International Symposium on Fluid Flow Measurement/67531/metadc690499/m2/1/high... · REAL-TIME...

13
REAL-TIME MEASUREMENT OF VEHICLE EXHAUST GAS FLOW J. E. Hardy, T. E. McKnight, and J. 0. Hylton Oak Ridge National Laboratory* R. D. Joy Retired from J-Tec Associates, Inc. Paper prepared for the International Symposium on Fluid Flow Measurement June28, 1999 7he submitted manuscript has been authored by il contractor of the U.S.Govemment under contract no. DE-AC05-960R22464. Accordingly, the U.S. Government retains a nonexclusive. royalty-free license to publish or reproduce the published form of this contribution. or allow others to do so. for U.S. Government purposes.” Research sponsored by the U.S. Department of Energy and performed at Oak Ridge National Laboratory, managed by Lockheed Martin Energy Research Corporation for the U.S. Department of Energy under contract DE-AC05-960EU2464.

Transcript of International Symposium on Fluid Flow Measurement/67531/metadc690499/m2/1/high... · REAL-TIME...

REAL-TIME MEASUREMENT OF VEHICLE EXHAUST GAS FLOW

J. E. Hardy, T. E. McKnight, and J. 0. Hylton Oak Ridge National Laboratory*

R. D. Joy Retired from J-Tec Associates, Inc.

Paper prepared for the International Symposium on Fluid Flow Measurement

June28, 1999

7 h e submitted manuscript has been authored by il contractor of the U.S.Govemment under contract no. DE-AC05-960R22464. Accordingly, the U.S. Government retains a nonexclusive. royalty-free license to publish or reproduce the published form of this contribution. or allow others to do so. for U.S. Government purposes.”

Research sponsored by the U.S. Department of Energy and performed at Oak Ridge National Laboratory, managed by Lockheed Martin Energy Research Corporation for the U.S. Department of Energy under contract DE-AC05-960EU2464.

DISCLAIMER

This report was prepared a s an account of work sponsored by an agency of t h e United States Government. Neither t h e United States Government nor any agency thereof, nor any of their employees, make any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of.the United States Government or any agency thereof.

DISCLAIMER

Portions of this document may be illegible in electronic image products. Images are produced from the best available original document.

REAL-TIME MEASUREMENT OF VEHICLE EXHAUST GAS FLOW J. E. Hardy, T. E. McKnight, J. 0. Hylton

Oak Ridge National Laboratory R. D. Joy

Retired from J-Tee Associates, Inc.

ABSTRACT

A flow measurement system was developed to measure, in real-time, the exhaust gas flow from vehicles. This new system was based on the vortex shedding principle using ultrasonic detectors for sensing the shed vortices. The flow meter was designed to measure flow over a range of 1 to 366 Ips with an inaccuracy of 21% of reading. Additionally, the meter was engineered to cause minimal pressure drop (less than 125mm of water), to function in a high temperature environment (up to 650OC) with thermal transients of 15OC/s, and to have a response time of 0.1 seconds for a 10% to 90% step change. The flow meter was also configured to measure bi-directional flow. Several flow meter prototypes were fabricated, tested, and calibrated in air, simulated exhaust gas, and actual exhaust gas. Testing included gas temperatures to 6OO0C, step response experiments, and flow rates from 0 to 360 Ips in air and exhaust gas. Two prototypes have been tested extensively at NIST and two additional meters have been installed in exhaust gas flow lines for over one year. This new flow meter design has shown to be accurate, durable, fast responding, and to have a wide rangeability.

-

INTRODUCTION

A need was identified by the auto industry for an economical, real-time exhaust gas flow measurement system. New Environmental Protection Agency's (EPA) specifications for low concentration levels of pollutants in vehicle exhaust emissions provided the motivation for the development of sophisticated vehicle emissions testing programs. The advanced analytical tools being developed to measure the low concentration emissions depend on the ability to accurately measure the total exhaust gas flow in real time. While a near-tern benefit of this flow technology is the capability to certify a vehicle's ability to meet the new EPA regulations, additional possibilities include an aid to new engine development (real time flow and emissions data) and environmental monitoring for a variety of industries.

