LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian...

23
This is an official electronic document and secured by BMD Laboratory ©BMD Laboratory 2016 BMD LaboratoryProvider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected] LAPORAN PROGRAM UJI BANDING (PROFICIENCY TEST) ANTAR LABORATORIUM SKEMA KHUSUS SEMESTER I-2016 Bahan Uji: AIR MINUM DALAM KEMASAN I BATCH 3 (AMDK I-3) Ref. No. 002/UR/BMD/2016 PARAMETER UJI: Nitrat (NO 3 - ), Sulfat (SO 4 2- ), dan Klorida (Cl - ) FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/03/02-01-2016

Transcript of LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian...

Page 1: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

This is an official electronic document and secured by BMD Laboratory

©BMD Laboratory 2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

LAPORAN

PROGRAM UJI BANDING (PROFICIENCY TEST) ANTAR LABORATORIUM

SKEMA KHUSUS SEMESTER I-2016

Bahan Uji:

AIR MINUM DALAM KEMASAN I BATCH 3 (AMDK I-3)

Ref. No. 002/UR/BMD/2016

PARAMETER UJI:

Nitrat (NO3-), Sulfat (SO4

2-), dan Klorida (Cl

-)

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Page 2: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page i of iii

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

Ref. No. 002/UR/BMD/2016

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

KATA PENGANTAR

Puji syukur kiranya kita panjatkan kepada Alloh SWT yang berkenan melimpahkan

karunia-Nya sehingga ‘Program Uji Banding Antar Laboratorium (Proficiency Test) Skema

Khusus Semester I-2016’ untuk bahan uji Air Minum dalam Kemasan I Batch 3 (AMDK I-3)

dengan parameter uji nitrat (NO3-), sulfat (SO4

2-), dan klorida (Cl

-) telah berhasil dilaksanakan

dan diselesaikan sampai dengan diterbitkannya laporan ini.

Program uji banding antar laboratorium merupakan salah satu instrumen yang sangat

penting sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium

pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ISO/IEC 17025:2008. Dengan

mengikuti program uji banding, laboratorium dapat melakukan evaluasi kinerja pengujian yang

dilakukan dalam rangka melakukan peningkatan secara berkesinambungan (continuous

improvement) untuk memberikan layanan terbaik kepada pelanggan.

Program skema khusus kali ini diikuti oleh 10 (sepuluh) laboratorium dengan 9 (sembilan)

diantaranya telah terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditas Nasional (KAN).

Hasil pengolahan data pada skema uji banding AMDK I-3 ini sangat memuaskan secara

statistika. Sebagai SDPA (simpangan baku yang digunakan dalam evaluasi) untuk ketiga

parameter digunakan SD Robust yaitu simpangan baku yang diperoleh dari pengolahan data

secara statistika menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E).

Dalam program uji banding ini masih digunakan nilai konsensus (consensus value) dari

laboratorium peserta dalam melakukan evaluasi, sehingga tidak memuaskan dalam

ketertelusuran (traceability) secara metrologi. Akan tetapi BMD Laboratory meyakini bahwa hal

tersebut tidak mengurangi manfaat dari pelaksanaan program uji banding ini.

BMD Laboratory sebagai provider uji banding antar laboratorium menyampaikan apresiasi

setinggi-tingginya kepada seluruh pihak yang telah membantu dan menyukseskan kegiatan ini.

Khususnya kepada seluruh laboratorium peserta atas partisipasi aktifnya mengikuti program uji

Page 3: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page ii of iii

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

Ref. No. 002/UR/BMD/2016

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

banding AMDK I-3. Ucapan terimakasih juga disampaikan kepada tim ahli yang telah bekerja

keras mengolah data dan menyusun laporan ini. Akhirnya, BMD Laboratory menyampaikan

permohonan maaf sekiranya terdapat kekurangan dalam penyelenggaraan program uji banding

antar laboratorium. Kritik, saran, dan masukan yang membangun sangat kami harapkan untuk

menghadirkan program uji banding yang semakin baik ke depannya. Semoga laporan ini dapat

dimanfaatkan oleh laboratorium peserta dan pihak-pihak terkait.

Tangerang Selatan, 20 Mei 2016

Ir. Renaldo Moontri

CEO BMD Group

(Selaku Manajer Puncak BMD Laboratory)

Page 4: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page iii of iii

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

Ref. No. 002/UR/BMD/2016

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

DAFTAR ISI

KATA PENGANTAR ......................................................................................................................... i

DAFTAR ISI ................................................................................................................................... iii

I. PENDAHULUAN .................................................................................................................... 1

II. PELAKSANAAN ..................................................................................................................... 1

III. STATISTIKA PENGOLAHAN DATA ......................................................................................... 2

III.1 Penilaian Laboratorium ............................................................................................... 2

III.2 Penetapan Nilai Evaluasi Uji Banding .......................................................................... 2

III.3 Penetapan SDPA .......................................................................................................... 2

