LXi.pdf
-
Upload
luis-fernando-garcia-s -
Category
Documents
-
view
232 -
download
3
Transcript of LXi.pdf
HiSpeed LX/i0459
*
Operator Manual
2295148–100
Revision 0Original Draft in English
Copyright 2001by General Electric Company
WARNING
X–RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATORUNLESS MEASURES OF PROTECTION ARE STRICTLY OBSERVED
Though this equipment is built to the highest standards of electrical and mechanical safety, the useful x–ray beam becomesa source of danger in the hands of the unauthorized or unqualified operator. Excessive exposure to x–radiation causes damageto human tissue.
Therefore, adequate precautions must be taken to prevent unauthorized or unqualified persons from operating this equipmentor exposing themselves or others to its radiation.
Before operation, persons qualified and authorized to operate this equipment should be familiar with the Recommendationsof the International Commission on Radiological Protection, contained in Annals Number 26 of the ICRP, and with applicablenational standards.
iCT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Regulatory Requirements
This product complies with regulatory requirements of the following:
� European Directive 93/42/EEC concerning medical devices 0459
The configurations delivered:
– without any option or accessory,or
– exclusively with one or several options and/or accessories listedbelow, have been declared in compliance with the EuropeanDirective 93/42/EEC and therefore bear the CE marking:
� Green QSD 1990 Standard issued by MDD (Medical Devices Director-ate, at the UK Department of Health).
� Medical Device Good Manufacturing Practice Manual issued by theFDA (Food and Drug Administration, United States Department ofHealth).
� Requirements from Underwriters Laboratories, Inc. (UL), independentlaboratory.
� Requirements from Canadian Standards Association (CSA).
� Requirements from the International Electrotechnical Commission(IEC), international standards organization.
General Electric Medical Systems, the manufacturer of this product, is ISO9001 certified.
iiCT HiSpeed LX/i Operator Manual
Dir. 2295148–100Rev.0
EMC Performance
This CT system may characteristically cause electromagnetic interferencewith other equipment, either through air or connecting cables. The term EMC ( ElectroMagnetic Compatibility ) means the capability of theequipment, which curbs electromagnetic influence from other equipmentand at the same time does not affect other equipment with similar electro-magnetic radiation from itself. This product is designed to fully comply withthe group 1 EN60601–1–2 ( IEC60601–1–2 ), Class A, in EMC regulations.In terms of EMC performance, limit value 12dB relaxation in electromagnet-ic interference of permanently–installed medical system used in a lead–shielded area, is applied to this product.
Notice upon Installation of the Product
� Locate the equipment as far as possible from other electronic equip-ment.
� Be sure to use either any cables provided by GEYMS or ones desig-nated by GEYMS. Wire these cables following these installation proce-dures.
General Notice
� Designation of Peripheral Equipment Connectable to This Product
Avoid using other equipment than designated. Failure to comply with thisinstruction may result in poor EMC performance of the product.
� Notice against User Modification
Never modify this product. Unilateral user modification may cause degrada-tion in EMC performance. Modifications of product include ;
� Changes in cable ( length, material, wiring etc.)
� Changes in system installation/layout
� Changes in system configuration/components
� Changes in means of fixing the system/parts ( cover open/close,cover screwing )
� Operate the system with all covers closed. If you open any cover forsome reason, be sure to shut it before starting/resuming operation.Operating the system with any cover open could affect EMC perfor-mance.
iiiCT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Countermeasures against EMC–related Issues
Generally it is pretty difficult to grapple with EMC–related issues.It may take much time and cost.
General countermeasures
Electromagnetic interference with other equipment
� Electromagnetic interference may be alleviated by positioning otherequipment far from the system.
� Electromagnetic interference could be mitigated by changing therelative location/installation angle between the system and otherequipment.
� Electromagnetic interference may be eased by changing wiringlocations of power/signal cables of other equipment.
� Electromagnetic influence could be reduced by altering the path ofpower supply for other equipment.
Notice on Service
� Ensure all screws are tight after servicing. Loose screws may cause the degradation in EMC performance.
� In case the high frequency gasket of this system is broken, replace itwith a new one immediately.
ivCT HiSpeed LX/i Operator Manual
Dir. 2295148–100Rev.0
Blank page.
Safety
vCT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Safety !
WARNING!
FAILURE TO FOLLOW THE OPERATING INSTRUCTIONS ANDSAFETY PRECAUTIONS COULD RESULT IN INJURY TO THEPATIENT, YOURSELF, OR OTHERS.
General Safety
� Keep the patient in view at all times. Never leave the patient unattended.Always stay alert to safety concerns involving the patient’s condition andequipment operation.
� Check the lengths of all patient health lines (IV tubing, oxygen lines, etc.) andmake sure they accommodate cradle travel. Position these lines so theycannot catch on anything within the patient vicinity or between the table andgantry during cradle travel or gantry tilt.
� Always follow the exam procedures provided in the operator manual. Verifycorrect entry of identification, patient positioning, and other patient databefore proceeding with the exam. Incorrect procedures or patient data entrycould cause misinterpretation of the exam results.
� Conduct Image Quality checks and follow the maintenance schedule outlinedin your HiSpeed LX/i operator manual. Discontinue use of the equipmentwhenever you notice equipment damage, or a malfunction occurs. Do notuse the equipment until qualified service personnel correct the problem.Never use the equipment unless all the protective covers are in place.
Radiation Safety
� Use of controls or adjustments, or performance of procedures other thanthose specified herein may result in hazardous radiation exposure.
� Always use proper technique factors for each procedure to minimize X–Rayexposure while still producing the best diagnostic results.
� Be sure to have an operator wear X–ray protection vest when a work isneeded near the gantry while X–ray is being emitted.
Safety
viCT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Mechanical Safety
� Never open, or remove, the gantry covers. (Only qualified service personnelshould remove covers.)
� Do not enter the scan room when the gantry covers have been opened orremoved. Never allow a patient or staff member to enter the scan room whenthe gantry covers have been removed for maintenance or PM. (Never allowanyone but qualified service personnel to enter the scan room during gantrymaintenance or PM.)
� To prevent the pinching or crushing of extremities, keep hands and feet awayfrom the edge of the moving table top/cradle and its surrounding equipment.(Be especially careful when positioning patients who weigh more than 250pounds.)
� Physically assist all patients on and off the table, and into position on thecradle.
� While a patient is being loaded onto a CT table from a gurney, make sure viathe following measures that the gurney NEVER moves.
� Lock all the casters of the gurney.
� Hold the gurney very firmly.
If the gurney moves while loading a patient, it may create a gap between thetable and the gurney posing the danger of the patient falling through the gap.
� Return the gantry to the 0� upright position, latch the table, and set it at acomfortable height for patient loading and unloading.
� Avoid any patient contact with the CT gantry during gantry tilt and cradlemovement (manual or software driven). Once again, pay close attention tolarge patients; make sure you don’t pinch skin or extremities between thecradle and the gantry.
� The concentrated weight of short, heavy patients can cause the cradle tomake contact with the gantry. Make sure you don’t drive the cradle into thegantry cover, and make sure you don’t pinch skin or extremities between thecradle and the gantry.
� Check the lengths of all patient health lines (IV tubing, oxygen lines, etc.) andmake sure they accommodate cradle travel. Position these lines so theycannot catch on anything within the patient vicinity or between the table andgantry during cradle travel or gantry tilt.
Safety
viiCT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
� Do not use the table base as a foot rest. You could entrap and injure yourfoot while lowering the table.
� Don’t place your hands inside the gantry cover when tilting the gantry.The gantry can pinch or crush your hands!
� Only use the cradle extender to support the patient’s head or feet during ascan study. The cradle extender supports up to 75 pounds; the head holdersupports up to 30 pounds. Neither device supports the full weight of apatient. If you sit, stand or otherwise apply excessive pressure to thesedevices, they will break or come off the cradle, and may cause injury.
� Periodically check all accessories for damage and remove from service ifdamaged or cracked. Also check the accessory attachment plate fixed to theend of the cradle. Repair or replace if loose or damaged.
� The cradle has a maximum distributed load capacity of 450 lb. (205 kg).GE assures incremental accuracy and normal traverse speeds up to 300pounds. Exceeding the 450 lb. (205 kg) maximum limit could result indegraded positioning performance, increased table lowering speed,equipment damage and/or injury.
Safety
viiiCT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Electrical Safety
� Avoid all contact with any electrical conductors.
� Wait at least 10 minutes after the last scan before you turn off [Main Power].This gives the circulating oil sufficient time to cool the X–Ray tube.
Laser Beam Safety
� The laser beam used for patient positioning can cause eye injuries.Instruct a patient not to stare into the laser beam.
Software Safety
� Should a malfunction occur, or a patient condition develop that requires
interrupting a scan series, push (Abort) on the right end of the OC scan
panel to stop X–Ray.
� Wait for image annotation to complete before filming. If you press theexposure button too soon, the film records a composite of two different scans.
� DICOM protocol has a “dialect”, which may cause some troubles likedisappearances of some portion of image annotations, when connecting tothe station where data transfer is not confirmed.
Emergency Stop
� In the event of a hardware failure that could cause serious damage, such as
smoke, fire or unintentional cradle movement, press one of the red [Emergency Stop] switches located on the operator console, at the gantrycontrol panels, or on the X–ray Generator. Low power to the electroniccomponents in the computer and data acquisition system remains ON.
� When Emergency Stop is applied, the moving cradle and tilting gantry mayoverrun by less than 10 mm and less than 0.5 degrees, respectively.
GE provides training support. Contact your local GE salesrepresentative to arrange training sessions to meet your needs.
Safety
ixCT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Radio Waves Safety
Never use the following devices near this equipment.Use of these devices near this equipment may induce erratic function of the equipment.
Devices not to be used near this equipment
Devices which intrinsically transmit radio waves such as;cellular phone, transceiver, mobile radio transmitter and radio–controlled toy,etc.
Keep those devices power–off near this equipment.
Medical staff in charge of this equipment is required toinstruct technicians, patients and other people who may bearound the equipment to fully comply with the aboveregulation.
Your system also has a warning label on a console thatwarns of the above hazard.
NOTE�
NOTE�
Safety
xCT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Warning Labels
Labels on Keyboard
� The following labels are attached to the upper side of the keyboard.
Patient may be pinched between gantry and table duringPrescribed Tilt. To avoid pinching, before using PrescribedTilt, make sure that patient does not contact the gantry.
CAUTION
Press Stop Scan button to stop cradle motion or scanning.
CAUTION
This X–ray unit may be dangerous to patient and operatorunless safe exposure factors and instructions are observed.
WARNING
Safety
xiCT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Laser Alignment Light Caution/Warning
The HiSpeed system employs semiconductor laser alignment lights for patientlandmarking. This type of alignment light casts a thin red light on the patientfor the purpose of positioning and landmarking.
� The following label is attached at five locations of the gantry aperture.
LASER APERTURE
DO NOT STAREINTO BEAM
� The following labels are attached to the lower part of the gantry front cover.
LASER RADIATIONDO NOT STARE INTO BEAM
600–700nm LASER DIODE1 MILLIWATT MAXIMUM OUTPUT
CLASS 2 LASER PRODUCTIEC 825–1:1993
LASER RADIATIONDO NOT STARE INTO BEAM
600–700nm LASER DIODE1 MILLIWATT MAXIMUM OUTPUTCLASS II LASER PRODUCT
CAUTION: Exposing eyes to the laser alignment lights may result in eye injury.
Do not stare directly into the laser beam.
Instruct patients to close their eyes during landmarking to avoideye exposure to the alignment light while laser light is on.
Closely monitor all patients and prevent them from accidentallystaring into the beam.
Do not leave the laser beam on after you position the patient.
Safety
xiiCT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
� The following label is attached at the rear side of the gantry pedestal.
MADE FOR GENERAL ELECTRIC CO.MILWAUKEE, WISCONSIN BY
GE YOKOGAWA MEDICAL SYSTEMS, LTD7–127 ASAHIGAOKA 4–CHOME, HINOSHI, TOKYO, JAPAN
MODEL xxxxxSERIAL yyyyyMANUFACTURED n month, 19xx CLASS 1
SOURCE: 90A/380–480V3~ 50/60HzSOURCE: 30A/200V~ 50/60HzSOURCE: 30A/115V~
~50/60Hz
SOURCE: 15A/115V 50/60Hz
COMPLIES WITH RADIATIONPERFORMANCE STANDARDS, 21 CFR SUBCHAPTER J
CAUTION : Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
Safety
xiiiCT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Information of Laser Alignment Lights Beam
Beam source location
The following diagram shows the locations of the laser beam source, A, B, C,D, E, F, G and H.
PositioningLight C, F
PositioningLight A, D
Gantry Front
PositioningLight B, E
G H
X–axis direction
Y–axisdirection
External plane
Scan plane
Gantry front cover
Light A Light C
90 deg.
Beam radiation area
X–axis View
Gantry front cover
90 deg.
Safety
xivCT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scan plane
Gantry front cover
Light B
90 deg.
Y–axis View
External plane
Gantry front cover
Beam radiation area
4 deg.
Front View
Light D
Light F
80 deg.80 deg.
Beam radiation area
External plane
Safety
xvCT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
3 deg.
Front View
Light E
90 deg.
Beam radiation areaExternal plane
Front View
Light G 90 deg.
Beam radiation area
Light H90 deg.
55 deg. 35 deg.
Scan plane
Safety
xviCT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Cradle Caution
� The following label is attached to the end of the cradle.
Do not grasp the side of the cradle.
CAUTION
Accessory Caution
� The following labels are attached to the standard headholder and shallowhead holder.
Accessory may fall and cause injury if not latched tocradle. Make sure that accessory is latched to underside ofcradle.
CAUTION
Excessive weight can break accessory and cause injury.Do not load more than 34 Kg or 75 pounds.
CAUTION
Safety
xviiCT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
� The following labels are attached to the coronal headholder.
Accessory may fall and cause injury if not latched tocradle. Make sure that accessory is latched to underside ofcradle.
CAUTION
Excessive weight can break accessory and cause injury.Do not load more than 34 Kg or 75 pounds.
CAUTION
Do not hit the accessory against the gantry. Patient injuryor equipment damage could result.
CAUTION
� The following labels are attached to the cradle extender.
Accessory may fall and cause injury if not latched tocradle. Make sure that accessory is latched to underside ofcradle.
CAUTION
Excessive weight can break accessory and cause injury.Do not load more than 45 Kg or 99 pounds.
CAUTION
Safety
xviiiCT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Radio Wave Caution
� The following label is attached to the operator console.
Do not use the following devices near this equipment.Cellular phone, radio transceiver, mobile radio transmitter,radio–controlled toy, etc.
CAUTION
Use of these devices near this equipment could cause thisequipment to perform outside the published specifications.
Keep power to these devices turned off when near thisequipment.
Safety
xixCT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Precaution for The CT System Disposal
The elimination of machines and accessories must be in accordance withnational regulations for waste processing.
In the CT system there are certain materials used that could causeenvironmental hazards if disposed of improperly. Those materials include leadblocks in the gantry and oil in the tank and the X–ray tube. Be sure to contactGEMS service personnel prior to disposing of the CT system or any CTcomponent to ensure safe and proper disposal.
Packing Materials
The materials used to pack our equipment are recyclable. They must becollected and processed in accordance with the regulations in force for thecountry where the machines or accessories are unpacked.
Safety
xxCT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Technical documents
GE will provide CT users with the following documents at their request to helpthem repair malfunction on their own.
� Diagrams
� Components part lists
� Descriptions
� Calibration instructions
iCT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Table of Contents
Chapter 1
SYSTEM 1-1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Overview 1-1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General Procedure of CT Examination 1-1. . . . . . . . . . . . . . . . . . . . . . . Main Components 1-2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Gantry 1-2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Gantry Control Panel 1-4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Gantry Display Panel 1-7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table 1-9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Operator Console (OC) 1-11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
User Interface 1-12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Mouse 1-12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Keyboard 1-13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessories 1-20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard Accessories 1-20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Head Accessories 1-21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Body Accessories 1-25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
System Power On/Off 1-27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Chapter 2
SCAN TOP LEVEL SCREEN 2-1. . . . . . . . . . . . . . . . . . . . . . . . . . Icon Selection 2-2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Function of Each Icon 1 2-3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Function of Each Icon 2 2-4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . New Patient 2-5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Patient Schedule 1 2-6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Protocol Management 2-12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Auto Voice Record 1 2-13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Protocol Management 1 2-16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Retro Recon 1 2-21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Recon Management 1 2-25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tube Warmup 2-27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Scanner Utilities 1 2-28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Raw Data Function 2-28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . User Calibration 2-34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
iiCT HiSpeed LX/i Operator Manual
Dir. 2295148–100Rev. 0
Chapter 3
DISPLAY TOP LEVEL SCREEN 3-1. . . . . . . . . . . . . . . . . . . . . . . Primary/Secondary Viewport 3-2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Image Display Layout 3-3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Autoview Layouts 1 3-4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Review Layouts 3-6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . List/Select 1 3-7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Routine Display 1 3-9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Roam Zoom 3-10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Explicit Magnify 3-11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Flip/Rotate 3-12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ProView 3-13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . GSE (Gray Scale Enhancement) 3-14. . . . . . . . . . . . . . . . . . . . . . . . . Display Normal 3-14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . List/Select 3-15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ellipse ROI 3-15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measure Distance 3-16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Grid 3-16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cross Reference 3-17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . User Annotation 3-18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exam Pg / Series Pg 3-19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Hide/Show Graphics 3-20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Erase 3-20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Screen Save 3-20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurements 1 3-21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MIROI (Multiple Image ROI) 3-22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . Report Pixels 3-23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display Preference 1 3-24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Annotation Levels 3-25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Next/Prior Each Viewport 3-27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Next/Prior Series Binding 3-27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous Report Cursor 3-27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Manual Film Composer 3-28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto Film Composer 3-29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Accelerator Bar 1 3-30. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
iiiCT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Chapter 4
SCAN 4-1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Patient Positioning 1 4-1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . New Patient 1 4-4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Scout Scan 1 4-6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Auto Store 4-11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto Transfer 4-12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Axial/Helical Scan Prescription 1 4-13. . . . . . . . . . . . . . . . . . . . . . . . . . . . Axial/Helical Prescription (View/Edit) Screen 4-13. . . . . . . . . . . . . . . Continuous Scan Modes 4-14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Prescribe Scan Parameters 4-15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . Prospective Multiple Reconstruction 4-21. . . . . . . . . . . . . . . . . . . . . . Icons of the Axial/Helical Scan Prescription screen (1) 4-23. . . . . . Icons of the Axial/Helical Scan Prescription screen (2) 4-28. . . . . . Show Localizer 4-31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto mA 4-33. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SmartRecon (Option) 4-38. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Axial/Helical Scan in Progress 4-40. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Axial/Helical Scan End 4-41. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Priority Recon 4-42. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Next Series 4-43. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Repeat Series 4-44. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . One More 4-46. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Repeat Last Group 4-47. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Biopsy Scan 1 4-48. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SmartPrep (Option) 1 4-50. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Smart Addition (Option) 1 4-56. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cardiac Gating (Option) 1 4-61. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Chapter 5
FILMING 5-1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . AutoFilm 1 5-1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . AutoFilm Composer 5-10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Manual Film 5-11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Manual Film Composer 1 5-12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ivCT HiSpeed LX/i Operator Manual
Dir. 2295148–100Rev. 0
Chapter 6
QUALITY ASSURANCE 6-1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Phantom Description 6-2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Scan the QA Phantom 6-4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . High Contrast Spatial Resolution 6-6. . . . . . . . . . . . . . . . . . . . . . . . . . . . Slice Thickness 6-7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Low Contrast Detectability 6-9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Noise and Uniformity 6-10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Weighted CTDI 6-12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Dose and Performance 1 6-13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Dose and Performance 2 6-17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Dose and Sensitivity Profile at Phantom Center 6-21. . . . . . . . . . . . . . .
Chapter 7
TECHNICAL SPECIFICATIONS 7-1. . . . . . . . . . . . . . . . . . . . . . . Component Identification for HiSpeed 7-1. . . . . . . . . . . . . . . . . . . . . . . . Component Labels 7-2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MX165ZJ X–Ray Tube Assembly Information 7-3. . . . . . . . . . . . . . . . . MX165ZJ Diagnostic Source Assembly 7-3. . . . . . . . . . . . . . . . . . . . . . MX165ZJ CT Scan Ratings 7-4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MX165ZJ X–Ray Tube IEC Information 7-6. . . . . . . . . . . . . . . . . . . . . . MX165ZJ Cathode Emission and Filament Characteristics 7-7. . . . . . Tube MX165 ZJ Anode Heating and Cooling Curves 7-9. . . . . . . . . . MX165ZJ X–Ray Tube Assembly Information 7-10. . . . . . . . . . . . . . . . . Generator Specifications 7-13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The Attenuation Equivalent 7-15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Periodic Maintenance by Qualified Personnel 7-16. . . . . . . . . . . . . . . . . Maintenance by CT Users 7-18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Symbols and Classification 7-19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning 7-23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Index
Chapter 1
1-1CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
SYSTEM
Overview of CT System and Examination
Overview
The HiSpeed computed tomography (CT) system consists of four majorcomponents and some peripheral gear. Four major components are thegantry, table, operator console(OC) and power distribution unit.
The primary objective of this CT system is to obtain for medical diagnosesthe two or three–dimensional images of internal structure of human body.
General Procedures of CT Examination :
1. �The operator prescribes the examination(scan) procedures at the OCwhile the patient is positioned on the table in the scan room.
2. �As the operator initiates the scan procedures with the patient in thegantry aperture, the X–ray unit inside the gantry rotates around thepatient body emitting the fan–shaped X–ray beam to the body. TheX–ray penetrates the body and is received by the detector, then it isconverted to the electrical data.
3. �The data acquired in the above step is processed into the images bythe computer at the OC. Then, the images are initially displayed on theCRT of the OC and subject to manipulation for analysis. The images canalso be filmed later for diagnoses by a physician.
4. �The image data can be saved in some electronic media like optical diskfor later analysis and also be transferred to some other stations forremote diagnoses.
System
1-2CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Main Components
Gantry
EmergencyButton
Control Panel
EmergencyButton
Control Panel
PositioningLight
PositioningLight
Display Panel
PositioningLight &Breath Navi
Front
Function
The gantry incorporates the X–ray tube unit, the HiLight detector and DAS(Data Acquisition System) inside. It also provides the following functions.
� Display Panel
The display panel shows the readings of the gantry tilt, table height, positionof land mark, latch status, scannable range and tilt range.Refer to the Gantry Display Panel page for the function descriptions.
� Emergency Button
Pressing the emergency button stops every mechanical movement and theX–ray emission.
� Control Panel
The control panel incorporates the several buttons to mainly control themovements of the gantry and table. Each front and rear cover has twocontrol panels. Refer to the Gantry Control Panel page for the functiondescriptions.
� Positioning Light & Breath Navi
The laser beam will be emitted through here that will be used to position apatient. Breath Navi gives the visual breathing instructions to a patient withhearing problem.
System
1-3CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Gantry Specifications
� Aperture : 70 cm
� Tilt : +/– 30 degrees
� Tilt speed : 1 degree/second
� Focus to isocenter : 541 mm
� Focus to detector : 949 mm
� Rotation speed : 360 degrees in 0.8, 1.0, 1.5, 2.0, 3.0 seconds
X–ray Tube Specifications
� Anode heat storage capacity : 3.5 MHU
� Anode heat dissipation : 820 KHU/min. (maximum)
� Casing heat dissipation : 300 KHU/min. (constant)
� Dual focal spots : 0.9 mm�0.62 mm, 1.2 mm�1.2 mm (Focal size depends on slice thickness and mA.)
HiLight Detector Specifications
� Channels : 816
� View : 972
System
1-4CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Gantry Control Panel
Gantry Tilt
Cradle Out
Table Up
PositioningLight
Table Down
ExternalLandmark
Fast
IMS TableMove
Cradle In
Internal Landmark
Practice
Tilt Range &ScannableRangeDisplay
Gantry TiltLamp
Gantry TiltLamp
ExternalLandmarkLamp
LED
Function
The gantry control panel enables you to move or control the movement ofthe gantry, table or cradle with the following buttons.
� Cradle In/Out
Use this to move the cradle toward or away from the gantry aperture.
� Table Up/Down
Use this to move the table up or down.
� Fast
Use this to increase the speed of cradle in/out. This does not apply to tableup/down.
System
1-5CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
� Gantry Tilt
Use this to tilt the gantry toward or away from the cradle.
WARNING!
The touch sensor on the gantry cover works only during the remote tiltoperation at the OC. It does not work when the tilt operation is done at theControl Panel.
� Gantry Tilt Lamp
The system turns on this lamp when gantry tilt is needed.
� IMS Table Move
Use this to manually extend the cradle (IMS table). The IMS table can beoperated only at the gantry control panel. The LED above the IMS Movebutton indicates the status of the IMS table use (not at the home position).The scannable range sticker on the cradle shows the edge of scannablerange with IMS table fully stretched.
� Positioning Light
Use this to turn on or off the internal/external positioning lights.
WARNING!
The laser beam used for patient positioning can cause eye injuries. Be sure to instruct a patient never to stare into the beam.
� External Landmark
Use this to move the anatomy beneath the external positioning light. When the gantry tilt is set at zero degrees, this button must be depressed formore than 0.5 seconds. Make sure that the [Distance from Landmark]display on Gantry Display Panel shows the number around 190 mm. When the gantry is tilted, keep depressing the button until the ExternalLandmark Lamp gets dim in order for the system to adjust the table height.
� External Landmark Lamp
While this lamp is on, keep depressing the External Landmark button untilthis lamp gets dim to adjust the table height.
System
1-6CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
� Internal Landmark
Use this to move the anatomy beneath the internal light to adjust scanlocation zero.
� Tilt Range & Scannable Range
Use this to display the gantry tilt range and the scannable range at certaintable height.
� Practice
Use this to test the * Breath Navi function.
(*Breath Navi is the small device attached to the edge of the gantry aperturethat gives the visual breathing instructions to a patient with hearing problem.)
System
1-7CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Gantry Display Panel
PositioningLight On/Off Interference
Tilt/AngleRange
Distance from Landmark(mm)
�
X–ray On
ReleaseCradle LockDistance from Iso–Center
� � � �
Function
The gantry display panel shows the following readings and indications.
� X–ray On
The X–ray On lamp illuminates during X–ray exposure.
� Tilt / Angle Range
The Tilt/Angle Range field shows the reading of either the current tilt or thetilt range limits at the current table height/cradle extension. The numericalreading is preceded by either S (superior) or I (inferior).
� Distance from Landmark
The Distance from Landmark field shows the reading of the distance (mm)between the landmark and the anatomy in the X–ray path. The numericalreading is preceded by either S (superior) or I (inferior).
� Distance from Iso–Center
The Distance from Iso–Center field shows the reading of the distance (mm)between the iso–center and the top of the table.
System
1-8CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
� Release Cradle Lock
The Release Cradle Lock lamp turns on while the cradle is unlatched.
� Interference
The Interference lamp turns on when some mechanical interference occurs.
� Positioning Light On/Off
The Positioning Light turns on or off when the positioning light button on thecontrol panel is toggled.
System
1-9CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Table
Cradle
Latch Button
Speaker
Foot Switch(Table Up)
Home Position
Function
The table is used to load a patient for scanning.
� Cradle
The cradle moves into or out of the gantry aperture.
� Latch Button
The latch button is used to latch or unlatch the cradle. The unlatched cradlecan be manually slid. The Display Panel shows whether the cradle islatched.
� Foot Switch
The foot switch is used to automatically raise the table to the height of140cm., then move the cradle inward.
� Home Position
The home position pedal is used to automatically position the table to itslowest level and the gantry to zero tilt.
� Speaker
The speaker is used to deliver oral instructions to a patient.
System
1-10CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Table Specifications
� Vertical range : from 40 cm to 95 cm (cradle height from floor)
� Vertical elevation speed : 18.3 mm/second on average
� Maximum cradle travel : 1620 mm (Note: Table Height, Gantry Tilt, and scanning software determine thescannable range. )
� Cradle speed : 20 mm/second (Slow mode), 100 mm/second (Fast mode)
� Table load capacity : maximum 205 Kg (450 lbs.)
System
1-11CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Operator Console (OC)
Keyboard
Scan/Display Monitor
Mouse
MOD Drive
Main Switch
CD–ROM Drive
Function
The operator console (OC) is mainly used for the operator to set up the scanprocedures and process the resultant image data.
� Scan/Display Monitor
The 21–inch monitor on the OC can be mainly used for two purposes,scanning patients and displaying images.
� Keyboard/Mouse
Please refer to the User Interface page.
� MOD Drive
Image data can be stored in 2.3 GB MOD (magnetic optical disk).
Caution : It is highly recommended to always take back–up image databecause there may be a possibility of medium breakdown.
System
1-12CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
User Interface
The HiSpeed system employs the user interface that enables the operator toeasily interact with the OC workstation. The user interface consists of amouse and keyboard.
Mouse
Left button
Middle button
Right button
General operations of the mouse are as follows:
� “Click”
Press a button once then release it quickly.
