Instruction Manual Gel Doc and Chemi Doc
Transcript of Instruction Manual Gel Doc and Chemi Doc
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
1/50
Molecular Imager
Gel Doc XR System
Molecular Imager
ChemiDoc XRS System170-8170 170-8070
170-8071
Hardware Instruction Manual
10014084 Rev A
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
2/50
2
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
3/50
3
Important
Please read these instructions before attempting to install or operate the Molecular Imager Gel Doc XR
or ChemiDoc XRS.
Note: This instrument is suitable for research use only.
It must be used, therefore, only by specialized personnel that know the health risks associated with UV
radiation, and with the reagents that are normally used with this instrument. Use of the acrylic screen doesnot guarantee protection of the user from UV radiation. The use of protective eyeglasses or mask and
gloves, is strongly recommended.
Wichtig
Bitte lesen Sie die Anweisungen und machen Sie sich mit der Bedienungsweise vertraut, bevor Sie den
Hood benutzen.
Anmerkung: Dieses Geraet ist nur fuer Forschungszwecke geeignet.
Ausserdem ist die Benutzung nur fuer spezialisiertes Personal gedacht, das mit den Gesundheitsrisiken
vertraut ist, die an die UV-Strahlung gebunden sind und den Reagentien, die normalerweise mit diesem
Geraet benutzt werden.Die Benutzung eines Plexiglasschutzschildes garantiert dem Benutzer keinen Schutz
vor UV-Strahlung. Die Benutzung von Schutzbrille oder Schutzmaske ist strengstens empfohlen.
Warranty
The Gel DocXR and ChemiDoc XRS systems are warranted against defects in materials and workmanship
for 1 year. If any defect occurs in the instrument during this warranty period, Bio-Rad Laboratories, Inc
will repair or replace the defective parts at its discretion without charge. The following defects, however,
are specifically excluded:
Defects caused by improper operation Repair or modification done by anyone other than Bio-Rad Laboratories, Inc. or an authorized agent Use of spare parts supplied by anyone other than Bio-Rad Laboratories, Inc. Damage caused by accident or misuse Damage caused by disaster Corrosion caused by improper solvents or samples
Garantie
Die Garantie fuer den Hood betraegt 1 Jahr auf Herstellungs- und Materialfehler.
Bei Auftreten von Fehlern waehrend der Garantiezeit repariert oder ersetzt Bio-Rad Laboratories, Inc. die
fehlerhaften Teile auf eigene Kosten. Die folgenden Schaeden sind in jedem Falle ausgeschlossen: Schaeden durch unsachgemaesse Bedienung bedingt Instandsetzungen oder Veraenderungen durch nicht authorisiertes Personal von Bio-Rad
Laboratories, Inc.
Benutzung von Ersatzteilen, die nicht von Bio-Rad Laboratories, Inc. geliefert wurden Schaeden durch Naturkatastrophen verursacht Schaeden durch unsachgemaessen und fehlerhaften Gebrauch verursacht Korrosionsschaeden durch ungeeignete Loesungen oder Proben
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
4/50
4
Regulatory Notice
IMPORTANT: The Gel Doc XR and ChemiDoc XRS systems are designed and certified to meet
EN61010, the internationally accepted electrical safety standards, EMC regulations, and TUV. Certified
products are safe to use when operated in accordance with the instruction manual. This instrument should
not be modified or altered in any way. Modification or alteration of this instrument will:
Void the manufacturers warranty Void the regulatory certifications Create a potential safety hazard
NOTE: Bio-Rad Laboratories, Inc. is not responsible for any injury or damage caused by use of this
instrument for purposes other than those for which it is intended or by modifications of the instrument not
performed by Bio-Rad Laboratories, Inc., or an authorized agent.
