LA- U R- Approved for publicrelease; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative X-ray Imager Scott C. Evans Tom N. Archuleta John A. Oertel Peter J. Walsh 13th High-Temperature Plasma Diagnostics Conference NATIONAL LAB0RAT0 RY Los Alamos NationalLaboratory,an affirmative actionhqualopportunity emplcyer, is operated by the Universityof California for the US. Department of Energy under contractW-7405-ENG-36.By acceptance of thls article, the publisher recognizes that the US. Government retains a nonexclusive, royalty-freelicense to publlshor reproduce the publishedform of this mntrlbution,or to allow others to do so, for US. Governmentpurposes. LOS Alamos NationalLaboratory requests that the publisher identify this article as work performedunder the auspices of the U.S. Department of Energy. Los Alamos National Laboratorystrongly supports academic freedom and a researcher’s right to publish; as an institution, however, tho Laboratory does not endorse the viewpoint of a publicationor guarantee ib technical correctness. Form 836 (1 0/96)

Transcript of Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release;...

Page 1: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

LA- U R- Approved for public release; distribution is unlimited.

Title

Author(s)

Submitted to,

Los Alamos

The Quantitative X-ray Imager

Scott C. Evans Tom N. Archuleta John A. Oertel Peter J. Walsh

13th High-Temperature Plasma Diagnostics Conference

NATIONAL LAB0 RAT0 RY Los Alamos National Laboratory, an affirmative actionhqual opportunity emplcyer, is operated by the University of California for the US. Department of Energy under contract W-7405-ENG-36.By acceptance of thls article, the publisher recognizes that the US. Government retains a nonexclusive, royalty-free license to publlsh or reproduce the published form of this mntrlbution, or to allow others to do so, for US. Government purposes. LOS Alamos National Laboratory requests that the publisher identify this article as work performed under the auspices of the U.S. Department of Energy. Los Alamos National Laboratory strongly supports academic freedom and a researcher’s right to publish; as an institution, however, tho Laboratory does not endorse the viewpoint of a publication or guarantee ib technical correctness.

Form 836 (1 0/96)

About This Report
This official electronic version was created by scanning the best available paper or microfiche copy of the original report at a 300 dpi resolution. Original color illustrations appear as black and white images. For additional information or comments, contact: Library Without Walls Project Los Alamos National Laboratory Research Library Los Alamos, NM 87544 Phone: (505)667-4448 E-mail: [email protected]
Page 2: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

4

Quantitative X-ray Imager

Scott C. Evans, Tom N. Archuleta, John A. Oertel, and Peter J. Walsh

Los Alamos National Lab, Plasma Physics, P-24, Los Alamos, NM 87545

ABSTRACT

We report on development of a Quantitative X-ray Imager (QXI) for the National

Inertial Confinement Fusion (ICF) and Radiation Physics (RP) programs. The QXI is

based on a typical gated X-ray imager'72 used for a variety of experiments conducted in

the ICF and RP Programs. The electronics were specified by Los Alamos National

Laboratory (LANL) and built by Grant Applied Physics (GAP) of San Francisco, CA.3

The camera system was assembled in Los P~lamos with development beginning in 1997.

The camera consists of a gated microchannd plate (MCP) that is driven by high-speed

electronic pulsers, coupled to a phosphor coated fiberoptic faceplate and recorded on

film. 'The QXI has a two-strip and a four-strip detection head and has the ability to

individually bias the gain of each recording strip. The electronic timing of the QXI was

done at the Trident4 short pulse laboratory, single strip jitter determined to be < 25 ps.

Flatfielding of the photocathode across the MCP was also done at the Trident laser.

Spatial resolution was determined to be < 5 microns. The QXI has been used on

Cylindrical Implosion work at the LLNL Nova' laser facility, and on direct-drive cylinder

mix and indirect-drive high convergence implosion experiments at UWLLE OMEGA6

laser facility. The two-strip module allows the use of point projection backlighting at

large magnifications, as part of technique development for experiments on the NIF.

