Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable...

49
Lucid 1.3.6 & 1.3.7 In-Service Presentation DC-00172 Rev D

Transcript of Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable...

Page 1: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

Lucid 1.3.6 & 1.3.7 In-Service Presentation

DC-00172 Rev D

Page 2: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

2

• The Lucid M1 System is a medical ultrasound system intended for use as an adjunct to the standard clinical practices for measuring and displaying cerebral blood flow velocity within the major conducting arteries of the head and neck. Additionally, the Lucid M1 System measures the occurrence of transient emboli signals within the blood stream.

• The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.

• Neural Analytics, Neural Analytics Logo, Lucid, and NeuralBot are trademarks of Neural Analytics, Inc. © 2019 Neural Analytics, Inc.

Indications for Use

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 3: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

3

• The Lucid M1 System is intended to be used by technologists or physicians qualified by education or training in the use of Doppler ultrasound. It is intended to be used in hospitals, clinics or physician offices by prescription or by order of a physician trained in the use and interpretation of TCD Ultrasound.

• US Federal Law restricts this device to use by or on the order of a physician.

Intended Users

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 4: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

4

• This device is not intended to be used in fetal applications.

• This device is not intended to be used inside the sterile field.

Contraindications

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 5: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

5

The Lucid M1 System is compatible with the following accessories:

• Neural Analytics 2-MHz Transducer, Handheld (NA-PRB1)

• Neural Analytics 2-MHz Monitoring Transducer 8-foot (NA-PRB2)

• Neural Analytics Monitoring Headframe (NA-MOUNT2)

• AC Power Adaptor (NA-ACC1)

• Neural Analytics Remote Control (NA-ACC2)

• Ergotron TCD Cart – Model: SV43-1310-0 (NA-ACC3)

• Mouse and Keyboard (NA-ACC6)

• External Monitor - (NA-ACC9)

Compatibility

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 6: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

Set-up

All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 7: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

7

1. Unfold the kickstand. Position in a comfortable viewing position.

2. Connect the DC power cord between the Lucid M1 System (“System”) console and the AC adapter

3. Connect the AC cordset to the AC adapter and a standard AC power outlet.

4. Plug the transducer connector/s into the front panel.

5. Plug the Remote Control into the port on the front panel.

Hardware Set-up

Transducer Plugs

Remote Plug

AC Power Outlet

Kickstand

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 8: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

8

1. Press the power button on the left side of the system to turn on system

2. Plug in keyboard and mouse into the USB ports on the side panel

Hardware Set-up

Power Button

USB Ports (x4)

Audio Output Jack

HDMI Port

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 9: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

9

• The first time the device is used, an individual designated as an administrator should logon using the following User ID and Password:

• Default username: admin

• Default password: password

• The user will then be asked to change the password. This new password will be the primary administrator’s password.

Initial Login

DC-00172 Rev D

All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 10: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

10

There are four options for the Exam Display. Select the preferred option:

1. Spectrogram on the top, M-mode on the bottom

2. M-mode on the top, spectrogram on the bottom

3. Spectrogram only

4. M-mode only

Settings – Exams – Exam Layout

• Select the M-mode preferred M-mode option:

1. Y-axis increasing in value

2. Y-axis decreasing in value

Settings – Exams – Invert M-Mode

Snapshots and screenshots can optionally be included with the Report

Display Options

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 11: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

11

Protocols can be customized for Complete and

Limited exams. Users can select the desired vessel

sequence for the exam, define the starting depth and

power for each vessel.

• Settings – Custom Protocol Admin – Add

• Choose a unique name for the protocol and

choose whether it is for a Complete or Limited

exam. Clicking Add will display a vessel option box

and clicking the down arrow will display the entire

vessel list. Select the vessel, then choose the starting

depth and power. (Default depth for all vessels is 50

mm and Power is 80% for all high-power vessels and

10% for low power vessels.)

• Continue the sequence to complete the protocol,

then click Save.

Custom Protocol

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 12: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

12

To set up communication with a PACS, go to:

• Settings – PACS

• On this screen, the Server Address, Port ID, Remote AE and Local AE can be entered.

• Check the Test Connection, to ensure a proper connection.

