Kolida Kts 580 Service Manual En

download Kolida Kts 580 Service Manual En

of 42

description

manual Kolida Kts 580 Service Manual English

Transcript of Kolida Kts 580 Service Manual En

  • NTS - 960R SERIES TOTAL STATION SERVICE MANUAL SOUTH SURVEYING & MAPPING INSTRUMENT CO., LTD.

    2010

    SOUTH201021

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 1

    1 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Tools

    Oscillograph Multimeter Collimator

    Screwdriver Pliers AllenKey

    Tweezers Polishing Sticks Cotton

    Lubricant Alcohol withether GreenGrease

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 2

    2 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Disassemblyandassembly

    Preparation:

    Pleasepreparetoolsasfollows:screwdriver,tweezer,pliers,allenKeyandalcoholready.Thenwashyourhands.

    Useatraytocollectthescrews.

    Useanotebooktowritedownthepositionoftheconnectionpointsandthecolorofwireswhennecessary.

    Contents:

    ThischaptermainlyintroducedthestandarddisassemblyprocedureofNTS960R.(Absoluteencoding),Pleasefollowthesestepswhentakingaparttheinstrument,andalsotheassemblyprocedure.Refertotherelatedchaptersforrepairingorreplacement.Applyafinaltopcoatofvarnishonthescrewsafterassembly,exceptverticaldiskcover,EDMcover,slipringcoveretc.Thefactorymightupgradesomeofthespareparts,ifyoufindsomedifferencesbetweenthismanualandyourinstrument,pleasecontactus.

    Step

    1. Removethecoverofverticaldiscsystem

    Loosenoffthesixscrewsontheleftcoverwithascrewdriver,andthenopenthecovercarefully.

    2. LCDPanel

    Loosenoffthefourscrewswithascrewdriver.AndthenremovetheLCDcovercarefully.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 3

    3 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    BeforeremovingtheLCDcover,pleasebecarefultheflatcableontheback.Firstopentheflatcableclampaccordingtothedirectionofarrow,andthenpullouttheflatcable,bothoftheLCDandtheflatcablearefragile,pleasebecareful.

    Loosenoffthefourscrewswithascrewdriver.AndthenremovetheLCDcarefully.

    Opentheflatcableclampaccordingtothedirectionofarrow,andthenpullouttheflatcable.

    Loosenoffthefourscrewswithascrewdriver.AndthenremovetheLCDbackcovercarefully.

    Opentheflatcableclampaccordingtothedirectionofarrow,andthenpullouttheflatcable.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 4

    4 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    3.Anglemeasuringboard

    Refertostep1,removethecover.Loosenoffthefourscrewsontheanglemeasuringboardwithascrewdriver.

    Pullouttheplugscarefully,andrecordthepositionofeachplugforassembly.Pleasekeeptheanglemeasuringboardappropriatelyandsafely.

    4.Right Cover Refertostep2,removetheLCD,Cutthenyloncabletiewithpliers.Donotcutthewires.

    Loosenoffthesixscrewsontherightcoverandthenopenit.

    Whenopeningthecover,pleasebecarefulthewire.Pullouttheplug,thenyoucanremovethecoverfreely.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 5

    5 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    5.EDMCover

    LoosenoffthefourscrewsontheEDMcoverandthenremovetheEDMcovercarefully.Makesurethatthetelescopetangentscrewislockedupincaseofscuffingtheinstrumentwhenyouloosenthosescrews.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 6

    6 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    6.Distancemeasuringboard

    Settheinstrumenttelescopeinreversedposition,thenremovethecoverontopoftheEDM.

    Loosenoffthefourscrewsonthedistancemeasuringboardwithascrewdriver.

    Pullouttheplugscarefully,recordthepositionofeachplugforreference.Pulloutthesignalplugcarefullyaccordingtothedirectionofredarrow.

    Finshedthosestepsabove,reversetheboard,youwillseetherecevingfiberisstillconnectedwiththedistancemeasuringboard.Becarefulwiththefragilefiber.

