MENKAR Level indicator OPERATING INSTRUCTIONS © MECON GmbH BA / MENKAR / EN / 19-10

Transcript of Level indicator - MECON

Page 1: Level indicator - MECON

MENKARLevel indicator

OPERATING INSTRUCTIONS

© MECON GmbH BA / MENKAR / EN / 19-10

Page 2: Level indicator - MECON

2 © MECON GmbH BA / MENKAR / EN / 19-10

Content

1 SAFETYINSTRUCTIONS..............................................................................................41.1 Intended use .......................................................................................................................41.2 Certifications ......................................................................................................................51.3 Manufacturer’ssafetyinstructions ...............................................................................5

2 INSTRUMENTDESCRIPTION ......................................................................................62.1 Scopeofdelivery ...............................................................................................................62.2 Nameplate ...........................................................................................................................7

3 INSTALLATIONANDPRINCIPLEOFOPERATION ....................................................83.1 Installationnotes ...............................................................................................................83.2 Installationconditions ......................................................................................................83.3 Installationinstructions ...................................................................................................93.4 Principleofoperation .......................................................................................................93.5 Displayoftheactuallevel ................................................................................................103.6 Availablefloats ...................................................................................................................103.7 Changeofprocessconditions ........................................................................................103.8 Configuration ......................................................................................................................11

4 START-UP ......................................................................................................................11

5 TECHNICALDATA .........................................................................................................125.1 Max.pressure16bar,150lbs ..........................................................................................125.2 Max.pressure40bar,300lbs .........................................................................................145.3 Max.pressure250bar,1500lbs .....................................................................................155.4 Withoutsurgepipe(surfacetankmounting) ...............................................................165.5 WithsurgepipeØ54or60.3(surfacetankmounting) ..............................................175.6 With3guiderodsØ76orØ104(surfacetankmounting) ........................................185.7 WithsurgepipeØ76.1or88.9(surfacetankmounting) ............................................19

6 CONTACTORS ...............................................................................................................206.1 Generaluselevelswitch ..................................................................................................216.2 Intrinsicallysafelevelswitches(Exi) ............................................................................226.3 Pressure-safelevelswitches ..........................................................................................236.4 Reedchainforcontinuousdisplay.................................................................................24

Page 3: Level indicator - MECON

3© MECON GmbH BA / MENKAR / EN / 19-10

IMPRINT

Allrightsreserved.AnyreproductionorusagewithoutthewrittenauthorisationofMECONGmbH-evenpartialcontent-isstrictlyprohibited.Subjecttochangewithoutnotice.

Copyright2019MECONGmbH-Röntgenstr.105-50169Kerpen-Germany

7 SERVICE .........................................................................................................................267.1 Storage ................................................................................................................................267.2 Maintenance .......................................................................................................................267.3 Returningtheinstrumenttothemanufacturer ..........................................................267.4 Disposal ...............................................................................................................................27

8 FAULTREPORTING .......................................................................................................27

Page 4: Level indicator - MECON

4 © MECON GmbH BA / MENKAR / EN / 19-10

SAFETY INSTRUCTIONS

1 SAFETYINSTRUCTIONS

1.1 Intended use

MENKARserieslevelindicatorsaresuitableforcontinuousmeasurementandmonitoringofthelevelofaliquidinopenorclosedcontainersasrequired.

Theliquidlevelinthecontainertobemonitoredistransmittedbyafloatina1:1ratioasananaloguevalue.Themeasurementisindependentoftheoperatingpressure.Theinstru-mentsarenotedfortheir robustconstruction,trouble-freeusageandgoodreadability,evenfromfurtheraway.

Variousmagneticallyoperatedcontactunitsforcontrolandregulationprocesses(acous-ticandopticalsignals,motorandvalvecontrollersorsimilar)areavailableasaccessories.Apartfromrequiringacertainminimumdistance,thecontactscanbefittedinanyorderandinanynumber.

Therobustdesignmeansitcanevenbeusedintoughconditions.Varioustypesofprocessconnectionsandfloatsmeetthediverseconditionsfoundacrossindustry.

Special features: » Universalapplicationsforalmostalltypesofliquids » Simple,robustconstructionwithmagneticdatatransmission » Strong,magneticcouplingsystemwithoutanymechanicaltransmission elements

» Goodreadability,evenfromadistance » Canalsobesuppliedforuseunderhighpressuresandtemperatures » Numberofcontactsandarrangementofthecontactisdefinedbytheuser;andonlylimitedbythecontacthousingdimensions

» Nohydraulicconnectionbetweenindicatorandmedium » Easytoinstall » Highmeasurement/displayedrange.Alsoavailableinasplitdesignabove5000 mm. » Alsoavailablewithamagneticflapperdisplay,

Page 5: Level indicator - MECON

5© MECON GmbH BA / MENKAR / EN / 19-10

SAFETY INSTRUCTIONS

1.2 Certifications

CEmarking

ByaffixingtheCEmark,themanufacturercertifiesthattheMENKARlevelindicator,whereapplicable,complieswiththelegalrequirementsofthefollowingEUdirective:

» PressureEquipmentDirective2014/68/EU » ThepermittedmediaareGroup1gasesandliquids » ClassIII(ClassificationundertheconditionPS.V<1000(Article4.1.1a)

1.3 Manufacturer’ssafetyinstructions DisclaimerThemanufacturerisnotliablefordamagesofanykindcausedbytheuseoftheinstru-ment,including,butnotlimitedto,direct,indirect,incidental,punitiveandconsequentialdamages.

