ionic Surfactants by Conductometry -...
Transcript of ionic Surfactants by Conductometry -...
Abstract
A conductivity method for CMC determination of non
copolymer using C
alkyl chain length or anion on IL are optimized in such a way that
should be negligible effect on CMC values of non ionic surfactants
composed of different hydrophobic part and common polyethylene oxide
block. The obtained CMC values are compared with the CMC values
evaluated using classical conventional method such as s
Ionic Liquid as a Probe for the
Determination of C
ionic
Chapter 6
A conductivity method for CMC determination of non ionic surfactant/block
copolymer using C4PyCl is proposed in this Chapter. The concentration and
alkyl chain length or anion on IL are optimized in such a way that
should be negligible effect on CMC values of non ionic surfactants
composed of different hydrophobic part and common polyethylene oxide
block. The obtained CMC values are compared with the CMC values
evaluated using classical conventional method such as surface tension.
Ionic Liquid as a Probe for the
termination of CMC of Non
ionic Surfactants by
Conductometry
ionic surfactant/block
in this Chapter. The concentration and
alkyl chain length or anion on IL are optimized in such a way that there
should be negligible effect on CMC values of non ionic surfactants
composed of different hydrophobic part and common polyethylene oxide
block. The obtained CMC values are compared with the CMC values
urface tension.
Ionic Liquid as a Probe for the
of Non-
Surfactants by
Chapter 6
197
6.1 Introduction
Surfactants are amphiphilic molecules that contain a non polar segment, commonly called
“the tail” and polar segment called “the head” also called surface active agents. Because
of this amphiphilic character, they undergo aggregation. At low concentrations the
molecules exist as on individual entities, and at high concentration the molecules tends to
associate to from aggregates. Due to their unique properties and versatile availability,
surfactants have applications in many areas, including chemistry, biology, pharmacy,
material science etc.1,2
In polar solvent like water, this dual character of the amphiphilic leads to self-association
or micellization, micelles where the hydrophobic tails of the surfactant associate, leaving
the head group (hydrophilic) exposed to the solvent (Figure 6.1).
Figure 6.1 Schematic representations of block copolymer monomer and micelle
The hydrophobic tails of the aggregates from the ‘cone’ of the micelle, while the polar
head groups are located at the micelle-water interface in contact with and hydrated by a
number of water molecules. The transition form a monomeric solution to an aggregated
form can be seen as a change in the slope of plots against surfactants concentration of
many physical properties such as viscosity, conductance, surface tension etc.3-8 The
concentration above which micelles form (the change in physical property take place) is
called the critical micelle concentration (CMC). Once the micelles are formed, further
increase of the surfactant concentration does not significantly change the concentration of
the free monomer, above CMC, monomers and micelles exist in dynamic equilibrium.
Depending upon the chemical structure of the surfactant, its micelle can be cationic,
anionic, zwitterionic or . There are various methods for CMC determination
Chapter 6
198
based on the change in physical properties at micelle formation.3-8 The methods listed in
Table includes classical as well as some newly developed methods based on specific
instruments along with the properties or parameters obtained during these measurements.
Table 6.1 shows a list of methods used for CMC determination for all kind of
surfactants/amphiphilic block copolymer. These techniques are mainly classified into two
classes; (i) direct method where change in physical properties as a function of
concentration are measured (surface tension, density, conductivity, ultrasonic etc.), and
(ii) indirect method where probes or tracers are used and changes in their properties are
measured for determination of CMC (dye solubilization, fluorescence etc.).
Table 6.1 Various methods for the determination of CMC along with equipments and information deduced from each methods
Sr. No.
Methods Instrument Required
Physical Parameters
Information deduced
Applicable for surfactant
Ref.
1. Surface tension Tensiometer Surface tensing
CMC, γcmc, πcmc, G°mic, τmax, a
sm, pC20
Non-ionic, Ionic
9
2 Light Scattering SLS/DLS Light intensity
CMC, effective charge,
aggregation number
Non-ionic, Ionic
10
3 Conductivity Conductometer Conductivity CMC Ionic 11,12
4 Capillary electrophoresis
P/ACE 5000 HPCE electrophoresis
electrical current
CMC Ionic 13,14
5 Chromatography (LE-MEKC)
CE Instrumental setup, CE-971 UV-
Detector
Elution time CMC Ionic (Anionic)
15
6 Refractometry Ray light-Haber-Lowe type
Interferometer
Refractive indices
CMC Ionic, Non-ionic
16,42
7 Micro calorimetry
Micro calorimetry Heat capacity
CMC, ∆H, ∆S, ∆G
Ionic,
Non-ionic
17,18,
19,37
8 Density Densimeter Density CMC Ionic,
Non-ionic
20,21, 22
Chapter 6
199
Sr. No.
Methods Instrument Required
Physical Parameters
Information deduced
Applicable for surfactant
Ref.
