Transcript of InsuJet Instructions
-
8/18/2019 InsuJet Instructions
1/14
instructions for use
-
8/18/2019 InsuJet Instructions
2/14
INDEX
Attaching the nozzle and piston 4
Attaching the adaptor 7
Preparing the InsuJet™ for administration 12
Administrating insulin with the InsuJet™ 18
Removing the nozzle and piston for replacement 221 CARTRIDGE CAP
REMOVAL KEY
InsuJet™ BASIC STARTER PACK CONTAINS
1 INSTRUCTIONS FOR USE
1 ADDITIONAL INFORMATION
1 POUCH
EXTERNAL SUPPLIES: A TISSUE
1 NOZZLE & PISTON
1 NOZZLE CAP1 InsuJet™
PEN1 VIAL (10mL) ADAPTOR
1 CARTRIDGE (3mL) ADAPTOR
-
8/18/2019 InsuJet Instructions
3/14
4
ATTACHING THE NOZZLE AND PISTON
Whilst sliding the nozzle lock catch
upwards, turn the nozzle lock to the rightuntil the nozzle lock arrow points to the
padlock unlocked position.
Ensure that the triangular corners on the
nozzle and piston are aligned with the
triangular shape inside the nozzle lock before inserting the nozzle and piston.
Firmly push the nozzle and piston as far
as it will go straight into the opening in the nozzle lock of the InsuJet™.
UNLOCKED
POSITION
LOCKED
POSITION
-
8/18/2019 InsuJet Instructions
4/14
6
CARTRIDGE (3mL) ADAPTOR VIAL (10mL) ADAPTOR
CLICK!
You should hear and fee l a ‘click’ as the
nozzle and piston engage. At least 1/3rd
of the nozzle cap will no longer be visible.
Once you have fully inserted the nozzle
and piston, turn the nozzle lock to the left
until the nozzle lock arrow points to the
padlock locked position. The nozzle and
piston have now been installed correctly.
Vial adaptor 10mL
Cartridge adaptor 3mL
The instructions show a cartridge (3mL)
adaptor with an insulin cartridge
ATTACHING THE ADAPTOR
IMPORTANT
Ensure that the nozzle lockarrow points to the padlocklocked position.
-
8/18/2019 InsuJet Instructions
5/14
8
Screw the cartridge cap removal key onto
the coloured cartridge cap.
Penll® is a registered trademark of Novo Nordisk A/S.
NOTICE
For Novo Nordisk Penll® users:
Remove the colored cartridge cap by using
the cartridge cap removal key.
10mL vial
Several 3mL cartridges
-
8/18/2019 InsuJet Instructions
6/14
10
Discard the coloured cartridge cap with
your regular household waste.
The picture shows a cartridge adaptor (3
mL) with an insulin cartridge.
Push the adaptor over the end of the
insulin cartridge/vial, making sure that
the plastic needle of the adaptor pierces
the rubber stopper on top of the insulincartridge/vial.
Pull off the coloured cartridge cap from the
Penll® with a straight rm pull.
The adaptor remains on the cartridge/vial.
-
8/18/2019 InsuJet Instructions
7/14
12
click!
Attach the insulin cartridge/vial by pressing
the adaptor rmly onto the front end ofthe nozzle and piston. Tilting the adaptor
while pressing will ensure that the adaptor
attaches easily.
Remove the clear adaptor cover from the
adaptor. Ensure that the nozzle cap is
removed from the nozzle and piston.
On the back body of the InsuJet™ you will
notice the number ‘1’ with a directionalarrow. Wind the back body in the direction
of the arrow until you feel resistance. The
InsuJet™ is now ready to be lled with
insulin.
PREPARING THE InsuJet™ FOR ADMINISTRATION
-
8/18/2019 InsuJet Instructions
8/14
14
NOTICE
Always check for air bubbles.
Hold the back body stationary and turn
the front body in the direction of the arrow
until your desired dose of insulin appearsin the dose window. Always aspirate some
extra insulin into the nozzle. If you do not
see any air bubbles, 1 unit of insulin is
sufcient. If you see some air bubbles in
the nozzle, aspirate 3 to 5 units of insulin
extra depending on the size of the air
bubbles.
Hold the InsuJet™ with the insulin
cartridge/vial upwards and the dose
window facing you. On the front body of
the InsuJet™ you will notice the number ‘2’
with a directional arrow (2).
Disconnect the insulin cartridge/vial and
adaptor from the InsuJet™, snap theadaptor away from the nozzle and piston
and remove.
-
8/18/2019 InsuJet Instructions
9/14
16
Keeping the InsuJet™ upright. Tap the
nozzle with your nger tips, collecting all
air bubbles close to the opening.
Do not attempt to wind theInsuJet™ beyond the 50 unitsindication.
IMPORTANT
Do not attempt to wind the InsuJet™
beyond the 50 units indication. Attemptingto dismantle the InsuJet™ may cause
serious injury and may also damage the
InsuJet™ beyond repair.
Fold the tissue to your index nger and
keep it beside the opening of the nozzle.
Wind back to your required dose of insulin.
The air bubble and the excess of insulin
will flow through the hole in the nozzle.
Gently wipe the excess of insulin witha tissue from the side. Always read the
number of units in the dose window to
check you have the correct dose before
administration.
-
8/18/2019 InsuJet Instructions
10/14
18
ADMINISTER INSULIN WITH THE InsuJet™
STEP 1
Select your administration site. See the
‘InsuJet™ Additional information’ for
administration sites and administration
techniques.
To administer your insulin, follow steps 1 to
7 below:
STEP 2
Slide the red safety catch as far as it will
go and hold it in place.
STEP 3
Place the end of the nozzle and piston
against your chosen administrationsite, ensuring that the InsuJet™ is held
at right angles (90°) to the skin at the
administration site.
IMPORTANT
Ensure that the InsuJet™is held perpendicular to theadministration site.
-
8/18/2019 InsuJet Instructions
11/14
20
5 sec
STEP 5
Continue to hold the nozzle and pistonrmly against the skin for 5 seconds.
STEP 4
While you are holding the red safety catch
in place, push the back body towards
your skin. This will result in releasing the
safety mechanism and nally, in the insulin
administration.
STEP 7
Place the clear adaptor cover back onthe adaptor until the next time. Store your
insulin according the guidelines of your
insulin supplier.
STEP 6
Remove the InsuJet™ from the skin and
release the red safety catch.
-
8/18/2019 InsuJet Instructions
12/14
22
Remove the nozzle cap from the nozzle
and piston.
REMOVING THE NOZZLE AND PISTON FOR REPLACEMENT
Pull the nozzle out of the InsuJet™ with a
straight rm pull.
While sliding the nozzle lock catch
upwards, turn the nozzle lock to the right
until the nozzle lock arrow points to the
padlock unlocked position.
-
8/18/2019 InsuJet Instructions
13/14
24
Remove the piston. Discard the used
nozzle and piston with your regularhousehold waste.
Remove the piston of the InsuJet™ by
pushing it rmly to one side.
-
8/18/2019 InsuJet Instructions
14/14
012000-01-2.0
Diabetes Management International BVBeechavenue 127, Euro Ofces Building 3rd Floor,
1119 RB Schiphol Rijk, The NetherlandsTel: +31 (0) 20 316 01 [email protected] | www.insujet.com