Kimo AQ200
-
Upload
okstorebor -
Category
Documents
-
view
230 -
download
0
Transcript of Kimo AQ200
-
7/30/2019 Kimo AQ200
1/4
AQ 200
Air quality
Connection
AQ 200P Current voltage module CO2/Temp./Hygrometry probe
++
The instrument
Advantages
Interchangeable measurement
modules
User-friendly (Joystick navigation)
Large graphic display
Blue backlight
Up to 8,000 measurement points
Up to 6 measurementssimultaneously
Instrument/PC wirelesscommunication
Wireless probes
Option
Option
Thermocouple temperaturemodule 4 channelsOptio
nOptio
n
Interchangeablesmeasurement modules1 instrument = more than 1 rangeand 1 parameter available.
Wireless connection
Instrument / PCInstrument / probe
Smart-plus systemWireless or wired probes automaticallyrecognized when connected to the instrument.
Climatic conditions moduleHygro/Temp. /Air pressure Wired or wireless temperature probe
or
CO / temperature probefrom 0 to 500 ppm /from -20C to +80C
CO2
/ temperature probe
from 0 to 5,000 ppm /from -20C to +80C
CO2
/ temperature / Hygrometry probe
from 0 to 5,00 0 ppm /from -20C to +80C / from 5 to
95%RH
Option
Option
Supplied
withCalibrationcertificate
AQ 200 Current voltage module CO2/Temp. probez
AQ 200 O Current voltage module Omnidirectionnal probeAir velocity/Hygro/Temp
+
+
-
7/30/2019 Kimo AQ200
2/4
Selection of units
Minimum / maximum values and hold function
Storage
Air quality
Functions
Thermometer
Technical features
AIR QUALITY PROBES :
CO / temperatureCO
2/ temperature
CO2
/ temperature / Hygrometry
Audible Alarm (2 setpoints)
Selection of units
Minimum / maximum values and hold function
Storage
CLIMATIC CONDITIONS MODULE
Thermocouple module :
Connection : 4 inputs for compensated miniatureplug of thermocouple type K, J or T Class 1(as per IEC 584-3 norm)
Current / voltage module :
Connection : 2 stereo jacks
Graphic display 128x128 pixelsDim. 50 x 54 mm
Blue backlightDisplay of 6 measurements (including 4 simultaneously)
On the top :
2 secured mini-DIN connectors for SMART-Plus probes
Left side :
1 USB port for KIMO cable only
1 power supply plug
Air quality probe
CO2
: NDIR infrared sensor (Non dispersive- infrared)
CO : Electrochemical sensorTemperature : Pt100 class 1/3 DinHygrometry : capacitive hygrometry sensor
Climatic conditions module
Hygrometry : capacitive hygrometry sensor
Temperature : semiconductor temperature sensorAir pressure : capacitive sensor
Thermocouple probes : type K, J and T class 1Pt100 Smart-plus probes : Pt100 class 1/3 Din
THERMOCOUPLE MODULE
Dynamic delta T Audible Alarm (lower and upper setpoints) Selection of units
Minimum / maximum values and hold function
4-channel storage of thermocouple K, J and T Calculation of U-value
TEMPERATURE PROBES
Dynamic delta T Audible Alarm (lower and upper setpoints) Selection of units
Minimum / maximum values and hold function
Storage
Current / voltage module Adjustable ranges Minimum / maximum values and hold function
Storage
Multi-parameters recording
Manual and automatic storage
Memory : up to 8,000 measurement points or50 datasets
User-friendly with printing of customized report
Management of instruments pool, follow-up ofcalibration periods
Intervention planning Wired or wireless interface
Datalogger-10
OMNIDIRECTIONAL PROBEAir velocity Temperature Hygrometry
Sensing elements
AQ200 connection
Interchangeable measurement modules
Display
Housing
Keypad
Conformity
Power supply
Operating environment
Operating temperature
Storage temperature
Auto shut-off
Weight
Languages
ABS shock-proofIP54
Metal-coated
5 keys1 joystick
Electromagnetical compatibility
(NF EN 61326-1 norm)
4 alkaline batteries 1,5V LR6
Neutral gas
from 0 to +50 C
from -20 to +80 C
adjustable from 0 to 120 min
340 g
French, English, Dutch, German, Italian, Spanish,Portuguese, Swedish, Norwegian, Finn, Danish
Omnidirectionnal probe
Air velocity : Thermistance with a negative temperature coefficientHygrometry/Temps.: capacitive sensor, Pt100 1/3 DIN
-
7/30/2019 Kimo AQ200
3/4
Specifications
Measuring units Measuring range Accuracy* Resolutions
V, mA From 0 to 2,5 V
From 0 to 10 V
From 0 to 4/20 mA
2mV
10mV
0.01mA
C, F
pp m
From -20 to +80C 0.3% of reading 0.25C 0,1 C
From 0 to 200 ppmFrom 200 to 500 ppm
3 ppm 0,1 ppm
CURRENT / VOLTAGE
CO / Temperature
Dimensions
From -200 to 1300C
From -100 to 750C
From -200 to 400C
1,1C or 0,4% reading***
0,8C or 0,4% reading***
0,5C or 0,4% reading***
0,1 C
0,1 C
0,1 C
C, F K :
J :
T :
THERMOCOUPLE (See related datasheet)
Side view Front viewTop view
Climatic conditions
Hygro.