Page 3: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

I. INTRODUCTION

The Los Alamos National Laboratory (LANL) Inertial Confinement Fusion (ICF)

and Radiation Physics (RP) programs study both direct and indirect drive configurations,

in both, long and short pulse experiments. The LANL program address not only laser

driven experiments, but pulse power driven experiments also. Because the program is

diverse, a versatile, weI1-calibrated instnunent was required. The Quantitative X-ray

Imager (QXI) was developed to provide unique capabilities while remaining highly

calibrated for quantitative results. We define “Quantative” as repeated and detailed

characterization of all the instrument systems. This instrument has to remain in this

characterized state while being transported to numerous experimental facilities around the

country. X-ray imagers have proven usefbl at diagnosing high-speed interactions due to

their ability to resolve broadband x-rays at high temporal and spatial resolution, typically

40-80 ps and 5 microns, respectively.

I. DESCRIPTION

The QXI is based on a pinhole aperatured gated x-ray imager used for a variety of

experiments conducted in the ICF and RP programs. The QXI consists of an imaging

fiont end, a gated microchannel plate (MCP) x-ray detector, a phosphor coated fiberoptic

faceplate coupled to film for data readout, along with high-speed electronic pulsers to

drive the x-ray detector. The QXI is the latest of LANL’s fast fiamhg cameras and

represents the state-of-the-art in this technology. See Figure 1.

The QXI has a 2 strip and a 4 strip detection head and the ability to individually

bias each of the strips independently that is limited to a +/-540V difference between

Page 4: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

adjacent strips, This gives the QXI the unique ability to add or subtract gain to adjacent

strips. This is useful for enhancing weak or suppressing strong images caused timing,

MCP ohmic losses, or photocathode quantum efficiency (QE). This instrument was

designed to fit into a standard size “cart” arid electromechanically interface to diagnostic

insertion mechanisms at various laser facilities. This facilitates its use at the Nova (at

LLNL), OMEGA (at the University of Rochester Laboratory for Laser Energetics) and

Trident (here at LANL) laser facilities in six (SIM) or ten inch (TIM) manipulators. It

has 2X, 4X 8X and 12X magnification capabilities.

111. CHARACTERISTICS

The detection head contains a 40mrn diameter MCP having with micropore length

over diameter ratio of 40 (L/D =40). A micro stripline is then coated over the MCP,

typically 50A of Cr overlaid by 5000A of Cu, followed by lOOOA Au. A micro strip of

l5mm in width yields 6.5 ohms for the 2-strip head, and 6mm in width yields 12 ohms

for the 4-strip head. The MCP is proximity focused to a P-1 1 (blue) phosphor coated

fiberoptic screen that converts the pulsed electron cloud output to visible light that is

recorded on panchromatic 35mm reverse rolled film. This is then digitized by a

microdensitometer to give a film density. The electronics package is composed of a high

speed, high voltage pulse applied to drive the intensified unit. The pulser unit puts out a

raw unshaped pulse that is about -6kV into 50 ohms. This output is used feed to pulse

forming networks that shape the electrical gating pulse. Available pulse widths are

1 Oops, 200ps 500ps 1 ns and 2ns. The pulse is 3.5 to 4keV in amplitude, depending on

the pulse forming network chosen. This high-amplitude, high-speed voltage pulse acts to

Page 5: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

shutter the x-rays as the pulse propagates along the micro stripline. A DC bias can be

injected between the pulser and the head via blocking bias tees, (a 220pF in capacitor and

a 1 Mohm resistor to ground) which allows ;application of DC bias without reverse

biasing the pulse generator. An individually controlled bias supply enables each strip to

be DC bias +/- 1kV in 50 V steps, with a maximum +/-540V between adjacent strips, and

is set via switches on top of the supply. A Zener diode stack limits the voltage on any

two outputs to +/-540v, no matter what the bias setting on the power supply. This

prevents accidental high adjacent voltages, which could damage the detector. The delay

unit provides fiom 0 to 12.7 ns for each channel and is switchable in loops steps. Each

delay setting was set and measured on a Tektronics CSA803 sampling oscilloscope

(50pddiv) and delays were measured to be less than 25 ps. Rotary switches are located on

the top of the delay unit for setting the c0rrt:sponding delay times. Absolute trigger

timing of the individual pulser units was done at the Trident short pulse lab, using 7 ps