IV. HASIL UJI BANDING ANTAR LABORATORIUM ...................................................................... 3

IV.1 Parameter Nitrat (NO3-) .............................................................................................. 6

IV.2 Parameter Sulfat (SO42-

) ............................................................................................. 8

IV.3 Parameter Klorida (Cl-) ............................................................................................. 10

V. PENUTUP ........................................................................................................................... 12

VI. REFERENSI .......................................................................................................................... 12

Annex A Rekapitulasi Hasil z-score Program Uji Banding ......................................................... 13

Annex B Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus ......................................... 14

Annex C Evaluasi Distribusi Normal Data Program Uji Banding ............................................... 16

Page 5: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 1 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

I. PENDAHULUAN

Salah satu kebutuhan dasar manusia adalah air, terutama yang digunakan untuk minum.

Kualitas kesehatan manusia salah satunya sangat ditentukan oleh kualitas air yang

diminumnya. Oleh karena itu, air minum dalam kemasan (AMDK) merupakan salah satu

komoditi yang diatur persyaratan mutunya sesuai SNI 01-3553-2006. Persyaratan mutu

untuk parameter nitrat (NO3-), sulfat (SO4

2-), dan klorida (Cl-) seperti pada Tabel 1.

Tabel 1. Persyaratan mutu air minum dalam kemasan (air mineral)[1]

No. Parameter Satuan Persyaratan

1. Nitrat (NO3-) mg/L Maks. 45 mg/L

2. Sulfat (SO42-) mg/L Maks. 200 mg/L

3. Klorida (Cl-) mg/L Maks. 250 mg/L

Salah satu jaminan mutu bagi laboratorium pengujian dan kalibrasi sesuai persyaratan

dalam SNI ISO/IEC 17025:2008 adalah partisipasi dalam uji banding antar laboratorium atau

uji profisiensi[2] sebagai jaminan mutu eksternal (external quality control). Untuk itu BMD

Laboratory menyelenggarakan program uji banding antar laboratorium (proficiency test)

untuk bahan uji Air Minum dalam Kemasan I Batch 3 (AMDK I-3) dengan tujuan sebagai

berikut:

1. Sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium

peserta khususnya laboratorium yang menerapkan standard SNI ISO/IEC 17025:2008.

2. Membantu laboratorium dalam mengevaluasi unjuk kerja pengujian kadar nitrat (NO3-),

sulfat (SO42-), dan klorida (Cl-) untuk bahan uji air minum dalam kemasan.

II. PELAKSANAAN

Berikut adalah jadual pelaksanaan program uji banding antar laboratorium (proficiency test)

yang diselenggarakan oleh BMD Laboratory untuk bahan uji Air Minum dalam Kemasan I

Batch 3 (AMDK I-3):

Page 6: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 2 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

a. Penutupan pendaftaran: 28 Maret 2016

b. Pendistribusian bahan uji: 11 s.d. 25 April 2016

c. Pelaporan hasil analisis oleh laboratorium peserta: Maks. 09 Mei 2016

d. Pendistribusian laporan program uji banding oleh BMD Laboratory: Maks. 23 Mei 2016

III. STATISTIKA PENGOLAHAN DATA

III.1 Penilaian Laboratorium

Penilaian terhadap unjuk kerja laboratorium menggunakan z-score dengan tiga kriteria

penilaian sebagai berikut:

� Untuk |z-score|≤2,0 dikategorikan memuaskan diberi lambang ‘OK’

� Untuk 2,0<|z-score|<3,0 dikategorikan meragukan diberi lambang ‘$’

� Untuk |z-score|≥3,0 dikategorikan kurang memuaskan diberi lambang ‘$$’

III.2 Penetapan Nilai Evaluasi Uji Banding

Penetapan nilai evaluasi uji banding menggunakan dua pendekatan statistika yaitu:

� Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E)[3]. Nilai tersebut

merupakan nilai konsensus dari peserta (consensus value).

� Seleksi data menggunakan perangkat lunak Distribution AnalyzerTM, Taylor

Enterprises, Inc.[4]

III.3 Penetapan SDPA

SDPA (simpangan baku yang digunakan dalam evaluasi) ditetapkan menggunakan tiga

pendekatan statistika yaitu:

� Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E) atau SD Robust

� Simpangan Baku Horwitz[5] atau SD Horwitz

� SDPA yang ditetapkan (Assigned SDPA)

Page 7: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 3 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

IV. HASIL UJI BANDING ANTAR LABORATORIUM

Sebanyak 10 (sepuluh) laboratorium telah berpartisipasi dalam program uji banding antar

laboratorium (proficiency test) yang diselenggarakan oleh BMD Laboratory untuk bahan uji

Air Minum dalam Kemasan I Batch 3 (AMDK I-3) dengan parameter uji nitrat (NO3-), sulfat

(SO42-), dan klorida (Cl-). Kesembilan laboratorium tersebut merupakan laboratorium

terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditasi Nasional (KAN). Masing-masing

laboratorium diberikan 2 (dua) buah bahan uji dengan kombinasi kode seperti pada Tabel 2.

