columns IonPac AS15 Anion-Exchange Column Passion. Power. Productivity. 0 38 -0.3 μS Minutes 28425 Peak Concentration (μg/L) A B 1. Fluoride 0.018 0.48 2. Chloride 0.12 2.49 3. Nitrite 0.042 2.53 4. Carbonate 5. Sulfate 0.075 4.72 6. Bromide 2.36 7. Nitrate 0.15 2.58 8. Phosphate 2.15 1.6 1 3 4 6 7 8 2 5 A B The IonPac ® AS15 is a high-capacity, hydroxide-selective anion-exchange column designed for the determination of trace-level concentrations of inorganic anions and low molecular weight organic acids, including fluoride, glycolate, acetate, formate, chloride, nitrite, carbonate, sulfate, oxalate, bromide, nitrate, and phosphate in high-purity water matrices encountered in the semiconductor and power generation industries. The IonPac AS15 column offers excellent retention of fluoride from the water dip and uses a simple hydroxide eluent easily supplied by the Eluent Generator (EG). The IonPac AS15 column is available in formats from 0.4 to 4 mm, allowing use of capillary to analytical flow rates, and supporting advanced IC × IC applications. Superior Chromatographic Performance Universal column for trace-level concentrations of inorganic anions and low molecular weight organic acids in high-purity water matrices. The IonPac AS15-5µm (3 × 150 mm) column is recommended for fast separation of common inorganic anions and monovalent organic acids in 15 minutes. IonPac AS15-9µm (0.4 × 250, 2 × 250 and 4 × 250 mm) columns are recommended for higher capacity applications (25 minute analysis time). The AS15 Capillary column (0.4 × 250 mm) offers reduced eluent consumption, reduced operating costs, and excellent reproducibility. Column selectivity optimized for a 30 °C operating temperature to ensure reproducible retention times. Large-loop injections for ppb-level analysis. Sample preconcentration for ultra- trace (ppt-level) analysis using opti- mized AC15 concentrator column. Superior retention and resolution of early-eluting anions, including fluo- ride, glycolate, acetate, and formate. Simplified operation is provided by the EG, which requires only a deionized water source to produce potassium hydroxide eluent. Hydroxide eluent suppresses to a very low background for trace-level determinations. Compatible with organic solvents to enhance analyte solubility, to modify column selectivity, or for effective column cleanup. Direct transfer of 4 mm applications to the 0.4 mm, 3 mm, and 2 mm column formats for more eco- nomical operation, reduced eluent consumption, and increased mass sensitivity.

Transcript of IonPac AS15 Anion Exchange Column - Cromlab · 2015-10-01 · Trace Analysis is Easier with the...

Page 1: IonPac AS15 Anion Exchange Column - Cromlab · 2015-10-01 · Trace Analysis is Easier with the Eluent Generator Combining the IonPac AS15 column with the EG makes trace anion analysis

colu

mns IonPac AS15 Anion-Exchange Column

Passion. Power. Productivity.

0 38-0.3

µS

Minutes

28425

Peak Concentration (µg/L) A B1. Fluoride 0.018 0.48 2. Chloride 0.12 2.493. Nitrite 0.042 2.534. Carbonate — —5. Sulfate 0.075 4.726. Bromide — 2.367. Nitrate 0.15 2.588. Phosphate — 2.15

1.6

1

3

4 6 7

8

2

5

A

B

The IonPac® AS15 is a high-capacity, hydroxide-selective anion-exchange column designed for the determination of trace-level concentrations of inorganic anions and low molecular weight organic acids, including fluoride, glycolate, acetate, formate, chloride, nitrite, carbonate, sulfate, oxalate, bromide, nitrate, and phosphate in high-purity water matrices encountered in the semiconductor and power generation industries. The IonPac AS15 column offers excellent retention of fluoride from the water dip and uses a simple hydroxide eluent easily supplied by the Eluent Generator (EG).

The IonPac AS15 column is available in formats from 0.4 to 4 mm, allowing use of capillary to analytical flow rates, and supporting advanced IC × IC applications.

Superior Chromatographic Performance• Universalcolumnfortrace-level

concentrationsofinorganicanionsandlowmolecularweightorganicacidsinhigh-puritywatermatrices.

• TheIonPacAS15-5µm(3×150mm) columnisrecommendedforfastseparationofcommoninorganicanionsandmonovalentorganicacidsin15minutes.

