STANDARD-TECH Report No.: GZE170372-H-B Laboratory: Standard-Tech Co. Ltd Testing Center NVLAP CODE: 201011-0 Report Format Number STD/QR4909-A/2 AddressStandard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com 1 / 16 LM-79-08 Test Report For GREEN LOGIC LED ELECTRICAL SUPPLY INC (Brand Name: GLLUSA) ShenFuBao Industry Park,Bonded area,Futian District,Shenzhen,China Fuel Pump Canopy Luminaires Model name(s):33-CP1-760-8XX Representative (Tested) Model: 33-CP1-760-840 33-CP1-760-857 Model Difference: All construction and rating are the same, except CCT Test & Report By: Review By: Sevon Li Engineer: Sevon Li Manager: Tommy Liang Date:Apr.28,2017 Note: This report does not imply product certification, approval, or endorsement by NVLAP, NIST, or any agency of the Federal Government.

Transcript of LM-79-08 Test Report · 2018-04-16 · STANDARD-TECH Report No.: GZE170372-H-B Laboratory:...

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

1 / 16

LM-79-08 Test Report

For

GREEN LOGIC LED ELECTRICAL SUPPLY INC (Brand Name: GLLUSA)

ShenFuBao Industry Park,Bonded area,Futian District,Shenzhen,China

Fuel Pump Canopy Luminaires

Model name(s):33-CP1-760-8XX

Representative (Tested) Model: 33-CP1-760-84033-CP1-760-857

Model Difference: All construction and rating are the same, except CCT

Test & Report By: Review By:

Sevon Li Engineer: Sevon Li Manager: Tommy Liang

Date:Apr.28,2017

Note: This report does not imply product certification, approval, or endorsement by NVLAP, NIST, or

any agency of the Federal Government.

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

2 / 16

1.1 Product Information:

Organization Name GREEN LOGIC LED ELECTRICAL SUPPLY INC

Brand Name GLLUSAModel Number 33-CP1-760-8XXSKU (if available) --

Type of Luminaire

(for integral lamps, list base type and lamp type) Fuel Pump Canopy Luminaires

Rated Voltage / Frequency 120-277Vac, 50/60 Hz

Nominal Power 60W

Rated Initial Lamp Lumen --

Declared CCT 4000K,4500K,5000K,5700K

LED Manufacturer ShenZhen JuFei Optoelectronics Co., Ltd.

LED Model 2835 White SMD LED

Sample Number GZE170372-H-B1(4000K),B2(5700K)

Luminaire Aperture (for downlights) -- in.

Luminaire Length -- mm

Luminaires Width -- mm

Number of Units (modular products) N/A s

Photo

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

3 / 16

1.2 Test Specifications:

Date of Receipt Apr.10,2017

Date of Test Apr.20,2017

Test item

1. Total Luminous Flux

2. Luminous Distribution Intensity

3. Luminous Efficacy

4. Correlated Color Temperature

5. Color Rendering Index

6. Chromaticity Coordinate

7. Electrical Parameters

Reference Standard

1. IES LM-79-2008 Electrical and Photometric Measurements of

Solid-State Lighting Products

2. ANSI C78.377-2008 Specifications for the Chromaticity of Solid

State Lighting Products

3. CIE 13.3-1995 Method of Measuring and Specifying Colour

Rendering Properties of Light Sources

4. CIE 15-2004 Technical Report Colorimetry

5. IESNA LM-16-93 Practical Guide to Colorimetry of Light Source

6. IESNA TM-16-05 Technical Memorandum on Light Emitting

Diode (LED) Sources and Systems

Reference Work Instruction QD25

1.3 Test Methods

1) Photometric and Light Distribution Measurement – Goniophotometer Method:

Photometric parameters were measured using the goniophotometer and software. The ambient

temperature shall be maintained at 25° C ± 1° C, measured at a point not more than 1 m from the

sample and at the same height as the sample. The sample was operated at 120 or rated Volts AC, 60Hz.

It was stabilized before measurement was made. Luminous flux, luminaire efficacy, zonal lumen were

calculated from the software taken at 1° vertical intervals and 22.5° horizontal intervals.