9 Sound Velocity (Speed of sound)
Pulsecho-overlap technique
Speed of sound
CMC Ionic,
Non-ionic
23
10 Viscosity Viscometer Viscosity CMC Ionic, Non-ionic
24
11 Solubilization UV-visible spectrophotometer
Absorbance CMC, λmax Ionic, Non-ionic
25
12 NMR (Magnetic resonance method)
NMR δppm ∆G°m, ∆H°m, ∆S°m, CMC
Ionic, Non-ionic
26,42
13 Potentiometry Potentiometer with electrode
Emf CMC Ionic (Cationic)
27,36
14 Cyclic Volammetry
Cyclic Volameter (Three
compartment electrolytic cell,
SCE, Pt)
Potential, Current
CMC Non-ionic, Ionic
28,35
15 Polarography Sargent model XV Polarograph
Emf, id (diffusion
current)
CMC Ionic, Non-ionic
29,30
16 Fluorescence Shimadzy flurometer, Spectro
flurometer, Luminescence spectrometer
fluorescence intensity
(I/I3)
CMC Ionic, Non-ionic,
gemini surfactant
31,32,
42
17 Spectral method UV-spectrometer Absorbance CMC Non-ionic, Ionic
33,34
18 Equilibrium dialysis
Spectrophotometer Optical density
CMC Ionic 38
19 Kinetics Baunch & Lomb monochrometer
Time CMC 39
20 Impedance spectroscopy
technique
Schlumberger
(SI-1286)
Z', Z" Conductivity
CMC Ionic, Non-ionic,
Mixed (Ionic & Non-ionic)
40
21 SANS Neutron Scattering instrument
(SANS)
Scattering Intensity
Micellar shape and size,
Aggregation number, volume
fraction, poly dispersity
Ionic, Non-ionic
41
Chapter 6
200
In direct methods, surface tension (ST) and ionic conductivity are the most popular
methods. For CMC determination of surfactant solutions, the surface tension is the most
widely used technique since the surface tensiometer is easy to operate and the evaluation
of CMC from surface tension vs concentration curve is clear. Beside this, ST is suitable
for all kind of surfactant solution like ionic, nonionic, zwitterionic, gemini and mixed
surfactants. Along with ST, ionic conductivity is also extensively employed method for
determining the CMC, but it is strictly limited for ionic surfactant solution only. A clear
slope change which corresponds to CMC is observed in the specific conductivity (κ) vs
concentration curve. Conductance measurement has been employed for the evaluation of
CMC of gemini surfactants,43 and mixed surfactant systems,44 in aqueous and non
aqueous media.45 It is well known that conductance method is not appropriate for the
CMC determination of nonionic surfactants, but Svitova and Pletnev46 used this method
for the nonionic surfactant solutions and observed that a change in slope of straight line in
the specific conductance vs concentration diagram was not clear. The inflection point
becomes clear when an electrolyte is added to the solution of nonionic surfactants which
normally affects the CMC value of the same. Except the conductance method all the
major techniques used for CMC determination requires sophisticated professional
equipments.
On the other side, in indirect methods, dye solubilization and fluorescence
spectrophotometry are the most widely used methods. In case of dye solubilization
method, dyes, used as probe are added to the surfactant solutions and the change in λmax
of probe is measured as a function of surfactant concentration. But it is well documented
that, the presence of dye in the spectral change method causes a little lowering in CMC of
surfactants which may be explained in terms of induced micelle formation.47 Similarly, in
case of fluorescence spectrophotometry method, fluorescence probes (e.g. rhodamine B,
pyrene etc) have been used for CMC determination, which causes similar problem like in
dye solubilization method. Therefore, it is a necessity to develop an environment friendly
method in which use of a probe or tracer should have very negligible effect on the
micellization of nonionic surfactants.
Ionic liquids (ILs) which are considered as salts with a melting point below 100 °C, have
attracted considerable interest during the past decade due to their unique properties, such
as low vapor pressure, wide liquid range, good conductivity and large electrochemical
Chapter 6
201
window.48 Due to their unique features, they have become very much attractive in fields
such as catalysis, in basic electrochemical studies of organic compounds and inorganic
compounds, formation of metal nanostructures, analytical chemistry, including sensor,
bioanalytical chemistry and for electrochemical biosensors.49 This variety of applications
shows that ILs offers a high solvating, yet non-coordinating medium in which a number
of organic and inorganic solutes may be dissolved which make it an appropriate media to
work with. In the current understanding, a typical IL is composed of a bulky organic
cation and an inorganic or organic anion. The ability to tune the physicochemical
properties of ILs by changing the structure of the ions has led to their being called
“designer” solvents.50 The properties of both cation and anion are useful tools for fine-
tuning the properties of the resulting IL for desired properties. The choice of anion along
with alkyl chain length on cation decides the nature of ILs in terms of hydrophilic and
hydrophobic character which in turn predicts the physical properties of ILs which are
very important for various studies.
In view of this, we present a simple and easy approach for determining CMC values of
nonionic surfactants using IL as probe in conductivity method. The method is validated
for surfactants with different molecular weights having variety of hydrophobic part and
also different hydrophobic to hydrophilic ratio. The CMC values obtained by this method
were also compared with that obtained by other methods. SANS, DLS and 1H NMR are
used to understand the interaction between IL and surfactant as well as its effect on
micellar size and shape.
6.2 Experimental
The detailed specifications of all nonionic surfactants/block copolymers are given in
Table 6.2. The CMC measurements were examined at 30 oC ± 0.1 oC by testing the
nonionic surfactants, Triton X-100 (>99% purity, Fluka), Brij-98 (>99% purity, Sigma-
Aldrich), nonionic silicon surfactants (Degussa, Germany), Amphiphilic block
copolymers Pluronics® F127, P123, P105 (Sigma-Aldrich) and (EO-BO-EO) B – 1 and
B – 2 (Dow Corning Chemicals, USA). The ionic liquid C4PyCl (99.99+, Merck) was
used as a probe for the determination of CMC of all above mentioned surfactants. All the
chemicals were used as received. All surfactant solutions were prepared in de-ionized
water having conductivity ~ 5 × 10-6 S.cm-1 to which constant volume of C4PyCl was
Chapter 6
202
added in all solutions in such a way that the net concentration of C4PyCl becomes
15 mM. The conductivity of the aqueous surfactant solutions in presence of IL were
measured using digital conductometer (Equiptronics, India). The conductivity cell (cell
constant, 0.1 cm-1) was immersed in surfactant solutions which were kept in constant
temperature (~ 30 oC) water bath. The measurements were repeated three times for each
concentration of surfactants. The surface tensions of all solutions were also measured by
surface tension on tensiometer (Data Physics, Germany) at the same temperature. The
specific and relative viscosities for surfactants/polymers with and without IL were
measured using Ubbelhold Viscometer in constant temperature bath at 30 oC.
