LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg...

31
LB-CAL-013 E mirates I nternational A ccreditation C entre has accredited Gulf Calibration & Technical Services Ghala Industrial Area, Muscat- Sultanate of Oman In accordance with the requirements of ISO/IEC 17025:2017 General requirements for the competence of testing and calibration laboratories to undertake the calibration in the attached accreditation scope This Accreditation is invalid without the attached accreditation scope and shall remain in force within the validity period printed below, subject to continuing compliance with the requirements of the accreditation criteria. Validity: 30-03-2021 to 10-09-2022 Initial Accreditation Date: 11-09-2013

Transcript of LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg...

Page 1: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

LB-CAL-013

Emirates International Accreditation Centre

has accredited

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

In accordance with the requirements of

ISO/IEC 17025:2017

General requirements for the competence of testing and calibration laboratories

to undertake the calibration in the attached accreditation scope

This Accreditation is invalid without the attached accreditation scope and shall remain in force within the validity

period printed below, subject to continuing compliance with the requirements of the accreditation criteria.

Validity: 30-03-2021 to 10-09-2022

Initial Accreditation Date: 11-09-2013

Page 2: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeLB-CAL-013

Date: 30-03-2021

Scope Issue No. Date

Temperature 5

Balance 4

Force 5

Mass 4

Dimensional 4

Pressure 5

Electrical 2

Volume 2

Humidity 2

Temperature 4

Balance 3

Force 4

Mass 3

Dimensional 3

Pressure 4

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Valid to: 10-09-2022

Extension in scope and first issuance under the name of

EIAC (which was formerly known as DAC)

First issuance under the name of EIAC (which was

formerly known as DAC)

3/30/2021

Re-issued due to: transition to the new version of ISO/IEC

17025:2017 and comply with the new accreditation

number format

Re-issued due to: transition to the new version of ISO/IEC

17025:2017 and comply with the new accreditation

number format

Re-issued due to: structural amendments made to

methods only, transition to the new version of ISO/IEC

17025:2017 and comply with the new accreditation

number format

Re-issued due to: transition to the new version of ISO/IEC

17025:2017 and comply with the new accreditation

number format

Re-issued due to: structural amendments made to

methods only, transition to the new version of ISO/IEC

17025:2017 and comply with the new accreditation

number format

Re-issued due to: structural amendments made to

methods only, transition to the new version of ISO/IEC

17025:2017 and comply with the new accreditation

number format

Accreditation History

Details

12/5/2019

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26661 of 2

Page 3: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeLB-CAL-013

Date: 30-03-2021

Scope Issue No. Date

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Valid to: 10-09-2022

3/30/2021Re-issued due to: transition to the new version of ISO/IEC

Accreditation History

Details

Electrical 1

Volume1

Humidity1

Granted accreditation from EIAC 12/5/2019

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26662 of 2

Page 4: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope Temperature Calibration

LB-CAL-013

Issue no.: 05

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

0 °C 0.02 °C

-20 °C to 150 °C 0.05 °C

>150 °C to 250 °C 0.07 °C

30 °C to 250 °C 0.3 °C

>250 °C to 400 °C 0.6 °C

>400 °C to 600°C 1.2 °C

0 °C 0.05 °C

-20 °C to 150 °C 0.1 °C

>150 °C to 250 °C 0.4 °C

>250 °C to 400 °C 0.6 °C

>400 °C to 600 °C 0.8 °C

>600 °C to 1100 °C 2.0 °C

-20 °C to 250 °C 0.25 °C

>250 °C to 400 °C 0.5 °C

>400 °C to 600 °C 1.0 °C

Liquid-in-glass

thermometers

GCTS-SOP-10 -20 °C to 250 °C 0.25 °C Laboratory

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Digital Thermometers

With PRT sensor

GCTS-SOP-08

In liquid bath

Laboratory

Digital Thermometers

With PRT sensor

GCTS-SOP-08

In dry block

Laboratory

Digital Thermometers

With TC sensor

GCTS-SOP-08

In liquid bath

Laboratory

Digital Thermometers

With TC sensor

GCTS-SOP-08

In dry block

Laboratory

Dial thermometers GCTS-SOP-09 Laboratory/

Customers

Premises

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26661 of 2

Page 5: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope Temperature Calibration