• IonPacAS15-9µm(0.4×250, 2×250and4×250mm)columnsarerecommendedforhigher capacityapplications(25minuteanalysistime).

• TheAS15Capillarycolumn (0.4×250mm)offersreducedeluent consumption,reducedoperatingcosts,andexcellentreproducibility.

• Columnselectivityoptimizedfora30°Coperatingtemperaturetoensurereproducibleretentiontimes.

• Large-loopinjectionsforppb-levelanalysis.

• Samplepreconcentrationforultra-trace(ppt-level)analysisusingopti-mizedAC15concentratorcolumn.

• Superiorretentionandresolutionofearly-elutinganions,includingfluo-ride,glycolate,acetate,andformate.

• SimplifiedoperationisprovidedbytheEG,whichrequiresonlyadeionizedwatersourcetoproducepotassiumhydroxideeluent.

• Hydroxideeluentsuppressestoaverylowbackgroundfortrace-leveldeterminations.

• Compatiblewithorganicsolventstoenhanceanalytesolubility,tomodifycolumnselectivity,orforeffectivecolumncleanup.

• Directtransferof4mmapplicationstothe0.4mm,3mm,and2mm columnformatsformoreeco-nomicaloperation,reducedeluentconsumption,andincreasedmasssensitivity.

Anuta
New Stamp
Page 2: IonPac AS15 Anion Exchange Column - Cromlab · 2015-10-01 · Trace Analysis is Easier with the Eluent Generator Combining the IonPac AS15 column with the EG makes trace anion analysis

2

High-Efficiency Particle StructureTheIonPacAS15packingisa

uniquestructurecomposedofahighlycrosslinkedcoreandananion-exchangelayergraftedtothesurface,asshowninFigure1.TheAS15columnsubstrateiscomposedofamacroporousresinbead (5µmdiameterforthe3mmcolumnand9µmforthe0.4mm,4mm, and2mmcolumns)consistingof ethylvinylbenzenecrosslinkedwith55%divinylbenzene.

Theanion-exchangelayerisfunctionalizedwithalkanolquaternaryammoniumgroupsandhasacontrolledthickness,whichresultsinexcellentmasstransfercharacteristicsand consequently,highlyefficientpeaks.

Versatile CapabilityTheIonPacAS15columnis

idealfortheisocraticseparationoffluoride,chloride,carbonate,nitrite,sulfate,bromide,nitrate,andphosphateinhighpuritywatermatricesusingasimplehydroxideeluentandsuppressedconductivitydetection.Fluorideiswellresolvedfromthewaterdipwith elutionofhighlyretainedanalytes suchasphosphatein15minusingthe 3×150mmcolumnand<25min usingthe0.4×250mmor4×250mmcolumns,asillustratedinFigure2.

TheIonPacAS15-5µm (3×150mm)columnusesthesamealkanolammoniumfunctionalgroupsastheAS15-9µm(4×250mm)column,butdiffersincolumnformatandresinparticlesize.Asaresult,the3mmcolumnprovidestheadvantagesoffasteranalysistimeforthecommoninorganicanions,reducedeluentconsumption,andimprovedminimumdetectionlimitscomparedtothe4mmcolumn.TheAS15-5µm(3×150mm)columnisrecommendedforthefastseparationofthecommoninorganicanionsandlowmolecularweightorganicacids.

Figure 1. Structure of an IonPac AS15-9µm packing particle.

Figure 2. Comparison of the IonPac AS15-5µm (3 × 150 mm), AS15-9µm (4 × 250 mm), and AS15-9µm (0.4 × 250 mm) columns for separation of inorganic anions.

Anion ExchangePolymer layer

Highly Crosslinked Core

4.5m

13786

16122Minutes

10

0 5 10 15 20 25

0 5 10 15 20 25

0

12

3

45

6 7

1

2 45

6 7

8

8

B

C

2

8

765

41

0

10

3

A

3

Column: A) IonPac AG15, AS15-5µm (3 × 150 mm) B) IonPac AG15, AS15-9µm (4 × 250 mm) C) IonPac AG15, AS15 (0.4 × 250 mm)Eluent: A) 42 mM Potassium hydroxide B) 40 mM Potassium hydroxide C) 38 mM Potassium Hydroxide Eluent Source: EGC-KOHTemperature: 30 °CFlow Rate: A) 0.5 mL/min B) 1.2 mL/min C) 0.012 mL/minInj. Volume: A) 5 µL B) 10 µL C) 400 nLDetection: Suppressed conductivity, ASRS®-ULTRA, AutoSuppression® recycle mode

