Introduction to Variable Area Flowmeters

5
eFunda: Introduction to Variable Area Flowmeters Variable Area Flowmeter About Us | Directory | Career | News | Chat | InfoStore | Industrial | Ask an Expert Design Home Sensors Sensor Home Instruments/Devices Methods/Principles Displacement Stress & Strain Pressure Fluid Flow Flowmeter Introduction Categorization Selection Coriolis Differential Pressure Magnetic Positive Displacement Target Thermal Turbine Ultrasonic Transit Time Doppler Variable Area Vortex Temperature Resources Bibliography Login Free Magazines Adhesives & Sealants Control Design Metal Finishing Modern Plastics Ceramic Industry Desktop Engineering Tooling & Production more... Search for Home Membership Palm Store Forum Search Member Calculators Materials Design Processes Units Formulas Math Overview A Variable area flowmeter's cross section area available to the flow varies with the flow rate. Under a (nearly) constant pressure drop, the higher the volume flow rate, the higher the flow path area. Variable area flowmeters can further be categorized into three types Rotameter: The most widely used variable area flowmeter. Movable Vane Meter: http://www.efunda.com/DesignStandards/sensors/flowmeters/flowmeter_va.cfm (1 of 5)29/09/2007 15:51:03

description

flowmeters

Transcript of Introduction to Variable Area Flowmeters

  • eFunda: Introduction to Variable Area Flowmeters

    Variable Area Flowmeter About Us | Directory | Career | News | Chat | InfoStore | Industrial | Ask an Expert

    Design Home

    SensorsSensor HomeInstruments/DevicesMethods/PrinciplesDisplacementStress & StrainPressureFluid FlowFlowmeter Introduction

    Categorization

    Selection

    Coriolis

    Differential Pressure

    Magnetic

    Positive Displacement

    Target

    Thermal

    Turbine

    Ultrasonic

    Transit Time

    Doppler

    Variable Area Vortex

    Temperature

    ResourcesBibliography

    Login

    Free Magazines

    Adhesives & Sealants

    Control Design

    Metal Finishing

    Modern Plastics

    Ceramic Industry

    Desktop Engineering

    Tooling & Production

    more...

    Search for

    Home Membership Palm Store Forum Search Member Calculators

    Materials

    Design

    Processes

    Units

    Formulas

    Math

    Overview

    A Variable area flowmeter's cross section area available to the flow varies with the flow rate. Under a (nearly) constant pressure drop, the higher the volume flow rate, the higher the flow path area.

    Variable area flowmeters can further be categorized into three types

    l Rotameter: The most widely used variable area flowmeter.

    l Movable Vane Meter:

    http://www.efunda.com/DesignStandards/sensors/flowmeters/flowmeter_va.cfm (1 of 5)29/09/2007 15:51:03

  • eFunda: Introduction to Variable Area Flowmeters

    l Weir, Flume:

    Variable area flowmeters are robust, low cost, simple, however less accurate, compared to other more complicated flowmeters.

    Top of Page

    Further Information

    Rotameter is the most widely used variable area flowmeter. Its principle of operation is discussed below. A movable vane meter has similar principle of operation, except the "moving piston (float)" of rotameter is now a swing open valve and the weight of the float is replaced by the spring force applied on the valve.

    Rotameter:

    A rotameter is mounted vertically with the narrow end at the bottom and the tube tappers into a wider top. The flow comes from the bottom and pushes the float inside the rotameter up to a point that the weight of the float is in balance with the force exerted by the flow. The annular area between the float and the tube wall is then related to the volume flow rate.

    As long as the fluid speed is substantially subsonic (V < mach 0.3), the incompressible Bernoulli's equation applies.

    where g is the gravity acceleration constant (9.81 m/s2 or 32.2 ft/s2), V is the velocity of the fluid, and z is the height above an arbitrary datum. C remains constant along

    http://www.efunda.com/DesignStandards/sensors/flowmeters/flowmeter_va.cfm (2 of 5)29/09/2007 15:51:03

  • eFunda: Introduction to Variable Area Flowmeters

    any streamline in the flow, but varies from streamline to streamline. If the flow is irrotational, then C has the same value for all streamlines.

    Applying this equation to a streamline traveling up the axis of the vertical tube gives,

    where subscript a represents the position right below the float, b is the balanced point of the float, usually the top of the float, V is the flow velocity, p is pressure, and is the density. A shorter form of the above equationi is

    where hf is the hight of the float or the distance from the bottom to the indicator of

    the float that depends on the float design.

    From continuity, the volume flow rate at a is the same as the volume flow rate at b, i.

    e., , which implies

    Please note that is the annular area between the float and the tube

    wall, not the whole cross section area at b.

    Hence, the velocity Vb can be substituted out of the Bernoulli's equation to give,

    The pressure drop is mostly resulting from the weight of the float

    where the subscript f represents the float, Vf is the volume, Af is the cross section

    area, and f is the density of the float.

    Solving for the volumetric flow rate Q, we have

    http://www.efunda.com/DesignStandards/sensors/flowmeters/flowmeter_va.cfm (3 of 5)29/09/2007 15:51:03

  • eFunda: Introduction to Variable Area Flowmeters

    Ideal, inviscid fluids would obey the above equation. The small amount of energy converted into heat within viscous boundary layers tends to somewhat lower the actual velocity of real fluids. A discharge coefficient C is typically introduced to account for the viscosity of fluids,

    C is found to depend on the Reynolds Number of the flow.

    For a given design, the cross section areas Aa(z) and Ab(z) of the rotameter are

    functions of the hight z, and the geometry (hf, Af, Vf) and the density ( f) of the float

    are also known. If the density of the fluid is measured and the readout of the position (z) of the float in the rotameter is available, the volume flow rate Q can be calculated from this formula:

    The mass flow rate can be easily found by multiplying Q with the fluid density ,

    Top of Page

    Common Specifications

    Common specifications for commercially available variable area flowmeters are listed below:

    http://www.efunda.com/DesignStandards/sensors/flowmeters/flowmeter_va.cfm (4 of 5)29/09/2007 15:51:03

  • eFunda: Introduction to Variable Area Flowmeters

    Fluid Phase: Please refer to flowmeter selection for specific sub-categories

    Score Phase Condition

    Gas Clean

    Liquid Clean

    Open Channel

    Gas Dirty

    Liquid Corrosive

    Dirty

    Steam Saturated

    : Recommended : Limited applicability

    Line Size: Mostly used in lines size 100 mm (4 inch) and below, but may go up to 2000 mm (80 inch) in line size or even for open channels.

    Turndown Ratio: 10 ~ 100 : 1

    Top of Page

    Pros and Cons

    Pros:

    - Very low initial set up cost

    - Simple, robust

    - Low, nearly constant, pressure drop

    Cons:

    - Moderate accuracy at best

    - Not suitable for low flow rate

    - Some variable area flowmeters can not be used in non/low gravity environments

    - Rotameters must be mounted vertically

    Top of Page

    Membership About Us Privacy Disclaimer Contact Advertise

    Copyright 2007 eFunda

    http://www.efunda.com/DesignStandards/sensors/flowmeters/flowmeter_va.cfm (5 of 5)29/09/2007 15:51:03