    Loosenalittlethesmallscrewonthetopofthereceivingfiberbasewithallenkey,thenpulloutthefibercarefully,andremovethemainboard.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 7

    7 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    7.Fibers

    WeuseforficatefiberontheNTS350series.Asshowninthefollowingfigure,thereisanmixedreceivingendandwhichdividesintotwofibers;oneisforinsidepathreceiving,theotherforoutsidepathreceiving.

    Refertostep5,6.RemovethetwocoversofEDMandthedistancemeasuringboard.Ononesidewithmotorsandreceivingfiber,youcanfindasmallscrew;loosenitalittletillyoucanpullouttheinsidepathfiber.Asshowninfigure.

    TurntheEDMover;loosenoffthefourscrewsonthemetalcover,andpullthefiberoutcarefullyfromtheholeinthemiddleofthemetalboard.

    mixedreceivingend outsidepathreceivingfiberinsidepathreceivingfiber

    Outsidepathfiberbase

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 8

    8 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Asshowninfigure,loosenalittlethesmallscrewandthenpullouttheoutsidepathfiber,tillnowweremovethewholefiber.Sincethefiberisveryeasilybroken,pleasekeepitappropriatelyandsafely.

    8.Verticalabsoluteencodingdiscsystems

    Refertostep1and3,removetheverticaldisccoverandtheanglemeasuringboard.Loosenoffthe4screwsinAareaandtheother4inBarea.Nowyoucantakeofftheverticaldiscdustcover.

    Loosenoffthe4screwsinCareaandtheother3inDarea.

    Intheredcirclearea,thereare4blackflatheadCCDholdingscrews.Donotloosenthemwhenremovetheverticaldisk.

    A B

    c

    D

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 9

    9 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    FllowthestepstoassembletheVerticalabsoluteencodingdiscsystem:

    AssemblethetwoverticalangleCCDsensorfirst.

    Attentiontothislittletrianglemarkontheedgeoftheverticaldisc,thisisthe0anglemark.

    Asshowninfigure,setthetelescopeinnormal,assembletheverticaldisksystemintotheappropriatepositioncarefully.AligntheholesforscrewsbetweentheEDMandtheverticaldisk,andatthesametimemakesurethesmalltrianglemarkisnearbythelowerCCD.Thisisveryimportant,ifthetrianglemarkisnotintherightpositionafterassemblethedisksystem,theverticalanglevaluereadingwillbewrong.

    Tightenallthescrewswhenfinishedthosestepsabove.Anddonotforgetthewasherduringassembling.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 10

    10 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    9.VerticalCCDunit

    Refertostep1,3,8,removetheverticaldisk

    coverDistancemeasuringboardandverticaldisksystem.Loosenoffthe4screwstodisassembletheCCDunit.Inthesamewaytoremoveanotheroneintheoppositeposition.

    10.CCDBracket

    Refertostep1,3,8,and9,removeverticaldiskcover,anglemeasuringboard,verticaldisksystem,andverticalCCDunit.Thenturnovertheverticaldisk,thereisacoppernut,loosenitoffwithpliers.

    Takeoutverticaldiscfrombracket.Pleasekeepthemappropriatelyandsafely.

    11.Verticaltangentscrew

    Refertostep4,removetheslipringcover.Loosenalittlethetwosmallscrewswithallenkey.Presstheendofverticalbrakeringsystemaccordingtothedirectionofredarrow,atthesametimepullouttheverticaltangentscrewcarefully.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 11

    11 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    12.Slipring

    Refertostep1,removetheverticaldiskcoveranddisconnecttheslipringplugformtheanglemeasuringboard.Refertostep2,removetheLeftsideLCDandthebackcover.Refertostep4,cutthecabletieandpullouttheslipringwire.Thenloosenoffthetwoscrewsasshowninfigure.

    Refertostep6.Pulloutthe4pinslipringplugfromthedistancemeasuringboard.Recordthecolorandthepositionofeachwirethenpullthemoutcarefullywithtweezers,pleasekeeptheplugappropriatelyandsafely.

    Loosenoffthe3screwsasshowninfigurethenpullouttheslipringunit.

    3.

    4.

    5.

    6.