Awarrantyappliestoeveryproductpurchasedfromthemanufacturer,accordingtotherelevantproductdocumentationandourvalidtermsandconditions.Themanufacturerreservestherighttorevisethecontentofthedocuments, includingthisdisclaimer,withoutnotice,andisnotliableinanywayforanypossibleconsequencesofsuchchanges.

ProductliabilityandwarrantyTheresponsibilityforensuringthatthemeasuringinstrumentissuitablefortheparticu-larapplicationrestssolelywiththeoperator.MECONGmbHassumesno liabilityfortheconsequencesofmisuse,modificationsorrepairsthatwerecarriedoutbythecustomerwithoutpriorconsultation.

Incaseofacomplaint thedisputedelementsmustbecleanedofanyhazardoussub-stancesandbereturnedtothemanufacturer,unlessotherwiseagreed.

GeneralinformationTopreventinjurytotheuserordamagetotheinstrument,itisnecessarythatyoureadtheseoperatinginstructionscarefullybeforestartingtouseit.

Thishandbookis intendedtocoverthecorrect installation,operationandmaintenanceof the instrument.Specialdesignsfor special applicationsandcustommodelsarenotcoveredbythisdocumentation.

0035

Page 6: Level indicator - MECON

6 © MECON GmbH BA / MENKAR / EN / 19-10

INSTRUMENT DESCRIPTION

2 INSTRUMENTDESCRIPTION

2.1 Scopeofdelivery

MENKARwithlocalindicator,capanddrainplugFloatValve(vent/drain)-optionalOperatinginstructionsOptional:Magneticflapper(mounted),contacts,reedchain

Fig. 1 Scope of delivery

Page 7: Level indicator - MECON

7© MECON GmbH BA / MENKAR / EN / 19-10

INSTRUMENT DESCRIPTION

2.2 Nameplate

Type DevicenameDrawingNo. DrawingnumberSerialNo. SerialnumberConnection ProcessconnectionCtoC Nozzledistance(centretocentre)Tsmin./max. Temperaturerange(min./max.)ofmediumin°Cpmin./max. Pressure(min/max)pmax.atTsmax. MaximumpermissibleoperatingpressureatTsmax.Material MaterialusedforthewettedpartsSpecificdensity RelativedensityCE CEmarking,classificationaccordingtothe

PressureEquipmentDirectiveTAGNo. TAGnumber

Page 8: Level indicator - MECON

8 © MECON GmbH BA / MENKAR / EN / 19-10

INSTALLATION AND PRINCIPLE OF OPERATION

3 INSTALLATIONANDPRINCIPLEOFOPERATION

3.1 Installation notes

Information!

All instruments are carefully checked for proper functionality before shipment.

Immediately on receipt, check the outer packing carefully for damage or signs of improper handling. Report any damage to the carrier and your local sales staff. In such cases, a description of the damage, the type and the serial number of the instrument should be given.

Unpack the instrument carefully to avoid damage.

Check the delivery for completeness on the basis of the delivery note and the match between the valve and the float.

Check the nameplate to ensure that the delivered level indicator meets with your order.

3.2 Installation conditions

Thefollowingpointsmustbeobservedwheninstallingtheinstrument.

» TheMENKARlevelindicatormustbemountedverticallyonthetank. » Itmustbeensuredthatanymagneticfieldsgeneratedbyotherinstrumentsdonotaffectthemeasurements.

» Selectscrewsandgaskets(tobeprovidedbythecustomer)thatmeetthepressurespecificationsoftheconnectionflangeandtheoperatingpressure.

» Theprocessconnections(flanges)mustfit,i.e.theymustbecentredandparallelandproperlyboltedontopreventanyunnecessarymechanicalstressbeingappliedtotheinstallation.

» Thetankmustbefreeofanycontamination.Theuseofshut-offelementsisrecom-mendede.g.taps,valves,etc.,installedbetweentankandlevelindicator,sothatthelevelindicatorcanbecleanedindependentlyofthetank.Thedrainplug(ifordered)inthebottomflangeshouldalsobereplacedwithadraintapfittedwithadrainpipe.

Page 9: Level indicator - MECON

9© MECON GmbH BA / MENKAR / EN / 19-10

INSTALLATION AND PRINCIPLE OF OPERATION

Thefollowingpointsshouldbenotedwheninstallingtheinstrumentsinthepipeline:

» Removethetransportlockfromthevalve. » Beforeinstallation,checkthatthefloatinthevalvecanmovefreelywithoutjamming.

