normal functions rendering operator intervention

    unnecessary.

    The display provides information about the current temperature setpoint, as well as, the current

    chuck temperature. By manipulating the keyboard, the display provides informationrelating to output power and tuning.

    TEMPERATURE CONTROLLER KEYBOARD

    YES / NO Manually controls set point. Use these dual purpose keys with

    the display shown above. The YES key increases the set point;

    the NO key decreases the set point as indicated by an "up" ar-

    row and a "down" arrow, respectively.

    In addition, use these keys to answer questions about informa-

    tion required by the temperature controller.

    TUNE / RETURN Use during controller tuning and calibration. Unless a change

    is dictated, the operator need not use this key.

    Ye s No

    TuneReturn

    ParamDisplay

    StartStop Last

    110.3 COUT1 SP 110.5

    REM

  • 8/4/2019 M2000 Hotplate

    12/43

    Page 12

    Section 2.

    SYSTEM OPERATION:

    TEMPERATURE CONTROLLER KEYBOARD continued

    PARAMeters / DISPLAY Changes the type of information presented on the tem-

    perature-controller display. The normal display shown

    above reveals the current chuck temperature, 110.3 C;

    and the current set point (SP), 110.5 C.

    The left side of the readout displays "OUT 1" since atthis moment the chuck heating element is on. "REM" on

    the right indicates active remote communications.

    Pressing PARAM / DISPLAY once changes to a display

    similar to the one depicted in Figure 1.

    Figure 1 displays the current output power level of the

    heater (OUT 1) as well as chuck temperature, 110.3 C.

    110.3 COUT 1 70%

    Ye s No

    TuneReturn

    Param

    DisplayStartStop Last

    110.3 COUT1 SP 110.5

    REM

    Figure 1.

  • 8/4/2019 M2000 Hotplate

    13/43

    Page 13

    Section 3.

    SYSTEM CONFIGURATION:

    This section presents the system configuration for the Series 2000 Hotplate.

    The configuration procedure conditions the machine to accept substrates from or deliver sub-

    strates to automatic cassette elevators or other Series 2000 Hotplates.

    The configuration procedure instructs the system as to the maximum, acceptable, temperature

    error (in percent) before initiating substrate baking.

    NOTE--In order to insure proper functioning of the CONF key, set the keyswitch on the front

    panel to PROG.

    NOTE--The default values provided represent examples; values may be different in a given ma-

    chine.

    PROCEDURE

    1. From the model number display (CEE-----SERIES 2000), press CONF to place the

    system in configure mode.

    DISPLAY

    ACTION

    CEE --- SERIES 2000

    PRESS CONF KEY

    (CONF)

    POWER OFFRUN PROG.

    ON

  • 8/4/2019 M2000 Hotplate

    14/43

    Page 14

    Section 3.

    SYSTEM CONFIGURATION:

    PROCEDURE continued

    The display reads "CONFIGURE ENTRY 0," where "0" indicates that the system is

    currently configured for manual entry.

    CONFIGURE ENTRY 0

    PRESS ENTER

    (ENTER)

    At any time, a flashing number or cursor signals the system waits for input from the

    operator. Pressing ENTER leaves the value unchanged.

    Pressing CLEAR changes the value followed by "1 ENTER" for cassette elevator entryor "2 ENTER" for receiving substrates from another Series 2000 Hotplate.

    CONFIGURE ENTRY 0

    PRESS 1 FOR ELEVATOR

    (CLEAR-1-ENTER)

    CONFIGURE ENTRY 0

    PRESS 2 FOR SUBSTRATES

    (CLEAR-2-ENTER)

    2. After configuring the entry type, next configure the exit type. The display reads

    "CONFIGURE EXIT 0," where again, "0" indicates that the system is currently

    configured for manual exit type.

    CONFIGURE ENTRY 0

    PRESS ENTER

    (ENTER)

    Retain or replace this value in the same manner as for the entry type.

  • 8/4/2019 M2000 Hotplate

    15/43

    Page 15

    Section 3.

    SYSTEM CONFIGURATION:

    PROCEDURE continued

    If configuration of both the entry and exit is manual loading and unloading,

    respectively, the system will skip the next two steps which pertain only to systemsusing cassette elevators.

    3. If the operator selects cassette elevators for either the entry or exit, the system displays

    "CASSETTE PITCH 3/16" indicating that the current configuration calls for cassettes

    with 3/16" spacing between substrates.

    CASSETTE PITCH 3/16"

    PRESS ENTER

    (ENTER)

    Pressing ENTER retains this value.

    Pressing CLEAR followed by the desired numeral and ENTER, changes this value.

    Acceptable values range from one-sixteenths to eight-sixteenths of an inch spacing, i.e.

    1/16" to 1/2."

    CASSETTE PITCH 3/16"

    PRESS # FOR SIZE

    (CLEAR-1-1-6-ENTER)

    4. The display reads "ENTER DIAM INCHES 3" where "3" represents the current,

    configured, substrate size. Current values may be retained or changed as in the entry

    type configuration. Acceptable values range from three to six inches.