    13.Verticalbrakering

    Refertostep4,11,and12,removetheslipringcover,telescopetangentscrew,andtheslipring.Loosenoffthe3screwsasshowninfigure,andremovetheverticalbrakering.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 12

    12 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    14.Axlesleeve

    Refertostep4,10,11and12,removetheslipringcover,verticaltelescopetangentscrew,slipring,andtheverticalarrestinggearring.Loosenoffthe4screwsasshowninfigure,holdtheEDM,andpullouttheaxlesleevewithyourhands.Ifitblocks,donotforceremovetheaxlesleeve,pleasesendtheinstrumentbacktothefactory.

    15.EDM

    Refertostep114,turntheEDMtoreverseposition,holdtheEDMwithbothofyourhands,andtakeitoutaccordingtothedirectionofredarrow.Becareful,DonotdroptheEDM.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 13

    13 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    16.Compensator

    Refertostep1and4,removetheverticaldiskcover,compensatorplugontheanglemeasuringboard,leftsideLCDandbackcover,thencutthecabletie,pulloutthecompensatorwire,opentheslipringcover.

    Loosenoffthe2screwsasshowninfigure,thenyoucantakeoutthecompensatorfreely.

    17.Horizontaltangentscrew

    Refertostep2,removetherightsideLCDandbackcover.

    Loosenthetwolittlescrewswithallenkey;presstheendofhorizontalbrakeringsystemaccordingtothedirectionofredarrow,atthesametimepulloutthetangentscrewcarefully.

    18.Horizontaldiscandverticalaxleunit

    Refertostep1,2,3,4,5,and17,removetheverticaldiskcover,slipringcover,bothoftheLCDsandbackcovers,cutoffthecabletie,andpullouttheCCDplugs,removeoneEDMcoverandthehorizontaltangentscrew.

    Takeoffthe4rubbercover.

    Loosenoffthe3screwswithallenkey.

    disassemblethetribrach.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 14

    14 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Turnovertheinstrumentloosenoffthescrewsonthebackcover.

    Useallenkeyloosenoffthe3baseholdingscrews,thentakeoffthebase,thenloosenoffthe6dustcoverholdingscrewsandtakeoffthedustcover.

    Holdtheinstrumentbodywithbothofyourhands,andthenmoveitupcarefully.Payattentionaboutthehorizontalbrakeringsystem,andtheCCDwire.

    19.HorizontalCCDunit

    Refertostep18,disassemblethehorizontaldiscandverticalaxleunit,loosenoffthescrewsontheCCDunit,thenyoucanremovetheCCDunit.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 15

    15 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    20.Verticalaxleunit

    Refertostep18and19,,disassemblethehorizontaldiscandverticalaxleunit,thendisassembletheCCDunit,disassemblethetribrach.

    Turnoverthehorizontaldiscandverticalaxleunit,loosenoffthescrewsonthebackcover.Becarefulwiththehorizontalangledisc.

    loosenoffthecoppernutwithallenkey.

    Movethebaseupaccordingtothedirectionofredarrow.Youwillfindtheverticalaxleandthebaseareseparated.

    Tips:keepthebottomofthebaseonthetopmakesurethesteelballswillnotdrop,iflostanysteelball,thewholesteelballsmustbechangedbecausetheyareuniquesetforeveryinstrument.Afteryouremovethebasefromtheaxle,youshouldpickupallthesteelballsintoasafetycontainerimmediately.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 16

    16 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    21.Platevial

    Loosenofftheplatevialadjusting.

    Refertostep3,removetheleftLCDalongwithitsbackcover,loosenoffthescrewsasshowninfigure,thenremovetheplatevial.

    22.Opticalplummet

    Refertostep1,removetheLeftcover.Usescrewdriverloosenoffthe2screwsontheOpticalplummet.

    Pullouttheopticalplummet.

    Pullouttheopticalplummetcarefully,andkeepitappropriatelyandsafely.

    23.Eyepieceandcrosshair

    Screwofftheeyepiececover,screwofftheeyepiece,loosenoffthefourscrewsinside,youcanremovethecrosshairfreely.