Peaks: 1. Fluoride 2 mg/L (ppm) 2. Chloride 5 3. Carbonate – 4. Nitrite 10 5. Sulfate 10 6. Bromide 20 7. Nitrate 20 8. Phosphate 30

7

0 10 20 30

0-1

µS

µS

µS

Thehigh-capacity IonPacAS15-9µm0.4×250mm, 4×250mmand2×250mmcolumnsarerecommendedforhighercapacityapplicationsandprovidemaximumresolutionoftracecomponentsinthepresenceofhighconcentrationsof otheranions.

Page 3: IonPac AS15 Anion Exchange Column - Cromlab · 2015-10-01 · Trace Analysis is Easier with the Eluent Generator Combining the IonPac AS15 column with the EG makes trace anion analysis

3

AS15 Capillary FormatTheIonPacAS15Capillary

column(0.4×250mm)ispackedwiththesamematerialastheequivalentstandardboreversion(producingthesameperformanceasa4mmcolumn),butrequiresonly1/100ththeeluentflowrate.Thecapillaryformatofferstheadvantageoflesseluentconsumptionprovidingreducedoperatingcosts.Figure3illustratestheseparationoftracelevelinorganicanionsusingtheIonPacAS15capillarycolumnafterconcentrationontheIonSwift®MAC-100Concentratorcolumn.Excellentretentiontimereproducibilitycanbeachievedwith thecapillaryformat.

Large Loop Injections for µg/L (ppb) Level Analysis

TheIonPacAS15columnselectivityandhighcapacityallowthedeterminationoftracelevelsofinorganicanionsandlowmolecularweightorganicacidsinhigh-puritywatermatricesusingalarge-loopinjection.Figure4illustratestheseparationofinorganicanionsandlowmolecularweightorganicacidsinahigh-puritywatersampleusingthistechnique.Separationwasachieved usingapotassiumhydroxidegradientwithsuppressedconductivitydetectionatatemperatureof30°C.Lowµg/L(ppb)levelsoftheseanalytescaneasilybequantifiedusinga1mLinjectiononanIonPacAS15-5µm(3×150mm) column.Theconductivityofthehydroxideeluentcanbesuppressedto averylowbackground,facilitatingtrace-levelanalysis.

Figure 4. Determination of inorganic anions and low molecular weight organic acids in high-purity water using a large-loop injection and a potassium hydroxide gradient on an IonPac AS15-5µm (3 × 150 mm) column.

284250 38Minutes

1.6

µS

-0.3

Column: IonPac AS15-9µm (0.4 × 250 mm)Eluent Source: Capillary EGC-KOH cartridge Eluent: 7 mM KOH (0 to 10 minutes), 7 to 32 mM KOH (10 to 16 minutes), 32 to 50 mM KOH (16 to 30 minutes), 50 to 65 mM (30 to 33 minutes), 7 mM KOH (33 to 38 minutes) Flow Rate: 12 µL/minTemperature: 30 °CDetection: Suppressed conductivity, ACES 300, Anion Capillary Electrolytic Suppressor, AutoSuppression recycle modeConcentrator: IonSwift MAC-100 Concentrator (0.5 mm × 80 mm)Inj. Volume: 180 µL Sample: A: Deionized water B: Deionized water spiked with 0.5 µg/L Fluoride, 2.5 µg/L chloride, nitrite, bromide, nitrate, and phosphate, 5.0 µg/L sulfate

Peak Concentration (µg/L) A B1. Fluoride 0.018 0.48 2. Chloride 0.12 2.493. Nitrite 0.042 2.534. Carbonate — —5. Sulfate 0.075 4.726. Bromide — 2.367. Nitrate 0.15 2.588. Phosphate — 2.15

1

3

4 6

7 8

2

5

A

B

Figure 3. Separation of inorganic anions at trace concentrations using the IonPac AS15 Capillary column and the IonSwift MAC-100 Concentrator column.