Tabel 2. Kombinasi Kode Bahan Uji

Rekapitulasi hasil pengujian oleh laboratorium peserta disajikan seperti pada Tabel 3.

Seluruh data hasil uji banding dapat diolah dengan baik menggunakan Robust Statistic

Algoritma A Mean sesuai ISO 13528: 2005(E). Selain itu dilakukan juga evaluasi data

menggunakan perangkat lunak Distribution AnalyzerTM guna memastikan bahwa data yang

diolah mengikuti pola distribusi normal pada tingkat kepercayaan 95%. Nilai evaluasi uji

No. Kode Bahan Uji Status Akreditasi

1. BMD-0201-16B3 BMD-0211-16B3 Ya

2. BMD-0202-16B3 BMD-0212-16B3 Ya

3. BMD-0203-16B3 BMD-0213-16B3 Ya

4. BMD-0204-16B3 BMD-0214-16B3 Ya

5. BMD-0205-16B3 BMD-0215-16B3 Ya

6. BMD-0206-16B3 BMD-0216-16B3 Tidak

7. BMD-0207-16B3 BMD-0217-16B3 Ya

8. BMD-0208-16B3 BMD-0218-16B3 Ya

9. BMD-0209-16B3 BMD-0219-16B3 Ya

10. BMD-0210-16B3 BMD-0220-16B3 Ya

Page 8: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 4 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

banding (assigned value) merupakan nilai konsensus (consensus value) dari laboratorium

peserta sehingga nilai tersebut bukan merupakan nilai acuan yang tertelusur secara

metrologi ke sistem satuan internasional (SI).

Tabel 3. Rekapitulasi Hasil Pengujian Laboratorium Peserta

No. Kode Lab. Nitrat (NO3

-) Sulfat (SO4

2-) Klorida (Cl

-)

mg.L-1

mg.L-1

mg.L-1

1. BMD-0201-16B3 14,24±0,261 19,50±1,586 26,80±1,692

Metoda SNI 01 3554 2006 SNI 01 3554 2006 SNI 01 3554 2006

2. BMD-0202-16B3 14,40±0,15 19,63±1,25 26,23

Metoda SNI 01 3554 2006 SNI 01 3554 2006 SNI 01 3554 2006

3. BMD-0203-16B3 14,22±0,263 19,74±1,603 27,44±1,726

Metoda SNI 01-3554-2006 SNI 01-3554-2006 SNI 01-3554-2006

4. BMD-0204-16B3 14,82±0,02 18,99±0,20 25,30±0,42

Metoda SNI 01-3554-2006 SNI 01-3554-2006 SNI 01-3554-2006

5. BMD-0205-16B3 14,36±0,62 15,97±1,22 29,64±0,65

Metoda Spectrophotometry/ UV

Screening Spectrophotometry/ UV

Turbidity Method Argentometri

6. BMD-0206-16B3 0,44±0,00 24±0,60 39,05±1,43

Metoda Spektrofotometri Spektrofotometri Argentometri

7. BMD-0207-16B3 15,27±0,11 20,68±0,34 23,76±0,03

Metoda APHA 2005 : 4500

NO3-NB SNI 6989.20-2009 SNI 6989.19-2009

8. BMD-0208-16B3 15,12 20,34 27,22

Metoda Ion Chromatography Ion Chromatography Ion Chromatography

9. BMD-0209-16B3 0,36 23,29 23,14

Metoda Spektrofotometri SNI 06-6989-20-2004 SNI 06-6989-19-2009

Page 9: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 5 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

No. Kode Lab. Nitrat (NO3

-) Sulfat (SO4

2-) Klorida (Cl

-)