    Summary:ThisisthedisassembleprocessofNTS350series,someofthesparepartsdisassembleprocessarenotmentioned.forexample:luminotron,motor,theinsidepathprism,telescopictubeetc.mostofthesepartsneedspecialtoolforadjustingwhenassemble.Sowedonotrecommendyoutodissemblethembyyourself.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 17

    17 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Commonproblemforanglemeasuring

    Errorcode17Adjustingthegapbetweencoppernutandaxlesleeve

    mechanicalfailure

    TightenCenterscrew

    Error1

    HorizontalCCD(adjustingorreplace)

    Electronicfailure

    Replaceanglemeasuringboard

    Error2telescoperotatedtoofast(protectingerror)

    Error3Mainbodyrotatedtoofast.

    Error4,Error7VerticalCCDunit

    A:badsignal(adjustingthesignal)

    B:dustondisc(cleanthedisc)

    C:CCDunitbroken,ordustonthesurfaceofCCD(replaceorclean)

    D:Sunradiationonthedisc.

    E:Mainboardbroken.

    ERROR5,Error6HorizontalCCDunit

    A:badsignal(adjustingthesignal)

    B:dustondisc(cleanthedisc)

    CCCDunitbroken,ordustonthesurfaceofCCD(replaceorclean)

    D:Sunradiationonthedisc.

    E:Mainboardbroken.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 18

    18 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Batteryvoltage

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 19

    19 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    3Daub polishing powder (Green grease) on the vertical axle.

    4Set the polishing tool on the vertical axle then rotate it 2-3 rounds clockwise.

    5Takeouttheaxle,andcleanthepolishingpowder.

    Polishingaxlesleeve.

    6Daub polishing powder on the polishing stick then put it into the axle sleeve and rotate it 2-3 rounds clockwise.

    .

    7Take out the polishing stick and clean the axle sleeve.

    8Assemble the axle and axle sleeve (base unit). If the axle can rotate smoothly, that means the polishing process succeed.

    9Iftheverticalaxlestillblocked,pleaserepeatthesteps28.

    Axleandaxlesleevepositionchanged(fixposition)

    Horizontalaxisblocked

    Axle(Rusty,Moldy,Tilt)(polishing)

    Polishingprocess:

    Whenthehorizontalaxleblocked,pleasedonottrytorotatethetelescope,itwilldamagethehorizontalaxleandaxlesleeve.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 20

    20 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    1Refertostep10indissemblechapter.Disassemblethehorizontalaxleandaxlesleeve.

    2Cleanthelubricantonthehorizontalaxle.

    3Daub polishing powder (Green grease) on the horizontal axle.

    4Set the polishing tool on the horizontal axle then rotate it 2-3 rounds clockwise.

    5Takeouttheaxle,andcleanthepolishingpowder.

    Polishingaxlesleeve.

    6Daub polishing powder on the polishing stick then put it into the axle sleeve and rotate it 2-3 rounds clockwise.

    7Take out the polishing stick and clean the axle sleeve.

    8Assemble the axle and axle sleeve. If the axle can rotate smoothly, that means the polishing process succeed.

    9Ifthehorizontalaxlestillblocked,pleaserepeatthesteps28.

    DisplayunitOnedisplayunitfailurewillcausebothofthemcantdisplay.Pleasechangeitonebyonetofindoutwhichoneisbroken.

    CommunicationportfaultmaintenanceMakesurethecommunicationcableisgoodandconnectionbetweencomputerandTSissteady.The

    communicationportiswellconnectedwiththemainboard.

    Andcheckthebaud rate. If the fault still can not be ruled out, please change the main board.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 21

    21 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    The name for angle-measuring board

    Upper CCD unit socket

    Bottom CCD unit socket

    Slip-ring socket

    Compensator socket

    Horizontal CCD unit socket for short wire

    Horizontal CCD unit socket for short wire

    LCD socket

    COM socket

    Power supply socket

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 22

    22 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    EDMfaultmaintenance

    EDMtestprocess1Checkoutthemotorsareworkingornotafterpowerontheinstrument.(itsoundslikeclick).Ifnot:

    1CheckthepowersupplytoEDMboard.Itshouldbe6V.ifthereisnopowersupply

    1checktheslipringisshortcircuitornot,orbroken,replacetheslipring.

    2checkthepowersupplyfromanglemainboardtoslipring.Ifthereisnopowersupply,replacetheanglemainboard.

