LC-110/H Reference Manual

Transcript of LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text...

Page 1: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

LC-110/H Reference Manual

Page 2: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2 Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.3 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

2. Calibrator Interface and Operation . . . . . . . . . . . . . . . . . . . . . 52.1 Milliamp Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.2 Milliamp Simulate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72.3 Milliamp Measure without 24V power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82.4 Milliamp Measure with 24V power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92.5 Volts Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

3. Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

4. Calibrator Setup Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114.1 Setting Auto Ramp Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114.2 Setting Auto Step Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124.3 Enabling Valve Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124.4 Enabling the HART Resistor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134.5 Selecting mA Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134.6 Contrast Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134.7 Configuring Auto Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144.8 HART Write Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

5. Advanced Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155.1 Step and Ramp Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155.2 Valve Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165.3 HART 250Ω Resistor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

6. HART Device Communications . . . . . . . . . . . . . . . . . . . . . . . . 176.1. HART Connection Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176.2 Communications Setup and Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196.3 Connecting to a HART Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216.4 Disconnect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236.5 Function Select Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236.6 Display Setup & Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236.7 Modify Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246.8 Trim, Set, & Zero Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286.9 Device Diagnostic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326.10 Data Log & Config Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 336.11 HART Modem Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 386.12 Error Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

7. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 457.1 Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 457.2 Cleaning the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 457.3 Fuse Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

8. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

9. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Page 3: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

1

IntroductionThe Martel LC-110 and LC-110H are mA (loop) calibrators designed to be simple to use and to offer the highest accuracy of any loop calibrator in their class. The LC-110H differs from the standard LC-110 in that it incorporates HART communications and supports a select subset of the HART universal and common practice commands. This unique feature allows the LC-110H to be used as both a loop calibrator and basic function communicator.

The calibrator supports the following functions:

•Currentmeasurementandsourcing,includingselectable24Vsupply

•Voltagemeasurement

•Valvetestcapability

•SelectableHART250Ω loop resistor

•Outputstepandramp

The calibrator has the following features:

•Largeeasytoreaddisplayformeasurementsanddataentry

•Digitalknobwithselectabledecadecontrolforquickandeasydata entry

•Interactivemenus

•Inputandoutputjacksprotectedbyselfresettingfusesprovidingprotectionto240VAC

Inthecommunicatormodetheuserisabletoreadbasicdeviceinformation,performdiagnostictests,andtrimthecalibrationonmostHARTenabledtransmitters.Inthepastthiscouldonlybedonewithadedicated communicator.

TheLC-110HcanalsobeusedasaHARTmodem,allowingPChostapplications to perform more sophisticated diagnostics and setup. HARTformatcommandsfromthehostarepassedtothetransmitter,andresponsesarepassedback.

1.1 Customer ServiceCorporate Office:

www.martelcorp.come-mail: [email protected]

Tel:(603)434-1433800-821-0023Fax:(603)434-1653Martel Electronics 3CorporateParkDrive Derry,NH03038

Page 4: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

2

1.2 Standard EquipmentChecktoseeifyourcalibratoriscomplete.Itshouldinclude:

•Calibrator

•6xAAAbatteries(installed)

•Testleads

- The LC-110 is supplied with one set of leads with alligator clips installed

-TheLC-110Hissuppliedwithtwosetsoftests.Onesetwith alligator clips and the other set with mini-grabber style ends

•Owner’sManual

•NISTCertificate

1.3 Safety information

Symbols Used

ThefollowingtableliststheInternationalElectricalSymbols.Someorall of these symbols may be used on the instrument or in this manual.

SymbolDescription

AC (Alternating Current)

AC-DC

Battery

CEComplieswithEuropeanUnionDirectives

DC

DoubleInsulated

ElectricShock

Fuse

Page 5: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

3

PEGround

HotSurface(BurnHazard)

ReadtheUser’sManual(ImportantInformation)

Off

On

CanadianStandardsAssociation

The following definitions apply to the terms “Warning” and “Caution”.

•“Warning”identifiesconditionsandactionsthatmayposehazardsto the user.

•“Caution”identifiesconditionsandactionsthatmaydamagetheinstrument being used.

Usethecalibratoronlyasspecifiedinthismanual,otherwiseinjuryand damage to the calibrator may occur.

WarningTo avoid possible electric shock or personal injury:

•Donotapplymorethantheratedvoltage.Seespecificationsforsupported ranges.

•Followallequipmentsafetyprocedures.

•Nevertouchtheprobetoavoltagesourcewhenthetestleadsareplugged into the current terminals.

•Donotusethecalibratorifitisdamaged.Beforeyouusethecalibrator,inspectthecase.Lookforcracksormissingplastic.Payparticular attention to the insulation surrounding the connectors.

•Selecttheproperfunctionandrangeforyourmeasurement.

•Makesurethebatterycoverisclosedandlatchedbeforeyouoperate the calibrator.

•Removetestleadsfromthecalibratorbeforeyouopenthebatterydoor.

Page 6: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

4

•Inspectthetestleadsfordamagedinsulationorexposedmetal.Checktestleadscontinuity.Replacedamagedtestleadsbeforeyou use the calibrator.

•Whenusingtheprobes,keepyourfingersawayfromtheprobecontacts. Keep your fingers behind the finger guards on the probes.

•Connectthecommontestleadbeforeyouconnectthelivetestlead.Whenyoudisconnecttestleads,disconnectthelivetestleadfirst.

•Donotusethecalibratorifitoperatesabnormally.Protectionmaybeimpaired.Whenindoubt,havethecalibratorserviced.

•Donotoperatethecalibratoraroundexplosivegas,vapor,ordust.

•Disconnecttestleadsbeforechangingtoanothermeasureorsource function.

•Whenservicingthecalibrator,useonlyspecifiedreplacementparts.

•Toavoidfalsereadings,whichcouldleadtopossibleelectricshockorpersonalinjury,replacethebatteryassoonasthebatteryindicator appears.

•Toavoidaviolentreleaseofpressureinapressurizedsystem,shutoff the valve and slowly bleed off the pressure before you attach the pressure module to the pressure line.

CautionTo avoid possible damage to calibrator or to equipment under test:

•Usetheproperjacks,function,andrangeforyourmeasurementorsourcing application.

Page 7: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

5

2. Calibrator Interface and Operation

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

Source 0.00%

4.000mA

Simultaneous % Display4mA = 0.00%20mA = 100.00%

Cursor

100% Key: Sets Output to 20mA

25% Key: Steps 4, 8, 12, 16, 20mA

0% Key: Sets Output to 4mA

Digital Knob Sets Output ValuePress to Move Cursor

MenuPress to Enter Setup MenuPress Again to Exit

Step / Ramp Enable

Backlight

On / Off

Mode

Figure 1

Figure1istypicalofthemaindisplayforalloperatingmodes.Itshows the present measurement or output value. The individual operating modes are described below.

UsetheMENU/EXITkeytodisplaythemainmenudescribedinsection3toselecttheprimaryoperatingmodeofthecalibrator,toaccessthecalibratorsetupfunctions,ortoenterHARTmode.

Inoutputmode:

•Presstheknobtomovethedecadecursor.

•Rotatetheknobtoincrementordecrementtheoutputinstepsindicated by the selected decade.

•Usethe3percentkeystodirectlysettheoutputtopresetvalues.

•Usethestep/rampkeytoselectandstoptheseadvancedmodeswhicharedescribedinsection5.1.

Page 8: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

6

2.1 Milliamp SourceInthemASourcemodethecalibratoroutputsasignalfrom0to24mAintoaloadofupto1000ohms(750ohmsiftheinternalHARTresistor is switched on).

