Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary...
Transcript of Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary...
![Page 1: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/1.jpg)
Keller America Inc User’s Guide Rev. 07/20
![Page 2: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/2.jpg)
![Page 3: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/3.jpg)
Keller America Inc User’s Guide Rev. 07/20 Table of ContentsI. CompanyInformation..................................................................................2II. ContactInformation……......……………………………………………………………………..2III. ProductApplicability.………....………………………………………………………………….2IV. ProductOverview…………………………………………………………………………………….3V. Lightning/SurgeProtection.………………………………………………………………………4VI. Options/Accessories.………………………………………………………………………………5VII. Installation……....………………………………………………………………………………………6VIII. Useandcare...…………….…………………………………………………………………………..8IX. TermsandConditions…....…………………………………………………………………………9X. Appendix A. Generic2-wireCurrentLoopConfiguration………………………………A-1 B. Generic3-wireVoltageOutputConfiguration…………………………..A-1 i. Econoline 1. 4-20mA,Cable..................………………………………….......A-2 2. 4-20mA,mPmConnector..................………………………A-2 3. 4-20mA,M12Connector...................………………………A-2 4. 0.5-4.5VDC,Cable................…………………………………..A-3 5. 0.5-4.5VDC,mPmConnector................……………………A-3 6. 0.5-4.5VDC,M12Connector................……………………A-3 ii. Valueline 1. 4-20mA,Cable................…………………………………………A-4 2. 4-20mA,DIN43650Connector................…………………A-4 3. 4-20mA,MIL-C-26482Connector.................……………A-5 4. 0-5/10VDC,Cable.................……………………………………A-4 5. 0-5/10VDC,DIN43650Connector..................……………A-4 6. 0-5/10VDC,MIL-C-26482Connector.................………..A-5 iii. Preciseline 1. 4-20mA+RS485,Cable................…………………………….A-6 2. 0-5/10VDC+RS485,Cable.................………………………..A-63.SDI-12+RS485,Cable.........................................…………A-6 4. 4-20mA+RS485,MIL-C-26482Connector...................A-7 5. 0-5/10VDC+RS485,MIL-C-26482Connector.....………..A-7 6. SDI-12+RS485,DIN43650Connector........................…A-6 C. SubmersibleLevelTransmitters iv. Levelgage 1. 4-20mA,Cable................…………………………………………A-8 2. 0-5/10VDC,Cable.................………………………………….…A-8 v. LevelRat 1. 4-20mA+RS485,Cable..............………………………………A-8 2. 0-5/10VDC+RS485,Cable..............…………………………A-8 vi. Acculevel 1. 4-20mA+RS485,Cable.………………..............………………A-8 2. 0-5/10VDC+RS485,Cable..............…………………………A-8 3. SDI-12+RS485,Cable..................................................A-9 vii. Microlevel/Nanolevel 1. 4-20mA+RS485,Cable...………………………………...........A-9 D. InstallationTips.………………………………………..…………..............…A9-11
![Page 4: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/4.jpg)
Keller America Inc User’s Guide Rev. 07/20
I. Company Information
KellerAmericaIncistheNorthAmericansubsidiaryofKellerAGfürDruckmesstechnik,whichcontinuesastheleaderinlevelandpressuresensing.Fromhumblebeginnings,Kellernowprovidesitsproductsandexpertisethroughaworldwidenetworkoffacto-ry-trainedprofessionals.
Withtheadventofinexpensive,miniaturizedmicroprocessors,Kellercontinuestoad-vancethestate-of-the-artinfullypackagedpressuretransmitters,withTotalErrorBandperformancenotpossiblejustafewshortyearsago.Annualsensorproductionnowexceeds1millionpieces.
DoingbusinesswithKelleriseasyandmostinquiriesareansweredwithin24hours.StandarddeliveryonU.S.-assembledproductsis3daysARO.Weofferconvenienttermswithapprovedcreditandacceptmostmajorcreditcards.
AtKeller,providingcustomerswithsuperiorvalueisawayoflife.
II. Contact Information
KellerAmerica,Inc.351BellKingRdNewportNews,Virginia,23606P:877-253-5537F:757-596-6659Web: www.kelleramerica.comEmail: [email protected]
III. Product Applicability:Unlessotherwisenoted,thismanualcontainsuserinformationforthefollowinglevelandpressureinstrumentationprovidedbyKellerAmerica:Acculevel™,Levelgage™,LevelRat™,Microlevel™,Econolevel™,andNanolevel™submersibleleveltransmitters;Econoline™,Valueline™,andPreciseline™abovegroundpressuretransmitters.
Electronic copies of this user’s guide can be found at kelleramerica.com.
Page2
![Page 5: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/5.jpg)
Keller America Inc User’s Guide Rev. 07/20
IV. Product Overview - Level Transmitters
Levelgage-TheKellerLevelgageisageneralpurposesubmersibleleveltransmitteravailablein1%and0.5%BRTEBaccuracy.Constructedfrom316Lstainlesssteel,itisideallysuitedformostfreshwatermonitoringapplications.Withstandardconstruction(i.e.EPDMandXLPE)materials,theLevelgagehasbeencertifiedtoNSF61andNSF372standardsforuseindrinkingwater.Itisavailableincustomrangesfrom0…3ftWCto0…900ftWC.Cablelengthsarecuttouserspecifications.TheLevelgageisalsoavailableinvariousoutputconfigurations,including4…20mA,0…5VDC,and0…10VDC.