PACS

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 13: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

13

Materials

1. Connect the USB Isolator to one of the USB ports on the side of the system

2. Connect the USB to Ethernet adapter to the USB Isolator

3. Connect the system running the AMQP server to the Ethernet adapter; it is recommended that the system running the AMQP server is IEC 60601-1 compliant or is connected to an isolation transformer that meets IEC 60601-1

Digital Streaming

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 14: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

14

• Login as an Admin User

• Navigate to the Digital Streaming tab in the Settings screen

• Configure the Server Address and credentials• Enter the Server IP Address in the Server Address field

• Enter the login credentials for the AMQP server

• User name in the User Name field

• Password in the Password field

• Press Test Connection to test the configuration and network connection• If the system fails to connect

• review the login credentials for accuracy

• verify the network cables are connected

• verify the AMQP server is running

• Select the desired data from the Allow Transfer section; select at least one to enable the Digital Streaming feature

• Press Save to continue

Digital Streaming - Continued

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 15: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

15

Patient Search:

• The Patient Search screen allows searching for a Patient Exam Records stored on the Lucid M1 System. Patient Exam Records can be searched for by a variety of different fields:

• First Name, Last Name

• Patient ID/ MRN

• Gender

• Date of Birth

• Exams in the past (1, 7, 30 days)

Patient Information

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 16: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

16

New Patient:

• To enter information for a New Patient

• Press the New button on the Patient Search screen. The system will navigate to the Patient Setup screen.

• Enter the desired patient information.

• Note that the fields with the red “*” must be filled out. Press Save/Exit to continue.

Patient Information

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 17: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

17

Delete a Patient:

• To Delete a Patient Exam Record:

• Select the Patient and press the Delete button.

• Press Yes to confirm the deletion.

• Please note that all Exam Records associated with the Patient will be deleted, including all reports

Patient Information

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 18: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

Exam Screens

DC-00172 Rev B 18

Page 19: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

19

Unilateral Exam Screen

TICThermal index (for cranial bone) – This is an estimate of a rise in bone temperature due to acoustic power. If the TIC exceeds 2.5, the TCD will turn off Vessel and Depth

Exam TypeThis icon indicates the type of exam

Digital Streaming

Home PageThis icon returns the display to the Home Page

Patient IDThe patient name and MRN are shown here

Exam TimerThe running time of the

exam is show here

Clock Lock ScreenIn the event of an interrupted exam, this icon will lock the screen until the user returns

Battery Status

Volume

This is a 4-way toggle

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 20: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

20

Bilateral Exam Screen

User EventsWhen this icon is present, the user can add events notes during an exam

Probe Power

IndicatorThe probe icons indicate that the probe is active. The icon will be greyed out if the

probe is inactive.

Active ChannelThe “active” channel is highlighted and the TCD parameters for this channel can be adjusted

M-mode/Velocity

Trend ToggleWhen this icon is present, the M-mode display can be toggled to display velocity trend lines

DC-00172 Rev D

All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 21: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

21

M-Mode Display

M-Mode display shows the direction and intensity of the blood flow velocity in a given depth range.

Red indicates blood flow toward the insonating transducer. Blue indicates blood flow velocity away from the insonating transducer.

Brighter colors indicate greater intensity of the signal. The sample depth is displayed as a horizontal line on the M-Mode graph.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 22: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

22

Spectrogram Display

The Spectrogram display shows the Doppler spectral waveform of the blood flow

velocity at the sampled depth.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 23: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

23

Velocity Display

During diagnostic exams four velocity measurements and calculations are displayed.

During monitoring exams two displayed for each channel.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 24: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

24

Scanning Controls: Depth

• The Depth setting adjusts the

depth or distance away

from the probe (in mm) at

which the velocity

spectrogram is produced.

• The depth box on the M-

mode display can be

dragged up and down to

change the depth. From the

remote, the depth is

changed by using the depth

button.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 25: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

25

Scanning Controls: Gain

• The Gain control adjusts

the intensity of the

Spectrogram display

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 26: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

26

Scanning Controls: Baseline

• The Baseline adjustment is

used to adjust the 0 cm/s

line on the vertical axis.

• This can be moved up or

down to capture the entire

spectral waveform.

• The baseline box on the

spectrogram can be

dragged up or down to

adjust the baseline position.

• From the remote, the

baseline position is changed

by using the Baseline button

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 27: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

27

Scanning Controls: Power Adjustment

• The Power control sets the

acoustic power.

• The power is adjustable

from 0 to 100%.

• This can be adjusted from

the screen or from the

remote control.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 28: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

28

Scanning Controls: Sample Volume

• The Sample Volume adjusts

the volume of space centered

at the Depth setting where the

Spectrogram is produced

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 29: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

29

Scanning Controls: Scale

• When the Scale button is

pressed on the screen or

mouse, the vertical axis of

the Spectrogram display

can be adjusted to capture

the full velocity envelope

and spectrogram on the

display.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 30: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

30

Scanning Controls: Acoustic Power

• The Acoustic Power ON/OFF control

toggles the ultrasound energy ON or

OFF.”

• When the ultrasound energy is ON, the

spectrogram and M-mode will actively

display data.