The significant design goals for this flow measurement systems were:

0 Flow range of 1 to 146 Ips, 0 Exhaust gas temperatures from -3OOC to 65OoC, 0 Maximum pressure drop across the meter of 127mm of water, 0 Response time (10% to 90% of a step) of 0.1 second, 0 Maximum temperature transient of 15OC per second, and 0 Accuracy of +1% of reading.

A commercially available flow meter did not exist, even after over 40 years of various attempts. Oak Ridge National Laboratory (ORNL) and J-Tec Associates, Inc. teamed together to

develop a new flow measurement device to meet the above criteria [ 11. This team received valuable insights from the auto industry through the Environmental Research Consortium (ERC).

MEASUREMENT TECHNIQUE

The vortex shedding principle [2,3] was chosen to make the required flow velocity measurement. This method works by placing a non-streamline bluffbody in the flow path. Above a critical Reynolds number, Von Karman vortices are shed from the bluff body. The vortices are shed from alternating sides of the downstream edge of the bluff body or strut and are essentially the same distance apart regardless of the gas properties. The size and shape of the bluff body was chosen such that the commencement of the vortex creation will occur at a flow velocity below the minimum design flow. For this particular application, circular cylinders were chosen for the struts. Downstream of the struts, diametrically opposed, were ultrasonic transmitters and receivers. An ultrasonic beam was transmitted across the pipe through the vortex path. As the shed vortices pass through the beam, the sonic rays are deflected due to the vectorial addition of the rotational velocity of the vortex and the sonic velocity of the sound beam. This effect causes the ultrasonic beam to be alternately "focused" and "unfocused or spread" as the vortices pass. This produces varying intensity of the ultrasonic energy that is received and sensed by the detector. The fiequency of the varying energy is the vortex shedding frequency. The vortex frequency is related to flow velocity as shown in Equation 1,

Vortex Shedding Frequency x Bluff Body Diameter

Strouhal Number V = Eq. 1

where V is the flow velocity and the Strouhal Number is a non-dimensional term that becomes a constant value at Reynolds' numbers > 10,000.

To be able to measure bi-directional flow, in case reverse flow occurs in the exhaust line, three struts were placed in the flow stream with two sets of transmitterddetectors. The arrangement of the bluff bodies and ultrasonic transducers are shown in Figure 1.

FLOW

BLUFF BODIES

0

f

VORTICES

ULTRASONIC TRANSDUCERS

Figure 1. Schematic Diagram of Bluff Bodies, Transducers, and Vortices

In forward flow as shown in Fig. 1, the vortices shed by the first 3.2mm-diameter strut are detected by the first set of ultrasonic transducers (A). The vortices shed by the middle strut (5 mm in diameter) are sensed by the second set of transducers (€3). Referring to Eq. 1, the smaller diameter strut will produce a higher shedding frequency for the same velocity as compared to the larger, center bluff body. With transducer pair A measuring the higher frequency than pair B, the flow direction is known, lefi to right. If the flow reverses direction, then transducer pair B will have the higher frequency and the flow direction is known to be from the right to left.

MEASUREMENT SYSTEM DESIGN

To accommodate the dynamic range of flow from 1 to 366 Ips, two separate flow meters were used; one was 50 mm in diameter and the second one was 76 mm in diameter. The 50-mm diameter flow tube measurement range was 0.9 to 146 Ips and the 76-mm diameter meter had a measurement range of 7 to 366 Ips. To prevent condensation inside the flow tube, which would cause a significant measurement error, the flow tube was heated with externally mounted heat tape to approximately 90°C, which was above the exhaust gas dew point temperature of -7OOC. A pressure sensor and thermocouple were installed with the flow meter so the actual flow rate could be converted to standard flow conditions. EPA requires the flow rate to be presented in standard volumetric units.

Pressure transients are fairly slow in a vehicle exhaust system but temperature transients are quite rapid, 15OC per second. If the temperature measurement device is not fast enough to keep up with the transients, errors in correcting the actual flow to standard flow could be as high as 5%. To mitigate this possible error, a 1-mm, type-N, exposed junction thermocouple was selected. A thermal model [4] was used to size the thermocouple, its material, and the need for one concentric radiation shield.