0 2 4 6 8 10 12 14Minutes

2.0

µS

Column: IonPac AG15, AS15-5µm (3 × 150 mm)Eluent : 10 mM Potassium hydroxide for 4 min to 60 mM at 11 min Eluent Source: EG40Temperature: 30 °CFlow Rate: 0.7 mL/minInj. Volume : 1000 µLDetection: Suppressed conductivity, ASRS-ULTRA 2-mm, AutoSuppression recycle mode

Peaks: 1. Fluoride 1 µg/L (ppb) 2. Glycolate 4 3. Acetate 4 4. Formate 2 5. Chloride 2 6. Nitrite 2 7. Carbonate – 8. Sulfate 2 9. Oxalate 2 10. Bromide 4 11. Nitrate 4 12. Phosphate 6

0

2 46

810

121

3 5

7

9 11

16395

Page 4: IonPac AS15 Anion Exchange Column - Cromlab · 2015-10-01 · Trace Analysis is Easier with the Eluent Generator Combining the IonPac AS15 column with the EG makes trace anion analysis

4

Figure 6. Determination of trace levels of chloride and sulfate in high purity water using a large loop injection on the IonPac AS15-5µm (3 × 150 mm) column.

AC15 Concentrator Column for Preconcentration for Ultratrace (ng/L) Analysis

Monitoringanioniccontaminantsatlowng/L(ppt)levelsinultrapurewaterisacriticalissueinthesemicon-ductorandpowergenerationindustries.Usingpreconcentrationtechniques,low-pptlevelsofinorganicanionsandlowmolecularweightorganicacidscanbedeterminedeasily.Figure5illustrates thistechniqueusinganIonPacAC15concentratorcolumn,whichisdesignedforreproduciblerecoveriesoftraceanionssuchaschlorideandsulfate fromhigh-puritywater.Usingsamplepreconcentration,low-pptlevelsof inorganicanionsandlowmolecularweightorganicacidscanberoutinelymeasuredinpowerplantandsemicon-ductorpurewater.

Trace Analysis is Easier with the Eluent Generator

CombiningtheIonPacAS15 columnwiththeEGmakestraceanionanalysiseveneasier.TheEGmakes potassiumhydroxideeluentelectrolyti-callyfromwater.Theuseofthisultra-purehydroxideasaneluentresultsinalowbackgroundandstablebaseline.Routineppb-levelanalysisofinorganicanionsandlowmolecularweightorgan-icacidsbydirectinjectionispossiblewiththeIonPacAS15columnusing theEG,asillustratedinFigures4and5.

Trace Analysis of Inorganic Anions and Low Molecular Weight Organic Acids in High Ionic Strength Matrices

Highcapacitycolumnsallowlargeloopinjectionswhilemaintainingthepeakshapesoftracecomponents,eveninthepresenceofhighconcentrationsofotheranions(e.g.carbonate)thatmaybepresentinthehigh-puritywatermatricesencounteredinthesemiconductorandpowergenerationindustries.

Figure 5. Determination of inorganic anions and low molecular weight organic acids from a 20-mL sample volume of high-purity water using the IonPac AS15-9µm (2 × 250 mm) column.

Figure6showstheuseofthe IonPacAS15-5µm(3×150mm) columnfortracechlorideandsulfateinahigh-puritywatermatrixcontaininghighconcentrationsofcarbonate.Usinganisocraticpotassiumhydroxideeluent,carbonateiswellresolvedfromchlorideandsulfateanddoesnotinterferewithquantification.Forhigh-puritywater applicationsthatrequiredetermination ofafewspecificanalytes,afastrunusingtheAS15-5µm(3×150mm)columnisrecommended.

Forapplicationsrequiringhighercapacity,theAS15-9µmcolumnsarerecommended.The0.4mm,4mm, and2mmIonPacAS15-9µmcolumnshaveveryhighcapacities,providing excellentresolutionofinorganicanionsincomplexsamplematrices.Figure7 comparestheseparationofinorganicanionsinawastewatersampleontheIonPacAS15-5µm(3×150mm)andAS15-9µm(4×250mm)columns.