Figure2showsthemaindisplayandtypicalconnectionsforthismode.

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

Source 0.00%

4.000mA

4-20mA INPUT DEVICERECORDER/INDICATORSCADA SYSTEM ETC.

(+)

(–)

4 to 20mA

Figure 2

•Presstheknobtomovethedecadecursor.

•Rotatetheknobtoincrementordecrementtheoutputinstepsindicated by the selected decade.

•Usethe3percentkeystodirectlysettheoutputtopresetvalues.

•Usethestep/rampkeytoselectandstoptheseadvancedmodeswhicharedescribedinsection5.1.Whenautomaticsteporrampis active one of the following is shown in the lower left corner:

Automatic step:

Automatic ramp: ∧

Page 9: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

7

•PressMENU/EXITtoenterthemainmenudescribedinsection3.

•Thetext‘ValveTest’isshowninthelowercenterwhenthevalvetestfunctionhasbeenenabledontheSetupMenu,section4.3.Valvetestisdescribedinmoredetailinsection5.2.

•Thetext‘250Ω’textisshowninthelowerrightcornerwhentheHARTresistorhasbeenenabledontheSetupMenu,section4.4.

2.2 Milliamp SimulateInthemASimulatemodethecalibratorfunctionslikea2-wiretransmitterbycontrollingtheloopcurrentfromanexternalpowersupply. This function is a great way to test a loop with the transmitter removed.

Figure3showsthemaindisplayandtypicalconnectionsforthismode.

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

Simulate 0.00%

4.000mA

POWER SUPPLY30 VDC MAX

(+)

(–)

4 to 20mA

4-20mAINPUT DEVICE

(+)

(–)

Figure 3

•Presstheknobtomovethedecadecursor.

•Rotatetheknobtoincrementordecrementtheoutputinstepsindicated by the selected decade.

•Usethe3percentkeystodirectlysettheoutputtopresetvalues.

•Usethestep/rampkeytoselectandstoptheseadvancedmodeswhicharedescribedinsection5.1.Whenautomaticsteporrampis active one of the following is shown in the lower left corner:

Page 10: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

8

Automatic step:

Automatic ramp: ∧

•PressMENU/EXITtoenterthemainmenudescribedinsection3.

•Thetext‘ValveTest’isshowninthelowercenterwhenthevalvetestfunctionhasbeenenabledontheSetupMenu,section4.3.Valvetestisdescribedinmoredetailinsection5.2.

•Thetext‘250Ω’textisshowninthelowerrightcornerwhentheHARTresistorhasbeenenabledontheSetupMenu,section4.4.

2.3 Milliamp Measure without 24V powerInthemAMeasuremode,thecalibratordisplaystheloopcurrent.

Figure4showsthemaindisplayandtypicalconnectionsforthismode.

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

Measure 0.00%

4.000mA

4-20mA OUTPUT DEVICE

(+)

(–)

4 to 20mA

Figure 4

•PressMENU/EXITtoenterthemainmenudescribedinsection3.

•Thetext‘250Ω’textisshowninthelowerrightcornerwhentheHARTresistorhasbeenenabledontheSetupMenu,section4.4.

Page 11: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

9

2.4 Milliamp Measure with 24V powerInthemAMeasurewith24Vmode,thecalibratoroutputs24VDCwhile displaying the loop current. The mode is useful for powering a transmitter without the need for a separate power supply.

Figure5showsthemaindisplayandtypicalconnectionsforthismode.

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

Measure 24V 0.00%

4.000mA

Figure 5

•PressMENU/EXITtoenterthemainmenudescribedinsection3.

•Thetext‘250Ω’textisshowninthelowerrightcornerwhentheHARTresistorhasbeenenabledontheSetupMenu,section4.4.

2.5 Volts MeasureIntheVoltsMeasuremode,thecalibratordisplaystheloopvoltage.

Figure6showsthemaindisplayandtypicalconnectionsforthismode.

•PressMENU/EXITtoenterthemainmenudescribedinsection3.

Page 12: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

10

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

Measure

10.000V

VOLTAGE OUTPUT DEVICE

(+)

(–)

Figure 6

3. Main MenuThe Main Menu is used to select the primary operating mode of the calibrator,toaccessthecalibratorsetupfunctions,andtoenterHARTmode.

mASource mASimulate mA Measure mAMeasurewith24V VoltsMeasure CalibratorSetupMenu HART Comm. Menu

Rotatetheknobtoselectanactionbymovingthereversevideohighlightupanddown.Presstheknobtoperformtheselectedaction.PressMENU/EXITtoreturntothemaindisplaywithoutperforming an action.

The first five actions change the operating mode accordingly and returntothemaindisplaydescribedinsection2.

The“CalibratorSetupMenu”actionisdescribedinsection4.

The “HART Comm. Menu” action is described in section 6. The ‘HART’menuitemonlyappearsontheLC-110Hdevice.

Page 13: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

11

4. Calibrator Setup MenuThe calibrator setup menu consists of two screens. The second screenisreachedbyselecting‘OtherParameters’onthefirstscreen.

Auto Ramp Time AutoStepTime ValveTest HART250Ω Resistor OtherParameters

mASpan Contrast AutoShutdownTime HARTWriteEnable

Rotatetheknobtoselectanactionbymovingthereversevideohighlightupanddown.Presstheknobtoperformtheselectedaction.PressMENU/EXITtoreturntothemaindisplaywithoutperforming an action.

Note:The‘HARTWriteEnable’menuitemonlyappearsontheLC-110H device.

4.1 Setting Auto Ramp Time

Auto Ramp Time

10SEC.Press&holdtoSave

This function sets the full scale ramp time for the mA ramp feature describedinmoredetailinsection5.1.Thevaluecanbesetfrom5to 300 seconds.

Rotatetheknobtoadjustthevalue.Pressandholdtheknobtosaveit.PressMENU/EXITtorestorethepreviousvalueandreturntothemain display.

Page 14: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

12

4.2 Setting Auto Step Time

Auto Step Time

10SEC.Press&holdtoSave

ThisfunctionsetsthestepintervaltimeforthemAAutoStepfeaturedescribedinmoredetailinsection5.1.Thevaluecanbesetfrom5to 300 seconds.

Rotatetheknobtoadjustthevalue.Pressandholdtheknobtosaveit.PressMENU/EXITtorestorethepreviousvalueandreturntothemain display.

4.3 Enabling Valve Test

Valve Test Enable

OFFON

This function enables and disables the valve test feature described in moredetailinsection5.2.

Rotatetheknobtomovethereversevideohighlighttothedesiredselection.Presstheknobtosavethehighlightedselection.PressMENU/EXITtorestorethepreviousselectionandreturntothemaindisplay.

Page 15: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

13

4.4 Enabling the HART Resistor

HART Resistor Enable

OFFON

This function enables and disables the HART resistor described in moredetailinsection5.3.

Rotatetheknobtomovethereversevideohighlighttothedesiredselection.Presstheknobtosavethehighlightedselection.PressMENU/EXITtorestorethepreviousselectionandreturntothemaindisplay.

4.5 Selecting mA Span

mASpanSelection 4to20mA 0to20mA

This function selects the mA span used to calculate the percent of spanfieldonthemaindisplay,andtosetthevalueusedbythe0%key.

Rotatetheknobtomovethereversevideohighlighttothedesiredselection.Presstheknobtosavethehighlightedselection.PressMENU/EXITtorestorethepreviousselectionandreturntothemaindisplay.

4.6 Contrast Adjustment

Contrast

Samplenormalvideo

Samplereversevideo

This function sets the display contrast.

