Laser-Sintering of PEEK The development of cranial implant geometry IUM 2011, Bad Wörishofen, April 2011 Jörg Lenz

Transcript of Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of...

Page 1: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

Laser-Sintering

of PEEK The development of cranial implant geometry

IUM 2011, Bad Wörishofen, April 2011

Jörg Lenz

Page 2: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbH

design evaluationmechanical properties and FEMpolymer infiltration trialssterilisation and biocompatibility testinganimal studiessocio-economic evaluation

EOS AgendaLaser-sintering of PEEK

1

2

3

4

EOS and Custom-IMD

EOSINT P 800 (High Temperature Laser-Sintering System)

Cranial implant geometry

Conclusions

Page 3: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbH

EOS and Custom-IMD

Cranial, dental, spinal

22 partners from 7 European countries (12 SMEs, 8 Research institutions, 2 hospitals)

EOS was involved in the cranial case

SME supply chain integration for enhanced fully customizable implants

A research project funded by the EC in the FP6 program

www.customimd.eu

Page 4: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbH

EOS and Custom-IMD

Surgeons chose PEEK over PEEKHT due to years of proven implant performance and existing FDA Masterfile

one of the bio-passive (non-interacting with body tissues) polymers of particular interest to the project

biocompatible; x-ray and CT translucent; chemically resistant / bio-stable; easily sterilizable (radiation, gas and steam); has excellent wear properties and does not induce osteolysis (macrophage induce bone degeneration)

Why PEEK (polyetherketone)?

Source: EOS GmbH

www.customimd.eu

Page 5: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbH

EOSINT P 800

www.customimd.eu

Developed for processing high temperature melting polymers

High Temperature

Laser-Sintering

System

Supported

by the EC in the Sixth Framework program

Source: EOS GmbH

Page 6: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbH

EOSINT P 800

Optical windowExchangeable build frameSystem GUI

High Temperature

Laser-Sintering

System

Exposure Recoating device Front window

Source: EOS GmbH

Page 7: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Ti Cranial Implant

Cranial

implant

geometryDesign evaluation

Idea: a solid PEEK plate similar to the titanium bone plates already known

Page 8: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Cranial

implant

geometryDesign evaluation

design from Ti could not be transferred 1:1thickness and fixing had to be modified (patent pending AZM)

First version

Page 9: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbH

Design evaluation

Source: EOS, AZM

Cranial

implant

geometryDesign evaluation

PEEK scaffold for improved bone in-growthvariation of different mesh designs and sizesgoals: high strength, small deformation, infiltration ability, bone growth

Second version

Page 10: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Cranial

implant

geometryDesign evaluation

various software applications, new mesh designsolid border (rim) for better fitting into the skull and higher strengthcontour following mesh – no stress peaks

Iterations

Page 11: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Cranial

implant

geometryDesign evaluation

design with discontinuous rimhigh strength, low deformationperfect fit and excellent bone in-growth expected

Final version

Page 12: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS

According

to DIN EN ISO 527-1A4.184[MPa]Tensile

modulus

According

to ISO 178108[MPa]Flexural

strength

1,30[g/cm³]density

According

to DIN EN ISO 527-1A

72,73 (e=1,95%)[MPa]Tensile

strength

According

to ISO 604133[MPa]Compression

strength

RemarksValueUnitValue

Cranial

implant

geometryMechanical properties and FEM

Values evaluated with standard test coupons, considering anisotropy

Page 13: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, Within

Cranial

implant

geometryMechanical properties and FEM

Stress distribution when applying 70 MPa

Page 14: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, Within

Cranial

implant

geometry

Page 15: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, Within

Cranial

implant

geometryMechanical properties and FEM

Displacement under load (70 MPa)

Page 16: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, Within

Cranial

implant

geometry

Page 17: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: ASCAMM

Cranial

implant

geometryMechanical properties and FEM

results range from 16-20 MPa for first mesh up to >100 MPa for the final version

Experimental testing

Page 18: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, Suprapolix, Xpand

Cranial

implant

geometry

SupraB bio-absorbable polymer/osteoconductive hydroxylapatite compound infiltration into different mesh geometry

Polymer infiltration trials

Polymer infiltration

Page 19: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, TNO, Suprapolix, Xpand

Cranial

implant

geometry

Compound infiltration into the final geometry by hot pressing performed by TNOOnly possible due to the unprecedented low melt viscosity of the SupraB polymer

Polymer infiltration in real geometry

Polymer infiltration

Page 20: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, LasMed, Rapra

Value Unit Beforesterilization

Aftersterilization Remarks

Compression

strength [MPa] 133 137 According

to ISO 604

Tensile

strength [MPa] 72,73 (e=1,95%)

71,75 (e=1,95%) According

to DIN EN ISO 527-1A

density [g/cm³] 1,30 1,30

Tensile

modulus [MPa] 4.184 4.107 According

to DIN EN ISO 527-1A

Flexural

strength [MPa] 108 111 According

to ISO 178

Cranial

implant

geometrySterilization

PEEK samples of representative shape, from laser-sintering, cleaned, vacuum packed6MeV EB accelerator at TU of Lodz, standard sterilization dose of 25 kGy ± 0.6 kGy

despite the fact that PEEK can withstand all common sterilization techniques without abrupt changes in its properties

that was mainly due to high reliability, short time of sterilization and parametric release of EB technique

Page 21: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, LasMed, Rapra

Instrument: Polymer Laboratories GPC120, with PL GPC-AS MT heated autosamplerColumns: PLgel guard plus 2 x mixed bed-B, 30 cm, 10 μmSolvent: 50:50 (w/w) 1,2,4-trichlorobenzene:phenol with anti-oxidantFlow-rate: 0.8 mL/min (nominal) Temperature: PL GPC120: 115oC (nominal) PL GPC-AS MT: 80oC (nominal) Detector: refractive index

Cranial

implant

geometrySterilization

“…

the sterilization process used in this case has not affected the molecular weight of the PEEK samples and so it can be concluded that no detectable degradation (chain scission or cross-linking) has resulted.”

