INTERSTER BOWIE & DICK AUTOCLAVE TEST PACK - KITEMARK€¦ · o Process indicator on pack conforms...

2
1300 995 999 | [email protected] www.sentrymedical.com.au The Interster Bowie & Dick Autoclave Test Pack aligns with the requirements of AS4187 as a steam penetration, air removal test. Result record sheet included. PRODUCT FEATURES o 7kg mimic o Does not contain heavy metals o Kitemark - BS EN-ISO 11140-4 o Conforms to EN-ISO 11140-1 (Type 2) o Process indicator on pack conforms to EN-ISO 11140-1 (Type 1) o Contains a specially developed indicator ink which is affixed through a silkscreen printing process SUITABLE FOR o Autoclave sterilisers REORDER FEATURE PCS/UNIT UNITS/CTN 3FSKS610860 BSI Kitemark 40 1 INTERSTER BOWIE & DICK AUTOCLAVE TEST PACK - KITEMARK SMHO-All-BRO-0142_REV3_06_20 All product informaon is correct at date of publicaon NON STERILE LATEX

Transcript of INTERSTER BOWIE & DICK AUTOCLAVE TEST PACK - KITEMARK€¦ · o Process indicator on pack conforms...

  • 1300 995 999 | [email protected]

    The Interster Bowie & Dick Autoclave Test Pack aligns with the requirements of AS4187 as a steam penetration, air removal test. Result record sheet included.

    PRODUCT FEATURES o 7kg mimic o Does not contain heavy metals o Kitemark - BS EN-ISO 11140-4 o Conforms to EN-ISO 11140-1 (Type 2) o Process indicator on pack conforms to

    EN-ISO 11140-1 (Type 1) o Contains a specially developed indicator ink which

    is affixed through a silkscreen printing process

    SUITABLE FOR o Autoclave sterilisers

    REORDER FEATURE PCS/UNIT UNITS/CTN

    3FSKS610860 BSI Kitemark 40 1

    INTERSTER BOWIE & DICK AUTOCLAVE TEST PACK - KITEMARK

    SMHO

    -All-

    BRO

    -014

    2_RE

    V3_0

    6_20

    All p

    rodu

    ct in

    form

    ation

    is c

    orre

    ct a

    t dat

    e of

    pub

    licati

    on

    NONSTERILE

    LATEX

  • Test results Bowie & Dick Type Autoclave Lean Pack:

    Before use

    Not properly changed sheetA clearly defined mark is visible.Explanation: presence of air or non-condensable gases. Possible causes: poor air removal, leakage or gas in the steam and sub-atmospheric pulsing.

    Completely changed sheetMinimum even colour change that must be achieved.

    Not properly changed sheetAn indistinct mark, almost always in the middle, is visible.Explanation: superheated steam. Possible causes: the test pack has been placed in the sterilizer without immediately starting the process(desiccation), too high steam velocities or steam supply pressure.

    Not properly changed sheetPossible causes: a too short sterilization time or too low temperature.

    Not properly changed sheetCompletely changed sheet except for the corners. Possible causes: wet steam.

    Do not use the above shown pictures as a reference colour

    Interster International B.V.Samsonweg 2, 1521 RC Wormerveer T: +31 75 627 22 41 - E: [email protected]

    www.interster.nl/exportrange

    Art. no. 3FSKS610860 - 3,5 min.

    02-

    2016