2) Chromaticity Measurement – Sphere-Spectroradiometer Method:

Chromaticity parameters were measured using an integrating sphere, a spectroradiometer and software.

The ambient temperature condition inside the sphere was maintained at 25º C ± 1º C. The sample

measurements were made using a spectroradiometer connected by a fiber optic cable and detector

through the detector port of the integrating sphere. The sample was operated at 120 or rated Volts AC,

60Hz. It was stabilized before measurement was made. Chromaticity coordinates, correlated color

temperature and color rendering index were calculated from the spectral power distribution taken at 5

nm intervals over the range of 380 to 780 nm.

3) Electrical Measurements:

Electrical parameters were measured using power meters incorporated in goniophotometer or

sphere-spectroradiometer system. The ambient temperature surrounding the sample was maintained at

25º C ± 1º C. The sample was operated at 120 or rated Volts AC, 60Hz. It was stabilized before

measurement was made. Voltage, frequency, current, power, power factor and total harmonic distortion

were measured by and read from the power meter.

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

4 / 16

2. Electrical, Photometric and Chromaticity Measurements

(Refer to Work Instruction QD25)

Test date 2017-04-20 Test Ambient: 25.2 º C

Test Orientation As intended Stabilization Time (min) 90

Model Number 33-CP1-760-840

Electrical Measurement:

Sample No. Voltage (Vac) Frequency (Hz ) Current (A) Power (W) Power Factor THD %

GZE170372-

H-B1

120.0 60 0.4638 55.57 0.9985 3.50

277.0 60 0.2074 54.80 0.9537 5.92

DLC Pass Criteria >= 0.9(-3%) <= 20(+5)

Chromaticity Measurement - Sphere-Spectroradiometer Method:

Parameter Result Special Color Rendering Indices

Test Voltage (V) 120.0 R1 74 R9 0

Frequency (Hz) 60 R2 82 R10 55

CCT (K) 3959 R3 87 R11 71

Duv 0.0012 R4 75 R12 44

Chromaticity (x, y) x=0.3831 y=0.3810 R5 73 R13 75

Chromaticity (u’, v’) u'=0.2252 v'=0.5038 R6 73 R14 92

Color Rendering Index (CRI) 75.9 R7 84 R15 69

R9 0 R8 59 -- --

Photometric Measurement – Goniophotometer Method:

Parameter Result DLC V4.0 Pass Criteria

Test Voltage (V) 120.0 277.0 --

Frequency (Hz) 60 60

Total Luminous (lm) 6592.7 6502.9 >=2000(-10%)

Luminous Efficacy (lm/W) 118.64 118.67 Standard: >=

95(-3%)

Premium: >=

115(-3%) Most Worst Luminous/Highest Watts 117.02

Zonal lumens in the 0-40° zone (%) 44 -- ≥40%(-3%)

Zonal lumens in the 40-70° zone (%) 43.9 -- ≥40%(-3%)

Beam Angle (°) 106.2 -- --

Center Beam Candle Power (cd) 2360 -- --

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

5 / 16

Zonal Lumen Tabulation

Zonal Lumen Summary

Zone Lumens % Luminaire

0-30 1,796.2 27.2%

0-40 2,903.4 44%

0-60 5,025.0 76.2%

60-90 1,451.8 22%

70-100 705.7 10.7%

90-120 95.8 1.5%

0-90 6,476.8 98.3%

90-180 115.2 1.7%

0-180 6,592.1 100%

Lumens Per Zone

Zone Lumens % Total Zone Lumens % Total

0-10 222.9 3.4% 90-100 27.7 0.4%

10-20 632.4 9.6% 100-110 40.9 0.6%

20-30 941.0 14.3% 110-120 27.1 0.4%

30-40 1,107.2 16.8% 120-130 13.0 0.2%

40-50 1,121.2 17.0% 130-140 3.6 0.1%

50-60 1,000.4 15.2% 140-150 1.2 0%

60-70 773.9 11.7% 150-160 0.9 0%

70-80 478.9 7.3% 160-170 0.6 0%

80-90 199.0 3.0% 170-180 0.2 0%

Spectral Power Distribution & Chromaticity Diagram

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

6 / 16

Photometric Data

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

7 / 16

Candela Table - Type C

0 22.5 45 67.5 90 112.5 135 157.5 180 202.5 225 247.5 270 292.5 315 337.5 360

0 2360 2360 2360 2360 2360 2360 2360 2360 2360 2360 2360 2360 2360 2360 2360 2360 2360