� “Double / triple click”
Click two/three times quickly.
� “Click and drag”
Press and hold the button and move the mouse across the pad.
Main functions of each button are as follows:
� Left button
To select any of icon, item, function, etc.
� Middle button
To change window width and/or level by moving the mouse left/right and/orup/down respectively with the button pressed.
� Right button
To scroll the image or modify the image zoom factor.
System
1-13CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Keyboard
The workstation includes a keyboard for entering text at on–screen promptsand numeric data in data entry fields.
The upper part of the keyboard includes scan–related buttons andsafety–related buttons.
Move toScan
StopMove
PrescribedTilt
StartScan Pause
X–ray on
StopScan
EmergencyStop
Talk (Intercom)Mic.
Volume
Reset
Functions of each button and key are described on the following pages.
System
1-14CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scan–related buttons
� Prescribed Tilt
Prescribed Tilt enables you to tilt the gantry to the prescribed angle withoutoperating at the gantry panel.
This button starts to flash following the [Move to Scan] button when thegantry tilt is prescribed at the View/Edit Screen and the [Confirm] icon isselected.
When this button is flashing, you can hold it down to remotely tilt the gantrytill it automatically stops at the prescribed angle (maximum � 30 degrees).
Note : The gantry tilts only while the button is being held down.
WARNING!
During this remote tilt operation there may be a danger of the gantry coverhitting the patient unless the operator is very closely watching the tiltmovement. The touch sensor on the gantry cover is supposed to stop thetilt movement when it detects the physical contact with the patient body.However, that sensor is just an auxiliary safety device. Prior to the actualtilt the operator must ensure without relying on the sensor that the gantrycover will not come in contact with the patient body.Breath Navi (look guide) does not have a function of touch sensor.The touch sensor on the gantry cover works only during the remote tiltoperation at the OC. It does not work when the tilt operation is done at theControl Panel.The front touch sensor (option) must be used in case it is hard to see thefront gantry cover from the console.See the next page for the locations the front and rear touch sensors arebuilt in.
System
1-15CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
The locations and sizes of Front/Rear Touch Sensor
Front
75 deg.
SC* : Scan Center
370mm545mm
Front Touch Sensor(optional purchase)
SC*
Rear
SC* : Scan Center
664mm
Rear Touch Sensor
579mm
SC*
Note : After the Touch Sensor worked and stopped the gantry, tilt the gantryto the opposite direction to release it.
System
1-16CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
� Move to Scan
Press this button when lit to move the cradle to the prescribed scan startlocation.
� Stop Move
Press this button to stop the cradle when it is traveling.
� Start Scan
Press this button when lit to start scans.
� Pause
Press this button to temporarily halt scans. Scans pause after completing thecurrent one scan. Press Start Scan button to resume scans. ( Note : It is best to use the Stop Scan button to stop the scan immediately. The use of the Pause button may not stop the scan immediately due to thesignal transmission timelag. )
� X–ray on
This button lights up during X–ray emission.
� Stop Scan
Press this button to stop the current or halted scan.
� Emergency Stop
This button cuts off the power to the equipment inside the scan room.All mechanical movements and X–ray radiation are halted.
Note : This button DOES NOT affect the power of OC and laser camera.
� Reset
This button restores the power to the equipment inside the scan room afterEmergency Stop is applied. ( Use a pen head or something like that to pressthis button. A finger must be too big.)
The whole system can be reset by turning on the main switch.
System
1-17CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Communication button/dials
� Talk (Intercom)
Hold down this button to give some verbal instructions through themicrophone to the patient during a study.
� Volume
Use these dials to adjust the voice volume of patient or operator.
System
1-18CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Keyboard keys
� Function key
F1, F2, F3 and F4 function keys are used for filming–related operations.Please refer to Chapter 5 [FILMING] for detailed information.
FilmSeries
FilmPage
FilmMID
FilmImage
�� �� �� ��
F5 function key is used to return to the window width and level that havebeen used most recently.
Previous
W/L
��
F6–F11 function keys are used to preset the specific window width and levelfor specific anatomical areas. The procedure is as follows.
Abdo– Head Lungmen
Media– Spine Drop onstinum
Verte–bra Auto Film
�� � � �� ��� ���
1. �Select the image as a primary one (green border) whose WW and WLyou wish to register as default.
2. �Press [Shift+F6–F11] to register those WW and WL as presetconditions.
System
1-19CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
� Image Analysis related keys
These keys are used for image analysis purposes. Please refer to Chapter 3[DISPLAY TOP LEVEL SCREEN] for each key’s function.
DispEraseNorm
EllipseROI
Trace GridOn/Off
MeasDist
UserAnnot
ZoomRoam
� Key top
These keys are mainly used to input texts/numbers and display images.
Enter
�����
Next ExamNext
Prior ExamPrior
– [Enter] :�Use this to conclude the entry.
– [���] : Use these to move forward or backward through items on desktop.
– [�������] : Use these to adjust window level and width of images.
– [Prior Exam/Prior] : Use this to return to prior exam or image.
– [Next Exam/Next] : Use this to go to next exam or image.
System
1-20CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Accessories
The system utilizes the various accessories to facilitate the scan.
Standard Accessories
� Cradle pad
� Cradle extender
� Arm board
� Standard head holder assembly
� Security straps
� Phantoms to calibrate and check performance
� Phantom holder
� Console chair
� Operator manual
Optional Accessories
� Coronal head holder assembly
� Shallow head holder assembly
WARNING!
Do not use any accessories other than those provided by GE. Use of otheraccessories may cause unexpected troubles or physical injuries.
Also, if anything other than GE accessories is placed in the path of x–raybeams, it may affect image quality.
System
1-21CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Head Accessories
Head Strap
Chin Strap
Coronal Head Holder Assembly (option)Coronal Head Cushion
Standard Head Holder Assembly
Coronal Head Neck Pad
(Front side) (Back side)
MADE FOR GENERAL ELECTRIC CO.MILWAUKEE, WISCONSIN BYGE YOKOGAWA MEDICAL SYSTEMS, LTD.7–127 ASAHIGAOKA 4–CHOME, HINOSHI TOKYO, JAPANMODEL 2201806SERIALMANUFACTURED OCTOBER 1997
MADE FOR GENERAL ELECTRIC CO.MILWAUKEE, WISCONSIN BYGE YOKOGAWA MEDICAL SYSTEMS, LTD.7–127 ASAHIGAOKA 4–CHOME, HINOSHI TOKYO, JAPANMODEL 2201805SERIALMANUFACTURED OCTOBER 1997
Label
Label
(Back side)
LabelMADE FOR GENERAL ELECTRIC CO.MILWAUKEE, WISCONSIN BYGE YOKOGAWA MEDICAL SYSTEMS, LTD.7–127 ASAHIGAOKA 4–CHOME, HINOSHI TOKYO, JAPANMODEL 2183827SERIALMANUFACTURED OCTOBER 1997
Shallow Head Holder (option)
Wedge (Large, Small)
System
1-22CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Standard Head Holder Assembly
� Standard Head Holder
Use during normal head studies, with a supine patient. Use the head band,chin strap, and coil spacers (wedges) to secure the head in the properposition. The material is metal–free.
� Head Holder Cushion
Place this cushion on the Standard head holder. Use it to cushion the headand hold it in position during the study. Two sizes are available.
� Head Band
Place the wide area of the head band between the head holder and headholder cushion, with the straps hanging on either side of the head holder,before the patient lays on the table. Move the patient’s head into position onthe cushion, cross each end of the band over the forehead, and fasten thepad of Velcro “loops” to the “hooks” on the back of the head holder. Use thehead band to position and stabilize the head during a study.
� Chin Band
Center the hole in the band over the patient’s chin, and fasten the pad ofVelcro “loops” to the “hooks” on the back of the head holder. Use the chinband to position and stabilize the head during a study.
Coronal Head Holder Assembly (option)
� Coronal Head Holder
Use this head holder during direct coronal studies. The patient lays in thesupine position, with the head tipped backward into the coronal head holder.Use the head band and chin strap to secure the head in the direct coronalposition. The material is metal–free.
� Neck Pad
Place this pad vertically in the Coronal head holder to cushion the rearhead.
WARNING!
Be extremely careful to avoid contact between the coronal head holderand the gantry while lowering the table in the aperture. Patient injury to theneck or equipment damage could result.
System
1-23CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Body Accessory Use
� Knee Pad Assembly
Place the knee pad on the cradle. Rest the patient’s legs, with knees slightlybent, on the knee pad –or– Rest the patient’s head on the knee pad, anduse the knee pad strap to secure the hands over the patient’s head, toprevent catching the patient’s fingers between the cradle and table surfaces.
� Knee Pad Strap
Attaches to the three Velcro “loop” patches on the knee pad with Velcro“hooks”; use to secure the legs or arms to the knee pad. (See above)
� Cradle Pad
Use during standard studies; attaches to the cradle with Velcro “hooks.”
� Cradle Extender
Attach the cradle extender to the end of the table nearest the gantry, toincrease the overall cradle length, and position the body area of interestinside the scannable range. Use the extender to support the patient’s heador feet during a body scan. The material is metal–free.
System
1-24CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Security Strap Kit
� Body Strap A
Use the 60 mm wide strap to position and secure the patient’s head to theHead Cushion; or use as needed to secure blankets or clothing. FollowBody Strap B directions to attach and secure the straps. One set providedwith system.
� Body Strap B
Use the 150 mm wide straps to secure the body to the cradle, to reducepatient movement; to secure blankets or clothing, to prevent contact with thecradle surface; or, to cover and secure the patient’s hands. Slide the nylonguide, opposite the Velcro ends of the straps, into the raceway attached toboth side edges of the cradle. Slide the strap along the length of the cradleto position it over the patient, and fasten the Velcro ends together to securethe patient. Three sets provided with the system
� Body Strap C
Use the 370 mm wide strap in the same manner as Body Strap B.One set provided with system.
� Extension Strap
Use the extension strap with the corresponding body straps to secure largeor obese patients. Attach Velcro “hooks” to “loops” on the body strap toincrease the length. Each set of body straps has an extension strap; fiveprovided with the system.
System
1-25CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Body Accessories
Knee Pad
Knee Pad Strap
Cradle Extender
Cradle Pad
(Back side)
Cradle Extender Pad
Extender Strap (15cm.)37cm. Body Strap
MADE FOR GENERAL ELECTRIC CO.MILWAUKEE, WISCONSIN BYGE YOKOGAWA MEDICAL SYSTEMS, LTD.7–127 ASAHIGAOKA 4–CHOME, HINOSHI TOKYO, JAPANMODEL 2201803SERIALMANUFACTURED OCTOBER 1997
Label
System
1-26CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Attach/Detach Head Holder and Cradle Extender
CAUTION
Prevent damage to metal–free accessories !Carefully examine the metal–free clasp assembly on theaccessory and the catch on the cradle before attempting toattach the accessory for the first time.
To attach an accessory:
� Make sure the clasp on the accessory is unlatched, and hanging freely.Hold the clasp in the unlatched position while you attach the accessory.
� Align the accessory tongue with the pocket at the end of the cradle.Keep fingers clear of the cradle. Slide the tongue all the way into thepocket until it rests snugly against the angled face of the cradle.
� Gently pull the clasp toward the floor and away from the gantry until ithooks the catch on the cradle. Push the hinged–end of the clasp towardthe cradle until it snaps into place.
To detach an accessory:
� Pull the hinged–end of the clasp toward the floor and away from thegantry until it snaps open and clears the catch.
� Gently slide the accessory toward the gantry until it clears the cradle.If the accessory offers any resistance, check to make sure the claspcleared the catch on the cradle.
System
1-27CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
System Power On/Off
How to turn on the system power ( �����er to maintain the system stability
it is highly recommended to shut down and reboot the system once a day.)
1. �Turn on the main switch located on the lower right side of the OC.
2. �The system will start up and the Scan/Display monitor on the OC willautomatically proceed to its Top Level screen.
How to turn off the system power ( �����er to maintain the system stability
it is highly recommended to shut down and reboot the system once a day.)
1. �Click on the [Shutdown] icon on the upper left corner of the screen.
Shutdown
2. �The system will automatically proceed to the point where you can turnoff the main switch.
Note: In case you turn off the main power to the system immediately afterremoving a great number of images, when you boot up the systemnext time, there still may be some images unremoved. When youwant to remove some 3,000 images, wait at least 5 minutes beforeturning off the power. It may take around 30 minutes to remove some18,000 images. To maintain the stability of the system, reboot thesystem once a day.
System
1-28CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Blank page
Chapter 2
2-1CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
SCAN TOP LEVEL SCREEN
After the system starts up, the following screen appears on the CRT asdefault Top Level screen. This screen will also appear when the [Scan] iconon the upper left corner is selected while in the either Display or ImageWorks mode.
Scan
Scan Top Level Screen
2-2CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Icon Selection
� When a certain icon is selected, it highlights in blue.
� Icons currently not available are displayed in gray.
� When the second icon is selected after the first one was selected, the firsticon highlights in green and its upper right corner folds.
Functions of each icon are described on the following pages.
Scan Top Level Screen
2-3CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Function of Each Icon 1
Display
���������
Scan
�� ���
��������
This icon leads to the scan screen.
This contains Reformation, 3D, Network and other advanced
image analysis applications. Refer to {Image Works} section.
This is only for service purpose. Users are not allowed totouch this.
Use this to shut down the system.
This icon leads to the Display screen.
The icons on the Display screen are described in the laterchapter.
�����
This icon leads to the InSite service.(This icon is not available unless InSiteoption is installed.)
Scan Top Level Screen
2-4CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Function of Each Icon 2
To see current messages
Remaining disk space
� Remaining Disk Space
This area shows the remaining disk space. ( The above example shows thatthe disk can hold 148 images of 512x512 matrix, which amounts to 595images of 256x256 matrix.)
� Reconstruction Status
This area shows the exam, series and image number of the exam that iscurrently being reconstructed.
� Archive Status
This area shows the status of images being archived.
� Network Status
This area shows the status of networking.
� Filming Status
This area shows the status of filming.
� Current Messages
Click here to open the Current Messages screen.
Scan Top Level Screen
2-5CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
New Patient
New Patient icon is located at the left side of the screen.
New Patient
Function�Select this icon first to begin a new patient exam.
Refer to the chapter 4 [Scan] for detailed information.
1. �Click on the [New Patient] icon to open the following New Patientscreen.
Patient Information
Exam Number
Accession Number
Patient ID
Patient Name
SexBirthdate
Age
Weight Kilograms Pounds
Referring Physician
Radiologist
Operator
History
Exam Description
Protocol Number
EndExam
Protocol Selection
2. �Enter necessary patient information.
Note : Patient ID must be put to proceed. (required)
3. �Choose a desired scan protocol from the Default Protocol area locatedon the right side of the New Patient screen.
4. �The system automatically proceeds to the View/Edit screen.
Scan Top Level Screen
2-6CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Patient Schedule 1
Patient Schedule icon is located below the New Patient icon.
Patient Schedule
Function�Patient Schedule allows you to enter patient data ahead of time.Once it is time to perform a specific patient exam, you can select from thecreated list, enter the patient ID or requisition number to call up the patientdata.
1. � Click on [Patient Schedule] to open the following Schedule list.
Patient ScheduleSchedule
Status Patient ID
C
N
C
C
C
8120
8255
8764
Patient Name
Kim McCquaer
Dave Sonaty
Laura Sepcic
Date Time Exam Description
Next PriorQuit Add DeletePreferencesSelect View
5300 Juliet Herman
2761 Brad Fox
Patient AllPatient More InfoEdit
PatientDelete
Selected
6/15/98
6/12/98
6/09/98
6/02/98
5/27/98
10:23AM
02:19PM
03:26PM
11:23AM
10:46AM
Req. Number
0654
0718
0788
8332
9102
On the left side of the column under Status, there are two letters. C standsfor Completed and N for New. The completed exam at the top of the list isthe last completed exam. The only active functions at this point are [Quit]�[Add Patient], [Delete All], [Preferences] and Next/Prior arrows.
Note : Active icons have clear black characters on them whereas inactiveicons have dim characters.
2. �Once a patient is selected, the other icons become active.
Functions of each icon are described on the following pages.
Scan Top Level Screen
2-7CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Patient Schedule 2
Select Patient
SelectPatient
Take the following steps to use [Select Patient].
1. �Select [New Patient].
2. �Select [Patient Schedule].
3. �Select the desired patient on the Schedule list.
4. �Select [Select Patient]. The patient data will then be placed in the NewPatient data area.
Note : The patient data will also be called up by typing in the patient ID.
View More Info
ViewMore Info
While your HiSpeed system is connected with HIS/RIS and if your HiSpeedsystem has Connect Pro option, select [View More Info] to get moreinformation on the patient.
Scan Top Level Screen
2-8CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Patient Schedule 3
Add Patient
AddPatient
Select [Add Patient] to open the following menu to add patients to theschedule. Patient data, date and time of the exam and the protocol can beentered here.
Patient Schedule
Accession Number
Patient ID
Patient Name
SexBirthdate
Age
Weight Kilograms Pounds
Referring Physician
Radiologist
Operator
History
Exam Description
Protocol Number
Cancel
Protocol Selection
Accept
Anatomical Selector
Date Time
– Accession Number : Within 16 characters
– Patient ID : Within 16 characters
– Patient Name : Within 64 characters
– Sex : M (Male) or F (Female)
– Birthdate : Year, Month, Day
– Age : Years, Months, Weeks, Days
– Weight : Kgs or Pounds
Scan Top Level Screen
2-9CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Patient Schedule 4
– Reference Physician : Within 32 characters
– Radiologist : Within 32 characters
– Operator : Within 3 characters
– History : Within 60 characters
– Exam Description : Within 22 characters
– Protocol Number : Within 5 characters
– Date : Exam date, Year, Month, Day
– Time : Exam time, Hour, Minute
Edit Patient
EditPatient
Select [Edit Patient] to bring up the Patient Information screen to edit patientdata.
Scan Top Level Screen
2-10CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Patient Schedule 5
Delete Selected
DeleteSelected
[Delete Selected] allows you to delete specific patients from the Schedulelist.
1. �To delete only one patient, click on that patient’s information, then select[Delete Selected].
2. �To delete multiple patients in order, click on one patient, then go to thelast patient and press and hold the Shift key, then click on that patient.Then, select [Delete Selected].
To delete multiple patients out of order, press and hold the Control key,then click on the desired patients and select [Delete Selected].
3. �A confirmation message “Are you sure you want to delete selectedpatients ?” appears. If OK, click on [OK].
Delete All
DeleteAll
[Delete All] allows you to delete all of the patients.
1. �The following screen appears after selecting [Delete All].
Delete All
AllCompleted
AllNew
All
OK Cancel
2. �Select [All Completed] to delete all scan–completed patients.Select [All New] to delete all new patients.Select [All] to delete both of the above patients.
Note : The default is [All Completed].
Scan Top Level Screen
2-11CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Patient Schedule 6
Preferences
Preferences
[Preferences] allows you to set the sort order and delete time.
1. �The following screen appears upon the selection of [Preferences].
Preferences
OK Cancel
Update ScheduleAutomatically ?
Yes No
Yes NoShow UpdateParameters ?
Sort By : Date/Time
Name ID
Delete Completed Exams After : Days
Default Update Parameters
Get Patient List For :ThisSystem
All CTSystems
AllSystems
With a Date Range :Today
Days Before Today
Days After Today
All Dates
2. �The patient list can be sorted by Date/Time, Name or ID.
3. �If you want the system to automatically delete completed exams, youcan click in the Days box and type in the number of days desired.The default is 0 (zero).
4. �Select [OK] to conclude the entry.
Scan Top Level Screen
2-12CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Protocol Management
Protocol Management icon is located at the right side of the New Patienticon.
Protocol Management
Function : Protocol Management includes two functions, which are [AutoVoice Record] and [Protocol Management].
1. �The following screen appears upon the selection of [ProtocolManagement].
Auto VoiceRecord
ProtocolManagement
Quit
2. �Select [Auto Voice Record] to set up Auto Voice or select [ProtocolManagement]
Select [Quit] to quit.
Functions of [Auto Voice Record] and [Protocol Management] are describedon the following pages.
Scan Top Level Screen
2-13CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Auto Voice Record 1
Function : This function allows you to give the patient recorded vocalinstructions.
1. �The following screen appears when you click on [Protocol Management]icon.
Auto VoiceRecord
ProtocolManagement
Quit
2. �The following Auto Voice Record screen appears upon the selection of[Auto Voice Record].
Auto Voice Record
���� ����������
����������
����������
1.2.3.4.5.6.7.8.9.
���������
������������
11.12.13.14.15.16.17.18.1920.
����
Create New Message Message Management
����� ���� !
�"�# Save Message
�"�# $�"�������
$���
���������� ������������
10.
Language Selection English
Twenty protocols in all are available. The first three protocols are preset inthe system and can not be changed. Other 17 protocols are programmable.
Scan Top Level Screen
2-14CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Auto Voice Record 2
As for the first three protocols, there are nine language options.In order to choose any one of them, click on the [English] button on the AutoVoice Record menu on the previous page to open the following menu.
Auto Voice Language Selection
English–male English–female Japanese
German Spanish
French
Mexican Spanish
Cancel
Italian
Korean
Click on any one of them for your purpose.
3. �To create a new Auto Voice instruction message, point and click in theName area where you wish to put the protocol. Type in the name andpress the Enter key.
After pressing the Enter key, the Pre and Post Message areas becomeactive.
4. �Go to the Create New Messages area to create new instructionmessages.
Record Stop
Play Save Message
0.0
Create New Messages
The only active icon at this point is [Record].
Scan Top Level Screen
2-15CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Auto Voice Record 3
5. �Once you are ready to record, select [Record]. The time clock to theright will start. The maximum recording time is 20 seconds.
6. �When finished with recording, select [Stop]. Select [Play] to listen towhat was recorded.
7. �To save the recorded message as a protocol, click on the Pre or PostMessage box and select [Save Message].
8. �The Message Management area allows you to play or delete an existingAuto Voice protocol. To delete a protocol, select the protocol name andthen select [Delete Set].
Play
Message Management
Delete Set
Done
To play a message, simply select the message box you wish to playand select [Play].
9. �To leave the Auto Voice Record menu, select [Done].
Scan Top Level Screen
2-16CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Protocol Management 1
Function : This function allows you to create, edit or delete scan protocols.
1. �The following screen appears when you click on [Protocol Management]icon.
Auto VoiceRecord
ProtocolManagement
Quit
2. �The following screen appears upon the selection of [ProtocolManagement] on the above screen. This is the Adult Protocol menu.The Infant Protocol menu is also selected by clicking on the Infant tab.
Quit
Protocol Management
Adult tabInfant tab
3. �Select a protocol or an anatomic region to open the menu on the nextpage.
Scan Top Level Screen
2-17CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Protocol Management 2
4. �You can build up to 15 protocols for each anatomical region.
Protocol Name Default
New Edit Delete
Done Cancel
Duplicate
5. �To duplicate (copy) an existing protocol, first choose the protocol andselect [Duplicate]. Then, you may slightly modify it to create a newprotocol that will have the similar parameters to the original protocol.
6. �To edit or delete an existing protocol, first choose the protocol andselect [Edit] or [Delete] respectively.
7. �To create a new protocol, select [New].
8. �After selecting [New], enter the protocol name on the highlighted lineand press the Enter key. To set a protocol as the default for the region,select the [Default] button next to the protocol. Only one protocol canbecome the default for the region.
9. �Next, select the Series Type that will begin the protocol and click on[OK].
Scout
OK Cancel
Axial
Select New Series Type
The following Protocol Setup screen appears.
Scan Top Level Screen
2-18CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Protocol Management 3
The following screen is an example of axial scan protocol setup.
Anatomical
Prior Next
Series Description
Prior Next Create
Split
Repeat���������
Cancel ���Delete
Protocol Parameters Setup Screen
Reference
Protocol :
Auto Transfer Auto FilmSetup
ShowLocalizer
AutoStore
AddGroup Current
Group
DeleteSelectedGroup
Images ScanType
StartLoc.
EndLoc.
No. ofImages
Thick(mm)
ImageInt’val Pitch Gantry
Tilt SFOV kV mA
Scan Timing FilmRecon
Series Series Series Series
Series:
ReconArea
CTDIw(mGy)
Axial
Axial
S50.0 S10.0 5 581–8 0 120 180
9–15 7S10.0 I60.0 10 10 0 120 160
Body
Body
Auto mA OFF
SmartReconOFF
10. �In the above screen, the [Scan] icon parameters (Thick, Image Int’val,pitch, etc.) are supposed to be filled. In order to enter Timing, Reconand Film parameters, click on the [Timing], [Recon] and [Film] icon,respectively. The menus for Timing, Recon and Film are shown on thefollowing pages.
Prior/Next : Prior or Next key appears on the screen only when morethan three groups are prescribed for one screen.
Scan Top Level Screen
2-19CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Protocol Management 4
The following Timing menu will appear when you click on the [Timing] icon.
PrepGroup ISD
(sec)
BreathHold(sec)
BreatheTime
AutoVoice
No.
Scan Timing Recon Film
(sec)
The following Recon menu will appear when you click on the [Recon] icon.
DFOV(cm)
R/LCenter(mm)
A/PCenter(mm)
ReconType
ImageFilter
MatrixSize
MotionCorrect
SpecialFilter
ShowRecon1
ShowRecon2
ShowRecon3
Y N N
Scan Timing Recon Film
The following Film menu will appear when you click on the [Film] icon.
AutoFilm
FrameFormat
Interval Flip Width1
Level1
MagFactor Rotate
UserAnno. GSE
Recon 1Film Set
1
Recon 1Film Set
2
Recon 2Film Set
1
Recon 2Film Set
2
Recon 3Film Set
1
Recon 3Film Set
2
Scan Timing Recon Film
Scan Top Level Screen
2-20CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Protocol Management 5
Select [Prior Series] to display the prior series protocols.
Select [Next Series] to display the next series protocols.
Select [Repeat Series] to add the same series protocols.
Select [Delete Series] to delete the displayed series protocols.
Refer to the Chapter 4 for the function of other icons or buttons.
11. �After all the scout view parameters have been entered, select [CreateNew Series] button on the front page screen to create the next series.Then, the following screen appears.
Scout
CreateBefore Cancel
Axial
Select New Series Type
CreateAfter
12. �Select the series type from [Scout] or [Axial].
Also, select [Create Before] or [Create After] to determine if the serieswill be put before or after the first series you built.
13. �Again, enter scan parameters for the series selected.
14. �After all the series and the parameters have been created, touch[Accept] button to save the protocol.
15. �You will then return to the beginning so that you may choose toduplicate, delete, edit or create a new protocol. Or you can select[Done] and then [Quit] to exit the protocol section.
Scan Top Level Screen
2-21CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Retro Recon 1
Retro Recon icon is located below the Patient Schedule icon.
Retro Recon
Function : Retrospective reconstruction allows you to change thereconstruction parameters that were previously used for the exam. ThroughRetro Recon you are allowed to change DFOV(Display Field of View),AP(Anterior/Posterior) and RL(Right/Left) Centering, Recon Type, MatrixSize, Segmenting and Peristaltic Motion Correction for axial series.
1. �First click on the [Retro Recon] icon. The following screen appearslisting up to 1000 scan data. Click on the [Update] button to update thelist or select [Quit] to exit.
Retro Recon List Select Screen
Quit
Retro Recon List Select
Exams
Select Update
Patient ID Patient Name Exam # Date Time SuiteSeries
Series # Series Description Landmark
Prior Next Prior Next
Series
2. �Point and click on the exam and series you wish to reconstructretrospectively and then choose [Select Series] at the bottom center.
The system automatically proceeds to the Retro Recon View/Edit screen.
Scan Top Level Screen
2-22CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Retro Recon 2
3. �After selecting [Select Series], the following screen appears displayingall of the groups in the selected series. Y(Yes) in the Retro columnmeans that group will be reconstructed. If you do not wish to recon thegroup, simply point and click on the “Y” icon for the group to toggle it toN(No).
When the [Images] icon is selected, the right side of the screen is as follows.
RetroScanType
Y
Y
ScanStart
Location
ScanEnd
LocationRetroStart
RetroEnd
No. ofImages
ImageInterval
GantryTilt
SFOV
Y
QuitList
Exams Confirm
Images ReconPrior Next
When the [Recon] icon is selected, the right side of the screen is as follows.
RetroScanType
Y
Y
ScanStart
Location
ScanEndLocation
R/LDFOV
Y
QuitList
Exams Confirm
Prior Next
Center(mm)
A/PCenter
ReconType(mm)
ImageFilter
MotionCorrect
ReconMode
MatrixSize
SpecialFilter
Images Recon
(cm.)
GraphicRetro
N
N
N
Scan Top Level Screen
2-23CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Retro Recon 3
4. �Enter the parameter changes and select [Confirm].