    ENTER DIAM INCHES 3

    PRESS # FOR SIZE

    (CLEAR-6-ENTER)

  • 8/4/2019 M2000 Hotplate

    16/43

    Page 16

    Section 3.

    SYSTEM CONFIGURATION:

    PROCEDURE continued

    3. At this point, the display reads "TEMP ERROR PRCNT 2.5," where "PRCNT 2.5"

    equals 2.5%; a value previously configured to be the maximum error tolerated beforeactually baking wafers. Current values may be retained or changed as in the entry type

    configuration.

    TEMP ERROR PRCNT 2.5

    PRESS ENTER

    (ENTER)

    With the value of 2.5% and a setpoint of 100.0 C, the system would not begin baking a

    wafer unless the chuck temperature equalled at least 97.5 C--but not more than 102.5

    C. Acceptable values range from 0.1% to 9.9%.

    Note: setting the error range to 0.0% effectively shuts off the temperature-checking

    function. Turn this function off in this manner only if unusually large temperature

    variations can be tolerated without ill effects.

    4. The display will now read "CHILL METHOD/ PROX.". The cool method can bechanged by pressing "CONT." or "PROX.".

    CHILL METHOD/PROX PRESS "CONT" OR

    "PROX"

    5. The display will now read "CHILL TIME 0M 30S". Minutes are changed firstfollowed by seconds.

    CHILL TIME 0M 30SENTER NEW MINUTES

    (CLEAR-2-ENTER)

    CHILL TIME 2M 30SENTER NEW SECONDS

    (CLEAR-15-ENTER)

  • 8/4/2019 M2000 Hotplate

    17/43

    Page 17

    Section 3.

    SYSTEM CONFIGURATION:

    PROCEDURE continued

    6. Now the display reads "STORE DATA 1/0." Up to this point, a temporary buffer stores

    any changes; the system configuration remains unchanged. To make all changespermanent, press "1 ENTER."

    STORE DATA 1/0

    STORE CHANGES

    (1-ENTER)

    Before initiating permanent changes, if uncertainty about any changes exists, or the

    operator desires to review the default settings, enter "0" to return to the first display

    "CONFIGURE ENTRY 0."

    7. After storing all changes, the system returns to the model number display.

    8. To quit the configuration mode, press RESET at any time. The system returns to the

    model number display without storing any changes to the system configuration.

    * * * * *

    This completes the system configuration procedure.

  • 8/4/2019 M2000 Hotplate

    18/43

    Page 18

    Section 4.

    COLD PLATE OPERATIONS

    This unit was purchased with a cold plate installed to speed the cooling after bake processing

    photomask plates. This section outlines the operation of this portion of the system.

    Programming of this operation is accomplished through the configuration routine. The last in-structions requested in the configuration routine are the style of cooling mode and the duration.

    Cold plate programming is accomplished by using the up arrow, down arrow, and the enter

    button.

    TEMPERATURE CONTROLLER

    Note: The display shown is for 0.1 (resolution) option installed. Machines with thestandard 1 resolution will not show the decimal.

    The cold plate is designed to maintain itself at or near room temperature +/- 10C. The

    temperature controller displays the cold chuck temperature at all times in degrees C. The error

    indicators show at a glance if the temperature is correct, higher or lower than the setpoint of

    the controller.

    The display provides information about the current temperature set point, as well as, the cur-

    rent chuck temperature. By manipulating the keyboard, the display reveals information related

    to output power and tuning.

    The temperature controller incorporates a red and green LED display. Display of the current

    chuck temperature and set point occurs at all times. The red LED display shows the current

    temperature. The green LED display shows the set point.

    SP1

    AL

    SP2

    F

    C

    INDEX ENTER

    PV

    SV

    MADE IN USA

    AL

    20.0

    20.0

    AL

  • 8/4/2019 M2000 Hotplate

    19/43

    Page 19

    Section 4.

    COLD PLATE OPERATIONS

    KEYBOARD FUNCTIONS

    INDEX/AL Pressing INDEX once allows the user to adjust the setpoint

    with the up and down arrows. ENTER must be pressed to make

    set point changes.

    UP ARROW Used in conjunction with the INDEX/AL key to alter the set

    point (desired temperature setting) of the controller.

    DOWN ARROW Used in conjunction with the INDEX/AL key to alter the set

    point of the controller. This key functions inversely to the UP

    ARROW key.

    ENTER ENTER stores the value or the item changed.

    In order to minimize set-up times when adjusting the system temperature, the UP ARROW and

    the DOWN ARROW keys change to a high-speed mode after depressing for a few seconds.

    SP1

    AL

    SP2

    F

    C

    INDEX ENTER

    PV

    SV

    MADE IN USA

    AL

    20.0

    20.0

    AL

  • 8/4/2019 M2000 Hotplate

    20/43

    Page 20

    Section 5.

    PROGRAMMING:

    The Series 2000 Hotplates stores ten user programs in battery-backed memory. Programs

    contain information such as bake temperature, up to three bake methods, and bake times.

    The operator individually modifies or reviews programs, numbered zero to nine.