Rotatetheknobtoadjustthecontrast.Therangeofvaluesisshownbythebargraph,withhighercontrastshownbyalongerbar.Thesamplenormalandreversevideotextlinesallowevaluationofboth

Page 16: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

14

textmodes.Presstheknobtosavethecontrastvalue.PressMENU/EXITtorestorethepreviousselectionandreturntothemaindisplay.

4.7 Configuring Auto Shutdown

Auto Shutdown Time

10MIN.

This function sets or disables the time before the unit automatically shutsdownifthekeypadisnotused.Thevaluecanbesetto‘Disabled’,orfrom1to30minutes.

Rotatetheknobtoadjustthevalue.Pressandholdtheknobtosaveit.PressMENU/EXITtorestorethepreviousvalueandreturntothemain display.

4.8 HART Write EnableThissettingisavailableonlyontheLC-110H.ItprotectsallfunctionsontheModifySetupmenu,andtheDeviceDiagnostic,Trim4mA,Trim20mA,SetFixedOutput,andPVZerofunctions.Thedefaultsettingis“ON”,butmaybeturnedoffiftheuserwantstoprotectagainstunauthorizeduse.

HART Write Enable

OFFON

Rotatetheknobtomovethereversevideohighlighttothedesiredselection.Presstheknobtosavethehighlightedselection.PressMENU/EXITtorestorethepreviousselectionandreturntothemaindisplay.

Beforethesettingischanged,apasswordisrequiredtoacceptit.Thepasswordissetto617atthefactory.Enterthepasswordonthefollowing display. The range of values that can be set is 000 to 999.

Page 17: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

15

Password

000 Press&holdtoSave

Presstheknobtomovethedecadecursor.Rotatetheknobtoadjustthevalueinstepsindicatedbythedecadeselected.Pressandholdtheknobtosavethewriteenablesetting.PressMENU/EXITtorestore the previous HART enable selection and return to the main display.

The write enable setting is saved only when the correct password is entered.Otherwiseanerrormessageisdisplayed.

5. Advanced FeaturesThe calibrator has several advanced features that are available throughtheCalibratorSetupMenus.Howthefeaturesareenabledand what they do is described below.

5.1 Step and Ramp OperationThe calibrator has several unique features that are helpful aids when doing milliamp calibrations.

Thepercentkeysallowtheusertosetthemilliampoutputto0%ofspan,100%ofspan,orstepitby25%ofspan.

The“Step/Ramp”keyallowshandsfreeoperationbyautomaticallysteppingorrampingthemilliampoutputfrom0%to100%andbackcontinuously.

The100%valueisalways20mA,butthe0%valuemaybe0mAor4mAdependingonthemAspansettingdescribedinsection4.5.The25%stepsizeiseither5mAor4mAaccordingly.

5.1.1 Manual Stepping

1. Usethemainmenutosetthecalibratortosourceorsimulatecurrent.

2. Pressthe“0%”keytosettheoutputto0%ofspan.

3. Pressthe“100%”keytosettheoutputto100%ofspan.

4. Pressthe“25%”keytosteptheoutputin25%ofspanincrements,from0%ofspanto100%ofspanandback.

Page 18: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

16

5.1.2 Automatic Stepping and Ramping

1. Usethemainmenutosetthecalibratortosourceorsimulatecurrent.

2. ThecalibratorhasseparateAutorampandsteptimes.Usethemenu system to set the ramp or step time as described in section4.1or4.2.

3. Pressthe“Step/Ramp”keyoncetocontinuallysteptheoutputfrom0%ofspanto100%ofspanandbackinincrementsof25%of span at the specified interval.

Pressthe“Step/Ramp”oncemoretogotoautoramp.

Pressoneofthepercentkeys,orthe“Step/Ramp”twicemore,toturn off the auto step and ramp.

4. Pressthe“Step/Ramp”keytwicetocontinuallyramptheoutputfrom0%ofspanto100%ofspanoverthespecifiedinterval,andthenbackoverthespecifiedinterval.

Pressoneofthepercentkeys,orthe“Step/Ramp”oncemore,toturn off the auto step and ramp.

5.2 Valve TestThe calibrator has a valve test mode for verifying proper operation of valves.Invalvetesttheusercansteptheoutputtothefollowingvalues:3.8mA,4.0mA,4.2mA,8.0mA,12.0mA,16.0mA,19.8mA,20.0mA,20.2mA,andback.

The valve test mA values are not affected by the mA span setting describedinsection4.5

1. Usethemainmenutosetthecalibratortosourceorsimulatecurrent.

2. Ifvalvetestisnotenabled,usethemenusystemtoenableitasdescribedinsection4.3.

3. Pressthe“Step/Ramp”keyorthe“25%”keytosteptheoutputto verify the proper valve operation.

4. Usethemenusystemtodisablevalvetestwhendone.

5.3 HART 250Ω ResistorThecalibratorhastheabilitytoinserta250Ω resistor in series with the power supply in order to facilitate the use of a HART communicator. The HART resistor is enabled through the menu systemasdescribedinsection4.4.

Page 19: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

17

6. HART Device CommunicationsThe HART functions are only available in the LC-110H.

Auto shutdown is disabled on entry to the HART Menus and is restoredtoitspreviousstateuponexitfromtheHARTmenus.

Loopcurrenttrimissupportedfortransmitterdevices,butisnotsupported for actuator devices.

AllfunctionsontheModifySetupmenu,andthedevicediagnostic,trim4mA,trim20mA,fixedoutput,andPVzerofunctionsmaybedisabledusingthe‘HARTWriteEnable’selectionontheCalibratorSetupMenu,describedinsection4.8.Ifanyofthesefunctionsarerequired,theymustbeenabledbeforeenteringtheHARTmenus.

6.1. HART Connection Diagrams

6.1.1 In Circuit, External Loop Power

In“mAMeasure”modetheLC-110Hisincircuitandlooppowerissuppliedexternallyasshownbelow.

24VSUPPLY

(+)

(–)

4 to 20mA

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

Measure 0.00%

4.000mA250Ω

Note:Inthissituationthe250ΩHARTresistorisenabled.Ifyourloopalreadyhas250Ω,donotenabletheHARTresistor.

Page 20: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

18

6.1.2 In Circuit, LC-110H Loop Power

In“mAMeasurewith24V”modetheLC-110Hisincircuitandtheloop power is supplied by the LC-110H as shown below.

(+)(–)

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

Measure 24V 0.00%

4.000mA250Ω

Note:Inthissituationthe250ΩHARTresistorisenabled.Ifyourloopalreadyhas250Ω,donotenabletheHARTresistor.

6.1.3 Across Circuit, Communicator Only

In“CommunicatorOnly”modetheLC-110Hisacrossthecircuitandthelooppowerissuppliedexternallyasshownbelow.

MARTEL ELECTRONICS

LC-110H Precision LoopCalibrator w/HART™

100%

25%

0%MENUEXIT

PT12234

Device Setup & DataWrite LRV & URV ValueTrim, Set, & ZeroDevice DiagnosticData Log & Config Log

24VSUPPLY

(+)

(–)

4 to 20mA RL≥250Ω

Note:IntheCommunicatorOnlyMode,theremustbe250Ωs resistance present in the loop.

Page 21: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

19

6.2 Communications Setup and SelectionForallmaindisplayoperatingmodesexceptmAmeasurewith24Vorvoltsmeasure,theoperatingmodeissettomAmeasureuponentry of the HART Comm. Menu.