mg.L-1

mg.L-1

mg.L-1

10. BMD-0210-16B3 14,82±0,84 22,16±1,31 28,86±1,0

Metoda SNI 3554-2015 SNI 3554-2015 SNI 3554-2015

11. BMD-0211-16B3 14,19±0,261 20,39±1,656 27,04±1,703

Metoda SNI 01 3554 2006 SNI 01 3554 2006 SNI 01 3554 2006

12. BMD-0212-16B3 14,41±0,16 19,71±1,26 26,32

Metoda SNI 01 3554 2006 SNI 01 3554 2006 SNI 01 3554 2006

13. BMD-0213-16B3 14,75±0,263 20,82±1,693 26,95±1,698

Metoda SNI 01-3554-2006 SNI 01-3554-2006 SNI 01-3554-2006

14. BMD-0214-16B3 14,88±0,02 18,50±0,20 25,19±0,42

Metoda SNI 01-3554-2006 SNI 01-3554-2006 SNI 01-3554-2006

15. BMD-0215-16B3 13,82±0,60 18,72±1,42 29,83±0,66

Metoda Spectrophotometry/ UV

Screening Spectrophotometry/ UV

Turbidity Method Argentometri

16. BMD-0216-16B3 0,44±0,00 26±0,60 35,50±1,43

Metoda Spektrofotometri Spektrofotometri Argentometri

17. BMD-0217-16B3 15,31±0,11 20,91±0,34 24,15±0,01

Metoda APHA 2005 : 4500

NO3-NB SNI 6989.20-2009 SNI 6989.19-2009

18. BMD-0218-16B3 15,12 20,36 27,30

Metoda Ion Chromatography Ion Chromatography Ion Chromatography

19. BMD-0219-16B3 0,45 23,13 23,14

Metoda Spektrofotometri SNI 06-6989-20-2004 SNI 06-6989-19-2009

20. BMD-0220-16B3 14,90±0,84 21,78±1,31 28,90±1,0

Metoda SNI 3554-2015 SNI 3554-2015 SNI 3554-2015

Page 10: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 6 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

IV.1 Parameter Nitrat (NO3-)

Sebanyak 20 (dua puluh) laporan hasil pengujian untuk parameter nitrat (NO3-) diterima

oleh BMD Laboratory dan 16 (enam belas) diantaranya disertai dengan laporan nilai

estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil

pengolahan data secara statistika diperoleh sebanyak 16 (enam belas) data yang dilaporkan

laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0), dan 4 (empat) data yang

dilaporkan laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-score|≥3,0). Hasil

pengolahan data disajikan pada Tabel 4 dan Gambar 1.

Tabel 4. Hasil Pengolahan Data Parameter Nitrat (NO3-)

No. Kode Peserta Nilai Laporan

Peserta (mg/L)

U, k=2

(mg/L)

Nilai Evaluasi Uji

Banding (mg/L)

SDPA

(mg/L) z-score Kategori

1. BMD-0201-16B3 14.24 0.261 14.67 0.48 -0.90 OK

2. BMD-0202-16B3 14.40 0.15 14.67 0.48 -0.56 OK

3. BMD-0203-16B3 14.22 0.263 14.67 0.48 -0.94 OK

4. BMD-0204-16B3 14.82 0.02 14.67 0.48 0.31 OK

5. BMD-0205-16B3 14.36 0.62 14.67 0.48 -0.65 OK

6. BMD-0206-16B3 0.44 0.00 14.67 0.48 -29.52 $$

7. BMD-0207-16B3 15.27 0.11 14.67 0.48 1.24 OK

8. BMD-0208-16B3 15.12 - 14.67 0.48 0.93 OK

9. BMD-0209-16B3 0.36 - 14.67 0.48 -29.69 $$

10. BMD-0210-16B3 14.82 0.84 14.67 0.48 0.31 OK

11. BMD-0211-16B3 14.19 0.261 14.67 0.48 -1.00 OK

12. BMD-0212-16B3 14.41 0.16 14.67 0.48 -0.54 OK

13. BMD-0213-16B3 14.75 0.263 14.67 0.48 0.16 OK

14. BMD-0214-16B3 14.88 0.02 14.67 0.48 0.43 OK

15. BMD-0215-16B3 13.82 0.60 14.67 0.48 -1.77 OK

16. BMD-0216-16B3 0.44 0.00 14.67 0.48 -29.52 $$

17. BMD-0217-16B3 15.31 0.11 14.67 0.48 1.32 OK

18. BMD-0218-16B3 15.12 - 14.67 0.48 0.93 OK

19. BMD-0219-16B3 0.45 - 14.67 0.48 -29.50 $$

20. BMD-2020-16B3 14.90 0.84 14.67 0.48 0.47 OK

Page 11: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 7 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

12.00

12.40

12.80

13.20

13.60

14.00

14.40

14.80

15.20

15.60

16.00

16.40

16.80

17.20

BM

D-0

209-

16B

3

BM

D-0

206-

16B

3

BM

D-0

216-

16B

3

BM

D-0

219-

16B

3

BM

D-0

215-

16B

3

BM

D-0

211-

16B

3

BM

D-0

203-

16B

3

BM

D-0

201-

16B

3

BM

D-0

205-

16B

3

BM

D-0

202-

16B

3

BM

D-0

212-

16B

3

BM

D-0

213-

16B

3

BM

D-0

204-

16B

3

BM

D-0

210-

16B

3

BM

D-0

214-

16B

3

BM

D-2

020-

16B

3

BM

D-0

208-

16B

3

BM

D-0

218-

16B

3

BM

D-0

207-

16B

3

BM

D-0

217-

16B

3

NO

3-(m

g/L

)

Lab. Code

Cons. value (X)

X+2SDPA

X+3SDPA

X-2SDPA

X-3SDPA

Cons. value = 14.67 mg/LAlgoritma A Mean ISO 13528:2005(E)SDPA = 0.48 mg/L (SD Robust)SD Horwitz = 1.57 mg/L

©BMD Laboratory

Keterangan:

Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium

peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust

Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis

menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C). SDPA merupakan SD

Robust, yaitu simpangan baku yang diperoleh dari pengolahan data secara statistika

menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E). U yaitu

ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.