    NOTE--In order to insure proper functioning of the PROG key, set the switch on the front pan-

    el to PROG.

    NOTE--The default values provided represent examples; values may be different in a given

    machine.

    FLOW CHART SYSTEM PROGRAMMING

    PRESS PROGRAM

    KEY

    CEE --- SERIES 2000

    PROG MODE/PROG #? 2

    ENTER PROGRAM #

    (CLEAR-0-ENTER)

    PGM 0/ TEMP 115.0

    ENTER NEW TEMP

    (CLEAR-1000-ENTER)

    PGM 0/ METHOD 1 CONT

    ENTER NEW METHOD

    (CLEAR-PROX-ENTER)

    PGM 0/ TIME 1 0M 10S

    ENTER NEW MINUTES

    (CLEAR-1-ENTER)

    PGM 0/ TIME 1 0M 10SENTER NEW SECONDS

    (CLEAR-45-ENTER)

    PGM 0/ METHOD 2 VAC

    ENTER NEW METHOD

    (CLEAR-CONT-ENTER)

    PGM 0/ TIME 2 0M 10S

    ACCEPT MINUTES

    (ENTER)

    PGM 0/ TIME 2 0M 15SENTER NEW SECONDS

    (CLEAR-20-ENTER)

    PGM 0/ METHOD 3 PROX

    ENTER NEW METHOD

    (CLEAR-VAC-ENTER)

    PGM 0/ TIME 3 0M 10S

    ACCEPT MINUTES

    (ENTER)

    PGM 0/ TIME 3 0M 10SACCEPT SECONDS

    (ENTER)

    IFNOBKE

    (CLEAR)SELECTED

    IFNOBKE

    (CLEAR)SELECTED

    CEE ---- SERIES 2000 PROGRAMMING

    END OF

    DISPLAYACTION

  • 8/4/2019 M2000 Hotplate

    21/43

    Page 21

    Section 5.

    PROGRAMMING:

    PROCEDURE

    1. The preceding page details the Flow Chart System Programming.

    DISPLAY

    ACTION

    From the series number display (CEE -----SERIES 2000), press the PROG key to install

    program mode.

    CEE --- SERIES 2000

    PRESS PROGRAM

    KEY

    The display reads "PROG MODE/PROG #? 2," where "2" represents the most recently

    used program (default value). To work with program number 0 instead of number 2,

    press CLEAR, followed by "0," then ENTER.

    PROG MODE/PROG #? 2

    ENTER PROGRAM #

    (CLEAR-0-ENTER)

    2. At this point, the display reads "PGM 0/ TEMP 115.0," where "115.0" indicates the

    current temperature setting (TEMP) for program number 0 (PGM 0). To alter this val-

    ue press CLEAR, followed by "1," " 4," " 5," " 5," then ENTER. This sets the tempera-

    ture of program 0 to 145.5 C.

    PGM 0/ TEMP 115.0

    ENTER NEW TEMP

    (CLEAR-1-4-5-5-ENTER)

    3. Next, the display indicates "PGM 0/ METHOD 1 CONT" to program the first of three

    possible bake methods as indicated by "1."

    PGM 0/ METHOD 1 CONT

    ENTER NEW METHOD

    (CLEAR-PROX-ENTER)

  • 8/4/2019 M2000 Hotplate

    22/43

    Page 22

    Section 5.

    PROGRAMMING:

    PROCEDURE continued

    In this example, the contact ("CONT") bake method is currently selected. To change to

    the proximity bake method, press CLEAR, followed by PROX, then ENTER.

    PGM 0/ METHOD 1 CONT

    ENTER NEW METHOD

    (CLEAR-PROX-ENTER)

    4. As currently revealed by the display, "PGM 0/ TIME 1 0M 10S," the bake time

    ("TIME") for method 1 ("1") is 0 minutes ("0M") and 10 seconds ("10S").

    PGM 0/ TIME 1 0M 10S

    ENTER NEW METHOD

    (CLEAR-1-ENTER)

    First, accept or change the bake time minutes value. To program minutes in the bake

    time to 1, press CLEAR, followed by "1," then ENTER.

    PGM 0/ TIME 1 0M 10S

    ENTER NEW MINUTES

    (CLEAR-1-ENTER)

    Next, accept or change the bake time seconds value. To program seconds in the bake

    time to 45, press CLEAR, followed by "45," then ENTER.

    PGM 0/ TIME 1 1M 10S

    ENTER NEW SECONDS

    (CLEAR-45-ENTER)

    As a result, total bake time for the first bake, method 1, equals 1 minute and 45 sec-

    onds.

    Acceptable values for bake times range from 0 to 9 minutes and 0 to 59 seconds.

  • 8/4/2019 M2000 Hotplate

    23/43

    Page 23

    Section 5.

    PROGRAMMING:

    PROCEDURE continued

    5. At this point, the operator may continue programming for multiple bakes. The same

    procedure applies to the second and third methods and bake times.

