BNL (Physics):BNL (Physics): B.Azmoun, A.Milov, R.Pisani, T.Sakaguchi, A.Sickles, S.Stoll, C.Woody B.Azmoun, A.Milov, R.Pisani, T.Sakaguchi, A.Sickles, S.Stoll, C.WoodyBNL (Instrumentation): J.Harder, P.O’Connor, V.Radeka, B.Yu J.Harder, P.O’Connor, V.Radeka, B.Yu Columbia Univ :Columbia Univ : C-Y. ChiC-Y. ChiSUNY: W.Anderson, A.Drees, Z. Citron, M.Durham, T.Hemmick, R.Hutter, B.Jacak, J.Kamin Weizmann:Weizmann: A.Dubey, Z.Fraenkel, A. Kozlov, M.Naglis, I.Ravinovich, D.Sharma, I.Tserruya A.Dubey, Z.Fraenkel, A. Kozlov, M.Naglis, I.Ravinovich, D.Sharma, I.Tserruya
Construction, Commissioning and Performance of a Hadron Blind Detector for
the PHENIX Experiment at RHIC
Itzhak TserruyaItzhak Tserruya IEEE-NSS, Hawaii, October 29, 2007IEEE-NSS, Hawaii, October 29, 2007 22
OutlineOutline
MotivationMotivation
ConceptConcept
ConstructionConstruction
Performance Performance
MotivationElectron pairs (or dileptons in general) are unique probes to study the matter formed in relativistic heavy ion collisions at RHIC:– best probe for chiral symmetry restoration and in-medium modifications of
light vector mesons , ω and – sensitive probe for thermal radiation:
QGP qqbar * e+e-
HG +- * e+e-
Experimental challenge:
huge combinatorial background arising from e+e- pairs from copiously produced from 0 Dalitz decay and conversions.
e+ e -
e+ e -
• Both members of the pair are needed to reconstruct a Dalitz decay or a conversion.
•Pair reconstruction limited by:– Low pT acceptance of outer PHENIX detector: ( p > 200MeV)– Limited geometrical acceptance of present PHENIX configuration
44
Upgrade Concept
Hardware* HBD in inner region* Inner coil (foreseen in original design) B0 for r 60cm Software * Identify electrons with p>200 MeV in PHENIX central arm detectors
* Match to HBD
* Reject if another electron is found in the HBD within opening angle < 200 mrad.
Strategy• Create a field free region close to the vertex to preserve opening angle of close pairs.
• Identify electrons in the field free region
• reject close pairs.
HBD ConceptHBD ConceptHBD concept:
♣♣ windowless Cherenkov detector (L=50cm) windowless Cherenkov detector (L=50cm)
♣♣ CFCF44 as radiator and detector gas as radiator and detector gas
♣♣ Proximity focus: Proximity focus:
detect circular blob not ringdetect circular blob not ring
50 cm
~1 cm
CF4 radiator
detector element
5 cmbeam axis
Why is it Hadron Blind? reverse bias between mesh and top GEM repels ionization charge away from multiplication area
Sensitive to UV and blind to traversing ionizing particles
E
hadronUV-photon
Detector element:Detector element:
♣♣ CsI reflective photocathodeCsI reflective photocathode
♣♣ Triple GEM with pad readoutTriple GEM with pad readout
66
All panels made of honeycomb & FR4 structure
Mylar window
Readout plane
Service panel
Triple GEM module with mesh grid
The Detector The Detector The detector fits under 3%X0 (vessel 0.92%, gas 0.54%, electronics ~1.5%) and it is leak tight to keep water out 0.12cc/min (~1 volume per year)!
Side panel
Sealing frame
HV terminals
Detector designed and built at the Weizmann Institute
FEEs Readout plane with 1152 hexagonal pads is made of Kapton in a single sheet to serve as gas seal
Each side has 12 (23x27cm2) triple GEM detectors stacks: Mesh electrode Top gold plated GEM for CsI Two standard GEMs pads
Two identical arms
Detector elements
GEM positioning elements are produced with 0.5mm mechanical tolerance.
Dead areas are minimized by stretching GEM foils on a 5mm frames and a support in the middle.