– Retro Start : Start location of retro recon
– Retro End : End location of retro recon
– Image Int’val : Image interval (in case of Helical)
– DFOV : Display Field of View ( 4.8cm to SFOV)
– DFOV A/P R/L Center : Anterior/Posterior and Right/Left center of DFOV
– Recon Type : SOFT, STND, STD+, DETL, CHEST, BONE, EDGE or PFRM
– Image Filter : Smooth (S1, S11, S2, S21, S3), Edge (E1, E2, E21, E22, E23, E3), Lung (L1, L2, L3)
– Motion Correct : Motion correction
– Recon Mode : (See next page.)
– Matrix Size : 256x256 or 512x512
– Special Filter : ANR (Advanced Noise Reduction) (1 or 2), AAR (Advanced Artifact Reduction) (1 or 2)
ANR filter allows you to reduce noise without compromising spatialresolution. It contributes to a decrease in standard deviation byapproximately 10%, which may be equivalent to one–step decrease inmA. Note : ANR is compatible only with SOFT, STND, and STD+.
AAR filter allows you to reduce artifacts normally shown around arms.
When you click on the [Special Filter] button or the each Special Filterfield, the following menu appears.
None ANR 1 ANR 2 CancelAAR 1 AAR 2
Select the desired Special Image Filter.
[ANR 2] has a stronger effect than [ANR 1]. Also, [AAR 2] has astronger effect than [AAR 1]. Select either of them or select [None] notto use these filters.
Scan Top Level Screen
2-24CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Retro Recon 4
– Graphic Retro : If you wish to display on the image the circle cursor tographically prescribe the retro recon area instead of typing in DFOV,R/L Center and A/P Center, toggle N (No) on each field to Y (Yes).
If the displayed circle cursor is too big to handle, type in the smallerenough DFOV to display the entire circle cursor, then change its sizeand location using one of four squares outside the cursor and thecenter dot of the cursor respectively to designate the recon area.
5. �Select the [Recon Mode] button to open the following screen.(*Smart Helical Option is needed to have [Plus] button.)Recon Mode section allows you to edit the scan type and the amount ofdata for reconstruction.
Cancel
Axials:
Recon Mode
Full
OK
Helicals:
Cines:
Segment Full Plus
Segment Full
2 3 4 5 6
The scan types available to edit will be active.
[Full] uses full–rotation data, whereas [Segment] uses partial–rotation data.
*Smart Helical Option offers [Plus] function that uses 1.6–rotation data.
Make the appropriate selection.
6. �Click on the [Confirm] button to start retro recon.
Newly retro–reconstructed images will be saved as a new series with“Retro” annotation.
7. �If you need to retro–recon another exam/series, select [List Exams]button. Select [Quit] to exit.
Scan Top Level Screen
2-25CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Recon Management 1
Recon Management icon is located at the right side of the Retro Recon icon.
Recon Mgmt
Function : Within the Recon Management function, you can have access toSuspended Entries and Recon Queue Management.
1. �Select [Recon Mgmt] to open the following screen.
Recon Management
Suspended EntriesPatient ID Exam # Series # Image # Type
DeleteSuspended
Entries
UnsuspendedEntries Prior Next
UpdateList
Recon Queue Management:
PauseQueue
ViewQueueEntries
RestartQueue
DeleteRetroEntries
Quit
Functions of each button are described on the following pages.
Scan Top Level Screen
2-26CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Recon Management 2
Suspended Entries
These are basically images that were not reconstructed for some reason andwere placed in a Queue.
� [Delete Suspended Entries]
Select this button to delete the suspended entries. Once the images aredeleted, they can not be reconstructed prospectively, but they can bereconstructed retrospectively.
� [Unsuspended Entries]
Select this button to try to reconstruct the images again after selecting imageentries.
� [Update List]
Select this button to update the Unsuspended Entries List if any entries needto be added while Recon Management was open.
Recon Queue Management
� [Pause Queue]
Select this button to stop the reconstruction process. This might be useful ifyou have just started a Retro Recon, but maybe incorrect parameters wereentered. Instead of having the system reconstruct all of the incorrect images,you can select [Pause Queue] to stop the reconstruction and then delete theimages that have not been yet reconstructed.
� [Restart Queue]
Select this button to restart the reconstruction queue after you stopped it forsome reason.
� [View Queue Entry]
Select this button to display a menu that shows what is in the reconstructionqueue. You can select [Cancel] to leave the View Queue function or select[Update List] to update the queue entries list.
� [Delete Retro Entries]
Select this button to delete entries that remain in the reconstruction queue.This is used in conjunction with [Pause Queue].
Scan Top Level Screen
2-27CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Tube Warm–up
Tube Warmup icon is located below the Retro Recon icon.
Tube Warmup
Function : Tube Warmup allows you to have access to the Tube Warmupand Daily Calibration procedures.
Note : Tube Warmup is required when more than two hours have passedsince the last scan or whenever the system recommends it.
1. �Click on the [ Tube Warmup] icon. The following screen will appear. [ Tube Warmup] button on the screen allows you to conduct only thetube warm–up procedure. [Daily Calibration] allows you to perform boththe tube warm–up and daily calibration procedures. [Daily Calibration] isselected by default.
��
Cancel OK
Make sure that no one is in the scan room and the gantry aperture isclear during warmup.
Tube Warmup
Daily Calibration
Tube Warmup
Warming up the tube as recommended will provide the high qualityimages and extend the life of the tube.Menu1
2. �Select the [ Tube Warmup] button, then [OK] to proceed to the tubewarmup procedure or select the [Cancel] button to exit.
Scan Top Level Screen
2-28CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Tube Warmup 2
3. �Upon selection of [OK], the following scan menu will appear. Select [Confirm] to initiate the tube warm–up.
Service Scan
Service Tool: Warm up Scan Exam ID: 218120
Scan Type Axial
Scan Time 2.0 secondsVoltage 80 kVCurrent 80 mA
Thickness 1.0 mmFocus Large
Remaining scan : 18
Press <Confirm> and Start scan> to start scan
ConfirmCancel
Menu 2
4. �Several seconds after selecting [Confirm], the [Start Scan] button on thekeyboard flashes. Press it to start X–ray radiation.
The system automatically performs four tube warm–up scans.
Note : When you select the [ Tube Warmup] button more than 30 hourslater since the last daily calibration, the following message will be postedto urge you to perform daily calibration.
Service ScanWarning
Please execute “Daily Calibration” to provide thehigh quality images.
Do you continue the tube warmup ?
Yes No
Selecting [Yes] leads you to Menu 2 and [No] to Menu 1.
Scan Top Level Screen
2-29CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Tube Warmup 3
5. �When you select the [Daily Calibration] at Menu 1, Menu 2 will appear.
Service Scan
Service Tool: Warm up Scan Exam ID: 218120
Scan Type Axial
Scan Time 2.0 secondsVoltage 80 kVCurrent 80 mA
Thickness 1.0 mmFocus Large
Remaining scan : 18
Press <Confirm> and Start scan> to start scan
ConfirmCancel
Menu 2
Upon selection of [Confirm], the system first performs the same scanprotocol as Tube Warmup and daily calibration later.
Scan Top Level Screen
2-30CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scanner Utilities 1Raw Data Function
Scanner Utilities icon is located at the right side of the Tube Warmup icon.
Scanner Utilities
Function : From Scanner Utilities you can have access to either RawdataFunction or User Calibration.
1. �Click on the [Scanner Utilities] icon. The following screen will appear.
Scanner Utilities
RawdataFunctions
Quit
UserCalibration
2. �Select [Rawdata Functions] to open the following default screen.This screen will also appear when you select [Save to MOD] from thepulldown menu of Function Selection. (See next page.)
File Function Selection Rawdata Selection
Save Rawdata to MOD
Select RawDataExam:
Exam # NameExam No:Series #
Series No:Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Start Save
Type Acqus Reserve Patient PositionDate Reserve
Scan Top Level Screen
2-31CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Scanner Utilities 2Raw Data Function
When you select the [File] button, the [Exit] button appears.Select [Exit] to exit from Rawdata Functions.
When you select the [Function Selection] button, the pulldownmenu appears which contains the [Save to MOD],[Restore/Delete from MOD], [Reserve/Release] and [InitializeMOD] buttons.
When you select the [Rawdata Selection] button, the pulldownmenu appears which contains the [Whole Exam], [WholeSeries] and [Individual Raw] buttons.
%�"�
����
%����������"������
�� ������&$
������'$�"����(����&$
���� �'��"����
������"�)���&$
����������"������
���"������
���"�������
���� ����"����
Scan Top Level Screen
2-32CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scanner Utilities 3Raw Data Function
Save Rawdata to MOD
When you select [Save to MOD] from the [Function Selection] pulldownmenu, the following screen will appear.
File Function Selection Rawdata Selection
Save Rawdata to MOD
Select RawDataExam:
Exam # NameExam No:Series #
Series No:Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Start Save
Type Acqus Reserve Patient PositionDate Reserve
1. �Select any of raw data that you wish to save to MOD (Magnetic OpticalDisk). You can also use [Whole Exam], [Whole Series] or [IndividualRaw] button from the [Rawdata Selection] pulldown menu.
2. �After selecting the raw data, click on the [Start Save] button to startsaving the raw data into the MOD.
Scan Top Level Screen
2-33CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Scanner Utilities 4Raw Data Function
Restore/Delete from MOD
When you select [Restore/Delete from MOD] from the [Function Selection]pulldown menu, the following screen appears.
File Function Selection Rawdata Selection
Restore/Delete from MOD
Select RawDataExam: Exam No:
Series No:Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Restore Delete
Exam # Name Series # Type Acqus Reserve Patient PositionDate Reserve
1. �Select any of raw data that you wish to restore from MOD to the systemdisk, or that you wish to delete from MOD. You can also use [WholeExam], [Whole Series] or [Individual Raw] button from the [RawdataSelection] pulldown menu.
2. �After selecting the raw data, click on the [ Restore] or [Delete].
Scan Top Level Screen
2-34CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scanner Utilities 5Raw Data Function
Reserve/Release Rawdata
When you select [Reserve/Release] from the [Function Selection] pulldownmenu, the following screen will appear.
File Function Selection Rawdata Selection
Reserve/Release Rawdata
Select RawDataExam: Exam No:
Series No:Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Reserve Release
Exam # Name Series # Type Acqus Reserve Patient PositionDate Reserve
1. �Select any of raw data that you wish to reserve or release.( “Reserve” rawdata means to protect rawdata from being deleted.“Release” rawdata means to release the protection.)
2. �After selecting the rawdata, select [Reserve] or [Release].
Scan Top Level Screen
2-35CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Scanner Utilities 6Raw Data Function
Initialize MOD
When you select [Initialize MOD] from the [Function Selection] pulldownmenu, the following screen will appear.
Initialize MOD
Initialize MOD will take approximately 3–4 minutes.
Press Yes to continue or No to quit.
Yes No
1. �Select [Yes] to start initializing the MOD.
2. �Select [No] to exit.
Scan Top Level Screen
2-36CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scanner Utilities 7User Calibration
1. �Select [User Calibration] to open the following screen.
Calibration
Sequence
AirCalibration
Air & PhantomCalibration
kV
80 120 140
Cancel OK
2. �Select either [Air Calibration] or [Air & Phantom Calibration].Also, select kV from 80, 120 or 140.
3. �Select [OK] to perform calibration scans, or [Cancel] to cancel.
Chapter 3
3-1CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
DISPLAY TOP LEVEL SCREEN
The following screen will appear when you click on the [Display] icon on theupper left corner of the screen while in the either Scan or Image Worksmode.
Image Area
Display
Display Top Level Screen
3-2CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Primary/Secondary Viewport
Focus Selection
“Focus” means which viewport is active for image display. There are twokinds of focus, which are primary and secondary focuses. Primary focusviewport has a blue border, whereas the secondary focus viewport has ayellow border.
Every image analysis or manipulation function is available in primary focus.Image analysis or manipulation functions available in the secondary focusare as follows: Window W/L, Magnification, Pan/Zoom, Flip/Rotate, DisplayNormal, and Erase Graphics.
You can select the primary or secondary focus by clicking on the desiredviewport as follows:
� Double click
To select primary focus viewport and deselect other viewports
� Single click
To select a new primary focus viewport and change the previous primary tosecondary
� Triple click
To select a new primary focus viewport and change all the other viewports tosecondary
Display Top Level Screen
3-3CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Image Display Layout
Layouts
On the Display desktop, there is a choice between two layouts for viewingimages. They are the Autoview Layouts and the Review Layouts. Duringscanning the Autoview Layouts will primarily be used.
Autoview Layouts
Click on the left side of this button to select Au-toview Layouts and on the right side for ReviewLayouts.
�����������
Clicking here with Autoview Layouts leads tothe screen format choice from three formats.
Clicking here with Review Layouts leads to thescreen format choice of five format for Full Screen orViewport.
Each layout is described on the following pages.
Display Top Level Screen
3-4CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Autoview Layouts 1
1. �Select Autoview Layouts at the Exam Rx Display menu to open theAutoview Layouts menu.
Autoview Layouts
Autoview Layouts menu
Scout Auto AutoFilmView
AutoLink
AutoView
AutoLink
AutoLink
AutoView
Back
Display Top Level Screen
3-5CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Autoview Layouts 2
There are four different types of viewport for six other Autoview Layouts.
� Autoview
This viewport always displays in real time the most recently reconstructedimage. Next/Prior buttons do not work here.
� Auto Reference
This viewport displays a Scout image with reference lines that show thelocations of Autoview images. Front Scout and side Scout can be toggled byNext/Prior buttons.
� Auto Film
This viewport displays an image for auto filming and occupies the upper leftquadrant. This can be done by selecting the folded corner at the top right ofthe viewport. To return the viewport to the cross reference display, select thefolded corner again.
� AL (Auto Link)
While reconstruction is active, the lower right viewport will automaticallydisplay the first image of the series being currently reconstructed. Theviewport will be annotated with AL to indicate that Auto Link is on. All displayfeatures and Next/Prior can be used with Auto Link.
� Free
This viewport can be used to display any image from other exam or series.
Display Top Level Screen
3-6CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Review Layouts
Multiple Image Display (MID)
Review Layouts can be used during scanning, but you will be unable to viewthe current autoviewed image or autofilming.
1. �Display an image in primary focus viewport.
2. �Select Review Layouts at the Exam Rx Display menu to open theReview Layouts menu.
Review Layouts
[ViewportFormat]
Review Layouts menu
[Full ScreenFormat]
3. �Select [Full Screen Format] to display the selected layout on the entirescreen. Select [Viewport Format] to display the selected layout inprimary focus viewport.
4. �Select one of the four formats.
Display Top Level Screen
3-7CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
List / Select 1
Selecting this function will give you a listing of the exams and otherinformation about the exams on the system disk.
1. �Select List / Select at the Exam Rx Display menu.
List / Select
2. �The following List/Select browser appears.
Note : The List/Select browser does not appear when the system diskcontains no data.
�
A: Examination ListB: Series ListC: Image List
Examinations : Exam #47, May 05 92, SMITH, JON
Exam
J.Herman
DateName Ser Imgs Description Mod
2 examinationsSeriesone series
60 images
Number of images stored: 1101, Available disk space: 948 256x256 image(s), or 236 512x512 image(s).
MidscnMatrix(mm)
Thck/SP(cm)SFOVImg Img Ctr Gntry
10.0/2
1
10.0/
512
(deg)
4
3
6
5
Archive
No
No
No
NoNo
No
TypeDescription Mod
Accept
Cancel
1 PROSP35123145 Jan 08 98
B.Fox18
�
�
Fmt A
S–IImg CtrR–L
Img CtrA–P
Res
50
50
50
50
50
50
CT CT
10.0/
10.0/
10.0/
10.0/
S 50.0S 45.0
+0.0
���
+0.0 512
512
512
512
512
+0.0
+0.0
+0.0
+0.0
Lung
Lung
Lung
Lung
Lung
Lung
S 40.0
S 35.0
S 30.0
S 25.5
R5.0R5.0
R5.0
R5.0
R5.0
R5.0
(mm)
A0.0A0.0
A0.0
A0.0
A0.0A0.0
Dec 23 97 CT
Display Top Level Screen
3-8CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
List / Select 2
3. �Select exam, series and image to display.
4. �Click on the [Accept].
Once all images are loaded, the first image in the series will bedisplayed in a primary focus viewport (blue border).
5. �Use [Next] or [Prior] key on the keyboard to go through images forwardor backward.
Click on the [Cancel] to exit without selecting anything.
Display Top Level Screen
3-9CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Routine Display 1
1. �Select [Routine Display] at the Exam Rx menu.
Routine Display
2. �The following Routine Display menu will appear.
Routine Display
ProView
GSE
ExplicitRoam
ZoomFlip/
Display List/
Ellipse Measure Grid
Cross User Exam Pg
Series Pg
HideErase Screen
Back
RotateMagnify
Normal Select
On/OffDistanceROI
AnnotationReference
SaveGraphics
3. �Click on [Back] to return to Exam Rx menu.
Function of each icon is described on the following pages.
Display Top Level Screen
3-10CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Routine Display 2
�Roam / Zoom
1. �Place the desired image in primary focus viewport.
2. �Select [Routine Display] at the Exam Rx menu to open the RoutineDisplay menu.
3. �Select [Roam] or [Zoom] on the Routine Display menu.
4. �After selecting [Roam], place the mouse cursor over the image you wishto roam and hold down the right mouse button, then move the mouse toroam the image. When you release the mouse button, the image will bedeposited on that point. Select [Display Normal] to return the image toits original position.
After selecting [Zoom], place the mouse cursor over the image you wishto zoom and hold down the right mouse button, then move the mouseright (or up) to magnify or left (or down) to minify the image. When yourelease the mouse button, the image will retain the zoom and themagnification factor will be displayed at the bottom of the annotation onthe upper right of the image. The valid magnification ranges are from0.5 to 8 and from1.0 to 4.0 in a 1024 matrix viewport.
Select [Display Normal] to return the image to its original size.
Display Top Level Screen
3-11CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Routine Display 3
Explicit Magnify
This function allows you to numerically set an image magnification factor.
1. �Select [Routine Display] and click on [Explicit Magnify] on the RoutineDisplay menu.
2. �The following Enter Magnification on Factor menu appears.
Enter Magnification Factor
0.5 1.0 1.5 2.0 Other:
Valid magnification factors are between 0.5 and 8.0.
Cancel
3. �Select one of the preset magnification factors or type in your own.The valid ranges are from 0.5 to 8.0 and from1.0 to 4.0 in a 1024 matrixviewport.
4. �This magnification will be applied to all viewports in focus.
Select [Display Normal] to return the image to its original size.
Display Top Level Screen
3-12CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Routine Display 4
Flip/Rotate
1. �Place the desired image in primary focus and select [Routine Display].
2. �Select [Flip/Rotate] on the Routine Display menu.
3. �The following Flip/Rotate menu appears.
FLIP/ROTATEFLIP:
FLR FTB FTB/FLR
ROTATE:
RR RL
Cancel
4. �Select one from FLR (Flip Left to Right), FTB (Flip Top to Bottom),FTB/FLR, RR (Rotate to Right) or RL (Rotate to Left).
5. �This selection will be applied to all viewports in focus.
Select [Display Normal] to return the image to its original orientation.
Display Top Level Screen
3-13CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Routine Display 5
ProView
This function allows you to select an edge enhancement or smoothing filterto be applied to images.
1. �Images must be in primary focus.
2. �Select [ProView] on the Routine Display menu.
3. �The following Select Image Filter menu appears.
Select Image Filter
Edge E1 E2 E21 Lung
S1 S11 S2 OffSmooth
Cancel
E22 E23 E3
S21 S3
4. �Click on any button of the selections.
The Edge enhancement filters are useful for filming bone windows. The Lungenhancement is designed specifically to film lung windows.
E1 : Weak E2 : Moderate E3 : Intense Lung : Lung window
Note : When a filter is applied, images are annotated as E1, E2, E3 or Lungabove the vertical tickmark scale on the right side of the image.
The Smooth filters are used to film soft tissue windows or when you need todecrease the appearance of noise on images.
S1 : Minimum S2 : Moderate S3 : Maximum Off : Without filter
Note : When a filter is applied, images are annotated as S1, S2 or S3 abovethe vertical tickmark scale on the right side of the image.
Note : These filters can also be selected on the Auto Film Tab card.
Note : These filters can not be applied to Scout images.
Display Top Level Screen
3-14CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Routine Display 6
GSE (Gray Scale Enhancement)
This function allows you to alter the slope and gamma curve. This is useful inhead studies to improve the gray–white matter differentiation.
1. �Images must be in primary focus.
2. �Select [GSE] on the Routine Display menu.
3. �The following Gray Scale menu appears.
Gray Scale
G1 G2 G3 Off
Cancel
4. �Click on any of the selections.
G1 : Weak G2 : Moderate G3 : Intense Off : No filter
Note : When a filter is applied, the image is annotated as G1, G2 or G3 justabove the vertical tickmark scale on the right side of the image.
Note : These filters can also be selected on the Auto Film Tab card.
Note : These filters can not be applied to Scout images.
Display Normal
1. �Images must be in primary focus.
2. �Select [Display Normal] to return an image to its normal state ifRoam/Zoom, Explicit Magnify, Flip/Rotate and ProView are used. It willalso erase graphics for Elliptical ROI, Measure Distance, Grid, CrossReference and User Annotations.
Display Top Level Screen
3-15CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Routine Display 7
List / Select
1. �When you select [List/Select], the browser appears.
2. �You can then select an image to view in a viewport or viewports,depending on what is in focus. Simply select an Exam, Series andImage, then select [Accept].
Ellipse ROI
1. �Images must be in primary focus.
2. �Select [Ellipse ROI] to perform a Region Of Interest measurements.
3. �The ellipse ROI cursor will be displayed in blue (active) on the image.To move the ROI cursor, place the mouse cursor in the center of theellipse, then click and drag it with the left mouse button. The other wayto move the ROI cursor is to place the mouse cursor on the number nextto the ellipse or on the line, then click and drag it with the left mousebutton. To change the size, click and drag the small box outside theROI cursor. To rotate the ROI cursor, click and rotate the solid cursor onthe line.
4. �Place the mouse cursor out of the ellipse, then, click the left mousebutton once to calculate the ROI of the ellipse. The ellipse ROI cursorturns white (inactive) and is deposited there.
Note : Once the ellipse ROI cursor turns white, it can not be modified, butyou can turn it blue again by placing the mouse cursor inside the ROI cursorand click the left mouse button once. Then, you can change the size andlocation of the ROI cursor again.
The figures of the Mean, Standard Deviation and Area are displayed at thebottom right of the image. The ROI measurements can be done many times,but the system will display only the last three results on the image.
Select [Erase] to erase the ROI.
Display Top Level Screen
3-16CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Routine Display 8
Measure Distance
1. �Images must be in primary focus.
2. �Select [Measure Distance].
3. �A line with two small boxes at each end will appear in blue (active).
4. �To move the line, place the mouse cursor on the line and hold down theleft mouse button, then drag the line to a new location. To change thelength of the line, place the mouse cursor over the small box and holddown the left mouse button, then drag the box until you get a desiredlength.
5. �Place the mouse cursor out of the line, then, click the left mouse buttononce to calculate the length of the line. The line turns white (inactive)and is deposited there.
Note : Once the line turns white, it can not be modified, but you can turn itblue again by placing the mouse cursor on the line and click the left mousebutton once. Then, you can change the length and location of the line again.
At the bottom right of the image area, a real time readout of the distance willbe shown. The Distance measurement can be done many times, but thesystem will display only the last three results on the image.
Select [Erase] to erase the line.
Grid
1. �Images must be in primary focus.
2. �Select [Grid On/Off].
3. �The grid appears on the image.
4. �To move the grid, first click left on the center of the grid. It becomesactive (blue), then drag it to the desired position. To deposit the grid,click outside the grid.
5. �To take the grid off, click on [Grid On/Off] again.
Display Top Level Screen
3-17CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Routine Display 9
Cross Reference
This function allows you to display the cross reference lines on a Scoutimage.
1. �Display a Scout image in primary focus.
2. �Select [Cross Reference]. The following Post Reference Lines menuappears.
Post Reference Lines
Series Number
Image Numbers All First/Last
Increment
OK Cancel
3. �[Series Number] : Enter the desired series number. [Image Number] : Enter the desired image number.[All] : Select this for all images. [First/Last] : Select this to show only the first and last image line. [Increment] : Select this to show image lines with incremental interval.
4. �Select [OK] to display the image lines on the Scout or select [Cancel] toexit.
Display Top Level Screen
3-18CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Routine Display 10
User Annotation
1. �Images must be in primary focus.
2. �Select [User Annotation] to show the following active (blue) arrow cursoron the image.
3. �You can move both the box and arrow at once by placing the mousecursor anywhere on the blue line, then click and hold down the leftmouse button, then move it. You can move just the arrow or box by thesame way except you would place the mouse cursor on just the areayou wish to move instead of the blue line.
4. �Type in the necessary information, then click the left mouse buttonoutside the box to conclude the type entry. Make sure when typing theannotation that the mouse cursor is inside the image area.
Display Top Level Screen
3-19CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Routine Display 11
Exam Pg / Series Pg
1. �Select [Exam Pg] or [Series Pg] to open the exam page or series pageof the exam in primary focus.
2. �The following is an example.
Text Page Box
Pat. name: T. YOKOGAWAPat. ID: Y8120Age: 34 Sex: M Weight: 65 Kg
Referred by:Radiologist:Operator:Exam Description:History:Contrast:
Ser Images Ref Scan Range Type Series Description
2 18 OM S 50 I115 PROSP Recon2
3 26 SN I65 I350 PROSP
Quit Manual Film Auto Film
Exam no: 81293July 17 97
Use �� buttons on the lower right to go through pages.
3. �Select [Manual Film] to put the page into Manual Film Composer orselect [Auto Film] to put the page into Auto Film Composer.
4. �Select [Quit] to exit from the page.
Display Top Level Screen
3-20CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Routine Display 12
Hide/Show Graphics
1. �Select [Hide Graphics] to temporarily hide any graphics, annotation ormeasurement.
2. �The [Hide Graphics] button works as a toggle. Once you have selectedHide, then the button changes to [Show Graphics], allowing you to showany graphics, annotation or measurement.
Erase
The Erase function allows you to erase annotations, cursors andmeasurements.
To erase, simply click on [Erase]. If you have multiple annotations or cursorson the image, the erase function will erase the active one (blue).
Screen Save
Screen Save will save everything that is on the image in primary focus.
1. �Images must be in primary focus.
2. �Select [Screen Save].
Note : Screen Saved image will be kept under series number 99.
Display Top Level Screen
3-21CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Measurements 1
[Measurements] provides access to the more advanced imagemeasurements
Measurements
Clicking on [Measurements] opens the following menu.
BoxROI
TraceROI
MeasureAngle
EllipseROI
MeasureDistance
GridOn/Off
HideGraphics
Erase ScreenSave
Back
MIROI ReportPixels
Display Top Level Screen
3-22CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Measurements 2
MIROI (Multiple Image ROI)
This function gives a graphical representation of CT number change overtime. This is useful for multiple scans done at one location or for segmentedreconstruction of a scan.
1. �Images must be in primary focus.
2. �Select [MIROI] from the Measurements menu.The following menu appears.
MIROI Analysis
Image Range All
Scale for HU Absolute Relative
Define Region Ellipse ROI Box ROI Trace ROI
OK Cancel
3. �The first selection on the menu is Image Range. The default is All.If you wish to select certain image range, first click on [All] to turn it off,then type in the number range.
4. �The next selection is Scale for HU (Hounsfield Units). The default isRelative, meaning the HU scale will display the first CT number as zeroand display the difference in CT numbers from one scan to another.The Absolute scale will display the actual CT numbers and thedifference from one scan to another.
5. �Define Region selection allows you to choose the Ellipse, Box or TraceROI.
6. �Choose [OK] to continue or [Cancel] to cancel the function.
Once you select [OK], the analysis will be automatically done with theresulting graph displayed.
You can use [Screen Save] to save the plotted graph.
Display Top Level Screen
3-23CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Measurements 3
Report Pixels
This function allows you to get individual pixel values readings from a certainsmall area.
1. �Select [Report Pixel] from the Measurements menu.
2. �An 11x11 pixel box will appear along with the Report Pixels menu.The size of the pixel box can not be changed but you can move the boxto the desired location by clicking and dragging on the letters RP or onthe box itself.
3. �Once you placed the box, select [OK].
Report Pixels
Position Box and Select OK
OK Cancel
A graph like the following example will be displayed showing the pixel valueswithin the box.
Report PixelsID: 8120
I 84 I 85 I 86 I 87 I 88 I 89 I 90 I 91 I 92 I 93 I94
I 165I 166I 167I 168I 169I 170I 171I 172I 173I 174I 175
����������������������������������������������������������������
����������������������������������������������������������������
����������������������������������������������������������������
����������������������������������������������������������������
�������������������������������������������������������������������������������������������������������������������������������
���������������������������������������������������������������
����������������������������������������������������������������
����������������������������������������������������������������
����������������������������������������������������������������
����������������������������������������������������������������
You can save this chart by [Screen Save].
Display Top Level Screen
3-24CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Display Preference 1
Display Preference provides access to functions that affect how images willbe displayed.