The operating mode remains the same as the main display when it is mAmeasurewith24V.Ifvoltsmeasurewasselectedinthemaindisplay,themenuwilldefaulttocommunicatormodeonlyandthe250Ωresistorselectionshows“n/a”.The250Ω resistor selection is not editable in communicator mode.

Onexitneitherthemodenortheresistorwillchangefromthelastselections made in the HART Communications Menu.

Themodeand250Ω resistor settings must correspond to how the testleadsarehookedupbeforeconnectingorsimulatingamodem.

HART Comm. Menu

4.123mA Mode: mA 250ΩResistor:ON HARTCONNECT SIMULATEHARTMODEM

Rotatetheknobtoselectanactionbymovingthereversevideohighlightupanddown.Presstheknobtoperformtheselectedaction.PressMENU/EXITtodisconnectfromHARTmodeandreturnto the main display without performing an action.

Themodeand250Ωresistoractionsaredescribedinsections6.2.1and6.2.2respectively.

The“HARTCONNECT”actionandsubsequentoperationsaredescribed in sections 6.3 through 6.10. An error is displayed and no actionistakenifthemeasuredinputisoutofrange,OL,or–OL.

The“SIMULATEHARTMODEM”actionisdescribedinsection6.11.

6.2.1 Select Mode

Mode

mA Measure mAMeasurewith24V CommunicatorOnly

Page 22: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

20

Rotatetheknobtomovethereversevideohighlighttothedesiredselection.PresstheknobtochangethemodetothehighlightedselectionandreturntotheSetupandSelectiondisplayinsection6.2.PressMENU/EXITtorestorethepreviousselectionandreturntotheSetupandSelectiondisplayinsection6.2.

In“mAMeasure”modetheLC-110Hisincircuitandlooppowerissuppliedexternallyasshowninsection6.1.1.

In“mAMeasurewith24V”modetheLC-110HisincircuitandthelooppowerissuppliedbytheLC-110Hasshownin6.1.2.

In“CommunicatorOnly”modetheLC-110Hisacrossthecircuitandthelooppowerissuppliedexternallyasshownin6.1.3.The250Ω Resistorselectionwilldefaultto“n/a”.

6.2.2 Select 250Ω Resistor

250Ω Resistor

OFFON

Rotatetheknobtomovethereversevideohighlighttothedesiredselection.PresstheknobtochangetheresistortothehighlightedselectionandreturntotheSetupandSelectiondisplayinsection6.2.PressMENU/EXITtorestorethepreviousselectionandreturntotheSetupandSelectiondisplayinsection6.2.

Page 23: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

21

6.3 Connecting to a HART DeviceBeforeanyoperationmaybeperformedwithaHARTdevice,itmustbe located on the loop. This is done by polling all of the possible device addresses and selecting a device from those addresses that respond.

IfaHARTprotocolrevision5orearlierdeviceisfoundontheloop,pollingstopsatpolladdress15,otherwiseitcontinuestopolladdress63.Pollingstopsafteratotalof10devicesarefoundontheloop.

Ifmultipledevicesarefoundontheloop,ataglistisdisplayedforselecting the device of interest.

Ifonlyonedeviceisfoundontheloop,itbecomesthedeviceofinterest by default.

Onceadeviceofinterestisdeterminedallrelevantdataisreadfromthedeviceandtheoperationsdescribedinsections6.5through6.10become available.

6.3.1 Polling

The polling starts immediately and the display changes to the followingwherethestringofdotsisextendedoncepersecondtoshow the operation is progressing.

PollingLoop ...........

Found2devices

Pressknobtoskip remainder

The number of devices found on the loop so far during polling is displayed.

Theknobmaybepressedtostoppollingearlyifitisknownthatallofthedevicesontheloophavebeenfound.PressMENU/EXITtostoppolling,disconnectfromHARTmode,andreturntothemaindisplay.

An error is displayed if no devices are found.

Ifmultipledevicesarefound,alistoftagsisdisplayedforselectingthe device of interest.

Ifonlyonedeviceisfoundthetagselectionstepisskipped.

Page 24: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

22

6.3.2 Tag Selection

The tag selection display lists all of the long tag names found during polling. Tag names may span two lines if necessary to show all of thetext.

Ifthelongtagnameisnotavailable,oritisblank,theshorttagnameisused.Iftheshorttagnameisblank,thetext“Polladdressx”isused.

Ifnecessary,thedisplayspansmorethanonescreen.Thelowerleftcorner contains the symbol when additional screens follow the presentone,thesymbol when additional screens precede the presentone,andbothwhenadditionalscreensfollowandprecedethepresentone.Ifthereisonlyonescreen,thereisnosymbol.

Long Tag 1 on two lines

LongTag2 Long Tag 3 LongTag4 on two lines

Rotatetheknobtomovethereversevideohighlighttothedesiredtag.Presstheknobtoselectthehighlightedtagandproceedtothedataacquisitiondisplay.PressMENU/EXITtodisconnectfromHARTmode and return to the main display.

6.3.3 Data Acquisition

This display is shown while acquiring all of the configuration data fromthedevice.Thestringofdotsisextendedoncepersecondtoshow the operation is progressing. The ♥ symbol in the upper right cornerflashestoindicateaworkingconnection.

AcquiringData ♥

............

Tag name

The display shows the name of the tag being accessed.

PressMENU/EXITtostopdataacquisition,disconnectfromHARTmode,andreturntothemaindisplay.

Page 25: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

23

Whendataacquisitioniscomplete,theFunctionSelectMenudescribedinsection6.5isdisplayed.

6.4 DisconnectThis display is shown before returning to the main display to allow the user to disconnect the LC-110H from the loop.

DisconnectfromLoop

PressMenu/Exitto exitHARTmode

6.5 Function Select Menu

TagName♥

DisplaySetup&Data ModifySetup Trim,Set,&Zero DeviceDiagnostic DataLog&ConfigLog

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Rotatetheknobtomovethereversevideohighlighttothedesiredaction.Presstheknobtoselectthehighlightedactionandproceedto the corresponding display described in sections 6.6 through 6.10. PressMENU/EXITtodisconnectfromHARTmodeandreturntothemain display.

6.6 Display Setup & DataThisdisplayspans11screenswiththefollowinggeneralformat.ItdisplaysallofthedataretrievedduringtheDataAcquisitionprocedure described in section 6.3.3.

Page 26: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

24

Sample Screen Image

PT12234♥

PT12234

PVUnit:inWC68F PV:226.0inWC PVmA:13.016mA PV%:56.35%

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Eachscreencontainsupto6itemsofdata.Anitemmayspanmorethanonelineifnecessarytodisplaythefulltext.IfadataitemisnotsupportedintheHARTdevice,itismarked“n/a”,notavailable.Dataitems that change dynamically in the HART device are updated as often as possible on the screens.

The lower left corner contains the symbol when additional screens followthepresentone,thesymbol when additional screens precedethepresentone,andbothwhenadditionalscreensfollowand precede the present one.

Rotatetheknobtostepbetweenscreens.PressMENU/EXITtoreturntotheFunctionSelectmenuinsection6.5.

6.7 Modify SetupIftheHARTwritecommandsarenotenabledpersection4.8,noneofthese functions are available and an error message is displayed instead of the following screen.

TagName♥

WriteLRV WriteURV WritePVUnit ModifyTags,Msg,Descr

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Page 27: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

25

Rotatetheknobtomovethereversevideohighlighttothedesiredaction.Presstheknobtoselectthehighlightedactionandproceedtothecorrespondingdisplaydescribedinsection6.7.1through6.7.4.PressMENU/EXITtoreturntotheFunctionSelectmenuinsection6.5.