Detector construction involves ~350 gluing operations per box
Jig for box assembly
Itzhak TserruyaItzhak Tserruya IEEE-NSS, Hawaii, October 29, 2007IEEE-NSS, Hawaii, October 29, 2007 88
CsI evaporation and detector assembly in clean tent at Stony Brook”
CsI Evaporator and quantum efficiency
measurement(on loan from INFN)Can make up to 4 photocathodes in
one shot
6 men-post glove box, continuous gas
recirculation & heating
O2 < 5 ppmH2O < 10 ppm
Laminar Flow Table for GEM
assembly
High Vacuum GEM storage
Class 10-100 ( N < 0.5 mm particles/m3)
Detector assembly
99
HBD Engineering Run
The HBD was commissioned during the 2007 RHIC run.
After overcoming an initial HV problem, the detector operated smoothly at a gas gain of 3000-6000 for several months.
The CF4 recirculation gas system worked very smoothly. The oxygen and water content of the gas were monitored at the input as well as at the output of each vessel. In addition the gas transparency was monitored with a monochromator system. A reasonable transmittance of ~80% was achieved at a gas flow of 4 lpm.
The entire readout chain (both analog and digital) worked smoothly.
The excellent noise performance of the device (pedestal rms corresponding to 0.15 fC or 0.2 p.e. at a gain of 5000) allowed online implementation of a simple zero suppression algorithm to reduce the data volume.
A few billion minimum bias Au+Au collisions at √sNN = 200 GeV were collected and are presently being analyzed.
Typical noise performance
Itzhak TserruyaItzhak Tserruya IEEE-NSS, Hawaii, October 29, 2007IEEE-NSS, Hawaii, October 29, 2007 1010
Tracking & position resolution
Run 226502 ES4 at 3600V FB
Position resolution:
z ≈ ≈ 1 cm
Dictated by pad size: hexagon a = 1.55 cm(2a/√12 = 0.9 cm)
Hadrons selected in central arm:
Vertex +/- 20 cm< 50 tracks3 matching to PC3 and EMCaln0 < 0EMC energy < 0.5
Projected onto HBD:Z in HBD +/- 2 cm in HBD +/- 25 mrad
Itzhak TserruyaItzhak Tserruya IEEE-NSS, Hawaii, October 29, 2007IEEE-NSS, Hawaii, October 29, 2007 1111
Hadron Blindness & e-h separationHadron suppression illustrated by
comparing hadron spectra in FB and RB(same number of central tracks)
Pulse height
Hadron rejection factor
Pulse height
Electron - hadron separation (RB)
Strong suppression of hadron signal while
keeping efficient detection of photoelectrons
at reverse drift field
1212
Electron detection efficiency Identify e in central arm using RICH and EMCalIdentify e in central arm using RICH and EMCal Project central arm track to HBD Project central arm track to HBD Relative e-detection efficiency in HBD obtained by varying Relative e-detection efficiency in HBD obtained by varying
the charge threshold of the closest (matched) pad the charge threshold of the closest (matched) pad
EN3 G ≈ 3300 EN3 G ≈ 3300
(several runs at “nominal” voltage) (several runs at “nominal” voltage)
All modules <G> ≈ 6600 All modules <G> ≈ 6600
(Run 237092 “nominal” voltage + 100V)(Run 237092 “nominal” voltage + 100V)
~4 p.e. ~4 p.e.
Efficiency drop at pad threshold larger than about 4p.e. probably due to Efficiency drop at pad threshold larger than about 4p.e. probably due to electrons converted in the gas near the GEMs. Needs further study. electrons converted in the gas near the GEMs. Needs further study.
Itzhak TserruyaItzhak Tserruya IEEE-NSS, Hawaii, October 29, 2007IEEE-NSS, Hawaii, October 29, 2007 1313
SummarySummary
Low-mass eLow-mass e++ee-- pairs is a significant observable to diagnose the pairs is a significant observable to diagnose the matter formed at RHIC.matter formed at RHIC.
A novel HBD detector has been constructed and installed in the PHENIX set-up
A commissioning run took place in spring 2007
Preliminary analysis of data show:
Clear separation between e and h
Hadron rejection factor
Good electron detection efficiency