Table 6.2 Molecular formula, molecular weight and % of EO for various surfactants/block copolymers
Sr. No.
Surfactant/
Block copolymer
Formula Mol. wt.
gm.mol-1
% EO
1 P123 (EO)20(PO)70(EO)20 5750 30
2 P105 (EO)99(PO)69(EO)99 6500 50
3 F127 (EO)37(PO)56(EO)37 12600 70
4 B – 1 C4H9O-(BO)n-(EO)m-OH
m = 16, n = 9
1500 50
5 B – 2 HO-(EO)m-(BO)n-(EO)m-OH
m = 43, n = 14
5000 80
6 Triton X - 100 C8H17−C6H4−O(CH2CH2O)9.5H 625 80
7 Brij - 98 C18H35−(OCH2CH2)nOH, n~20 1149 80
8 SS – 1
For SS – 1 m = 5, n = 13, x = 12, y = 0
For SS – 2 m = 5, n = 20, x = 10, y = 4
4360 100
9 SS - 2 5600 75
EO = Ethylene Oxide, PO = Propylene Oxide, BO = Butylene Oxide
Chapter 6
203
SANS measurements were carried out on the surfactant solutions prepared in D2O at the
SANS facility at DHRUVA reactor, Trombay. The mean incident wavelength was 5.2 Å
with ∆λ/λ = 15%. The scattering was measured in the scattering vector (q) range of
0.017–0.3 Å-1. The measured SANS data were corrected for the background, empty cell
contributions, and transmission and were placed on an absolute scale using standard
protocols. The detailed data analysis were made using core-shell hard sphere model.51
DLS measurements were carried out in the H2O medium using a Malvern 4800 Autosizer
with 7132 digital correlator. The light source was an argon ion laser operated at 514.5 nm
with a maximum output power of 2 W. The average decay rate was obtained by analyzing
the electric field autocorrelation function, g1(τ) vs time data using a modified cumulants
method as has recently been proposed.52a The apparent equivalent hydrodynamic radii of
the micelles were calculated using the Stokes-Einstein relationship. All 1H NMR
experiments were conducted on a Bruker Avance 400 spectrometer at a frequency of
400.13 MHz in D2O and TMS as internal standard.
6.3 Result and Discussion
6.3.1 Concentration Optimization for Ionic Liquid
In the determination of CMC by conductometry method, selection of IL is crucial and we
have selected pyridinium based IL with alkyl chain having four carbon atoms. In recent
study carried out by us,52b where the effect of alkyl chain length on cationic head group in
CnPyCl (n = 4, 6, 8), effect of various cationic head groups and anions of various ILs on
the micellization of P123 in aqueous solutions was studied. Among all the variable
parameters, alkyl chain having four carbon atoms and pyridinium head groups have
almost negligible effect on the micelles of P123. In addition to this, C4PyCl IL was
unable to form stable aggregates while IL with C6 and C8 alkyl chain form stable
aggregates at room temperature.53 The concentration of IL is selected in such a way that it
falls very well below to the CMC value of ILs and it should have negligible effect on
morphology of micelles and less interaction with hydrophobic/hydrophilic blocks i.e. the
concentration of IL can only charge the micelles so that it has an impact of conductivity
in pre-micellar and micellar phase. In order to optimize the concentration of IL (C4PyCl),
we started with P123 (EO20-PO70-EO20) system which is well known for formation of
well organized spherical micelles.54
Chapter 6
204
The SANS and DLS methods were used to measure the effect of IL on morphology of
P123 micelles in aqueous solutions while NMR study gives an idea about the location as
well as interaction of IL with surfactants. Figure 6.2 represent the SANS curves for 15%
(w/w) of P123 solutions (well above from CMC value i.e. 0.015% (w/w)) in presence of
different concentrations of IL at 30 oC.
0.01 0.110-1
100
101
102
P123 P123 + 100 mM C4PyCl
dΣ/Σ/ Σ/Σ/
dΩ
, Ω
, Ω
, Ω
, cm
-1
q, Å-10.3
Figure 6.2 SANS spectra of 15% (w/w) P123 in D2O and 100 mM C4PyCl
Figure 6.2 reveals that upon addition of 100 mM IL, there was no appreciable change in
SANS profile which can be easily observed by no change in peak position as well as
intensity. The SANS curves were fitted to hard sphere core-shell model52 and the
parameters obtained from the fits are depicted in Table 6.3.