6.7.1 Write LRV

AwarningtochangethelooptoMANUALbeforeproceedingisdisplayed.Presstheknobtoproceed.PressMENU/EXITtoreturntotheModifySetupmenuinsection6.7.

ThepresentLRVvalueandunitsaredisplayed..

TagName♥

WriteLRV:degreesF

300.5Press&holdtoWrite

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

•Presstheknobtomovethedecadecursor.

•Rotatetheknobtoincrementordecrementthevalueinstepsindicated by the selected decade.

•PressandholdtheknobtosendthenewvaluetotheHARTdevice,remainingonthisdisplay.AnerrorisdisplayediftheHARTdevicerejectsthevalueforanyreason.

•PressMENU/EXITtoreturntotheModifySetupmenuinsection6.7.AremindertochangethelooptoAUTOMATICisdisplayedfirst.

6.7.2 Write URV

AwarningtochangethelooptoMANUALbeforeproceedingisdisplayed.Presstheknobtoproceed.PressMENU/EXITtoreturntotheModifySetupmenuinsection6.7.

ThepresentURVvalueandunitsaredisplayed.

Page 28: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

26

TagName♥

WriteURV:degreesF

300.5Press&holdtoWrite

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

•Presstheknobtomovethedecadecursor.

•Rotatetheknobtoincrementordecrementthevalueinstepsindicated by the selected decade.

•PressandholdtheknobtosendthenewvaluetotheHARTdevice,remainingonthisdisplay.AnerrorisdisplayediftheHARTdevicerejectsthevalueforanyreason.

•PressMENU/EXITtoreturntotheModifySetupmenuinsection6.7.AremindertochangethelooptoAUTOMATICisdisplayedfirst.

6.7.3 Write PV Unit

AwarningtochangethelooptoMANUALbeforeproceedingisdisplayed.Presstheknobtoproceed.PressMENU/EXITtoreturntotheModifySetupmenuinsection6.7.

The available units are displayed with the present unit highlighted in reverse video. The available units listed correspond to the type of unitpresentlyselected.Forexample,ifthepresentunitisPSIonlypressureunitsarelisted,ifthepresentunitis°Conlytemperatureunits are listed.

TagName♥

SelectPVUnit: degree C degreeF degree R degree K Press&holdtoWrite

Page 29: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

27

The tag name is truncated to fit on one line if necessary. The ♥ symbol intheupperrightcornerflashestoindicateaworkingHARTconnection.

• Rotatetheknobtomovethereversevideohighlighttothedesired unit.

• PressandholdtheknobtosendthenewvaluetotheHARTdevice,returningtotheModifySetupmenuafterdisplayingaremindertochangethelooptoAUTOMATIC.AnerrorisdisplayediftheHARTdevicerejectsthevalueforanyreason.

• PressMENU/EXITtoreturntotheModifySetupmenuinsection6.7.AremindertochangethelooptoAUTOMATICisdisplayedfirst.

6.7.4 Modify Tags, Msg, Descr

Selecttheitemtobemodifiedfromthefollowingmenu.

TagName♥

WriteShortTag Write Long Tag WriteDescriptor Write Message

The tag name is truncated to fit on one line if necessary. The ♥ symbol intheupperrightcornerflashestoindicateaworkingHARTconnection.

Rotatetheknobtomovethereversevideocursortothedesireditem.Presstheknobtomovetotheeditscreenbelow.Theexampleshown is for a long tag.

Truncated Long Tag ♥

Long tag: Furnace3oxygentemp erature

Press for help Press&holdtoWrite

The tag name is truncated to fit on one line if necessary. The ♥ symbol intheupperrightcornerflashestoindicateaworkingHARTconnection.

Page 30: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

28

The character selected for change is indicated by an underline cursor.

Shorttagisupto8characterslongfromalimitedcharactersetthatdoes not include lower case.

Longtagisupto32characterslongfromthefullcharacterset.

Descriptorisupto16characterslongfromalimitedcharactersetthat does not include lower case.

Messageisupto32characterslongfromalimitedcharactersetthatdoes not include lower case.

• Rotatetheknobtomodifytheselectedcharacterbyscrollingthrough the available character set.

• PressandholdtheknobtosendthenewvaluetotheHARTdevice,returningtotheModifySetupmenu.AnerrorisdisplayediftheHARTdevicerejectsthevalueforanyreason.

• Presstheknobtomovethecursoronecharactertotheright.

• Pressthe“25%”keytomovethecursoronecharactertotheleft.

• Pressthe“0%”keytodeletetheselectedcharacter.

• Pressthe“100%”keytoinsertaspaceinfrontoftheselectedcharacter.

• Pressthe“Step/Ramp” keytodisplayahelpscreen.PresstheknoborMENU/EXITonthehelpscreentoresumethemodification at the same position.

• PressMENU/EXITtoreturntotheModifySetupmenuinsection6.7.

6.8 Trim, Set, & Zero MenuIftheHARTwritecommandsarenotenabledpersection4.8,noneofthese functions are available and an error message is displayed instead of the following screen.

TagName♥

TrimD/Aat4mA TrimD/Aat20mA SetfixedmAOutput PVZero

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Page 31: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

29

Rotatetheknobtomovethereversevideohighlighttothedesiredaction.Presstheknobtoselectthehighlightedactionandproceedtothecorrespondingdisplaydescribedinsections6.8.1through6.8.4.PressMENU/EXITtoreturntotheFunctionSelectmenuinsection6.5.

6.8.1 Trim 4 mA

Iftheoperatingmodeis‘CommunicatorOnly’,thisfunctionisnotavailable and an error message is displayed instead of the following screens.

AwarningtochangethelooptoMANUALbeforeproceedingisdisplayed.Presstheknobtoproceed.PressMENU/EXITtoreturntotheTrim,Set&Zeromenuinsection6.8.

WhiletheHARTdeviceischangedtofixedoutputmode,aninformation display is shown. An error is displayed if the HART devicerejectsthemodechangecommandforanyreason.Oncethemodechangeissuccessful,thefollowingisdisplayed.

TagName♥

TrimD/Aat4mA LC110H:

4.065mAPressknobtotrim

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Whilewaitingfortheoutputtosettleat4mA,thescreenindicatestheLC-110H measurement. The measurement is updated once a second.

•PresstheknobtotrimtheHARTdevice,remainingonthisdisplayto evaluate the result. An error is displayed if the HART device rejectsthetrimcommandforanyreason.

•PressMENU/EXITtochangetheHARTdevicetonormaloutputmodeandreturntotheTrim,Set&Zeromenuinsection6.8.AremindertochangethelooptoAUTOMATICisdisplayedfirst.Anerrorisdisplayedifthedevicerejectsthemodechangecommandfor any reason.

Page 32: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

30

6.8.2 Trim 20 mA

Iftheoperatingmodeis‘CommunicatorOnly’,thisfunctionisnotavailable and an error message is displayed instead of the following screens.

AwarningtochangethelooptoMANUALbeforeproceedingisdisplayed.Presstheknobtoproceed.PressMENU/EXITtoreturntotheTrim,Set&Zeromenuinsection6.8.

WhiletheHARTdeviceischangedtofixedoutputmode,aninformation display is shown. An error is displayed if the HART devicerejectsthemodechangecommandforanyreason.Oncethemodechangeissuccessful,thefollowingisdisplayed.

TagName♥

TrimD/Aat20mA LC110H:

19.965mAPressknobtotrim

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Whilewaitingfortheoutputtosettleat20mA,thescreenindicatesthe LC-110H measurement. The measurement is updated once a second.

•PresstheknobtotrimtheHARTdevice,remainingonthisdisplayto evaluate the result. An error is displayed if the HART device rejectsthetrimcommandforanyreason.