1 2361 2364 2363 2369 2375 2348 2348 2351 2360 2364 2363 2368 2374 2348 2348 2351 2361

2 2359 2363 2362 2368 2373 2347 2347 2350 2359 2363 2361 2366 2373 2346 2347 2350 2359

3 2357 2361 2359 2365 2370 2345 2344 2346 2356 2360 2359 2364 2370 2343 2345 2347 2357

4 2353 2358 2356 2362 2368 2341 2341 2344 2352 2356 2355 2360 2367 2340 2341 2344 2353

5 2349 2353 2352 2358 2363 2336 2337 2338 2347 2352 2350 2356 2363 2336 2337 2339 2349

6 2343 2348 2346 2352 2358 2331 2331 2332 2342 2345 2345 2350 2357 2330 2330 2333 2343

7 2337 2341 2341 2345 2350 2325 2324 2327 2334 2339 2338 2343 2349 2325 2324 2328 2337

8 2329 2334 2332 2337 2342 2318 2317 2318 2326 2331 2329 2336 2342 2316 2317 2320 2329

9 2321 2325 2323 2328 2334 2309 2308 2310 2318 2321 2321 2326 2332 2306 2308 2312 2321

10 2310 2314 2314 2319 2323 2298 2299 2299 2307 2312 2310 2318 2323 2297 2299 2301 2310

11 2299 2305 2303 2307 2313 2288 2287 2287 2297 2299 2300 2305 2311 2286 2287 2291 2299

12 2288 2291 2292 2297 2299 2275 2274 2276 2283 2289 2287 2292 2298 2275 2277 2279 2288

13 2274 2281 2279 2283 2287 2263 2263 2262 2269 2274 2272 2280 2287 2261 2262 2265 2274

14 2261 2265 2263 2268 2273 2248 2248 2250 2256 2258 2259 2265 2271 2246 2248 2253 2261

15 2246 2249 2250 2254 2256 2232 2234 2233 2240 2245 2243 2248 2257 2233 2235 2236 2246

16 2226 2236 2233 2237 2242 2218 2217 2216 2225 2227 2228 2234 2240 2215 2217 2222 2226

17 2212 2217 2218 2222 2223 2202 2199 2200 2207 2212 2210 2215 2224 2200 2203 2204 2212

18 2193 2202 2199 2203 2207 2184 2183 2181 2186 2192 2194 2200 2205 2180 2184 2185 2193

19 2173 2181 2183 2187 2187 2168 2163 2164 2170 2171 2174 2179 2184 2165 2164 2169 2173

20 2155 2165 2162 2165 2169 2147 2145 2143 2148 2154 2151 2157 2166 2144 2146 2147 2155

21 2132 2142 2144 2148 2147 2124 2123 2125 2129 2131 2133 2140 2144 2126 2124 2129 2132

22 2114 2123 2121 2124 2128 2105 2100 2101 2106 2112 2110 2116 2125 2103 2106 2106 2114

23 2090 2099 2097 2100 2104 2080 2081 2077 2081 2087 2090 2096 2101 2079 2082 2087 2090

24 2071 2074 2077 2079 2084 2060 2055 2056 2060 2062 2065 2071 2080 2058 2062 2062 2071

25 2044 2054 2051 2053 2057 2035 2034 2030 2034 2041 2044 2050 2054 2033 2036 2042 2044

26 2023 2027 2030 2032 2036 2014 2008 2009 2012 2013 2017 2023 2032 2012 2015 2016 2023

27 1995 2005 2003 2004 2008 1986 1986 1981 1984 1991 1989 1995 2004 1984 1987 1994 1995

28 1972 1977 1980 1983 1985 1957 1958 1958 1961 1962 1966 1973 1976 1962 1965 1966 1972