Gambar 1. Grafik Hasil Uji Banding Parameter Nitrat (NO3-)

Page 12: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 8 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

IV.2 Parameter Sulfat (SO42-

)

Sebanyak 20 (dua puluh) laporan hasil pengujian untuk parameter sulfat (SO42-) diterima

oleh BMD Laboratory dan 16 (enam belas) diantaranya disertai dengan laporan nilai

estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil

pengolahan data secara statistika diperoleh sebanyak 18 (delapan belas) data yang

dilaporkan laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0), 1 (satu) data

yang dilaporkan laboratorium masuk dalam kategori ‘meragukan’ (2,0<|z-score|<3,0), dan 1

(satu) data yang dilaporkan laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-

score|≥3,0). Hasil pengolahan data disajikan pada Tabel 5 dan Gambar 2.

Tabel 5. Hasil Pengolahan Data Parameter Sulfat (SO42-)

No. Kode Peserta Nilai Laporan

Peserta (mg/L)

U, k=2

(mg/L)

Nilai Evaluasi Uji

Banding (mg/L)

SDPA

(mg/L) z-score Kategori

1. BMD-0201-16B3 19.50 1.586 20.58 1.78 -0.61 OK

2. BMD-0202-16B3 19.63 1.25 20.58 1.78 -0.54 OK

3. BMD-0203-16B3 19.74 1.603 20.58 1.78 -0.47 OK

4. BMD-0204-16B3 18.99 0.20 20.58 1.78 -0.89 OK

5. BMD-0205-16B3 15.97 1.22 20.58 1.78 -2.59 $

6. BMD-0206-16B3 24 0.60 20.58 1.78 1.92 OK

7. BMD-0207-16B3 20.68 0.34 20.58 1.78 0.05 OK

8. BMD-0208-16B3 20.34 - 20.58 1.78 -0.14 OK

9. BMD-0209-16B3 23.29 - 20.58 1.78 1.52 OK

10. BMD-0210-16B3 22.16 1.31 20.58 1.78 0.89 OK

11. BMD-0211-16B3 20.39 1.656 20.58 1.78 -0.11 OK

12. BMD-0212-16B3 19.71 1.26 20.58 1.78 -0.49 OK

13. BMD-0213-16B3 20.82 1.693 20.58 1.78 0.13 OK

14. BMD-0214-16B3 18.50 0.20 20.58 1.78 -1.17 OK

15. BMD-0215-16B3 18.72 1.42 20.58 1.78 -1.05 OK

16. BMD-0216-16B3 26 0.60 20.58 1.78 3.04 $$

17. BMD-0217-16B3 20.91 0.34 20.58 1.78 0.18 OK

18. BMD-0218-16B3 20.36 - 20.58 1.78 -0.12 OK

19. BMD-0219-16B3 23.13 - 20.58 1.78 1.43 OK

20. BMD-2020-16B3 21.78 1.31 20.58 1.78 0.67 OK

Page 13: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 9 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

12.00

12.90

13.80

14.70

15.60

16.50

17.40

18.30

19.20

20.10

21.00

21.90

22.80

23.70

24.60

25.50

26.40

27.30

28.20

29.10

30.00

BM

D-0

205-

16B

3

BM

D-0

214-

16B

3

BM

D-0

215-

16B

3

BM

D-0

204-

16B

3

BM

D-0

201-

16B

3

BM

D-0

202-

16B

3

BM

D-0

212-

16B

3

BM

D-0

203-

16B

3

BM

D-0

208-

16B

3

BM

D-0

218-

16B

3

BM

D-0

211-

16B

3

BM

D-0

207-

16B

3

BM

D-0

213-

16B

3

BM

D-0

217-

16B

3

BM

D-2

020-

16B

3

BM

D-0

210-

16B

3

BM

D-0

219-

16B

3

BM

D-0

209-

16B

3

BM

D-0

206-

16B

3

BM

D-0

216-

16B

3

SO

42

-(m

g/L

)

Lab. Code

Cons. value (X)

X+2SDPA

X+3SDPA

X-2SDPA

X-3SDPA

Cons. value = 20.58 mg/LAlgoritma A Mean ISO 13528:2005(E)SDPA = 1.78 mg/L (SD Robust)SD Horwitz = 2.09 mg/L

©BMD Laboratory

Keterangan:

Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium

peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust

Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis

menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C). SDPA merupakan SD

Robust, yaitu simpangan baku yang diperoleh dari pengolahan data secara statistika

menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E). U yaitu

ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.

Gambar 2. Grafik Hasil Uji Banding Parameter Sulfat (SO42-)

Page 14: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 10 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

IV.3 Parameter Klorida (Cl-)

Sebanyak 20 (dua puluh) laporan hasil pengujian untuk parameter klorida (Cl-) diterima oleh

BMD Laboratory dan 14 (empat belas) diantaranya disertai dengan laporan nilai estimasi

ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan

data secara statistika diperoleh sebanyak 18 (delapan belas) data yang dilaporkan

laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0) dan 2 (dua) data yang

dilaporkan laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-score|≥3,0). Hasil

pengolahan data disajikan pada Tabel 6 dan Gambar 3.