Acculevel-TheAcculevelfromKellerisahighaccuracysubmersibleleveltransmitter.IncorporatingallofthefeaturesoftheLevelgageincludingtheNSF61andNSF372certifications,theAcculevelhasadditionalfeaturesthatprovideenhancedfunctionality,includingdualoutputs;i.e.,oneanalog(4…20mA,0…5VDC,0…10VDC)andonedigital(RS485).TheRS485interfaceprovideslevelandtemperatureinformationandallowstheusertorescaletheanalogoutputfrom10…110%ofthebasicrange.Additionally,theAcculeveliscapableofgreateraccuracythantheLevelgage,withastandardaccuracyof0.25%BRTEBandanoptional0.1%BRTEB.TheAcculevelcanbemanufacturedin316Lstainlesssteelortitaniumforincreasedresistancetocorrosion,notablyfromseawater.
Acculevel SDI -TheAcculevelSDIisaversionoftheAcculevelhighaccuracysubmersibleleveltransmitterthatisavailablewithSDI-12orRS485digitalcommunicationprotocols.Itisavailablewithstandardaccuracyof±0.1%FSTEBinrangesupto900ftWCandoptionalUSGSOSWaccuracyforrangesupto70ftWC.TheAcculevelSDIcanbemanu-facturedinstandard316Lstainlesssteeloroptionaltitaniumforincreasedresistancetocorrosion,notablyfromseawater,carriestheNSF61andNSF372approvals.
LevelRat -TheKellerLevelRatisspecificallydesignedtoprovide0.5%BRTEBaccuracyforlevelindicationingrease-ladenenvironmentscommonlyfoundinliftstationsservingcommercialconcerns.TheLevelRatdesigntakesastepawayfromthelargeandbulkyprotectioncagedesignsbyutilizingauniquediaphragmmaterialthatoffersthenon-stickpropertiesofTeflonandprovidesaddedresistancetodamagefromdebris.TheLevelRatprovidesdualoutputs;yourchoiceofanalogoutputs(4…20mA,0…5VDC,or0…10VDC)plusonedigital(RS485)forlevelrangesfrom0…5ftWCto0…100ftWC.
Microlevel-TheMicrolevelcombinesthecapabilityoftheAcculevelwithaclassleading0.63”(16mm)outsidediameter.TheMicrolevelisavailableinfullscalerangesfrom0…3ftWCthrough0…900ftWCandprovides0.25%BRTEBaccuracy.LiketheAcculevel,italsoprovidesboth4…20mAanalogandRS485Digitaloutput.TheRS485interfaceprovideslevelandtemperatureinformation.Italsoallowstheusertorescaletheanalogoutputfrom10…110%ofthebasicrange.
Nanolevel-TheNanolevelisuniquelydesignedforhighaccuracymeasurementofverylowpressureranges.TheNanolevelisavailableinfullscalerangesfrom0…4inWCthrough0…10ftWCandisavailablewithstandardaccuracy0.25%BRTEBand0.1%BRTEBoptional.LiketheAcculevel,italsoprovidesboth4…20mAanalogandRS485Digitaloutput.TheRS485interfaceprovideslevelandtemperatureinformation.Italsoallowstheusertorescaletheanalogoutputfrom10…110%ofthebasicrange.
Page3
![Page 6: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/6.jpg)
Keller America Inc User’s Guide Rev. 07/20
Econolevel-TheEconolevelbyKellerisalowcostgeneralpurposesubmersibleleveltransmitterdesignedforhighvolumeapplications.Itisavailableinnominalpressurerangesfortankandwelllevelmonitoringapplications.TheEconolevelprovidesa4…20mAanalogoutputthatiscompatiblewithmostSCADAsystemsandPLCs.Mini-mumorderquantitiesapply.
IV. Product Overview cont’d - Pressure Transmitters
Econoline-TheEconolinebyKellerisalowcostgeneralpurposepressuretransmitter.Itisavailableinnominalpressurerangesbetween30psiand10,000psiandwithathread-ed¼”NPTpressureconnection.Idealfortank,pump,andpipepressuremonitoringapplication,itprovidesastandard4…20mAanalogoutputcompatiblewithmostSCADAsystemsandPLCs.Anoptional0.5…4.5VDC(ratiometricornon-ratiometric)analogout-putisalsoavailableonrequest.
Valueline-TheValuelinebyKellerisahighlyversatileandaccuratepressuretransmitter.Itisavailableincustompressurerangesfrom0…2through0…15,000PSIandyourchoiceofelectricalconnectionsincludingattachedcable,DIN43650,andMIL-C-26482connec-tors.Theattachedcableversionisfullysubmersible,makingittheidealchoicewheninstalledinavalvepitorotherenvironmentpronetoperiodicflooding.TheValuelinealsocarriestheNSF61andNSF372approvalsfordrinkingwaterapplications.
Preciseline -ThePreciselineisourmostaccurateabove-groundpressuretransmitter.Withstandardaccuracyof0.25%BRTEBandoptional0.1%BRTEB,a¼”NPTpressureconnection,anddualoutputs(analoganddigital)thePreciselineoffersversatilityindesignandapplication.LiketheAcculevel,theanalogoutputcanberescaledviatheRS485interface.TheRS485interfacealsoallowstemperaturedatatoberetrievediftheapplicationdemands.LiketheValueline,thePreciselineisincludedintheNSF61and372approvals.
Preciseline SDI-ThePreciselineSDIishighlyprecisepressuretransmitterthatisavail-ablewithSDI-12orRS485digitalcommunicationprotocols.Itisavailablewithstandardaccuracyof±0.1%FSTEBinrangesupto900ftWCandoptionalUSGSOSWaccuracyforrangesupto70ftWC.ItcomeswithyourchoiceofattachedcableorMIL-C-26482connectorelectricalconnections.Thecableversionisfullysubmersible,makingittheidealchoiceinenvironmentspronetoperiodicflooding,whiletheNSF61andNSF372approvalsmakethePreciselineSDIsuitableforuseindrinkingwaterapplications.