1. �Select [Display Preference] from the Exam Rx menu.
Display Preference
2. �The following Display Preference menu appears.
AnnotationLevels
Auto MinifyAutoEnlarge
InverseVideo
Next/PriorEachViewport
Next/PriorSeriesBinding
ContinuousReportCursor
����
�Not available) �Not available)
Display Top Level Screen
3-25CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Display Preference 2
Annotation Levels
This function allows you to change the annotation levels of the screen andfilm.
1. �Select [Annotation Levels] from the Display Preference menu.The following menu appears.
Annotation Levels
Screen
No Annotation
Partial Annotation
Full Annotation
Custom Annotation
No Annotation
Partial Annotation
Full Annotation
Custom Annotation
Film
Customize Customize
Cancel
2. �You can change the annotation level by clicking on the diamond or thewords for Screen and Film. The selected choice will be highlighted inblue.
– No Annotation : Only window width and level
– Partial Annotation : Exam number, image number, anatomicreference, table location, patient name, date of exam, RAS coordinate,and the measuring scale
– Full Annotation : All image annotations (Default)
– Custom Annotation : Customized annotation
Display Top Level Screen
3-26CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Display Preference 3
3. �Select [Customize] to review or edit the annotation levels of AdvancedAnnotation. The following menu appears.
Customize
1. Hospital Name2. Patient Name3. Patient ID4. ESI Information5. Scan Time6. Scan Parameters7. View Parameters8. Top Marker9. Bottom Marker10. Left Marker11. Right MarkerW. WindowingG. Graphic Annotation
OK Cancel
4. �Click on � to select the items of annotation.
Display Top Level Screen
3-27CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Display Preference 4
Next/Prior Each Viewport
This function is used with a multiple image display. With this function on,when you select to look at the next or prior set of images, you will only getone new image each time you select next or prior.
With this function off, selecting the next or prior image in the multiple imagedisplay will give you a new set of images. The default is off.
Next/Prior Series Binding
With this function on, you can go to the next or prior series within an exam ifyou reach the end of a series using the next or prior image functions.With this function off, when you reach the end of a series, selecting nextimage will return you to the first image of that series.
Continuous Report Cursor
This function allows you to see the RAS coordinates when you place themouse cursor on an image.
When the function is on, place the mouse cursor on the area of interest.The coordinate readout will be displayed at the bottom of the imageannotation, on the upper left of the viewport. You will notice a V, denoting theCT number of the one pixel located directly under the cursor.This information does not show up when filmed.
Display Top Level Screen
3-28CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Manual Film Composer
Refer to the later chapter for the detailed function description.
1. �Select [Manual Film Composer] from the Exam Rx Display menu.
[Manual FilmComposer]
2. �The following Film Composer will appear.
Laser CameraFormats
Film Composer
Options
Clear
Current status is :
Display Top Level Screen
3-29CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Auto Film Composer
Refer to the later chapter for the detailed function description.
1. �Select [Auto Film Composer]
[Auto FilmComposer]
2. �The following Auto Film Composer will appear.
Auto Film
Clear
Display Top Level Screen
3-30CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Accelerator Bar 1
There is an Accelerator Bar at the bottom of the Exam Rx menu.
Accelerator Bar
You can type in here various commands to invoke various functions.
filter fi
gray scaleenhance
gse
e1, e2, e3, s1, s2, s3, lung, off
��������������
exam � � �!�"#$
image i i<image#>
next exam ne
next series ns
%����� ��!�" %�
screen annota-tion none an
&���' annota-tion partial
�%
&���'��''ta-tion full ��
&���'��''�(���'�&� �"
�&
�''����'���%
��% ')��*��!�!�!
������� ����
�+�%�+�������,� �+�
�+�%��%�-��"
��-
�""�'. �--�������' �'��'�
Display Top Level Screen
3-31CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Accelerator Bar 2
rotate left rl
rotate right rr
normal '
& &�++ !�%�!�+$� �%�!�+$
windowlevel
wl
��'.���.�,
��
����'����+��'.��'�
��
/" / ��&��$
�'��� � ��
�'�+� �'� !�$ �$ !�$ �$ !�$ �$
&� ���� !� 0 1 ���� #$
&� refappend
!��
&� ������� '!�
��&2�"��2 �" '����
��&2�"��2�����&�+
�"� '����
��&2�"��2,��/'��+
�", '����
���. ���. '����
� ����''� ��
��� �����%,�& ��
��� ���++����%,�&
���
,�.�����%,�& ,�
,�����%,�& �
%��'�� ���� %�
��!��%����!�"
��
�""�'. �--�������' �'��'�
��!��%��� ����
�
��!��%�����
�%�
Display Top Level Screen
3-32CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Accelerator Bar 3
mouse modescroll
mmr
�""�'. �--�������' Contents
mouse modezoom
mmz
user testpage ��%
���� �-�'.�'� - '����
r imagematte �"����
e imagematte �"����
&���'� ��� &' ���
-+�'2�����%�� -+�'2
��%���%�!�+ �%
������'&��&��
��34 �,���&����"������ ���� �4�'�������'&���"���
Chapter 4
4-1CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
SCAN
Patient Positioning 1The scan procedure starts with patient positioning.Image quality depends on proper patient positioning.Poor patient positioning may result in artifacts and inaccurate CT value.
WARNING!
Avoid patient injury ! Use the security straps provided with the system toimmobilize the patient. The security straps secure the patient to the tableand prevent the patient from falling to the floor. Never leave a patientunattended while the patient is in the scan room. A violent patient maybreak the straps or become unsecured.
Be extremely careful with patient IV or oxygen lines, patient’s hands andfeet and loose clothing to avoid patient injury during table travel.
The cradle has a maximum distribution load capacity of 450 pounds (205Kg). Exceeding this load limit could affect some mechanical performance.
Scan
4-2CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Patient Positioning 2
1.�In order to safely lay a patient on the cradle, make sure first that thecradle is locked. (Refer to Chapter 1 for the function of Gantry DisplayPanel.)
Positioning lightInterference
TiltDistance from Landmark (mm)
�
������������� ����
X–ray On
Releasecradle lockDistance from iso–center On/Off
angle/range
� � � �
2.�Press a Positioning Light button on the Gantry Control Panel to light a
laser marker . (Refer to Chapter 1 for the function of Gantry ControlPanel.)
WARNING!
The laser beam used for patient positioning can cause eye injuries.Be sure to instruct a patient never to stare into the beam.
3.�Match the laser marker to an anatomical landmark of the patient byusing control buttons on the Gantry Control Panel.
Internallaser
�������
���
190 mm
Tomographicplane
Patient positioning ends when the table location figure is displayed on thedisplay panel.
Scan
4-3CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Patient Positioning 3
�����Main landmarks for CT examination
The following landmarks are usually used in CT examinations.
��
��
�
��
��
��
��
��
��
GB : Glabella�
OM : Orbital Meatal line�
EM : External Auditory Meatus
SN : Sternal Notch�
XY : Xyphoid�
CM : Costal Margin
IC : Iliac Crest�
UB : Umbilicus
SP : Symphysis Pubis�
Scan
4-4CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
New Patient 1
Select [New Patient] to initiate a new examination.
New Patient
1.�Select [New Patient] to open the following Patient Information/ProtocolSelection screen.
Patient Information
Exam Number
Accession Number
Patient ID
Patient Name
SexBirthdate
Age
Weight Kilograms Pounds
Referring Physician
Radiologist
Operator
History
Exam Description
Protocol Number
EndExam
Protocol Selection
Req.Proc.ID
2.�Enter the patient demographic data.
Note : As a minimum, the patient ID must be entered.
– Exam Number : Within 12 characters
– Accession Number : Within 16 characters
– Patient ID : Within 16 characters
– Patient Name : Within 64 characters
Scan
4-5CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
New Patient 2
– Sex : M (Male) or F (Female)
– Birthdate : Year, Month, Day
– Age : Years, Months, Weeks, Days
– Weight : Kgs or Pounds
– Referring Physician : Within 32 characters
– Radiologist : Within 32 characters
– Operator : Within 3 characters
– History : Within 60 characters
– Exam Description : Within 22 characters
– Protocol Number : Within 5 characters
Note : Patient Name and Exam Description will be posted in Image WorksBrowser.
3.�Each of the ten anatomic regions contains 15 protocols. To select thedefault protocol for the region, point and click on the protocol next to thenumber. To view and select one of the 15 protocols for the region, pointand click on the region itself on the full body icon.
4.�When scanning a pediatric patient, the Pediatric Anatomical Selector isinitiated by touching the pediatric icon. This will minify the AdultAnatomical Selector.
Upon the selection of the protocol, the system automatically proceeds to theScout prescription screen on the next page.
Scan
4-6CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scout Scan 1
Scout Prescription Screen
When you finish entering the necessary data on the New Patient screen, thefollowing screen will open for Scout scan prescription.
Name : ID :
Anatomical
Series Description
Select Next Create Repeat�������
End One
Reference
Protocol :
Auto Transfer
AutoStore
AddScout
DeleteSelected
NewSeries Series
Scout
StartLoc.
EndLoc. kV mA Scout
Plane
AutoVoiceNo.
ScoutNum
1
ScanType
Scout
2 Scout
Image Area
Exam: Series:
Exam Protocol More������
Note : Make sure the patient’s position in the protocol you have selectedmatches that on the CT table. Compare your patient’s position with theon–screen picture in the upper left corner of the screen. Changes to thepatient’s position in the protocol can be made by Arrow icons on thepicture. In order to reverse the position of Head–first or Feet–first, point andclick on the icon. In order to change the position of Decubitus left/right orSupine/Prone, point and click on the or icon.
Before entering or modifying any starting or ending locations for a Scoutview, the system will determine if an anatomic reference point has beenestablished. If it has not, the starting and ending location parameters will behighlighted in red. To proceed, you must establish an anatomic referencepoint by using the internal or external landmark buttons on the gantry control.
Scan
4-7CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Scout Scan 2
Image Area : Images will be displayed here. This image area can beenlarged to full screen by clicking on the small square icon located in theupper right corner.
Icons on Scout prescription screen
Add Scout
AddScout
Select this to add the same Scout as the last one.
Delete Selected Scout
DeleteSelectedScout
Select this to delete the selected Scout series.
End Exam
EndExam
Select this to end the current exam.
Select New Protocol
SelectNew
Protocol
Select this to return to the Protocol Selection screen and choose a newprotocol for a different region.
Next Series
NextSeries
Select this to choose a next series in the protocol.
Scan
4-8CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scout Scan 3
Create New Series
CreateNew
Series
Select this to create a new series.
Repeat Series
RepeatSeries
Select this to repeat the current series.
One More
OneMore
Select this to repeat the current Scout series.
Confirm
Confirm
Select this icon to accept the prescription and proceed to X–ray radiation. This icon will change to the following [Pause] icon during the scan.
Pause
Pause
Select this icon to temporarily halt the scan. This icon will change to the[Resume] icon shown below during the pause.
Resume
Resume
Select this icon to resume the scan.
Scan
4-9CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Scout Scan 4
1.�Enter or modify each parameter for a Scout scan.
– [Scan Type] : Select Scout.
– [Start Loc.] : Scan start location
– [End Loc.] : Scan end location
Note : Scout locations must be preceded with S (superior) or I(inferior). “Superior” means the direction towards patient’s head fromthe zero mm location, namely reference point. “Inferior” means thedirection towards patient’s feet from the zero mm location. Scoutscans are performed from the superior to the inferior regardless of thepatient position.
Note : The valid Scout scan range is between 50mm and 1000mm.
– [kV] : 80,120,140 kV�Default 120 kV�
– [mA] : 60–200mA (at 80 kV), 10–200 mA (at 120 kV), 10–160 mA (at 140 kV)
– [Scout Plane] : azimuth: 5–degree increment (However, to be able tographically prescribe axial or helical scan locations, the Scout planesmust be zero, 90, 180 or 270 degrees only.)
Note : If the data field is highlighted in orange, it means the original data hasbeen modified. Also, if the data field is highlighted in red, it means thesystem can not accept that entry.
2.�After entering or modifying parameters, select [Confirm] to initiate aScout scan. The following Scan Progress screen appears.
Scan Progress
EndExam
NextSeries
RepeatSeries
OneMore
PriorityRecon
RepeatLast
Group
0 10
SecondsNumber Type kV mA Start
S65.0001 120 80 S65.0 I120.0
Scanning Delay Timer Patient Handling
Pause
I120.0Scout
Plane End
0
Exam:128 Series:1
BiopsyRx
Scan
4-10CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scout Scan 5
3.�Press [Move to Scan] button on the key board when it flashes.
Move to Scan Stop Move Start Scan Pause Stop Scan
Scan Related Buttons on the Keyboard
X–ray on
4.�Press [Start Scan] button when it is lit.
The system automatically starts the prescribed Scout scan.
Note : A scan will be automatically aborted unless you press [Stop Scan]within 30 seconds after it is lit.
5.�After finishing the first Scout scan, you may select [Repeat Series]button on the Scan Progress screen to repeat the same Scout series.
6.�If there is an axial series in the protocol you used, you can proceed toAxial/Helical View/Edit screen by clicking on [Next Series] button on theScan Progress screen.
7.�If you need to return to the Scout prescription screen, select thefollowing [Return to Rx] icon.
��
CAUTION
Measurements of Distance and/or Angle on the Scout image do notnecessarily yield accurate data. Do not use those data to avoidmisdiagnosis.
End of Scout scan
If you wish to end the exam after the Scout scan, select [End Exam] on theScan Progress screen.
Scan
4-11CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Auto Store
Function : Auto Store enables you to automatically store images to themedia just after reconstruction.
First make sure that an archive medium is placed in the MOD drive.
MOD Drive
1.�Click on [Auto Store] on the Axial View/Edit screen.
Name : ID :
Anatomical
Series Description
Reference
Protocol :
Auto Transfer Auto FilmSetup
ShowLocalizer
AutoStore
AutoStore
Exam: Series:
ReconArea
AutomA OFF
SmartReconOFF
2.�When Auto Store is on, all series will be automatically stored into theMOD after completion of reconstruction.
Scan
4-12CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Auto Transfer
Function : Auto Transfer enables you to automatically transfer reconstructedimages to another view station.
1.�Click on [Auto Transfer] on the Axial View/Edit screen to open the AutoTransfer Remote Host Selection menu.
OK
Host 1
Host 2
Host 3
Host 4
CT2CT3AW1AW2MR1MR2
Auto Transfer Remote Host Selection
Exam Image
Exam Image
Exam Image
Exam Image
CancelPrior Next
2.�After clicking on [Host x], select host(s) from the list to which you wish totransfer the data. You can select up to four hosts (Host1–Host4).
3.�After selecting host(s), choose either [Exam] or [Image] as the unit oftransfer data. Then, select [OK] to close the menu.
4.�In case of [Exam] selection, the system automatically starts to transferdata immediately after selecting [End Exam], whereas in case of [Image]the system automatically starts to transfer data immediately aftercompletion of reconstruction.
Note : This function can be applied to only Scout and prospective images.Other types of images, such as 3D, retrospective, or screen–saved, shouldbe manually transmitted.
Scan
4-13CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 1
Axial/Helical Prescription (View/Edit) Screen
After Scout scan, you may proceed to the following Axial/Helical Prescriptionscreen to perform Axial, Helical or Cine scans.
Name : ID :
Anatomical
Prior Next
Series Description
New Next Create
Split
Repeat�������
End
������
One
Reference
Protocol :
Auto Transfer Auto FilmSetup
ShowLocalizer
AutoStore
AddGroup Current
Group
DeleteSelectedGroup
Images ScanType
StartLoc.
EndLoc.
No. ofImages
Thick(mm)
ImageInt’val Pitch Gantry
Tilt SFOV kV mA CTDIw
Scan Timing FilmRecon
Select
Series Series More
Preview
Optimize
Exam Protocol
PriorityRecon
Rx
BiopsyRx
����������
Exam: Series:
ReconArea
(mGy)
SmartPrep
SmartRecon
AutomA OFF
OFF
Image Area : Images will be displayed here. This image area can beenlarged to full screen by clicking on the small square icon located in theupper right corner.
Prior/Next : Prior or Next key appears on the screen only when more thanthree groups are prescribed.
Scan
4-14CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 2
Continuous Scan Modes
This HiSpeed system provides continuous multiple 360–degree scan modescalled Helical or Cine.
Helical Scan : Continuous multiple 360–degree data acquisition is achievedwith continuous table travel. Images taken with this mode are annotated asHE (Helical).
Cine Scan : Continuous multiple 360–degree data acquisition is achievedwithout table travel. Images taken with this mode are annotated as CI (Cine).
Clinical Benefits of Helical/Cine Scans
� Scan speed
Since there is no interscan delay between each slice acquisition, scan speedis much faster than conventional axial scans.
Faster scan also helps reduce the total amount of contrast agent.
� No misregistration of anatomy
Since one scan sequence is performed during patient’s breath hold, itcontributes to very little misregistration of anatomy.
Scan
4-15CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 3
Prescribe Scan Parameters
1.�Enter the parameters or modify them in the protocol if necessary.The following screen appears as default or when you click on the [Scan]icon. The screen changes when you click on the [Timing], [Recon] or[Film] icon to each own parameters one.
New Next Create
Split
Repeat���
����
End One
AddGroup Current
Group
DeleteSelectedGroup
Images ScanType
StartLoc.
EndLoc.
No. ofImages
Thick(mm)
ImageInt’val Pitch Gantry
TiltSFOV kV mA
CTDIw
Scan Timing FilmRecon
Select
Series Series More
Preview
Optimize
Exam ProtocolPriorityRecon
Rx
BiopsyRx
������
(mGy)
SmartPrep
� When you wish to use the common parameters for all scan groups as toScan Type, Thick, Image Int’val, Pitch, Gantry Tilt, SFOV, kV or mA, clickon each column head and then select the alternative.
� When you wish to select parameters separately for each group, click oneach data field.
� Each parameter is described on the following page.
Scan
4-16CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 4
– [Scan Type]���Click here to open the following menu. (Your system may not have [0.8] Rotation Time.)
Select the desired Scan Type
Scan Type Axial Helical Cine
Rotation Time 1.0 2.0 3.0
Rotation Length Segment Full
1.50.8
Plus
OK Cancel
Scan Type: Axial, Helical�� Cine
Rotation Time (second): Time needed for 360–degree rotation ( “3.0” unavailable for Helical)
Rotation Length: Amount of data for one scan(Smart Helical option offers [Plus] that uses 1.6–rotation data.)
– [Start Loc.] : Scan start location preceded with S (Superior) or I (Inferior). 0.1 increment with Helical
– [End Loc]���Scan end location preceded with S (Superior) or I (Inferior). 0.1 increment with Helical
– [No. of Images]���The number of images will be automatically calculated in the end of data entry.
– [Thick (mm)] : Slice thickness: 1, 2, 3, 5, 7, 10 mm
– [Image Int’val] : Image interval
– [Pitch] : The ratio between the slice thickness and the distance acradle travels during one rotation time in Helical scan (0.1 increment)
– [Gantry Tilt] : � 30 degrees (by 0.5 degrees)
– [SFOV] : Scan Field of View: Ped–Head (18 cm), Head (25 cm), Small (25 cm), Medium (35cm), Large (50cm), Shoulder–L (50 cm)
– [kV] : 80, 120, 140 kV
– [mA] : 60–300(350) mA (at 80 kV), 10–300(350) mA (at 120 kV)� 10–250 mA (at 140 kV), 5 mA increment
– [CTDIw] : Dose index for each group will be automatically calculated and shown.
Scan
4-17CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 5
2.�Click on the [Timing] icon to open the following time parameters menu.
PrepGroup
ISD(sec)
BreathHold(sec)
BreatheTime
AutoVoiceNo.
Scan Timing Recon Film
(sec) (sec)
Click on each parameter item and select or type in the value.
– [Prep Group (Delay)] : Delay time between pressing [Start Scan]button and the start of X–ray radiation (0–360 seconds)
– [ISD] : Inter–scan delay (1–300 seconds)
– [Breath Hold] :�Duration in which a patient is required to hold breath
– [Breathe Time]���Patient breathing time
– [Auto Voice No.] :�Auto Voice setting number
Scan
4-18CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 6
3.�Click on the [Recon] icon to open the following reconstructionparameters menu.
DFOV
Scan Timing Recon Film
(cm)R/L
Center(mm)
A/PCenter(mm)
ReconType
ImageFilter
MatrixSize
MotionCorrect
SpecialFilter
ShowRecon1
ShowRecon2
ShowRecon3
Y N N
Click on each parameter item and select or type in the value.
– [DFOV] :�Display Field of View: 48–500 mm (by 1 mm)
– [R/L Center] :�Reconstruction center in the right or left direction preceded with R (Right) or L (Left)
– [A/P Center] :�Reconstruction center in the anterior or posterior direction preceded with A (Anterior) or P (Posterior)
– [Recon Type] :�SOFT, STND, STD+, DETL, CHST, BONE, EDGE or PFRM
– [Image Filters] :�Smooth (S1, S11, S2, S21, S3), Edge (E1, E2, E21, E22, E23, E3), Lung (L1, L2, L3)Note : These filters are not available when CHST is selected at [ReconType].
– [Matrix Size] :�256�256 or 512�512
– [Motion Correct] : Motion correction
Note : Motion Correct does not apply to the following cases.Helical/Cine scans and Segment Rotation Length
Scan
4-19CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 7
– [Special Filter] : ANR (Advanced Noise Reduction) (1 or 2) AAR (Advanced Artifact Reduction) (1 or 2)
ANR filter allows you to reduce noise without compromising spatialresolution. It contributes to a decrease in standard deviation byapproximately 10%, which may be equivalent to one–step decrease inmA. Note : ANR is compatible only with SOFT, STND, STD+.
AAR filter allows you to reduce artifacts shown around arms.
When you click on the [Special Filter] button or the each Special Filterfield, the following menu appears.
None ANR 1 ANR 2 CancelAAR 1 AAR 2
Select the desired Special Image Filter.
[ANR 2] has a stronger effect than [ANR 1]. Also, [AAR 2] has astronger effect than [AAR 1]. Select either of them or select [None]not to use this filter.
– [Show Recon 1, 2, 3] : These buttons are used for prospective multiple reconstruction.
Scan
4-20CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 8
4.�Click on the [Film] icon to open the following menu for auto filming.You may skip this procedure if you do not use autofilm.
Scan Timing Recon Film
AutoFilm
FrameFormat
Interval Flip Width1
Level1
MagFactor Rotate User
Anno.GSE
Recon 1Film Set
1
Recon 1Film Set
2
Recon 2Film Set
1
Recon 2Film Set
2
Recon 3Film Set
1
Recon 3Film Set
2
Click on each button to enter or select data.
– [Auto Film] :�Toggle between autofilm On and Off.
– [Frame Format] :�Four formats
– [Interval] : 1–5 sheets
– [Flip] : FTB (Flip to Bottom), FTB/FLR (Flip to Bottom/Flip to Left Right), FLR (Flip Left Right)
– [Width 1] : First choice of window width (1–4,096)
– [Level 1] :�First choice of window level (–1,024 to 3,072)
– [Mag Factor] : 0.5–2.0 times
– [Rotate] :�Left 90 degrees, right 90 degrees, 180 degrees
– [User Anno] : User annotation
– [GSE] :�Gray Scale Enhancement (G1, G2, G3, Off)
Scan
4-21CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 9Prospective Multiple Reconstruction
Function : This enables you to prospectively prescribe up to two additionalreconstruction sets for each scan group within one series.
Make sure that the [Recon] icon is selected on the right side of View/Editscreen to prescribe Prospective Multiple Reconstruction.
DFOV
Scan Timing Recon Film
(cm)
R/LCenter(mm)
A/PCenter(mm)
ReconType
ImageFilter
MatrixSize
MotionCorrect
SpecialFilter
ShowRecon1
ShowRecon2
ShowRecon3
Y N N
ReconY/N
N
N
There are three buttons at the bottom of the Recon menu.They are [Show Recon 1], [Show Recon 2] and [Show Recon 3].[Show Recon1 Y] is highlighted in blue and contains recon parametersprescribed beforehand.
1.�Select [Show Recon2 N] to prescribe the second recon set.
2.�Once [Show Recon2 N ] is selected, there appears [Recon Y/N] columnat the left end of the Recon menu. These parameters are default fromRecon1. You may change some or all of those parameters. The alteredparameters will become those of Recon2. Lastly, toggle N (No) to Y(Yes) to allow the group to be reconstructed with those parameters.[Show Recon2 N] will be changed to [Show Recon2 Y] and highlightedin blue.
– [DFOV] :�Display Field of View: 48–500 mm (by 1 mm)
– [R/L Center] :�Reconstruction center in the right or left direction preceded with R (Right) or L (Left)
– [A/P Center] :�Reconstruction center in the anterior or posterior direction preceded with A (Anterior) or P (Posterior)
Scan
4-22CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 10Prospective Multiple Reconstruction
– [Recon Type] :�SOFT, STND, STD+, DETL, CHEST, BONE, EDGE or PERM
– [Image Filters] :�Smooth (S1, S11, S2, S21, S3), Edge (E1, E2, E21, E22, E23, E3), Lung (L1, L2, L3)
– [Matrix Size] :�256�256 or 512�512
– [Motion Correct] : Motion correction
– [Special Filter] : ANR (Advanced Noise Reduction) (1 or 2) AAR (Advanced Artifact Reduction) (1 or 2)
ANR filter allows you to reduce noise without compromising spatialresolution. It contributes to a decrease in standard deviation byapproximately 10%, which may be equivalent to one–step decrease inmA. Note : ANR is compatible only with SOFT, STND, STD+.
AAR filter allows you to reduce artifacts shown around arms.
When you click on the [Special Filter] button or the each Special Filterfield, the following menu appears.
None ANR 1 ANR 2 CancelAAR 1 AAR 2
Select the desired Special Image Filter.
[ANR 2] has a stronger effect than [ANR 1]. Also, [AAR 2] has astronger effect than [AAR 1]. Select either of them or select [None]not to use this filter.
When you need to return to the parameters of Recon1 , simply select [ShowRecon1].
3.�If you need to perform the third reconstruction, select [Show Recon3],then take the same steps as Recon2.
4.�Select [Confirm] to initiate a scan.
Note : During the scan, only the set of Recon1 will be reconstructed. In orderto activate Recon2 or 3, select any of the following buttons.[End Exam], [Create New Series], [Next Series], [Select New Protocol],or [Repeat Series]
Scan
4-23CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 11
Icons of the Axial/Helical Scan Prescription screen (1)
Autofilm Setup
AutofilmSetup
Select this icon to open the following menu for autofilming.
Laser Camera
OK
Autofilm Setup
Destination
Format Size
Normal Slide
Copies
�
Film Direction
Top to Bottom
Auto Start Auto Print
Yes Yes
Exam Page Series Page
Yes Yes
Scout XRef–Scout
Auto Film Composer
e/s/i
Show Grayscale
Yes
Cancel
Please refer to the later chapter of Filming for detailed descriptions.
Show Localizer
ShowLocalizer
Click this icon to display the Scout image with graphical lines that show thescan locations prescribed beforehand. You can graphically modify thestart/end location, interval, image number, or DFOV. Those changes will bereflected on the scan prescription screen in orange color.
Please refer to the later pages for detailed descriptions.
Scan
4-24CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 12
Recon Area
ReconArea
Select this to display a circle cursor over the image, then change the sizeand location of the cursor to designate the recon area instead of typing inDFOV, R/L Center and A/P Center at the Axial/Prescription screen.
Add Group
AddGroup
First select a certain scan group, then click on this to add that group justafter the current group.
Split Current Group
SplitCurrentGroup
Select this to split the selected scan group.
1.�First select the desired scan group you wish to split.
2.�Then, click on the above icon to open the following menu.
Split Group
Enter location or image number
OK Cancel
3.�Enter the location or image number at which you wish to split the group.Then, select [OK].
Note : If you enter an invalid number, the warning message will appear.“Image number is outside of group image range.”
Scan
4-25CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 13
Delete Selected Group
DeleteSelectedGroup
First, select a scan group that you wish to delete, then click on this.
Biospy Rx
BiopsyRx
Select this to prescribe Biopsy Rx.
Refer to the later pages for the function description of Biopsy Rx.
Smart Prep
PrepSmart
Select this to prescribe Smart Prep.
Refer to the later pages for the function description of Smart Prep.
Scan
4-26CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 14
Optimize Rx
Optimize
Rx
When the system decides that some change in scan techniques is needed tocomplete the entire scan, this icon lights up in red. At this point you mayenter the following Technique Optimize screen by clicking on it.
Tube cooling is needed at image ; 26
Optimize Method :
Group byGroup
UpfrontDelay(sec) mA
240
GroupDelay(sec)
Image
1–45
46–65
Technique Optimize screen example
Tube cooling is needed at image : At the top of the screen the messageappears that informs you at which image the tube cooling will be needed withthe current scan techniques. In the above example it is 26.