Table 6.3 Values of Core radius, RC, Shell Thickness, RS, Hard Sphere Radius, RHS,
Polydispersity, σ, Volume fraction, φ, Hydrodynamic Radius, Rh, for 15% (w/v) aqueous solutions of P123 in presence and absence of C4PyCl
System
Bkg
(cm-1)
RC
(Å)
RS
(Å)
RHS
(Å)
σσσσ
Volume fraction,
φφφφ
Rh
(Å)a
P123 + D2O 0.14 36.3 ± 0.2 6.8 ± 0.1 79.7 ± 0.3 0.49 0.152 94.5 ± 0.4
P123 + 50 mM C4PyCl
0.15 36.2 ± 0.2 6.7 ± 0.1 79.5 ± 0.3 0.49 0.153 94.1 ± 0.4
P123 + 100 mM C4PyCl
0.18 36.8 ± 0.2 6.7 ± 0.1 79.2 ± 0.3 0.48 0.157 93.1 ± 0.3
a Obtained from DLS measurements
Chapter 6
205
Table indicates that up to 100 mM concentration of IL, there is no change in size and
shape of P123 micelles (only little increase in background was found which is expected
due to the incoherent scattering from IL), which indicates that the IL has no prominent
interaction with PEO/PPO parts of P123. The same was also confirmed with the help of
DLS measurements. The plots of the intensity correlation function for 15% (w/w) P123 in
absence and presence of IL (50 mM) at a scattering angle of 90o is shown in Figure 6.3.
Figure reveals that the addition of 50mM IL was unable to make any shift in the
correlation function vs time plot which manifests pristine micellar morphology of P123.
Analysis of the same is carried out using the regularized non-negatively constrained
method, CONTIN55 which shows a unimodal distribution of relaxation rate. The
hydrodynamic radius (Rh) parameters obtained by this method are depicted in column 8 of
Table 6.3 and it clearly shows negligible effect of 50 mM IL concentration on the P123
solution. From the above measurements and reasonable discussion, we decided to take 15
mM of IL as an appropriate concentration for CMC determination using conductometry
method.
10 100
0.0
0.1
0.2
0.3
0.4
0.5
0.6
g1 , ττ ττ
Time, µµµµs
P123 P123 + 25 mM C4PyCl
600
Figure 6.3 Representative plot of the intensity correlation function for 50 mM C4PyCl with 15% (w/w) P123 at a scattering angle of 90°. The solid line is a fit to the data using the method of CONTIN
In order to know the interaction and position of ILs in micelles, the concentration
dependent 1H NMR spectra for P123 block copolymer in presence of 15 mM C4PyCl are
measured. (Figure 6.4). The concentration of block copolymers were selected in such a
way that they were below and above CMC values. From the 1H NMR spectra, the δppm
Chapter 6
values at 4.80, 3.61, 3.49, 1.24 corresponds to HDO, EO
respectively which are in good agreement with previous assignments
signals are attributed to IL.
Figure 6.4 1H NMR spectra of Pluronic P123 in presence of 15 mM CCMC value
The presence of different multiplets at lower concentration (below CMC) of P123 is
because of efficient motion of polymer
values at 4.80, 3.61, 3.49, 1.24 corresponds to HDO, EO –CH2–, PO
respectively which are in good agreement with previous assignments
signals are attributed to IL.53b
NMR spectra of Pluronic P123 in presence of 15 mM C
The presence of different multiplets at lower concentration (below CMC) of P123 is
because of efficient motion of polymeric chain which indicates that even in presence of
Below CMC
Above CMC
206
, PO –CH2–, PO –CH3
respectively which are in good agreement with previous assignments56 and the remaining
NMR spectra of Pluronic P123 in presence of 15 mM C4PyCl below and above
The presence of different multiplets at lower concentration (below CMC) of P123 is
ic chain which indicates that even in presence of
CMC
Above CMC
Chapter 6
207
IL, all segments of the solvated polymer can move freely. This has marked effect on the
conductivity of IL. When the concentration is increased above a certain value (CMC), the
chemical shifts of PO –CH3 and PO –CH2– were slightly shifted towards downfield while,
PO –CH– was slightly shifted towards upfield, which indicates that PPO domain of P123
micelle is not completely interacting with the ILs.56 In order to find out the position of IL
in micelles we have monitored the change in chemical shifts of terminal methyl (–CH3)
group of butyl chain as well as aromatic ring hydrogen of head group of IL (Figure 6.5).
Figure 6.5 Aggregation of Pluronic in presence of 15mM C4PyCl by 1H NMR Spectroscopy. δppm is the observed chemical shift (a) for aromatic proton (b) for the proton of the terminal methyl unit
δ (ppm)
Pure 15mM C4PyCl
Above CMC
(a) (b)
Chapter 6
208
Figure reveals, increase in the chemical shifts of both types of hydrogen. The observed
downfield shift of the terminal –CH3 of butyl chain protons is a manifestation of increase
in hydration i.e. it might not enter the apolar PPO core but reside at the interface of
PPO/PEO where it is in the influence of hydration. Moreover, the broadening of the peaks
after the CMC values indicates reduction in the mobility of –CH3 of butyl chain57 which
ultimately affects the conductivity of IL upon micellization.
6.3.2 CMC determination by Conductometry
As discussed in experimental section, all surfactant solutions were prepared in aqueous
solution of 15 mM C4PyCl and have specific conductivity 948 µScm-1 at 30 oC. The
conductivity of the C4PyCl solutions changes upon addition of nonionic surfactants. The
conductivity curves of the C4PyCl in the presence of the increasing amount of nonionic
surfactants were studied and found to be changing markedly like surface tension. The
specific conductivity vs concentration plots for the aqueous solution of surfactants
(having different hydrophobic groups) in presence of 15 mM C4PyCl are shown in
Figure 6.6. For the comparison purpose we have put surface tension data of the same
systems in the Figure 6.6. A perusal of figure reveals that a marked change in
conductivity of C4PyCl was noticed until the stable formation of micelles (i.e. complete
aggregation of surfactant molecules). The conductivity remains almost constant upon
further addition of surfactant above the CMC values. The concentration at which
conductivity attained the constant values is equal to CMC. The exact CMC values are
obtained by plotting the conductivity of the C4PyCl-surfactant mixtures vs concentration
of the surfactants. The obtained CMC values for different surfactants from conductivity
method along with surface tension and other methods are depicted in Table 6.4.