•PressMENU/EXITtochangetheHARTdevicetonormaloutputmodeandreturntotheTrim,Set&Zeromenuinsection6.8.AremindertochangethelooptoAUTOMATICisdisplayedfirst.Anerrorisdisplayedifthedevicerejectsthemodechangecommandfor any reason.

6.8.3 Set Fixed mA Output

Iftheoperatingmodeis‘CommunicatorOnly’,thisfunctionisnotavailable and an error message is displayed instead of the following screens.

AwarningtochangethelooptoMANUALbeforeproceedingisdisplayed.Presstheknobtoproceed.PressMENU/EXITtoreturntotheTrim,Set&Zeromenuinsection6.8.

Page 33: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

31

WhiletheHARTdeviceischangedtofixedoutputmode,aninformation display is shown. An error is displayed if the HART devicerejectsthemodechangecommandforanyreason.Oncethemodechangeissuccessful,thefollowingisdisplayed.

TagName♥

LC110H:4.065mA

Selectfixedoutput:

4.000 mAPress&holdtoSend

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Thescreenisusedtosetafixedoutputandmonitortheresultwiththe LC-110H measurement. The measurement is updated once a second.

Therangeofvaluesthatcanbesetis3.0to21.0mA.

•Presstheknobtomovethedecadecursor.

•Rotatetheknobtoincrementordecrementthevalueinstepsindicated by the selected decade.

•PressandholdtheknobtosendthenewvaluetotheHARTdevice,remainingonthisdisplay.AnerrorisdisplayediftheHARTdevicerejectsthevalueforanyreason.

•PressMENU/EXITtochangetheHARTdevicetonormaloutputmodeandreturntotheTrim,Set&Zeromenuinsection6.8.AremindertochangethelooptoAUTOMATICisdisplayedfirst.Anerrorisdisplayedifthedevicerejectsthemodechangecommandfor any reason.

6.8.4 PV Zero

AwarningtochangethelooptoMANUALbeforeproceedingisdisplayed.Presstheknobtoproceed.PressMENU/EXITtoreturntotheTrim,Set&Zeromenuinsection6.8.

ThisscreenrequeststheusertosetthePVprocessinputsignaltozeroandmonitorstheresultwiththeLC-110Hmeasurement.Themeasurement is updated once a second.

Page 34: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

32

Whentheoperatingmodeis‘CommunicatorOnly’,themAmeasurementisnotavailableandthemessage‘mAnotavailable’,‘inComm.Onlymode’isdisplayedinstead.

TagName♥

SetPVinputtozero LC110H:

4.065mAPressknobtozero

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

•PresstheknobtozerotheHARTdevice,remainingonthisdisplayto evaluate the result. An error is displayed if the HART device rejectsthezerocommandforanyreason.

•PressMENU/EXITtoreturntotheTrim,Set&Zeromenuinsection6.8.AremindertorestorethePVprocessinputsignaltonormaloperatingconfigurationandchangethelooptoAUTOMATICisdisplayed first.

6.9 Device DiagnosticIftheHARTwritecommandsarenotenabledpersection4.8,thisfunction is not available and an error message is displayed instead of the following screens.

AwarningtochangethelooptoMANUALbeforeproceedingisdisplayed.Presstheknobtoproceed.PressMENU/EXITtoreturntotheFunctionSelectmenuinsection6.5.

TagName♥

SelfTest

Pressknobtotest

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Page 35: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

33

Presstheknobtostartthetest.Whiletheselftestisperformedthebottomlinechangesto“Testing”andastringofdotsisextendedonce per second to show the operation is progressing.

Attheendofthetestthefollowingscreenisdisplayed.Itshowseither“Noerrors”,orthefirsterrorreported.

TagName♥

SelfTestResults

RAM fault

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

The lower left corner contains the symbol when additional errors followthepresentone,thesymbol when additional errors precede thepresentone,andbothwhenadditionalerrorsfollowandprecedethepresentone.Ifthereisnoerrororonlyoneerror,thereisnosymbol.

Rotatetheknobtostepbetweenerrors.

PressMENU/EXITtoreturntotheFunctionSelectmenuinsection6.5.AremindertochangethelooptoAUTOMATICisdisplayedfirst.

6.10 Data Log & Config LogConfigurationLogandDataLogareavailablewhenconnectedtoaHARTdevice.TousetheDataLogandConfigLogfunctionsrequiresthepurchaseoftheoptionalBetaLOGHARTsoftwareapplication.BetaLOGHARTallowstheusertouploadthedatafilesfromtheLC-110HandstorethemonaPCas.TXT,.CSVorExcelfiles.

BetaLOGHARTincludesthenecessarycommunicationscableandUSBdrivertosupportcalibratortoPCcommunications.Additionalcables can be purchased separately.

Page 36: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

34

TagName♥

Configuration Log DataLog

The tag name is truncated to fit on one line if necessary. The ♥ symbolintheupperrightcornerflashestoindicateaworkingHARTconnection.

Rotatetheknobtomovethereversevideohighlighttothedesiredaction.Presstheknobtoselectthehighlightedactionandproceedtothecorrespondingdisplaydescribedinsections6.10.1or6.10.2.PressMENU/EXITtoreturntotheFunctionSelectmenuinsection6.5.

6.10.1 Configuration Log

Configurationdataforupto20tagsmaybestoredforlaterrecall.The configuration data saved is the same as that shown on the DeviceDatascreen,section6.6.

The initial display spans more than one screen and shows a list of the tags contained in storage. Tag names span two lines if they exceedthelengthofasingleline.Ifastoragepositionisnotused,thetagnameareadisplays‘<empty>’.Afterastoragepositionisselected,datamaybesavedtoit,recalledfromit,erased,orsenttotheRS232port.ThedatamayalsobeuploadedtoapersonalcomputerviatheBetaLOGHARTWindowsapplication.WerecommendthatyouusePN:1920049(LemotoUSBCable)forusewiththeBetaLOGHARTSoftware.ItincorporatesaUSB/SerialadapterusinganFTDIchipset.ThiscombinationprovidesconsistentandreliablecommunicationsbetweentheLC-110HandPC.TheUSBDriversareincludedontheBetaLOGHARTSoftwareCD.

Configuration Log ♥ 8Tag1 on two lines 9<empty> 10 Tag 3 11Tag4

Page 37: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

35

The ♥symbolintheupperrightcornerflashestoindicateaworkingHART connection.

The lower left corner contains the symbol when additional screens followthepresentone,thesymbol when additional screens precedethepresentone,andbothwhenadditionalscreensfollowand precede the present one.

Rotatetheknobtomovethereversevideohighlighttothedesiredstorageposition.Presstheknobtoselectthestorageposition.PressMENU/EXITtoreturntotheLogSelectmenuinsection6.10.

Onceastoragepositionhasbeenselected,thefollowingmenuisdisplayed to select the operation to be performed.

Position:♥

3 Tag on two lines

SAVE RECALL ERASE SEND

The ♥symbolintheupperrightcornerflashestoindicateaworkingHART connection.

The number and contents of the storage position are shown at the top.Thetagnumberis‘<empty>’ifthestoragepositionisempty.

Rotatetheknobtomovethereversevideohighlighttothedesiredoperation.Presstheknobtoperformtheoperation.PressMENU/EXITtoreturntotheLogSelectmenuinsection6.10.

SAVEoperation:

•Ifthepositionisempty,savethepresentdeviceconfigurationdata into the storage position.

•Ifthepositionisinuse,confirmthattheexistingdataistobereplaced with the present tag data before saving it into the storage position.