For use in the auto industry to test vehicles in their emission test stands, both the 50-mm and 76-rnm flow meters were housed in one unit with a user interface located remotely. The ultrasonic transducers used have a Curie temperature of 365OC. Therefore cooling air was provided to limit their temperature to 100°C. A photograph of the overall system is shown in Figure 2.

Figure 2. Photograph of the prototype Exhaust Gas Flow

I

Meter

FLOW METER TESTING AND CALIBRATION

The time response of the entire measurement system was determined by “instantaneously” changing the airflow with fast acting shutoff and bypass valves. The flow was varied from 0% to 50% of MI scale (FS), from 50% to 0% FS, and from 25% to 50% FS. The time response (10% to 90% of the input step size) was less than 100 ms for all cases. A 90-ms filter in the system software primarily determined the response time. A typical response curve is illustrated in Figure 3.

Time Response

is < looms (10 to 90% of step)

Time (100 mddiv)

Figure 3. Time response curve for a step input

Initial air testing of the 50-mm and 76-mm flow meters indicated that the device output was linear with flow rate and had wide rangeability. The 50-mm test results are shown in Fig. 4.

3000

? 2500

0 20 40 60 80 100 120

Flow Rate - WS

Figure 4. Frequency Output Vs Flow for the 50-mm Prototype Flow Meter

The 50-mm flow meter was also calibrated in air to verify the accuracy of the prototype. The data for the low end of the calibration run is shown in Figure 5.

s 61 F

0

L

L Q) c,

5 56 Q) L Y

s

50

45 40

35

30

25 20

15 10 5

0 0 10 20 30

Standard Air Flow (SLPS)

40 50

Figure 5. Air Flow Calibration of the 50-mm Vortex Flow Meter

Additional testing in an engine test stand showed that the vortex flow meter could withstand temperatures up to 6OO0C and function properly. However, measurement system problems were discovered that were attributed to pressure fluctuations in the exhaust line. This particular test stand did not have a muffler or catalytic converter so pressure fluctuations due to valve movements were not dampened before reaching the vortex flow meter. The signal processing algorithms and filtering ahead of the ultrasonic receiver were modified to handle the system dynamics. Subsequent testing on the engine test stand indicated that the modifications corrected for the excessive noise on the detector output signal.

A 50-rnm diameter prototype vortex flow meter was sent to NIST to undergo calibration testing in air and simulated exhaust gas flow. The NIST flow stand can run at temperatures up to 260°C with actual exhaust gas constituents. The prototype sent to NIST did not have pressure or temperature measurements or tube wall heaters. The prototype measured the actual volumetric gas flow.

Data taken at NIST in room temperature air (2OoC) is shown in Figure 6. The data was plotted in terms of output frequency divided by the flow rate (the slope of the frequency Vs flow curve) versus the NIST determined flow rate to better show the results. For flow rates above about 5 Ips the variation in the frequency/flow value is less than 22%. For flows <3 Ips the variability in the frequency/flow term was 230%. These variations for flows under 5 Ips are illustrated graphically in Figure 7.

35

330 0 - 25 > 0 LIJ

LIJ 15

E!

3 E 2 10

* 20

n c = 5 0

0 0 10 20 30 40 50 60 70 80

FLOW RATE - LPS

Figure 6. NIST data for the 50-mm flow meter using air at 20°C

5%

4%

3%

8 1% s

0% z

-1% a

z -2% w 0

3

-3%

4%

-5% 0 10 20 30 40 50 60 70 80

FLOW RATE - LPS

Figure 7. Uncertainty in data for the 50-mm flow meter using air at 20°C

The prototype flow meter was then tested with air and exhaust gas at higher temperatures. The data from these experiments are shown in Figures 8-10. The variation in the output./flow term for air at 12OoC became larger than +2% below 7 Ips, slightly higher than at 2OoC data. At air temperatures of 26OoC, the variability becomes > 2% for flows below -17 Ips (Figure 9.) Figure 10 contains data with simulated exhaust gas. The variability in the output&low term is larger than 22% at flows lower than 6.5 Ips, approximately the same as the 12OoC-ak case. It appears that the meter functions the same in air as in exhaust gas. It was later determined that excessive transducer cooling airflow caused the apparent rise in the frequency divided by flow data at low meter flow rates. When a later NIST test was made on the 76-mm unit this problem did not appear.