It can be clearly seen that the CMC values obtained by variation in conductivity of
C4PyCl is in good agreement with the literature data as well as obtained by surface
tension method. Unlike other additive methods where dyes or organic molecules were
used as probe, this method does not make any effect on the micellization processes. In
order to check versatility of this method, we have selected block copolymers/surfactants
in such a way that they have different hydrophobic moieties and along with this some
industrially important surfactants like Triton X-100, Brij-98 are also used. The change in
conductivity of the C4PyCl may be due to the restriction in the motion of IL.
Chapter 6
209
0.000 0.001 0.002 0.003 0.004 0.005 0.006 0.007
905
910
915
920
925
Conductivity Surfacetension
Conc. of Surfactant, g/dl
Sp
.Co
nd
uct
ance
, κ, κ,
κ,
κ, µµ µµ
S/c
m
P123
CMC
(a)
35
40
45
50
Su
rface tensio
n, γγ γγ, m
Nm
-1
0.00 0.01 0.02 0.03 0.04 0.05 0.06 0.07910
915
920
925
930
935
940
945
Conductivity Surface tension
Conc. of Surfactant, g/dl
Sp
.Co
nd
uct
ance
, κ, κ,
κ,
κ, µµ µµ
S/c
m
B - 2
CMC
(b)
30
35
40
45
50
55
Su
rface tensio
n, γγ γγ, m
Nm
-1
0.00 0.01 0.02 0.03 0.04 0.05 0.06
910
915
920
925
930
Conductivity Surfacetension
Conc. of Surfactant, g/dl
Sp
.Co
nd
uct
ance
, κ,κ, κ,κ, µµ µµ
S/c
m
CMC
30
35
40
45
50
55
Su
rface tensio
n, γγ γγ, m
Nm
-1TX-100 (c)
0.00 0.01 0.02 0.03 0.04 0.05
915
920
925
930
935
940
Conductivity Surface tension
Conc. of Surfactant, g/dl
Sp
.Co
nd
uct
ance
, κ, κ,
κ,
κ, µµ µµ
S/c
m
CMC
(d)
45
50
55
60
65
70
Su
rface tensio
n, γ,γ, γ,γ, m
Nm
-1
Brij-98
0.00 0.01 0.02 0.03 0.04 0.05 0.06910
915
920
925
930
935
940
Conductivity Surface tension
Conc. of Surfactant, g/dl
Sp
.co
nd
uct
ance
, κ, κ,
κ,
κ, µµ µµ
S/c
m
CMC
(e)
30
35
40
45
50
Su
rface tensio
n, γγ γγ, m
Nm
-1
SS - 1
Figure 6.6 Specific conductance vs. concentration of surfactants at 30 oC, (a) P123, (b) B – 2, (c) Triton X-100, (d) Brij – 98 and (e) SS – 1 (conductivity of 15 mM C4PyCl is 948 µS/cm2 in water)
As discussed in previous section, through 1H NMR study we found that the IL molecules
are located at hydrated PEO/PPO interface, which restrict the motion of IL molecules
which in turn decreases the conductivity of the IL.
Chapter 6
210
Table 6.4 Critical Micelle Concentration of surfactants and block copolymers obtained from Conductivity method and surface tension method at 30oC
Surfactants Critical Micelle Concentration (g.dl-1)
Surface tension Conductivity Lit. Ref.
P123 5.7±0.3 × 10-3 5.9±0.3 × 10-3 5.0 × 10-3 [a]
1.9 × 10-3[b]
60
61
P105 3.7±0.2 × 10-2 3.5±0.2 × 10-2 2.5 × 10-2 [a] 60
F127 4.0±0.2 × 10-2 3.9±0.2 × 10-2 3.9 × 10-2 [b] 61
B - 1 1.9±0.1 × 10-3 1.5 ±0.1× 10-3 2.2 × 10-3 [b] 62
B - 2 2.5±0.1 × 10-2 2.3±0.1 × 10-2 2.9 × 10-2 [b] 62
Triton X-100 2.6±0.1 × 10-2 2.5±0.1 × 10-2 1.4 × 10-2 [c]
1.8 × 10-2 [d]
63
64
Brij-98 2.5±0.1 × 10-2 2.4±0.1 × 10-2 3.0 × 10-2 [b] 65
SS - 1 4.2±0.2 × 10-2 3.8±0.2 × 10-2 4.2 × 10-2 [b] 66
SS - 2 1.1±0.1 × 10-2 1.1±0.1 × 10-2 2.5 × 10-2 [b]
@25 oC
67
a Dye solubilization, bSurface tension, c Kinetic approach, dCyclic voltammetry
It is very well documented that, Pluronic block copolymers in surface tension curves
shows two distinct break points58 and generally second break point is considered as the
CMC value. Alexandridis et. al.59 reasoned that Pluronic EPE copolymer molecules at
very low copolymer bulk concentration adsorbs at the air/water interface in an extended
conformation. And when the bulk concentration increases, the interface is covered fully
with the copolymer molecules, thus changing the extended conformation to an inverted
‘∩’ with two ends blocks contact with water and the middle hydrophobic block
protruding into air. This change in conformation would manifest as first break point in
surface tension-concentration plots. Other indirect methods like dye solubilization, cyclic
voltammetry, fluorescence spectroscopy etc. were insensitive to measure double break
points. Figure 6.6(a) and Figure 6.7(a, b) shows the formation of double break point in
Pluronic block copolymers.60,61 Figures reveal that the first break point observed by
conductivity method is also in good agreement with the one obtained by ST, which
indicates that the method is sensitive towards this kind of critical phenomenon. The
double break point in conductivity measurements can be explain as; the quaternary
ammonium ion of C4PyCl is known for making coordination bonds with etheric oxygen
of PEO and PPO.