3.3 Installation instructions » Theeffectivepressureoftheinstallation(themaximumpressureallowedbythepres-surereliefvalve)mustneverexceedthemaximumallowablepressurePSindicatedonthenameplate.

» Theusermustensurethatthematerials(guidetube,floats,seals,etc.)thatcomeintocontactwiththefluidarecompatiblewiththisfluidandmeettheageingcharacter-isticsofthefluidandthemeasurementenvironment.Theappropriateinformationiseithergiveninthemanualorlistedinthecontractunderownspecifications.

» Theexternalpressure(Pext)mustequaltheatmosphericpressure(Patm).

3.4 Principle of operation

TheMENKARlevelindicatorsoperatewithasystemofcommunicatingtubesorthefloatprinciple:

Themediumentersfromthetankthroughthelowerconnectingtubeinthestandpipe.Thefloatlocatedinthestandpipethenfloatsintheliquid,itsheight(seeFig.1)correspondingdirectlytothelevelinthetank.

Eachmovementofthefloat(whenthelevelchanges)istransmittedtothemagneticflap-persbyamagnetinsertedinthefloat(rotationfromwhitetored)-othercoloursonre-quest-theredsideindicatingthelevel.

Asanoption,contactdevicesareavailablethatarealsoactu-atedbythefloat.

Themagnetic transmissionsystemworkswithoutcontact inallcases.

Fig. 1: Principle of operation

Page 10: Level indicator - MECON

10 © MECON GmbH BA / MENKAR / EN / 19-10

INSTALLATION AND PRINCIPLE OF OPERATION

3.5 Displayoftheactuallevel

Thefloatisequippedwitharingsystemmadeofpermanentmagnetsinordertotransmittheliquidlevel.

Fordesignreasons,theminimumlevelinthemeasuringtubeisindicatedbytheaxisofthelowerlateralconnectingflange,i.e.thezeropointformsthecentrelineofthelowerconnectingflange.

Intheeventofamajorchangeintheproductdensityspecifiedintheorder,orwhenin-stallinganotherfloat,thescaleontheMENKARmayneedtobeadjustedtoensureacor-rectreading.Insuchacase,pleasecontactMECONGmbH.

3.6 Availablefloats

Allmagneticlevelindicatorsareequippedwithafloat.Asstandard,thefloatissuppliedinstainlesssteel,butcanalsobeorderedintitanium,Hastelloy,PVC-C,PVC-U,PPorPVDF.Thefloatmusthavesufficientbuoyancy,sotheprocessconditionsshouldbeconsideredwhenmakingtheselection.

Thefollowinginformationisrequiredforchoosingtherightfloat: » Medium » Density » Operatingpressure » Operatingtemperature

Thelowestselectabledensityis380kg/m³,dependingontheprocessconditions.Inthecaseofaggressiveliquids,thefloatcanalsobeorderedwithalining.Examplesinclude:E-CTFE,PFAandETFE.

3.7 Changeofprocessconditions

IftheuserwishestousetheMENKARtomeasureanotherproduct,thefollowingpointsshouldbeconsidered:

» ContactMECONGmbHforinformationaboutthecompatibilityofinstrumentsandproducts,especiallyiftheinstrumentistobeusedinhazardousareas.

» Whereapplicable,ensurethattherequirementsofthePressureEquipmentDirective2014/68/EUareadheredto.

» Theimmersiondepthofthefloatincreaseswithdecreasingdensityoftheproduct.Thisdepthalsodependsonthemodeloffloatandthematerialused(stainlesssteelortitanium).

Page 11: Level indicator - MECON

11© MECON GmbH BA / MENKAR / EN / 19-10

START-UP

3.8 ConfigurationMENKARlevelindicatorsaremanufacturedinlengthsofupto5,000mm,dependingonthetype,andwithasplitdesignforlongerlengthsinordertofacilitatetransport.Thecon-nectionismadeusingflanges.Thefollowingpointsmustbeobservedtakingaccountoftheparticularoperatingdataandlocalconditions: » Typeoffitting » Typeanddesignofinstrument » Length/measuringrange » Scaletype(%,volumeorheightunits):Forscalesinvolumeunits,theusermustprovideanoutflowtable.Thescalecanalsoberetrofittedandinstalledbytheuser.Forrectangularandcylindricalcontainerswithflatorcurvedbases,inahorizontalorverticalposition,thescalecanbecalculatedatthefactoryinunitsofvolume(atreimbursementofcost).Forthis,thegeometricdimensionsofthecontainermustbespecified.

» Gatevalve:Itisadvisabletoprovideagatevalveintheconnectingpipes,sothatthelevelindicatorcanbecleaned,removedorre-installed.

Additionaldata: » Typeofmedium(withdensity) » Operatingpressureandoperatingtemperature » Nominaldiameteranddesignoftheconnectionflanges » Nominalwidthanddesignoffrontandblankflanges,ifdifferentfromstandard » Sealingmaterial.

