Transcript of Introduction to the 8000 Hearing Aid Test System
![Page 1: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/1.jpg)
Introduction to the 8000 Hearing Aid
Test System
By: Kristina FryeFrye Electronics, Inc.
![Page 2: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/2.jpg)
Outline Coupler Measurements
Coupler Directional ANSI/IEC Coupler Multicurve Advanced Coupler: Enhanced DSP, Attack &
Release, I/O, Battery Current
![Page 3: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/3.jpg)
Coupler Directional
F3 – Advanced Coupler F5 – Coupler Directional
![Page 4: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/4.jpg)
Directional Measurements
Polar Plot
Difference (0° vs 135°)
![Page 5: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/5.jpg)
Simultaneous, Manual, and Auto Modes In Simultaneous Mode, five polar plot curves are
measured simultaneously A Composite source signal is used to take the
measurements. The response at specified frequencies is graphed
In Manual Mode, the user selects the curve to measure. A Puretone signal at the user specified frequency or a Composite source is used.
In Auto Mode, the user can pre-save manual measurements and run curves automatically
![Page 6: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/6.jpg)
WinCHAP Data Grab
![Page 7: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/7.jpg)
Basic Coupler Measurements
ANSI 03 — F4Coupler Multicurve — F1
![Page 8: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/8.jpg)
ANSI 03 Settings
Choose ear
Aid type
Frequencies Telecoil
![Page 9: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/9.jpg)
ANSI 03 ResultsOSPL90
Response curve
Max
OSPL90-HFA
Distortion
FOG
TelecoilEIN
Battery current
Telecoil I/O curves
Attack & Release
![Page 10: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/10.jpg)
Troubleshooter ANSI
![Page 11: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/11.jpg)
Coupler Multicurve Features
Realistic frequency response measurements
Variety of signal types: Composite, Digital Speech, Pure-tone sweeps
Display 10 different curves simultaneously
Switch back and forth between gain and output. All curves translated.
![Page 12: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/12.jpg)
Coupler Screen Operation
Choose Ear
Select Curve Source Type
Amplitude
![Page 13: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/13.jpg)
Example
50 dB
60 dB
70 dB
80 dB
![Page 14: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/14.jpg)
Troubleshooter Coupler Multicurve
![Page 15: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/15.jpg)
Testing Digital Aids Digital Speech Type of Composite signal Modulated to allow testing of noise
reducing digital aids Used with ANSI or ICRA speech spectra
![Page 16: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/16.jpg)
Example of digital aid test
Digital Speech
Composite
![Page 17: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/17.jpg)
Testing Open Fit Hearing Aids The Open Fit Coupler was designed to
Provide a realistic response for open fit hearing aids.
Provide an easy mechanism for attaching an open fit hearing aid to a coupler
The Open Fit Coupler is a non-standard sized coupler not designed to test to manufacturing specifications
![Page 18: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/18.jpg)
Attaching the Open Fit Coupler
CRV 1 = Open Fit CouplerCRV 2 = HA-1CRV 3 = HA-2 w/attachmentCRV 4 = KEMAR
![Page 19: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/19.jpg)
Coupler Automation Set up own test sequence Select signal type for each curve Select signal amplitude for each curve Add any desired pauses in test sequence Save and auto load test sequence
![Page 20: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/20.jpg)
Programming Auto Mode
F6 — Program Mode
Help window w/instruction
s
F5 — Source TypeAmplitude
START
STOP
Advance to next curve without pause
Advance to next curve with pause
![Page 21: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/21.jpg)
Example of Auto Test
Digital Speech @ 50, 65, 80 dB
Pure-tone sweep @ 90 dB
F6 — Fully Automatic mode
![Page 22: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/22.jpg)
Saving and Loading Settings
Automatically load saved curve group and put analyzer in Auto Mode
![Page 23: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/23.jpg)
Troubleshooter ProfilerGreat for baseline testing… save time!
![Page 24: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/24.jpg)
Advanced Coupler Battery Current [F1] Coupler I/O [F2] Attack & Release [F3] Enhanced DSP [F4]
Function key operations available in the Advanced Coupler Navigation Screen
![Page 25: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/25.jpg)
Battery Current Screen Measure Battery Current as a function of
frequency Measure Battery Current as a function of
amplitude Estimate battery life of hearing aid
![Page 26: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/26.jpg)
Coupler I/O Test Results
Frequency
Test Results
F5: Source Type
![Page 27: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/27.jpg)
Attack & Release Test Screen Visual time graph of Attack and Release
tests Scale graphs independently to zoom in on
results Use Composite or Pure-tone Choose any frequency from 200-8000 Hz
in 100 Hz intervals Customize amplitudes used in tests
![Page 28: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/28.jpg)
Attack & Release Results
Frequency
F5: Source type
F6 & F7: Change graph scales
![Page 29: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/29.jpg)
Enhanced DSP Designed to measure the "group delay" of a
DSP aid. In other words, the actual digital processing time of a hearing aid. 0 ms for analog aids 1-12 ms for DSP Great way to check if aid is analog or DSP
Also tests the "phase" of a hearing aid. This is useful for binaural fittings. Are the aids working together?
![Page 30: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/30.jpg)
Technical Details: Group Delay
An impulse signal is delivered to the hearing aid. The time of the maximum peak of the response is considered the group delay. The 7000 system delay is subtracted from this value to obtain the final measurement.
![Page 31: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/31.jpg)
Technical Details: Phase
One pulse of a 1000 Hz cosine signal is delivered to the aid. This signal is 1 msec long. The first 20 msec of the aid's response to the signal is displayed in graphical format.
![Page 32: Introduction to the 8000 Hearing Aid Test System](https://reader036.fdocuments.us/reader036/viewer/2022062407/56812aa0550346895d8e643a/html5/thumbnails/32.jpg)
ExampleGroup Delay Phase