V. Lightning / Surge Protection:
Asofthiswriting,Kelleristheonlypressure/leveltransmittermanufacturerintheworldtoprovide,atnoadditionalcharge,alifetimewarrantyagainstdamagecausedbylight-ningand/orelectricalsurge.Thisguaranteeappliestothe4…20mAoutputversionofthefollowingproducts:Levelgage,Acculevel,LevelRat,Microlevel,Valueline,andPreciseline.Thiswarrantyapplieseveninthecaseofadirectlightningstrike.Theuser’sliabilityislimitedtoshippingcoststoKellerAmerica.
Page4
![Page 7: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/7.jpg)
Keller America Inc User’s Guide Rev. 07/20
VI. Options /Accessories:
Titanium Construction -OptionalfortheAcculevelandPreciseline.Highlyrecommend-edforbrackishandseawaterapplications.
External Surge Protector-Recommendedforlightning/surgeprotectionofuser’spowersupply/readout/PLC,etc.TobeusedinconjunctionwithLevelgage,Acculevel,LevelRat,Econolevel,Microlevel,Nanolevel,Valueline,orPreciselinetransmitters.
½” NPT Pipe conduit fitting -½”NPTmalefitting,allowsrigidmountingto½”conduitforLevelgage,Acculevel,orLevelRatsubmersibletransmitters.ItcanalsobeaddedtoourPreciselineandValuelinepressuretransmitters.
Drying Tube Assembly-Cleartubefilledwithindicatingdesiccant,attachesdirectlytocableventtube,interceptswatervapor.Highlyrecommendedwhenoperatingisinhighhumidityconditions.Mustbeperiodicallyrenewedasdesiccantbecomessaturated,turningcolorfromblue(dry)topink(saturated).
Bellows Assembly-Alternativetothedryingtube,thisaneroidbellowsattachestocableventtubeandrequiresnoperiodicmaintenance.Notethattheuseofthebellowsresultsinaclosedreferencepressuresystemsubjecttozeroshifterrorsinducedbychangingtemperaturesofupto0.0004psi/ºC.Notrecommendedforusewithaccuracyrequirementsof±0.25%FSorbetter,orranges<2psi.
Electrical Connectors -mPm393isoptionalfortheEconoline,DIN43650andMIL-C26482areoptionalforValuelineandPreciseline.ThemPm393andMIL-C26482matingconnectorsfeaturesoldercupswhiletheDIN43650incorporatesscrewterminals.
RS485 Converter/Cable -FortheuserwhowishestocommunicatewiththeirAcculevel,LevelRat,Nanolevel,Microlevel,orPreciselinetransmitterviatheRS485digitalinterface.AvailableinRS232andUSBversions,alongwithcableadapters,asneeded,andCCS30Software(alsoavailablefordownload,freeofchargefromwww.kelleramerica.com).Consultfactoryfordetails.
Termination Enclosure-Convenientoptioncomplementinggauge-typepressure/leveltransmitters,whereitisdesiredtoterminatethetransmittercableclosetothemea-surementpoint.ItincludesaNEMA4Xclearfrontenclosure(7.9X4.7X3.5inches)withtwo,liquid-tightcablefittings(onein,oneout),aterminalstrip,andampleroomformountingadryingtubeorbellowsassembly.
Stabilizing Weight-ZincpropshaftanodeadaptedtofitØ21mmO.D.ofLevelgage,Acculevel,andLevelRatsubmersibleleveltransmitters.Aidsincorrosionresistanceaswellashelpsensurethatthecableremainstautinturbulentconditions.
Cable Hanger-Singleeyefor0.230”diametercable.Designedtogripthecablejacketandprovideasuspensionpointforinstallationofcabledtransmitters.
Pressure Test Adapters -Designedtofacilitatecalibrationandmaintenancetesting,thesespecialtyadaptersallowforconnectionofthevariousKellerleveltransmitterstoapressurecalibrator.
Page5
![Page 8: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/8.jpg)
Keller America Inc User’s Guide Rev. 07/20
Digital Display / Meter -TheTRIDENTdisplaybyPrecisionDigitalisaversatileandcosteffectiveprocessmeterprovidinganeasytouseandreaddisplay/controllerforyourlevelandpressuremeasurementapplication.Formoreinformationandtechnicalsup-port,pleasecontactPrecisionDigitalat(800)343-1001oronlineatwww.predig.com.
VII. Installation:
Thefollowingisimportantinstallationandgeneralmaintenanceinformationforsub-mersibletransmitters.PleasecontactKellerAmericaforadditionalinstruction.
Supply Voltage-Adequatesupplyvoltageiscriticaltoensureproperoperationof4-20mApressureandleveltransmitters.Withouttheminimumrequiredvoltageavailableatthetransmitter,thetransmitterwillnotoutputthecorrectanalogvalue.
Manyanalogtransmitterswillappeartooperateproperlyevenwhenthesupplyvoltageisnotadequatetopowertheloopwhenthetransmittershouldbeoutputting20mA.Forexample,a10voltsupplymayappeartobeenoughtopowerananalogtransmitterwhenitisoutputting4mAwithzeropressureapplied,butasthetransmitter’soutputincreaseswithincreasingpressure,voltagedropsacrossotherdevicesintheloop(ana-loginputdevices,cableand/orexternalbarrierdevices)mayreducethesupplyvoltagetothetransmitterandpreventitfromprovidingthecorrectoutputaboveacertaininputpressure/levelthreshold.
