14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 1
Irradiation in KarlsruheIrradiation in Karlsruhe
• The Karlsruhe-Probestations
• Procedure of Irradiation
• Results
• Future annealing studies
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 2
The ProbestationThe Probestation
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 3
Cold measurementsCold measurements
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 4
The SoftwareThe Software
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 5
ProcedureProcedure• Measurements of sensors
– @ 20°C– @ rel. humidity < 30%– IV & CV– Ileak – Rpoly (2V) – CC (100Hz)
– Pinhole (10V) – Rint (0-0,5V) >10G
– Cint (1MHz)
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 6
ProcedureProcedure
• Bias connections for irradiation (conductive rubber & bonds)
• Irradiation– 34 MeV-protons– Fluence
• Annealing – @ 60 °C & 80 min– Mistake: annealing of sensors in the envelopes
• Same measurements as before irradiation, but– @ -10°C & dew point << -10°C (39°C)
21
21 14
,14
, 108,0103cm
nbarreloutereqcm
nbarrelinnereq
MeVMeV
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 7
Alpha @ -10°CAlpha @ -10°C
cmA
VI
C eq
C 1720 10420
??10
10
eq
C
VI
C 2010 20
10 C
C
II
C
From literature
cmA
CKK
CeTI
eTIC
KKBkeV
CTBkeV
C
CTBkeV
C
e 1820
2
293263
2010
102,12631
2931
24,1
2024,1
2200
1024,1
2100
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 8
Leakage current before and Leakage current before and after irradiationafter irradiation
• Guardring connected (for defined Volume)• Increase of leakage current as expected• Breakdown voltage >1200V
irradII
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 9
Depletion voltage before and Depletion voltage before and after irradiationafter irradiation
• Annealed @ 60°C and 80 min• But in the envelopes !
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 10
Leakage current after IrradiationLeakage current after Irradiation
Low FluenceHigh Fluence
Increase of leakage current depends on fluence as aspected
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 11
Rpoly after irradiationRpoly after irradiation
0V
1V
100V
No dependence on bias conditions !
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 12
CC after IrradiationCC after Irradiation0V 100V
1V No dependence on bias conditions !
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 13
Cint after irradiationCint after irradiation
1V
No increase of Cint seen
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 14
IQC-rampsIQC-ramps
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 15
Annealing curvesAnnealing curves
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 16
Annealing @ 50°C ?Annealing @ 50°C ?• Problem of annealing studies @ 60 °C
• high stability of temperature needed• Effects of heating up and cooling down become important
• Problem of annealing studies @ 40 °C• long time constants
T[°C] -10 -7 0 10 20 40 60 80
a 306d 180d 53d 10d 55h 4h 19min 2min
b 1/134 1/78 1/23 1/5 1 16 174 1490
T[°C] -10 0 10 20 40 60 80 100
y 516y 61y 8y 475d 17d 1260min 92min 9min
b 1/396 1/47 1/6 1/1 29 544 7430 76650
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 17
Time constants @ 50°CTime constants @ 50°C• Fit from values in the Moll-Thesis:
• Beneficial annealing: 65 min• Reverse annealing: 5162 min
Annealing steps: 15 min ?
14.3.2002 Irradiation Workshop Karlsruhe A. Furgeri 18
ConclusionsConclusions
• Measurements @ -10°C after irradiation
• Annealing without envelopes !
• No dependence of stripvalues from bias conditions
• Next irradiations without bias ?