LB-CAL-013

Issue no.: 05

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Digital Thermometers GCTS-SOP-08 Laboratory50 °C to 300 °C 2.0 °C

>300 °C to 500 °C 3.0 °C

Environmental chambers,

Ovens

GCTS-SOP-11 ­20 °C to 300 °C 0.7 °C

Incubators

(5 sensors)

GCTS-SOP-13 Ambient to 100 °C 0.3 °C

Autoclaves (temperature GCTS-SOP-38 100 °C to 140 °C 0.3 °C Laboratory/

Customers

Premises­20 °C to 95 °C 0.25 °C

>95 °C to 200 °C 0.4 °C

Ambient to 300 °C 1 °C

300 °C to 600 °C 3 °C

600 °C to 800 °C 6 °C

800 °C to 1200 °C 10 °C

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Liquid bath

(5 sensors)

GCTS-SOP-12 Laboratory/

Customers

Premises

Muffle Furnace

(1 sensor)

GCTS-SOP-16

IR thermometers GCTS-SOP-40 Laboratory

Chillers, Freezers

(9 sensors)

GCTS-SOP-11 .-20 °C to 20 °C 0.5 °C Laboratory/

Customers

Premises

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26662 of 2

Page 6: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeBalance Calibration

LB-CAL-013

Issue no.: 04

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Up to 5 kg 5x 10-6 of FS

>5 kg up to 30 kg 1x 10-5 of FS

>30 kg up to 1000 kg 5x10-5 of FS

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Mass/ Electronic Balances Use of mass standards, acc.

to EURAMET cg 18:2015;

ASTM E898-20

GCTS-SOP-01

Customers

Premises

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26661 of 1

Page 7: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeDimensional Calibration

LB-CAL-013

Issue no.: 04

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

All types of Calipers

Vernier

Dial

Electronic

GCTS -SOP 05

based on BS EN 887:2008

Up to 300 mm Overall Performance

10+ (15 x *L) μm

Laboratory

External Micrometer GCTS-SOP 06

based on BS EN 870:2008

Up to 100 mm 2 μm Laboratory

Dial Gauges GCTS -SOP 07

based on BS 907:2008

Up to 25 mm 4 μm Laboratory

Below 4 mm 4 µm Laboratory

Above 4 mm 20 µm Laboratory/

Customer

Premises

Performance Testing for

Laboratory Test Sieves

ASTM E2427 & UKAS LAB

22

Below 4 mm 1% Laboratory/

Customer

Premises

Feeler Gauges GCTS-SOP-39 based on BS

957:2008

0.03 mm – 1.00 mm 10 µm Laboratory

GCTS-SOP-14 based on

ASTM E11-17 / ISO 3310-

2:2013(E) / ISO 3310-

1:2016(E)

Laboratory Test Sieves

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26661 of 2

Page 8: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeDimensional Calibration

LB-CAL-013

Issue no.: 04

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

All types of Depth

Gauges.(Dial/ Vernier

/Electronic)

GCTS-SOP-43 based on BS

EN ISO 13385-01:2011

Up to 100 mm 10 µm Laboratory

Depth Micrometers GCTS-SOP-44 based on BS

6468:2008

Up to 100 mm 10 µm Laboratory

Profile Projectors GCTS-SOP-45 based on JIS

B 7184 : 1999

Up to 300 mm ( (x-y axis) 2 µm (Linear), 10 min of

ARC (Angular) & 0.5%

(Magnification)