    3checkthebatterysupplythepowervoltagetoanglemainboardornot.

    2Checkthemotorsworkingvoltage,itshouldbe2.5V,ifnot,replacetheEDMmainboard,checkthemotorsvoltageintheteststate(factorymode).

    1Modemotorandinside/outsidepathmotorvoltagewillbealways2.5V.

    2TestlightReducemotorwhenitisworking.

    3Checkwhetherthemotorisbrokenornot.Ifbroken,replaceanewone.

    2Checkthelasercomesoutornot,ifnotTheemittingtubeworkingvoltageshouldbe3v.Ifthereisnopowersupplyfrommeasuringboardtoemittingtube,replacethemeasuringboard.Ifthepowersupplyisgood,thenreplacetheemittingtube.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 23

    23 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    3Checkthesignallevel:Poweron,press[systemsetup],andthen,select[systeminformation],thenpress[F1]5times[F4]5times.Enterthetestmode.

    Whentestmodeactived,thescreenwillshows3newoptions:

    1. 1.ADJCOEFFICIENT2. EDMADJUSTING3. INSTRUMENTACCURACY

    Select[EDMadjusting]option,itshowstheEDMversionandcurrentworkingvoltage.

    ThecorrectworkingvoltageiswrittenontheEDMboard.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 24

    24 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Ifthecurrentworkingvoltageismuchdifferentthancorrectworkingvoltage,thenpress[next]pressF1keyswitchtostate3select[Hiv]tochangetheworkingvoltage,afterthatpresssettosave.

    2 InNonPrismmode,shotonthewhitewall,outsidepathsignalshouldbe80150mv(Thesignalisrelativitybythereflectoranddistance.)ifthesignalisveryweak,thefibermaybebroken,replacethefiber.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 25

    25 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    3Insidepathsignalshouldbe10001500mv.(PrismorNonPrismmode)ifnot,loosentheinsidefilterholdingscrewnearthefiber.Turntheinsidepathfiltertochangethesignal;tightenthescrewwhenthesignalintheworkingrange.

    Press[tape]toswitchthetapefrom05,theirsignalshouldbe10001500mv,ifoneofthemnotundertherage,replacethemeasuringboard.

    Ifallthetapescannotadjusttotherightrange,replacethefiberormeasuringboard.

    4InSTATE1,switchthemeasuremodetoPRISM.Themeasuremodemotorwilldropdownthefilter.Thefilterwillblockthelaserandonlyallowsasmallquantitylasercomethroughthefilterfromthegapinthecenter.

    Whenthelasercomesout,thegapshouldbeinthecenterofthelaserblockedonthefilter.

    Ifnot,loosenthefixscrewstoadjustthepositionuntilthegapinthecenterofthelaser,andtightenthefixscrews,putsomevarnishonfixscrews.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 26

    26 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    4CheckandAdjust3axis.(beforeadjust3axis,makesure2Ciscorrect.RefertototalstationCalibrationthe4thsection)

    3axes:sightaxis,emittingaxis,receivingaxis.(The3axisshouldbematchtogether,ifnot,itwillcausetheinstrumentcannotmeasurelongdistance.

    1.checkandadjustEmittingaxis

    1Setthetotalstationonthetripodandlevelit.Putacrosshairboard50Meterawayfromtotalstation.Turnontheinstrument,enterthetestmode.SelectNonPrisminState1toletthelasercomesoutfromtheLenscenter.

    2Setupatheodolitebesidethetotalstation.Useforobservethelaserdotonthecrosshairboard.

    3MatchthetwocrosshairstogetherbetweenTOTALSTATIONandcrosshairboard.

    4Adjustscrew1,2whichonthebaseboardtomovethelaserpointtotheleftorright.loosenscrew3,4thenturntheredirectionmirrortomovethelaserpointupordown,untilthelaserdotinthecenterofthecrosshair.

    5loosentheholdingscrewthenTurnthefocusprismclockwisetomakelaserpointfocus.Tighttheholdingscrew.

    12 34

    FocusPrism

    Redirectionmirror

    HoldingSCREW

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 27

    27 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    2CheckandadjustReceivingaxis

    Checkthereceivingaxistoseeifitismatchedtogetherwiththesightaxis.ifnot,adjustthereceivingfiberbase.