RECALLoperation:

•Ifthepositionisempty,anerrormessageisdisplayed.

•Ifthepositionisinuse,displaythedatainasequenceofscreensidenticaltotheDeviceDatascreen,section6.6.Return from those screens to this screen.

Page 38: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

36

ERASEoperation:

•Ifthepositionisempty,anerrormessageisdisplayed.

•Ifthepositionisinuse,confirmthattheexistingdataistobepermanently erased before doing so.

SENDoperation:

•Ifthepositionisempty,anerrormessageisdisplayed.

•Ifthepositionisinuse,sendthedatatotheRS232portinareport format.

6.10.2 Data Log

ProcessdatamaybestoredforasingletagforlateruploadtoapersonalcomputerviatheBetaLOGHARTWindowsapplication.

WerecommendthatyouusePN:1920049(LemotoUSBCable)forusewiththeBetaLOGHARTSoftware.ItincorporatesaUSB/SerialadapterusinganFTDIchipset.ThiscombinationprovidesconsistentandreliablecommunicationsbetweentheLC-110HandPC.TheUSBDriversareincludedontheBetaLOGHARTSoftwareCD.

Datamaybeloggedinmultiplesessions,butallsessionsmustbefrom the same HART device as determined by the long tag name. A differentloggingintervalmaybeselectedforeachsession.EachdatasamplecontainstheLC-110Hmeasurement,devicemA,andallfour process variables.

Thedataloggingcapacityis9810recordsless2overheadrecordsfor each session. The number of sessions is limited to 99.

Recordsfree:9810♥

Tag on two lines

START ERASE Battery:7.6V

The ♥symbolintheupperrightcornerflashestoindicateaworkingHART connection.

Thenumberoffreerecordsisshownonthefirstline.Ifdatahasalreadybeenlogged,thetagnumberisshownbelowit.

The present battery voltage is displayed at the bottom to allow the user to decide whether the batteries should be changed before starting the log. The log is stopped before the LC-110H is turned off whenthebatteryreachesitslowvoltageautomaticshutofflimitof5.6volts.

Page 39: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

37

Rotatetheknobtomovethereversevideohighlighttothedesiredoperation.Presstheknobtoperformtheoperation.PressMENU/EXITtoreturntotheLogSelectmenuinsection6.10.

STARToperation:

•Ifnofreerecordsremain,nofreesessionsremain,orthepresent HART device does not match the HART device already logged,anerrormessageisdisplayed.

•Otherwise,proceedtointervalselectiondescribedbelow.

ERASEoperation:

•Ifthereisnologgeddata,anerrormessageisdisplayed.

•Otherwise,confirmthattheexistingdataistobepermanentlyerased before doing so.

Selectthelogginginterval.

SelectInterval♥

1 second 2seconds 5seconds 10 seconds 30 seconds 60 seconds

The ♥symbolintheupperrightcornerflashestoindicateaworkingHART connection.

Rotatetheknobtomovethereversevideohighlighttothedesiredinterval.Presstheknobtostartloggingatthatinterval.PressMENU/EXITtoreturntothefirstdatalogscreen.

While logging the following is displayed to monitor progress and to allow logging to be stopped.

Logging ♥ Interval:30seconds Elapsed:2:15:25 Recordsused:270 Recordsfree:9430 LC110H6.288mA PVmA:6.317mA

The ♥symbolintheupperrightcornerflashestoindicateaworkingHART connection.

Page 40: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

38

PressMENU/EXITtostoploggingandreturntothefirstdatalogscreen.

The displayed data items are:

•Thetoplineindicateswhetherloggingisinprogress(Logging)orisstopped(Stopped).Loggingstopsautomaticallywhenstorageisfull,orbeforetheLC-110Histurnedoffwhenthebatteryreachesitslowvoltageautomaticshutofflimitof5.6volts.

•Intervalistheitempreviouslyselected.

•Elapsedisthetimesincethelogwasstarted,updatedeachtime a new sample is saved.

•Recordsusedisthetotalnumberusedtodateforallsessions,updated each time a new sample is saved.

•Recordsfreeisthetotalnumberthatremainunused,updatedeach time a new sample is saved.

•LC110Histhepresentmeasurement,updatedasoftenaspossible.

•PVmAisthelastHARTdevicemeasurement,updatedasoftenas possible.

6.11 HART Modem Simulation

SimulateHARTModem ♥

4.123mA

TheLEMOcommunicationsportissetto1200baud,8databitsplusoddparity,and1stopbitpertheHARTstandard.AlldatareceivedontheLEMOportissenttotheHARTmodemasis,andalldatareceivedfromtheHARTmodemissenttotheLEMOportasis.

The ♥ symbol in the upper right flashes once for each set of data receivedontheLEMOport.

TheLEMOportisrestoredtonormal9600baudoperationwhenthesimulation is stopped.

PressMENU/EXITtostopthemodemsimulation,disconnectfromHART mode and return to the main display.

WerecommendthatyouusePN:1920049(LemotoUSBcable)whenusingtheLC-110HasaHARTmodem.ItincorporatesaUSB/SerialadapterusinganFTDIchipset.Itprovidesconsistentand

Page 41: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

39

reliablecommunicationsbetweentheLC-110HandPC.USBdriversareincludedonaCDwiththecablewhenpurchasedfromMartel.

6.12 Error DisplaysThere are two classes of errors which may occur during HART communications. The first class contains errors that are reported by the loop device in its response. The second class contains errors that are detected by the LC110H when receiving responses from the loop device.

Loop Device Errors

These errors are reported by the loop device in its response to the LC110H,pertheHARTprotocol.

Theyareindicatedbythetext“LoopDeviceerror:”atthetopofthedisplay. This is followed by one of the following error messages.

1) Invalid selection

The data value sent with the HART command is not allowed by this loop device. The command was ignored by the loop device. HART responsecode2.

2) Value too large

The data value sent with the HART command is greater than the range allowed by this loop device. The command was ignored by the loop device. HART response code 3.

3) Value too small

The data value sent with the HART command is less than the range allowed by this loop device. The command was ignored by the loop device.HARTresponsecode4.

4) Message too short

The HART command received contains fewer characters than expectedbythisloopdevice.Thecommandwasignoredbytheloopdevice.HARTresponsecode5.

5) Device specific error

The loop device has detected a device specific error. The LC110H does not support the reporting of details on device specific errors. The command was ignored by the loop device. HART response code 6.

Page 42: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

40

6) Write protected

The loop device is write protected. The LC110H can not alter the write protect status of a device. The command was ignored by the loopdevice.HARTresponsecode7.

7) Set to nearest possible value

The data value sent with the HART command was rounded or truncated due to limitations within the loop device. The command was accepted by the loop device. The resultant data values should bechecked.HARTresponsecode8.

8) Update in progress

The loop device is updating the data value on a delayed basis. The commandwasacceptedbytheloopdevice.HARTresponsecode8.

9) Incorrect loop mA mode or value

Theloopdeviceisnotinfixedcurrentmode,orthemAvaluesupplied is invalid. The command was ignored by the loop device. HART response code 9.

10) LRV too high

TheLRVvaluereceivedisgreaterthantheUpperTransducerLimitofthisloopdevice,orsomeotherphysicaldevicelimitationisexceeded.Thecommandwasignoredbytheloopdevice.HARTresponse code 9.

11) Applied process too high

The data value received is greater than some device limit. The command was ignored by the loop device. HART response code 9.

12) LRV too low

TheLRVvaluereceivedislessthantheLowerTransducerLimitofthisloopdevice,orsomeotherdevicephysicallimitationisexceeded.Thecommandwasignoredbytheloopdevice.HARTresponse code 10.