Laboratory/

Customer

Premises

Bore Gauge GCTS-SOP-47 Up to 2 mm travel 3 µm Laboratory

Extensometers: GCTS-SOP-48 based on

ASTM-E-83 & ISO 9513

Up to 50 mm (Extension) &

Gauge Length- up to 600

mm

5 µm Customer

Premises

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26662 of 2

Page 9: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeForce Calibration

LB-CAL-013

Issue no.: 05

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Force/ Material Testing

Machines in Compression

Verification/ Calibration of

the force measuring system

by force proving

instruments in

compression, acc. ISO 7500-

1: 2018

Alignment of upper

machine platen acc. EN

12390-4:2019

GCTS-SOP-02

300 kN up to 3000 Kn 0.27 % of reading Customer

Premises

Force/ Flexural Strength

Material Testing Machines

Verification/Calibration of

the force measuring system

by force proving

instruments in

compression, acc. ISO 7500-

1: 2018

GCTS-SOP-02

10 kN up to 250 kN 0.27 % of reading Customer

Premises

Force/ Proving Ring, Load

cell with indicator for CBR,

Marshall testing machine

Verification/ Calibration of

the force measuring system

by force proving

instruments in

compression, acc. ISO 7500-

1: 2018

GCTS-SOP-30

5 kN to 50 kN 1% of reading Laboratory/

Customer

Premises

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26661 of 1

user
Highlight
user
Highlight
Page 10: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopePressure Calibration

LB-CAL-013

Issue no.: 05

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gas Pressure

(Negative Gauge)/ Digital

& Analogue Indicating

Devices

Comparison method, acc.

DKD R-6.1:2014

GCTS-SOP-04

100 kPa to 0 kPa 0.05 kPa Laboratory

Gas Pressure (gauge)/

Digital & Analogue

Indicating Devices

Comparison method, acc.

DKD R-6.1:2014

GCTS-SOP-03

0.1 to 3.5 MPa 1 kPa Laboratory

Hydraulic Pressure

(gauge)/ Digital &

Analogue Indicating Devices

Pressure generated by dead

weight tester, acc. DKD R-

6.1:2014 and EURAMET cg

17 v4: 2019

GCTS-SOP-31

0.1 MPa to 140 MPa 8 x 10 -5 pe + 50 Pa Laboratory

Hydraulic Pressure (gauge)

/ Digital & Analogue

Indicating Devices

Comparison method, acc.

DKD R-6.1:2014

GCTS-SOP-03

0.1 MPa to 300 MPa 0.25% of F.S Laboratory/

Customers

Premises

Gas Pressure (gauge)/

Digital & Analogue

Indicating Devices

Pressure generated by dead

weight tester, acc. DKD R-

6.1:2014 and EURAMET cg

17 v4: 2019

GCTS-SOP-031

0.1 MPa to 12 MPa 0.01 % of FS Laboratory

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26661 of 1

Page 11: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeMass Calibration

LB-CAL-013

Issue no.: 04

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

1 mg up to 5 kgOIML

class F2

1/3 of MPE Class F2 Laboratory

10 kg to 20 kg OIML

class M1

1/3 of MPE Class M1 Laboratory

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Valid to: 10-09-2022 Date: 30-03-2021

Mass

Conventional Mass

Substitution Weighing

with Air buoyancy Corr.

OIML R 111

[email protected]

www.eiac.gov.aeOne Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

1 of 1

Page 12: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeVolume Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

1 to 100 µl 0.19 μl

100 to 500 µl 0.22 μl

500 to 1000 µl 0.30 μl

1 to 10 ml 3.3 µl

1 to 50 ml 0.05 ml

> 50 ml to 500 ml 0.41 ml

˃500 to 2000 ml 1.3 ml

1 to 10 ml 0.06 ml

5 to 25 ml 0.3 ml

10 to 100 ml 0.73 ml

25 to 250 ml 0.85 ml

50 to 500 ml 3.2 ml

100 to 1000 ml 6.7 ml

200 to 2000 ml 10 ml

0.1 to 1 ml 0.012 ml

>1 ml to 10 ml 0.02 ml

˃10 ml to 50 ml 0.04 ml

Volumetric Glass Wares

(Measuring cylinders)

GCTS-SOP-25 based on

ISO 4787 & EURAMET cg19

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Volumetric Glass Wares

(Graduated Pipettes.)