4 START-UP

Theoperatinginstructionsaretobereadintheirentiretybeforeinstallationandcommis-sioning.Installationandrepairmayonlybecarriedoutbysuitablepersonnel!Downloadingthisdocumentfromthehomepageofwww.mecon.deandprintingitoutispermittedwhenitisusedwithoneofourlevelindicators.Thismanualmaynotberepro-ducedor transmittedbyphotocopyingor any other recording techniques or any othermeanswithoutthepriorwrittenconsentofMECONGmbH.

Great carehasbeen taken in thedevelopment andpreparationof thismanual.Never-theless,mistakescannotbefullyeliminated.Thecompanyandtheauthorarenotabletoacceptanylegalorotherliabilityforincorrectfunctionalityorinformationandanyconse-quencesthatmayarise.

MECONGmbHoffersnowarranty,expressorimplied,astoitssuitabilityforanypurposeotherthanthatdescribedinthismanual.

Page 12: Level indicator - MECON

12 © MECON GmbH BA / MENKAR / EN / 19-10

START-UP

5 TECHNICALDATA

5.1 Max.pressure16bar,150lbs

Wereservetherighttochangetechnicaldataasaresultoftechnologicalprogress.Thelatestinformationonthisproductcanbefoundatwww.mecon.Youcanalsocontactoursalesdepartmentviae-mailatinfo@mecon.de.

MENKAR LMENKAR D

Page 13: Level indicator - MECON

13© MECON GmbH BA / MENKAR / EN / 19-10

TECHNICAL DATA

Type D-16 / D 150 - L-16 / L-150

Material Stainlesssteel1.4404(316L),stainlesssteel1.4301(304),PP,PVC,

PVDF,PE,Monel,Titanium,Hastelloy,6Mo

Pipe 60.3x2mmor60.3x2.77mm

Pressure Max.16bar/150lbs

Temperature Max.+400°C

Measuringlength Upto5500mmin1piece,longerinseveralparts

Indicator rail Polycarbonate(maxtemp.+105°C,temporary+120°C)

Process connection DINDN15–DN32/PN16,B=75mm

ANSI½”–1¼”150#RF(RTJ),B=85mm

Weldingend/thread(external/internalthread),W=70mm

DN40–DN50andANSI1.½”–2”on1”pipe,B=130mm

Drain ¼",½"or¾"plugorvalveBSP/NPTorflange

Sideentryasabove

AdditionalflangeaccordingtoDINorANSI

Without

Sealingmaterial PTFE,aramides,graphite,spirallywound

Upper side ¼",½"or¾"plugorvalveBSP/NPTorflange

FlangeDN50/PN16orANSI2"150#RF

FlangeDN25/PN16(likedrain)

Without

Float Densitymin.380kg/m³

Pathlength Densitymin.920kg/m³A=210mm(*)

Densitymin.830kg/m³A=245mm(*)

Densitymin.720kg/m³A=295mm(*)

Densitymin.660kg/m³A=350mm(*)

Additionalreinforcement SpigotspacingCtoC>3metresforoffshore

SpigotdistanceCtoC>4metresforonshore

MENKAR Stainlesssteel

Marking NameplateaccordingtostandarddesigninVA

PEDmarkinguptoCategoryIIIstandard

Certifications MaterialcertificateEN102043.1+drawing

Pressure test

NACE MR 01.75 / ISO 15156

WPS/PQRstandardmaterial

Special Insulation,steamjacket,spring,electricheater

(*) Special length (also shorter) available on request.

Page 14: Level indicator - MECON

14 © MECON GmbH BA / MENKAR / EN / 19-10

TECHNICAL DATA

Type D-40/D-300,L-40/L-300

Material Stainlesssteel1.4404(316L),stainlesssteel1.4301(304),PP,PVC,

PVDF,PE,Monel,Titanium,Hastelloy,6Mo

Pipe 60.3x2mmor60.3x2.77mm

Pressure 40bar/300lbs

Temperature Max.+400°C

Measuringlength Upto5500mmin1piece,longerinseveralparts

Indicator rail Polycarbonate(maxtemp.+105°C,temporary+120°C)

AluminiumwithstainlesssteelVA4flaps

CompletelystainlesssteelVA4

Process connec-

tionDINDN15–DN32/PN40,B=75mm

ANSI½”–1¼”300#RF(RTJ),B=85mm

Weldingend/thread(external/internalthread),W=70mm

DN40–DN50andANSI1.½”–2”on1”pipe,B=130mm

Drain ¼",½"or¾"plugorvalveBSP/NPTorflange

Sideentryasabove

AdditionalflangeaccordingtoDINorANSI

Without

Sealingmaterial PTFE,aramides,graphite,spirallywound

Upper side ¼",½"or¾"plugorvalveBSP/NPTorflange

Sideentryasabove

AdditionalflangeaccordingtoDINorANSI

Without

Float Densitymin.390kg/m³

Pathlength Densitymin.920kg/m³A=210mm(*)

Densitymin.830kg/m³A=245mm(*)

Densitymin.720kg/m³A=295mm(*)

Densitymin.660kg/m³A=350mm(*)

Additional rein-

forcementSpigotspacingCtoC>3metresforoffshore

SpigotdistanceCtoC>4metresforonshore

MENKAR Stainlesssteel

Marking NameplateaccordingtostandarddesigninVA

PEDmarkinguptoCategoryIIIstandard

Certifications MaterialcertificateEN102043.1+drawing

Pressure test

NACE MR 01.75 / ISO 15156

WPS/PQRstandardmaterial

Special Insulation,steamjacket,spring,electricheater

(*) Special length (also shorter) available on request.