1

23

4

5

789 10

11

4.0

µS

0

0 10 20Minutes

5 15

6

14871

Sample Volume: 20 mL, preconcentrated Detection: Suppressed conductivity, ASRS-ULTRA 2-mm, AutoSuppression® recycle mode

Peaks: 1. Fluoride 0.1 µg/L (ppb) 2. Acetate 0.1 3. Formate 0.1 4. Chloride 0.1 5. Nitrite 0.1 6. Carbonate – 7. Sulfate 0.1 8. Oxalate 0.1 9. Bromide 0.1 10. Nitrate 0.1 11. Phosphate 0.1

Column: IonPac AG15, AS15-9µm (2 × 250 mm)ConcentratorColumn: IonPac AC15Eluent: Potassium hydroxide, 10 mM from 0 to 4 min, 10 mM to 40 mM from 4 min to 14 min, 40 mM to 60 mM from 14 to 18 minEluent Source: EG40Temperature: 30 °CFlow Rate: 0.5 mL/min

0 5Minutes

3.0

µS

10

0

16394

1

2

3

Column: IonPac AG15, AS15-5µm (3 × 150 mm)Eluent: 42 mM Potassium hydroxideEluent Source: EG40Temperature: 30 °CFlow Rate: 0.5 mL/min

Inj. Volume: 1000 µL Detection: Suppressed conductivity, ASRS-ULTRA 2-mm, AutoSuppression recycle mode

Peaks: 1. Chloride 1 µg/L (ppb) 2. Carbonate 1000 3. Sulfate 1

Page 5: IonPac AS15 Anion Exchange Column - Cromlab · 2015-10-01 · Trace Analysis is Easier with the Eluent Generator Combining the IonPac AS15 column with the EG makes trace anion analysis

5

Figure 7. Analysis of an industrial wastewater sample on the IonPac AS15-5µm (3 × 150 mm) and AS15-9µm (4 × 250 mm) columns.

Figure 8. Separation of weakly retained organic acids using a potassium hydroxide gradient on the IonPac AS15-5µm (3 × 150 mm) column.

Determination of Inorganic Anions and Low Molecular Weight Organic Acids

Lowmolecularweightorganicacidsandmono-anddivalentinorganicanionscommonlyencounteredinthechemical,semiconductor,andpowergenerationindustriescanbedeterminedinasinglerunusingtheIonPacAS15-5µm (3×150mm)column.Figure8 illustratestheseparationofweaklyretainedanionssuchasacetate,formate,propionate,andbutyrateinasimulatedindustrialchemicalsampleusinga hydroxidegradientatacontrolled temperatureof30°C.

System RecommendationsTheIonPacAS15columnis

recommendedforusewithDionex ICS-2100orICS-5000ICsystemsequippedwithanEG.TheAS15columncanalsobeusedwitholderDionexICsystemsequippedwithanEGoranRFC-30Reagent-Free®Controller.TheEGisusedtoautomaticallyproducepotassiumhydroxidegradientsfromdeionizedwater.

Suppressor RecommendationsForoptimumeaseofuseand

economy,usetheIonPacAS15columnwiththeASRS®300AnionSelf-Regen-eratingSuppressor®orACES™ Anion CapillaryElectrolyticSuppressor300.OperatetheAS15atanelevated temperature(30°C)toensure reproducibleretentiontimes.

Anion Trap ColumnsWhenusingtheEGforeluent

delivery,aCR-ATCcolumnshouldbeinstalledbetweentheEGcartridgeandtheEGdegasmodule.Alternately,4mm,3mm,and2mmsystemscanusetheATC-HCcolumn,installedbetweenthepumpoutletandinletoftheEluGencartridgeintheEGmodule.Whenperformingsodiumhydroxide-gradientapplicationsontheIonPacAS15 columnusinghand-preparedbottledeluents,theATC-3shouldbeinstalledbetweenthegradientpumpandtheinjectionvalvetoremoveanionic contaminantsfromtheeluent.

0

0

3.0

µS 45

0 5 10 15 20 25 30Minutes

0

3.0

µS

2

45

16278

3

21

13

Column: A) IonPac AG15, AS15-5µm (3 × 150 mm) B) IonPac AG15, AS15-9µm (4 × 250 mm)Eluent: A) 10 mM Potassium hydroxide for 4 min to 60 mM at 9 min B) 8 mM Potassium hydroxide for 6 min to 55 mM at 18 minEluent Source: EG40Temperature: 30 °CFlow Rate: A) 0.7 mL/min B) 1.6 mL/minSample Prep.: Industrial wastewater, 1:10 dilutionInj. Volume: A) 5 µL B) 10 µLDetection: Suppressed conductivity, ASRS-ULTRA, AutoSuppression recycle mode