Optimize Method : Optimize Method tells you how the change in techniquewill occur. In the above example it says “Group by Group” meaning thetechnique must be changed by scan group. Also, in the above example,since the image 26 belongs to the first group, all the images in the first groupfrom 1 to 45 are equally subject to changes in technique. The system willcalculate and display up to three factors for the scans to be able to continue.Those three factors are;
– Upfront Delay (sec) :�Pre–scan delay
– mA :�Tube current used for that scan group
– Group Delay (sec) :�Delay time between scan groups
Rectifying any one of those three factors will display a message “Tubecooling is No Longer Needed” enabling the system to continue scans.
Notice: When the system estimates the cooling time to be more than 600seconds, it shows “––” in the [Upfront Delay] box.
Finally select [Optimize in Progress] highlighted in blue to exit.
Scan
4-27CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 15
Preview
Preview
This function enables you to graphically view prior to the scan the interactionbetween scan process and the combination of Breath Hold and BreatheTime.
When you click on this button after axial scan prescriptions, the followingscreen appears.
0 10 30 40
SecondsImages Type kV mA Time Thk Tilt Start
001–010 120 250 10.0 5 S80.0 S30.0S80.0 S30.0
BiopsyRx
20End
S0.0Helical
Preview Screen
011–022 120 250 12.0 5 S10.0S0.0Helical I50.0
023–027 120 250 5.0 5 I60.0S0.0Helical I85.0
S10.0 I50.0
I60.0 I85.0
Prior Next
ScrollStep
If the prescribed scan sequence is too long for a displayed time scale, useStep or Scroll button to view the entire sequence for the corresponding timescale. (Step/Scroll buttons appear on the screen only when too long scansequence is prescribed.)
When more than three groups are prescribed, use Prior or Next key to gothrough the pages. (Prior or Next key appears on the screen only when morethan three groups are prescribed.)
Scan
4-28CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 16
Icons of the Axial/Helical Scan Prescription screen (2)
End Exam
EndExam
Select this to end the current exam.
Select New Protocol
SelectNew
Protocol
Click on this to select another protocol.
Next Series
NextSeries
Click on this to select another series within the protocol. Please refer to thelater page for the function description.
Create New Series
CreateNew
Series
Select this to open the following menu and create a new series that does notexist in a protocol.
Select New Series Type
OK Cancel
Scout Axial
ATTENTION: This will remove all unscanned groups.
Scan
4-29CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 17
Repeat Series
RepeatSeries
Select this to repeat the current series. Please refer to the later page for thefunction description.
Priority Recon
PriorityRecon
Select this to give the priority to a certain image to be reconstructed earlierthan other images. Refer to the later page for the details.
One More
OneMore
Select this to repeat the last one scan. Please refer to the later page for thefunction description.
Confirm
Confirm
Select this icon to proceed to X–ray radiation after confirming all scanprescriptions. This icon will change to the [Pause] icon shown below duringthe scan.
Scan
4-30CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 18
Pause
Pause
Select this icon to temporarily halt the scan. Scans pause after completingthe current one scan. This icon will change to the [Resume] icon shownbelow. (Note : It is best to use the Stop Scan button on the key board to stopthe scan immediately. The use of the Pause icon may not stop the scanimmediately due to the signal transmission timelag.)
Resume
Resume
Select this icon to resume the scan.
Intravenous contrast icon
��
Select this icon when you use intravenous contrast agent.
Gastrointestinal contrast icon
��
Select this icon when you use gastrointestinal contrast agent.
Scan
4-31CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 19Show Localizer
Function : Show Localizer allows you to graphically show the prescribedscan lines on the Scout image.
1.�Click on [Show Localizer] located on the upper left corner of theView/Edit screen.
Name : ID :
Anatomical
Series Description
Reference
Protocol :
Auto Transfer Auto FilmSetup
ShowLocalizer
AutoStore
ShowLocalizer
View/Edit screen
Exam: Series:
ReconArea
AutomA OFF
SmartReconOFF
2.�The Scout image appears with lines that show prescribed scan locations.
3.�In order to adjust the slice location, click and hold on the “X” mark inblue, then drag the entire group horizontally to the desired location.
4.�In order to adjust the DFOV center, click and hold on the “X” mark in bluewith the Shift key pressed down, then drag the entire group vertically tothe desired location.
Scan
4-32CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 20Show Localizer
5.�In order to adjust the location of only one group, first doubleclick on thegroup to select, then do the above operation. (When you select a certaingroup by doubleclicking on it, it turns blue.)
6.�In order to add or delete a slice at the top of the group, click and hold onthe small blue solid square and move the mouse to the desireddirection.
If you wish to add or delete a slice at the bottom of the group, click andhold on the small blue hollow square and move the mouse to thedesired direction.
7.�In order to adjust the size of DFOV, click and hold on the small diamond,then extend it or contract it.
8.�In order to adjust the gantry tilt, click and hold on the small circle, then tiltit to the desired angle.
Note : Blue lines on the axial images show the mechanical center (thepair of two slices) of the detector, not the slice.
Note : When the gantry tilt is applied on the Localizer, the last image ineach scan group will be automatically removed. This is because with the matrix type detector tilting the gantry increasesslice intervals which cause the slice overlaps between groups as thesystem keeps the original start locations for each scan group. To avoidthe slice overlaps the system automatically removes the slice(s)depending on the scan mode and the tilt angle.
Scan
4-33CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 21Auto mA
Function : The Auto mA feature enables the system to reduce mA andpatient dose with negligible effect on image quality. The system varies mAaccording to patient shape and a predetermined algorithm for scan location.
Principle : The Auto mA feature requires one Scout scan, either AP or LATin the diagram below, to obtain necessary information on X–ray attenuation.
�������
����������� !��"#�$
%!&�
%!&�
'����������� !��"(#%$
The system automatically increases mA at the scan locations where X–rayattenuation is relatively high and decreases mA at the scan locations whereX–ray attenuation is relatively low. In the example below, the mA is set low atthe start scan location (lung area), then the mA will increase at the locationof liver area, then it will decline at the location of abdomen area.
�
)#
����� ���
����� ���
����� ���� �
CTDIW : CTDIW is displayed by using pre–programmed loading factors thatdetermine the maximum possible CTDIW. If you need to know CTDIW for eachAxial scan group, first confirm the mA in “mA Table” for Auto mA Clipping,then calculate CTDIW according to the way mentioned at the CTDIW sectionin Chapter 6.
Scan
4-34CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 22Auto mA
Prescribe Auto mA
One Scout data is necessary to prescribe the Auto mA.(Note : When more than one Scout scan is available like AP and LAT, theAuto mA feature uses the last Scout scan as the necessary data.)
1.�Click on the [AutomA OFF] located on the upper left corner of theView/Edit screen. (Default setting is “OFF“.)
Name : ID :
Anatomical
Series Description
Reference
Protocol :
Auto Transfer Auto FilmSetup
ShowLocalizer
AutoStore
AutomA OFF
View/Edit screen
Exam: Series:
ReconArea
AutomA OFF
SmartReconOFF
2.�The following [Select AutomA Mode] menu appears.
LowDoseNormalIQOff MaxmA Cancel
Select AutomA Mode
Select one of the four modes, IQ, Normal, LowDose or MaxmA. The menu disappears upon the selection.
– [IQ] : Select this mode when high image quality is needed. The possible highest mA of this mode is the highest among IQ,Normal and LowDose modes.[AutomA OFF] will change to [AutomA IQ].
– [Normal] : This mode is most commonly recommended.[AutomA OFF] will change to [AutomA Normal].
Scan
4-35CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 23Auto mA
– [LowDose] : Select this mode when the lowest patient dose isrequired. The possible highest mA of this mode is the lowest amongIQ, Normal and LowDose modes.[AutomA OFF] will change to [AutomA LOWDOSE].
– [MaxmA] : Select this mode to define your own desired maximum mAother than IQ, Normal or LowDose mode.
– [Off] : Select this when you wish to end any of AutomA modes afterusing it.
– [Cancel] : Select this to cancel the AutomA mode.
Automatic mA Range Selection
In case of [IQ], [Normal] or [LowDose] mode:
The system calculates proper Auto mA values and draws a chart like thebelow one (The chart is not displayed on the screen.). Based on this chart,the system automatically applies either one of the two mA ranges (10–190mA / 100–maximum mA ) to each scan group.
Auto mA Calculation (example)mA
100
10scan location
200
The two ranges do not alternate with each other within a single scan group.In some cases this causes the system–calculated mA to be clipped(cut) for acertain slice. The following are examples.
Example 1 : With the [10–190mA] range selected, even if 210mA iscalculated for a certain slice, only 190mA will be applied to the slice. In other words, mA values more than 190mA are clipped.
Example 2 : With the [100–max.mA] range selected, even if 80mA iscalculated for a certain slice, 100mA will be applied to the slice. In other words, mA values less than 100mA are clipped.
Scan
4-36CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 24Auto mA
Auto mA Clipping
The CT operator can intentionally set the upper limit of mA in the modes ofIQ, Normal and LowDose. This feature is useful for tube cooling wait.Enter the desired mA value into the [mA] field. The following charts show the example of Clipping by Operator at 200mA.
mA
100
50
150
200
250
300
Slice 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Without Clipping (example)
Clipping by Operator @200mA (example)
mA
100
50
150
200
250
300
Slice 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Automatic Max mA Range Selection
In case of [MaxmA] mode:
Based on the mA value prescribed at [MaxmA] mode, the systemautomatically selects either one of the two mA ranges as the below chartshows.
mA input
10–190 10–190
Automatically selected range
200–max. 100–max.
Scan
4-37CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 25Auto mA
The status of ON/OFF or mode of the AutomA feature is shown onthe bottom of the Preview screen.
3.�The following menu appears upon the selection of [mA] on the View/Editscreen.
)#�%�&�*��+,�+���,��*����-�.� �����
�������/���������)#
When [AutomA OFF] is posted, the [mA Table] key looks dimindicating it is not functional.
Select or enter the desired maximum mA.
When any one of four modes (IQ, Normal, LowDose, MaxmA) isselected, all mA choice boxes look dim, whereas the [mA Table] keylooks solid. Click on the [mA Table] key to open the following mATable Information menu.
)#
�
+
*
0
,
.
1
-
'
��
��
�2
�,�
�,�
�,�
�0�
�0�
�*�
���
���
���
���
���
)#�%�&����3 �)��� �
����4
"���)5�$
The above menu shows the expected mA at each scan location.Click on [OK] to close the menu.
Scan
4-38CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan Prescription 26SmartRecon (Option)
Function : The SmartRecon feature can be used along with Helical or Cinescans to generate real–time reconstructed images at the rate of six imagesper second with 256x256 matrix.
Notice : Images taken with the SmartRecon feature are not recommendedfor diagnosis since 256x256 matrix is used.
Conditions :
� 512x512 matrix is not available for SmartRecon.
� SmartRecon can not be used for ordinary axial scans.
� SmartRecon is always succeeded by ordinary prospective reconstruction.(SmartRecon can not be prescribed alone.)
� SmartRecon is prescribed by series.
� The available scan time for SmartRecon is 0.8 or 1.0 second.
� While performing SmartRecon, the system halts all network and archivingtasks, also all image analyses like 3D, Reformat, etc. are suspended.
� During SmartRecon all the function buttons on the desktop except for[Shut Down] are not usable.
� Images generated by SmartRecon will not be stored into the system diskbut into the memory buffer.
� The AutomA feature can be applied to SmartRecon.
� The [Priority Recon] button can not be applied to SmartRecon.
Scan
4-39CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan Prescription 27SmartRecon (Option)
1.�Click on [SmartRecon OFF] located on the upper left corner of theView/Edit screen. (Default setting is “OFF“.)
Name : ID :
Anatomical
Series Description
Reference
Protocol :
Auto Transfer Auto FilmSetup
ShowLocalizer
AutoStore
View/Edit screen
Exam: Series:
ReconArea
AutomA OFF
SmartReconOFF
SmartReconOFF
[SmartRecon OFF] turns to [SmartRecon ON].
2.�As the Helical or Cine scans proceed, the system reconstructs anddisplays images at the rate of six images/second.
3.�Upon the completion of the SmartRecon the operator may return to theView/Edit screen to start the prospective reconstruction.
Scan
4-40CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Axial/Helical Scan in Progress
1.�Click on the [Confirm] icon after confirming all prescribed scanparameters.
Select Next Create RepeatNewProtocol
NewSeries
End
Confirm
One Priority
Confirm���67����������
Series SeriesExam ReconMore
2.�Press the [Move to Scan] button when lit, then, press the [Start Scan]button to start a scan.
3.�The system automatically proceeds to the Scan Progress screen.
Scan Progress
RepeatSeries
OneMore
RepeatLast
Group
Pause
0 10 30 40
SecondsImages Type kV mA Time Thk Tilt Start
Scanning Delay Timer Patient Handling
S80.0 S30.0
BiopsyRx
20End
EndExam
NextSeries
PriorityRecon
Exam:285 Series:2
001–010 120 250 10.0 5 S80.0 S30.0S0.0Helical
011–022 120 250 12.0 5 S10.0S0.0Helical I50.0
023–027 120 250 5.0 5 I60.0S0.0Helical I85.0
S10.0 I50.0
I60.0 I85.0
Note : A warning message will appear in the Patient Handling box if thethe scanning object is off center.
Note : A scan will be automatically aborted unless you press [StartScan] within 30 seconds after it is lit.
4.�The system automatically performs the prescribed axial or helical scan.You can monitor the progress of the scan on the above Scan Progressscreen.
Scan
4-41CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Axial/Helical Scan End
Upon the completion of the scan, the message “All scans completed” willappear in the Scanning box of Scan Progress screen. If you wish to end thecurrent exam here, take the following steps.
1.�After the scan, click on [End Exam] located on the lower left corner ofthe Scan Progress screen.
Scan Progress
EndExam
RepeatSeries
OneMore
PriorityRecon
RepeatLast
Group
Scanning Delay Timer Patient Handling
BiopsyRx
End Exam
All scans completed.
NextSeries
Exam:285 Series:20 10 30 40
SecondsImages Type kV mA Time Thk Tilt Start
20End
023–027 120 250 5.0 5 I60.0S0.0Helical I85.0I60.0 I85.0
2.�Or, you may select “Return to View/Edit screen” icon located on the rightside of Scan Progress screen to return to the View/Edit screen.�
�� Return to View/Edit screen
3.�Select [End Exam] on the lower left corner of the View/Edit screen toend the current exam. The screen automatically returns to Top Levelscreen.
Scan
4-42CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Priority Recon
Function : Priority Recon allows you to designate and reconstruct an imageearlier than any other image in the reconstruction queue. Upon the selectionof this function, the system will start to reconstruct the designated image justafter reconstructing the current image.
1.�In the scan process, if you wish to reconstruct a certain image earlierthan any other image, select [Priority Recon] located on the bottom rightof Scan Progress screen.
Scan Progress
Scanning Delay Timer Patient Handling
BiopsyRx
Priority Recon
EndExam
RepeatSeries
OneMore
PriorityRecon
RepeatLast
Group
NextSeries
Exam:285 Series:20 10 30 40
SecondsImages Type kV mA Time Thk Tilt Start
20End
023–027 120 250 5.0 5 I60.0S0.0Helical I85.0I60.0 I85.0
Pause
The system starts to reconstruct the designated image just aftercompletion of the current reconstruction.
2.�After completion of the whole scan, if you wish to give the reconstructionpriority to the lastly scanned image, select [Priority Recon] on thebottom right of the View/Edit screen.
SelectNext Create RepeatNew
ProtocolNew
Series
End
Confirm
OneMore
Priority
Priority Recon���67����������
Series SeriesExam Recon
Scan
4-43CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Next Series
Function : Next Series allows you to directly proceed to a scan prescriptionscreen if the next axial or helical series is included in the protocol.
1.�Click on [Next Series] on the following scan end screen.
Scan Progress
Scanning Delay Timer Patient Handling
BiopsyRx
Next Series
All scans completed.
EndExam
RepeatSeries
OneMore
PriorityRecon
RepeatLast
Group
NextSeries
Exam:285 Series:20 10 30 40
SecondsImages Type kV mA Time Thk Tilt Start
20End
023–027 120 250 5.0 5 I60.0S0.0Helical I85.0I60.0 I85.0
2.�The system automatically proceeds to the View/Edit screen that showsthe parameters of the series in the protocol. You may modifyparameters.
NewNext Create
Split
Repeat�������
End
������
One
AddGroup Current
Group
DeleteSelectedGroup
Images ScanType
StartLoc.
EndLoc.
No. ofImages
Thick(mm)
ImageInt’val Pitch Gantry
TiltSFOV kV mA
Scan Timing FilmRecon
Select
Series Series More
Preview
Optimize
Exam Protocol
PriorityRecon
Rx
BiopsyRx
CTDIw(mGy)
���67�����������
3.�After confirming the parameters, select [Confirm] to start scan.
Scan
4-44CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Repeat Series 1
Function : Repeat Series allows you to select any one of the previouslyscanned series within the same exam.
1.�Click on [Repeat Series] on the following scan end screen.
Scan Progress
Scanning Delay Timer Patient Handling
BiopsyRx
Repeat Series
All scans completed.
EndExam
RepeatSeries
OneMore
PriorityRecon
RepeatLast
Group
NextSeries
Exam:285 Series:2
0 10 30 40
SecondsImages Type kV mA Time Thk Tilt Start
20End
023–027 120 250 5.0 5 I60.0S0.0Helical I85.0I60.0 I85.0
2.�There appears the following Select Previous Series menu which lists allthe previously scanned series that have the same Patient Position/Orientation (supine/prone, head first/feet first) as the most recentlyscanned one. The most recently scanned series tops the list and getshighlighted.
Select Previous Series
OK Cancel
Series 8
Series 7
Series 6
Series 5
Series 4
Series 3
Description of Series 8
Description of Series 7
Description of Series 6
Description of Series 5
Description of Series 4
Description of Series 3
Highlighted
The list can show up to ten series on one screen. When it containsmore than ten series, there appear(s) Prior and/or Next button on thescreen.
Scan
4-45CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Repeat Series 2
After selecting and highlighting any one of the series, select [OK] toconfirm the selection. To cancel this step, click on [Cancel].
3.�After the selection of [OK], the system automatically proceeds to theView/Edit screen that shows the parameters of the series chosen in theSelect Previous Series menu. You may modify parameters here.
NewNext Create
Split
Repeat�������
End
������
One
AddGroup Current
Group
DeleteSelectedGroup
Images ScanType
StartLoc.
EndLoc.
No. ofImages
Thick(mm)
ImageInt’val Pitch Gantry
TiltSFOV kV mA
Scan Timing FilmRecon
Select
Series Series More
Preview
Optimize
Exam Protocol
PriorityRecon
Rx
BiopsyRx
CTDIw(mGy)
���67�����������
4.�After confirming the parameters, select [Confirm] to start scan.
Scan
4-46CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
One More
Function : One More allows you to scan one more slice that will haveexactly the same parameters as the most recently scanned image.
1.�Click on [One More] on the following scan end screen.
Scan Progress
Scanning Delay Timer Patient Handling
BiopsyRx
One More
All scans completed.
EndExam
RepeatSeries
OneMore
PriorityRecon
RepeatLast
Group
NextSeries
Exam:285 Series:20 10 30 40
SecondsImages Type kV mA Time Thk Tilt Start
20End
023–027 120 250 5.0 5 I60.0S0.0Helical I85.0I60.0 I85.0
2.�Press [Move to Scan] button when lit.
3.�Then, press [Start Scan] button when lit to start a scan.
Scan
4-47CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Repeat Last Group
Function : Repeat Last Group allows you to scan the most recently scannedgroup again with the same condition.
1.�Click on [Repeat Last Group] on the following scan end screen.
Scan Progress
Scanning Delay Timer Patient Handling
BiopsyRx
Repeat Last Group
All scans completed.
EndExam
RepeatSeries
OneMore
PriorityRecon
RepeatLast
Group
NextSeries
Exam:285 Series:20 10 30 40
SecondsImages Type kV mA Time Thk Tilt Start
20End
023–027 120 250 5.0 5 I60.0S0.0Helical I85.0I60.0 I85.0
2.�Press [Move to Scan] button when lit.
3.�Then, press [Start Scan] button when lit to start a scan.
Note : Newly scanned series will have the same series number as theprevious one.
Scan
4-48CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Biopsy Scan 1
Function : The Biopsy Rx feature allows you to easily repeat the scanlocation during the biopsy procedures.
Biopsy Rx prescription
1.�Biopsy Rx can be accessed through the [Biopsy Rx] icon on the rightside of Axial/Helical Prescription screen.
Repeat One
Thick(mm)
ImageInt’val Pitch Gantry
TiltSFOV kV mA
CTDIw
Scan Timing FilmRecon
Series More
Preview
Optimize
PriorityRecon
Rx
BiopsyRx
������
(mGy)
SmartPrep
BiopsyRx
2.�The following screen appears upon the selection of [Biopsy Rx].
Biopsy Rx
Biopsy Reference Superior Centered Inferior
Biopsy Location Get AlignmentLight Location Internal External
Number of Images Gantry Tilt
Thickness
Helical Pitch Image Interval
ConfirmBiopsy Rx Cancel
Scan
4-49CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Biopsy Scan 2
3.�In order to determine the reference centering location in a biopsy scan,select [Superior], [Centered] or [Inferior] at Biopsy Reference field.
[Superior] means scanning from the landmark toward patient’s head.
[Centered] means scanning around the landmark.
[Inferior] means scanning from the landmark toward patient’s feet.
4.�At Biopsy Location field, enter the location of the biopsy scan to beperformed.
Or
5.�When the internal light is used, select [Internal], or, when the externallight is used, select [External] at Get Alignment Light Location field.
6.�Enter the following parameters.
– Number of Images :
– Gantry Tilt :
– Thickness :
– Helical Pitch :
– Image Interval :
7.�Click on the [Confirm Biopsy Rx] button.
8.�Press [Move to Scan] button when it lights up. Then, press [Start Scan]to start the biopsy scans.
Scan
4-50CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Smart Prep (Option) 1
Function : The Smart Prep feature allows you to monitor contrastenhancement change during injection, to assure the acquisition of axial orhelical scans while optimum levels of contrast are present.
Three phases of Smart Prep
Smart Prep consists of three phases, which are a Baseline, Monitor andScan Phase.
� Baseline phase
This phase requires one unenhanced scan through a particular area ofanatomy that will be the point of interest to monitor.
� Monitor phase
This phase requires the operator to take up to 20 scans at one location whilethe IV (introvenous) contrast is being administered. You will be able to watchthese scans on the CRT as the scans are proceeding.
� Scan phase
This phase performs your scan prescription that will be started by theoperator when the optimum level of contrast enhancement is reached, eitherby visual evaluation or by a preset threshold selection.
Parameters of Smart Prep
Some of the parameters used during the Smart Prep prescription are presetand unchangeable. They are 256 matrix, Soft Tissue algorithm and 0.6–secscan time.
Note : Because monitoring scans will not be used for diagnostic purposes,these techniques are used to make sure that the patient receive the minimaldose.
Autovoice
During the Smart Prep the Autovoice function will only be available after thescan phase is initiated. So, the operator may have to give oral breathinginstructions through the intercom during Baseline and Monitor phase and atthe beginning of Scan phase.
Scout and scan prescription
When you perform Smart Prep, the Prep Delay can be excluded fromparameters. Even if you use it, Smart Prep will override it by putting “SP” inplace of the number.
Scan
4-51CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Smart Prep 3
Smart Prep prescription
1.�Smart Prep can be accessed through the [Smart Prep] icon on theAxial/Helical Prescription screen.
Repeat One
Thick(mm)
ImageInt’val Pitch Gantry
TiltSFOV kV mA
CTDIw
Scan Timing FilmRecon
Series More
Preview
Optimize
PriorityRecon
Rx
BiopsyRx
������
(mGy)
SmartPrep
SmartPrep
2.�When the Smart Prep is not incorporated in the protocol, toggle the [Off]button to [On] on the Smart Prep screen.
To [On]Smart Prep screen
Smart Prep
Off
MonitorLocation mA
MonitoringDelay
MonitoringISD
EnhancementThreshold
Scan PhaseDelay
ShowLocalizer
Accept Cancel
Each parameter is described on the following page.
Scan
4-52CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Smart Prep 4
3.�Enter the following parameters for the Smart Prep prescription.
– Monitor Location : Location of monitoring scan
– mA : Tube current for Base Line and all Monitoring scans(40–100mA�by 5 mA�
– Monitoring Delay : The delay before the Monitoring scan begins(0–90 seconds, by 0.1 sec.�This delay works in conjunction with theadministration of IV contrast.
– Monitoring ISD : The delay between each monitoring scan(3–90 seconds�by 0.1 sec.)
– Enhancement Threshold : The difference in CT value between theBaseline ROI and the one at which you wish to start the Scan Phase.( For example: If you want to start the scan phase when the CT valueof the area of interst reaches 70, then assuming the Baseline ROI is30, the Enhancement Threshold will be 40.)
– Scan Phase Delay : The delay between the time you press [StartScan] button and the time the actual scan begins (3–90 seconds, by0.1 sec.)
Note : The slice selection at Scan Phase may affect Scan Phasedelay. The selection of 7/10mm with Helical or 7/10mm x 2i will affectthe delay. This is because the selection of only up to 5mm x 2 is usedat Monitoring Scan.
– Show Localizer : Select this to display the Scout image with a line onit. With this line, you will be setting the location of the Baseline andsubsequent Monitoring scans.
Note : If there is the difference between Monitoring location and Scan Phasestart location, the Scan Phase will be delayed due to the cradle travel. Forexample, it takes some four seconds for the cradle to travel 300mm.
Scan
4-53CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Smart Prep 5
4.�After completing all the entries, select [Accept]. The system will return tothe View/Edit screen.
5.�Click on [Confirm] and press [Start Scan] when lit to initiate the Baselinescan.
Scan Progress screen shows the Baseline group, Monitor group and scanprescription.
Scan Progress
Scanning Delay Timer Patient Handling
BiopsyRx
EndExam
Monitor Scan PriorityRecon
RepeatLast
Group
NextSeries
Exam:285 Series:2
Pause
Phase Phase
0 10 30 4020Images Type kV mA Time Thk Tilt Start End
001–001 120 040 0.6 10 S0.0 S0.0S0.0
S0.0Axial
002–021 120 040 12.0 10 S0.0S0.0Axial S0.0
001–002 120 080 5.0 10 S0.0S0.0 Axial I10.0
50 60
S0.0
S0.0
Seconds
Note : When you select [Accept] on the Smart Prep prescription screen, thefollowing Timing menu shows “SP” in the Prep Group (delay) field, meaningthe Smart Prep Monitoring delay has been set.
PrepGroup
ISD(sec)
BreathHold(sec)
BreatheTime
AutoVoiceNo.
Scan Timing Recon Film
(sec)
SP
SP
Scan
4-54CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Smart Prep 6
6.�After the Baseline scan, its image will be displayed. Also, six displayfunctions for Smart Prep will be presented.
Smart Prep Display
ZoomDisplayNormal
Ellipse
Hide Erase Explicit
Baseline image
ROI
MagGraphics
Refer to the Chapter 3 [Display] for the each display function.
7.�You can calculate up to three ROIs by selecting [Ellipse ROI].
8.�After calculating ROIs, click on [Monitor Phase] and injectsimultaneously the IV contrast.
9.�After you press the [Start Scan] button, the system will automaticallyproceed to the Monitoring scans through the Monitoring delay.
� The Display desktop screen will now look similar to the following one.
Baseline image
Graph
Time
Most recently
reconstructed image
� In the lower right quadrant the Baseline image with ROIs is displayed.
� The lower left quadrant will be displaying in real time the time when eachmonitoring scan is acquired, based on the onset of the monitoring delay.It also displays each of the ROI values of that scan.
� The upper right quadrant will be displaying in real time the most recentlyreconstructed image.
� The upper left quadrant will be displaying in real time the enhancementthreshold graph, comparing the ROI of each monitoring scan with the timefrom the start of the monitoring delay. If you did not take any ROI on theBaseline image, this quadrant will be blank.
Scan
4-55CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Smart Prep 7
� The Scan desktop screen will now look similar to the following one.
Graph Most recently��� ��!������)���
���8��6
10.�As the monitoring scans are being acquired, you can watch the linesclimbing toward the enhancement threshold on the graph in the upperleft quadrant.
+� 0� .�
�0�
TT: Threshold
Upper left quadrant
11.�When the line depicting the ROI gets close to or at the thresholdenhancement, select [Scan Phase] on the Scan Progress screen toinitiate the Scan Phase.
Note : If the scan location of Monitoring Phase does not match the startlocation of Scan Phase, the scan start will be delayed by the time the cradletakes to move. For example, it takes some four seconds for the cradle tomove 300 mm. It is highly recommended that the scan location of MonitoringPhase match the start location of Scan Phase.
Note : When the system initiates the Scan Phase, the real time calculation ofthe Smart Prep quadrant will be stopped. A screen save will be used forlater inspection.
Scan
4-56CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Smart Addition (Option) 1
Function
The Smart Addition feature allows you to prospectively add two or moreAxial/Helical images into one image. This is beneficial for brain studies sincethe added images will generate much less artifact particularly aroundposterior fossa area.