29 1944 1954 1951 1954 1956 1934 1928 1929 1932 1939 1937 1943 1952 1933 1936 1942 1944

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

8 / 16

30 1914 1924 1928 1931 1926 1904 1905 1899 1907 1908 1913 1919 1922 1909 1913 1913 1914

31 1890 1899 1897 1900 1901 1880 1874 1874 1877 1884 1883 1889 1898 1879 1883 1889 1890

32 1858 1869 1873 1875 1869 1849 1850 1844 1845 1852 1857 1864 1866 1854 1858 1858 1858

33 1833 1843 1841 1842 1844 1824 1818 1818 1820 1821 1826 1832 1841 1823 1827 1827 1833

34 1801 1812 1815 1815 1811 1791 1792 1786 1787 1795 1793 1806 1808 1791 1795 1801 1801

35 1775 1786 1783 1782 1785 1765 1760 1760 1761 1762 1767 1773 1782 1765 1769 1769 1775

36 1742 1752 1750 1750 1752 1732 1726 1726 1728 1735 1734 1740 1749 1732 1736 1742 1742

37 1715 1726 1723 1725 1725 1698 1699 1693 1700 1701 1707 1713 1716 1705 1710 1709 1715

38 1681 1691 1689 1693 1690 1672 1665 1665 1666 1674 1673 1679 1688 1672 1676 1682 1681

39 1653 1657 1662 1666 1662 1637 1638 1631 1638 1639 1645 1652 1654 1645 1648 1647 1653

40 1618 1629 1627 1630 1627 1610 1603 1603 1603 1604 1610 1617 1626 1610 1615 1620 1618

41 1590 1594 1599 1600 1599 1574 1575 1568 1567 1575 1575 1589 1590 1583 1587 1585 1590

42 1555 1566 1563 1561 1563 1546 1540 1540 1539 1540 1546 1554 1555 1548 1552 1557 1555

43 1519 1530 1535 1531 1527 1511 1504 1504 1503 1511 1511 1518 1526 1519 1524 1522 1519

44 1490 1501 1499 1501 1499 1475 1475 1468 1474 1475 1482 1489 1491 1484 1488 1486 1490

45 1455 1465 1470 1465 1462 1446 1439 1439 1438 1446 1446 1453 1461 1455 1460 1458 1455

46 1426 1436 1434 1436 1433 1410 1410 1403 1401 1409 1417 1423 1425 1420 1424 1421 1426

47 1390 1399 1405 1399 1396 1381 1374 1374 1372 1373 1380 1387 1396 1391 1388 1393 1390

48 1360 1370 1368 1370 1367 1344 1344 1337 1335 1344 1344 1358 1359 1355 1359 1356 1360

49 1324 1334 1339 1333 1330 1308 1308 1308 1305 1307 1315 1322 1322 1326 1323 1327 1324

50 1294 1304 1302 1304 1300 1278 1271 1271 1269 1277 1278 1285 1293 1290 1293 1291 1294

51 1257 1267 1265 1267 1263 1242 1242 1234 1239 1240 1248 1255 1256 1253 1257 1261 1257

52 1228 1230 1235 1230 1233 1212 1205 1204 1202 1203 1211 1219 1227 1224 1227 1225 1228

53 1191 1201 1198 1200 1196 1175 1175 1168 1165 1174 1174 1189 1190 1188 1191 1195 1191

54 1161 1163 1168 1163 1166 1146 1138 1138 1135 1136 1145 1152 1160 1158 1162 1159 1161

55 1124 1134 1131 1133 1128 1109 1108 1101 1098 1107 1108 1115 1123 1122 1125 1130 1124

56 1095 1097 1102 1096 1099 1072 1071 1071 1068 1070 1078 1086 1086 1092 1096 1093 1095

57 1057 1067 1064 1066 1061 1042 1034 1034 1031 1040 1041 1049 1056 1055 1059 1063 1057

58 1028 1030 1035 1029 1024 1005 1005 997 993 1003 1011 1019 1019 1027 1029 1026 1028