5.2 Max.pressure40bar,300lbs

MENKAR L

MENKAR D

Page 15: Level indicator - MECON

15© MECON GmbH BA / MENKAR / EN / 19-10

TECHNICAL DATA

Type D-64 / D-100 / D-160 / D-600 / D-900 / D-1500 / D-2500

Material Stainlesssteel1.4404(316L)/1.4571(316Ti)

Pipe 60.3x2.77mm/60.3x3.91mm

Pressure Max.250bar

Temperature Max.+450°C

Measuringlength Upto5500mmin1piece,longerinseveralparts

Indicator rail Polycarbonate(maxtemp.+105°C,temporary+120°C)

WithstainlesssteelVA4flaps

CompletelystainlesssteelVA4

Process connec-

tionDINDN15–DN32/PN64–PN160,B=80mm

ANSI½”–1¼”600#–2500#RF–RTJ,B=85mm

Weldingend/thread(external/internalthread),W=75mm

DN40–DN50andANSI1.½”–2”on1”pipe,W=130mm

Drain ¼”,½”or¾”plugsBSPorNPT

¼",½"or¾"valve

AdditionalflangeaccordingtoDINorANSI

Without

Sealingmaterial PTFE,aramides,graphite,spirallywound

Upperflange

(possiblyventi-

lation)

¼",½"or¾"plugorvalveBSP/NPTorflange

¼",½"or¾"valve

AdditionalflangeaccordingtoDINorANSI

Without

Float Densitymin.610kg/m³

Pathlength A=dependingonpressureandtemperature

Additional

reinforcementSpigotspacingCtoC>3metresforoffshore

SpigotdistanceCtoC>4metresforonshore

MENKAR High&lowinstainlesssteel

Marking NameplateaccordingtostandarddesigninVA

PEDmarkinguptoCategoryIIIstandard

Certifications MaterialcertificateEN102043.1+drawing

Pressure test

NACE MR 01.75 / ISO 15156

WPS/PQRstandardmaterial

Special Insulation,spring,electricheater

5.3 Max.pressure250bar,1500lbs

MENKAR D

Page 16: Level indicator - MECON

16 © MECON GmbH BA / MENKAR / EN / 19-10

TECHNICAL DATA

Type F-00A / F-00B

Material Stainlesssteel1.4404(316L),othersonrequest

Pipe 60.3x2mm(uppertank)

Surgepipe without

Pressure Max.60bar(dependingontype)

Temperature Max.+350°C

Measuringlength Max.5500mm

Indicator rail Polycarbonate(maxtemp.+105°C,temporary+120°C)

AluminiumwithstainlesssteelVA4flaps

CompletelystainlesssteelVA4

Process connec-

tionDINDN50–DN150/PN40,B

ANSI 2” – 6” 150# RF

ANSI 2” – 6” 300# RF

ANSI 2” – 6” 600# RF

Upper side ¼",½"or¾"plugorvalveBSP/NPT,flange

FloatF-00A FloatOD52mm

Densitymin.480kg/m³

Densitydependingonthemeasuringlength

Withameasuringlengthof1mforstandardfloats:

Fromdensity1210kg/m³A=115mm

Fromdensity1030kg/m³A=185mm

Fromdensity810kg/m³A=205mm

Fromdensity670kg/m³A=255mm

FloatF-00B FloatOD67or72mm

Densitymin.380kg/m³

Densitydependingonthemeasuringlength

Forameasuringlengthof1mforstandardfloats(OD72mm):

Fromdensity970kg/m³A=100mm

Fromdensity690kg/m³A=150mm

Fromdensity570kg/m³A=200mm

Fromdensity500kg/m³A=250mm

MENKAR Stainlesssteel

Marking NameplateaccordingtostandarddesigninVA

Certifications MaterialcertificateEN102043.1+drawing

Pressure test

BVcertificate

NACE MR 01.75 / ISO 15156

WPS/PQRstandardmaterial

5.4 Withoutsurgepipe(surfacetankmounting)

MENKAR F00

Page 17: Level indicator - MECON

17© MECON GmbH BA / MENKAR / EN / 19-10

TECHNICAL DATA

Type F-01 / F-01A

Material Stainlesssteel1.4404(316L),othersonrequest

Pipe 60.3x2mm(uppertank)

Surgepipe Pipes54or60.3

Pressure Max.60bar(dependingontype)

Temperature Max.+350°C

Measuringlength Max.5500mm

Indicator rail Polycarbonate(maxtemp.+105°C,temporary+120°C)

AluminiumwithstainlesssteelVA4flaps

CompletelystainlesssteelVA4

Process connec-

tionDINDN50–DN150/PN40,B

ANSI 2” – 6” 150# RF

ANSI 2” – 6” 300# RF

ANSI 2” – 6” 600# RF

Upper side ¼",½"or¾"plugorvalveBSP/NPT,flange

FloatF-01 SurgepipeOD60.3,floatOD52

Densitymin.480kg/m³

Densitydependsonthemeasuringlength.