KellerAmerica’s4-20mApressureandleveltransmittersfeaturemicroprocessorbasedsignalconditioning.Duringpowerup,thecircuitperformsachecksequencewhichde-termineswhetherthereissufficientsupplyvoltagetopoweralldevicesontheloop,bysettingtheoutputto~110%ofthemaximumvalue,i.e.,~22.5mA.Ifthesupplyvoltageisnotsufficienttosupply22.5mAtothecircuit,thenthemaximumpossiblecurrentwillbeseenontheanalogoutput,e.g.,17mA,andthetransmitterwillnotinitiatenormaloperation.
Thebenefitofthistechnologyisthattotalloopimpedanceisaccountedforpriortoplacingtheequipmentintoservice,preventingfalseindicationswhenthevoltagesupplyisinsufficienttosupporttheloopwithmaximumpressureapplied.
Mostcurrentloopscontainanaloginputdevices,indicatorsorothercomponentshavinginputimpedanceswhichmustbeconsideredwhencalculatingthesupplyvoltagesneed-ed.Forexampleananaloginputdevicewitha250OhmInputImpedancewillrequireanadditionalvoltageof5.5VDC(250x0.022=5.5)inadditiontotheminimumsupplyvoltagenecessaryforthetransmittertooperateproperlyovertheentirerange.
Page6
Total LoopImpedance
IndicatorAnalog Input
Device
Cable
ResistanceIS Barrier
4 - 20mA outputLevel / Pressure
Transmitter
![Page 9: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/9.jpg)
Keller America Inc User’s Guide Rev. 07/20
Thecalculationsbelowareusefulinidentifyingtheminimumsupplyvoltageneededforacurrentloopwithadditionallineimpedancesorconversely,themaximumcurrentloopimpedanceallowedforagivensupplyvoltage.
Minimumsupplyvoltagewithlightningprotectionoptioninstalled: 11VDC+(Totalloopimpedancex0.022)
Minimumsupplyvoltagewithoutlightningprotectionoptioninstalled: 8VDC+(TotalloopImpedancex0.022)
Maximumallowableloopimpedancewithlightningprotectionoption installed: (SupplyVDC–11VDC)/0.022
Maximumallowableloopimpedancewithoutlightningprotectionoption installed: (SupplyVDC–8VDC)/0.022
Transmitter Anchoring-ItisrecommendedthatKellersubmersibletransmittersbeinstalledinastillingwellorattachedtorigidconduitviaaconduitfittingintegraltothetransmitter,inordertopreventdamagetothetransmitterfromimpactwithimmovableobjects.Itisnotadvisabletotiethetransmittertoapumportopiping,asanyproblemwiththetransmittercouldrequirethatthepumpbepulledfromtheinstallation.
Someapplicationsrequirethetransmittertobesuspendedwithoutaprotectivestillingwellorconduitattachment.Inallinstallations,careshouldbetakentopreventdamagetothesubmersiblecable.
Transmitter Submersion -Damagetosubmersiblecablecanleadtofailureofthetrans-mitter.Kelleremploysaruggedcablejacketmaterialstominimizetheriskofcutsandabrasion.Still,takecarewhenloweringyourtransmitterintothewell,makingsurethecabledoesnotdragoversharpedges.Avoiddroppingthetransmitterfromthesurface.Donotsubmergeelectricalconnectionendofcable.
Condensation protection -Kellerhasoptimizedthesizeofthecableventtominimizetheoccurrenceofwatervaporincursion.Inareasofhighhumidity,itmaybedesirabletouseaDryingTubeAssembly(desiccant)orBellowsAssemblytopreventwatervaporfromenteringtheventtube.ContactKellerAmericafororderinginformation.
Bending of Cable-Ourjacketedcableisquiteflexible.However,caremustbetakentoensuretheventtubeintegraltothecableisnotcrimpedwhenbendingthecabletosuityourinstallation.Itisrecommendedthatthecablenotbebenttoaradiussmallerthan1inch.
Cable Compression -Manyusersemployacompressionfittingtosecureourcableasitentersajunctionbox.Caremustbetakenthatthefittingisnotovertightened,causingdamage to the cable and/or crimping the vent tube.
Page7
![Page 10: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/10.jpg)
Keller America Inc User’s Guide Rev. 07/20
Position Sensitivity-Thetransmittershouldbeinstalledinaverticalposition,other-wiseitmayexhibitanoffset.Ifthetransmittermustbeinstalledinanypositionotherthanvertical,measuretheoutputwithnopressureappliedpriortoconnectiontoyourdisplay,PLC,orcontroller.Usethemeasuredvalueforyourzeropoint.
VIII. Use and Care:
Safe Handling-SafehandlingofKELLERpressuremeasurementdevicesisaccomplishedifanominalamountofcareistaken.Thingstoavoidare:•Sharpimpactagainsthardsurfaces•Contactwithchemicalsknowntobecorrosivetothematerialsofconstruction•ProbingofpressuresensingmembranewithANYTHING
Limits of Pressure -KELLERpressuresensors,transducersandtransmittersaredesignedtowithstandacertainamountofoverpressurewithoutdamageorcalibrationshift.Itcanrangefrom15Xforthelowerpressurerangesto1.1Xforthehighestranges.Thisvalueisdifferentforeachproductandisreferredtointheproductliteratureas“Overpressure”.Itistheuser’sresponsibilitytoensurethattheproperKELLERproductischo-senfortheparticularpressureconditionsexpected.
Environmental conditions-EachKELLERproductisdesignedtobecompatiblewithaparticularenvironment.Itistheuser’sresponsibilitytoensurethattheKELLERproductisnotexposedtoanenvironmentalconditionforwhichitisnotdesigned.Thesecondi-tionscanincludeoperatingtemperaturerangeandexposuretohigh-pressurewaterjets,medianotcompatiblewiththematerialsofconstruction,submergenceoftransmittersnotdesignedforthatpurpose,orpotentiallyexplosiveatmospheres.AKELLERengi-neercanhelptheuserdeterminethecorrectchoiceofenclosuretosuittheparticularapplication.