21 lnm pulse. This allows us to set the sycronisity of the four strips when all strip delays

are set to zero. Figure 2 demonstrates the sjmnchronous timing of all four channels when

timed to turn on at the same time. Strip to strip timing is better than 25 ps. Single strip

jitter was observed at Trident also and determined to be +/-15ps. Flatfielding test were

done at Trident, using the main 2 a (532 mi) beam at 150 joules to illuminate an Au disk

with a Ins square pulse to generate a flat top x-ray pulse. The QXI is then timed to

record in the middle of this 1 ns pulse. ‘This shows the photocathode response as a

function of position on the MCP. See Figure 3. Spatial resolution was observed under th+

same laser conditions, using a backlit Au grid, having a 25um period (6 um bar and 19

um spacing). The camera can easily resolve the 6 urn bars using a 10 um pinhole at 12X

Page 6: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

magnification. Calculations indicate that resolutions of less that 5 microns are achievable

with 5 micron pinholes.’ See Figure 4

In order to maintain its calibrative state, the QXI is tested both before and after

experimental shot campaigns. The pulser units are measured for both peak voltage height

and relative timing between strips. MCP current draw is measured in picoAmps as a

function of DC bias voltage, which can indicate possible fiacture or loose particulate

damage that can occur in transport. The QXI is also flatfielded a DC x-ray source where

pre and post campaign images are compared. A flatfield is also done if there has been a

photocathode or phosphor screen change. A11 or one of these has demonstrated to affect

the instruments sensitivity and resolution.

IV CONCLUSIONS

The need for a highly calibrated x-ray imager that is transportable to

multiple facilities countrywide is apparent. QXI currently fits that bill and has

been used at Trident in Los Alamos, NM, Nova laser facility, in Livermore, CA

and OMEGA laser facility, in Rochester, NY. Transporting under various

mechanical and varying humidities, while keeping it calibrated represents quite a

challenge. The versatility of the QXI along with the multi-head configurations

has proven this camera a good choice for x-ray imaging. Work is continuing on

further development of this camera,,

Page 7: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

11. ACKNOWLEDGEMENTS

The authors would like to express our appreciation for the support of the LANL

Diagnostics Engineering Team, the Trident, NOVA and OMEGA laser facilities and

all their technical support teams. This work was supported to the U. S. DOE under

contract No. W7405-ENG-36.

‘J.A.Oerte1, T. Archuleta, S. Evans J. Jimerson, T. Sedillo and R.G. Watt, “GatedX-ray

Images of Nova Hohlraums, ” Defense Research Review Journal, 6, Jan-July (1 994),

UCRL #53880-6-1&2.

20.L. Landen, P.M. Bell, J.A. Oertel, J.J. Satariano, and K. K. Bradley, Gain Uniformity,

saturation and depletion in micro-channel plate x-ray fkaming cameras, SPIE Proceedings

in Ultra High-speed Photography, Videography and Photonics, 1993 (unpublished), Vol.

2002

3Grant Applied Physics, 101 Lombard Street #911 W San Francisco, Ca

41nformation on the Trident laser facility can be found at

htttx//harry. lanl. ~ovrPlasmaPhysics/trident.~

’Nova laser facility, Lawrence Livermore National Laboratory, Livermore Ca.

‘Information on the OMEGA laser facility can be found at http://www.Ile.rochester.edu/

’G.O. Reynolds and J. H. Ward, SPIE 3 (October/November 1966)

Page 8: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative
Page 9: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative
Page 10: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

Figure 3a

Flatfklcl of 4 strip xnodtrlc acquired at Trident Laser Facility

Page 11: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

Figure 3b Flatfield data on the 2 strip module acquired Trident Laser Facility

Page 12: Los Alamos - UNT Digital Library/67531/metadc721196/... · LA- U R- Approved for public release; distribution is unlimited. Title Author(s) Submitted to, Los Alamos The Quantitative

Figure 4

Au resolution Grid with a 25urn period(6urnbars, 19um spacing)