Conditions/Restrictions
The following conditions/restrictions must be met to use the Smart Additionoption.
� Scan Type : Axial or Helical (360–degree data)
� Segment and Helical Plus incompatible with this option
� Available Helical pitch : 1.0 – 3.0
� This option can be applied to group by group.
� Only contiguous images (Interval = Thickness) to be added
� Possible thickness of added image : 2, 3, 4, 5, 6, 7, 10 mm
� Possible number of images to be added : 2, 3, 4, 5, 6, 7, 10
� When Smart Addition is prescribed for [Recon 1], both [Recon 2] and[Recon 3] will be automatically designated for Smart Addition. Also, when Smart Addition is not prescribed for [Recon 1], Smart Additioncan not be prescribed for [Recon 2] nor [Recon3].
Procedure
1.�Click on the [ Thick (mm)] key in the Scan Tab menu. The Image Thickness selection menu opens (See next page).
Thick(mm)
ImageInt’val Pitch Gantry
TiltSFOV kV mA
CTDIw
Scan Timing FilmRecon
(mGy)
Thick
Scan
4-57CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Smart Addition 2
Smart Addition
Cancel1 2 3 5 7 10
Select the desired Thickness (mm)
Fig.1 Image Thickness selection menu
Note : The Smart Addition key is not available unless your CT systemhas Smart Addition option installed. Also, even if your CT system hasthe Smart Addition option installed, the Smart Addition key remains dim(inactive) unless the right Scan Type is selected.
2.�In order to activate the Smart Addition option, click on the [SmartAddition] key. Then, the Fig.2 Smart Addition Image Thicknessselection menu opens.
Smart Addition
2 3 5 6 7
Image Thickness (mm)
4 10
1 2 53
Beam Thickness (mm) Multiplication Factor
�5
OK Cancel
Fig.2 Smart Addition Image Thickness selection menu
Select the desired Thickness (mm)
On this menu select Beam Thickness (original image thickness) andImage Thickness (thickness of added images), then the MultiplicationFactor (number of images to be added) is automatically determined.
Since the Multiplication Factor is an integer, there are restrictions on thecombination of Beam Thickness and Image Thickness. For example,when 2mm is selected as Beam Thickness, Image Thickness of 3, 5,and 7 are unavailable.
The example of Fig.2 shows 1mm and 5mm are selected as BeamThickness and Image Thickness, respectively. So, the MultiplicationFactor is automatically set to 5.
Scan
4-58CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Smart Addition 3
3.�Click on [OK] to accept the selection. Then, the Fig.2 menu disappearsand the selected thickness is posted in the [ Thickness] field on theView/Edit screen like the example below.
51x5
Example
The example shows that Beam Thickness is 1mm, Multiplication Factor5 and Image Thickness 5mm.
Click on [Smart Addition] if you need to return to Fig.1 menu.
Click on [Cancel] to remove the Fig.2 menu and return to the View/Editscreen.
4.�Prescribe all the other parameters as well.
[Interval] in Smart Addition means the interval between “added” images.The possible maximum value is the image thickness of added imageand the possible minimum value is Beam Thickness.
In the example below, 1mm silces are added by three images. When the possible maximum value of 3mm is set at [Interval], addedimages are combined like image numbers of [1, 2, 3], [4, 5, 6], [7, 8, 9]and [10, 11, 12]. In case of the possible minimum value of 1mm, addedimages are combined like [1, 2, 3], [2, 3, 4], [3, 4, 5] and [4, 5, 6].
Beam Thickness : 1mm
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
Interval : 3mm
Smart Addition Example : 1mm x 3
Image number
5.�Proceed to the scan after entering all the necessary parameters.
Added images have the image annotations of ADDx (x: MultiplicationFactor) next to the slice thickness annotation.
Scan
4-59CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Smart Addition Retro Recon (Option) 1
Function
With Smart Addition option images can be retrospectively added from theraw data obtained from the usual scans.
Conditions/Restrictions
� Data obtained with 7mm or 10mm slice thickness can not be used.
� Recon Mode : Full
� Contiguous image data ( no overlap, no gap)
� Available Helical pitch : 1.0 – 3.0
� Two or more images data
When the above conditions/restrictions are not completely observed, thekeys under [ Thick (mm)] field in the Image tab are dimly displayed indicatingSmart Addition retro recon is unavailable.
RetroScanType
Y
Y
ScanStart
Location
ScanEnd
LocationRetroStart
RetroEnd
No. ofImages
ImageInterval
GantryTilt
SFOV
QuitList
Exams Confirm
Images Recon
Thick(mm)
Thick key
Scan
4-60CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Smart Addition Retro Recon 2
1.�After confirming the [Thick] key is boldly displayed (active), click on the[Thick] key to open the following Fig.1 Image Thickness selection menu.
Select the desired Image Thickness
Cancel1 2 103 4 5 6 7
Fig.1 Image Thickness selection menu
Since the number of images to be added is an integer, the slicethickness of the added image depends on the beam thickness(thickness of original image). The chart below shows beam thicknessand valid/invalid added image thickness.
1 mm 1, 2, 3, 4, 5, 6, 7, 10 mm
2 mm 2, 4, 6, 10 mm 1, 3, 5, 7 mm
3 mm 3, 6 mm 1, 2, 3, 5, 7, 10 mm
5 mm 5, 10 mm 1, 2, 3, 4, 6, 7 mm
Beam Thickness Valid Thickness Invalid Thickness
On the real menu of Fig.1 the valid thickness numbers are displayedboldly whereas the invalid thickness numbers are displayed dimlydepending on the beam thickness.
2.�Select the desired image thickness from the Fig.1 menu or click on[Cancel].
3.�After entering all necessary parameters click on [Confirm] to startretrospective reconstruction of added images.
Added images have the image annotations of ADDx (x: MultiplicationFactor) next to the slice thickness annotation.
Scan
4-61CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Cardiac Gating (Option) 1
Introduction
A primary cause of heart problems is the buildup of plaque inside thecoronary arteries. At present there is no established noninvasive means ofdetecting and measuring these deposits. Recent research has focused onthe calcium which accumulates in some of these plaque deposits. Although calcium may not be present in all plaque deposits, a significantcorrelation between volume and clinical risk seems to exist.
Unlike the plaque itself the calcification can be seen in X–ray images. With a CT scan of the entire heart, volume/density calcifications on the areasof calcification may be used to quantitatively assess the patient’s level ofcoronary risk.
A difficult aspect of calcification scoring from CT images is the motion of theheart. The images need to be acquired during a portion of the heart cyclewhen the heart is as still as possible to minimize blurring and othermotion–induced artifacts.
Theory
Cardiac Gating images are retrospectively created from the scan raw dataobtained during the R to R interval. The R to R interval is the time betweenmaximum patient heart contractions (R–wave peaks). Each R to R intervalrepresents the length of one cardiac cycle.
R–R IntervalR R
QS
QS
R–wave is used for triggering because of its strong electrical signal andcorrelation of the muscle activity of the heart.
Scan
4-62CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Cardiac Gating 2
Basic Steps
Cardiac Gating scans and subsequent CACS (Coronary Artery CalcificationScoring) analysis are implemented through the following steps.
1.�Heart Rate Display Setup
2.�ECG Setup
3.�Patient Preparation
4.�Scan Prescription and Initiation
5.�Retrospective Reconstruction for Cardiac Gating
6.�Data transfer to Advantage Workstation (AW)
7.�CACS analysis on AW (Refer to the [SmartScore] manual.)
Heart Rate Display Setup
bpm
Heart Rate Display
GE Medical Systems
Front
Dimmer
Cord
CAUTION
Do not use this display asa patient monitor to preventmisdiagnosis.
Rear
Heart Rate Display can be placed on or near the OC. This device calculatesand displays the patient’s heart rate by R–pulse from the ECG, thus thedisplayed heart rate is not the same as the ECG shows, though it is stillaccurate enough to determine a suitable scan pitch.
The heart rate is displayed in 2 or 3–digit number ranging from 15 bpm to200 bpm. When the actual heart rate is less than 15 or more than 200, itshows “0 bpm” or “blinking 200 bpm”, respectively.
After turning on the switch, there is no user operation.
Scan
4-63CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Cardiac Gating 3ECG Setup
Before initiating scans for Cardiac Gating, make sure an ECG monitor setupis complete for the system to receive trigger signals synchronized with Rpulses.
The diagram below shows the connection box and cable that connect the CTsystem and the ECG. (Refer to the ECG manual for its use.)
Connection Box
Cable BoxTo ECG
Cable
Bottom of Gantry right side (from front)
Note : If the system detects no signal from the ECG monitor for threeseconds during the scan, the system will automatically abort the scan andpost a message to inform you that you should check the connection betweenthe system and the ECG monitor.
Patient Preparation
Refer to the SmartScore manual on how to place ECG leads on the patient.Note : The patient must have the ability to hold his/her breath for 30 to 40seconds (typically), although it may be necessary to hold longer than 40seconds.
Scan
4-64CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Cardiac Gating 4
Scan Prescription and Initiation
1.�Scout scan should be done as a prerequisite for Cardiac Gating.Obtain AP and lateral scout to localize the heart.
2.�In order to activate Cardiac Gating function, first select [Create NewSeries] button at the bottom of View/Edit screen.
Name : ID :
Anatomical
Prior Next
Series Description
New Next Create
Split
Repeat�������
End
������
One
Reference
Protocol :
Auto Transfer Auto FilmSetup
ShowLocalizer
AutoStore
AddGroup Current
Group
DeleteSelectedGroup
Images ScanType
StartLoc.
EndLoc.
No. ofImages
Thick(mm)
ImageInt’val Pitch Gantry
Tilt SFOV kV mA CTDIw
Scan Timing FilmRecon
Select
Series Series More
Preview
Optimize
Exam Protocol
PriorityRecon
Rx
����������
Exam: Series:
ReconArea
(mGy)
SmartRecon
AutomA OFF
OFF
View/Edit Screen
[Create New Series]
Scan
4-65CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Cardiac Gating 5
3.�The following menu opens to provide access to [Cardiac Gating]. (Note : [Cardiac Gating] button will not be available unless your CTsystem has Cardiac Gating option installed.)
Scout Axial
Select New Series Type
ATTENTION : This will remove all unscanned groups.
OK Cancel
CardiacGating
Optional
Click on [Cardiac Gating] button, then [OK] to proceed. Or, click on [Cancel] to exit.
4.�Upon selection of [Cardiac Gating], [Scan Type] at the View/Edit screenwill automatically post [Cardiac Full 1.5sec], which indicates [Helical Full1.5sec]. Then, the scan time should be changed from 1.5sec to thefastest scan time available on the system such as 0.7sec or 0.8secdepending on your system. (Note: The system does not automatically choose the fastest scan timeas default.)
Images ScanType
StartLoc.
EndLoc.
No. ofImages
Thick(mm)
ImageInt’val Pitch Gantry
TiltSFOV kV mA
Scan Timing FilmRecon
CTDIw(mGy)
CardiacFull
0.8sec
CardiacFull
0.8 sec
ScanType
Scan
4-66CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Cardiac Gating 6
5.�Prescribe the other scan parameters as well.
Be sure to scan from base of heart to apex in one breath–hold.Scanning from superior to inferior acquires images of the base of theheart earlier in the breath–hold, increasing the likelihood of minimizingmotion effects.
The use of the following parameters is highly recommended togenerate enough images to produce one image per cardiac cycleacross the entire heart volume.
� 0.7 ( or 0.8) seconds scan time, single–breath Helical full scan
� 120 kV, 200 mA
� 3 mm slice thickness
� Large SFOV
� 25 cm DFOV
� 0.7 :1 pitch for heart rate of 60 BMP(beats per min.) Increase pitch as heart rate increases. Refer to the table/formula onthe appendix page of this manual to calculate or select pitch.
� The default kernel should be STND. An error has been made in theprotocol which selects the STND+ kernel. Be sure to change thekernel in the protocol to STND before performing Calcium Scoringexams.
Note : The image S/N ratio tends to deteriorate for big patients. This may make the scoring less accurate.
6.�Initiate the scanning.
Note : When the system does not receive signals from the ECG for 3seconds or longer, the system will automatically abort the scan anddisplay the message of “SCAN ABORTED BY NO GATING SIGNAL” atthe status area. Check the connection between the system and theECG monitor, then try to scan again.
Note : Cardiac Gating scans prospectively generate ordinary Helicalimages. However, special retrospective reconstructions must be doneto create images for CACS (Coronary Artery Calcification Scoring)purpose. Refer to the following pages for those retrospectivereconstruction procedures.
Scan
4-67CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Cardiac Gating 7
Retrospective Reconstruction for Cardiac Gating
You can prospectively obtain ordinary Helical images through Cardiac Gatingscans. However, special retrospective reconstructions must be done tocreate images for CACS (Coronary Artery Calcification Scoring) purpose.
1.�After performing Cardiac scans, proceed to Retro Recon session toobtain the necessary images.
Retro Recon
2.�At the Retro Recon parameter screen, you will see three newparameters for Cardiac Gating. They are [ Trigger Delay (%)], [ImagesPer R to R] and [ Time Between Images].
Retro ScanType
Y
ScanStart
Location
ScanEnd
LocationRetroStart
RetroEnd
No. ofImages
GantryTilt
SFOV
Images Recon
TriggerDelay
ImagesPer
R to R
TimeBetweenImages
ImagesPer
R to R
TimeBetweenImages
TriggerDelay
(%)
(%)
Scan
4-68CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Cardiac Gating 8
3.�Each of [ Trigger Delay(%)], [ Images Per R to R] and [ Time BetweenImages] has its own default figure, but you can also have choices. To select or enter other figures than default figures, click on the eachcolumn head.
� [ Trigger Delay(%)]
The [ Trigger Delay(%)] defines, in a percentage, where in the cycle themiddle image is acquired (which should be in the diastolic phase tominimize the cardiac motion.) The default is 70%. For example, if youhave requested 3 images, if the [R to R Interval] is 900 ms, and if youuse the default trigger delay of 70 %, the image 2 will be acquired at630 ms. The valid range for trigger delay is 0 to 99%. To specify thedelay, click the box and type in the percentage value (without “%”mark).
Enter the desired Trigger Delay(%).
Cancel
� [Images Per R to R]
[Images per R to R] specifies the number of images created with eachheartbeat. The default is 3, but choices are 1, 3, or 5 images. To makea selection, click the button. A box will open. In the box, click the buttonfor 1, 3, or 5. Click [Cancel] to close the box without making a selection.
Select the desired Images per R to R Interval.
Cancel1 3 5
Scan
4-69CT HiSpeed LX/i Operator ManualDir. 2295148–100Rev. 0
Cardiac Gating 9
[ Time Between Images (ms)] spaces the images in the [R to RInterval]. The default is 50 ms, but options are 50 and 100 ms. To continue the previous example, if you have specified 3 images with atrigger delay of 70% for the [R to R Interval] of 900 ms, and if youspecify 100 ms as the [ Time Between Images], the images will bereconstructed at 530 ms, 630 ms, and 730 ms.
Select the desired Time Between Images (ms).
50 100 Cancel
To make your selection, click the button. A box will open. In the box,click the button for 50 or 100. Click [Cancel] to close the box withoutmaking selections.
4.�After retrospectively reconstructing images, you need to transfer thoseimages to Advantage Workstation for CACS (Coronary ArteryCalcification Scoring) analysis. Refer to the SmartScore manual for theCACS analysis.
Scan
4-70CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Cardiac Gating 10
Appendix : Pitch Selections for Cardiac Gating
Without Overlap
Pitch = ST × bpm ÷ 60 = bpm ÷ 75
( ST : rotational scan time, bpm : beats per minute)
Table : Pitch Selection for various Heart Rate (with 2 mm slice thickness & scan time of 0.7 or 0.8 sec)
HeartRate
R–to–R Interval
Pitch
(bpm) (ms)
50
55
60
65
70
75
80
85
90
1200
1090
1000
923
857
800
750
705
666
0.6
0.6
0.7
0.8
0.8
0.9
0.9
1.0
1.0
(with 0.7sec)
0.6
0.7
0.8
0.8
0.9
1.0
1.0
1.1
1.2
(with 0.8sec)Pitch
Chapter 5
5-1CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
FILMING
AutoFilm 1Function : AutoFilm provides a wide variety of options so you can tailor thefilming to a specific need for your scan protocol or edit the existing filmingparameters as needed.
There are two pieces to the AutoFilm setup. One is setting the parametersfor the film, which can be accessed through [AutoFilm Setup] button on theView/Edit screen. The other piece is setting the parameters for the image,which can be accessed through the [Film] tab card.
Name : ID :
Anatomical
Series Description
Reference
Protocol:
Auto Transfer Auto FilmSetup
ShowLocalizer
AutoStore
AutoFilm Setup
View/Edit screen
Exam: Series:
ReconArea
Scan Timing Recon Film
AutoFilm
FrameFormat
Interval Flip Width1
Level1
MagFactor
Rotate UserAnno.
GSE
Recon 1Film Set
1
Recon 1Film Set
2
Recon 2Film Set
1
Recon 2Film Set
2
Recon 3Film Set
1
Recon 3Film Set
2
Film tab card
Filming
5-2CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
AutoFilm 2
1. �To open the Autofilm Setup page, select the [AutoFilm Setup] icon atthe top of the scan monitor while in the Axial/Helical series. Theseparameters will be set up per series or temporarily edited while in theprotocol.
Laser Camera
OK
Autofilm Setup
Destination
Format Size
Normal Slide
Copies
�
Film Direction
Top to Bottom
Auto Start Auto Print
Yes Yes
Exam Page Series Page
Yes Yes
Scout XRef–Scout
Auto Film Composer
e/s/i
Show Grayscale
Yes
Cancel
Note: Once a scan has been acquired, you can not go back to the AutofilmSetup page, unless you select a new series or a new protocol.
Description of each function is as follows;
� Format
There are 12 options for film format. Click on one of them.
� Film Direction
You can have the images filmed Top–to–Bottom or Bottom–to–Top. Click onthe blue highlighted arrow to toggle.
Filming
5-3CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
AutoFilm 3
� Destination
This determines what type of printer the images will be filmed on. Click on���buttons to select the destination (printer). The current destination isdisplayed in the message area.
� Size
You can choose to film either in the normal setting or slide setting.
� Copies
You can choose how many copies you want the camera to print. Select thenumber you want by either clicking in the box and typing in the number orclicking on ���arrows to increase or decrease the number. The validrange is from 1 to 99.
� Exam Page/Series Page
This allows you to film the Exam Page and/or Series Page. Toggle betweenYes and No. These pages will be filmed at the end of the film and will not beadded to the film until a new series or [End Exam] is selected.
� Scout
This allows you to autofilm a Scout image. Select Scout icon to open thefollowing menu.
Film Scout
Scout Series Number
Scout Number
Magnification Factor
Yes No
Window Width
Window Level
Accept
In order to autofilm a Scout image, select [Yes] first, then enter otherparameters. Lastly select [Accept].
Note: The valid range for Magnification Factor is from 0.5 to 8.0. If a Scoutimage exceeds 500mm in length, the magnification factor must be less than1.0 to view the entire Scout image.
Filming
5-4CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
AutoFilm 4
� XRef–Scout
This allows you to film a Scout image with cross–reference lines that showaxial scan locations. Select [XRef–Scout] icon to open the following menu.
Film Scout With Reference
Scout Series Number
Scout Number
Magnification Factor
Yes No
Window Width
Window Level
Accept
Image Range All First/Last
At Image Range selection, select [All] for axial images and [First/Last] for thefirst and last axial image.
Note: The valid range for Magnification Factor is from 0.5 to 8.0. If a Scoutimage exceeds 500mm in length, the magnification factor must be less than1.0 to view the entire Scout image.
After entering all parameters, select [Accept].
� Show Gray Scale
This allows you to choose whether to have the gray scale displayed on thefilm. Simply toggle between Yes and No.
Note: This selection is not available if your laser camera interface is digital.
� Auto Film Composer
This selection will determine which will be captured on the AutoFilmcomposer, an image itself or numbers of exam/series/image. Select either[Image] icon or [e/s/i] icon.
Filming
5-5CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
AutoFilm 5
� Auto Start
If you select [Auto Start] icon to Yes, the following menu will appear. You canselect whether to have your film sets automatically start filming.
AutoFilm Setup
Film Set 1
Auto Start Film Sets
NoAuto start
Auto startNew Sheet
Auto startSame Sheet
Film Set 2
Film Set 3
NoAuto start
Auto startNew Sheet
Auto startSame Sheet
NoAuto start
Auto startNew Sheet
Auto startSame Sheet
Accept Cancel
If you select [No Auto start] on the above menu, the system will notautomatically start filming. In this case the operator is to decide when to startfilming.
If Autofilm is on and the Autofilm viewport is showing, once you startscanning, you can choose from one of the three options at the bottom of theAutofilm window. The following three icons are those options.
Start New Sheet
StartNew Sheet
[Start New Sheet] will start a new sheet of film with the format that has beenselected from the Autofilm Setup page.
Continue Same Sheet
ContinueSame Sheet
[Continue Same Sheet] will continue filming on the current sheet in theAutofilm composer with the same format as the Autofilm composer.
Cancel Film Series
CancelFilm Series
[Cancel Film Series] will stop Autofilm for that series.
Filming
5-6CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
AutoFilm 6
If you select [No Autostart] with autofilm on but the autofilm viewport is notshown, the following message will appear.
Cancel
Exam: 4Series: 1
is ready for auto filming
Show FilmViewport
If you select [Show Film Viewport], the autofilm viewport will be displayed,then you can select [Start New Sheet], [Continue Same Sheet] or [CancelFilm Series].
Autofilm viewport
StartNew Sheet
ContinueSame Sheet
CancelFilm Series
If you select [Autostart New Sheet], the Autofilm process automatically uses[Start New Sheet] function.
If you select [Autostart Same Sheet], the Autofilm process automaticallyuses [Continue Same Sheet] function.
After making all the necessary selections, select [Accept] to continue orselect [Cancel] to cancel. In both cases, the system returns to the AutofilmSetup page.
� Auto Print
[Auto Print] is a toggle button between Yes and No. If you select Yes, thenthe last film of the exam will be automatically printed, whether the Autofilmcomposer is filled or not. If you select No, then you will need to select Printon the Autofilm composer.
2. �On the Autofilm Setup page, select [OK] to accept all entries. Or, select[Cancel] to cancel.
Filming
5-7CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
AutoFilm 7
The Autofilm parameters for the images can be prescribed or modified byselecting the Film Tab Card on the View/Edit screen.
1. �Click on the Film Tab Card to open the following menu for Autofilmingparameters entry. You may skip this step if you do not intend to performAutofilming.
Scan Timing Recon Film
AutoFilm
FrameFormat
Interval Flip Width1
Level1
MagFactor
Rotate UserAnno.
GSE
Recon 1Film Set
1
Recon 1Film Set
2
Recon 2Film Set
1
Recon 2Film Set
2
Recon 3Film Set
1
Recon 3Film Set
2
Film Tab Card
2. �Click on each button and select or type in a parameter.
� Auto Film
Select On or Off on the following menu after clicking on [Auto Film].
Auto Film
On Off Cancel
� Frame Format
Select one of the four formats in the following menu by clicking on it.
MID Format
Cancel
� Interval
Select one of the five choices as to which image to film.1= every image, 2=every other image, 3= every third image, and so on.
Filming
5-8CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
AutoFilm 8
� Flip
Select one of the following options. [FTB] (Flip Top to Bottom), [FTB/FLR](Flip Top to Bottom/Flip Left to Right), [FLR] (Flip Left to Right), [None]
Flip
FTB FTB/FLR FLR None Cancel
� Width 1
The first choice for window width (1� 4096)
� Level 1
The first choice for window level (–1024� 3072)
� Mag Factor
Valid image magnification factor range is from 0.5 to 4.0
� Rotate
Right 90 degrees, left 90 degrees, or 180 degrees
Rotation
None Cancel
� User Anno(tation)
Maximum four lines
User Annotation
Accept Cancel
Filming
5-9CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
AutoFilm 9
� GSE (Gray Scale Enhancement)
Select one of the four choices.
Gray Scale
G1 G2 G3 Off
Cancel
G1 : lowest contrast G2 : medium contrast G3 : highest contrastOff : no effect
The name of the grayscale will be annotated just above the tick mark.
Filming
5-10CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
AutoFilm Composer
Click on the [AutoFilm Composer] icon in the Display mode to display thefollowing Autofilm Composer. You do not necessarily have to display thiscomposer during autofilming.
You can move the Autofilm Composer on the screen by holding the cursoranywhere in the title bar area, then dragging the composer to the desiredlocation.
Auto Film
Clear
Pause Filming
Film format reflects the one prescribed on the AutoFilm Setup page.
Each port can contain either an image or a set of exam, series and imagenumber depending on the prescription on the AutoFilm Setup page.
Click on the [AutoFilm Composer] icon again to remove the composer.
Filming
5-11CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Manual Film
Function : Manual Film allows you to manually film images.
The Manual Film Composer can be initiated from several locations includingthe Exam Rx Desktop, Image Works browser, Image Works viewer andImage Works miniviewer.
[Manual FilmComposer]
Exam Rx Display menu
[Review Layouts][Autoview
Layouts]
When you perform Manual Filming, you have to use one or two viewports inthe Review Layouts, or one or two viewports in the bottom of the AutoviewLayouts.
Filming
5-12CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Manual Film Composer 1
Upon the selection of [Manual Film Composer], the following Manual FilmComposer will appear.
You can move the Manual Film Composer on the screen by holding thecursor anywhere in the title bar area, then dragging the composer to thedesired location.
Laser CameraFormats
Film Composer
Options
Clear
Current status is :
close button
� Film Composer close button
In order to close the Film Composer, click on the small button at the upperright corner of the composer, or click on the [Manual Film Composer] iconagain.
� Laser Camera
The destination of printing will be changed. If you wish to have a destinationother than a laser camera, place the cursor over the word “Laser Camera”,press and hold down the left mouse button to open a pull–down menu, thenselect a desired destination.
� Formats
Click on any one of the twelve formats. It will be reflected on the composer.
Filming
5-13CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Manual Film Composer 2
If you select the [Option] icon, the following Print optionsmenu appears.
������������
����������
��
����������������
��
����������
��
����������������
�
���
������������
��
�����������
�����
�������������
���� !�"�
#�� $�����
Select On from the Slide format pull–down menu to employ aslide format.
Note : If the selected printing machine does not support slideformat, the Slide format button is shown in gray, indicating itis not usable.
Select On from the Grayscale pull–down menu to print agrayscale onto a film.
Note : If the selected printing machine does not supportgrayscale, the Grayscale button is shown in gray, indicating itis not usable.
Select On from the Auto Printing pull–down menu toautomatically start printing immediately after all the frames arefilled.
����
Slide format:
Off
On
Grayscale:
Off
On
Auto printing:
OffOn
Filming
5-14CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Manual Film Composer 3
Select On from the Auto clear page pull–down menu toautomatically clear all the frames after the completion ofprinting.
Select either [E/S/I] or [Image] from the Icon labels pull–downmenu to put either Exam/Series/Image number or Image itself,resepctively, into the frame.
Click on the [�] or [�] button to increase or decreaserespectively the number of copies. Or, type in the desirednumber into the numerical entry field.
When filming function keys are being used, this selection willdetermine the order with which the images will be printed intothe frames of the composer.
Select [Done] to accept the entries and close the Printoptions window.
Auto clear page:
Off
On
Icon labels:
�����
Number of copies:
� � �
Expose order:
Left/RightTop/BottomRight/LeftBottom/Top
���
Filming
5-15CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Manual Film Composer 4
If the Auto clear function is off under the Print options menu,select [Clear] on the Film Composer to clear the page ofimages to start over or start a new sheet.
The message “Clear the current page” will appear. Select [OK] to clear orclick on [Cancel] to cancel.
Note : If the Auto clear page is on, the Film Composer will automaticallydisappear after printing.
If the Auto printing function is off under the Print optionsmenu, select [Print] on the Film Composer to print the currentpage of images.
The status line at the bottom of the composer will start the message“Printing...” and the images will be sent to the printing queue. Once thequeue is filled with images, the printing will start.
As the Film Composer communicates with the camera, thestatus is posted at the bottom of the composer, for example,“print queue empty”, “printing”, “film supply low”, or “outputdevice not connected”.
Note : Messages posted in green mean everything is OK. Messages inyellow are warnings. Messages in red mean a problem exists.
Note : When there exists an arrow button in the message area, you can getthe details by clicking on it.
%����
&��
Filming
5-16CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Manual Film Composer 5
Image removal from Film Composer
1. �If you wish to remove an image on the Film Composer window, click onthat image.
2. �There appears a message “Do you really want to delete this image'”
3. �Select [Yes] to delete, or select [No] not to delete.
Load images by drag and drop
1. �In order to load an image onto the Film Composer, click and hold thecursor on the image, then drag it to an image frame, then release thecursor to deposit the image.
2. �Repeat the above steps as needed.
Load images by F1 function key
Another way to load images onto the Film Composer is as follows.
1. �Move the cursor onto the desired image and press the F1 key.