59 991 1000 997 999 994 976 968 967 964 966 974 982 990 990 993 997 991

60 954 963 968 962 956 939 938 930 926 936 937 953 953 961 963 960 954

61 924 933 931 933 926 902 901 901 896 899 908 916 916 924 927 924 924

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

9 / 16

62 887 896 901 895 889 872 864 864 859 869 871 879 886 895 890 894 887

63 857 867 864 866 859 835 835 827 829 832 841 850 850 859 861 858 857

64 821 830 834 829 822 806 798 797 792 795 804 813 820 822 824 828 821

65 791 800 797 792 792 769 768 760 755 765 775 783 783 793 795 792 791

66 754 763 760 762 755 740 731 731 725 729 738 747 754 757 759 763 754

67 725 734 731 725 725 703 702 694 688 699 702 717 717 728 730 726 725

68 688 697 694 696 687 667 665 665 659 663 672 681 680 691 693 697 688

69 659 660 665 659 658 637 629 628 622 633 636 644 651 662 664 660 659

70 622 631 628 630 621 601 600 592 593 597 607 615 614 627 628 631 622

71 586 595 599 593 591 572 564 563 556 561 571 579 585 598 600 595 586

72 557 566 563 564 555 536 535 527 520 532 535 550 549 563 564 566 557

73 521 530 534 528 519 508 500 498 491 496 507 515 513 534 536 531 521

74 492 501 499 500 490 473 464 463 455 468 471 480 485 499 500 502 492

75 457 466 470 464 454 438 437 428 427 433 443 452 450 471 473 468 457

76 428 438 435 436 426 410 402 401 392 405 409 417 422 437 438 433 428

77 394 403 408 402 391 377 375 367 358 371 382 390 387 409 411 405 394

78 367 376 374 375 364 350 342 340 332 338 349 357 360 376 377 372 367

79 333 343 347 342 331 318 317 308 299 312 317 331 328 343 344 345 333

80 307 317 315 310 305 287 286 283 274 281 292 299 296 318 319 313 307

81 276 285 284 285 274 263 256 253 244 257 262 269 271 287 288 288 276

82 252 255 260 255 250 235 234 226 216 229 239 246 242 263 264 258 252

83 223 233 232 233 222 214 208 205 195 202 212 219 220 235 236 235 223

84 201 206 211 207 197 190 189 181 171 183 188 194 195 214 214 208 201

85 176 186 186 188 178 168 167 159 150 161 170 176 172 189 190 184 176

86 154 163 169 166 157 153 148 144 135 142 150 156 155 172 167 166 154

87 138 144 149 147 143 136 135 128 119 129 134 142 138 152 151 146 138

88 121 130 132 134 124 116 120 117 109 115 121 123 117 134 134 132 121

89 110 116 119 112 102 98 98 98 91 100 104 100 99 118 121 116 110

90 95 101 96 92 84 72 72 69 67 73 73 78 75 96 102 99 95

91 70 72 70 64 60 49 38 45 48 44 37 47 55 73 68 75 70

92 49 47 34 36 41 23 8 20 26 24 12 20 30 43 39 43 49

93 25 20 5 17 20 6 3 5 9 7 3 6 11 17 7 21 25

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

10 / 16

94 9 4 2 3 5 4 3 3 3 2 3 4 3 5 2 4 9

95 2 2 3 4 2 5 6 2 2 2 3 6 3 3 3 2 2

96 2 2 9 7 2 13 17 2 2 2 13 16 6 6 13 2 2

97 2 4 22 16 5 22 29 8 2 5 24 27 16 15 22 8 2

98 12 21 34 25 19 34 34 24 13 19 35 37 33 27 32 22 12

99 30 38 34 37 35 43 32 36 25 35 35 46 47 36 32 38 30

100 43 48 31 44 49 52 30 48 41 45 33 50 63 46 29 50 43

101 56 53 29 51 65 55 28 50 51 51 31 48 74 51 27 52 56

102 57 50 27 51 74 53 26 48 55 49 30 46 74 50 25 49 57

103 54 47 25 48 74 51 24 45 53 47 28 44 72 47 23 47 54

104 52 45 24 46 72 48 22 43 51 44 27 35 70 45 22 45 52

105 49 43 23 42 71 42 22 41 49 42 26 27 65 38 21 42 49

106 47 40 22 35 68 31 22 39 46 40 25 19 50 27 20 40 47

107 44 38 22 23 61 22 22 37 44 38 25 17 37 19 20 38 44

108 42 36 21 16 45 16 22 34 42 35 24 20 32 16 20 35 42

109 40 34 21 15 32 17 22 32 39 33 24 29 32 20 20 33 40

110 38 32 21 21 29 25 21 30 37 32 23 34 41 27 20 31 38

111 35 29 20 29 31 32 21 29 35 30 23 35 52 35 19 29 35

112 33 27 20 36 42 38 21 27 33 28 22 35 55 36 19 27 33

113 31 26 19 36 50 38 20 24 31 26 21 35 53 36 19 25 31

114 29 23 18 35 52 37 19 23 28 24 20 35 52 36 18 23 29

