Transcript of LPAC Compatibility of MP membranes and O 3 Membranes as O 3 contactors Materials under study...
LPAC
Compatibility of MP membranes and O3
Membranes as O3 contactors
Materials under study
Experimental Procedure
Compatible or not compatible ?
Conclusions
LPAC
Membranes for O3 contacting: Several benefits!
O3 in water treatment:
MP membranes feasible for O3 mass transfer?
production mass
transfer
water
treatment
off-gas
treatment
- Different contactor types, characteristics & performance
- Membrane contactors:• off-gas can be recycled,
• bubble-free,
• scaling,
• moderate to fast mass transfer (high kLa values)
LPAC
Compatibility of MP membranes and O3
Membranes as O3 contactors
Materials under study
Experimental Procedure
Compatible or not compatible ?
Conclusions
LPAC
Original and chemically modified membrane materials under study
MP polymeric ultra-filtration membrane tubes
Membrane test assembly:• PVC outer tube, MP membranes inside
• Water flow inside the membranes, O2/O3 flow outside the membranes
• 1 Assembly with original membrane tubes
1 Assembly with chemically modified membranes
water
O2/O3
LPAC
Compatibility of MP membranes and O3
Membranes as O3 contactors
Materials under study
Experimental Procedure
Compatible or not compatible ?
Conclusions
LPAC
Methods: Ozonation of membrane tubes
Set-up: Batch ozonation of MP membrane tubes
O2
pump
MFC
OG
OS
FC2
FC1
LPAC
Methods: Ozonation of membrane tubes
Experimental conditions:
• 5 pieces from 3 different membrane tubes
• Visual observation of damage + weigth before - after
• Milli-Q water, near-neutral pH, RT
• Ozonation: 3.5 + 7.3 + 6.0 hours, [O3]liq = 8.2–8.6 mg/L
• Monitoring of pH, ORP, conductivity, [O3]gas, [O3]liq, T,
UV/VIS
LPAC
Methods: Test assembly
Test assembly:• Milli-Q water, near-neutral pH, RT
• Water flow: peristaltic pump
• [O3]gas = ~ 50 g/m3
Set-up:
O2
pump
MFC
OG
OS
FC2
FC1
LPAC
Compatibility of MP membranes and O3
Membranes as O3 contactors
Materials under study
Experimental Procedure
Compatible or not compatible ?
Conclusions
LPAC
Ozonation of membrane tubes: clear indication of degradation
Visual observations (after ± 17 hours ozonation):• Dispersion of white particles in solution
• Increased surface roughness, loss of flexibility
Mass balance:
mass loss after ozonation periods points at degradation
UV/VIS + Conductivity:
build-up of degradation products in solution
pH:
decrease in pH caused by formation of carboxylic acids (?)
LPAC
Test assemblies: MP membranes incompatible with O3
Both test assemblies: damage after < 50 minutes O3 flow!
LPAC
Compatibility of MP membranes and O3
Membranes as O3 contactors
Materials under study
Experimental Procedure
Compatible or not compatible ?
Conclusions
LPAC
MP membranes and O3 incompatible
Nor original, neither modified MP ultra-filtration membranes are resistant to O3
MP membranes cannot be used for O3 contacting despite low [O3]liq in disinfection (typically 0.1 – 1 mg/L)
degradation = concentration + exposure !
Ozonated water streams need to be separated from filter units in the dishwashers!