lecture2_massspec

26
Mass spectrometry 8/31/10 hat are the principles behind MS? hat do all MS instruments have in common? hat are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction systems, mass analyzers 3) popular combinations in geosciences JJ Thomson’s cathode ray tube, 1897

Transcript of lecture2_massspec

Page 1: lecture2_massspec

Mass spectrometry 8/31/10

What are the principles behind MS?

What do all MS instruments have in common?

What are the different types of MS? Lecture outline:1) Introduction to mass spectrometry

2) sample introduction systems, mass analyzers

3) popular combinations in geosciences

JJ Thomson’s cathode ray tube, 1897

Page 2: lecture2_massspec

Introduction to Mass Spectrometry

Sample introduction

IonizationMinimize collisions, interferences

Separatemasses

Count ionsCollect results

Nier-type mass spec

Page 3: lecture2_massspec

Basic equations of mass spectrometry

21

2mv zV

2 /F mv R

F Bzv2 /mv R Bzv

2 2/ / 2m z B R V

Ion’s kinetic E function of accelerating voltage (V) and charge (z).

Centrifugal force

Applied magnetic field

balance as ion goes through flight tube

Fundamental equation of mass spectrometry

Combine equations to obtain:

Change ‘mass-to-charge’ (m/z) ratio bychanging V or changing B.

NOTE: if B, V, z constant, then:

r m

Page 4: lecture2_massspec

If: B in gaussr in centimetersm in amuV in voltsz in electronic chargethen….

What magnetic field strength would be required to focus a beam of CO2+ ions on

a collector of a mass spectrometer whose analyzer tube as a radius of 31.45cm,assuming a voltage of 1000V?

Change your magnetic field strength by -10%, what voltage puts the CO2 ionsinto the collector?

V

rBx

z

m 22510825.4

Page 5: lecture2_massspec

Examples of mass spec data output

You can scan in B or V to sweep massesacross a single detector.

OR

You can put different masses intomultiple cups without changing B or V.

Ex: B

Page 6: lecture2_massspec

Sample Introduction Systems (aka “front ends”)

1) Gas source (lighter elements)dual inlet - sample purified and measured with standard gas at identical conditions

precisions ~ ±0.005%continous flow - sample volatized and purified (by EA or GC) and injected into

mass spec in He carrier gas, standards measured before and after,precisions ~ 0.005-0.01%

2) Solid source (heavier elements)TIMS - sample loaded onto Re filament, heated to ~1500°C, precisions ~0.001%laser ablation - sample surface sealed under vacuum, then sputtered with laser

precisions ~0.01%?

3) Inductively coupled plasma (all elements, Li to U)ICPMS - sample converted to liquid form,

converted to fine aerosol in nebulizer,injected into ~5000K plasma torch

Page 7: lecture2_massspec

Ionization occurs in the ‘source’

Electron Ionization

Gas stream passes through beam of e-,positive ions generated.

Thermal IonizationPlasma: Gas stream passes through plasmamaintained by RF current and Ar.

Themal: Filament heated to ~1500C

Page 8: lecture2_massspec

Mass Analyzers - the quadrupole vs. magnetic sector

Quadrupole:Changes DC and RFvoltages to isolatea given m/z ion.PRO: cheap, fast, easy

Magnetic Sector:Changes B and V to focusa given m/z into detector.PRO: turn in geometry means

less ‘dark noise’,higher precision,

Page 9: lecture2_massspec

Two types of ion detectors

A) Faraday collector - long life, stable, for signals > 2-3e6 cps

B) Electron multiplier - limited life, linearity issues, high-precision, signals < 2e6 cps

Page 10: lecture2_massspec

Popular combinationsGas source1) Dual inlet isotope ratio mass spec (at GT, Lynch-Steiglitz and Cobb)

- O, C, H ratio analyses

2) Elemental analyzer IRMS (at GT, Montoya)- N, C, S ratio analyses

3) Gas chromatograph IRMS (at GT, ????)- compound-specific ratio analyses

Solid source1) Thermal Ionization mass spec (multi-collector)

- heavy metals, REE

ICP1) ICP quadrupole mass spec (at GT, Taillefert)

- trace metal analysis

2) Single collector magnetic sector ICPMS- higher-precision trace metal analysis

2) Multi-collector ICPMS- U/Th dating, TIMS replacement

Micromass IsoProbe - MC-ICPMS

Page 11: lecture2_massspec

Inductively Coupled Plasma Mass Spectrometry

sample cone

skimmer cone

instrument housing

mass/chargediscriminator

detector

atmosphericpressure

“fore” vacuum10-4 bar

high vacuum10-7 bar

Shared componentsof all ICPMS machines

Page 12: lecture2_massspec

or magneticsector

Faraday cupand ion counter (electron multiplier)

1. Quadrupole ICPMS- measure concentrations

as low as several ppt

- no fuss sample preparation(dissolve in 5% HNO3)

- get beam intensityvs. mass/charge ratio

Page 13: lecture2_massspec

2. High resolution ICPMSaka double-focusing ICPMSaka magnetic sector ICPMS

- same front end as Q-ICPMS

- combines magnet welectrostatic analyzer

electrostaticanalyzer

separatesions by charge

magnetseparates

ion by mass

Faraday cupand EM

High-resolution ICPMS

Page 14: lecture2_massspec

Multi-collector ICPMS

3. MC-ICPMS

- same front end as other ICPMS

- same magnet and ES as HR-ICPMS

- multiple detectors spaced 1amuapart enable simultaneous

measurement of many (~7) isotopes

-good for what kinds of systems?