115 27 21 17 35 50 36 18 21 26 22 19 34 50 35 17 21 27

116 24 20 16 34 48 35 17 19 24 20 18 33 48 34 17 20 24

117 22 17 15 33 46 34 16 17 22 18 17 33 46 34 16 18 22

118 20 16 14 32 44 33 15 15 21 16 15 32 44 33 15 16 20

119 18 14 13 31 42 32 14 13 18 15 14 31 42 32 14 14 18

120 17 12 12 30 40 31 13 12 17 13 13 30 40 31 13 12 17

121 15 11 11 29 38 30 12 10 15 11 12 29 38 29 12 11 15

122 13 9 11 28 36 28 12 9 13 10 11 28 36 28 11 9 13

123 11 7 10 26 35 27 11 7 12 8 11 26 35 27 11 8 11

124 10 6 10 25 33 26 11 6 10 7 10 25 33 26 11 6 10

125 8 5 9 24 31 24 10 5 9 6 3 24 31 25 10 5 8

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

11 / 16

126 7 4 9 23 29 23 9 4 7 5 2 23 30 24 9 4 7

127 5 3 7 22 27 22 9 3 6 3 2 22 28 22 3 3 5

128 4 2 2 20 25 21 5 2 5 2 2 21 27 21 1 2 4

129 3 1 1 19 24 19 2 2 3 2 2 19 25 20 1 1 3

130 2 1 1 18 22 18 2 1 3 2 2 18 23 19 1 1 2

131 1 1 1 16 21 17 2 2 2 2 2 17 22 17 1 1 1

132 1 1 1 15 19 15 2 1 2 2 5 16 20 16 1 1 1

133 1 1 1 14 18 14 2 1 2 2 4 6 19 15 1 1 1

134 1 1 1 13 17 13 2 2 2 2 4 2 17 6 4 1 1

135 1 1 2 9 15 11 2 2 2 2 3 2 15 1 4 1 1

136 1 1 2 2 14 2 2 2 2 2 3 2 4 1 3 2 1

137 1 1 2 2 12 2 2 2 2 2 2 2 2 1 2 1 1

138 1 1 1 2 3 2 2 2 2 2 2 2 2 1 2 2 1

139 1 1 1 1 2 2 2 2 2 2 2 2 2 1 2 2 1

140 2 1 1 2 2 2 2 2 2 2 2 4 2 1 1 2 2

141 2 1 2 2 2 2 2 2 2 2 2 5 2 3 1 1 2

142 2 1 1 2 2 2 2 2 2 2 2 4 2 4 1 2 2

143 2 2 2 3 2 2 2 2 2 2 2 3 4 4 1 1 2

144 2 2 2 2 2 2 2 2 2 2 2 3 4 3 2 2 2

145 2 2 2 2 2 2 2 2 2 2 2 2 3 2 2 2 2

146 2 2 2 2 2 2 2 2 2 2 2 2 3 2 2 2 2

147 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

148 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

149 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

150 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

151 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

152 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

153 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

154 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

155 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

156 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

157 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

12 / 16

158 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

159 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

160 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

161 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

162 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

163 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

164 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

165 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

166 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

167 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

168 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

169 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

170 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

171 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

172 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

173 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

174 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

175 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

176 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

177 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

178 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

179 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

180 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

13 / 16

2.2 Electrical, Photometric and Chromaticity Measurements

(Refer to Work Instruction QD25)