13) Applied process too low

The data value received is less than some device limit. The command was ignored by the loop device. HART response code 10.

Page 43: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

41

14) URV too high

TheURVvaluereceivedisgreaterthantheUpperTransducerLimitofthis loop device. The command was ignored by the loop device. HART response code 11.

15) Loop mA not active

The loop device is in multi-drop mode. The command was ignored by the loop device. HART response code 11.

16) URV too low

TheURVvaluereceivedislessthantheLowerTransducerLimitofthis loop device. The command was ignored by the loop device. HARTresponsecode12.

17) LRV & URV out of limits

TheLRVand/orURVvaluereceivedisoutsideoftheTransducerLimitsofthisloopdevice,orsomeotherdevicephysicallimitationisexceeded.Thecommandwasignoredbytheloopdevice.HARTresponse code 13.

18) Span too small

ThespanoftheLRVandURVvaluesreceivedislessthantheMinimumSpanallowedbythisloopdevice.Thecommandwasignoredbytheloopdevice.HARTresponsecode14.

19) Access restricted

The HART command received is not allowed when the loop device in its current operating mode. The command was ignored by the loop device. HART response code 16.

20) Invalid units code

The units code received is not allowed by the loop device in its current operating mode. The command was ignored by the loop device.HARTresponsecode18.

21) Invalid span

The data values in the HART command would result in an invalid span for this loop device. The command was ignored by the loop device.HARTresponsecode29.

Page 44: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

42

22) Busy

The loop device is temporarily busy performing some internal operation. The command was ignored by the loop device. Try the commandagainlater.HARTresponsecode32.

23) Command not implemented

The HART command has not been implemented in this loop device. The command was ignored by the loop device. HART response code64.

24) Character parity

The loop device detected a character parity error in one or more characters received. The command was ignored by the loop device. Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.HARTresponsecode192(hexC0).

25) Character over run

The loop device detected a general communications error. The command was ignored by the loop device. Try the command again. Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.HARTresponsecode160(hexA0).

26) Character framing

The loop device detected a character framing error in one or more characters received. The command was ignored by the loop device. Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.HARTresponsecode144(hex90).

27) Check byte bad

The command received by the loop device did not contain a valid checkbytepertheHARTprotocoldefinition.Thecommandwasignoredbytheloopdevice.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.HARTresponsecode136(hex88).

28) Buffer overflow

The loop device detected a general communications error. The command was ignored by the loop device. Try the command again. Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.HARTresponsecode130(hex82).

Page 45: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

43

LC110H Detected Errors

These errors are detected by the LC110H when receiving responses from the loop device.

Theyareindicatedbythetext“LC110Herror:”atthetopofthedisplay. This is followed by one of the following error messages.

a) No response

The loop device did not respond to the last HART command sent. Checkthepowerandconnectionstotheloopdevice.Trythecommand again.

b) Character parity

The response received from the loop device contained one or more characterswithaninvalidparitycheck.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.

c) Character frame

The response received from the loop device contained one or more characterswithaninvalidframeformat.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.

Page 46: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

44

d) Character overrun

A general communications error occurred during the response receivedfromtheloopdevice.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.

e) Buffer overrun

A general communications error occurred during the response receivedfromtheloopdevice.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.

f) Framing bad

The structure of the response received from the loop device did not matchtheHARTprotocoldefinition.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.

g) Check byte bad

The response received from the loop device did not contain a valid checkbytepertheHARTprotocoldefinition.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloop device.

h) Invalid address

The response received from the loop device contained a HART deviceaddressdifferentfromthatsent.Ensuretherearenoburstmodedevicesbroadcastingontheloop.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.

i) Invalid command

The response received from the loop device contained a HART commandcodedifferentfromthatsent.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.

j) Invalid fixed data

The structure of the response received from the loop device did not matchtheHARTprotocoldefinition.Trythecommandagain.Iftheerrorpersists,checkthepowerandconnectionstotheloopdevice.

Page 47: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

45

7. Maintenance

7.1 Replacing BatteriesReplace batteries as soon as the battery indicator turns on to avoid falsemeasurement.IfthebatteriesdischargetoodeeplytheLC-110willautomaticallyshutdowntoavoidbatteryleakage.

Note:UseonlyAAAsizealkaline,lithiumbatteries,orrechargeableNiMhcells.

7.2 Cleaning the Unit

WarningToavoidpersonalinjuryordamagetothecalibrator,useonlythespecified replacement parts and do not allow water into the case.

CautionToavoiddamagingtheplasticlensandcase,donotusesolventsorabrasive cleaners. Clean the calibrator with a soft cloth dampened with water or water and mild soap.

7.3 Fuse ProtectionTheLC-110Seriesisprotectedwithaninternalpolyfuseagainstovercurrentconditions.Thefusewillautomaticallyreset,usuallywithinafewseconds.Donotattempttoopentheunitandtroubleshoot. This will void the product warranty.

Page 48: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

46

8. Specifications

Functions: mAsource,mAsimulate,mAread,mAread/looppower,andvoltsread.

Ranges: mA(0to24mA)andVolts(0to30VDC)

Resolution: 1uAonmArangesand1mVonvoltagerange

Accuracy: 0.01%ofreading±2LSDallranges (@23°±5°C)

OperatingTempRange: -10°Cto55°C(14°Fto131°F)

Humidity Range: 10to95%non-condensing

Storage: -20°Cto60°C(-4°Fto140°F)

Stability: 20ppmofF.S./°Cfrom-10°Cto18°Cand28°Cto55°C

Display: 128x64pixels,LCDGraphicw/backlight, .34”highdigits

Power: 6AAAalkaline,lithium,orNiMHbatteries

Battery Life: ≥40hourscontinuoususe(measuremode)Loop Compliance Voltage:

24VDC@20mA

LoopDriveCapability: 1200ΩwithoutHARTresistor,950Ω with HART resistor

IPRating: IEC60529:IP40

Over-VoltageProtection: 240VAC

OverloadCurrentProtection:

28mADC

EMC: EN61326AnnexA(PortableInstruments)

Dimensions(LxWxD): 6”x3.6”x1.3”(15cmx9cmx3cm)

Weight: 9.5ounces(0.3kg)

IncludedAccessories: NISTtraceablecalibrationcertificatewithdata,batteries,testleads,andmanual

9. WarrantyMartelElectronicsCorporationwarrantsallproductsagainstmaterialdefectsandworkmanshipforaperiodoftwelve(12)monthsafterthedateofshipment.Problemsordefectsthatarisefrommisuseorabuseoftheinstrumentarenotcovered.Ifanyproductistobereturned,a“ReturnMaterialAuthorization”formcanbeobtainedfromour website www.martelcorp.com under customer service. You can

Page 49: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

47

alsocall1-800-821-0023tohaveaformfaxed.Martelwillnotberesponsiblefordamageasaresultofpoorreturnpackaging.Outofwarrantyrepairsandrecalibrationwillbesubjecttospecificcharges.UndernocircumstanceswillMartelElectronicsbeliableforanydevice or circumstance beyond the value of the product.

Page 50: LC-110/Hs8933.pcdn.co/wp-content/uploads/LC-110H-Loop-Calibrator-Rev-B-8-13.pdf · • The text ‘250Ω’ text is shown in the lower right corner when the HART resistor has been

0220024RevB8/13

www.martelcalibrators.come-mail: [email protected]

Tel:(603)434-1433800-821-0023 Fax:(603)434-1653

Martel Electronics 3CorporateParkDrive Derry,NH03038