• When the ultrasound energy is OFF, the

spectrogram and M-mode will stop

displaying data.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 31: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

31

Scanning Controls: Filter

• The Filter control adjusts the low pass

filter frequency.

• For signals that have low frequency

noise, the filter can be adjusted to

screen out that noise.

• There is no Filter button on the

remote.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 32: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

32

Scanning Controls: Sweep Speed

• The Sweep Speed is used to adjust the

horizontal axis of the Spectrogram and

M-mode display and can be used to

capture the desired Spectrogram

waveforms.

• The Sweep speed can be toggled from

4, 8 and 12 sec.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 33: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

33

Scanning Controls: Vessel Label

During an exam, the vessel label is displayed at the top of the screen and above the snapshot review box.

Use the up/down arrows on the screen or on the remote to advance to the next vessel in an exam protocol

or return to a previous label.

Vessel label list

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 34: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

34

Scanning Controls: Envelope

• The envelope icon is used to toggle

the envelope between:

+ Flow

- Flow

Auto or None

• When on Auto, the envelope position

is linked to the expected direction of

flow for each vessel

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 35: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

35

Scanning Controls: Baseline Shift

• When the Envelope is set to Auto, the spectrogram

baseline will shift slightly down for vessels where normal

flow is expected to be toward the probe and shift slightly

up when flow is expected to be away from the probe.

This provides additional room for the spectrogram

without aliasing.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 36: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

36

Scanning Controls: Screen Freeze

• To freeze the screen to enable

annotating a snapshot or taking

manual velocity measurements during

an exam, press the green freeze icon

on the screen or press and hold the

snapshot icon on the remote.

• When the envelope is on, the Mean

Velocity is calculated by averaging

the velocities from every slice of the

spectrogram. When the envelope is

turned off and manual cursors are

used to measure the Max and Min

Velocities, the Mean Velocity is

estimated using the formula:

• (Max V * .66) + (Min V * .33) = Mean

Velocity

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 37: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

37

Scanning Controls: Screenshots

• Full screen screenshots are taken every time a snapshot

is taken.

• A Screenshot taken during a unilateral exam will show

the single channel.

• A Screenshot taken during a bilateral exam with show

both channels.

• A Screenshot differs from a Screen Freeze in that a

Screenshot cannot be edited or annotated.

• Screenshots can be included in the report by checking

the “Include Screenshot Images” check box on the Exam

tab in Settings.

• When checked, Screenshots are printed two per page in

the report.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 38: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

Exams and Reports

DC-00172 Rev D 38

Page 39: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

39

Perform TCD Exam

• The Exam Procedures Screen can be accessed

directly from the Home Page by pressing “Run” or

from the Patient Records screen by pressing “New

Exam”.

• If an exam is started from the Home Page, the

patient will be assigned after the exam is

completed.

• From the Patient Records page, the patient is

assigned to the exam from the start.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 40: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

40

Exam Procedures Screen

The Exam Procedures Screen is organized by row. The top row includes

the Diagnostic Exams, the second row includes the Monitoring Exams.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 41: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

41

Diagnostic Exam: Complete

The Complete exam is intended to be used to

measure velocities in multiple vessels at multiple

depths

• A factory set protocol is pre-loaded for a

Complete exam. This protocol includes vessel

and depth sequencing for the major cerebral

arteries

• If custom protocols have been added, the

protocol list will be displayed, and the user must

select the protocol to be used

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 42: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

42

Diagnostic Exam: Limited

The Limited Exam is intended to be used to monitor daily velocity changes

in cerebral arteries to detect vasospasm

• The Limited Exam is intended to be used to monitor daily velocity

changes in cerebral arteries to detect vasospasm.

• A factory set protocol named “Limited” is pre-loaded for the Limited

Exam. The right and left MCA, the right and left ICA and the Basilar

arteries are included in this protocol. Graphs will be generated for the

vessel pairs and the Basilar Artery.

• If additional vessel pairs are desired, a custom protocol can be

generated to plot velocity changes in other vessels.

• When a protocol is selected for a patient, it should remain constant for

the duration of the monitoring.

• Snapshots should be taken in the usual manner and the software will

select the snapshot with the highest mean velocity for each vessel, for

each day.

• Each data point will be plotted daily for up to 14 days.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 43: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

43

Monitoring Exams: Exam Replay

• Green and orange circles indicate when a

snapshot was taken. Squares show when an

event was entered.

Monitoring Exams, Bubble Studies and Emboli Exams are all recorded exams that can be replayed in

post-processing.

• The snapshot icon

can be used to take

a snapshot from the

recording

• Play, Pause, Fast

Forward and Rewind

buttons enable

review of the

recorded exam.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 44: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

44

Monitoring Exams: Monitoring

The Monitoring Exam is intended to be used to monitor velocity changes in

real time.