Test date 2017-04-25 Test Ambient: 25.2 º C

Test Orientation As intended Stabilization Time (min) 90

Model Number 33-CP1-760-857

Electrical Measurement:

Sample No. Voltage (Vac) Frequency (Hz ) Current (A) Power (W) Power Factor THD %

GZE170372-

H-B2

120.0 60 0.4700 55.97 0.9924 5.48

277.0 60 0.2093 55.42 0.9558 7.56

DLC Pass Criteria >= 0.9(-3%) <= 20(+5)

Chromaticity Measurement - Sphere-Spectroradiometer Method:

Parameter Result Special Color Rendering Indices

Test Voltage (V) 120.0 R1 77 R9 0

Frequency (Hz) 60 R2 82 R10 55

CCT (K) 5660 R3 83 R11 76

Duv -0.0005 R4 79 R12 49

Chromaticity (x, y) x=0.3289 y=0.3371 R5 77 R13 77

Chromaticity (u’, v’) u'=0.2060 v'=0.4750 R6 74 R14 90

Color Rendering Index (CRI) 77.7 R7 85 R15 73

R9 0 R8 66 -- --

Photometric Measurement – Sphere-Spectroradiometer Method:

Parameter Result DLC V4.0 Pass Criteria

Test Voltage (V) 120.0 277.0 --

Frequency (Hz) 60 60

Total Luminous (lm) 6752.8 6701.4 >=2000(-10%)

Luminous Efficacy (lm/W) 120.65 120.92 Standard: >=

95(-3%)

Premium: >=

115(-3%) Most Worst Luminous/Highest Watts 119.73

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

14 / 16

Spectral Power Distribution & Chromaticity Diagram

STANDARD-TECH Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

15 / 16

2.3 Performance Assessment:

Model name CCT(K) Total Luminous

(lm) Power (W)

Luminous

Efficacy (lm/W)

33-CP1-760-840 4000K 6592.7 55.57 118.64

33-CP1-760-845 4500K 6646.1*1

55.77*2

119.17*3

33-CP1-760-850 5000K 6699.4*1

55.77*2

120.13*3

33-CP1-760-857 5700K 6752.8 55.97 120.65

*1: This value is calculated and the calculation formula is as below:

6646.1= (6752.8-6592.7)/3+6592.7

6699.4= (6752.8-6592.7)/3+6646.1

*2: This value is calculated and the calculation formula is as below:

55.77= (55.57+55.97)/2

*3: This value is calculated and the calculation formula is as below:

119.17=6646.1/55.77

120.13= 6699.4/55.77

STANDARD-TECH

Report No.: GZE170372-H-B

Laboratory: Standard-Tech Co. Ltd Testing Center

NVLAP CODE: 201011-0

Report Format Number STD/QR4909-A/2

Address:Standard-Tech Building, No.6 Guanhong Road,Guangzhou Science City, Guangzhou 510663, China

Tel: 8620-3229 0320 Fax: 8620-32290422 http://www.standard-tech.com

16 / 16

3. Test Equipment

Equipment ID Equipment Name Last Calibration Date Next Calibration Date

ST-R-336 2 meter Integrating Sphere 2016-07-01 2017-06-30

ST-R-331 Spectral analysis system HAAS-2000 2016-07-01 2017-06-30

D204 Standard Lamp 2016-07-01 2017-06-30

PF2010 Power Meter for Integrating Sphere 2016-07-01 2017-06-30

EE-09 Goniophotometer system 2016-07-01 2017-06-30

D908S Standard Lamp 2016-07-01 2017-06-30

PF210 Power Meter for Goniophotometer 2016-07-01 2017-06-30

ST-R-181A Temperature Tester 2016-07-01 2017-06-30

Uncertainty:

Photometric Measurement (Sphere):1.74%

Chromaticity Measurement(Sphere):14.3K

Photometric Measurement(Goniophotometer):1.62%

****** END OF REPORT ******