Tabel 6. Hasil Pengolahan Data Parameter Klorida (Cl-)

No. Kode Peserta Nilai Laporan

Peserta (mg/L)

U, k=2

(mg/L)

Nilai Evaluasi Uji

Banding (mg/L)

SDPA

(mg/L) z-score Kategori

1. BMD-0201-16B3 26.80 1.692 26.52 2.34 0.12 OK

2. BMD-0202-16B3 26.23 - 26.52 2.34 -0.13 OK

3. BMD-0203-16B3 27.44 1.726 26.52 2.34 0.39 OK

4. BMD-0204-16B3 25.30 0.42 26.52 2.34 -0.52 OK

5. BMD-0205-16B3 29.64 0.65 26.52 2.34 1.33 OK

6. BMD-0206-16B3 39.05 1.43 26.52 2.34 5.36 $$

7. BMD-0207-16B3 23.76 0.03 26.52 2.34 -1.18 OK

8. BMD-0208-16B3 27.22 - 26.52 2.34 0.30 OK

9. BMD-0209-16B3 23.14 - 26.52 2.34 -1.45 OK

10. BMD-0210-16B3 28.86 1.0 26.52 2.34 1.00 OK

11. BMD-0211-16B3 27.04 1.703 26.52 2.34 0.22 OK

12. BMD-0212-16B3 26.32 - 26.52 2.34 -0.09 OK

13. BMD-0213-16B3 26.95 1.698 26.52 2.34 0.18 OK

14. BMD-0214-16B3 25.19 0.42 26.52 2.34 -0.57 OK

15. BMD-0215-16B3 29.83 0.66 26.52 2.34 1.41 OK

16. BMD-0216-16B3 35.50 1.43 26.52 2.34 3.84 $$

17. BMD-0217-16B3 24.15 0.01 26.52 2.34 -1.01 OK

18. BMD-0218-16B3 27.30 - 26.52 2.34 0.33 OK

19. BMD-0219-16B3 23.14 - 26.52 2.34 -1.45 OK

20. BMD-2020-16B3 28.90 1.0 26.52 2.34 1.02 OK

Page 15: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 11 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

16.00 17.20 18.40 19.60 20.80 22.00 23.20 24.40 25.60 26.80 28.00 29.20 30.40 31.60 32.80 34.00 35.20 36.40 37.60 38.80 40.00 41.20

BM

D-0

219-

16B

3

BM

D-0

209-

16B

3

BM

D-0

207-

16B

3

BM

D-0

217-

16B

3

BM

D-0

214-

16B

3

BM

D-0

204-

16B

3

BM

D-0

202-

16B

3

BM

D-0

212-

16B

3

BM

D-0

201-

16B

3

BM

D-0

213-

16B

3

BM

D-0

211-

16B

3

BM

D-0

208-

16B

3

BM

D-0

218-

16B

3

BM

D-0

203-

16B

3

BM

D-0

210-

16B

3

BM

D-2

020-

16B

3

BM

D-0

205-

16B

3

BM

D-0

215-

16B

3

BM

D-0

216-

16B

3

BM

D-0

206-

16B

3

Cl- (

mg

/L)

Lab. Code

Cons. value (X)

X+2SDPA

X+3SDPA

X-2SDPA

X-3SDPA

Cons. value = 26.52 mg/LAlgoritma A Mean ISO 13528:2005(E)SDPA = 2.34 mg/L (SD Robust)SD Horwitz = 2.59 mg/L

©BMD Laboratory

Keterangan:

Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium

peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust

Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis

menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C). SDPA merupakan SD

Robust, yaitu simpangan baku yang diperoleh dari pengolahan data secara statistika

menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E). U yaitu

ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.

Gambar 3. Grafik Hasil Uji Banding Parameter Klorida (Cl-)

Page 16: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 12 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

V. PENUTUP

Program uji banding antar laboratorium merupakan salah satu sarana jaminan mutu

eksternal bagi laboratorium pengujian dan kalibrasi. Hasil dari program uji banding ini

diharapkan dapat digunakan sebagai sarana evaluasi bagi laboratorium untuk meningkatkan

kinerja dalam kegiatan pengujian yang dilakukan.