    PGM 0/METHOD 2 VAC PGM 0/METHOD 3 PROX

    ENTER NEW METHOD ENTER NEW METHOD

    (CLEAR-CONT-ENTER) (CLEAR-VAC-ENTER)

    PGM 0/TIME 2 0M 10S PGM 0/TIME 3 0M 10S

    ACCEPT MINUTES ACCEPT MINUTES

    (ENTER) (ENTER)

    PGM O TIME 2 0M 10S PGM O TIME 3 0M 10S

    ENTER NEW SECONDS ACCEPT SECONDS

    (CLEAR-20-ENTER) (ENTER)

    END OF

    PROGRAMMING

    CEE --- SERIES 2000

    If a multistage bake is unnecessary, press CLEAR for the second bake method. Thedisplay will read "NOBKE" indicating no additional bake methods.

    PGM 0/METHOD 2 VAC PGM 0/METHOD 3 PROX

    ENTER NEW METHOD ENTER NEW METHOD

    (CLEAR-CONT-ENTER) (CLEAR-VAC-ENTER)

    IF NOBKE (CLEAR) SELECTED

    IF NOBKE (CLEAR) SELECTED

    END OF

    PROGRAMMING

    CEE --- SERIES 2000

    To end the program at this point, press the ENTER key. To program dual method

    bakes, enter CLEAR, "NOBKE," for the third method.

  • 8/4/2019 M2000 Hotplate

    24/43

    Page 24

    Section 5.

    PROGRAMMING:

    PROCEDURE continued

    6. After programming the third bake method and time or selecting a NOBKE, the system

    stores all program data in memory.

    NOTE: The current program in memory is not replaced until completing the program.

    If at any time during programming the operator is unsure of the values selected or

    would prefer to use the default values, press RESET, and no changes occur to the origi-

    nal program. Press PROG again to start over from the beginning.

  • 8/4/2019 M2000 Hotplate

    25/43

    Page 25

    Section 6.

    RUNNING PROGRAMS:

    Running user programs on the Series 2000 Hotplates is not difficult. The hotplate system itself

    handles most of the work; the user need only supply substrates and instruct the system as to

    which program to run.

    Units equipped with cassette elevators accept many substrates at once. Except to replace a

    cassette for a completed batch, the Series 2000 Hotplate requires no operator intervention.

    Except when noted, the following procedures apply to both manual and cassette elevator

    systems.

    Warning: The surface of the chuck is hot; never touch the chuck surface. Never touch

    substrates after baking until sufficient time for cooling elapses.

    PROCEDURE

    1. From the model number display (CEE -----SERIES 2000), press the RUN key. This

    initiates the run mode. The display will read "RUN PROG#? _," where the flashing

    cursor indicates that the system

    waits for input from the operator.

    CEE --- SERIES 2000

    PRESS RUN KEY

    RUN PROG #?_

    ENTER PROGRAM #

    No default value exists in this display; however, it is possible to find out the last pro-

    gram entered or run.

    RUN PROG #?_

    PRESS PROGRAM KEY

    (PROG)

    PROGRAM NUMBER 4

    RELEASE PROGRAM KEY

    Pressing the PROG key anytime during run mode displays "PROGRAM NUMBER 4"

    where, "4" indicates the most recently used program.

  • 8/4/2019 M2000 Hotplate

    26/43

    Page 26

    Section 6.

    RUNNING PROGRAMS:

    Warning: The surface of the chuck is hot; never touch the chuck surface. Never touch

    substrates after baking until sufficient time for cooling elapses.

    PROCEDURE continued

    To enter a user program, press the appropriate numerical key "2," then ENTER.

    RUN PROG #?_

    ENTER PROGRAM #

    (2-ENTER)

    2. The system controller instructs the temperature controller as to the temperature setpoint

    of the current program. If disabled on the temperature controller keyboard, the system

    controller enables the heat element.

    In addition, the system controller compares the chuck temperature to the error bounds

    established in the configuration procedure (see section 3 for more information).

    3. If the chuck temperature falls under the acceptable temperature range, specified in the

    system configuration, the display reads "UNDER TEMP / OVERRIDE _." This indi-

    cates that the chuck temperature is currently below the normal operating range; at thispoint, the system allows the operator to override the error.

    UNDER TEMP/OVERRIDE_

    ENTER START TO OVERRIDE

    (START)

    Pressing START at this point overrides the system and begins baking wafers.

    Similarly, if the current chuck temperature is too high, the display reads "OVER TEMP

    / OVERRIDE_" indicating that the current chuck temperature falls out of range. At this

    point, the system allows the operator to override the error.

  • 8/4/2019 M2000 Hotplate

    27/43

    Page 27

    Section 6.

    RUNNING PROGRAMS:

    Warning: The surface of the chuck is hot; never touch the chuck surface. Never touch

    substrates after baking until sufficient time for cooling elapses.

    PROCEDURE continued

    Pressing START at this point overrides the system and begins baking wafers.

    OVER TEMP/OVERRIDE_

    ENTER START TO OVERRIDE

    (START)

    Before baking each wafer, the system checks the chuck surface temperature. To insti-

    tute a global override, during the configuration procedure set the allowed percent error

    to 0% (see Section 3 for more information). When the temperature has fallen within

    specifications or has been overridden, the system will begin the program.