GCTS-SOP-25 based on

ISO 4787 & EURAMET cg19

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Laboratory

Laboratory

Laboratory

Laboratory

Piston Operated Volumetric

Apparatus (Micro-Pipettes

& Pipettes)

GCTS-SOP-51 based on

ISO 8655-6, ISO/TR 20461

& ISO/TR 20461

Corr1:2008

Volumetric Glass Wares

(Measuring flasks)

GCTS-SOP-25 based on

ISO 4787 & EURAMET cg19

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26661 of 2

Page 13: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeVolume Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

LaboratoryPiston Operated Volumetric GCTS-SOP-51 based on Volumetric Glass Wares

(Bulb pipettes-Single

Volume)

GCTS-SOP-25 based on

ISO 4787 & EURAMET cg19

1 ml to 50 ml 0.03 ml Laboratory

Volumetric Glass Wares

(Graduated Burettes.)

GCTS-SOP-25 based on

ISO 4787 & EURAMET cg19

0.1 ml 1 ml to 50 ml 0.03 ml Laboratory

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26662 of 2

Page 14: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation ScopeHumidity Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

10- 50 °C 0.4 °C

35- 80 % rH at 20°C 3 % rH

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Direct Reading Thermo /

Hygrometers including

Data loggers

GCTS-SOP-27 Laboratory

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

[email protected]