Withameasuringlengthof1mforstandardfloats:

Fromdensity1160kg/m³A=150mm

Fromdensity1030kg/m³A=185mm

Fromdensity810kg/m³A=205mm

Fromdensity670kg/m³A=255mm

FloatF-01A SurgepipeOD54,floatOD47

Densitymin.600kg/m³

Densitydependingonthemeasuringlength

Withameasuringlengthof1mforstandardfloats:

Fromdensity1050kg/m³A=150mm

Fromdensity910kg/m³A=200mm

Fromdensity800kg/m³A=250mm

Fromdensity730kg/m³A=300mm

MENKAR Stainlesssteel

Marking NameplateaccordingtostandarddesigninVA

Certifications MaterialcertificateEN102043.1+drawing

Pressure test

BVcertificate

NACE MR 01.75 / ISO 15156

WPS/PQRstandardmaterial

5.5 WithsurgepipeØ54or60.3(surfacetankmounting)

MENKAR F-01A

Page 18: Level indicator - MECON

18 © MECON GmbH BA / MENKAR / EN / 19-10

TECHNICAL DATA

Type F-02 / F-04

Material Stainlesssteel1.4404(316L),othersonrequest

Pipe 60.3x2mm(uppertank)

Surgepipe 3guiderodsØ76orØ104

Pressure Max.60bar(dependingontype)

Temperature Max.+350°C

Measuringlength Max.5500mm

Indicator rail Polycarbonate(maxtemp.+105°C,temporary+120°C)

AluminiumwithstainlesssteelVA4flaps

CompletelystainlesssteelVA4

Process connec-

tionDINDN50–DN150/PN40,B

ANSI 2” – 6” 150# RF

ANSI 2” – 6” 300# RF

ANSI 2” – 6” 600# RF

Upper side ¼",½"or¾"plugorvalveBSP/NPT,flange

FloatF-02 3guiderodsØ76,floatOD52mm

Densitymin.480kg/m³

Densitydependingonthemeasuringlength

Withameasuringlengthof1mforstandardfloats:

Fromdensity1160kg/m³A=150mm

Fromdensity1030kg/m³A=185mm

Fromdensity810kg/m³A=205mm

Fromdensity670kg/m³A=255mm

FloatF-04 3guiderodsØ104,floatOD72mm

Densitymin.380kg/m³

Densitydependingonthemeasuringlength

Withameasuringlengthof1mforstandardfloats:

Fromdensity970kg/m³A=100mm

Fromdensity690kg/m³A=150mm

Fromdensity570kg/m³A=200mm

Fromdensity500kg/m³A=250mm

MENKAR Stainlesssteel

Marking NameplateaccordingtostandarddesigninVA

Certifications MaterialcertificateEN102043.1+drawing

Pressure test

BVcertificate

NACE MR 01.75 / ISO 15156

WPS/PQRstandardmaterial

5.6 With3guiderodsØ76orØ104(surfacetankmounting)

MENKAR F-02

Page 19: Level indicator - MECON

19© MECON GmbH BA / MENKAR / EN / 19-10

TECHNICAL DATA

Type F-03A / F-03B

Material Stainlesssteel1.4404(316L),othersonrequest

Pipe 60.3x2mm(uppertank)

Surgepipe Pipes76.1or88.9

Pressure Max.20bar(dependingontype)

Temperature Max.+350°C

Measuringlength Max.5500mm

Indicator rail Polycarbonate(maxtemp.+105°C,temporary+120°C)

AluminiumwithstainlesssteelVA4flaps

CompletelystainlesssteelVA4

Process connec-

tionDINDN50–DN150/PN40,B

ANSI 2” – 6” 150# RF

ANSI 2” – 6” 300# RF

ANSI 2” – 6” 600# RF

Upper side ¼",½"or¾"plugorvalveBSP/NPT,flange

FloatF-03A Pipe76.1,floatOD67

Densitymin.470kg/m³

Densitydependingonthemeasuringlength

Withameasuringlengthof1mforstandardfloats:

Fromdensity1050kg/m³A=100mm

Fromdensity760kg/m³A=150mm

Fromdensity630kg/m³A=200mm

Fromdensity560kg/m³A=250mm

FloatF-03B Pipe88.9,floatOD72

Densitymin.380kg/m³

Densitydependingonthemeasuringlength

Withameasuringlengthof1mforstandardfloats:

Fromdensity970kg/m³A=100mm

Fromdensity690kg/m³A=150mm

Fromdensity570kg/m³A=200mm

Fromdensity500kg/m³A=250mm

MENKAR Stainlesssteel

Marking NameplateaccordingtostandarddesigninVA

Certifications MaterialcertificateEN102043.1+drawing

Pressure test

BVcertificate

NACE MR 01.75 / ISO 15156

WPS/PQRstandardmaterial

5.7 WithsurgepipeØ76.1or88.9(surfacetankmounting)

MENKAR F-03

Page 20: Level indicator - MECON

20 © MECON GmbH BA / MENKAR / EN / 19-10

CONTACTORS

6 Contactors

Youwillreceiveelectricalsignalsbymountingoneormoremagneticswitchesonthelevelindicator.Withseveralswitchesyoucan,forexample,outputapumpcontrolcommand(pumpon/off)and/oranalarm(high/low).Levelswitchesforgeneraluseaswellasswitchesforhazardousareascanbesupplied.

Important!

The power supply must be switched off before wiring the instrument.

The density of the liquid can change due to changing pressure / temper-ature. In this case, the float can indicate a different level. In the case of intrinsically safe systems, it may be necessary to provide electrical galvanic isolation.

The process temperature of the switches may depend on the isolation options of the level indicator.

Operating manual:1. Mount the switch in the correct position.2. Move the float from the bottom to the top and back.3. Check the function of the switch.4. If necessary, change the connections of the switch if another function

is required.5. Connect the switch to the mains.

The integrated cables must be protected against pulling, twisting and mechanical damage.

Overvoltageprotectionforreedswitch:

Capacitiveloads(inthecaseofextremelylongcableguides)andlamploadsmaycausehighstartingcurrents,whichcangreatlyreducethelifeoftheswitchcontactswhenclos-ing. It is recommendedthatanovervoltageprotectioncircuit isused inserieswiththeswitchandascloseaspossibletotheswitch.

Fornormal signal circuits, thecapacity in thecable canbeneglectedbecause severalhundredmetresofcablemustbeconnectedtotheswitchtocausedamage.

Page 21: Level indicator - MECON

21© MECON GmbH BA / MENKAR / EN / 19-10

CONTACTORS

Type HLS-15 LMS-Ha2 LMS-Ha1

Function SPDT SPDT SPDT

System Reed switch bi-stable Reed switch bi-stable Reed switch bi-stable

Max. rated 2.5 A / 60 W / 60 VA 0.8 A / 60 W / 40 VA 2 A / 40 W / 100 VA

Supply voltage 10 - 230V 10 - 230V 10 - 230V

Temperature range -25… +95 °C -40… +180 °C -50… +380 °C

Lifetime 1 x 109 1 x 108 1 x 107

Housing protection IP66 / 67 and IP68 IP 65 IP 67

Connection 5 metres of PVC cable M16 cable gland. M16 cable gland.

Dimensions 65 x 25 x 15 mm 100 x 75 x 40 mm 95 x 65 x 54 mm

Material Resin Aluminium housing AlSi housing

Circuit blackbrownblue

Options Temperature up to +130 °C Temperature up to +130 °C

6.1 Generaluselevelswitch

1 2 3

Page 22: Level indicator - MECON

22 © MECON GmbH BA / MENKAR / EN / 19-10

Type HLS-25i LMS-Ha1E

Function SPDT SPDT

System Reed switch bi-stable Micro switch

Max. rated 250 mA / 1.3 W / 1.3 VA 0.5 A / 20 W / 30 VA

Supply voltage 10 - 30 V 10 - 24 V

Temperature range -25… +100 °C -50… +380 °C

Lifetime 1 x 109 1 x 107

Housing protection IP66 / 67 and IP68 IP 67

Connection 5 metres of PVC cable M20 cable gland. (blue)

Dimensions 80 x 25 x 20 mm 95 x 65 x 54 mm

Material Stainless steel VA4 housing AlSi housing

Circuit blackbrownblue

Options M16 cable gland. (blue)

Approval II 1 GD Exia IIC T6 GaII 1 GD Exia IIC T85°C IP 66/67 Da

Ex i "Simple resource"

6.2 Intrinsicallysafelevelswitches(Exi)

Page 23: Level indicator - MECON

23© MECON GmbH BA / MENKAR / EN / 19-10

Type HLS-25d LMS-HaD

Function SPDT SPDT

System Reed switch bi-stable Micro switch

Max. rated 2.5 A / 60 W / 45 VA 2 A / 40 W / 100 VA

Supply voltage 10 - 230 V 10 - 230 V

Temperature range -25… +100 °C -40… +380 °C

Lifetime 1 x 109 1 x 107

Housing protection IP66 / 67 and IP68 IP66 / 67 and IP68

Connection 5 metres of PVC cable ¾ "NPT or M20 x 1.5 max 1.5 mm2

Dimensions 90 x 25 x 20 mm 130 x 130 x 90 mm

Material Stainless steel VA4 housing Aluminium housing

Circuit blackbrownblue

1 = brown, 2 = blue, 3 = black

Options Stainless steel VA4 housing2 x SPDT

Approval II 2 GD Exd IIC T6 GbII 2 GD Ex tb IIIC T85 °C Db

II 2G Ex d IIC T3..T4 GbII 2D Ex tb IIIC T135 °C..T200 °C Db

6.3 Pressure-safelevelswitches

Page 24: Level indicator - MECON

24 © MECON GmbH BA / MENKAR / EN / 19-10

6.4 Reedchainforcontinuousdisplay

Itispossibletogeneratea4-20mAsignalbyusingareedchain.Thereedchainismount-edasstandardontheentirelengthoftheindicator.