Peaks: 1. Chloride 2. Carbonate 3. Sulfate 4. Bromide 5. Chlorate

A

B

0 2 4 6 8 10 12 14 16

Minutes

4.0

µS

0

1

2

3

4

5

6

7

89

10

1112

1314

15

16

17

Peaks: 1. Fluoride 1 mg/L (ppm) 2. Glycolate 5 3. Acetate 5 4. Formate 5 5. Propionate 5 6. Chloride 2 7. Nitrite 2 8. Butyrate 5 9. Carbonate 30 10. Sulfite 2 11. Sulfate 2 12. Oxalate 3 13. Bromide 5 14. Nitrate 5 15. Phosphate 10 16. Chlorate 10 17. Thiosulfate 10

Column: IonPac AG15, AS15-5µm (3 × 150 mm)Eluent: 10 mM Potassium hydroxide for 4 min to 60 mM at 9 minEluent Source: EG40Temperature: 30 °CFlow Rate: 0.7 mL/minInj. Volume: 5 µL Detection: Suppressed conductivity, ASRS-ULTRA 2-mm, AutoSuppression recycle mode

16393

Concentrator ColumnsForconcentratorwork,whena

singlepistonpumpsuchastheAXPpump(pulsedamperrequired)isusedforsampledelivery,usetheIonPacAG15Guardcolumn,ultratraceanionconcentratorcolumns(UTAC-LP1,UTAC-ULP1,UTAC-XLP1,UTAC-LP2, UTAC-ULP2,orUTAC-XLP2),or thetraceanionconcentratorcolumns

(TAC-ULP1orTAC-2).UsetheUTAC-LP1,UTAC-LP2orthe TAC-LP1whenthesampleisdeliveredwithasyringeorwithalow-pressureautosamplersuchastheAS-DV.

Forconcentratorworkwith a0.4mmcapillarycolumn,usetheIonPacAG15CapillaryGuard columnortheIonSwiftMAC-100 Concentratorcolumn.

Page 6: IonPac AS15 Anion Exchange Column - Cromlab · 2015-10-01 · Trace Analysis is Easier with the Eluent Generator Combining the IonPac AS15 column with the EG makes trace anion analysis

6

SPECIFICATIONS

Dimensions: IonPacAS15CapillaryColumn: 0.4×250mm

IonPacAG15CapillaryGuardColumn: 0.4×50mm

IonPacAS15AnalyticalColumn: 2×250,3×150,and4×250mm

IonPacAG15GuardColumn: 2×50,3×30,and4×50mm

IonPacAC15ConcentratorColumn: 2×50and4×50mm

Maximum Operating Pressure: 4000psi(StandardorMicrobore) 5000psi(Capillary)

Mobile Phase Compatibility: pH0–14;0–100%HPLCsolvents

Substrate Characteristics: BeadDiameter: 5µm(3×150mm)

9µm(0.4×250,2×250and 4 × 250 mm)

13µm(AC15)

PoreSize: 100Å(AS15,AG15) Microporous,<10Å(AC15) Crosslinking(%DVB):55%

Latex Characteristics (AC15): LatexCrosslinking:6% LatexDiameter:85nm

Ion-Exchange Group: Surface-functionalizedwithalkanol quaternaryammoniumion(AS15, AG15)Latex-functionalizedwith alkanolquaternaryammoniumion (AC15)

Functional Group Characteristics: Medium–highhydrophobicity(AS15,AG15) Verylowhydrophobicity(AC15)

Capacity: 2.25µeq(0.4×250mmcapillarycolumn) 0.45µeq(0.4×50mmcapillaryguardcolumn) 56.25µeq(2×250mmanalyticalcolumn) 11.25µeq(2×50mmguardcolumn) 2.2µeq(2×50mmconcentratorcolumn) 70µeq(3×150mmanalyticalcolumn) 14µeq(3×30mmguardcolumn) 225µeq(4×250mmanalyticalcolumn) 45µeq(4×50mmguardcolumn) 9µeq(4×50mmconcentratorcolumn)

AC15 Void Volume: 9µL(2×50mm) 210µL(4×50mm)

Column Construction:PEEK®with10–32threadedferrule-styleend fittings.Allcomponentsarenonmetallic.