• This exam can be performed unilaterally with a hand-held probe or

bilaterally with a headframe and (2) monitoring probes.

• In the Exam mode, the M-mode display can be toggled with velocity

trend lines by pressing the M-mode/Trend button.

• Trend lines for Mean, Min and Max velocities as well as the Pulsatility

Index (PI) can be plotted.

• Events can be included in the exam recording by clicking the User Events

icon. Each event will be displayed on the report.

• The first page of the report will show the velocity trend graphs, and

snapshots can be included on subsequent pages if desired.

• The Monitoring Exam differs from an Emboli exam in the ability to change

the PRF to optimize the scale for different velocities.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 45: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

45

Monitoring Exams: Vaso-motor Reactivity

The Vaso-Motor Reactivity (VMR) Exam is intended to measure

the cerebral vascular reserve and generate a VMR Index.

• This exam can be performed unilaterally with a hand-held

probe or bilaterally with a headframe and (2) monitoring

probes.

• The difference between a VMR exam and the monitoring

exam is that there are pre-defined events in the VMR exam.

• When this exam is selected, there are three protocol

options:

• CO2 Inhalation

• CO2 Inhalation with Hyperventilation

• Breath Holding.

• When the desired protocol is selected, the exam screen will

be displayed. Vessels should be located in the usual

manner and the exam started with the Start Exam button.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 46: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

46

• When the exam is started the first pre-defined event will appear in the event text box. When this button is clicked, a snapshot will be taken and the next event in the protocol will appear. After the last event has been captured, the display will advance to the VMR Review page. Do not press the “End Exam” button to end the exam, the study will terminate automatically.

• From the Exam Review page, each event marker can be dragged to the exact location of the event occurrence. The VMR index will be displayed

on the right side of the trend graph.

Monitoring Exams: Vaso-motor Reactivity

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 47: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

47

Monitoring Exams: Bubble Study

The Bubble Study exam is intended to be used to detect arterial micro bubbles from an intravenous injection.

Detection of bubbles in cerebral arteries may indicate a right to left vascular shunt. This exam is frequently

used to detect a Patent Foramen Ovale.

• This exam can be performed either unilaterally or bilaterally.

• After the vessel or vessels are located, select “Start Exam” to begin the recording.

• With the TCD recording running, the intravenous micro bubble injection can be given. Follow standard

“At Rest” and “With Valsalva Strain” injection/Valsalva timing protocols.

• Continue the recording for at least one minute after the injection. Snapshots of representative bubble

activity can be taken during the exam or after the exam in Exam Review.

• Click “Exit Exam” to stop the recording and advance to the Exam Review screen.

• Replay the recording to visualize any bubbles.

• Pause the recording and take a snapshot of segment with the maximum number of bubbles for the “At

Rest” and “With Valsalva” segments of the study. If only one injection is given, the software will assume it

was with Valsalva.

• Use the back icon to return to the Exam Review: Snapshots page to select and identify the snapshot(s) to

be included in the report.

• Enter the actual or estimated number of bubbles for each snapshot.

• The software will convert the bubble count to a Spencer Logarithmic Scale shunt grade.

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 48: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

48

Monitoring Exams: Emboli Study

The Emboli Exam is intended to be used to detect transient emboli in cerebral arteries.

This exam is typically performed bilaterally with a headframe.

1. After the target vessels are located and the probe position secured, click “Start Exam” to begin the

recording.

2. Auto-detection of emboli is always on in this exam and embolic activity will be captured with auto-

generated snapshots.

3. Use event markers to document events.

4. Press “Exit Exam” to stop the recording and advance to Exam Review: Snapshots.

5. Use the snapshot flyout to display the embolic events and view each embolus. If the embolic

event is determined to be artifact, use the check box on the left of the snapshot and the delete

button to delete the snapshot. Multiple snapshots can be deleted simultaneously.

6. All remaining snapshots will be included in the emboli count.

7. Use the blue check marks to identify any snapshots to be included in the report.

8. The recording can be viewed by selecting “Replay Exam”. This will display the entire recording, all

snapshots of emboli and any event markers.

9. The right, left and total emboli counts will be displayed in the report.

10. The Emboli Exam has the PRF set to 8 KHz, for optimal embolic detection

DC-00172 Rev D All steps taken from Lucid™ 1.3 System Manual (60-00118)

Page 49: Lucid 1.3.6 & 1.3.7 In-Service Presentation · 7 1. Unfold the kickstand. Position in a comfortable viewing position. 2. Connect the DC power cord between the Lucid M1 System (“System”)

DC-00172 Rev D 49

If you need help or encounter any issues with your Neural Analytics device, please reach out to our service team at:

877-638-7251 ext 733

Issue Reporting