    4. For systems configured for manual substrate loading, the display reads "READY

    PRESS START." Press START to initiate the program.

    READY PRESS START

    PRESS START TO BEGIN

    (START)

    Systems configured for auto cassette loading skip this step.

    5. First, the system checks verifies that no substrates remain on the exit track (right-hand

    belts). If present, the display reads "PLEASE CLEAR EXIT." Remove any substrates

    from the exit track.

    PLEASE CLEAR EXIT

    REMOVE SUBSTRATE

  • 8/4/2019 M2000 Hotplate

    28/43

    Page 28

    Section 6.

    RUNNING PROGRAMS:

    Warning: The surface of the chuck is hot; never touch the chuck surface. Never touch

    substrates after baking until sufficient time for cooling elapses.

    PROCEDURE continued

    6a. Manual systems:

    At this point, the display reads "PLEASE LOAD ENTRY."

    PLEASE LOAD ENTRY

    PLACE SUBSTRATE

    ON BELT

    The entry track runs for approximately twenty seconds; after which if no substrate is de-

    tected, the track turns off, and the system reverts to step 4. Place a substrate on the en-

    try track if empty. The track carries the substrate past the entry track detector and onto

    the chuck.

    After detecting the substrate on the chuck, the Series 2000 Hotplate bakes the

    substrate as programmed. The wafer unloads at the far end of the exit track.

    Remove the finished substrate from the exit track: load a new substrate onto the entry

    track. Continue this process until completion of all substrates. For more information

    on baking styles, refer to Process Description, Section 6.

  • 8/4/2019 M2000 Hotplate

    29/43

    Page 29

    Section 6.

    RUNNING PROGRAMS:

    Warning: The surface of the chuck is hot; never touch the chuck surface. Never touch

    substrates after baking until sufficient time for cooling elapses.

    PROCEDURE continued

    6b. Cassette elevator systems:

    A two-second delay occurs in unloading the substrate from the entry (left) elevator. The

    entry track starts allowing a single "quickie" substrate to proceed through the automatic

    system. This circumvents loading the substrate into a cassette or reconfiguring the sys-

    tem for manual operation.

    After the two-second delay, the system controller requests a substrate from the entry ele-

    vator. The entry elevator unloads the substrate onto the chuck. After detecting the sub-

    strate on the chuck, the Series 2000 Hotplate bakes the wafer as programmed. The sub-

    strate unloads at the far end of the exit track.

    Upon reaching the exit detector (located at the far end of the exit track), the exit eleva-

    tor unloads the substrate from the exit track.

    The process automatically continues until: 1) an empty entry cassette or 2) a full exit

    cassette. For more information on baking styles, refer to process description, Section 6.

    7. To abort processing in the middle of a batch, press the STOP key. The system halts af-

    ter finishing the current substrate. Acknowledging this, the display reads "STOPPED -

    PRESS START." Press START and the system continues as before.

    STOPPED - PRESS START

    PRESS START TO CONTINUE

    (START)

    To stop processing and exit the run mode, press RESET. If performing a reset anytime

    other than during a stopped or idle system, clear any substrates from the chuck and exit

    track. No memory of the substrates will exist; the substrates will not bake accurately.

    8. For a complete description of possible errors, see Errors and Recovery, Section 8.

  • 8/4/2019 M2000 Hotplate

    30/43

    Page 30

    Section 7.

    PROCESS DESCRIPTION:

    Many advantages with hotplates exist compared to convection-type ovens: decreased baketime, increased reproducibility, and better film quality. This section describes some ofthese differences and sets a few guidelines for incorporating hotplates.

    CONVENTIONAL OVENS

    Stratification, the formation of different temperature zones within a unit, is a problemassociated with convection ovens. Stratification can severely affect film quality andreproducibility.

    The heating rate of a substrate depends not onlyon the air flow past the substrate but on the

    proximity to another cold substrate. Hence,the heating rate for a cassette of substratesplaced in an oven--for each substrate--willbe less than if each substrate baked alone inthe oven.

    In addition, substrates near the ends of a cassette heatfaster than substrates in the middle producing non-uniformity of heating across substrates.

    Particle generation occurs with standard ovens. In aforced-air, convection oven, substrates arecommonly exposed to a flow of unfiltered air forat least 30 minutes.

    Considerable particulate contamination during resin film cures occurs in substrates. The sub-strates are vulnerable since the film may still contain solvent; flying particles will adhereto the surface during this "soft" state.

    150

    155154

    153152

    151150

    148149

    7/Min3/Min

    7/Min

    UNEVEN HEATING

  • 8/4/2019 M2000 Hotplate

    31/43

    Page 31

    Section 7.

    PROCESS DESCRIPTION:

    Another disadvantage in normal oven bakingresults from baking substrates from the"outside in." A skin forms on the surface ofthe film trapping solvents. Upon

    vaporizing, these solvents form blisters orbubbles resulting in adhesion loss or evenbulk film failure. This problem prevails inprocesses involving thick film resins, e.g.polyimides.

