Interrogation and programming workshop part I - Pacemaker follow-up LING Hon-Lun Associate Sales Manager Cardiac Rhythm Disease Management (CRDM)

Transcript of Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ”...

Page 1: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Interrogation and programming

workshop part I

- Pacemaker follow-up

LING Hon-Lun

Associate Sales Manager

Cardiac Rhythm Disease Management (CRDM)

Page 2: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Interrogation and programming

workshop part II

Hands-on session

Medtronic helpers

Page 3: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Aims Of Pacemaker Follow Up

• Determine battery status

• Check lead integrity

• Check Sensing Threshold

• Check Pacing Threshold

• Analyse diagnostics

• Assess current parameters

• Obtain documentation

Page 4: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Device interrogation

Page 5: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Question?

What is the most important piece attached

to an pacemaker system?

Page 6: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

The PatientThe PatientThe PatientThe PatientThe PatientThe PatientThe PatientThe Patient

Treat the patient, not the device

Page 7: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Interrogation

Presenting Rhythm

& Rate

Battery Status

Lead impedance

Threshold

Sensing

Observation

Program & Print

Page 8: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

P Presenting Rhythm and Rate

B Battery Status

L Lead Status

S Sensing

T Threshold

O Observation, Data, and Events

P Program & Print

“PBL STOP”

Page 9: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

‘PBL STOP’ – Your acronym for

a standardized follow-up

Page 10: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Programming step by step

• Determine battery status

• Check lead integrity

• Check Sensing Threshold

• Check Pacing Threshold

• Analyse diagnostics

• Assess current parameters

• Obtain documentation

Page 11: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Device battery status

Page 12: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Programming step by step

• Determine battery status

• Check lead integrity

• Check Sensing Threshold

• Check Pacing Threshold

• Analyse diagnostics

• Assess current parameters

• Obtain documentation

Page 13: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Check lead integrity

Page 14: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Lead integrity – in office test

Page 15: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Lead integrity – in office test

Page 16: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Programming step by step

• Determine battery status

• Check lead integrity

• Check Sensing Threshold

• Check Pacing Threshold

• Analyse diagnostics

• Assess current parameters

• Obtain documentation

Page 17: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Sensing

Page 18: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Sensing – in office test

Page 19: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Programming step by step

• Determine battery status

• Check lead integrity

• Check Sensing Threshold

• Check Pacing Threshold

• Analyse diagnostics

• Assess current parameters

• Obtain documentation

Page 20: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Threshold

Page 21: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Threshold – in office test

Page 22: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

1. Pulse amplitude – the amount of voltage

delivered to the heart by the pacemaker to ensure capture

of the heart (strength of an output pulse)

2. Pulse width – time (duration) of the pacing pulse

2 Settings Used to Ensure Capture

GOAL Find the minimum amount of energy needed to

consistently capture the heart (stimulation threshold).

Page 23: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Amplitude Threshold Testing

Keeping pulse width stable, decrement voltage until loss of

capture

1.5V 1.0V .5V

@ 0.5ms

Page 24: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Pulse Width Threshold TestingKeeping voltage stable, decrement in pulse width

.30ms .20ms .1ms

@ 2 volts

.06ms

Page 25: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Programming step by step

• Determine battery status

• Check lead integrity

• Check Sensing Threshold

• Check Pacing Threshold

• Analyse diagnostics

• Assess current parameters

• Obtain documentation

Page 26: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Analyse diagnostics (Observation)

Page 27: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Analyse diagnostics (Observation)

Page 28: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Programming step by step

• Determine battery status

• Check lead integrity

• Check Sensing Threshold

• Check Pacing Threshold

• Analyse diagnostics

• Assess current parameters

• Obtain documentation

Page 29: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Assess current parameters (Program)

Page 30: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Assess current parameters (Program)

Page 31: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Assess current parameters (Program)

Page 32: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Programming step by step

• Determine battery status

• Check lead integrity

• Check Sensing Threshold

• Check Pacing Threshold

• Analyse diagnostics

• Assess current parameters

• Obtain documentation

Page 33: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

“P”

Programming &

Printing Final Reports

(obtain documentation)

Page 34: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Print final report

Page 35: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Printed final report – post f/u parameter

Page 36: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Print Reports

• Print Final Reports

– Sensing test results

– Threshold test results

– Key observations

– Any changes this session noted on final report (initial and

final)

– Documented all tested result and parameters

Page 37: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

P Presenting Rhythm and Rate

B Battery Status

L Lead Status

S Sensing

T Threshold

O Observation, Data, and Events

P Program & Print

“PBL STOP”

Page 38: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

PBL STOP

P

B

L

OT

P

S

Page 39: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Interrogation and programming

workshop part II

Hands-on session by Ray Chan

Medtronic helpers:

Poly Fu, Ray Chan, Brandon Chen, Chapman Lee, Fayzi Lee, Tina Ling, Shirley Chau, Cristina Leung

Page 40: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Hands-on session

• Hands-on for a pacemaker follow-up

• Divided into 5 groups each of 8-10 persons

• 5 tables of Medtronic simulators

• Two virtual patient A & B

• Each one perform both patient with printed reports

45mins

Post hands-on f/u observation~ 10mins

Page 41: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Patient historyPatient B

• Post implant 1 day

• Implanted with Medtronic dual chamber pacemaker

• RA & RV active fixation leads

• First degree AVB

Patient A

• Replacement done with a Medtronic dual chamber pacemaker over 1 year

• RA & RV old passive leads implanted in 2002

• Syncope, AF, long pause ~ 5 sec

Parameters during replacement

P wave: 2.0mV

R wave: 18mV

Pacing threshold:

RA: [email protected]

RV: [email protected]

Imp:

RA: 688Ω

RV: 1200Ω

Parameters during implant

P wave: 1.8mV

R wave: 10mV

Pacing threshold:

RA: [email protected]

RV: [email protected]

Imp:

RA: 632Ω

RV: 887Ω

Page 42: Interrogation and programming workshop part I - … · • Check lead integrity ... “PBL STOP ” ‘PBL STOP ... Interrogation and programming workshop part II Hands-on session

Post follow up observation

Patient A Old leads implanted 11yrs

old

RV lead parameters Impedance: 1200 > 4xxΩΩΩΩ

R wave: 18mV > 2.0mV

Threshold ↑↑↑↑ to 5.0V

Conclusion: RV parameters change showing a suspected RV lead insulation break and degradation

Clinical concern:

1. Pacing dependency?

2. Intact AV conduction?

3. RV lead replacement need?

Patient B Post implant one day with new RA &

RV active leads

RA lead parameters Impedance: normal

P wave: ~ 3.0mV (normal? EGM?)

RA threshold: n/a but capture RV@ 5.0V

Conclusion: RA parameters, ECG & EGM signals shown a suspected RA lead dislodgement to RV chamber

Clinical concern:1. RA reposition need

2. Active fixation on alternative RA position?

3. Passive lead?