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 26661 of 1

Page 15: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

0 mV to 600 mV 0.24 mV

>0.60 V to 6 V 8.9 mV

˃6.0 V to 60 V 89 mV

˃60.0 V to 600 V 0.89 V

˃600 V to 1000 V 4.12 V

0 mV to 100 mV 0.022 mV

˃ 0.10 V to 1 V 0.13 mV

˃ 1.0 V to 10 V 1.3 mV

˃ 10.0 V to 100 V 90 mV

˃ 100.0 V to 1000 V 0.13 V

DC High Voltage Direct Method using Fluke

179 DMM & High Voltage

Probe

U = Measured voltage value

1 kV to 30 kV 20 x 10-3 U Laboratory/

Customers

Premises

0 mV to 600 mV

@45 Hz to 1 kHz

0.46 mV

˃ 0.6 V to 6 V

@45 Hz to 1 kHz

0.073 V

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

DC Volatage

Direct Method using Fluke

179 DMM

DC Voltage Laboratory/

Customers

Premises

Calibration of calibrators

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Laboratory/

Customers

Premises

Direct Method using Fluke

179 DMM

AC Voltage

Customers

Premises

Direct Method using

Transmille 1000 A Portable

Multi-Function Calibrator

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

1 of 17

Page 16: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

AC Voltage Direct Method using Fluke

179 DMM

˃ 6.0 V to 60 V

@45 Hz to 1 kHz

0.73 V Laboratory/

Customers

˃ 60.0 V to 600 V

@45 Hz to 1 kHz

7.3 V

˃600 V to 1000 V

@45 Hz to 1 kHz

15.59 V

AC High Voltage Direct Method using Fluke

179 DMM & High Voltage

Probe

U = Measured voltage value

1 kV to 28 kV 20 x 10-3 U Laboratory/

Customers

Premises

1 Hz to 99.99 Hz 0.13 Hz

100.0 Hz to 999.9 Hz 1.3 Hz

1 kHz to 9.99 kHz 12.9 Hz

10.00 kHz to 99.99 kHz 130.0 Hz

0 Ω to 600 Ω 6.4 Ω

˃ 600 Ω to 6 kΩ 64 Ω

˃ 6.0 k Ω to 60 kΩ 0.64 kΩ

˃ 60.0 k Ω to 600 kΩ 6.4 kΩ

Laboratory/

Customers

Premises

Direct Method using Fluke

179 DMM

Resistance

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Laboratory/

Customers

Premises

Direct Method using Fluke

179 DMM

Frequency

Calibration of calibrators

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

2 of 17

Page 17: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

˃ 600 k Ω to 6 MΩ 64 kΩ

˃ 6.0 M Ω to 50 MΩ 0.54 MΩ

0 mA to 60 mA 0.73 mA

˃ 60.0 mA to 400 mA 5.0 mA

˃ 0.4 A to 6 A 73 mA

˃ 6.0 A to 10 A 0.15 A

0 mA to 60 mA

@45 Hz to 1 kHz

1.1 mA

˃ 60.0mA to 400 mA

@45 Hz to 1 kHz

7.3 mA

˃ 0.40 A to 6 A

@45 Hz to 1 kHz

0.11 A

˃ 6.0 A to 10 A

@45 Hz to 1 kHz

0.21 A

RTD

PT 100

Direct Method using Beamex

MC6 Multi-Function

Calibrator

-200 °C to 850 °C 0.5 °C Laboratory/

Customers

Premises

Thermocouple

K Type Calibration of

calibrators

Direct Method using

BEAMEX MC6 Multi-

Function Calibrator

-200 °C to 1370 °C 0.6 °C Laboratory/

Customers

Premises

Direct Method using Fluke

179 DMM

DC Current Laboratory/

Customers

Premises

Laboratory/

Customers

Premises

Direct Method using Fluke

179 DMM

AC Current

Calibration of calibrators

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Direct Method using Fluke

179 DMM

Resistance Laboratory/

Customers

Premises

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

3 of 17