    HOTPLATE BAKING

    A well designed hotplate

    insures uniform bakingacross the substrate.Since the substrateintimately contacts asurface of known,constant temperature, itheats at a ratedependent only uponthe bake style selectedand the thermal

    properties of the substrate.

    Increased throughput results from a faster warmup of the substrate. Bake times measured inseconds result rather than minutes or hours as in conventional ovens.

    Reduced vulnerability to particulate contamination is a major advantage to hotplate baking:only normal, clean room, ambient air passes over the substrate.

    No skin effect occurs on a hotplate since hotplate

    baking heats the substrate from the bottomup. This "inside out" approach offers advan-tages for thick films since solvents in thefilm nearest the substrate escape before thefilm surface seals over.

    Surface

    Temperature

    C

    140

    150

    160

    Left edge Right edgeCenter

    POSITION ON CHUCK SURFACE

    SURFACE TEMPERATURE UNIFORMITY

  • 8/4/2019 M2000 Hotplate

    32/43

    Page 32

    Section 7.

    PROCESS DESCRIPTION:

    BAKING STYLES

    Three baking styles exist on the Series 2000 Hotplates as well as combinations thereof. This

    section details the methods and the advantages of each.

    Proximity bake:

    In this method, substrates

    float on a pillow of nitro-

    gen blown through orifices

    in the chuck surface. A

    combination of heated gas

    and radiant heat from the

    chuck heats the substrate.

    This slower heating of the

    substrate reduces blister-

    ing and cracking of films

    incorporating fast drying

    solvents.

    Commonly used as a prebake

    stage and/or in combina-

    tion with hard-contact

    bake, the proximity bake

    makes two temperature

    bake schedules obsolete.

    Soft-contact bake:

    In this method, gravity alone

    holds the substrate against

    the chuck surface. Com-

    pared to proximity and

    hard-contact baking, more

    heat transfers to the sub-

    strate and slower warmup

    times occur, respectively.

    While this represents the least

    accurate bake style, this

    method finds some use as

    an intermediate betweenhard-contact and proximi-

    ty bakes.

    Hard-contact bake:

    Hard-contact bake represents

    the most accurate baking

    method for hotplates.

    Vacuum ports in the chuck

    intimately hold the sub-

    strate in place. This meth-

    od insures bake uniformity

    and minimizes bowing and

    warping of the substrate.

    A quicker warmup and more

    efficient heating produces

    faster throughput in short-er bake times. Selecting

    the VAC (vacuum) bake

    method initiates the hard-

    contact bake cycle--the

    most preferred method.

  • 8/4/2019 M2000 Hotplate

    33/43

    Page 33

    Section 7.

    PROCESS DESCRIPTION:

    REFLOW SOLDERING

    Hotplate baking heats the substrateand the solder without applying

    heat directly to the devices on

    board. Using a combination of

    proximity and hard-contact

    bakes, the bake profile can be

    adjusted to suit any process.

    SILICON WAFERS Cee Hotplates virtually eliminate the skin effect with

    thick films andsubstantially increase throughput. This chartpresents process examples for 2 commonly used res-ins: positive photoresist and polyimide. Do not useas a rigid guideline since the best method with a par-ticular application results through experimentation.

    APPLICATION OVEN BAKE HOTPLATE BAKE

    Positive photoresist 90 C/30 min 115 C/30 sec Hard contact

    Polyimide alpha (solvent removal) 90 C/30 min 150 C/15 sec Prox bake

    Polyimide beta (partial imidization) 135 C/30 min 150 C/90 sec Hard contact

    GALIUM ARSINIDESegments and pieces of GaAs wafers are commonly used in

    research and pilot lines for economic reasons. The Series

    2000 Hotplates are ideally suited for these circumstances.

    All Series 2000 Hotplates offer a proximity-bake mode.

    This is most useful for prewarming GaAs wafers before go-

    ing to a hard-contact bake and insures uniform heating with-

  • 8/4/2019 M2000 Hotplate

    34/43

    Page 34

    out thermal shock. Typical bake processes are identical to those provided above for Silicon.

    Section 7.

    PROCESS DESCRIPTION:

    PHOTOMASKS AND DISPLAYS

    Reproducibility and throughput are key issues

    both for photomask and for display

    makers. Because of the large thermal

    mass of these substrates, oven baking is

    slow and nonuniform. A proximity bake

    eliminates the back-side defects. The

    Series 2000 Hotplate can typically reduce

    bake times by 90%.

    Reproducibility is greatly improved since heating rate is not dependent on batch size. All

    substrates are baked individually.

    Particle contamination is an important factor for photomask makers and display makers alike,

    since both industries insist on zero or near zero defect yields. With the Series 2000 Hotplate,

    these substrates bake in the clean room environment for a fraction of the time necessary in a

    conventional oven.

    EXHAUST COVER

    The design of the exhaust cover promotes removal of vapors evolved from a substrate placed on

    the chuck without actually drawing air across the chuck surface.

    CHUCK SURFACE

    TO

    EXHAUST

    BLOWER

  • 8/4/2019 M2000 Hotplate

    35/43

    Page 35

    Section 7.