Ai r pr es su re
%R H From 5 to 95%HR Accuracy** (Repeatability, linearity,
hysteresis) :
1,8%RH (from 15C to 25C)
Factory calibration uncertainty: 0,88 %RH
Temperature dependence :
0.04 x (T-20) %RH (if T25C)
0,1 %RH
From 800 to 1100 hPa 3 hPa 1 hPahP a
Temp. From -20 to +80C 0.4% of reading 0.3C 0,1 CC, F
Pt100 Smart-plus orwireless probes (See related datasheet)
From -50 to 250C(According to model)
0,3% of reading 0.25C(According to model)
0,01 CC, F
C, F
pp m
From -20 to +80C 0.3% of reading 0.25C 0,1 C
From 0 to 5000 ppm 3% of reading or 50ppm 1 ppm
CO2 / Temperature
C, Fpp m
%R H
From -20 to +80C 0.3% of reading 0.25C 0,1 C
From 0 to 5000 ppm
From 5 to 95%HR
3% of reading or 50ppm 1 ppm
CO2
/ Temperature / Hygrometry
See Climatic conditions module 0,1 %RH
CO
Temp.
CO2
Temp.
CO2
Temp.
Hygro.
0,001 V
0,01 V
0,01 mA
+
*All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with required compensation. ** As per NFX 15-113 and the Charter 2000/2001 HYGROMETERS, GAL (Guaranteed Accuracy Limit) which has been calculated with a coverage factor value of 2 is 2.88%RH between 18 and 28C on the measuring range from 5 to 95%RH. Sensor drift is less than 1%RH/year.***The accuracy is expressed either by a deviation in C, or by a percentage of the value concerned. Only th e bigger value is considered.
1.5% of reading
Omnidirectionnal probe (Air velocity / Temperature / Hygrometry)
Temp.
Ai r ve lo ci ty
Hygro.
m/ s C, F
%R H From 5 to 95%HR
From -20 to +80C
From 0 to 5 m/s
See Climatic conditions module
3% of reading 0.05 m/s
0.3% of reading 0.25C
0,01 m/s
0,1 C
0,1 %RH
-
7/30/2019 Kimo AQ200
4/4
FTang
AQ200
12/01/11
RCS(24)Prigueux349282
095Non-contractualdocument
Wereservetherighttomodifythecharacteristicsofourproductswithoutpriornotice.
Supplied with ...
Accessories (See related datasheet)
Working principle
NDIR - Non dispersive infrared absorbance
A gas absorbs light at a specific wavelength, some o f the intensity emittedby the infrared source is absorbed by the gas sample. The amount of lightread by the IR sensor is inversely proportional to CO
2concentration.
Electrochemical sensor
DESCRIPTION
Current / voltage
Thermocouple temperature module
Climatic condition module
CO2
/ temperature probe
CO / temperature probe
CO2
/ temperature / Hygrometry probe
Thermocouple K, J and T probe
SMART-Plus Pt100 probe
Wireless Pt100 probe
Omnidirectional probe (Air velocity/temperature/hygrometry)
8 rechargeable batteries with charger
Current / voltage cable
Calibration certificate
Transport case
AQ 200
Supplied with Option
Electrochemical cell consists of a container, 2 electrodes, connexion wiresand an electrolyte. Carbon monoxide is oxidised at one electrode to CO
2
whilst oxygen is consumed at the other electrode. The current produced isproportional to CO concentration.
AOH-, HCO
3
-, CO3
2-
Gas
Air
Working electrode
Auxiliary electrode
Electrolyte
e-
Infrared light Gas sample
Lightintensity
Lightintensity
Infraredsensor
Large choice of temperatureprobes(See related datasheet) :
ambient contact penetration Food industry penetration General use
Warranty period
Instruments have 1-year guarantee for any manufacturingdefect (return to our After-Sales Service required for appraisal).
AQ 200 P
Datalogger-10 KPIJ 20 50 100 200 - 600Datalogger-10 KPIJ 20 50 100 200 - 600 RTS
CE 200 GST ADS
CTC-PJAC CHA
4 batteriescharger
Set of 4 LR6batteries
Datalogger-10 PC software fordata recording and processing.Wired (LPCF) or wireless(LPCR) interface.
Ammeter clamp withPVC cablelg. 2m and jackconnector.
Telescopic extension,length 1 m, bent at 90 formeasuring probe.
Hands-free protectivecover
Silicone heat conductivegrease for temperatureprobes
Adaptor forpower supply230 Vac
Input current orvoltage input PVCcable lg 2 m and
jack connector
AQ 200 S AQ 200 O