Page 15: lecture2_massspec

56Fevery low concentrations

in environmental samples,but high interest (why?)

Unfortunately, 56Fe has thesame atomic wt as ArO

(40Ar+16O)

Quadrupole measurement =INTERFERENCE!

Low vs. High – resolution ICPMS and Interferences

HR-ICPMS measurement =can distinguish 56Fe from ArO

NOTE: most elements can bedistinguished with a lowresolution quadrupole

Page 16: lecture2_massspec

The importance of standards in mass spectrometry

ICPMS: Can determine concentration to ~1% R.E. using calibration curve (below)

Can monitor Sensitivity (signal response for givensolution concentration) over time

unknown sample = 8.2e7 cps,

conc ~ 10.5ppb

Page 17: lecture2_massspec

REMEMBER: all mass spectrometers are “black boxes” we really have no idea what goes on from sample container to detector signal

Ex: you measure a count-rate of 10,000 cps for a given element, but you need to know how many atoms of that element, or its concentration, were in your sample

- measuring isotope ratios is a powerful approach because we can measuresamples against standards with known isotopic ratios (it’s much more difficult to change a material’s isotopic ratios than it is to change its elemental concentration!)

- isotope dilution takes advantage of ability to precisely measure ratios

- ALL measurements need to include blanks and standards (either concentration or ratio standards)

Page 18: lecture2_massspec

Isotope dilution principle

Isotope dilution is an analytical technique used in combination with mass spectrometryto determine the concentration of element x in unknown samples.

ex: Rb

A known amount of “spike” with known elemental concentration

and isotopic abundances(what’s the diff?)

is added to sample with unknownelemental concentration butknown isotopic abundances.

Requirements: 1) The sample has natural (or known) isotopic abundance (usually true).2) The spike and sample isotopic ratios are different.

Page 19: lecture2_massspec

More Commonly used ICPMS terms

Nebulization efficiency – the amount of solution that reaches the plasma (~1%)- varies with sample matrix

- surface tension, viscosity, and density of solution will affect neb. eff.- usually all standards, spikes, and samples are introduced as 2-5% HNO3

- an acid solution reduces complexation, surface adsorption

Matrix effects – the changes in ICP characteristics with variable matrices- largely black box (we see these effects, cannot wholly explain/predict them)

- you must carefully match the matrices of your standards/samples to obtain quantitative results

Ionization efficiency – the amount of ions produced per atoms introduced- depends on matrix, focusing of beam through cones, lenses

- usually no better than 1/1000

Page 20: lecture2_massspec

ICP-OES ICP-MS

Detection limit – defined as 3 x the S.D. of the signal as the concentration of the analyte

approaches 0 (measure stability at a variety of conc’s, extrapolate to 0; or measure

5% HNO3 blank solution)

ICP detection limits for a variety of elements

Page 21: lecture2_massspec

Ion microprobe(orSecondaryIonMassSpectrometry SIMS)

-use an ion beam (usually Cs+1) to “sputter” a sample surface; secondary ions fed into mass spec

20μm

Page 22: lecture2_massspec

Accelerator Mass Spectrometry

The AMS at University of Arizona (3MV)-prior to AMS samples were 14C-dated by counting the number of decays

- required large samples and long analysis times

-1977: Nelson et al. and Bennett et al. publish papers in Science demonstratingthe utility of attaching an accelerator to a conventional mass spectrometer

The AMS at LLNL (10MV)

Principle:You cannot quantitatively remove interferring ions to look for one 14C atom among several

quadrillion C atoms.Instead, you

a) destroy molecular ions (foil or gas)b) filter by the energy of the ions (detector)

to separate the needle in the haystack.

Page 23: lecture2_massspec

a) ION SOURCEgenerates negative

carbon ionsby Cs sputtering

b) INJECTOR MAGNETseparates ions by mass,

masses 12, 13, and 14 injected

http://www.physics.arizona.edu/ams/education/ams_principle.htm

c) ACCELERATORgenerates 2.5 million volts,

accelerates C- ions

d) TERMINALC- ions interact with

‘stripper’ gas Ar,become C+ ions,

molecular species CHdestroyed

e) ELECTROSTATIC DEFLECTORspecific charge of ions selected (3+)

f) MAGNETIC SEPARATION13C steered into cup, 14C

passes through to solid detector

g) Si BARRIER DETECTORpulse produced is proportional to the energy of ion, can

differentiate b/t 14C and other ions count rate for modern sample = 100cps

Page 24: lecture2_massspec

Hurdles in mass spectrometry

1) Abundance sensitivity - ratio of signal at massm to signal at m+1

- better with better vacuum- acceptable values: 1-3ppm at 1amu

2) Mass discrimination

- heavier atoms not ionized as efficiently as light atoms

- can contribute 1% errors to isotope values

- can correct with known (natural) isotope ratios within run, or with known standards between runs

Page 25: lecture2_massspec

3) Dark Noise - detector will register signal even without an ion beam- no vacuum is perfect

and- no detector is perfect

- must measure prior to run to get “instrument blank” if needed

4) Detector “gain” - what is the relationship between the electronic signal recorded by the detector and the number of ions that it has counted?- usually close to 1 after factory calibration- changes as detector “ages”- must quantify with standards

Cardinal rule of mass spectrometry:Your measurements are only as good as your STANDARDS!

Standards (both concentration and isotopic) can be purchased from NIST

Hurdles in mass spectrometry (cont.)

Page 26: lecture2_massspec

Ex: NBS-19, O, C carbonate isotopic standard