    PROCESS DESCRIPTION:BAKING PROCESS

    Loading, baking, and unloading on the Series 2000 Hotplates includes constant error checking

    and precise timing to ensure efficient, reliable operation.

    By placing the system in run mode and selecting a program, this informs the temperature

    controller of the new temperature set point. The system controller begins polling the

    temperature controller approximately every 1.5 seconds. If the current chuck temperature

    falls outside the specified range, the system informs the operator by displaying "UNDER

    TEMP / OVERRIDE_" or "OVER TEMP / OVERRIDE_."

    UNDER TEMP / OVERRIDE_

    PRESS START TO OVERRIDE

    (START)

    OVER TEMP / OVERRIDE_

    PRESS START TO OVERRIDE

    (START)

    At this point, the system allows overriding of the error to continue processing. After reaching

    the correct temperature or overriding the error, the system loads the first substrate.For a manually operated machine, the display reads "READY PRESS START" and waits for a

    response from the operator.

    READY PRESS START

    PRESS START TO BEGIN

    (START)

    After the operator presses the START key, the exit track checks for substrates.

    The Series 2000 Hotplates will not load the chuck with a substrate unless a clear exit pathway

    exists. This prevents overbaking of a substrate while waiting for the exit to clear or for the

    substrate on the exit to dump onto the floor.

  • 8/4/2019 M2000 Hotplate

    36/43

    Page 36

    Section 7.

    PROCESS DESCRIPTION:BAKING PROCESS continued

    When the exit clears, the entry track begins to run. Autoloading systems run this track for

    approximately 1.5 seconds before requesting a substrate from the entry elevator. This

    permits the insertion of a substrate into the baking line without having to load it into acassette.

    While the system controller waits for a substrate from the entry elevator, the display reads

    "LOADING TRACK." A delivery error occurs if the substrate fails to show up within the

    specified time; the entry track shuts down and sounds an alarm.

    After an autoloaded or manually loaded substrate reaches the entry track sensor, the display

    reads "LOADING CHUCK."

    Next, the gas transfer and gas pillow jets turn on to carry the substrate to the stop pins and

    chuck sensor. Again, an allotted time period exists for arrival of the substrate to the chuck

    sensor before displaying a delivery error.

    The substrate arrives at the chuck sensor and stop pins almost simultaneously. As soon as it is

    detected the transfer and gas pillow jets turn off, the vacuum valve opens for 20

    milliseconds, and the wafer halts.

    Initiation of the programmed bake method starts by setting the necessary valves: gas jets,

    proximity; vacuum, hard contact; or neither, soft contact. Multiple bake cycles consecutive-

    ly run until program completion.

    After completion of substrate baking, the stop pins lift out of the way; the transfer and gas

    pillow jets turn on; the exit track starts.

    Again a time-limit delay occurs to permit arrival of the substrate at the exit sensor before

    issuing a delivery error. Upon arrival at the exit, the stop pins drop; gas jets turn off; and

    the exit track stops.

    At this point, manual systems request the operator to remove the finished substrate from the exit

    track; autoloading systems order the exit elevator to unload a substrate. If the elevator

    signals a ready condition, the exit track starts again, and the wafer unloads.

  • 8/4/2019 M2000 Hotplate

    37/43

    Page 37

    Section 7.

    PROCESS DESCRIPTION:

    BAKING PROCESS continued

    Usually, no time lapses between the exit track stopping and starting up again. Display of a

    delivery error proceeds if the exit elevator does not receive a substrate within the allottedtime.

    After baking and unloading a substrate, the controller checks the chuck temperature. Unless

    previously overridden, the next substrate loads and runs in a like manner. Continuation of

    this process occurs until no substrates remain, or the operator halts the system.

  • 8/4/2019 M2000 Hotplate

    38/43

    Page 38

    Section 8.

    OPTIONS:

    The Series 2000 Hotplates are designed to be quite flexible in operation as well as in modifica-

    tion. This section describes both the OPT ("option") key which utilizes custom or special-

    ized software and the hardware options currently available for the Series 2000 Hotplates.

    THE OPTion KEY

    Current Series 2000 Hotplates ship with one optional function installed: the single step function

    Information on additional packages will be distributed as they become available.

    The single step function allows the operator to bake wafers without prior system programming.

    As a result, five or ten substrates may be baked with different methods and different bake

    times without having to preset ten user programs.

    All options are accessible from the model number display (CEE ----- SERIES 2000). Option 1

    contains the single step function.

    CEE --- SERIES 2000

    PRESS OPT KEY

    To enable this function, press the OPT key. The display reads "OPTION--." The system waits

    for input from the user.

    OPTION--

    ENTER OPT #

    Press "1," and the display reads "OPTION-- SSTEP."

    OPTION--

    ENTER OPT #

    (1-ENTER)

    Press ENTER; this engages the single step function which remains in effect until pressing the

    RESET key.

    OPTION-- SSTEP

    (ENTER)

  • 8/4/2019 M2000 Hotplate

    39/43

    Page 39

    Section 8.