Page 18: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

Thermocouple

J Type

Direct Method using Beamex

MC6 Multi-Function

Calibrator

-210 °C to 1200 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

T Type

Direct Method using Beamex

MC6 Multi-Function

Calibrator

-250 °C to 400 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

R Type Calibration of

calibrators

Direct Method using Beamex

MC6 Multi-Function

Calibrator

0 °C to 1760 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

S Type

Direct Method using Beamex

MC6 Multi-Function

Calibrator

0 °C to 1760 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

N Type

Direct Method using Beamex

MC6 Multi-Function

Calibrator

-200 °C to 1300 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

B Type

Direct Method using Beamex

MC6 Multi-Function

Calibrator

600 °C to 1820 °C 0.6 °C Laboratory/

Customers

Premises

Calibration of calibrators

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

4 of 17

Page 19: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

Thermocouple

E Type

Direct Method using

Beamex MC6 Multi-Function

Calibrator

-250 °C to 1000 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

L Type

Direct Method using Beamex

MC6 Multi-Function

Calibrator

-200 °C to 900 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

U Type

Direct Method using

Beamex MC6 Multi-Function

Calibrator

-200 °C to 600 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

C Type

Direct Method using

Beamex MC6 Multi-Function

Calibrator

0 °C to 2300 °C 0.6 °C Laboratory/

Customers

Premises

0 mV to 100 mV 0.2 mV

˃ 0.10 V to 1 V 0.77 mV

˃ 1.0 V to 10 V 14 mV

˃ 10.0 V to 100 V 0.17 V

˃ 100.0 V to 1000 V 2.4 V

Direct Method using

Transmille 1000A Portable

Multi-Function Calibrator

AC Voltage Customers

Premises

Calibration of calibrators

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Calibration of instruments

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

5 of 17

Page 20: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

0.0 mV to 200 mV

@ 45 Hz to 1 kHz

0.14 mV

˃0.20 V to 2 V

@ 45 Hz to 1 kHz

1.2 mV

˃ 2.0 V to 20 V

@ 45 Hz to 1 kHz

12 mV

˃ 20.0 V 200 V

@ 45 Hz to 1 kHz

0.14V

˃ 200.0 V to 1000 V

@ 45 Hz to 1 kHz

0.72 V

0.0 mV to 200 mV 0.016 mV

˃0.20 V to 2 V 0.2 mV

˃ 2.0 V to 20 V 1.5 mV

˃ 20.0 V 200 V 15 mV

˃ 200.0 V to 1000 V 81 mV

1 Hz to 100 Hz 0.0023 Hz

>100Hz to 1 kHz 0.02 Hz

>1 kHz to 10 kHz 0.23 Hz

>10 kHz to 20 kHz 0.46 Hz

>20 kHz to 50 kHz 1.2 Hz

>50 kHz to 100 kHz 2.3 Hz

Direct Method using

Transmille 3050A/ 3041A

Multi-Function Calibrator

AC Voltage Laboratory

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Direct Method using

Transmille 1000A Portable

Multi-Function Calibrator

Frequency Customers

Premises

Calibration of instruments

Direct Method using

Transmille 3050A/ 3041A

Multi-Function Calibrator

DC Voltage Laboratory

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

6 of 17

Page 21: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

1 Hz to 100 Hz 0.0023 Hz