VI. REFERENSI

[1] SNI 01-3553-2006

[2] SNI ISO/IEC 17025:2008

[3] ISO 13528:2005(E)

[4] Distribution AnalyzerTM, Copyright © by Taylor Enterprises, Inc. USA

[5] M. Thompson (2004) The Amazing Horwitz Function, AMC Technical Brief No.17, Royal

Society of Chemistry

Page 17: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 13 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

Annex A

Rekapitulasi Hasil z-score Program Uji Banding

No

. Kode Peserta

z-score

Nitrat (NO3-) Sulfat (SO4

2-) Klorida (Cl

-)

1. BMD-0201-16B3 -0.90 OK -0.61 OK 0.12 OK

2. BMD-0202-16B3 -0.56 OK -0.54 OK -0.13 OK

3. BMD-0203-16B3 -0.94 OK -0.47 OK 0.39 OK

4. BMD-0204-16B3 0.31 OK -0.89 OK -0.52 OK

5. BMD-0205-16B3 -0.65 OK -2.59 $ 1.33 OK

6. BMD-0206-16B3 -29.52 $$ 1.92 OK 5.36 $$

7. BMD-0207-16B3 1.24 OK 0.05 OK -1.18 OK

8. BMD-0208-16B3 0.93 OK -0.14 OK 0.30 OK

9. BMD-0209-16B3 -29.69 $$ 1.52 OK -1.45 OK

10. BMD-0210-16B3 0.31 OK 0.89 OK 1.00 OK

11. BMD-0211-16B3 -1.00 OK -0.11 OK 0.22 OK

12. BMD-0212-16B3 -0.54 OK -0.49 OK -0.09 OK

13. BMD-0213-16B3 0.16 OK 0.13 OK 0.18 OK

14. BMD-0214-16B3 0.43 OK -1.17 OK -0.57 OK

15. BMD-0215-16B3 -1.77 OK -1.05 OK 1.41 OK

16. BMD-0216-16B3 -29.52 $$ 3.04 $$ 3.84 $$

17. BMD-0217-16B3 1.32 OK 0.18 OK -1.01 OK

18. BMD-0218-16B3 0.93 OK -0.12 OK 0.33 OK

19. BMD-0219-16B3 -29.50 $$ 1.43 OK -1.45 OK

20. BMD-2020-16B3 0.47 OK 0.67 OK 1.02 OK

Keterangan:

OK : Kategori ‘memuaskan’

$ : Kategori ‘meragukan’

$$ : Kategori ‘kurang memuaskan’

Page 18: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 14 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

-14.50

-12.50

-10.50

-8.50

-6.50

-4.50

-2.50

-0.50

1.50

b (

mg

/L)

Lab. Code

Parameter Nitrat (NO3-)

-10.00

-8.00

-6.00

-4.00

-2.00

0.00

2.00

4.00

6.00

8.00

10.00

b (

mg

/L)

Lab. Code

Parameter Sulfat (SO42-)

Annex B

Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus

Page 19: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 15 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

-13.00

-11.00

-9.00

-7.00

-5.00

-3.00

-1.00

1.00

3.00

5.00

7.00

9.00

11.00

13.00

b (

mg

/L)

Lab. Code

Parameter Klorida (Cl-)

Page 20: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 16 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

Annex C

Evaluasi Distribusi Normal Data Program Uji Banding

Normal Distribution Analysis

(Parameter NO3-)

Powered by:

Distribution AnalyzerTM

(licensed software)

©Taylor Enterprises, Inc.

Sample Size = 20Average = 11.816Standard Deviation = 5.858Skewness = -1.61Excess Kurtosis = 0.67

Test of Fit: p-value = 0.0222 (SK All) Decision = Fail (SK Spec) Decision = No specs

0.3600 15.31007.8350

No Kode PesertaNilai Laporan

Peserta (mg/L)

1 BMD-0201-16B3 14.242 BMD-0202-16B3 14.403 BMD-0203-16B3 14.224 BMD-0204-16B3 14.825 BMD-0205-16B3 14.366 BMD-0206-16B3 0.447 BMD-0207-16B3 15.27

8 BMD-0208-16B3 15.129 BMD-0209-16B3 0.36

10 BMD-0210-16B3 14.8211 BMD-0211-16B3 14.1912 BMD-0212-16B3 14.4113 BMD-0213-16B3 14.7514 BMD-0214-16B3 14.8815 BMD-0215-16B3 13.8216 BMD-0216-16B3 0.4417 BMD-0217-16B3 15.3118 BMD-0218-16B3 15.1219 BMD-0219-16B3 0.4520 BMD-2020-16B3 14.90

Normal Distribution Analysis

(Parameter NO3-)

Powered by:

Distribution AnalyzerTM

(licensed software)

©Taylor Enterprises, Inc.

No Kode PesertaNilai Laporan

Peserta (mg/L)

1 BMD-0201-16B3 14.242 BMD-0202-16B3 14.403 BMD-0203-16B3 14.224 BMD-0204-16B3 14.825 BMD-0205-16B3 14.366 BMD-0207-16B3 15.27

7 BMD-0208-16B3 15.128 BMD-0210-16B3 14.829 BMD-0211-16B3 14.19

10 BMD-0212-16B3 14.4111 BMD-0213-16B3 14.7512 BMD-0214-16B3 14.8813 BMD-0215-16B3 13.8214 BMD-0217-16B3 15.3115 BMD-0218-16B3 15.1216 BMD-2020-16B3 14.90

Sample Size = 16Average = 14.664Standard Deviation = 0.440Skewness = -0.20Excess Kurtosis = -0.91

Test of Fit: p-value = 0.5767 (SK All) Decision = Pass (SK Spec) Decision = No specs

With 95% confidence more than 99% of the values are between 12.9884 and 16.3404

13.8200 15.310014.5650

BMD-0206-16B3,

BMD-0209-16B3,

BMD-0216-16B3,

BMD-0219-16B3 were rejected

Page 21: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 17 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

Normal Distribution Analysis

(Parameter SO42-

)

Powered by:

Distribution AnalyzerTM

(licensed software)

©Taylor Enterprises, Inc.

