Gesellschaft zur Förderung der Qualitätssicherung in medizinischen Laboratorien
INSTAND e.V.
JCTLM Member activities
Prof. Dr. Michael Spannagl Dr. Patricia Kaiser
2
EQAS organizer worldwide
infectious disease
transplant immunology transfusion coagulation haematology … POCT
3
0
2000
4000
6000
8000
10000
12000
1993 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015
Number of participants from 1993-2015
german participants foreign participants
4
RiliBÄK - Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Testing
Section B1
5
Section A
RiliBÄK - Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Testing
Section B1
6
Section A
RiliBÄK - Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Testing
Sections B1 B2 B3 B4 B5
7
New Guideline for Quality Assurance of the German Medical Association (RiliBÄK)
-> mandatory for all medical laboratories in Germany
Section A Fundamental requirements for quality assurance
Effective since 1 April 2008 Update 2011
Specific Section B 1 Quantitative analyses in laboratory medicine - internal QA - external QA
Effective since 1 April 2008
Specific Section B 2 Qualitative analyses in laboratory medicine - internal QA - external QA
Effective since 1 July 2011
Specific Section B 3 Direct detection and characterization of pathogens - internal QA - external QA
Effective: 1 June 2015
Specific Section B 4 Analyses of ejaculate - internal QA - external QA
Effective since 1 July 2011
Specific Section B 5 Molecular- and cytogenetic analyses - internal QA - external QA
Effective since 1 October 2011
reimbursement
8
New Guideline for Quality Assurance of the German Medical Association (RiliBÄK)
-> mandatory for all medical laboratories in Germany (extract from annex B1)
9
Reference laboratory
Accredited as calibration laboratory according to ISO 17025 and ISO 15195
2006 Cholesterol 2009 Chloride, Potassium, Sodium, ALT, AST, GGT, CK, LDH Creatinine, Urea, Cortisol, Total Protein, HbA1c, Theophyline 2012 Calcium, Lithium, Magnesium, Uric Acid, Total Gycerides, Testosterone, Thyroxine 2015 Glucose, 17-ß-Estradiol, Progesterone, Digoxin, Digitoxin
RMP
at IN
STAN
D e.
V.
RiliBÄK requirements
pH, Phosphate Total hemoglobin, cell counting
10
Interlaboratory CV Glucose
EQAS results 2002 - 2015
0
1
2
3
4
5
6
7
8
coef
ficie
nt o
f var
iatio
n [%
] all manufacturers
target values set by RMP
devi
atio
n fr
om ta
rget
val
ue [%
]
187 112
209 197
Deviation from target value EQAS for HbA1c with lyophilized and fresh
whole blood samples
EQAS run
Jan 2014 Juli 2014 Okt 2014 May 2015
lyophilized samples
fresh samples
12
HbA1c EQAS results
January 2009 lyophilized samples
May 2015 fresh blood samples
Sam
ple
2 Hb
A1c
[mm
ol/m
ol]
Sam
ple
2 Hb
A1c
[mm
ol/m
ol]
Sample 1 HbA1c [mmol/mol] Sample 1 HbA1c [mmol/mol] acceptability range ± 18 % from target value
13
Participation in intercomparison studies
RELA – IFCC ring trials
IFCC HbA1c Network Laboratories
campaigns
14
Further activities
Commutability
IFCC Working Groups
WG on HbA2 standardisation
15
Further activities
Reference method for
cell counting
Reference method for
Total Hemoglobine
EMRP project „metalloproteins“
EMPIR call 2015
Coagulation factor VIII Replacement therapy 1950 - 2015
17
USA/Canada
OS Assay WHO 7. IS
Xyntha 1000 IU 131 μg
Europe
CS Assay WHO 7. IS
ReFacto AF 1000 IU 95 μg
Xyntha/ReFacto AF
Worldwide
CS Assay WHO 6. IS
ReFacto 1000 IU 83 μg
previous ReFacto
ReFacto AF vs. Xyntha
18
LC-MS/MS chromatogram of recomb. FVIII after triptic digestion
Perspectives
• High quality reference materials • Commutability • To convince the „real world“ of the traceability
concept
Thank you for your attention