Linear Accelerator Quality Control
Using Clinical Treatment Delivery:
A Feasibility Study
Charles M. Able, MS
Assistant Professor, Radiation Oncology
![Page 2: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/2.jpg)
Objective
▪ To determine if dose distributions resulting from
clinical treatment delivery are sensitive enough to
detect operational changes in mechanical parameters
at the level specified by TG-40.
▪ To evaluate if treatment delivery accuracy can be
determined using process behavior charts developed
from clinical treatment delivery dose distributions
![Page 3: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/3.jpg)
Background
▪ Statistical Process Control (SPC)
▪ W. A. Shewhart, Bell Labs, 1924
▪ Application of probability and statistics to quality
control of mass production
▪ Introduced the control chart (or process behavior
chart)
![Page 4: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/4.jpg)
Clinical Treatment Plans
![Page 5: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/5.jpg)
Materials and Methods
▪ 3 clinical treatment plans
▪ Mechanical Parameters manipulated at 0.5, 1.0 & 1.5
times the TG-40 Tolerance
▪ Gantry angle
▪ Collimator angle
▪ Table Height
▪ Table Lateral
▪ Field Size
![Page 6: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/6.jpg)
Materials and Methods
![Page 7: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/7.jpg)
Results
![Page 8: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/8.jpg)
F Test of Dose Difference Distribution
F Test @ 0.01 Significance Level-Dose Difference Distribution
Modification Magnitude 4 Field Box Whole Pelvis Prostate IMRT
none none
Gantry 0.5 deg
Gantry 1.0 deg
Gantry 1.5 deg
Collimator 0.5 deg
Collimator 1.0 deg
Collimator 1.5 deg
Table Height 1.0 mm
Table Height 2.0 mm
Table Height 3.0 mm
Table Lateral 1.0 mm
Table Lateral 2.0 mm
Table Lateral 3.0 mm
Field Size 1.0 mm N/A
Field Size 2.0 mm N/A
Field Size 3.0 mm N/A
Color Key Not Significant Significant @0.01
![Page 9: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/9.jpg)
Results
▪ Vertical and Horizontal profiles across the penumbra
were extracted and evaluated.
![Page 10: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/10.jpg)
Results
4 Field Box: Table Lateral
Horizontal Profile
-30
-25
-20
-15
-10
-5
0
5
10
15
20
25
30
0 2 4 6 8 10 12 14 16
Horizontal Position (cm)
Perc
enta
ge D
iffe
rence
NoModification
1mm
2mm
3mm
![Page 11: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/11.jpg)
Chi-Square Test Horizontal Profile
Chi-Square Test @ 0.05 Significance Level - Horizontal Profiles
Modification Magnitude 4 Field Box Whole Pelvis Prostate IMRT
none none
Gantry 0.5 deg
Gantry 1.0 deg
Gantry 1.5 deg
Collimator 0.5 deg
Collimator 1.0 deg
Collimator 1.5 deg
Table Height 1.0 mm
Table Height 2.0 mm
Table Height 3.0 mm
Table Lateral 1.0 mm
Table Lateral 2.0 mm
Table Lateral 3.0 mm
Field Size 1.0 mm N/A
Field Size 2.0 mm N/A
Field Size 3.0 mm N/A
Color Key Not Significant Significant @ 0.05
![Page 12: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/12.jpg)
Chi-Square Test-Vertical Profile
Chi-Square Test @ 0.05 Significance Level - Vertical Profile
Modification Magnitude 4 Field Box Whole Pelvis Prostate IMRT
none none
Gantry 0.5 deg
Gantry 1.0 deg
Gantry 1.5 deg
Collimator 0.5 deg
Collimator 1.0 deg
Collimator 1.5 deg
Table Height 1.0 mm
Table Height 2.0 mm
Table Height 3.0 mm
Table Lateral 1.0 mm
Table Lateral 2.0 mm
Table Lateral 3.0 mm
Field Size 1.0 mm N/A
Field Size 2.0 mm N/A
Field Size 3.0 mm N/A
Color Key Not Significant Significant @ 0.05
![Page 13: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/13.jpg)
Results
![Page 14: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/14.jpg)
Results
![Page 15: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/15.jpg)
Conclusion
▪ Clinical treatment delivery is sensitive enough to
detect inaccuracies in mechanical parameters
▪ Strategically sampled clinical treatment dose
distributions can be used to provide quality control
for mechanical parameters using statistical process
control methodology.
![Page 16: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/16.jpg)
Thank You For Your Attention
![Page 17: Linear Accelerator Quality Control Using Clinical ......process control for radiotherapy quality assurance” Med Phys 32 (9) 2777-2786. Pawlicki, R., Whitaker, M.“Variation and](https://reader033.fdocuments.us/reader033/viewer/2022060502/5f1bbf8cdd58dd73be515e1b/html5/thumbnails/17.jpg)
References
▪ Kutcher, G.J., Coia, L., Gillin, M., et al.
“Comprehensive QA for radiation oncology: Report of
radiation therapy committee task group 40” Med Phys
21 (4) 581-618.
▪ Pawlicki, R., Whitaker, M., Boyer, A.L. “Statistical
process control for radiotherapy quality assurance”
Med Phys 32 (9) 2777-2786.
▪ Pawlicki, R., Whitaker, M.“Variation and control of
process behavior” Int. J. Radiat Oncol Biol Phys 71(1)
supl s210-s214.