Chapter 6
211
As mentioned above, at very low concentration of polymer, hydrophobic part will reside
at air/water interface thus it is less available in to bulk solution, which results into rare
chances of interaction of quaternary ions with lone pair of oxygen from PPO i.e. it is
paving the free mobility of ionic species of IL which leads to less change in conductivity.
At high concentration of polymer, due to the saturation of PPO molecules at air/water
interface, they are more available in bulk in addition to PEO and ions of IL can interact
with PPO which ultimately reduces the mobility and resulting into suppression of
conductivity. Therefore, we can say that our method is simple as well as sensitive towards
such kind of changes along with the determination of CMC values of nonionic surfactants
where sophisticated instrumentation is not required.
0.00 0.01 0.02 0.03 0.04 0.05 0.06
965
970
975
980
985
990
Conductivity Surface tension
Conc. of Surfactant, g/dl
Sp
.Co
nd
uct
ance
, κ,
κ,
κ,
κ, µµ µµ
S/c
m
P105
CMC
(a)
35
40
45
50
55
60
65
70
Su
rface tensio
n, γγ γγ, m
Nm
-1
0.00 0.01 0.02 0.03 0.04 0.05 0.06 0.07910
915
920
925
930
935
Conductivity Surface tension
Conc. of Surfactant, g/dl
Sp
.Co
nd
uct
ance
, κ,
κ,
κ,
κ, µµ µµ
S/c
m
F127
CMC
(b)
40
45
50
55
60
Su
rface tensio
n, γγ γγ, m
Nm
-1
0.00 0.01 0.02 0.03 0.04
915
920
925
930
Conductivity Surface tension
Conc. of Surfactant, g/dl
Sp
.Co
nd
uct
ance
, κ,κ, κ,κ, µµ µµ
S/c
m
CMC
(c)
20
30
40
50
60
70
Su
rface tensio
n, γγ γγ, m
Nm
-1
SS-2
0.000 0.001 0.002 0.003 0.004 0.005
910
915
920
925
930
935
Conductivity Surface tension
Conc. of Surfactant, g/dl
Sp
.Co
nd
uct
ance
, κ κ κ κ, µµ µµ
S/c
m
B-1 (d)
35
40
45
50
55
Su
rface tensio
n, γγ γγ, m
Nm
-1
CMC
Figure 6.7 CMC measurements from Conductivity and Surface tension Curve of (a) P105, (b) F127, (c) SS-2, (d) B-1 (conductivity of 15 mM C4PyCl is 948 µS/cm2 in water)
It is worth to mentioned that the viscosity effect on conductivity of ILs has not been
consider in the change of conductivity as a function of concentration of block
Chapter 6
copolymers/surfactants because the variation in viscosity of polymer with concentration is
not prominent.(Figure 6.8)
0 10 20 30
0.005
0.010
0.015
0.020
0.025
ηη ηη sp
/c
Conc. of C
Figure 6.8 Reduced viscosities of (a) 5% (w/w) P123 in various concentration of C(b) various concentration of P123 in water at 30
Figure 6.9 1H NMR spectra of pure P123 in D
copolymers/surfactants because the variation in viscosity of polymer with concentration is
(Figure 6.8)
40 50 60 70 80 90 100 110
Conc. of C4PyCl, mM
5% w/w P123 + C4PyCl
0 1 20.0
0.5
1.0
1.5
2.0
2.5
3.0
ηη ηηsp
/c
Conc. of P123, gm/dl
Reduced viscosities of (a) 5% (w/w) P123 in various concentration of C(b) various concentration of P123 in water at 30 oC
H NMR spectra of pure P123 in D2O at above CMC
212
copolymers/surfactants because the variation in viscosity of polymer with concentration is
3 4 5 6
Conc. of P123, gm/dl
P123 in water
Reduced viscosities of (a) 5% (w/w) P123 in various concentration of C4PyCl,
Chapter 6
213
Figure 6.10 1H NMR spectra of 150 mM pure IL (C4PyCl) in D2O
6.4 Conclusions
A simple and novel method has been devised for the determination of Critical Micelle
Concentration values for nonionic surfactants using conductometry where IL, C4PyCl was
used as a probe. Selection of IL and optimization of its concentration was carried out
using SANS and DLS measurements. 1H NMR measurements gave an insight into the
location as well as interaction of IL inside the micelles of surfactants. From these
experiments we have found that IL does not affect the morphology of micelles by making
strong interaction with hydrophobic or hydrophilic part of the surfactant. This method
was checked for various nonionic surfactants having different hydrophobic parts and
hydrophobic/hydrophilic ratio. The CMC values obtained by this method are in good
agreement with the one obtained by other versatile methods and its sensitivity was proved
by observing double break points in the Pluronic block copolymer system. Thus, we can
say that the conductivity method can be employed for the determination of CMC of
nonionic surfactants using IL as conducting probe.