    RefertoChapter:disassemblystep6,7,pulloutthemixreceivingendfromthemeasuringboard.

    Focusthetotalstationtoinfinity.

    Asshowninfigure,useatheodoliteaimatthetotalstationlens,turnthetheodolitezoomringutillyoucanseetotalstationcrosshairclearly.

    Aimthemixreceivingendtoalight.

    Asshowninfigure,wecanseealightdotthroughthetheodoliteeyepiece.

    Ifyoucannotseethelightdot,youneedtoadjusttheoutsidepathreceivingfiberto

    focusthelightdotuntilyoucanseeitclearly.

    LoosentheoutsidepathreceivingfiberholdingScrew,

    movethefiberupanddowntofocusthelightdot,whenfinished,

    Tightentheholdingscrew.

    Loosenthreescrewsonthefiberbaseasshowninfigure.Movethefiberbasetomakethelightdotinthecenterofthetotalstationcrosshair.

    Finally,tightenthosescrewsandputvarnishonthem.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 28

    28 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    5TestPointontheEDMBoard1thesocketandplugontheEDMboard

    EmittingFibersignalandPowersupplyplug.

    Slipringsocket

    3MotorsPowerSupplysocket

    Opticalcouplingunitsocket

    forficatefibermixreceivingendsocket

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 29

    29 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    2EDMboardtestpoint

    Workingvoltagetestpoint,sameasthestickerontheEDMboard(thisboardis173.2V)

    Emittingfiberworkingvoltage,3.3V

    BatteryPowersupply6V.

    Sameasslipring

    CPUpowersupply3.6V

    Powersupplyformeasuretape

    Workingvoltageis4.9v

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 30

    30 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    6Opticalcoupler Opticalcouplerisausedforestimatethedistancebetweentotalstationandtarget.TheemittingdiodeandreceivingdiodeontheopticalcouplerconnectedtogetherandlookslikeaU.MakesuretheblackedgeofthelightfilterinsidetheU.(theemittingdiodeandreceivingdiode.)

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 31

    31 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Commonfaultsanderror.

    (1)

    Outsidepathsignaltooweak,fiberbrokenorEDMboardbroken,replacefiberorEDMboard

    (2)

    ItdoesntmeasureunderPrismMode,butcanmeasureunderNonprismmode.fiberbrokenorEDMboardbroken,replacefiberorEDMboard

    (3)

    UnderNonprismmode,itcanmeasureinsometimes,butsometimesnot.insidepathmotoragainstwithotherparts,fixthemotorposition.orEDMboardbroken,replaceit.

    (4)

    ItdoesntmeasureunderNonPrismmode.Checkthefiberandupgradetheprogram,checktheEDMboard.

    (5)

    Measurespeedisslow.UpgradetheEDMprogram,connectthePC,PressF3andpoweron,upgradetheprogram.

    (6)

    Itdoesntmeasuredistance.Checktheinstrumentstepbystepaccordingthischapter.

    (7)

    Aimattheprismcenter,theinstrumentdoesntshootdistance,offsetthecrosshairfromtheprismcenter,theinstrumentstartshootdistance.

    Theinstrumentonlymeasureshortdistance,butdoesntmeasurelongdistance.

    CheckandAdjust3axis

    (8)

    Themeasureresultisnotstable,insidepathsignalistoostrong,adjustthesignalaccordingthischapterCheckthesignallevel

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 32

    32 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    BA

    C

    BA

    C

    Calibrations

    Preparation:

    Pleasepreparescrewdriver,adjustingpin,varnish,hexagonwrench,leveltheinstrumentonthecollimator,washyourhandsandgetready.

    Content:

    Thischaptermainlyintroducesyouthestandardadjustingprocess,thefollowingstepsshouldbefinishedonthecollimator.Excepttheadjustmentofopticalplummet,theotherstepsshouldfollowthegivenorder;otherwiseyoucantgetthewrongresult.