˃10 Hz to 1 kHz 0.023 Hz

˃ 1.0 kHz to 10.0 kHz 0.23 Hz

˃ 10.0 kHz to 100 kHz 2.3 Hz

˃ 100.0 kHz to 1.0 MHz 20 Hz

˃ 1.0 MHz to 10 MHz 0.23 kHz

0 µA to 100 µA 0.07 μA

˃ 0.10 mA to 1 mA 0.46 μA

˃ 1.0 mA to 10 mA 4.6 μA

˃ 10.0 mA to 100 mA 46 µA

˃ 0.10 A to 1 A 0.52 mA

˃ 1.0 A to 10 A 8.3 mA

DC Current Direct Method using

Portable Multi-Function

Calibrator

(Transmille 1000A) & 2, 10

& 50 turn current coil

I = Measured current value

10 A to1000 A 10 x 10-3 I Customers

Premises

Frequency Laboratory

Customers

Premises

Direct Method using

Transmille 1000A Portable

Multi-Function Calibrator

DC Current

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Calibration of instruments

Direct Method using

Transmille 3050A/ 3041A

Multi-Function Calibrator

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

7 of 17

Page 22: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

0.0 µA to 200 µA 0.05 μA

˃ 0.20 mA to 2 mA 0.3 μA

˃ 2.0 mA to 20 mA 3.3 μA

˃ 20.0 mA to 200 mA 40 µA

˃ 0.20 A to 2 A 1.3 mA

˃ 2.0 A to 30 A 13 mA

DC Current Direct Method using

Portable Multi-Function

Calibrator

(TRANSMILLE

3050A/3041A) & 2, 10 & 50

turn current coil

I = Measured current value

10 A to 1500 A 10 x 10-3 I Laboratory

0 µA to 100 µA @45 Hz to 1

kHz

0.6 μA

˃ 0.10 mA to 1 mA

@45 Hz to 1 kHz

2.6 μA

Direct Method using

Transmille 1000A Portable

Multi-Function Calibrator

AC Current Customers

Premises

Calibration of instruments

Direct Method using

Transmille 3050A/ 3041A

Multi-Function Calibrator

LaboratoryDC Current

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

8 of 17

Page 23: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

˃ 1.0 mA to 10 mA

@45 Hz to 1 kHz

22 µA

˃ 10.0 mA to 100 mA

@45 Hz to 1 kHz

0.22 mA

˃ 0.10 A to 1 A

@45 Hz to 1 kHz

2.2 mA

˃ 1.0 A to 10 A

@45 Hz to 1 kHz

0.03 A

AC Current Direct Method using

Portable Multi-Function

Calibrator

Transmille 1000A) & 2, 10 &

50 turn current coil

I = Measured current value

10 A to1000 A

@45 Hz to 200 Hz

10 x 10-3 I Customers

Premises

AC Current Direct Method using

Portable Multi-Function

Calibrator

Transmille 3050A/3041A) &

2, 10 & 50 turn current coil

I = Measured current value

10 A to 1500 A

@45 Hz to 200 Hz

10 x 10-3 I Laboratory

Calibration of instruments

Direct Method using

Transmille 1000A Portable

Multi-Function Calibrator

AC Current Customers

Premises

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

9 of 17

Page 24: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

0.0 µA to 200 µA

@45 Hz to 1kHz

0.52 μA

˃ 0.20 mA to 2 mA

@45 Hz to 1kHz

2.2 μA

˃ 2.0 mA to 20 mA

@45 Hz to 1kHz

21 µA

˃ 20.0 mA to 200 mA @45

Hz to 1kHz

0.21 mA

˃ 0.20 A to 2 A

@45 Hz to 1kHz

2.7 mA

˃ 2.0 A to 30 A

@45 Hz to 1kHz

56 mA

10 nF 0.1 nF

100 nF 1.0 nF

1 µF 20 nF

10 nF 0.052 nF

20 nF 0.11 nF

50 nF 0.24 nF

100 nF 0.48 nF

1 µF 4.7 nF

Customers

Premises

Direct Method using

Transmille 1000A Portable

Multi-Function Calibrator

Capacetance

Direct Method using

Transmille 3050A/ 3041A

Multi-Function Calibrator

(Fixed Values only)