    OPTIONS:

    THE OPTion KEY continued

    Next, press the RUN key to enter run mode as in Section 5.

    CEE --- SERIES 2000

    PRESS RUN KEY

    The display reads "RUN PROG# ? _ SSTEP" indicating that the selected program will be

    run in single step mode.

    RUN PROG# ? _ SSTEP

    ENTER PROGRAM #

    In actuality, the system only uses bake temperature from the preset program.

    At which point the temperature falls within range or override occurs, the operator chooses the

    system bake method. Choose a method "PROX," "CONT," or "VAC," and press ENTER.

    PROG# SSTEP

    ENTER NEW METHOD

    (PROX, CONT, OR VAC-ENTER)

    The substrate loads onto the chuck and bakes as usual; except, the display counts up from 0.

    After reaching the desired bake time, press the STOP key: the substrate unloads from the

    chuck.

    Processing continues in this fashion until pressing the RESET key which restores the normal

    operating mode.

  • 8/4/2019 M2000 Hotplate

    40/43

    Page 40

    Section 8.

    OPTIONS:

    HARDWARE OPTIONS

    1. CASSETTE ELEVATORS: A pair of our cassette elevators converts a manual Se-

    ries 2000 Hotplate into a fully autoloading system. In-terfaces are already built into the machine.

    2. FIELD INSTALLABLE Retrofits, which can be installed in the field, allow the

    RETROFITS FOR baking of substrates other than those for which the

    CHANGING SUBSTRATE system was originally designed. Preferred by research

    SIZE: laboratories and pilot lines.

    3. MULTI-SUBSTRATE This chuck adjusts to handle four different

    CHUCK: substrate sizes. If installed, an appendix to this manu-

    al details the operation of this option.

  • 8/4/2019 M2000 Hotplate

    41/43

    Page 41

    Section 9.

    ERRORS AND THEIR RECOVERY:

    The Series 2000 Hotplates contain extensive error checking routines. These minimize problems

    while in use, as well as, point out potential problems before they become serious.

    The following lists the possible error messages ("display messages"), their cause, and the cor-

    rective actions necessary.

    ERROR: CAUSE: ACTION:

    Delivery error entry Substrate did not leave the Reset substrate on entry track and

    entry track as specified. press START.

    Delivery error chuck Substrate did not arrive Reset substrate on entry track and

    at chuck as specified. press START. May indicate

    improper leveling of machine.

    Delivery error exit Substrate did not arrive Reset substrate on exit track and

    at the exit track as press START. May indicate

    specified. chuck needs cleaning.

    Wafer lost entry Substrate did not transfer Replace substrate into cassette.

    from elevator to May indicate poor alignment

    entry track. of elevator to unit.

    Wafer lost exit Substrate did not transfer Replace substrate into cassette.

    from exit track to elevator. May indicate poor alignment

    of elevator to unit.

    Temperature controller Electrical malfunction in Turn system off; Wait 5 seconds;

    not responsive ("Temp temperature controller Turn back on; Check for loose

    contr not resp") cables.

    Temperature controller Error in communication to Press START to re-attempt.

    communication error temperature controller

    ("Temp contr com error")

  • 8/4/2019 M2000 Hotplate

    42/43

    Page 42

    Section 9.

    ERRORS AND THEIR RECOVERY:

    ERROR: CAUSE: ACTION:

    Temperature controller Temperature controller not Enable remote communication,

    disabled ("Temp contr able to comply with see Section 2, Temperaturedisabled") command. Controller.

    Temperature controller Unforeseen error with Contact Cee Engineering Dept.

    error ("Temp contr temperature controller.

    error")

    Cassette full exit Maximum capacity reached Replace full exit cassette with an

    in the exit cassette. empty cassette.

    Cassette empty entry No substrates in entry Replace empty cassette with a

    full

    cassette cassette.

    Reset Cassette (entry/exit) Elevator does not recognize Lift cassette from platform and

    cassette, or cassette is replace.

    misaligned.

    Need cassette (entry/exit) Cassette missing Place cassette on platform.

    Elevator error (entry/exit) Unforeseen error with Contact Cee Engineering Dept.

    elevator

    System dead, no lights No power to outlet Check electrical outlet for

    on temperature controller presence of 120 Volts, AC.

    Blown fuse Replace fuse (located at the back

    of unit).

    Nitrogen and/or Blockage in tubing Check connections to the

    vacuum not functional unit for possible kinks or

    obstructions.

  • 8/4/2019 M2000 Hotplate

    43/43

    Section 9.

    ERRORS AND THEIR RECOVERY:

    ERROR: CAUSE: ACTION:

    Temperature controller Low voltage Check electrical outlet forreads "E1" presence of 120 Volts, AC.

    Temperature controller Electrical problems Remove power for 5 seconds and

    reads "E2," "E3," could be caused by restore, otherwise contact Cee

    "E5," or "E6" noise on power line. Engineering Dept.

    Temperature controller Calibration lost Contact Cee Engineering Dept.

    reads "E4"

    Cee A di i i f B S i I P O B GG R ll MO 65402