    Process:

    1. PlateVial

    Leveltheinstrumentonthecollimator.Asshowninthefigure,rotateinstrument,makeplatelevelparallelinthecenteroflevelingscrewAandB,andthenadjustslevelingscrewtomakeplatebubblecentered.Turntheinstrumentfor90degrees,adjustthethirdlevelingscrew,andmaketheplatebubblecentered.Repeatthesteps1and2,maketheplatebubblecenteredinabovetwopositions.

    Rotateinstrument180.Iftheplatebubblenotcentered,adjustthefixscrew,movetheplatebubbleahalfofthedifference;

    A B

    C

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 33

    33 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    2. CircularVial

    Afterexaminingtheplatebubble,circularbubbleshouldbecenteredtoo(ornotbeyondthereticlecircularity),whichindicatesthecircularbubbleisverticaltoverticalaxis.ifnottheadjustmentstepsareasfollows.

    1. Loosenthescrew(oneortwo)oppositewithbubbledeflectivedirection;

    2. Tweaktightlythescrewonthedirectionaccordantdeflectiveuntilcircularbubbleiscentered;

    3. Adjustthreeadjustmentscrewsforseveraltimesuntilcircularbubbleiscentered;

    4. Theforcepowerfixingthreeadjustmentscrewsmustbeconsistentwhencircularleveliscenteredatlast.

    3. CrossHair

    Makesurethecollimatorcrosshairisgood.Trytomatchthetotalstationcrosshaircenterandcollimatorcrosshaircentertogether.Ifthetwocrosshairmatchedperfectly,meansthetotalstationcrosshairisgood.

    Asshowninfigure.Ifthetotalstationcrosshairistilt,thenyouhavetocalibrateit.

    Takeofftheeyepiececover,Loosenthethreecrosshairholdingscrews;movethecrosshairbasetomatchthetwocrosshairtogetherthenTightenscrews,andthendaubalittlevarnishonthosescrews.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 34

    34 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    4. 2C(PerpendicularitybetweenSightAxisandHorizontalAxis)

    Leveltheinstrument;aimatthecollimator,matchthetotalstationcrosshairwithoneofthehorizontalscalesonthecollimatorcrosshair.

    0setthehorizontalangle.

    Reversethetelescopeandaimatthesamescale,andreadthehorizontalanglevalue,whichshouldbe1808.Asshowninthefigure,thevalueofHRis1800040.

    itmeansthe2Cis+40,wehavetoadjustit.

    NTS960Rseriestotalstationhastwowaytocalibrate2C

    Solution1:mechanicalcalibration

    Turnthehorizontaltangentscrewtomakethehorizontalanglevalueto18000'20"halfofthe2Cvalue.Seetheshownfigure.

    lookatthecrosshair,thecrosshairmovedalittlefromthescale(grayarea),takeofftheeyepiececoverandadjusttheleftandrightadjustingscrewstomakethecrosshairbacktomatchthescaleagain.Theprincipleisloosenoneadjustingscrewbeforetightenanother.

    Aftertheadjustment,checkthe2Cagain,ifitstillover8,pleasedoitagain.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 35

    35 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    When2C>1,itwillaffectthe3axis.First,youhavetofixthe2C,thencheckandadjustthe3axis,otherwiseitwillaffectedthedistancemeasurement.

    Solution2Electroniccalibration(onlycanusewhen2C

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 36

    36 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Press[set]toendthecalibrationprocessandthesystemwillshowtheresult.

    Presssettosave.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 37

    37 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    5.HighlowDifference(PerpendicularitybetweenHorizontalAxisandVerticalAxis)InspectionStepsandCriterion:

    Onthefirstposition(mainboardisonyourleftside),aimatthehorizontalcollimator,matchthetotalstationcrosshairwithoneofthehorizontalscalesonthecollimatorcrosshair.Asshowninfigure.

    RotatetheEDM,Aimatthelowercollimator,notedowntheintersectionbetweentotalstationcrosshairandcollimatorcrosshair.

    Onthesecondposition(mainboardisonyourrightside),dothesameprocedureasabove,notedowntheintersectionbetweentotalstationcrosshairandcollimatorcrosshair.

    Thedifferencebetweenthetwointersectionsishighlowdifference,whichshouldbelessthan10(1divisiononthecollimatoris30)

    RefertoDisassemblypart4takeofftherightcover,Loosenthefourscrewsalittle.(notice:DONOTloosentoomuch)Accordingtotheredarrow,movetheaxlesleeveupanddowntocalibratethehighlowdifference.

