Lec 4_Fetal Heart Rate Monitoring Review
Transcript of Lec 4_Fetal Heart Rate Monitoring Review
-
7/31/2019 Lec 4_Fetal Heart Rate Monitoring Review
1/7
Fetal heart rate monitoring (From Obstetrics, MD Consult E book)
- goal of intrapartum monitoring is to detect hypoxia in labor and try nonoperativeinterventions to correct it
o detect severity and duration of hypoxiais C section necessary?
o Many fetuses develop intermittent hypoxia which never progresses tometabolic acidosis need to intervene b4 tissue damage or fetal death
occurs- External monitoring is done mainly from Doppler (US)
o Computer can eliminate most noise and artifacts
o Done when membrane are intact and shouldnt be ruptured or fetal skin
should not be punctured (in instances to decrease vertical transmission of
disease to fetus like if mom is infected with HIV, hep C, HSV)
o Contractions are monitored externally via tocodynamometer
tocodynamometer depicts the frequency of the contractions accurately, butthe strength of the contractions only relatively, since it cannot measure
actual intrauterine pressure. In addition, the apparent duration of thecontraction varies with the sensitivity of the monitor, which is negativelyaffected by variables such as maternal obesity and premature gestationalage
o In general, when the FHR is reassuring and there is an adequate tracing and when
the progress of labor is adequate, the external devices are fine. When betterquality FHR monitoring is required or it becomes important to accurately assessuterine contraction duration and intensity, then internal device(s) may benecessary.
o
- Internal monitoring is indicated to get high quality, accurate, continuous FHR
tracing
o Particularly important if mom is obese, premature fetus or very active fetus
(hard to get good Doppler reading)
o Uses spiral electrode which amplifies signal
-
7/31/2019 Lec 4_Fetal Heart Rate Monitoring Review
2/7
o Note: if fetus is dead, machine may amplify maternal ECG signals which
can be misinterpreted as fetal bradyo Contractions can be more accurately monitored via an intrauterine pressure
catheter than tocodynamometer
o
- FHR monitoringo under normal circumstances during labor, the only variable that alters fetal
oxygenation is the temporary interruption in blood flow to the placenta that occursas a result of the compression of the spiral arteries by the wall of the uterus at thepeak of the contraction
o Under normal circumstances, the fetus tolerates these periods of stasis well
without a significant change in its oxygen content. Contractions that are unusuallylong or unusually strong may, however, result in transient periods of fetalhypoxemia
o
Fetal perfusion also depends on uterine perfusion: laboring women can developsupine hypotension due to fetal compression of vena cava, redistribution ofmaternal blood flow from the uterus with regional anesthesia, maternalhemorrhage (placenta previa or abruptio placentae), microvascular disease (HTN,preeclampsia/eclampsia, diabetic vasculopathy, postmaturity)
o While the placenta functions as the fetal lung, the umbilical cord functions as its
trachea, leading oxygen to the baby and carbon dioxide (CO2 ) away. Alteration inumbilical cord blood flow is a very common occurrence during labor, either fromdirect compression or from stretch
Direct: compression from babys body against uterine wall
Stretch as baby enters birth canal
- Tachycardiao baseline FHR is typically between 120 and 160 bpm- the baseline heart rate is
largely a function of vagal activity (as baby ages, increased vagal activity=decreased heart rate- premies have higher HR than full term babies)
o two most common causes of tachycardia are maternal fever and drugs that directly
raise the fetal heart rate
The FHR rises approximately 10 bpm for each 1F increase in maternaltemperature
Drugs that elevate the fetal heart rate fall into one of two categories:vagolytic and -sympathomimetic but these drugs rarely raise FHR above160
-
7/31/2019 Lec 4_Fetal Heart Rate Monitoring Review
3/7
Other less common causes of fetal tachycardia include fetalhyperthyroidism, fetal anemia, fetal heart failure, and fetaltachyarrhythmias. As fetal hypoxia becomes progressively worse andpersists over time, fetal tachycardia often develops.
- Bradycardiao FHR less than 120 bpm is termed a bradycardia
o baseline bradycardia is usually innocuous, whereas a prolonged deceleration toless than 120 bpm lasting more than 60 to 90 seconds may often indicatesignificant fetal hypoxia
o range of 90 to 120 bpm, a bradycardia may often be a normal variant, and these
fetuses are usually bradycardic after birth, but are otherwise well oxygenated andnormal
o maternal hypothermia may cause fetal bradycardia. This is commonly seen with
patients on magnesium sulfate (MgSO4 ) who are vasodilated and has also beendescribed with maternal hypoglycemia and hypothermia, many drugs can alsocause fetal brady.
o A fetal baseline heart rate in the range of 80 bpm or less, especially with reduced
variability, may be caused by a complete heart blocko a patient is admitted with a baseline bradycardia, one should also consider the
possibility of maternal heart rate being recorded with a dead fetus
- Variabilityo differences in heart rate from beat to beat are recorded as "variability" reflected
visually as a line that fluctuates above and below the baseline. This variability is areflection of neuromodulation of the FHR by an intact and active CNS and alsoreflects normal cardiac responsiveness
o Short-term variability is the beat-to-beat irregularity in the FHR and is caused by
the difference in rates between successive beats of the FHRo Long-term variability is the waviness of the FHR tracing, and is generally seen in
three to five cycles per minute.o When the fetus is alert and active, the FHR variability is normal or increased.
When the fetus is obtunded, due to whatever cause, the variability is reduced.
