SC-FAST Seawater Analysis M. Paul Field, Institute of Marine and Coastal Sciences Rutgers The State University of New Jersey

Transcript of M. Paul Field, Institute of Marine and Coastal Sciences...

Page 1: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

SC-FAST Seawater Analysis

M. Paul Field,

Institute of Marine and Coastal SciencesRutgers The State University of New Jersey

Page 2: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

SC-2 with FAST Flow Injection

Reduce analysis timeReduce memory effectsReduce ICP-MS cone salting and injector pluggingImprove instrument efficiency

Page 3: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

SC-Fast Sea Water analysis on Element-XR

10% sea water (SW) diluted with 3% HNO3. (Real samples)Y internal standardAnalysis Time

SC-FAST 10 sec load time; 2 min analysis time.SIS 1 min load time; 2 min analysis time; 1 min rinse

120 unknown samples.70 standards, blanks and CRMs.Matrix match Std curve4 point curve,1 Blank, 1 NASS-5 every 12 unknownsTotal 180 injections of SW and 10 injections of acid blank (only SW injections plotted).No cleaning of cones between day

Page 4: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Pressurized Gas Delivery System of Carrier Solution

Pulse free deliveryControlled low flow ratesNo peristaltic pump tubingEnsures low background

Page 5: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

System set-ups

µFlow (160µL/min)

PFA spray Chamber

1500W

Tuned on 10% SW

10 sec uptake, 2 min analysis 10% SW

2min 10 sec/sample 120 unknowns = 7.25 hrs

µFlow (160µL/min)

PFA spray Chamber

1500W

Tuned on 10% SW

1 min uptake, 2 min analysis,1 min wash

10% SW

4 mins/sample 120 unknowns =14hrs

Sea spray (600µL/min)

Glass spray Chamber

1500W

Tuned on 10% SW

1 min uptake, 2 min analysis,1 min wash

10% SW

4 mins/sample 120 unknowns =14hrs

SC-FAST + CH4PFA MicroFlowStandard Set-up

Page 6: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

0.00

0.25

0.50

0.75

1.00

1.25

0 75 150 225 300 375 450 525 600 675 750

Number of analysis (10% SW)

Rel

ativ

e se

nsiti

v

Day 1 Day 2 Day 4Day 3

Sensitivity Drift (Y)

NebulizerSeaspray(600 µL/min)

NebulizerPFA µFlow (160µL/min)

SC-Fast + CH4(160µL/min)

Page 7: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Cone DepositionCones after Sea Spray

120 analysisCones after µFlow

540 analysisCones after CH4 SC-Fast

740 analysis

• 30% reduction in salt loading

• More effective rinsing between samples

• <20% Drift

Page 8: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Analysis Time

7.5sec25secDead Time/sample

6hr 45minTime saved per run =

7hr 15min14hrTotal time

190190No. of samples

2min 10sec4minSample Measurement time

SC-FASTStandard Introduction System

Page 9: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Tuning

1ppb Tune, PFA-ST flow rate 160µL/minPFA Spray chamber with CH4 addition

In U UO %

Page 10: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

NASS-5 (n=38) 4 Days

0

2

4

6

8

10

12

14

16

18

20

P V Mn Mo Ba U AsElement

Con

cent

ratio

n (p

p

CertifiedDet (n=38)

P = 17.3 ± 0.9 (2σ )

Ba = 5.3 ± 1.6 (2σ )

Errors are < ± 6% 2σ

Page 11: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

SC-Fast:Detection Limits

Detection limits (160µL/min, PFA spraychamber )SC-FAST with CH4 addition

Blank DL(ppt) (ppt)

P MR 67.50 5.00V MR 0.18 0.18

Mn MR 1.00 0.50Mo MR 0.64 1.11Ba MR 0.40 0.49U MR 0.00 0.04

As HR 0.21 0.30

Element Resolution

Page 12: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Summary

Maintains instrument sensitivity, accuracy and stabilityMore effective rinsing of introduction systemReduces total matrix reaching conesExtends time between cone maintenanceDramatically reduces total analysis time

Page 13: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Future Applications SC-FAST

Automated LC-ICPMS interfaceMultiple injections per single analysis

Minimize sample loading to the ICPMS cones of aggressive samples (e.g. H3PO4)Multiple collision/reaction cell conditions