Richtlinien
WICHTIG: Dieses Geraet, von Bio-Rad Laboratories, Inc. konstruiert, besitzt das Zertifikat ENC1010
anerkannt von dem internationalen elektrischen Sicherheitsstandard EMC und unterliegt dessen Richtlinienund TUV. Die Geraete sind sicher im Gebrauch, wenn die vorliegende Bedienungs- anleitung beachtet wird
Dieses Geraet darf auf keinen Fall veraendert werden. Eventuelle Veraenderungen fuehren zu:
Ungueltigkeit der Herstellergarantie Ungueltigkeit des Zertifikats Eventuelles Sicherheitsrisiko
Bemerkung: Bio-Rad Laboratories, Inc. haftet nicht fuer Verletzungen an Personen oder Sachschaeden,
hervorgerufen durch eine nicht vorhergesehene Benutzung des Geraetes. Ebenso uebernimmt Bio-Rad
Laboratories, Inc. keine Haftung fuer Veraenderungen, die von nicht authorisierten Personen durchgefuehrt
wurden.
General Precautions
Please read the instruction manual carefully The instrument must be used only for the intended purpose of gel documentation in research
laboratories
The instrument must be connected to a power source line which MUST BE grounded and protectedby a circuit breaker
Do not pour liquids directly on or inside the instrument Switch off all the lights immediately after use Clean the transilluminator platen after use
Generelle Vorsichtsmassnahmen
Bitte die Bedienungsanleitung aufmerksam lesen Das Geraet darf nur fuer die vorgesehenen Applikationen eingesetzt werden, d.h.
Geldokumentation in Forschungslaboratorien
Fuer die Stromversorgung des Hood muss eine geerdtete Steckdose benutzt werden Keine Fluessigkeiten auf oder in das Geraet giessen Die Lampe sofort nach dem Gebrauch ausschalten Den UV-Filter des transilluminators nach Gebrauch reinigen
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
5/50
5
TABLE OF CONTENTS
SECTION 1 INTRODUCTION ......................................................................................................................................................6
1.1 COMPUTERREQUIREMENTS .......................................................................................................................................................6
SECTION 2 IMPORTANT SAFETY INFORMATION..............................................................................................................7
2.1IMPORTANTNOTICE....................................................................................................................................................................7
2.2POWER SAFETYINFORMATION.....................................................................................................................................................7
SECTION 3 PRODUCT DESCRIPTION......................................................................................................................................8
3.1CCDCAMERA ANDLENSES .........................................................................................................................................................8
3.2DARKROOMENCLOSURE,EMISSIONFILTERS, ANDILLUMINATIONSOURCES .................................................................................8
3.3USB AND FIREWIRECONNECTIONS .............................................................................................................................................9
3.4QUANTITYONESOFTWARE..........................................................................................................................................................9
3.5PRINTER (OPTIONAL) ..................................................................................................................................................................9
3.6UNPACKING THESYSTEM...........................................................................................................................................................10
SECTION 4 GETTING STARTED..............................................................................................................................................14
4.1SELECTING ALOCATION............................................................................................................................................................14
4.2ASSEMBLING THESYSTEM..........................................................................................................................................................14
4.2.1 ASSEMBLING THE UNIVERSAL HOOD II...................................................................................................................14
4.3GELDOCXRSYSTEM...............................................................................................................................................................154.3.1INSTALLING THEGELDOCXRCAMERA .......................................................................................................................15
4.3.2INSTALLATION OFQUANTITYONESOFTWARE FOR GELDOCXR...................................................................................16
4.3.3 GELDOCXRSYSTEMCABLECONFIGURATION............................................................................................................20
4.4CHEMIDOC XRSSYSTEM ........................................................................................................................................................22
4.4.1INSTALLATION OFQUANTITYONESOFTWARE FOR CHEMIDOCXRS ..............................................................................23
4.4.2CHEMIDOCXRSSYSTEMCABLECONFIGURATION........................................................................................................28
4.4.3FIREWIRECARDINSTALLATION....................................................................................................................................29
4.5INSTALLING THEWHITELIGHTTRANSILLUMINATOR (OPTIONAL)................................................................................................294.6INSTALLING THEOPTIONAL 17MM OR 25 MMWIDE-ANGLELENSES (CHEMIDOCXRSSYSTEM ONLY) ........................................30