50

45 40

0 L

0 0 20 40 60 80 100 120

FLOW RATE - LPS

Figure 8. NIST data for the 50-mm flow meter using air at 120'C

50

45 0 2 40

E 30

L

0 0 20 40 60 80 100 120 140

FLOW RATE - LPS

Figure 9. NIST data for the 50-mm flow meter using air at 26OoC

3 0 J L 4

40

35

30

25

20

15

10

5

0 0 10 20 30

FLOW RATE - LPS

40 50

Figure 10. NIST data for the 50-mm flow meter using simulated exhaust gas at 12OoC

In attempting to understand why temperature was affecting the flow meter performance and gas mixture was not, the vertical temperature profile was measured in the flow duct. Significant variations in the gas temperature were measured at the top and bottom of the flow duct as shown in Figure 11. In the NIST test stand, the meter was preceded by a straight pipe about 6-m long, which was heated to the same temperature as the gas flow. As the flow travel down the pipe the higher temperature gas drifted towards the top of the pipe. It is believed this temperature stratification caused the flow profile to become distorted with higher flows towards the top of the pipe and hence the strut. There were different flow rates along the vortex strut and this caused some uncertainty in the vortex formation and thus, the flow measurement. As the flow increases and better mixing occurs, the temperature difference becomes smaller. When the temperature difference was less than 5'C, the outpudflow value variation was <2%. This phenomenon held true for the 26OoC air data and the 120'C exhaust gas data. To achieve higher accuracy at the low flows, better mixing is required.

To test this theory a flow straightener that was designed to mix the flow was added upstream of the 76-mm vortex flow meter.

i- o 20 a

0 20 40 60 80 100 120

FLOW RATE - LPS

Figure 1 1. Temperature profile in flow duct for air at 12OoC

Temperature difference data with the straightener in and out of the flow duct for air at 12OoC is shown in Figure 12. The mixer reduced the temperature difference at the lower flows ( 4 0 Ips) by 2OoC to 3OoC. At about 12 Ips the temperature difference was the same with or without the mixer. For flows above 18 Ips the mixer cause a higher temperature difference than occurred with no mixer in place.

120

a

t- 3 0

-2 0 0 20 40 6 0 80 100 120 140

FLOW R A T E IN LPS

Figure 12. Temperature Difference with and without a flow mixer for air at 120'C

ability of the flow meter to measure accurately is affected by temperature distribution in the flow duct. If the temperature difference between the top and the bottom of the flow duct is less than 5OC, the meter will maintain its stated accuracy. The six-meter heated upstream pipe in the MST test stand likely caused much of the temperature stratification. In most applications it is unlikely that long, heated pipe lengths will be used which would create the stratification problem. Ifthere were a long horizontal heated pipe, then installing the flow meter in a vertical pipe section wouid minimize or prevent this problem.

An interesting result that was uncovered during the testing at NIST was that a flow straightenerhixer did not necessarily improve the temperature distribution within the flow duct. It may be that the device straightens the flow but is not effective at mixing the flow. Only one test was conducted with the mixer but the data indicated that the mixer was of help at low flow rates and a detriment at higher flows.

REFERENCES

1.

2.

3.

4.

Hardy, J. E., et al. (1997). Exhaust Gas Flow Measurement System, C/ORNL93-0232.

Miller, R W., (1983). Flow Measurement Engineering Handbook, McGraw-Hill, pp. 14- _ _ 18.

Joy, R. D., (1984). Ultrasonic Vortex Flowmeters. Proceedings of the 3dh International Imtmmentation Symposium, ISA.

Moffat, R. J., (1 962). Gas Temperature Measurement Temperature-Its Measurement and Control in Science and ItidisQ, 3(2).

ACKNOWLEDGEMENTS

The authors would like to acknowledge the support of the Laboratory Technology Research Program of the Ofice of Science in the U.S. Department of Energy. The authors would also like to express their appreciation to Craig A. Morgan of the Chrysler Proving Grounds, colleagues at J-TEC particularly Greg Miltner and ORNL, and the assistance of personnel at NIST. This work was completed under contract DEAC05-960R22464 with the Oak Ridge National Laboratory, managed by Lockheed Martin Energy Research Corporation.