Electrical conditions -EachKELLERproductisdesignedtooperateproperlywithinaspecificrangeofelectricalconditions.Thespecificproductlabeldefinestherating(s),ifany,thatappliestotheproducttowhichitisaffixed.Alltransmittersaredesignedtowithstandreversepolarityaswellasovervoltagetoacertainextent.Itistheuser’sresponsibilitytoensurethatallelectricalconnectionsaremadetotheKELLERproductsinaccordancewithKELLERrecommendationsaswellaslocalelectricalcode.Wirecolorsorconnectorpin-outsareeitherprintedonthelabelaffixedtotheproductorprovidedseparately.
Cleaning-WARNING!Undernocircumstancesshouldthemembrane,pressureinputport,ornosecaptotheKELLERdevicebeprobedwithanyobject.Damagetothesens-ingmembraneispermanentand,inmostcases,requiresrepairorreplacement.
Regardingmedia-isolatedproducts,shouldthepressureinputornosecaptothesensor,transducer,ortransmitterbecomefouled,itmaybecleanedinthefollowingmanner:
Thedeviceshouldbeslowlyloweredmembrane-endintoasolutionofwarm,soapywater.Careshouldbetakennottosubmergetheentiredevice,unlessitisspecificallydesignedforcontinuoussubmergence.Agitateinthesolutionfor20-30secondsoruntil
Page8
![Page 11: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/11.jpg)
Keller America Inc User’s Guide Rev. 07/20
theinput/nosecaptothedeviceisclear.Finishbystirringinfreshwater.Wipedrywithasoftragortowel.
Asoapscumandhard-waterstainremovermayalsobeused,ifnecessary,butonlyaftercompatibilitywithanyo-ringsealsintheKELLERproductisdetermined.Followthesolventmanufacturer’srecommendationsforsafehandling.
IX. Terms and Conditions:
Offers -Unlessotherwiseindicated,quotedpricesarevalidfor60daysfromthedateofthequotationandapplytoproductsthatarescheduledforshipmentwithintwelvemonthsfromthedateoftheBuyer’spurchaseoratKellerAmerica,Inc.shippingcapabil-itiesatthetimetheorderisentered,whicheverislater.
Acceptance of order-Anypurchaseordertobeacceptedhastorefertoaquotation,whetherdeliveredbysurfacemail,e-mailorfacsimile.Alloralordersmustbeconfirmedinwriting.Byreferringtothisquotation,BuyeracceptsandadoptstheGeneralTermsandConditionsofSale(“TermsandConditions“)totheexclusionofanyadditionalordifferenttermsappearinginBuyer’spurchaseorderandwaivesanyrightBuyermayhavetoenforceanysuchadditionalordifferentterms.OurConfirmationofOrderorInvoicevalidatesorders.Ordersforcustom-configuredproductsarebuilttoyourspecifi-cationandthereforemaynotbecanceledoncetheorderhasbeenconfirmed.
Date of shipment-TheestimatedshippingcapabilitystatedonthequotationisgivensolelyfortheBuyer’sinformationanddoesnotconstituteacommitmenttodeliverproductsinaccordanceherewith.Buyermayrequestaspecificshippingdateorshippingschedule.KellerAmerica,Inc.willscheduleshipmentsbaseduponBuyer’srequestandKellerAmerica,Inc.’sshippingcapabilitiesatthetimeBuyer’spurchaseorderispro-cessed,atwhichtimeKellerAmerica,Inc.willissuetoBuyeraformalAcknowledgmentofOrderthatwillindicatetheestimatedshippingdate(s).Aftertheshipmentshavebeenscheduled,BuyermaynotcancelorpostponeascheduledshipmentunlessBuyersub-mitsitsrequestinwritingandKellerAmerica,Inc.consentsbyissuinganewacknowl-edgment.
Anyrequesttocancelortorescheduletheshipmentthatisreceivedlessthan21daysfromthedatescheduledfortheshipmentoftheproductscoveredbytherequestmayberejectedasuntimely,ormaybeaccepteduponpaymentofthecancellationchargeof20%orareschedulingchargeof10%ofthesalespriceshownonthisquotationatKellerAmerica,Inc.’soption.KellerAmerica,Inc.willuseitsbesteffortstoshiponorevenbe-foretheestimatedshippingdatesindicated,butwillnotbeliableforanydelayorfailuretodeliver.KellerAmerica,Inc.shallnotbeliableforanyspecial,incidentalorconsequen-tialdamagesresultingfordeliverydelaysorinabilitytodeliver. AllgoodsormaterialssuppliedbyKellerAmerica,Inc.remainitspropertyuntiltotalpay-mentisreceived.AllsuchgoodsandmaterialsareatthesoleriskoftheBuyerandintheeventofbeingdamaged,destroyedorlostafterdelivery.KellerAmerica,Inc.isentitledtoreceivethetotalpaymentofthosegoods.
Page9
![Page 12: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/12.jpg)
Keller America Inc User’s Guide Rev. 07/20
Price -Unlessotherwisespecifiedinthequotation,ourpricesareinUSdollars(USD).Theyareexclusiveoffreightcostsandofallstateandlocalsales,use,excise,privilegeandsimilartaxes.SuchtaxesimposedonKellerAmerica,Inc.orwhichKellerAmerica,Inc.hasadutytocollectinconnectionwiththesaleordeliveryoftheproductsde-scribedonthequotationshallbepaidbyBuyerandwillappearasseparateitemsontheinvoice.
