LOD Determination With Noiseless Detectors
Transcript of LOD Determination With Noiseless Detectors
LOD Determination With Noiseless Detectors
1
Dr. Michael LouisJordi Labs
April 2, 2020
Limit of Detection
Introduction of noiseless detectors
Methods for analysis of LOD
Most practical methods for noiseless detection
Outline
2
Limits of Detection
3
• Limits of detection and quantitation are critical for analytical instrumentation and validation of analytical methods
• Necessary in determining toxicological thresholds for compounds of interest across a wide variety of industries including pharmaceutical, medical devices, food contact materials, as well as other consumer products
Detection and Quantitation Limits
4
Limit of Detection (LOD)• USP/International Conference on Harmonization (ICH)-Lowest amount of
analyte in the sample, which can be detected but not necessarily quantitated under stated experimental conditions
• US FDA-Not explicitly described
Limit of Quantitation (LOQ)• USP/International Conference on Harmonization (ICH)-Lowest amount of
analyte in the sample, which can be quantitatively determined with suitable precision and accuracy
• US FDA-The lowest amount of analyte that can be quantitatively determined with suitable precision and accuracy also called LLOQ (Lower Limit of Quantification)
Detection and Quantitation Limits
5
Limit of Detection/Quantitation (LOD/LOQ)• By visual evaluation• Based on S/N ratio (LOD-
3:1/LOQ-10:1)• Based on standard
deviation of response and slope (LOD-3.3σ/s-LOQ-10σ/s)
• Traditionally use of S/N has been used to determine LOD/LOQ
Triple Quadrupole (QqQ)
6
• First QqQ developed in late 1970’s
• Significantly improved detection limits for Gas Chromatography (GC), Liquid Chromatography (LC) and Inductively Coupled Plasma (ICP) analysis
• Use of tandem quads, allows for significant noise reduction increasing sensitivity of target analyte
• Little to no noise, required additional means for evaluating LOD
LCMS/GCMS QqQ
7
• LC-QqQ and GC-QqQ can both operate using multiple reaction monitoring (MRM) where parent ions are isolated in quad one then are further fragmented in the collision cell and product ions are selected in quad two significantly reducing background noise
Ionization
SourceQ1 Q2
Collision
CellDetector
ICP-MS QqQ
8
• ICP-MS QqQ utilizes tandem quadrupoles to isolate different elements which may have the same m/z increasing sensitivity for element of interest
• Examples include removal of 204Hg+
interference with 204Pb+ or trace detection of 129I
Ionization Source Q1 Q2Collision Cell Detector
Filter Desired Ion
Reactive Gas in
Collision Cell
Filter Desired Ion
Limit of Detection
9
• Analysis of a number of blanks will generate a distribution where Lc can be determined as a level above the blank
• Alpha represents the probability of a false positive• At Lc the chances of a false negative are 50%
https://www.intechopen.com/books/advances-in-gas-chromatography/limit-of-detection-and-limit-of-quantification-determination-in-gas-chromatography
Limit of Detection
10
• Limit of detection must be increased to a level where the risk of a false negative is reduced to an acceptable level
• LOD will be higher than Lc based on the standard deviation of the distribution and acceptable amount of error
• This is where S/N of 2-3 times noise arises
https://www.intechopen.com/books/advances-in-gas-chromatography/limit-of-detection-and-limit-of-quantification-determination-in-gas-chromatography
Signal-to-Noise
11
• Used for methods which produce back ground noise
• Baseline 20X the peak width at half height is analyzed
• S/N=2H/h• LOD=3:1• LOQ=10:1
h
H
S/N=2H/h
Blank Determination
12
• Used when blank gives non-zero standard deviation (noise)• Overlap of distributions of LOD and Blank can result in both false positives and false
negatives• LOD=Xb+3Sb
• LOQ=Xb+10Sb
Noise in Noiseless Detectors
13
• Decreased background can result in more varied noise regions
• Utilization of traditional signal-to-noise approach wont work for MRM acquisition as highly variable baseline noise results in highly variable LOD and LOQ determination
• Alternative methods for LOD and LOQ determination are necessary
Red-10 ppm in Scan
Blue-100 ppb in MRM
Baseline Noise in MRM
14
• In noiseless detectors selection of background noise region becomes extremely critical• Different locations can give highly variable amounts of background noise resulting in
inconsistent signal-to-noise ratios for the same analyte
Noise Region (a) (b) (c)
Time (min) 3-4 4.5-5.5 7.5-8.5
Signal to Noise 126 583 3594
Linear Regression
15
• Most applicable to methods where background noise is absent• LOD=3Sa/b• LOQ=10Sa/b• Standard deviation of the slope can be calculated from y-residuals
Instrument Detection Limit Testing Level
16
• Repeated analysis of a low concentration of analyte allows for statistical determination of LOD of instrument
• Critical that concentration analyzed is not too high or not too low
• Optimal %RSD 8-20%
7
9
11
13
15
17
19
0 10 20 30 40 50
% R
SD
Concentration (ppb)
Instrument Detection Limit
17
• Repeated analysis of a low concentration of analyte allows for statistical determination of LOD of instrument
• Critical that concentration analyzed is not too high or not too low
• Optimal %RSD 8-20%
7
9
11
13
15
17
19
0 10 20 30 40 50
% R
SD
Concentration (ppb)
Instrument Detection Limit
18
• Analysis of 7+ injections of the same concentration at predetermined level (8%-20% RSD) allows for instrument detection limit calculation to be performed
• IDL=t-value(99% CI)*RSD*Concentration/100
8 injections-99% CI-2.896
RSD-8.83%
𝐼𝐷𝐿 =2.896 ∗8.83%∗1 𝑝𝑝𝑏
100=256 ppt
Effect of Standard Deviation on IDL
19
• Standard deviation of analyzed concentration directly correlated to IDL• Smaller standard deviation given the same alpha and instrument sensitivity results
in a lower IDL
Mean Standard Response
Instrument Detection Limit
Quantity of Analyte
Sig
nal R
esponse
Mean Standard Response
Instrument Detection Limit
Quantity of Analyte
Sig
nal R
esponse
Method Detection Limit
20
• According to US EPA-The method detection limit (MDL) is defined as the minimum concentration of a substance that can be measured and reported with 99% confidence that the analyte concentration is greater than zero and is determined from analysis of a sample in a given matrix containing the analyte.
• Due to the possibility of matrix suppression or enhancement the method detection limit (MDL) is a more appropriate measure of LOD than using IDL
• Analyzing 7+ spiked samples, the MDL is calculated in the same manner as the IDL• Practical Quantitation Limit (PQL) is 5-10 times the MDL
Jordi Labs Analysis Using Noiseless Detection
21
• https://jordilabs.com/knb/case-studies/medical-grade-polymer/
• https://jordilabs.com/knb/case-studies/detecting-aflatoxins-medical-device-el/
In Summary
22
• No noise detectors require different methods to evaluate detection limits due to inconsistent or nonexistent baseline noise
• Statistical evaluation or linear regression approaches are most appropriate in noiseless detection modes
• USP, ICH, and FDA guidelines serve to inform determination of detection limit in noiseless detectors
23
Thank You!