    Asshowninthefigure,adjusttheaxlesleevetillthecrosshairmovetothewhiteline.(Halfofthedifference)

    Afterhighlowdifference,thefourscrewsshouldbetighttomakesurethehorizontalaxisissteady,daubsomevanishonthescrews,anddotheplatevialcalibrationafter.

    LEFTHR

    LEFTLOW

    RIGHTHR

    RIGHTLOW

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 38

    38 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    6.Compensator

    Makesuretheplatevialisgoodandleveltheinstrumentonthecollimator,poweronandpressesc,youcanseethemodel,numberandversionofsoftware.

    SwitchontheTS,enter[systemsetup][systeminformation]pressF1fivetimesthenF45timestoactivethetestmode.

    Select[ADJCoefficient],the[compensationvalues]showstheelectronicbubbleoffsetvalue.

    XdirectioncalibrationloosentwoscrewsinAareaalittleknocktheleftandrightsideofthecompensatormaketheX=000'00",thentightenthescrews.

    YdirectioncalibrationloosenscrewCturningscrewB,untilY=000'00",thentightenthescrews.

    A A B C

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 39

    39 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    7. IAngle(VerticalAngleIndexerror)

    Leveltheinstrumentandaimatthecollimator(firstposition),asshowninfigure,matchthetotalstationcrosshairwithoneoftheverticalscalesonthecollimatorcrosshair.

    AsshowninthefiguretheV1is900000.

    Reversethetelescopeandaimatthesamescale,notedowntheverticalanglevalue.

    AsshowninfiguretheV2is2700050.

    Iangle=(v1+v2360)/2.Iangleshould

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 40

    40 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    Reversethetelescope,aimatthecollimatorasshowninfigure,matchthetotalstationcrosshairwiththesameverticalscalesonthecollimator.

    Presssettoendthecalibrationprocessandthesystemwillshowtheresultonthescreen.

    Presssettosave.

  • SOUTHSURVEYING&MAPPINGINSTRUMENTCO.,LTD. 41

    41 NTS960RSERIESTOTALSTATIONSERVICEMANUAL

    8. OpticalPlummetInspectionStepsandCriterion

    Settheinstrumentonatripod.Drawacrosshaironapieceofwhitepaperandputitontheground.Focustheopticalplummettoseethecrosshairclearlyandmovethepaper(orinstrument)tomatchtheopticalplummetcenterandcrosshaircentertogether.

    Rotatetheinstrumentevery90toseeiftheopticalplummetcenterstillmatchedwithcrosshaircenterornot.

    Ifnot,Asshowninfigure,rotateinstrument180,checktheopticalplummetcenteroffsetdistance.

    Asshowninthefigure,takeoffopticalplummeteyepiececoveroff,adjustthefourscrews.

    Movetheopticalplummetcenterahalftothecrosshaircenter.

    Movethepaper(orinstrument)tomatchtheopticalplummetcenterandcrosshaircentertogetherandchecktheopticalplummetasabove.

    Repeattheabovestepsuntiltheopticalplummetisgood.

    COVERTOOLSDisassembly1.Left cover2.LCD panel3. Angle-measuring board4.Right cover5.EDM cover6.EDM board7.Fiber8.Vertical absolute encoding disc9.Vertical CCD unit10.CCD bracket11.Vertical tangent screw12.Slip ring13.Axle Sleeve14.Axle Sleeve15.EDM16.Compensator17.Horizontal tangent screw18.Horizontal disc and vertical axle unit19.Horizontal CCD unit20.Vertical axle unit21.Plate vial22.Optical plummet23.Eyepiece and crosshair

    Common problem for angle measuringError code 1-7No display or can't power onAxle blockedAngle-measuring board socket name

    EDM fault and test process1.Check motors2.Check laser3.Check signal4.Check 3 axis5.Check onboard test point6.Check Optical couplerCommon faults and error

    Calibrations1.Plate vial2.Circular vial3.Crosshair4.2C5.High-low difference6.Compensator7.Index error8.Optical plummet