Chemischer Verification: High temperature Gel Permeation Chromatography (GPC)

Page 22: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, LasMed, Rapra

“It has not been possible using either ATR IR or Pyrolysis

IR, to detect any chemical differences between the sterilized and unsterilized PEEK samples. There are no apparent differences in molecular weight, or molecular weight distribution, between the sterilized and unsterilized PEEK samples as well.”

sterilized unsterilized

Cranial

implant

geometrySterilization

Infrared spectroscopy (ATR IR)

Page 23: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, LasMed, Rapra

Cranial

implant

geometrySterilization

The conclusions withdrawn stated that laser-sintered PEEK material has perfect properties for application as biomaterial for various purposes that include cranial implant

Sterilisation by EB irradiation at 25 kGy does not affect the physical and chemical properties of the polymer, nor influence in any way it faultless biocompatibility and bioinertness

Conclusion

Page 24: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, LasMed, Tecnalia

Cranial

implant

geometry

non-cytotoxic according to ISO standard 10993-5 “Biological evaluation of medical devices - Part 5: Test for in vitro cytotoxicity”

non-haemolytic properties according to ISO standard 10993-4 “Biological evaluation of medical devices - Part 4: Selection of test for interactions with blood” and ASTM 756-00 “Standard Practice for Assessment of Hemolytic Properties of Materials”.

not causing pyrogenicity according to FDA limits, procedure following the recommendations in ISO standard 10993-12 “Sample Preparation and Reference Materials”

non-irritant according to Inasmet inner procedure, based on ISO 10993-11, and EpiDermTM kit supplied from MatTek corporation company (USA)

This test was validated by ECVAM (European Centre for the Validation of Alternative Methods) in November 2008, in replacement of in vivo rabbit Draize test

not causing any sensitization response according to ISO standard 10993-10

Summarized results of the biocompatibility testing

Page 25: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM, Suprapolix, Xpand

Cranial

implant

geometryAnimal testing

solid,

Open mesh,

mesh with compound infiltration

Geometry to be implantedin mini pigs skull

Page 26: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Cranial

implant

geometryAnimal testing

Page 27: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Animal testing

Cranial

implant

geometry

Page 28: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Animal testing

Cranial

implant

geometry

Page 29: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Animal testing

Cranial

implant

geometry

Page 30: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Animal testing

Cranial

implant

geometry

Page 31: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS

Micro-CT of samples

Cranial

implant

geometryAnimal testing

Page 32: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: University Hasselt

Cranial

implant

geometryAnimal testing

Analysis of surface roughnessScanning Electron Microscopy

Page 33: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS

Cranial

implant

geometry

PEEK Mash and Composite

Animal testing

Page 34: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

improve health gain reduced risk of side-effects

increase patient satisfaction customised treatment & reduced side-effects

reduce costs reduced production costs (via reduced labour & material costs);attractive to hospitals and other stakeholders

be more cost-effective potentially cheaper & greater/same effectiveness

become the preferred treatment combination of: costs, health gain, cost-effectiveness

Compared to current care, a custom IMD implant has the potential to:

Cranial

implant

geometrySocio-economic evaluation

Page 35: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS, AZM

Conclusions

Laser-Sintering of PEEK implant geometry was successfully performed for the first time.

All requirements were met regarding strength, sterilizations and biocompatibility.

Scaffold implant structures could be built that are impossible using conventional technology.

The concept of hybrid cranial implants (composite of polymers and ceramics) opens new ways in the reconstruction of skull defects).

Introduction of laser-sintered PEEK, polymer-ceramic compositesand new design techniques was proven to be an excellent combination for reconstruction of skull defects.

Page 36: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbHSource: EOS

Thanks

to…

AZM University hospital Maastricht (The Netherlands) & the University of Hasselt (Belgium) for the medical advise and analysis.

Suprapolix & Xpand biotechnology (The Netherlands) for providing bio-absorbable polymers and bioceramics.

ASCAMM (Spain) for the biomechanical validation of cranial implants.

TNO (The Netherlands) for the scaffold ideas and the infiltration trials.

LasMed (Poland), Tecnalia (Spain) and Rapra (United Kingdom) for providing consultancy and tests on sterilization methods and biocompatibility.

Within (United Kingdom) for providing designs and FEM analysis.

Partners

Page 37: Laser-Sintering of PEEK · The development of cranial implant geometry. IUM 2011, ... short time of sterilization and parametric release of EB technique . EOS ... Laser-Sintering

EOS 2011 – Joerg Lenz

Source: EOS GmbH

Laser-Sintering

of PEEK The development of cranial implant geometry