Shipment / freight -DeliveriesareF.O.B.NewportNews,VA.TitleandriskoflossshallpasstoBuyerupontenderoftheproductsbyKellerAmerica,Inc.toacommoncarrier.InabsenceofspecificwritteninstructionfromBuyer,KellerAmerica,Inc.willselectthecommoncarrier,butKellerAmerica,Inc.shallnottherebyincuranyliabilityinconnec-tionwithshipment.Buyershallberesponsibleforanyfreightcharge.Declaredvalueforeachshipmentwillbeamaximumof$100regardlessoftheactualvalueofthegoods,notwithstandingwritteninstructionsfromtheBuyerspecifyingahigherdeclaredvalue.Iftheproductsareshippedfreightprepaid,BuyershallpayKellerAmerica,Inc.theap-propriatefreightcharges,whichwillbeshownasseparateitemsontheinvoice.
Payment -TermsareindicatedonourConfirmationofOrderandonourInvoices.KellerAmerica,Inc.willsubmitaninvoicetoBuyerforeachshipmentatthetimeofshipment.Exceptasotherwiseprovidedonthequotation,BuyershallpaytheamountinvoicedbyKellerAmerica,Inc.within30daysfromthedateoftheinvoice.IfinthejudgmentofKellerAmerica,Inc.thefinancialconditionorpaymentrecordofBuyeratanytimedoesnotjustifyshipmentunderthepaymenttermsspecifiedabove,KellerAmerica,Inc.mayrefusetoshipunlessitreceivespaymentinadvance,oratitsoption,paymentupondelivery.2%interestpermonthischargedonoverdueaccounts.
Documentation -Theinformationgiveninourdocumentation,printedmatter,datasheetsandpricelistsiswithoutcommitment.Thisinformationspecifiestheproductbutisnowarranty,unlessagreedinwriting.
Warranties -KellerAmerica,Inc.warrantsthattheproductsthatitsellsaredeliveredfreefromdefectsinmaterialandworkmanship.KellerAmerica,Inc.’sliabilityunderthiswarrantyislimitedtoreplacingorrepairingorissuingacreditnote,atitsoption,foranyproductwhichisreturnedtothefactory,transportationchargesprepaid,andwhichisdeterminedbyKellerAmerica,Inc.tobedefective.Thiswarrantydoesnotapplytobatteriesandaccumulatorsandtoanyproductwhichhasbeensubjectedtoordam-agedduetomisuse,misapplication,negligenceoraccident,orwhichhasbeenrepairedoralteredwithoutexpresspriorandwrittenconsentfromKellerAmerica,Inc.Inordertoobtainserviceunderthetermsofthewarranty,BuyermustnotifyKellerAmerica,Inc.ofanydefectsbeforetheexpirationofthewarrantyperiodandmakesuitablearrangementsfortheperformanceofservices.Inallcases,BuyershallberesponsibleforpackagingandshippingtheproductstotheKellerAmerica,Inc.plantwithshippingchargesprepaid.KellerAmerica,Inc.shallpayforthereturnofanyproductstoBuyeriftheshipmentistoalocationwithinthecontinentalUSA.Buyershallberesponsibleforpayingallshippingduties,taxes,andotherchargesforproductsreturnedtoanyotherlocation.KellerAmerica,Inc.willprovideon-siteserviceonlyuponprioragreementanduponpaymentofalltravelexpensesbyBuyer.
Page10
![Page 13: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/13.jpg)
Keller America Inc User’s Guide Rev. 07/20
WarrantyperiodfortheLevelgage,LevelRat,Acculevel,Microlevel,Preciseline,andVal-uelineproductsis24(twenty-four)monthsfromdateofshipment.Warrantyperiodforallotherproductshereinis12(twelve)monthsfromdateofshipment.Warrantyrepaircoversallapplicablepartsandlabor.Thiswarrantyisgiveninlieuofanyotherwarranty,expressorimplied.KellerAmerica,Inc.explicitlydisclaimsanyimpliedwarrantiesofmerchantabilityandfitnessforaparticularpurpose.Therearenowarranties,expressorimplied,thatextendbeyondthedescriptionherein.ThesoleandexclusiveremedyforanyclaimsagainstKellerAmerica,Inc.shallbethewarrantydescribedinthisdocument.Alldamages,directorconsequential,limitedtothedescribedwarrantyareexcluded.
Repairs-UnlessotherwiserequestedinwritingbytheBuyer,allproductsreturnedtoKellerAmerica,Inc.underthetermsofthewarrantywillbecheckedandanalyzedinor-dertodeterminethecauseofthedefault(s)claimedbytheBuyer.AreportwillthenbesubmittedtotheBuyerpointingoutthenatureofthedefault(s),thepartyresponsibleforthedefault(s)andthequotationoftherepair,ifneeded.Forfurtherrepairinstruc-tionsgotowww.kelleramerica.com.
Complaints-AllclaimsordisputesmustbemadeinwritingtoKellerAmerica,Inc.amaximumof60daysfromreceiptofthegoods,includingdiscoveryoffaultsnotpre-viouslyapparent.Ifthewarrantyclaimsarejustified,KellerAmerica,Inc.isfreeatitsdiscretiontorepair,replaceorissuecredit.Nofurthercompensationfordamageswillbemade.
AnydisputesorclaimsofBuyermustbeinitiatedinapropercourtorotheradjudicativebody,asapplicable,withinone(1)yearfromthedateofshipmentbyKellerAmerica,Inc.,oritsrepresentative(s),orsuchclaimshallbedeemedinvalidorexpiredandcannotberenewed.Totheextentallowed,thislimitationperiodshalltrumpanyapplicablestatutorylimitationsperiodthatmaystatealongerperiod.