Since severe hypoxia, especially when it reaches the level of metabolic acidosis,will always depress the CNS, normal variability reliably indicates the absence ofsevere hypoxia and acidosis
o there are many things that cause the CNS to be depressed- not just hypoxia or
metabolic acidosis- include fetal sleep cycles; drugs, especially CNS depressants;fetal anomalies, especially of the CNS; and previous insults that have damagedthe fetal brain
o interpretation of variability is both subjective and nonspecific
- Decelerationso Early decelerations are shallow, symmetric, uniform decelerations with onset and
return that are gradual, resulting in a -shaped decelerationo They begin early in the contraction, have their nadir coincident with the peak of the
contraction, and return to the baseline by the time the contraction is overo They are thought to be caused by compression of the fetal head by the uterine
cervix as it overrides the anterior fontanel of the cranium
o They do not indicate fetal hypoxia and are only significant in that they may be
easily confused with late decelerations because of their similar shape and depth
-
7/31/2019 Lec 4_Fetal Heart Rate Monitoring Review
4/7
o
- Late decelerationso similar in appearance to early decelerations. They too are of gradual onset and
return, -shaped, and generally descend below the baseline no more than 30 to40 bpm, in contrast to early decelerations, late decelerations are delayed in timingrelative to the contraction. They begin usually about 30 seconds after the onset ofthe contraction or even at or after its peak. Their nadir is after the peak of thecontraction. FHR variability may be unchanged or even increased during thedecelerations
o
Late decelerations are generally said to be caused by "uteroplacentalinsufficiency."
o is generally agreed that the depth of the late deceleration cannot be used to judge
the severity of the hypoxia. Because of the mechanisms causing these changes,late decelerations always indicate fetal hypoxia
o Excessive uterine contractions, usually seen with oxytocin, are the single most
common cause of late decelerations. In these situations, the duration ofinterruption of uterine blood flow is prolonged and the hypoxia is more than thenormal fetus can endure, and late decelerations are expressed. The most commonpathologic conditions of the placenta associated with late decelerations are thosecharacterized by either microvascular disease in the placenta or local vasospasmcompromising blood flow and thus exchange. Common causes includepostmaturity, maternal hypertension (chronic hypertension or preeclampsia),
collagen vascular diseases, and diabetes mellitus in its more advanced stages.Severe maternal anemia or maternal hypoxemia may compromise oxygen deliveryand result in late decelerations. Conversely, chronic fetal anemia may diminishfetal oxygen uptake and be associated with late decelerations.
o metabolic acidosis is usually not mixed with a respiratory acidosis (CO2 is usually
eliminated despite there being hypoxia) . The only common exception to this iswith abruptio placentae, where CO2 retention is seen with late decelerations
- Variable decelerations
-
7/31/2019 Lec 4_Fetal Heart Rate Monitoring Review
5/7
o most common type of decelerations seen in the laboring patient
o synonymous with umbilical cord compression, and anything that results in the
interruption of blood flow within the umbilical cord
two most common causes of late decelerations in labor are excessiveuterine contractions (usually caused by oxytocin) and maternalhypotension
o They are variable in all ways: size, shape, depth, duration, and timing relative tothe contraction. The onset is usually abrupt and sharp.
decelerations are often recognized by the accelerations that precede andfollow the decelerations.
o The depth and duration are proportional to the severity and duration of interruption
of cord blood flow.o duration of a variable deceleration should be limited to 2 minutes, and that beyond
2 minutes, it should be called a prolonged deceleration
-
7/31/2019 Lec 4_Fetal Heart Rate Monitoring Review
6/7
- Prolonged decelerationso isolated decelerations lasting 90 to 120 seconds or more
o those lasting 2 to 10 minutes and that beyond 10 minutes this is a baseline changeo decelerations may be caused by virtually any of the mechanisms previously
described, but usually are of a more profound and sustained nature. Prolongedumbilical cord compression, profound placental insufficiency, or possibly evensustained head compression may lead to prolonged decelerations
o presence of and severity of hypoxia are thought to correlate with the following
variables with a prolonged deceleration: the depth and duration of the deceleration,how abruptly it returns to baseline, how much variability is lost during thedeceleration, and whether there is a rebound tachycardia and loss of variabilityfollowing the deceleration
o more profound stimuli that may result in this type of deceleration include prolapsed
umbilical cord or other forms of prolonged cord compression, prolonged uterinehyperstimulation, hypotension following conduction anesthesia, severe degrees of
abruptio placentae, paracervical anesthesia, an eclamptic seizure, and rapiddescent through the birth canal
- accelerationso periodic changes of the FHR above the baseline, usually lasting no longer then 30-
90 seconds
o all accelerations are a physiologic response to fetal movement- absence of
accelerations means only that the baby is not moving
o in fetuses having otherwise nonreassuring FHR patterns, the presence of
accelerations, using this definition, virtually always ruled out a pH less than 7.20 onscalp sampling. If there is no acceleration in the face of an otherwisenonreassuring FHR, most studies have shown that about 50 percent of thesefetuses have an acidotic pH value on scalp sampling. However, the absence ofaccelerations in a fetus without a nonreassuring deceleration pattern rarely
indicates fetal acidosis.
-
7/31/2019 Lec 4_Fetal Heart Rate Monitoring Review
7/7
- Sinusoidal patternso Rare but BAD
o strongly associated with fetal hypoxia, most often seen in the presence of severe
fetal anemiao criteria for identifying a sinusoidal FHR include (1) a stable baseline FHR of 120 to
160 bpm with regular sine wave-like oscillations, (2) an amplitude of 5 to 15 bpm,(3) a frequency of 2 to 5 cycles/min, (4) fixed or absent short-term variability, (5)oscillation of the sine wave above and below the baseline, and (6) absence ofaccelerations
o note: the below is not actually sinusoidal pattern, but actually pseudosinusoidal but
it is the oly pic I could get from this text book- pseudosinusoidal waves look likesinusoidal patterns but do not meet all 6 criteria for the sinusoidal pattern
-