Page 7: IonPac AS15 Anion Exchange Column - Cromlab · 2015-10-01 · Trace Analysis is Easier with the Eluent Generator Combining the IonPac AS15 column with the EG makes trace anion analysis

ORDERING INFORMATION

IntheU.S.,call(800)346-6390orcontacttheDionexRegionalOfficenearestyou. OutsidetheU.S.,orderthroughyourlocalDionexofficeordistributor.Referthefollowingpartnumbers:

AS15 Analytical and Guard Columns Part Number

IonPacAS15-9µmCapillaryColumn(0.4×250mm)..............................................................................................................................075662

IonPacAG15-9µmGuardColumn(0.4×50mm)....................................................................................................................................075663

IonPacAS15-9µmAnalyticalColumn(4×250mm)...............................................................................................................................053940

IonPacAG15-9µmGuardColumn(4×50mm).......................................................................................................................................053942

IonPacAS15-5µmAnalyticalColumn(3×150mm)...............................................................................................................................057594

IonPacAG15-5µmGuardColumn(3×30mm).......................................................................................................................................057597

IonPacAS15-9µmAnalyticalColumn(2×250mm)...............................................................................................................................053941

IonPacAG15-9µmGuardColumn(2×50mm).......................................................................................................................................053943

Trap Columns

CR-ATCContinuouslyRegeneratedAnionTrapColumn(forusewithsystemsequippedwithaneluentgeneratororRFC-30Reagent-Free

Controller)..................................................................................................................................................................................................060477

CR-ATCContinuouslyRegeneratedAnionTrapColumn(Capillary)(forusewithCapillaryAnionColumns).....................................072078

ATC-3AnionTrapColumn(9×24mm)(forusewith4mmcolumns)...................................................................................................059660

ATC-3AnionTrapColumn(4×35mm)(forusewith2mmcolumns)...................................................................................................059661

ATC-HCAnionTrapColumn(forusewiththeEG).................................................................................................................................059604

Trace Anion Concentrator Columns

IonPacAC15ConcentratorColumn(2×50mm;forusewith2-mmand3-mmcolumns)......................................................................055695

IonPacAC15ConcentratorColumn(4×50mm).....................................................................................................................................055694

IonSwiftMAC-100MonolithAnionConcentratorColumn(0.5×80mm)..............................................................................................074702

TAC-LP1TraceAnionConcentrator(4×35mm)....................................................................................................................................046026

TAC-ULP1TraceAnionConcentrator(5×23mm).................................................................................................................................061400

UTAC-LP1UltraTraceAnionConcentratorLowPressure(4×35mm)................................................................................................063079

UTAC-ULP1UltraTraceAnionConcentratorUltraLowPressure(5×23mm).....................................................................................063475

UTAC-XLP1UltraTraceAnionConcentratorExtremelyLowPressure(6×16mm).............................................................................063459

UTAC-LP2UltraTraceAnionConcentratorLowPressure(4×35mm).................................................................................................072779

UTAC-ULP2UltraTraceAnionConcentratorUltraLowPressure(5×23mm).....................................................................................072780

UTAC-XLP2UltraTraceAnionConcentratorExtremelyLowPressure(6×16mm).............................................................................072781

ACES, ASRS, AutoSuppression, IonPac, IonSwift, and Reagent-Free are registered trademarks of Dionex Corporation. PEEK is a trademark of Victrex PLC.

North America

U.S./Canada (847) 295-7500 South America

Brazil (55) 11 3731 5140

Europe

Austria (43) 1 616 51 25 Benelux (31) 20 683 9768 (32) 3 353 4294 Denmark (45) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 Ireland (353) 1 644 0064 Italy (39) 02 51 62 1267 Sweden (46) 8 473 3380 Switzerland (41) 62 205 9966 United Kingdom (44) 1276 691722

Asia Pacific

Australia (61) 2 9420 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 2653 2580 Singapore (65) 6289 1190Taiwan (886) 2 8751 6655

Dionex Corporation

1228 Titan Way P.O. Box 3603 Sunnyvale, CA 94088-3603 (408) 737-0700 www.dionex.com

LPN 1000-05 PDF 3/11©2011 Dionex Corporation

pixie.hranilovich
Text Box
079970
pixie.hranilovich
Text Box
079932
pixie.hranilovich
Text Box
079917
pixie.hranilovich
Text Box
079918