Limitation of responsibility -Ourresponsibilityconcernsthecorrectfunctionofourproductsonly.Itcannotbeextendedtothewholesysteminwhichtheyareused.Ourresponsibilityislimitedtothereplacement,repairorreimbursementofthegoodsweagreearedefectiveornon-conforming.Theclaimmustbeinwritingwithin60daysfromreceiptofthegoods.
Place of jurisdiction / applicable law -ThecontractmadebyacceptanceofthisoffershallbedeemedmadeintheStateofVirginiaandshallbegovernedbyandconstruedinaccordancewiththelawsofthatstatewithoutreferencetoorapplicationofanyconflictsoflawsprinciplesandwithoutconsiderationoftheplaceofexecution.BuyerexpresslyagreestothesubjectmatterandpersonaljurisdictionoftheCircuitCourtsfortheCityofNewportNews,Virginia,orthefederalDistrictCourtfortheEasternDistrictofVirginia,NewportNewsDivision.
Assignment-NeitherthisoffernoranycontractresultingtherefrommaybeassignedortransferredinwholeorpartwithoutthepriorwrittenconsentofKellerAmerica,Inc.NoassignmentortransferinviolationofthisprovisionshallbevalidorbindingonKellerAmerica,Inc.
Page11
![Page 14: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/14.jpg)
Keller America Inc User’s Guide Rev. 07/20
Attorney’s fees-UponanybreachofthisAgreement,thereasonableattorney’sfeesandcostsofthesubstantiallyprevailingparty,whetherbylitigationorsettlement,shallbepaid by the breaching party.
Page12
![Page 15: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/15.jpg)
Keller America Inc User’s Guide Rev. 07/20
X: Appendix
+ SUPPLY
OUTPUT/GROUND
2-WIRE
TRANSMITTER
POWER
SUPPLY
DISPLAY/
CONTROL
+
-
-
+
2-WIRE CURRENT LOOP
3-WIRE VOLTAGE OUTPUT
+ SUPPLY
+ OUTPUT
3-WIRE
TRANSMITTER
POWER
SUPPLY
DISPLAY/
CONTROL
+
-
-
+
GROUND
A-1
PleaserefertoSectionVII:Installationforminimumsupplyvoltageinformation.
![Page 16: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/16.jpg)
Keller America Inc User’s Guide Rev. 07/20
12
3
4
DRAIN/CASE GROUND
BLACK: + SUPPLY
WHITE: OUTPUT/GROUND
ECONOLINE w/ PVC CABLE...4-20mA
ECONOLINE w/mPm393...4-20mA
DRAIN/CASE GROUND
BROWN or BLACK: + SUPPLY
WHITE: OUTPUT/GROUND
ECONOLINE w/ PUR CABLE...4-20mA
CASE
+ SUPPLY
OUTPUT/GND
VIEW OF REAR OF TRANSMITTER
ECONOLINE w/M12...4-20mA
+ SUPPLY
OUTPUT/GND
VIEW OF REAR OF TRANSMITTER
4 3
2 1
A-2
NOTES: 1: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.
![Page 17: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/17.jpg)
Keller America Inc User’s Guide Rev. 07/20
12
3
4
DRAIN
BLACK: + SUPPLY
WHITE: GROUND
RED: + OUTPUT
ECONOLINE w/CABLE...0,5-4,5VDC
ECONOLINE w/mPm393...0,5-4,5VDC
CASE
+ SUPPLY
GROUND
VIEW OF REAR OF TRANSMITTER
+ OUTPUT
ECONOLINE w/M12...0,5-4,5VDC
+ SUPPLY
GROUND
VIEW OF REAR OF TRANSMITTER
4 3
2 1
+OUT
NOTES: 1: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.
A-3
![Page 18: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/18.jpg)
Keller America Inc User’s Guide Rev. 07/20
NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/ SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.
VALUELINE w/ATTACHED CABLE...4-20mA
VALUELINE w/ATTACHED CABLE...0-5...10VDC
VALUELINE w/DIN43650...4-20mA
VIEW OF REAR OF TRANSMITTER
VALUELINE w/DIN43650...0-5...10VDCVALUELINE
+ OUTPUT
VIEW OF REAR OF TRANSMITTER
3
12
3
12
DRAIN
BLACK: + SUPPLY
WHITE: OUTPUT/GROUND
DRAIN
BLACK: + SUPPLY
WHITE: GROUND
RED: + OUTPUT
+ SUPPLY
OUTPUT/GROUND
SEE NOTE 1.
+ SUPPLY
GROUND
A-4
![Page 19: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/19.jpg)
Keller America Inc User’s Guide Rev. 07/20
AB
CD
E
F
AB
CD
E
F
+ SUPPLY
OUTPUT/GROUND
VALUELINE W/MIL-C-26482...4-20mA
REAR VIEW OF TRANSMITTER
VALUELINE W/MIL-C-26482...0-5...10VDC
+ SUPPLY
GROUND
+ OUTPUT
REAR VIEW OF TRANSMITTER
E: DRAIN
A-5
NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/ SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.
![Page 20: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/20.jpg)
Keller America Inc User’s Guide Rev. 07/20
DRAIN (NOTE 1)
BLACK: + SUPPLY
WHITE: OUTPUT/GROUND
BLUE: RS485A
YELLOW: RS485B
DRAIN (NOTE 1)
BLACK: + SUPPLY
WHITE: GROUND
RED: + OUTPUT
BLUE: RS485A
YELLOW: RS485B
PRECISELINE w/ATTACHED CABLE...4-20mA + RS485
PRECISELINE w/ATTACHED CABLE...0-5...10VDC + RS485
DRAIN (NOTE 1)
BLACK: GROUND
WHITE: SDI-12
RED: + VCC
BLUE: RS485A
YELLOW: RS485B
PRECISELINE SDI w/ATTACHED CABLE...SDI-12+ RS485
A-6
NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/ SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.