No Kode PesertaNilai Laporan

Peserta (mg/L)

1 BMD-0201-16B3 19.502 BMD-0202-16B3 19.633 BMD-0203-16B3 19.744 BMD-0204-16B3 18.995 BMD-0205-16B3 15.976 BMD-0206-16B3 247 BMD-0207-16B3 20.68

8 BMD-0208-16B3 20.349 BMD-0209-16B3 23.29

10 BMD-0210-16B3 22.1611 BMD-0211-16B3 20.3912 BMD-0212-16B3 19.7113 BMD-0213-16B3 20.8214 BMD-0214-16B3 18.5015 BMD-0215-16B3 18.7216 BMD-0216-16B3 2617 BMD-0217-16B3 20.9118 BMD-0218-16B3 20.3619 BMD-0219-16B3 23.1320 BMD-2020-16B3 21.78

Sample Size = 20Average = 20.731Standard Deviation = 2.222Skewness = 0.42Excess Kurtosis = 1.01

Test of Fit: p-value = 0.3295 (SK All) Decision = Pass (SK Spec) Decision = No specs

With 95% confidence more than 99% of the values are between 12.6991 and 28.7629

15.9700 26.000020.9850

Normal Distribution Analysis

(Parameter Cl-)

Powered by:

Distribution AnalyzerTM

(licensed software)

©Taylor Enterprises, Inc.

No Kode PesertaNilai Laporan

Peserta (mg/L)

1 BMD-0201-16B3 26.802 BMD-0202-16B3 26.233 BMD-0203-16B3 27.444 BMD-0204-16B3 25.305 BMD-0205-16B3 29.646 BMD-0206-16B3 39.057 BMD-0207-16B3 23.76

8 BMD-0208-16B3 27.229 BMD-0209-16B3 23.14

10 BMD-0210-16B3 28.8611 BMD-0211-16B3 27.0412 BMD-0212-16B3 26.3213 BMD-0213-16B3 26.9514 BMD-0214-16B3 25.1915 BMD-0215-16B3 29.8316 BMD-0216-16B3 35.5017 BMD-0217-16B3 24.1518 BMD-0218-16B3 27.3019 BMD-0219-16B3 23.1420 BMD-2020-16B3 28.90

Sample Size = 20Average = 27.588Standard Deviation = 3.897Skewness = 1.70Excess Kurtosis = 3.48

Test of Fit: p-value = 0.0041 (SK All) Decision = Fail (SK Spec) Decision = No specs

23.1400 39.050031.0950

Page 22: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

Page 18 of 18

FR.06 Laporan Program Uji Banding (Final Report)

Ed./Rev./Date: 01/03/02-01-2016

Ref. No. 002/UR/BMD/2016

BMD Laboratory–Provider of Proficiency Testing Program

Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316

Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]

www.ujiprofisiensi.com

Normal Distribution Analysis

(Parameter Cl-)

Powered by:

Distribution AnalyzerTM

(licensed software)

©Taylor Enterprises, Inc.

BMD-0206-16B3,

BMD-0216-16B3 were rejected

No Kode PesertaNilai Laporan

Peserta (mg/L)

1 BMD-0201-16B3 26.802 BMD-0202-16B3 26.233 BMD-0203-16B3 27.444 BMD-0204-16B3 25.305 BMD-0205-16B3 29.646 BMD-0207-16B3 23.76

7 BMD-0208-16B3 27.228 BMD-0209-16B3 23.149 BMD-0210-16B3 28.86

10 BMD-0211-16B3 27.0411 BMD-0212-16B3 26.3212 BMD-0213-16B3 26.9513 BMD-0214-16B3 25.1914 BMD-0215-16B3 29.8315 BMD-0217-16B3 24.1516 BMD-0218-16B3 27.3017 BMD-0219-16B3 23.1418 BMD-2020-16B3 28.90

Sample Size = 18Average = 26.512Standard Deviation = 2.083Skewness = -0.14Excess Kurtosis = -0.79

Test of Fit: p-value = 0.6823 (SK All) Decision = Pass (SK Spec) Decision = No specs

With 95% confidence more than 99% of the values are between 18.7995 and 34.2238

23.1400 29.830026.4850

Page 23: LAPORAN PROGRAM UJI BANDING (PROFICIENCY … … · Nitrat (NO 3 -), Sulfat (SO 4 2- ... pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ... Hasil pengolahan data pada

©BMD Laboratory 2016