Capacitance Laboratory

Calibration of instruments

Direct Method using

TRANSMILLE 3050A/

3041A Multi-Function

Calibrator

AC Current

45 Hz to 1kHz

Laboratory

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

10 of 17

Page 25: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

0 Ω to 10 Ω 0.07 Ω

˃ 10 Ω to 100 Ω 0.08 Ω

˃ 100.0 Ω to 1 kΩ 0.29 Ω

˃1.0 kΩ to 10 kΩ 2.4 Ω

>10.0 KΩ to 100 KΩ 0.024 kΩ

>100.1 KΩ to 1 MΩ 0.25 kΩ

˃1 MΩ to 10 MΩ 4.1 kΩ

10 Ω 13 mΩ

100 Ω 16 mΩ

1 kΩ 0.12 Ω

10 kΩ 1.2 Ω

100 kΩ 12 Ω

1 MΩ 0.22 kΩ

10 MΩ 7.4 kΩ

100 MΩ 0.75 MΩ

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Direct Method using

Transmille 3050A/ 3041A

Multi-Function Calibrator

Resistance

Fixed values only-2 WIRE

Laboratory

Calibration of instruments

Customers

Premises

Simulation Method using

Transmille 1000A

Portable Multi-Function

Calibrator

Resistance

Resistance

Fixed values only-2 WIRE

Direct Method using

Transmille 3050A/ 3041A

Multi-Function Calibrator

Laboratory

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

11 of 17

Page 26: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

0.01 MΩ to 5 MΩ 10 x 10-3 R

˃ 5 MΩ to 2 GΩ 39 x 10-3 R

100mΩ to 900 mΩ 40 mΩ

1 Ω to 2 Ω 50 mΩ

˃2 Ω to 900 Ω 20 x 10-3 R

1 kΩ to 50 kΩ 20 x 10-3 R

10 mA to 100 mA 2.0 mA

>100 mA to 300 mA 5.0 mA

AC Voltage Out Direct Method using 3200B

Electrical Test Calibrator

U = Measured voltage value

100 V to 400 V@50 Hz 5.0 x 10-3U Laboratory

Insulation Resistance Voltage

Measure

Direct Method using 3200B

Electrical Test Calibrator

U = Measured voltage value

50 V to 1000 V 30 x 10-3 U Laboratory

Calibration of instruments

LaboratoryDirect Method using 3200B

Electrical Test Calibrator

R = Measured resistance

value

Continuity Resistance

Direct Method using 3200B

Electrical Test Calibrator

Continuity Current Laboratory

Direct Method using 3200B

Electrical Test Calibrator

R = Measured resistance

value

Insulation Resistance Laboratory

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

12 of 17

Page 27: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

Insulation Resistance DC

Current Measurement

Direct Method using 3200B

Electrical Test Calibrator

0.5 mA & 1 mA 20 µA Laboratory

0.05 Ω to 10.0 Ω 0.5 Ω

100 Ω 6.0 Ω

900.0 Ω 10 Ω

2 mA to 10 mA 0.2 mA

10.01 mA to 30 mA 0.5 mA

30.01 mA to 100 mA 1.6 mA

100.01 mA to 300 mA 4.5 mA

300.01 mA to 1000 mA 15 mA

1000.01 mA to 3000 mA 30 mA

20 ms to 390 ms 1.3 ms

900 ms 20 ms

PAT Insulation Resistance Direct Method using 3200B

Electrical Test Calibrator

R = Measured resistance

value

1 MΩ to 10 MΩ 10 x 10-3 R Laboratory

LaboratoryDirect Method using 3200B

Electrical Test CalibratorR

RCD Trip Time

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Calibration of instruments

Direct Method using 3200B

Electrical Test Calibrator

LaboratoryLoop Impedance Resistance

Measurement

Direct Method using 3200B

Electrical Test Calibrator

RCD Current Laboratory

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

13 of 17

Page 28: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

0.05 Ω to 1.0 Ω 20 mΩ

˃1 Ω to 900.0 Ω 20 x 10-3 R

0 mA to 500 mA 16 mA

1 A to 10 A 0.2 A

11 A to 30 A 0.6 A

PAT Leakage Current Direct Method using 3200B

Electrical Test Calibrator

2 mA to 8 mA

@ 240 V

20 µA Laboratory

RTD

(PT 100 Source Mode)

Simulation Method using

Beamex MC6 Multi-Function

Calibrator

­ 200 °C to 850 °C 0.5 °C Laboratory/

Customers

Premises

Thermocouple

K Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

-200 °C to 1370 °C 0.6 °C Laboratory/

Customers

Premises

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Calibration of instruments

Direct Method using

Electrical Test Calibrator

R = Measured resistance

value

PAT Earth Bond Resistance Laboratory

Direct Method using

Electrical Test Calibrator

Earth Bond Current

Measurement

Laboratory

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

14 of 17

Page 29: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

Thermocouple

J Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

-210 °C to 1200 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

T Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

-250 °C to 400 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

R Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

0 °C to 1760 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

S Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

0 °C to 1760 °C 0.6 °C Laboratory/

Customers

Premises

Calibration of instruments

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

15 of 17

Page 30: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

Thermocouple

N Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

-200 °C to 1300 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

B Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

600 °C to 1820 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

E Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

-250 °C to 1000 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

L Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

-200 °C to 900 °C 0.6 °C Laboratory/

Customers

Premises

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Calibration of instruments

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.

The Trade Centre District, Dubai-United Arab Emirates

Office: +971 4 872 2666

16 of 17

Page 31: LB-CAL-013 · 2021. 5. 29. · Mass/ Electronic Balances Use of mass standards, acc. to EURAMET cg 18:2015; ASTM E898-20 GCTS-SOP-01 Customers Premises Gulf Calibration & Technical

Accreditation Scope

Electrical Calibration

LB-CAL-013

Issue no.: 02

Calibration Field/

Measuring QualityCalibration Method Range and Specification

Calibration Measurement

Capability

(CMC)*

Location

Gulf Calibration & Technical Services

Ghala Industrial Area, Muscat- Sultanate of Oman

Date: 30-03-2021 Valid to: 10-09-2022

Calibration of calibrators

Thermocouple

U Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

-200 °C to 600 °C 0.6 °C Laboratory/

Customers

Premises

Thermocouple

C Type

Simulation Method using

Transmille 3050A Multi-

Function Calibrator &

Thermocouple Simulator/

Beamex MC6 Multi-Function

Calibrator

0 °C to 2300 °C 0.6 °C Laboratory/

Customers

Premises

RPM

Tachometers

Non-Contact

GCTS-SOP-54 using

Transmille EA015 Multi-

Function Workstation

240 rpm to 20,000 rpm 2 rpm Laboratory

* Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing

more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded

uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration

performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

Calibration of instruments

[email protected]

www.eiac.gov.ae

One Central-Offices 2- Level 3, Suit No.: 302, 303 308.