![Page 21: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/21.jpg)
Keller America Inc User’s Guide Rev. 07/20
PRECISELINE w/MIL-C-26482...4-20mA + RS485
REAR VIEW OF TRANSMITTER
PRECISELINE w/MIL-C-26482...0-5...10VDC + RS485
REAR VIEW OF TRANSMITTER
A: + SUPPLY
C: OUTPUT/GROUND
A: + SUPPLY
C: GROUND
B: + OUTPUT
D: RS485A
F: RS485B
D: RS485A
F: RS485B
E: DRAIN
AB
CD
E
F
AB
CD
E
F
PRECISELINE SDI w/MIL-C-26482...SDI-12 + RS485
REAR VIEW OF TRANSMITTER
A: + SUPPLY
C: GROUND
B: SDI-12F: RS485B
D: RS485A
AB
CD
E
F
A-7
NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/ SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.
![Page 22: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/22.jpg)
Keller America Inc User’s Guide Rev. 07/20
NOTES: 1: FOR LIGHTNING/SURGE PROTECTION TO BE EFFECTIVE, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. 2: LIGHTNING PROTECTION NOT AVAILABLE ON ECONOLEVEL. DRAIN WIRE CONNECTION NOT REQUIRED. 3: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.
LEVELGAGE & ECONOLEVEL w/ATTACHED CABLE...4-20mA
LEVELGAGE w/ATTACHED CABLE...0-5...10VDC
ACCULEVEL & LEVELRAT w/ATTACHED CABLE...4-20mA + RS485
ACCULEVEL & LEVELRAT w/ATTACHED CABLE...0-5...10VDC + RS485
DRAIN (NOTE 1,2)
BLACK: + SUPPLY
WHITE: OUTPUT/GROUND
DRAIN (NOTE 1)
BLACK: + SUPPLY
WHITE: GROUND
RED: + OUTPUT
DRAIN (NOTE 1)
BLACK: + SUPPLY
WHITE: OUTPUT/GROUND
BLUE: RS485A
YELLOW: RS485B
DRAIN (NOTE 1)
BLACK: + SUPPLY
WHITE: GROUND
RED: + OUTPUT
BLUE: RS485A
YELLOW: RS485B
A-8
![Page 23: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/23.jpg)
Keller America Inc User’s Guide Rev. 07/20
NOTES: 1: FOR LIGHTNING/SURGE PROTECTION TO BE EFFECTIVE, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.
MICROLEVEL & NANOLEVEL w/ATTACHED CABLE...4-20mA + RS485
DRAIN (NOTE 1)
BLACK: + SUPPLY
WHITE: OUTPUT/GROUND
BLUE: RS485A
YELLOW: RS485B
DRAIN (NOTE 1)
BLACK: GROUND
WHITE: SDI-12
RED: + VCC
BLUE: RS485A
YELLOW: RS485B
ACCULEVEL SDI w/ATTACHED CABLE...SDI-12+ RS485
A-9
![Page 24: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/24.jpg)
Keller America Inc User’s Guide Rev. 07/20
A-10
MINIMUM PIPE BEND RADIUS
Acculevel/ LevelgageØ0.83”
MicrolevelØ0.63”
Schedule40
1”IPS(Ø1.315x0.133”w) 22” 18”
1.5”IPS(Ø1.900”x0.145”w) 4” 5”
2”IPS(Ø2.375x0.154”w) 3” 3”
Schedule80
1”IPS (Ø1.315x0.179”w) 22” 22”
1.5”IPS(Ø1.315x0.200”w) 4” 6”
2”IPS(Ø2.375x0.218”w) 4” 4”
![Page 25: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/25.jpg)
Keller America Inc User’s Guide Rev. 07/20
A-11
![Page 26: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/26.jpg)
Keller America Inc User’s Guide Rev. 07/20
A-12
Signal cable
Transmitter
Vent tube
Bellows Assembly
Drying Tube Assembly
VENT TUBE TERMINATION - DRYING TUBE OR BELLOWS
Remove yellowcap after installation
NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DRYING TUBE ASSEMBLY - CLEAR TUBE FILLED WITH INDICATING DESICCANT, ATTACHES DIRECTLY TO CABLE VENT TUBE, INTERCEPTS WATER VAPOR. HIGHLY RECOMMENDED WHEN OPERATING IS IN HIGH HUMIDITY CONDITIONS. MUST BE PERIODICALLY RENEWED AS DESICCANT BECOMES SATURATED, TURNING COLOR FROM BLUE (DRY) TO PINK (SATURATED). REMOVE YELLOW CAP FROM FREE END AFTER INSTALLATION. 3: BELLOWS ASSEMBLY - ALTERNATIVE TO THE DRYING TUBE, THIS ANEROID BELLOWS ATTACHES TO CABLE VENT TUBE AND REQUIRES NO PERIODIC MAINTENANCE. RECOMMENDED WHERE A SLIGHT SACRIFICE IN ACCURACY CAN BE TOLERATED.
![Page 27: Keller America Inc User s Guide Rev. 02/20 · Keller America Inc is the North American subsidiary of Keller AG für Druckmesstechnik, which continues as the leader in level and pressure](https://reader033.fdocuments.us/reader033/viewer/2022050109/5f4734330457a921352bfd70/html5/thumbnails/27.jpg)
Electronic copies of this user’s guide can be found at kelleramerica.com
BORN SWISS · MADE AMERICAN · KELLERAMERICA.COM