Chapter 6
214
6.5 References
1. (a) Fendler, J. H. Membrane Mimetic Chemistry; Wiley Interscience: New York, 1982.
(b) Fendler, J. H.; Fendler, E. H. Catalysis in Micellar and Macromolecular Systems;
Academic: New York, 1975. (c) Bunton, C. A.; Nome, F.; Quina, F. H.; Romsted, L. S.
Acc. Chem. Res. 1991, 24, 357.
2. (a) Soni, S. S.; Henderson, M. J.; Bardeau, J. F.; Gibaud, A. Adv. Mater. 2008, 20, 1493.
(b) Kresge, C. T.; Leonowicz, M. E.; Roth, W. J.; Vartuli, J. C.; Beck, J. S. Nature 1992,
359, 710.
3. (a) Calvo, E.; Bravo, R.; Amigo, A.; Gracia-Fadrique, J. Fluid Phase Equilibr. 2009, 282,
14-19. (b) Edwards, D. A.; Luthy, R. G.; Liu, Z. Environ. Sci. Technol. 1991, 25, 127.
4. (a) Anna, J. J. Colloid Interf. Sci. 2010, 346, 398; (b) Sohrabi, B.; Bazyari, A.;
Hashemianzadeh, M. Colloid Surface A. 2010, 364, 87.
5. Frindi, M.; Michels, B.; Levy, H.; Zana, R. Langmuir 1994, 10, 1140.
6. (a) Paillet, S.; Grassl, B.; Desbrieres, J. Anal. Chim. Acta 2009, 636, 236. (b) Bedells, A.
D.; Arafeh, R. M.; Yang, Z.; Attwood, D.; Heatley, F.; Pedget, J. C.; Price, C.; Booth, C.
J. Chem. Soc., Faraday Trans. 1993, 89, 1235.
7. (a) Tan, C. H.; Huang, Z. J.; Huang, X. G. Anal. Biochem. 2010, 401, 144. (b) Ekwall, P.;
Mandell, L.; Solyom, P. J. Colloid Interf. Sci. 1971, 35, 519.
8. Hoffmann, H.; Platz, G.; Rehage, S. W.; Ulbricht, W. Ber. Bunsenges, Phys. Chem. 1981,
85, 255.
9. Chakraborty, T.; Ghosh, S. Colloid Polym. Sci. 2007, 285, 1665.
10. (a) Debye P. J. Phys. Chem. 1949, 53, 1. (b) Phillips J. N.; Mysels, K. J. J. Phys. Chem.
1955, 59, 325. (c) Mysels, K. J.; Princen, L. H. J. Phys. Chem. 1959, 63, 1699.
11. Garcia-Mateos, I.; Velazquez, M. M.; Rodriguez, L. J. Langmuir 1990, 6, 1078.
12. Voeks, J. F.; Tartar, H. V. J. Phys. Chem. 1955, 59, 1190.
13. Cifuentes, A.; Bernal, J. L.; Diez-Masa, J. C. Anal. Chem. 1997, 69, 4271.
14. Terabe, S. Anal. Chem. 1985, 57, 834.
15. Chen, Z.; Lin, J. M.; Uchiyama, K.; Hobo, T. Anal. Chim. Acta. 2000, 403, 173.
16. Klevens, H. B. J. Phys. Chem. 1948, 52, 130.
17. Stodghill, S. P.; Smith, A. E.; O’Haver, J. H. Langmuir 2004, 20, 11387.
18. Kiraly, Z.; Dekany, I. J. Colloid Interf. Sci. 2001, 242, 214.
19. Majhi, P. R.; Moulik, S. P. Langmuir 1998, 14, 3986.
20. Wright, K. A.; Tartar, H. V. J. Am. Chem. Soc. 1939, 61, 544.
21. Grindley, B. J.; Bury, C. R. J. Chem. Soc. 1929, 679.
22. Bury, C. R.; Parry, G. A. J. Chem. Soc. 1935, 626.
23. Junquera, E.; Tardajos, G.; Aicart, E. Langmuir 1993, 9,1213.
Chapter 6
215
24. Poskanzer, A. M.; Goodrich, F. C. J. Phys. Chem. 1975, 79, 2122.
25. Vulliez, Le.; Normand, B.; Eisele, J. L. Anal. Biochem. 1993, 208, 241.
26. Muller, N.; Pellerin, J. H.; Chen, W. W. J. Phys. Chem. 1972, 76, 3012.
27. Vojtekova, M.; Kopecky, F.; Greksakova, O.; Oremusova, J. Collect. Czech. Chem.
Commun. 1994, 59, 99.