Important!

If necessary, mount the reed chain on the level indicator. The 4 mA setting (0%) is marked on the reed chain and should correspond to the lowest point of the lower process connection.Supply 12 - 30 VDC

Only + and - terminals should be used for the wiring. The other terminals (3, 4, 5 and 6) can only be used for the factory.

Page 25: Level indicator - MECON

25© MECON GmbH BA / MENKAR / EN / 19-10

Type Standard Exi Exd

Transmitter "SMART" type "SMART" type "SMART" type

Approval II 1G Ex ia IIC T4..T6 II 2G Ex db IIC T5..T1 GbII 2D Ex tb T100 °C..T350°C

Supply voltage 8 – 35 VDC 8 – 30 VDC 8 – 30 VDC

Temperature range -50… +350 °C -50… +350 °C -50… +350 °C

Measuring accuracy ± 5 mm ± 5 mm ± 5 mm

Pipe material Stainless steel1.4404 (316)

Stainless steel1.4404 (316)

Stainless steel1.4404 (316)

Max. measuring length 5.5 metres 5.5 metres 5.5 metres

Housing material ABS or aluminium Aluminium or VA Aluminium or stainless steel SS316

Type of protection IP 67 IP 67 IP 66 – IP 68

Connection M16 x 1,5 M20 x 1,5 ¾ "NPT, M20 x 1.5

Signal range 4 - 20 mA / 2 conductors 4 - 20 mA / 2 conductors 4 - 20 mA / 2 conductors

Options Higher accuracy (± 2.5 or ± 1 mm)

M16 x 1.5, M20 x 1.5; ½ "NPT; ¾ "NPT connection

HARD

IP 68 housing

PROFIBUS

FIELDBUS

Stainless steel VA4 housing

Housing with LCD display (also with optical buttons)

Signal range (ohm or V) output

Page 26: Level indicator - MECON

26 © MECON GmbH BA / MENKAR / EN / 19-10

7 SERVICE

7.1 Storage

Storetheinstrumentinadryanddust-freeplace.Keepawayfromdirectcontinuoussunandheat.Avoidexternalloadsbeingplacedontheinstrument.

Thepermissiblestoragetemperaturesforstandard instrumentswithelectricalcompo-nentsare-40°Cto+70°C.

7.2 Maintenance

Attention!

Appropriate safety precautions must be taken when removing the instru-ment from the pipeline. As a rule, new seals must be used when they are being reinstalled in the pipeline!

7.3 Returningtheinstrumenttothemanufacturer

Duetocarefulmanufacturingprocessesandfinalchecksontheinstrument,theMENKAR is expected to operatewithout any troublewhere both installation and operation are carriedoutinaccordancewiththeseinstructions

ShouldyouneverthelessneedtoreturnainstrumenttoMECONGmbHpleaseobservethefollowingpoints:

Attention!

In order to protect the environment and safeguard the health and safety of our personnel, all instruments sent to MECON GmbH for repair must be free of toxic and hazardous substances. This also applies to cavities in these instruments. If necessary, the customer is kindly requested to neutralise or rinse these instruments before returning them to MECON GmbH.The customer must confirm this by filling in an appropriate form which is available for download on the MECON website.

» https://www.mecon.de/en/device-returns/

Page 27: Level indicator - MECON

27© MECON GmbH BA / MENKAR / EN / 19-10

Attention!

If, in spite of these requirements, instruments are returned that contain toxic and dangerous substances, MECON GmbH is entitled to dispose of these substances at the expense of the customer without making any fur-ther enquiries.

7.4 Disposal

Attention!

Comply with the regulations in your country when disposing of instruments

8 FAULTREPORTING

Faultdescription Possiblecauseoffaultandrectification

Magneticindicatordoesnotreact Impuritiesblockingthefloat:Cleaningandcheckingifthefloatis

intact.

Floatdefective:Thefloatmustbereplaced.

Noelectricalsignal Checkelectricalconnectionsaccordingtowiringdiagram.

Page 28: Level indicator - MECON

28 © MECON GmbH BA / MENKAR / EN / 19-10

MECONGmbHRöntgenstr.10550169KerpenGermany

Tel.: +49(0)223760006-0Fax: +49(0)223760006-40E-mail: [email protected]

Web: www.mecon.de