2. �The image will be automatically placed into the next available frame
Note : This method may be faster than the drag & drop method, however,you can not jump the frame.
Filming
5-17CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Manual Film Composer 6
Page filming by F2 function key
You can use this function to load one sheet of images onto the FilmComposer.
Note : The Film Composer must be empty beforehand. If it is not empty, usethe [Clear] button to empty the Film Composer.
1. �Move the cursor over any one of the displayed images and press the F2key.
2. �The displayed images will be loaded onto the Film Composer in order ofthe displayed images.
Note : In order to perform page filming, the formats must be the samebetween the displayed images and the Film Composer. If the formats aredifferent, the format of the Film Composer automatically turns to that of thedisplayed images as you press the F2 key.
Note : When you are using the Viewer or Mini Viewer, you can also use the[Film Page <F2>] button to perform page filming.
MID (Multiple Image Display) filming by F3 function key
You can use this function to load the multiple image display onto one filmframe.
Note : When you perform MID filming, the image resolution will decrease alittle bit compared with the page filming.
1. �Move the cursor over any one of the displayed images and press the F3key.
2. �The displayed images will be loaded onto the Film Composer in order ofthe displayed images.
Note : When you are using the Viewer or Mini Viewer, you can also use the[Film MID <F3>] button to perform MID filming.
Filming
5-18CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Manual Film Composer 7
Series filming by F4 function key
The following menu will appear when you press the F4 key.
Format�Use Film Composer
� Viewer Format
Image Selection
1 ← → 181 18
Interval� Print all Images�
�
1/21/3
Current Print Job
No Current Job
Cancel All
Print Last Sheet No
Print Series Cancel
� Format
– Use Film Composer : Select this to use the same format as the currentFilm Composer.
– Viewer Format : Select this to change the Film Composer format tothat of Viewer Format.
� Image Selection
This determines the number of images in the series for filming.
Use the slide bar to set the number.
� Interval
Print all Images : all images
1/2 : every other image
1/3 : every three images
� Current Print Job
This area displays the list of jobs in the current queue.
You can cancel all the jobs with [Cancel All].
Filming
5-19CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Manual Film Composer 8
� Print Last Sheet
This selection determines whether the last sheet will be printed before it isfilled with images.
After completing all entries, select [Print Series] to start series filming.
Note : If you wish to cancel the filming after clicking on [Print Series], pressF4 and click on [Cancel All].
Filming
5-20CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Blank page
Chapter 6
6-1CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
QUALITY ASSURANCE
Overview
In order to assure consistent image quality over the lifetime of thediagnostic radiology equipment, users must establish and actively maintaina regular Quality Assurance (QA) program. These procedures ask you toscan a known material (usually a phantom) under a prescribed set ofconditions, and then compare your results with the predicted or optimumvalues. Because you repeat these tests frequently, if not daily, you noticechanges in image quality values before the problem becomes visible. Ifyou do notice a degradation in image quality or a change in QA values youcan schedule a site visit and let the service person or imaging physicist runmore sophisticated tests. Their early intervention could prevent a majorbreakdown.
User QA begins with baseline performance data obtained by performingthe QA tests as soon as the system meets operating system specifications.Take the first set of baseline performance data right after installation andupdate it any time the system undergoes an upgrade or a major repair thataffects image quality. An x–ray tube change is one example. Compareyour daily QA checks against these baselines. The Quality Assuranceprogram documents any change in image quality over time.
Although you can save baseline images to visually compare with your dailyQA checks, you don’t have to. The numerical data supplied during theactual testing provides the necessary objective data for comparison. Thissection contains a sheet titled QA DATA FORM that you can copy and useto record this numerical data.
Quality Assurance
6-2CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Phantom Description
Use the Quality Assurance and Performance Phantom provided with yourCT scanner to assess system performance and establish an ongoingQuality Assurance program. The phantom’s design provides maximumperformance information with minimum effort. This phantom measures sixaspects of image quality. It contains three sections, each corresponding toa single scan plane. The following illustration contains a list of the sectionsand corresponding tests.
Section 1
High Contrast ResolutionContrast ScaleSlice ThicknessPositioning Light Accuracy
Section 2
Low Contrast Detectability
Section 3
Noise and Uniformity
QA Schedule
The most effective Quality Assurance program involves obtaining basicperformance data once a day, or at least 2–3 times per week. You mustobtain data frequently and on a regular basis in order to detect anychanges in system performance that might occur before it affects clinicalimage quality. At minimum, acquire a single 10mm scan of Sections 1 and3 of the Performance Phantom each day.
Select On from the Grayscale pull–down menu to print agrayscale onto a film.
Note : If the selected printing machine does not supportgrayscale, the Grayscale button is shown in gray, indicating it isnot usable.
Grayscale:
Off
On
Quality Assurance
6-3CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Phantom Setup
Place the performance phantom on the phantom holder and level it. (Tapea small piece of cardboard or a washer to the phantom, if necessary toaccomplish this.) Position the phantom using the laser alignment lights asfollows:
1. Align the axial light to the circumferential line marking Section 1.
2. Align the coronal light to the horizontal lines on either side of the phantom.
3. Align the sagittal light (where it strikes the top of the phantom) to the verticalline on the face of the phantom.
4. Position the phantom and press the Internal Land button on the gantry.
The performance phantom contains three sections. When you correctlyfollow the positioning instructions listed above, Section 1 corresponds to0.0 mm table position, Section 2 (Low Contrast Detectability) to the 35.0mm location and Section 3 (Noise and Uniformity) corresponds to the 50.0mm location.
Circumferentialreference line Vertical reference line
Horizontal reference line
Horizontal reference line
Quality Assurance
6-4CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scan the QA Phantom
Follow the normal Single Scan protocol. Scan three locations, one for eachQA phantom section. If you set up the phantom as described on theprevious page, prescribe scan location 0.0 for section 1 of the phantom,35.0 for section 2 and 50.0 for section 3. Use the scan parameterssuggested in Table 1. You can use other parameters, but the performanceresults won’t match the data in this manual.
[NEW PATIENT] kVp – 120 CAL FOV – Small[HEAD FIRST] mA – 100 Recon FOV – 25cm
Time – 3 sec centered[HEAD] Thickness – 10mm*[SINGLE SCAN] Scan Mode – Single Recon Mode – Standard
SOFTKEY PROMPT SCAN VALUE RECONSTRUCTION
TABLE 1
*Check image thickness and positioning light accuracy by acquiring anumber of images and varying the slice thickness between scans.
Test and Analysis of the Phantom Images
Begin the analysis as soon as the Section 1 image display appears. Makecopies of the form on the following page and record the QA results there.Keep previously recorded QA results and compare them to the most recentanalysis for consistency.
Quality Assurance
6-5CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Contrast Scale
Section 1 of the phantom tests the contrast scale. CT assigns CTnumbers, also called (HU) Houndsfield Units, to the attenuation values ofX–Ray passing through a variety of material densities. The softwaremakes the attenuation visible by assigning shades of gray to groups ofnumbers you select with Window Width/ Level functions during imageDisplay. For test purposes, the CT values of water and acrylic in thephantom represent the standard against which you track your system’scontrast scale over time. The test for contrast scale follows:
1. Display a circle cursor (approximately 1 cm in diameter) from [Ellipse ROI] onthe image as shown in Figure 1. For consistency, use the same size cursorand location each time you perform this test.
2. Position the cursor on the Plexiglass block and click the left mouse button onceto calculate the ROI. Record the mean CT number on the QA Data Form.(Standard deviation record optional.)
3. Position the cursor over the water section and click the left mouse button onceto calculate the ROI. Record the mean CT number for water on the QA Dataform. (Standard deviation record optional.)
4. Subtract water’s CT number from Plexiglass’ CT number and record thedifference on the QA Data form.
FIGURE 1
Position 1cm ROI over water
Position 1 cm ROI over Plexiglas
Quality Assurance
6-6CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
High Contrast Spatial Resolution
Section 1 of the phantom contains six sets of bar patterns in a Plexiglassblock that you use to test high contrast spatial resolution. Each patternconsists of sets of equally sized bars and spaces, in the following sizes: 1.6mm, 1.3 mm, 1.0 mm, 0.8 mm, 0.6 mm, and 0.5 mm. Water fills the spacesand provides about 12% (120 HU) contrast. Examine the bar patterns todetermine the limiting resolution, defined here as the smallest bar pattern inwhich you see all five bars.
A more sensitive and quantitative method for assessing changes in systemresolution involves measuring the standard deviation of the pixel values ina single or multiple bar pattern. ROI standard deviation provides a goodindicator of system resolution and a sensitive method to detect changes insystem resolution. The recommended procedure follows:
1. If necessary, click on [Erase] to remove previous ROI data.
2. Display and position a box cursor from [Box ROI] over the largest (1.6 mm) barpattern. The cursor should fit within the bar pattern as shown in Figure 2.Adjust the size and position of the cursor as necessary.
3. Click the left mouse button once to calculate the ROI and record the standarddeviation on the QA data form.
4. (Optional) Repeat this procedure for the 1.3, 1.0, and 0.8 mm bar patterns.
FIGURE 2
Position box cursorover largest barpattern, and size ituntil it fits over thepattern.
Optional: repeatfor 1.3mm pat-tern
Optional: repeatfor 1.0mm pat-tern
Optional: repeatfor 0.8mm pat-tern
Quality Assurance
6-7CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Slice Thickness
Section 1 of the phantom also tests slice thickness. Both sides of theresolution block contain a pattern of air filled holes designed todemonstrate slice thickness. (See Figure 3.)
Air filledholes
FIGURE 3
The resolution block contains holes drilled 1 mm apart and aligned in thedirection of slice thickness (perpendicular to scan plane). Each visible holein the image represents 1 mm of beam thickness. The software assignsless negative CT numbers to partial hole images or holes located on theedge of the slice profile. To determine slice thickness, display theimage at the recommended window level and width, and count thevisible holes. Black holes in the image represent a full millimeter of slicethickness. Gray holes count as fractions of a millimeter; two equally grayholes count as a single 1 mm slice thickness.
Recommended window width: 300. Recommended window level:–100 for 3.0 mm slices, 0 for 5.0 mm, and +50 for 10.0 mm slices. Yourimage may show less detail than this example.
FIGURE 4
Adjust the window width and level, then count thelines, which represent theair filled holes.
Each black line represents one milli-meter of slice thickness. Gray linesrepresent fractions of a millimeter
Quality Assurance
6-8CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Positioning Light Accuracy (optional)
Refer to Figure 3 on the previous page: notice how the center hole in thehole patterns on both sides of the resolution block appear longer than theothers. The manufacturers drilled the center holes deeper to help youidentify them in the image. The center hole position corresponds preciselyto the black line scribed on the circumference of the phantom. When youuse an accurate Positioning light and align the phantom’s circumferentialline to the axial light, you’ll see a symmetrical hole pattern around thecenter (longer) hole in the slice thickness pattern. See Figure 5. For bestresults, use the 1.0 mm slice thickness.
FIGURE 5
Center hole positioncorresponds to black linearound circumference ofphantom
Align black line onphantom to positioninglight
Quality Assurance
6-9CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Low Contrast Detectability
Section 2 of the QA phantom tests low contrast detectability, defined hereas the smallest hole size visible for a given contrast level at a given dose.This phantom section contains a 0.75 mm thick polystyrene membranesuspended in water and pierced by a series of holes in the following sizes:10.0 mm, 7.5 mm, 5.0 mm, 3.0 mm, and 1.5 mm. The difference in CTnumbers between the water, and water plus plastic, equals the contrast inHoundsfield Units (HU). Divide the HU value by ten to obtain the contrastin percent. Measure the contrast between the plastic membrane and thesurrounding water in the following manner:
1. If necessary, click on [Erase] to remove previous ROI data.
2. Display and position a box cursor from [Box ROI] over the image. Adjust thecursor to a rectangle, approximately 1/2 cm high by 5 cm long, as shown inFigure 6.
3. First position the cursor over the polystyrene membrane above the holes.Click the left mouse button once to calculate the ROI. Record the mean CTnumber in the Low Contrast resolution box on the QA Data Form.
4. Next place the cursor in the water section above the membrane and click theleft mouse button once to calculate the ROI. Record the mean CT number.
5. Subtract the CT number of the water from the CT number of the membrane andrecord the difference.
6. Click on [Erase] to remove previous ROI data.
7. Repeat steps 3, 4, and 5. This time position the cursor below the membraneholes, then move it over the water area below the membrane.
8. Count and record the number of visible holes to determine contrast.
FIGURE 6
A. Position box cur-sor over polystyrenemembrane aboveholes, and take ROI.
B. Position box cur-sor over water abovemembrane and, takeROI.
C. Position box cur-sor over polystyrenemembrane belowholes, and take ROI.
D. Position box cursorover water below mem-brane, and take ROI.
Subtract B from A
Subtract D from C
Count visible holes
Quality Assurance
6-10CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Noise and Uniformity
Section 3 of the phantom tests noise and uniformity. Take a water–onlyscan in Section 3 to provide a uniform image by which to assess image CTnumber noise and uniformity. Enclose a region of interest, click the leftmouse button once to calculate the ROI, and the software calculates anddisplays the standard deviation or noise of the pixels inside. The softwareoften divides the HU noise values by 1000 (representing the contrast scalebetween air and water) and multiplies by 100 to convert HU to apercentage of water attenuation.
The procedure for noise and uniformity testing follows:
1. If necessary, click on [Erase] to remove previous ROI data.
2. Place a circle cursor approximately 2 cm in diameter on the center of the imageas shown in Figure 7. Adjust the size of the cursor as necessary.
3. Click the left mouse button once to calculate the ROI. Record the mean CTnumber and standard deviation on the QA Data Form.
4. (Optional) Repeat the above instructions placing the cursor at the 12 o’clockposition and once again at the 3 o’clock position.
Position circle cursor overthe center of the image,and take ROI.
Optional: Take an ROI at the12 o’clock position.
Optional: Take an ROI at the 3 o’clockposition.
FIGURE 7
Quality Assurance
6-11CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Typical Results and Allowable Variations
Because people determine clinical image quality, it remains subjective anddifficult to define. GE expects the standards of allowable variation in imagequality parameters to vary with the installation and image evaluator(s). GEencourages you to establish and follow a Quality Assurance (QA) programso you can discover any degradation of image quality before it effectsclinical images. Over time, institutions use the QA procedure to establish acorrelation between acceptable clinical image quality and acceptablevariations in the image performance indices included in the program. Thispage contains suggested allowable variations; don’t mistake them forabsolutes. Compare any parameter variation to the maximum deviationspecified in the next section called, Dose and Performance. Make sureyou used the prescribed technique, then inform service when the variationsreach the specified maximum deviation.
Contrast Scale
The difference in CT numbers between the Plexiglass resolution block andwater should equal 118, with a suggested allowable variation of 10%.
High Contrast Spatial Resolution
The standard deviation for an ROI in the 1.6 mm bar pattern should equal36 HU, with a suggested allowable variation of 20%.
Nominal Slice Thickness
Slice thickness should not vary from the expected value by more than 50%for thickness of 2.0mm or less and ± 1.0mm for thickness over 2.0mm,when evaluated according to instructions.
Low Contrast Detectability
Because this test relies upon the perceptual judgment of the personcounting visible and well–defined holes, we can’t suggest an allowablevariation. Rather, we suggest you choose a single, barely visible hole andclosely monitor that particular hole during subsequent testing fordegradation in this image parameter.
Noise and Uniformity
Mean should equal 0± 3Standard deviation should be less than 4.9
Quality Assurance
6-12CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Weighted CTDI100 (CTDIW) for MX165ZJ Tube
Explained below contains information that relates image quality to radiationdose, as required by the IEC standard, in compliance with theIEC60601–2–44, dated (1999–02). Please review this information.
Head 300 mAs (mGy)
80 kV
120 kV
140 kV
1 mm
28
70
94
2 mm
26
65
86
3 mm
28
68
91
5 mm
29
72
97
7 mm 10 mm
29 31
74 76
99 100
Body 195 mAs (mGy)
80 kV
120 kV
140 kV
1 mm
8
20
27
2 mm
7
19
26
3 mm
8
28
5 mm
8
21
30
7 mm 10 mm
9 9
22
30 31
20 23
With SmartBeam Filter
Head 300 mAs (mGy)
80 kV
120 kV
140 kV
1 mm
18
53
76
2 mm
16
47
67
3 mm
15
45
64
5 mm
15
45
64
7 mm 10 mm
15 15
45 46
64 65
Body 195 mAs (mGy)
80 kV
120 kV
140 kV
1 mm
4
16
24
2 mm
4
14
21
3 mm
4
20
5 mm
4
14
19
7 mm 10 mm
4 5
13
20 19
14 14
Quality Assurance
6-13CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Dose and Performance 1 for MX165ZJ Tube
Explained below contains information that relates image quality to radiationdose, as required by the federal government, in compliance with FederalRegulations 21CFR 1020.33(c). The dose measurement procedure isdescribed in the Code of Federal Regulations 21CFR 1020.33. The Codeof Federal Regulations can be obtained from the U.S. government printingoffice or can be viewed from the World Wide Web.
Statement of Typical Technique
HEAD BODY 25 cm Field of View (FOV) 50 cm FOV 120 kVp 120 kVp 150 mA 130 mA 2.0 sec scan time 1.5 sec scan time 10 mm slice thickness 10 mm slice thickness
Large Focus Large Focus
CT Dose Index (CTDI) For Typical Technique At Various Positions On ThePhantom Image.
POSITION HEAD BODY A 6.9 rad 1.3 rad B 7.4 rad 2.7 rad C 7.1 rad 2.5 rad D 6.9 rad 2.0 rad
E 7.2 rad 2.5 rad CTDI has no angular maximum near the surface for 360° scanning
A
B
C
D
E
Quality Assurance
6-14CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Dose and Performance 1 for MX165ZJ Tube with SmartBeamFilter
Explained below contains information that relates image quality to radiationdose, as required by the federal government, in compliance with FederalRegulations 21CFR 1020.33(c). The dose measurement procedure isdescribed in the Code of Federal Regulations 21CFR 1020.33. The Codeof Federal Regulations can be obtained from the U.S. government printingoffice or can be viewed from the World Wide Web.
Statement of Typical Technique
HEAD BODY 25 cm Field of View (FOV) 50 cm FOV 120 kVp 120 kVp 150 mA 130 mA 2.0 sec scan time 1.5 sec scan time 10 mm slice thickness 10 mm slice thickness
Large Focus Large Focus
CT Dose Index (CTDI) For Typical Technique At Various Positions On ThePhantom Image.
POSITION HEAD BODY A 4.4 rad 0.9 rad B 4.3 rad 1.6 rad C 4.2 rad 1.5 rad D 4.2 rad 1.3 rad
E 4.4 rad 1.5 rad CTDI has no angular maximum near the surface for 360° scanning
A
B
C
D
E
Quality Assurance
6-15CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
CTDI Over A Range of Techniques
Normalized to a value of 1 for typical technique and position A. (All othertechnique settings at typical value.)
POSITION HEAD BODY10 mA A 0.06 0.07350 mA A 2.33 2.700.8 sec A 0.40 0.523.0 sec A 1.52 2.017.0 mm A 0.89 0.815.0 mm A 0.75 0.623.0 mm A 0.56 0.382.0 mm A 0.44 0.311.0 mm A 0.34 0.1680 kV A 0.38 0.30140 kV A 1.39 1.48
Explained below contains information that relates image quality to radiationdose, as required by the federal government, in compliance with FederalRegulation 21CFR 1020.33(c). Please review this information.
Helical Dose For Typical Helical Technique
HEAD BODY 25 cm Field of View (FOV) 50 cm FOV 120 kVp 120 kVp 150 mA 130 mA 2.0 sec scan time 1.5 sec scan time 10 mm slice thickness 10 mm slice thickness 10 mm interval 10 mm interval
If the Helical mAs, Slice Count and Intvl selection equals the Axial mAs,Slice Count and Intvl selection, then Helical dose equals Axial dose.
Quality Assurance
6-16CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
CTDI Over A Range of Techniques with SmartBeam Filter
Normalized to a value of 1 for typical technique and position A. (All othertechnique settings at typical value.)
POSITION HEAD BODY10 mA A 0.07 0.07350 mA A 2.32 2.700.8 sec A 0.40 0.543.0 sec A 1.50 1.997.0 mm A 0.90 0.845.0 mm A 0.78 0.673.0 mm A 0.62 0.442.0 mm A 0.54 0.401.0 mm A 0.44 0.2480 kV A 0.32 0.27140 kV A 1.43 1.52
Explained below contains information that relates image quality to radiationdose, as required by the federal government, in compliance with FederalRegulation 21CFR 1020.33(c). Please review this information.
Helical Dose For Typical Helical Technique
HEAD BODY 25 cm Field of View (FOV) 50 cm FOV 120 kVp 120 kVp 150 mA 130 mA 2.0 sec scan time 1.5 sec scan time 10 mm slice thickness 10 mm slice thickness 10 mm interval 10 mm interval
If the Helical mAs, Slice Count and Intvl selection equals the Axial mAs,Slice Count and Intvl selection, then Helical dose equals Axial dose.
Quality Assurance
6-17CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Dose and Performance 2
Explained below contains information that relates image quality to radiationdose, as required by the IEC standard, in compliance with IEC60601–2–44(1999–02).
Statement of Typical Technique
HEAD BODY 25 cm Field of View (FOV) 50 cm FOV 120 kVp 120 kVp 150 mA 130 mA 2.0 sec scan time 1.5 sec scan time 10 mm slice thickness 10 mm slice thickness
Large Focus Large Focus
CTDI100 For Typical Technique At Various Positions On The Phantom Image.
POSITION HEAD BODY A 70 mGy 12 mGy B 79 mGy 28 mGy C 76 mGy 26 mGy D 73 mGy 21 mGy
E 77 mGy 26 mGy CTDI100 has no angular maximum near the surface for 360° scanning
A
B
C
D
E
Quality Assurance
6-18CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
CTDI100 Over A Range of Techniques
Normalized to a value of 1 for typical technique and position A. (All othertechnique settings at typical value.)
POSITION HEAD BODY10 mA A 0.06 0.07350 mA A 2.32 2.710.8 sec A 0.40 0.523.0 sec A 1.51 2.017.0 mm A 0.98 0.975.0 mm A 0.95 0.933.0 mm A 0.89 0.862.0 mm A 0.86 0.801.0 mm A 0.92 0.8180 kV A 0.36 0.13140 kV A 1.23 0.64
Quality Assurance
6-19CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Dose and Performance 2 with SmartBeam Filter
Explained below contains information that relates image quality to radiationdose, as required by the IEC standard, in compliance withIEC60601–2–44 (1999–02).
Statement of Typical Technique
HEAD BODY 25 cm Field of View (FOV) 50 cm FOV 120 kVp 120 kVp 150 mA 130 mA 2.0 sec scan time 1.5 sec scan time 10 mm slice thickness 10 mm slice thickness
Large Focus Large Focus
CTDI100 For Typical Technique At Various Positions On The Phantom Image.
POSITION HEAD BODY A 45 mGy 8,4 mGy B 46 mGy 17 mGy C 45 mGy 15 mGy D 45 mGy 13 mGy
E 47 mGy 16 mGy CTDI100 has no angular maximum near the surface for 360° scanning
A
B
C
D
E
Quality Assurance
6-20CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
CTDI100 Over A Range of Techniques with SmartBeam Filter
Normalized to a value of 1 for typical technique and position A. (All othertechnique settings at typical value.)
POSITION HEAD BODY10 mA A 0.07 0.08350 mA A 2.32 2.720.8 sec A 0.40 0.543.0 sec A 1.50 2.007.0 mm A 0.99 1.005.0 mm A 0.98 1.003.0 mm A 1.00 1.012.0 mm A 1.04 1.041.0 mm A 1.18 1.1980 kV A 0.31 0.13140 kV A 1.39 0.75
Quality Assurance
6-21CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Dose and Sensitivity Profile at Phantom Center
Position (mm)0
0.5
1.0Axial, Body, Center,120 kV, 130 mA,10 mm, 1.5 sec.
70 1400
Position (mm)0
0.5
1.0
70 1400
Position (mm)0
0.5
1.0
70 1400
Axial, Body, Center,120 kV, 130 mA, 7 mm, 1.5 sec.
Axial, Body, Center,120 kV, 130 mA,5 mm, 1.5 sec.
Dose Profile Sensitivity Profile
Quality Assurance
6-22CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Position (mm)0
0.5
1.0Axial, Body, Center,120 kV, 130 mA,3 mm, 1.5 sec.
70 1400
Position (mm)0
0.5
1.0
70 1400
Position (mm)0
0.5
1.0
70 1400
Axial, Body, Center,120 kV, 130 mA,2 mm, 1.5 sec.
Axial, Body, Center,120 kV, 130 mA,1mm, 1.5 sec.
Dose Profile Sensitivity Profile
Quality Assurance
6-23CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Position (mm)0
0.5
1.0Axial, Head, Center,120 kV, 150 mA,10 mm, 2.0 sec.
70 1400
Position (mm)0
0.5
1.0
70 1400
Position (mm)0
0.5
1.0
70 1400
Axial, Head, Center,120 kV, 150 mA, 7 mm, 2.0 sec.
Axial, Head, Center,120 kV, 150 mA,5 mm, 2.0 sec.
Dose Profile Sensitivity Profile
Quality Assurance
6-24CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Position (mm)0
0.5
1.0Axial, Head, Center,120 kV, 150 mA,3 mm, 2.0 sec.
70 1400
Position (mm)0
0.5
1.0
70 1400
Axial, Head, Center,120 kV, 150 mA, 2 mm, 2.0 sec.
Axial, Head, Center,120 kV, 150 mA, 1 mm, 2.0 sec.
Position (mm)0
0.5
1.0
70 1400
Dose Profile Sensitivity Profile
Quality Assurance
6-25CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Image Performance
Noise
At Typical Technique In Center Of Phantom Using Standard Algorithm
HEAD BODYσ = 0.49 % σ= 0.47 %
MTF
(same conditions as above)
100
80
60
40
20
0 2 4 6 8 10
100
80
60
40
20
0 2 4 6 8 10
HEAD BODY
Line pairs/cm Line pairs/cm
MODULATION
(%)
MODULATION
(%)
Nominal Tomographic Section Thickness HEAD BODY10.0 mm 10.0 mm
7.0 mm 7.0 mm 5.0 mm 5.0 mm 3.0 mm 3.0 mm 2.0 mm 2.0 mm 1.0 mm 1.0 mm
Sensitivity Profile
See previous pages
Quality Assurance
6-26CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Phantoms and Procedures
Dose
For best results, use the phantoms, dose profile and CTDI value calculationprocedures recommended in the CDRH final draft of “Routine ComplianceTesting for Computed Tomography X–Ray Systems” dated April 26, 1984.
Also, for best results, use the phantoms and CTDIW value calculationprocedures recommended in the IEC committee draft for vote ofIEC60601–2–44, dated August 8, 1997.
Performance
Each test uses a 25 cm water–filled acrylic phantom
Noise
Noise equals the standard deviation of an array of pixels contained in 674mm square region of interest (ROI) for Head and 2696 mm square ROI forBody. The software divides the standard deviation, expressed inHoundsfield Units, by 1000 (representing the contrast scale between airand water), then multiplies by 100 to give a value in percent.
Modulation Transfer Function (MTF)
A point spread function (PSF) image is obtained by scanning the GEperformance phantom (2100614) wire section. Software performs a twodimensional Fourier Transform on the PSF to obtain the MTF.
Slice
Use a wire ramp section of Catphan phantom, inclined 23�from the scanplane.
Sensitivity
23� from the scan plane to obtain sensitivity profiles.
Quality Assurance
6-27CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Deviations
In order to come up with “the maximum deviation,” manufacturers mustimagine every possible situation, however unlikely, that might occur withinthe entire user community. Our statements of deviation include a maximumdeviation to assure compliance with the regulation, as well as a statementof expected deviations (2σ) in the large majority of our systems.
CTDI and CTDIW Typical Techniques
The anticipated “maximum deviation” for CTDI and CTDIW equals ±40%.The expected deviation equals ±20%, except for the 40 mA or less and1mm techniques, where variation increases (up to a factor of two) due tothe inherent deviation in small values.
Dose Profile
Anticipate a “maximum deviation” of ± 30% or 2.0mm, whichever is larger,relating to dose profiles (FWHM). This value includes variability inherent inthe measurement of dose profile with TLD chips.
Performance
Noise : The noise squared (σ2) in a CT image is inversely proportional tothe x–ray dose used to make the image. The maximum deviationanticipated for image noise equals ±30%. Expected deviation equals±10%.
MTF: Expect deviations within ±10% for values on the MTF curvegenerated with data gathered according protocol. Maximum deviationsmay reach ±20% for other methodologies.
Sensitivity Profile: Expect the full width slice half maximum sensitivityprofiles to vary ±20% or 1.0 mm, whichever is larger, when measured witha wire ramp section of Catphan phantom, inclined 23�from the scan plane.If you use other methodologies, the maximum deviation may reach 1.5 mmfor all slice thicknesses, because these measurement errors have thegreatest effect on thin slices.