28. Mandal A. B.; Nair, B. U.; Ramaswamy, D. Langmuir 1988, 4, 736.
29. Colichman, E. L. J. Am. Chem. Soc. 1950, 72, 4036.
30. Colichman, E. L. J. Am. Chem. Soc. 1951, 73, 1795.
31. De Vendittis, E.; Palumbo, G.; Parlato, G.; Bocchini,V. Anal. Biochem. 1981, 115, 278.
32. Nakahara, Y.; Kida, T.; Nakatsuji, Y.; Akashi, M. Langmuir 2005, 21, 6688.
33. Mukerjee, P.; Mysels, K. J. J. Am. Chem. Soc. 1955, 77, 2937.
34. Corrin, M. L.; Harkins, W. J. Am. Chem. Soc. 1947, 69, 683.
35. Ogita, M.; Nagai, Y.; Mehta, M. A. Fujinami, T. Sensors and Actuators B 2000, 64, 147.
36. Singh, C. D.; Shibata, Y.; Ogita, M. Sensors and Actuators B 2003, 96, 130.
37. Dearden, L. V.; Woolley, E. M. J. Phys. Chem. 1987, 91, 4123.
38. Yang, J. T.; Foster, J. F. J. Phys. Chem. 1953, 57, 628.
39. Perkowski, J.; Mayer, J.; Ledakowicz, S. Colloids and Surface A 1995, 101, 103.
40. Chang, H. C.; Hwang, B. J,; Lin, Y. Y.; Chen, L. J.; Lin, S. Y. Rev. Sci. Instrum. 1998,
69, 2514.
41. Bowers, J.; Butts, C. P.; Martin, P. J.; Vergara-Gutierrez, M. C.; Heenan, R. K. Langmuir
2004, 20, 2191.
42. Singh, T.; Kumar, A. J. Phys. Chem. B 2007, 111, 7843.
43. Kabiruddin, D.; Koya, P. A.; Khan, Z. A. J. Colloid Interface Sci. 2010, 342, 340.
44. Munoz, M.; Rodriguez, A.; Graciani, M. der Maria; Moya, M. L. Langmuir 2004, 20,
10858.
45. Flores, M. V.; Voutsas, E. C.; Spiliotis, N.; Eccleston, G. M.; Bell, G.; Tassios, D. P.;
Halling, P. J. J. Colloid Interface Sci. 2001, 240, 277.
46. Svitova, N. A.; Pletnev, M. Y. Colloid J. USSR, 1987, 49, 901.
47. Mukerjee, P.; Mysels, K. J. J. Am. Chem. Soc. 1955, 77, 2937.
48. Percus, J. K.; Yevick, G. J. Phys. Rev. 1958, 110, 1.
49. Antonietti, M.; Kuang, D.; Samershly, B.; Zhou, Y. Angew. Chem. Int. Ed. 2004, 43,
4988.
50. (a) Smiglak, M.; Metlen, A.; Rogers, R. D. Acc. Chem. Res. 2007, 40, 1182.
(b) Plechkova, N. V.; Seddon, K. R. Chem. Soc. Rev. 2008, 37, 123.
Chapter 6
216
51. (a) Freemantle, M. Chem. Eng. News 1998, 76, 32. (b) Plechkova, N. V.; Seddon, K. R.
In Methods and Reagents for Green Chemistry: An Introduction; Tundo, P.; Perosa, A.;
Zecchini, F. Eds.; John Wiley & Sons, Inc.: New York, 2007, pp 105.
52. (a) Hassan, P. A.; Kulshreshtha, S. K. J. Colloid Interface Sci. 2006, 300, 744.
(b) Vekariya, R. L.; Aswal, V. K.; Hassan, P. A.; Soni, S. S. submitted to Phys. Chem.
Chem. Phys. (unpublished work 2013).
53. (a) Jungnickel, C.; Luczak, J.; Ranke, J.; Fernandez, J. F.; Muller, A.; Thoming, J.
Colloid Surf., A. 2008, 316, 278. (b) Sastry N. V.; Vaghela, N. M.; Macwan, P. M.; Soni,
S. S.; Aswal, V. K.; Gibaud, A. J. Colloid Interface Sci. 2012, 371, 52.
54. Soni, S. S.; Brotons, G.; Bellour, M.; Narayanan, T.; Gibaud, A. J. Phys. Chem. B 2006,
110, 15157.
55. Stephen W, P. Comput. Phys. Commun. 1982, 27, 213; Comput. Phys. Commun. 1982,
27, 229.
56. Ma, J. H.; Guo, C.; Tang, Y. L.; Chen, L.; Bahadur, P.; Liu, H. Z. J. Phys. Chem. B 2007,
111, 5155.
57. Nivaggioli, T.; Tsao, B.; Alexandridis, P.; Hatton, T. A. Langmuir 1995, 11, 119.
58. Chu, B.; Zhou, Z. Nonionic Surfactants Nace, V. M. Ed., Surf. Sci. Ser. 60, Marcel
Dekker, New York, 1996, Ch.3, pp. 67-143.
59. Alexandridis, P.; Athanassiou, V.; Fukuda, S.; Hatton, T. A. Langmuir 1994, 10, 2604.
60. Alexandridis, P.; Holzwarth, J. F.; Hatton, T. A. Macromolecules 1994, 27, 2414.
61. Wanka, G.; Hoffmann, H.; Ulbricht, W. Macromolecules 1994, 27, 4145.
62. Soni, S. S.; Sastry, N. V.; George, J.; Bohidar, H. B. Langmuir 2003, 19, 4597.
63. Perkowski, J.; Mayer, J.; Ledakowicz, S. Colloid Surface A. 1995, 101, 103.
64. Mandal, A. B.; Nair, B. U.; Ramaswamy, D. Langmuir 1988, 4, 736.
65. Kapoor, Y.; Howell, B. A.; Chauhan, A. Invest. Ophth. Vis. Sci. 2009, 50, 2727.
66. Soni, S. S.; Panjabi, S. H.; Sastry, N. V. Colloid Surface A. 2011, 377, 205.
67. Soni, S. S.; Sastry, N. V.; George, J.; Bohidar, H. B. J. Phys. Chem. B 2003, 107, 5382.