Quality Assurance
6-28CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Blank
Chapter 7
7-1CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
TECHNICAL SPECIFICATIONS
Component Identification for HiSpeed
Component Model Number Location of Name Plate CDRH Certified
Gantry 2200997,2200998,2200999,2201000
Rear lower center Yes
X–Ray TubeHousing
2199077–2 Housing surface Yes
X–Ray Tube In-sert
46–309300G3 Housing surface No
Collimator 2132959 Front of collimator Yes
Detector 2187725 Front left corner No
Table 2200192,2200291,2113694,2200290
Gantry side of pillar Yes
Cradle Extender 2201803 No
Operator Con-sole
2153675 Rear lower right corner Yes
PDU 2200194 Rear lower right corner Yes
Axial Headholder 2201806 No
Coronal Head-holder
2201805 No
25cm Phantom 46–237991G1 On holding block No
35cm Phantom 46–266754G1 On holding block No
42cm Phantom 46–241707P544 On holding block No
Specifications
7-2CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Component Labels
MODEL (Note 5)SERIALMANUFACTURED CLASS IVOLTS (Note 6) V~AMPS MOMENTARY (Note 7) AAMPS CONTINUOUS (Note 8) AkVA (Note 9) HZ (Note 10)
MADE FOR GENERAL ELECTRIC CO.MILWAUKEE WISCONSIN BYGE YOKOGAWA MEDICAL SYSTEMS, LTD.7–127 ASAHIGAOKA 4–CHOMEHINO–SHI, TOKYO, JAPAN
MODEL (Note 1)SERIALMANUFACTURED CLASS ISOURCE: (Note 2) A/ (Note 3) V ~(Note 4) HZSOURCE: (Note 2) A/ (Note 3) V ~(Note 4) HZSOURCE: (Note 2) A/ (Note 3) V ~(Note 4) HZSOURCE: (Note 2) A/ (Note 3) V ~(Note 4) HZ
MADE FOR GENERAL ELECTRIC CO.MILWAUKEE WISCONSIN BYGE YOKOGAWA MEDICAL SYSTEMS, LTD. 7–127 ASAHIGAOKA 4–CHOMEHINO–SHI, TOKYO, JAPAN
COMPONENT (Note 1)MODELNUMBER
(Note 2)SUPPLY CURRENT
(Note 3)SUPPLY VOLTAGE
(Note 4)SUPPLY FREQUENCY
Gantry 2200997, 2200998,2200999, 2201000
90A30A30A
15A
380–480V3
200V~115V~
115V~
50/60Hz50/60Hz50/60Hz50/60Hz
Table 2200192, 2200291,2113694, 2200290
10A 115V~ 50/60 Hz
Operator Console 2153675 20A 115V~ 50/60Hz
(SYSTEM)* HiSpeed LX/iHiSpeed FX/iHiSpeed DX/i
100A 380/400/415/440/460/480V3~
50/60 Hz
COMPONENT (Note 5)MODELNUMBER
(Note 6)SUPPLY VOLTAGE
(Note 7)SUPPLY CURRENT
(Note 8)SUPPLY CURRENT
(Note 9)
KVA
(Note 10)SUPPLY FREQUENCY
PDU 2200194 380/400/415/440/460/480V3~
100A 20A 65 KVA 50/60 Hz
Specifications
7-3CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
MX165ZJ X–Ray Tube Assembly Information
MX165ZJ Tube Unit Identification
System/Tube Catalog No.
Descriptiongraphiteanode
Casing Model Number
InsertModel Number
Insert Catalog No.
MX165 ZJ X–Ray tubeassembly
2199077–2 N.A.
Insert N.A. 46–309300G3 D3110T
Improvements in the heat handling capabilities of this unit may result in newcatalog and Model number assignments
MX165ZJ Diagnostic Source Assembly
Leakage Technique Factors
Tube Assembly Model No. 2199077 140 kV, 26 mA
Minimum Inherent Filtration (Collimator model No. 2132959, 2247199)
Minimum inherent filtration of 2.7 mm aluminum equivalent at 140 kV:� Tube Unit:
Tube Insert – 0.8 mm aluminum equivalent at 140 kVTube Housing – 0.4 mm aluminum equivalent at 140 kV
� Collimator (lower plate) – 0.5 mm aluminum
� Collimator (upper plate) – 1.0 mm aluminum
Minimum Inherent Filtration (Collimator model No. 2247203, 2222390,2247019)
Minimum inherent filtration of 6.2 mm aluminum equivalent:� Tube Unit:
Tube Insert – 0.8 mm aluminum equivalent at 140 kVTube Housing – 0.4 mm aluminum equivalent at 140 kV
� Collimator (lower plate) – 0.5 mm aluminum
� Collimator (upper plate) – 4.5 mm aluminum equivalent at 70 kV
Specifications
7-4CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
MX165ZJ CT Scan Ratings
These ratings apply to CT systems with computer controlled techniqueselection, scan mode, scan duration, and scan series. The software uses acooling algorithm to determine anode and tube unit temperatures andcooling delays. When necessary, the cooling algorithm delays scan initiationto avoid exceeding maximum anode or tube unit temperatures.
TARGET LOADING in Kilowatts FOR EACH SCAN TECHNIQUE
mA 80 KV 120kV 140 kV
60 4.8 7.2 8.4 80 6.4 9.6 11.2 100 8.0 12.0 14.0 130 10.4 15.6 18.2 160 12.8 19.2 22.4
200 16.0 24.0 28.0250
The following table lists cooling delay times (in seconds) used by thesoftware before each scan. These times assume maximum anode or tubeunit heat loading and include cycling the rotor for each scan.
Model No. MX165ZJ COOLING DELAY TIME at 80kV
80kV and 60mA S 132 132 80mA S 132 132100mA S 132 132130mA S 132 132160mA S 132 135200mA S 132 139250mA S 132 150
Scan Focus Anode Cooling Delays Technique Size 2.0 sec 3.0 sec
Specifications
7-5CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Model No. MX165ZJ COOLING DELAY TIME at 120kV
120kV and 60mA S 132 132 80mA S 132 132100mA S 132 135130mA S 132 138160mA S 135 138200mA S 249 307
Scan Focus Anode Cooling Delays Technique Size 2.0 sec 3.0 sec
Model No. MX165ZJ COOLING DELAY TIME at 140 kV
140kV and 60mA S 131 132 80mA S 132 135100mA S 133 139130mA S 133 142160mA S 191 232
Scan Focus Anode Cooling Delays Technique Size 2.0 sec 3.0 sec
Specifications
7-6CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
MX165ZJ X–Ray Tube IEC Information
X–ray Tube Housing
Type General Electric MX165 ZJ HSG
Model 2199077–2
X–ray Tube Insert Fitted
Type General Electric MX165 ZJ INSERT
Model 46–309300G3
X–ray Tube Assembly
In accordance with IEC 637/1979, the complete X–Ray tube assemblycarries two identification labels, one each for the housing and insert, markedwith the model types and numbers listed above.
X–ray Tube Insert Information
Type General Electric MX165 ZJ INSERT
Model 46–309300G3
Focal spot SE: 0.9mmW x 0.62mmL; LF: 1.2mmW x 1.2mmL
Target Material Tungsten/Rhenium Alloy Focal Track on graphite base
Target Angle 7�
Maximum Potential Difference 140 kVp
High Potential Generator : General Electric CT HiSpeed System ConstantPotential
X–Ray Tube Filament Supply: Maximum Voltage: 140kVMaximum Current : 350mA
Specifications
7-7CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
MX165ZJ Cathode Emission and Filament Characteristics
HiSpeed Emission Z1163 7� TargetLarge Focal Spot
��
���
����
��� ��� ��� ��� ��� ��� ��� �� �� ��� ��� ��� ��� ��� ��� ���
Filament Current (amps)
Tube Current (mA)
80 kVp 120 kVp 140 kVp� � �
HiSpeed EmissionZ1163 7� TargetSmall Focal Spot
��
���
����
��� ��� ��� ��� ��� ��� ��� �� �� ��� ��� ��� ��� ��� ��� ���
Filament Current (amps)
Tube Current (mA)
80 kVp 120 kVp 140 kVp� � �
Specifications
7-8CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Nominal Anode Input Power
This tube accommodates GE HiSpeed Computed Tomography Systems witha nominal anode input power of 42 kW for 3 seconds.
Maximum Anode Heat Capacity
3500kHU
Maximum Anode Heat Dissipation
820kHU/min
Specifications
7-9CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Tube MX165 ZJ Anode Heating and Cooling Curves
��
���
����
���
���
����
���
� �� �� � ��� ��� ��� �� ���
TIME (SECONDS)
HiSpeed Anode Heating Curve
35kW
10kW
20kW30kW
�
���
���
���
��
����
����
����
����
���
����
����
����
����
� � �� �� �� �� �� �� �� �� �� �� ��
TIME (MINUTES)
HiSpeed Anode Cooling Curve
Cooling and heating curves reflect maximum tube performance.System software monitors and controls tube operation.
Single Load Rating
Large Focal Spot: 42 kW for 3 secondsSmall Focal Spot: 24 kW for 3 seconds(Same as the nominal anode input power)
Serial Load Ratings
Controlled by the HiSpeed CT system operating software
Rotating Anode Supply
Designed to operate on the HiSpeed CT system (See accompanying systemdocumentation.)
NOTE�
Specifications
7-10CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
MX165ZJ X–Ray Tube Assembly Information
Labels : The X–Ray Tube Assembly carries two identification labels. Onelabel identifies the Model and serial numbers of components (X–Ray Tubeand Housing), and provides the date and location of assembly manufacture.The second label provides the name of the manufacturer. A third labelcertifies compliance with USA Federal regulation 21 CFR Sub chapter J, andlists the data and place of assembly manufacture.
Reference Axis: Normal to the window center.
Target Angle : 7�
Nominal Focal Spot Values : Small Focal Spot: 0.9 W x 0.62 L mm Large Focal Spot: 1.2 W x 1.2 L mm
Focal Spot Modulation Transfer Functions: MTF for X–Ray TubeAssembly Standard magnification = 1.3
���
���
���
���
���
���
���
��
��
���
���
��� ��� ��� ��� �� ��� ��� ��� ��� �� ��� ��� ��� ��� �� ���
MODULATION TRANSFER FUNCTIONLENGTH DIRECTION
0.9mm W x 0.62mm L FOCAL SPOTSTANDARD MAGNIFICATION 1.3
AEC 336/1982 – NEMA XR–5
SPATIAL FREQUENCY – LINE PAIRS/mm
���
���
���
���
���
���
���
��
��
���
���
��� ��� ��� ��� �� ��� ��� ��� ��� �� ��� ��� ��� ��� �� ���
MODULATION TRANSFER FUNCTIONLENGTH DIRECTION
1.2mm W x 1.2mm L FOCAL SPOTSTANDARD MAGNIFICATION 1.3
AEC 336/1982 – NEMA XR–5
SPATIAL FREQUENCY – LINE PAIRS/mm
Specifications
7-11CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
���
���
���
���
���
���
���
��
��
���
���
��� ��� ��� ��� �� ��� ��� ��� ��� �� ��� ��� ��� ��� �� ���
MODULATION TRANSFER FUNCTIONWIDTH DIRECTION
0.9mm W x 0.62mm L FOCAL SPOTSTANDARD MAGNIFICATION 1.3
AEC 336/1982 – NEMA XR–5
SPATIAL FREQUENCY – LINE PAIRS/mm
Maximum Potential Difference : 140 kVp
Inherent Filtration
1.2 mm Al at 140 kV Tube 0.8 mm AlIEC 522/1976 Housing 0.4 mm Al
Electrical Connections
See curves and diagramsEmissions CharacteristicsConnection stator, Thermal and Pressure overload switches
Principle Dimensions
Length 21.14 inches (53.7cm)Height 11.26 inches (28.6cm)Depth 29.25 inches (74.3cm)Weight 69.1 kg (± 10%)
X–Ray Tube Conditioning
The HiSpeed system software controls X–Ray tube conditioning. (Seeaccompanying system documentation.)
Maximum X–Ray Tube Assembly Heat Storage Capacity
3.5MHU
Continuous Heat Dissipation of X–Ray Tube Assembly
X–Ray Tube and Heat Exchanger 300kHU/min
Specifications
7-12CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Tube Assembly MX165ZJ Heating and Cooling Curves
�
���
����
����
����
����
� �� �� �� �� �� ��
MX165ZJ Housing Cooling Curve
Time (Minutes)
Sto
red
Hea
t (K
ilojo
ule
s)
�
���
����
���
����
� � �� �� �� �� �� �� ��
MX165ZJ Housing Heating CurvesTemperature Degrees Celsius
Time (Seconds)
6kW 5kW 4kW
3kW
2kW
1kW
Sto
red
Hea
t (K
ilojo
ule
s)Cooling and heating curves reflect maximum tube performance.System software monitors and controls tube operation.
Leakage Radiation – Loading Factors
Specified values of the loading factors determining measures for protectiveshielding of X–Ray Tube Assembly against leakage radiation according toIEC publication 407/1973 are 140 kVp – 26 mA
Classification
IEC publication 536 /1978 and IEC Publication 601–1 Class 1
Transportation and Storage
–20° to +70° C, up to 95% Relative Humidity (non–condensing)Commercial airlines accept X–Ray tube insert/housing shipments
Transport Packaging
Transport ONLY in packaging supplied by General Electric
NOTE�
Specifications
7-13CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Generator Specifications
Main Power Supply
� Line voltage (no–load) 415 or 380 or 480 VAC
� 3–phase, 50 or 60 Hz ±0.5 Hz.
� Phase–to–phase balance within 3% of lowest phase–to–phase voltage.
� Line regulation 5% or less at maximum technique factor.
� Maximum line current demand, 100 Amps RMS (50/60 Hz) at 140 kV, 300mA.
Generator Rating and Duty Cycle
� Maximum Technique: 140 kV, 300 mA
� Maximum Output Rating: 80kVA
� Generator Duty Cycle:
Maximum Technique Factor Max. Duty Cycle
140kV and 300mA 10%
kV, mA, and Timer Accuracy
Mode Condition Accuracy
Kilovoltage:Axial 120kV ± 7% of kV settingExcluding initial 5 msec
Milliamperage: Axial 10 to 350mA ± 5 % of mA setting or± 1mA (whichever is more)
Excluding initial 100 msec
Specifications
7-14CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Scan Time: Axial
Selected Scan Time (sec) Actual Scan Time (sec)
3.0 3.0 ± 10%
2.0 2.0 ± 10%
1.5 1.5 ± 10%
1.0 1.0 ± 10%
0.7(half scan) 0.7 ± 10%
kV : Subject to an additional ± 5% accuracy of instrumentation used for calibration and measurement.
mA : Subject to an additional ± 5% accuracy of instrumentation used for calibration and measurement.
Exposure : Subject to an additional ± 3% accuracy of instrumentation used
Time : for calibration and measurement.
Accuracy subject to the following conditions: 1. Line regulation 5% or less. 2. Line voltages within this specified range: 415 or 380 or 480 VAC +6% –5% (50 or 60Hz)3. System powered for at least 30 minutes prior tomeasurements. 4. Line voltage variations caused by external loads should notexceed 1% or last longer than 0.5 second while scanning.
Measurement Basis
Tube Potential : Precision voltage divider, Model No. 46–154966G1 reduceshigh voltage generated across anode and cathode by 1000:1. TubePotential equals the average kilovoltage generated during the exposure,excluding transients at the beginning and end. See Section 2 of the X–rayAlignment procedure in the Service documentation for the completemeasurement procedure.
Tube Current : Tube current equals the average milliamperage generatedduring x–ray exposure. See Section 2 of the X–Ray Alignment procedure inthe Service documentation for the complete procedure.
Scan Time: Exposure time intervals equal the time during which thekilovoltage equals or exceeds 75% of its peak value. Measure kilovoltagewith a precision voltage divider and an oscilloscope. Use the time base ofthe oscilloscope to measure exposure duration.
NOTE�
Specifications
7-15CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
The Attenuation Equivalent
Standard Head Holder : 0.3 mm aluminum equivalent at 100 kV
Shallow Head Holder : 0.35 mm aluminum equivalent at 100 kV
Coronal Head Holder : 0.8 mm aluminum equivalent at 100 kV
Cradle : 0.55 mm aluminum equivalent at 100 kV
Extender : 1.5 mm aluminum equivalent at 100 kV
CAUTION
To obtain the optimum results, make sure nothing is left in the pathof X–ray beam that may have adverse effects on examinations.
Specifications
7-16CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Periodic Maintenance by Qualified Personnel
Recommended PM frequency is 2 times a year for NP+/NP series scanners.The frequency may vary due to local ordinance and the usage of the system, thesystem availability etc. If you need more PMs, order extra copies of this document orcopy the schedules.
Priority Code
Pr (Priority) – Description
1 Safety and Regulatory
2 Image Quality
3 Procedures that make the system unavailable for scanning
4 Procedures that can be performed while the customer is scanning
Frequency Code
Fr (Frequency) – Description
S Semi–Annually (every 6 months)
A Annually (every 12 months)
Safety and Regulatory
ITEM SUB– DESCRIPTION Pr FrAnnual PM Schedule
ITEM SUBSYSTEM DESCRIPTION Pr Fr
A B
PM0101 System Verify Emergency OFF buttons (Gantry /Console / PDU)
1 S � �
PM0102 System Check x–ray ON lights or buzzer/opera-tion of scan abort
1 S � �
PM0103 System Caution Label Check 1 S � �
PM0105 System Check Error Logs 1 S � �
PM0301 Gantry Check the number of Gantry revolutions 1 S � �
PM0201 O.C Verify Audio Function 1 S � �
PM0501 Table Check Head Holder 1 S � �
PM0502 Table Check Table Cover 1 S � �
PM0503 Table Check Gap sponge 1 S � �
PM0504 Table Gap between Table Cover and Cradle 1 S � �
PM0505 Table Check Cradle Rail 1 S � �
PM0506 Table Holder Stability Check 1 S � �
PM0507 Table Check Touch Sensor operation 1 S � �
PM0508 Table Verify Unlatch Function 1 S � �
PM0509 Table Check Gantry / Table interlock 1 S � �
PM0607 XG Check Parts 1 S � �
Specifications
7-17CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Image Quality
ITEM SUB– DESCRIPTION Pr FrAnnual PM Schedule
ITEM SUBSYSTEM DESCRIPTION Pr Fr
A B
PM0104 System Image Check (Image Performance) 2 S � �
PM0302 Gantry Clean Mylar ring cover 2 S � �
PM0401 DAS Check/Clean Detector face 2 S � �
Subtotal 1:00 1:00
System Performance
ITEM SUB– DESCRIPTION Pr FrAnnual PM Schedule
ITEM SUBSYSTEM DESCRIPTION Pr Fr
A B
PM0106 System Perform Filter Curve Test 3 A 0:15PM0107 System Check Ground Cable Terminals 3 A 0:10PM0108 System Check DAS Count 3 A 0:15PM0109 System Image Performance 2 A 1:00PM0202 O.C Clean Air Filter 3 S 0:10 0:10PM0203 O.C Inspect FANs 3 S 0:10 0:10PM0204 O.C Verify Mouse for smooth operation 3 S 0:15 0:15PM0205 O.C Clean Display monitor/Console exterior 4 S 0:10 0:10PM0206 O.C Check Cables / Power Cable Terminals 3 S 0:10 0:10PM0303 Gantry Verify cables and hardware are tight 3 S 0:20 0:20PM0304 Gantry Gantry cover cleaning 4 S 0:10 0:10PM0305 Gantry Check Positioning lights 3 S 0:10 0:10PM0306 Gantry Check Gantry anchor / Gantry Isolation 3 A 0:05PM0307 Gantry Main bearing grease–up 3 A 0:20PM0308 Gantry Check drive belt for wear 3 A 0:05PM0309 Gantry Check RF Shoe Position 3 A 0:10 0:10PM0402 DAS Verify Detector Heater Control 3 S 0:05 0:05PM0403 DAS Inspect FANs 3 S 0:05 0:05PM0510 Table Clean Cradle tray / Table cover 4 S 0:15 0:15PM0511 Table Check for Oil Leak 3 S 0:05 0:05PM0512 Table Grease–up 3 A 0:20PM0513 Table Check Table anchors / Table isolation 3 A 0:05PM0514 Table Inspect FANs 3 S 0:10 0:10PM0515 Table Inspect Cradle Wire Tension 3 S 0:10 0:10PM0601 XG Check/Clean Radiator FAN 3 S 0:10 0:10PM0602 XG Check X–ray tube oil and HV connector 3 S 0:20 0:20PM0603 XG Check HV Tank and HV connector 3 S 0:20 0:20PM0604 XG Check Power cable connection 3 S 0:05 0:05PM0605 XG Check Tube over heat safety 3 S 0:15 0:15PM0606 XG Check KV and mA 3 A 1:00PM0701 PDU Inspect FANs 3 S 0:05 0:05PM0702 PDU Check Power Cable Terminals 3 A 0:05PM0703 PDU Check Power Line Voltage 3 A 0:05
Subtotal 4:30 6:55Total 5:30 7:55
Specifications
7-18CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Maintenance by CT Users
The following maintenance by CT users is highly recommended.
Periodically check if those items function properly.
� Emergency stop : Press the button to confirm the system stops.
� Cradle latch operation : Press the button to confirm the cradles latches.
� Clean covers
Specifications
7-19CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Symbols and Classification
Symbol Publication Description
~~~
3
3N
417–5032
335–1
335–1
417–5019
348
417–5008
417–5007
Alternating Current
Three–phase alternating current
Three–phase alternating current with neutral conductor
Protective earth (ground)
Attention, consult ACCOMPANYING DOCUMENTS
OFF (Power: disconnection from the mains)
ON (Power: connection to the mains)
Dangerous voltage
Direct Current
Emergency Stop
Warning sign
RADIATION of LASERAPPARATUS
Specifications
7-20CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Symbol Publication Description
417–5339
417–5009
X–Ray source assembly Emitting
Stand by
Type B Applied Part
Abort
Start
Table Set
Intercom
(On Operator Console)
Power ON
Stand by Light OFF
Light ON
Specifications
7-21CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Symbol Description
Microphone (Mic)
Contrast
Brightness
System storage prior to installation: Maintain storage temperature between –10o C and +60o
C (–14o to 140o F)
System storage prior to installation: Maintain non–condensing storage humidity below 95%
DO NOT store system longer than 90 days
System storage and shipment Maintain Air Pressure between 750 and 1060hPa
50°C
Air Pressure
700–1060hPa
Humidity 10–90%Excluding Condensation
Specifications
7-22CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Class 1 Equipment
Any permanently installed equipment containing operator or patientaccessible surfaces must provide backup protection against electric shock, incase the BASIC INSULATION fails. In addition to BASIC INSULATION,Class 1 equipment contains a direct connection to a PROTECTIVE (EARTH)CONDUCTOR which prevents shocks when a person touches a brokenpiece of equipment or touches two different equipment surfacessimultaneously.
Type B Equipment
CLASS I, II, or III EQUIPMENT or EQUIPMENT with INTERNALELECTRICAL POWER SOURCES provide an adequate degree of protectionagainst electric shock arising from (allowable) LEAKAGE CURRENTS or abreakdown in the reliability of the protective earth connection
Ordinary Equipment
Enclosed EQUIPMENT without protection against the ingress of water.
Operation Of Equipment
CONTINUOUS OPERATION WITH INTERMITTENT LOADING.
Operation in which EQUIPMENT is connected continuously to the SUPPLYMAINS. The stated permissible loading time is so short that the long termon–load operating temperature is not attained. The ensuing interval inloading is, however, not sufficiently long for cooling down to the long termno–load operating temperature.
EQUIPMENT not suitable for use in the presence of a FLAMMABLEANESTHETIC MIXTURE WITH AIR or WITH OXYGEN or NITROUS OXIDE.
Specifications
7-23CT HiSpeed LX/i Operator ManualDir. 2295148-100Rev. 0
Cleaning
The CT system is not waterproof. It is not designed to protect internalcomponents against the ingress of liquid.Decontamination or cleanliness of the CT system (i.e. gantry, table, consoleand accessories) is the sole responsibility of the health care provider owningand/or operating the CT system.
Tips for Cleaning : � Use soft cloth damped with (hot) water to remove any dirt or stains. � Even blood stains can be wiped off with damp soft cloth.
Caution :Avoid chemical damage to surfaces. Some detergent or cleaning agent maydamage the surface.
Disclaimer :GEMS bears no responsibility for sterilization of system surfaces.GEMS shall not be held liable for any contagion which may have stemmedfrom contaminated stains on the system surfaces.
Specifications
7-24CT HiSpeed LX/i Operator Manual
Dir. 2295148-100Rev. 0
Blank page
�
������������������
�����������������
���������������������
������������������������������
�������������������������������
��������������������
����!��������������� �
�����������"���
����#����$����"��
����%��&'����������"
�(���)�����������*����"�"�
+�(���������"�"�
,��-�����"�"�
.�������/������"�"
/�����'���0������"�"1
/�������������"�""
�(���)�������������.��2������"�"
����.��2������������"�"
�(���)�����������.�������������"���
3!���4������"�
3/����4������"���
3����4������"��5
3#����24������"��1
���0������"� "
��������"���
����$����������"� �
�����/(��"� 5
����������"��"
���$�����"� �
������+�&��������"� �
6�������������0������"� 5
*��*(����"� �
0������������������������"��
�������������"��"
7����%���������������"��
+�(���������"� �
,��-�����"� �
,�����8�/(��"� �
.������"� �
.��%��&��"� 1
.�������/������"� �
.��������%�-�������/���������������"� �
/����������"� "
/�������������"� �
/�������"��
������+�&.����������"� �
�9�&'�����8����"� ��"���
�9�&/�����: :���"���
�����.�����"� 5�"�5
�����/����;,�����<��"���
�������!�������"���
������������0������"� "
=��&)*�����������"���
�
�����������"�"1
�
�������.��������"
������7�),������"
*(������'������>����5
!�������"
0�����#�������5
0�����#���'�������5
7-�-�%�����5
7�������'������>����5
.���������2'�29�����5
.�����������5
#������)6�&�����"
#���/��2�?���������/��2�����5
�
6����.����#���@������� � 1
6������.���������
6�������$���7��A�����������
6�������$���'������>�����
7����$����������1
.���������2'�29�,�),$$����1
/������������'��>����1
#���)��2��/��2������
BA���,������
6������.��$���������� "
����������'�%������� 5
����������/��������������� 1
+�(�).����*��9=��&�������� 1
+�(�).���������������2���� 1
�
!7'-7+0��5��
����!�����5��
����!�������������5�"
����.������5��
�����������5�5
��������5��
6������������5��
*(��.�2�)������.�2���5��
!���6����������5�
!�������5�
�������5��
�9�&0����������5�"
��8���5��
B/�$A�������5�"
����!�������������5��
-�����!�����5���
-�����!�������������5��
!�$�������>����5���
! $�������>����5��1
!�$�������>����5��1
!"$�������>����5���
�
0���������
�������.��������
6������������$�������������
6������.��������
*���2�������������
0����������$�������������
.���������2'�29�?����9+�%����
BA���#��������$�������������
�
7��2�6������'����������
�
'���)����������1
-�����!��������������� �
-��������������� �
-7/,7;-�������7��2�/,7<����
/�����.�(������� �
+�&.�������� �5�"�"
�
,�������������������
��������)-���������
-,66��%������
����)6������-�����������
�
.������.���������2��"��
'������>���"��
.��������9������� ��
���.�������� ��
6���������� ��
6��������������� ��
*���.�������� ��
.��$�������� ���
������.�������� �1
=��&-���7�$��� �1
.������)���������=��&��������
.�������-���2������� ��
����=����/������� ���
.�������-���2������� ���
3!���4������ ���
3/����4������ ���
3����4������ ���
3#����24������ ���
C��'7#D����/�+�*�����
6������.��$�����������������"����1�����
��29���������������/�������������
7��2�.��$����������� 5
'�&��������6�����������������
+����������$�����������
.9�����6��������������
.9���������������
�����9�C�.9���������"
�����#9��>��������1
@��29����#67�����
�
/����-���2������� � 5
/����/������ � �
/�%��&'�����������
-�������7��2�6������;-76<�����
/������6�����������
�����/�$�����������1
6������+����������"
*������/,7�����5
*��������
*(��.2)������.2������
*(������-�2��$�������
!���)/����������
0���������
0�*�����"
����)�9�&0���9�������
'���)�����������5
-������6�������������
.��=��&������
/���)E��������
��������%�����
��������������������
�
������������������ � �
/�&6���!��������� � �
����������������� ��"
�����������"��
����������"�1
���$�����"��
������+�&��������"��
6��������������������"�1
*��*(����"�1
+�(���������"�1
,��-�����"��
.������"��
/�������������"��
/�������"��
����.��2������������"��
������+�&.���������"�1
������.�&��,�),$$���� �
�
#���������
�����������
!����&���9�����
����.������������
'���9����������
����>�������
#���������$�������������
�
����7����$���������
�������������
������������������������"
��������>��������"
����A��������������������
-����������