4.7DESCRIPTION OFSYSTEMFUNCTION AND SYSTEMINITIALIZATION..............................................................................................32
4.7.1CONTROL PANEL .........................................................................................................................................................32
SECTION 5 OPERATION OF THE UNIVERSAL HOOD II ..................................................................................................39
5.1OPERATING THEUNIT...............................................................................................................................................................39
5.1.1SWITCH ON THEUNIVERSALHOODII...........................................................................................................................39
5.1.2POSITIONYOUR GEL....................................................................................................................................................39
5.1.3ACQUIRING ANIMAGE..................................................................................................................................................39
5.1.4ACQUIRING ANIMAGEWITHFLATFIELDING ................................................................................................................39
5.2CUTTING GELS .........................................................................................................................................................................41
SECTION 6 TROUBLESHOOTING...........................................................................................................................................42
SECTION 7 ACCESSORIES AND REPLACEMENT PARTS ................................................................................................43
SECTION 8 MAINTENANCE AND PARTS REPLACEMENT..............................................................................................45
8.1EPIILLUMINATIONLAMPREPLACEMENT..................................................................................................................................458.2 FUSEREPLACEMENT................................................................................................................................................................45
8.3UVTRANSILLUMINATOR............................................................................................................................................................46
8.4LAMPREPLACEMENT................................................................................................................................................................46
APPENDIX A: TECHNICAL SPECIFICATIONS ....................................................................................................................48
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
6/50
6
Section 1 Introduction
The Bio-Rad Gel Doc XR and ChemiDoc
XRS gel documentation systems are easy-to-use, high-
performance systems. They use a charge coupled device (CCD) camera to capture images in real time,
which allows you to more accurately position and focus the image. With Bio-Rad Quantity One software
acquired images can be optimized, annotated, analyzed and printed.
The Bio-Rad Gel Doc XR and ChemiDoc XRS systems utilize an enclosure (Universal Hood II) that islight-tight and contains UV illumination and white light illumination. Please note that both the Gel Doc XR
and ChemiDoc XRS cameras have external power supplies.
1.1 Computer Requirements
This software will run under Windows XP, Vista Ultimate, or Mac OS X. The amount of computer memory
required for using the program is mainly determined by the size of the images you will scan and analyze.
Images scanned at high resolution can be quite large. For this reason, we recommend that you archive
images on a network file server or removable storage media.
Specifications PC MAC
Operating system Windows XP or Vista Ultimate Mac OSX 10.5.x
Base unit Intel Core 2 Duo, minimum 2.0
GHz
Power PC G4 or better
RAM >2 GB >2GB
Cache memory >2 MB > 2 MB
Hard disk space >120 GB >120 GB
Video card 128 MB video memory minimum,
capable of generating minimum
1024 x 768 pixels
128 MB video memory minimum,
capable of generating minimum
1024 x 768 pixelsCamera connection ports One full height empty PCI slot
required (ChemiDoc XRS only)
Available FireWire port
(ChemiDoc XRS only)
Communication ports 4 USB 2.0 ports 4 USB 2.0 ports
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
7/50
7
Section 2 Important Safety Information
2.1 Important Notice
Both the Gel Doc
XR and ChemiDoc
XRS imaging systems are intended for laboratory use only
Use of the Gel Doc XR, and ChemiDoc XRS systems involve UV illumination. Proper precautions
must be taken to avoid eye and skin exposure to the UV radiation. This instrument is meant for use only
by trained personnel that know the health risks associated with UV radiation and with reagents that arenormally used with this instrument. The acrylic shield provides some UV protection. However, it does
not guarantee complete protection, and it is designed to shield only the person working in front of the
system.
WARNING: The operator should wear appropriate safety glasses, or a
protective mask and gloves, in addition to using the UV safety shield provided
with this instrument.
Anmerkung: Dieses Geraet ist nur fuer Forschungszwecke geeignet.
Ausserdem ist die Benutzung nur fuer spezialisiertes Personal gedacht, das mit den Gesundheitsrisiken
vertraut ist, die an die UV-Strahlung gebunden sind und den Reagentien, die normalerweise mit diesemGeraet benutzt werden.
Die Benutzung eines Plexiglasschutzschildes garantiert dem Benutzer keinen Schutz vor UV-Strahlung
Die Benutzung von Schutzbrille oder Schutzmaske ist strengstens empfohlen.
2.2 Power Safety Information
a) Voltage Setting Information: The Universal Hood II has a power supply that automatically chooses thecorrect voltage for your country or region.
b)Fusing: The Universal Hood II has two user-serviceable fuses, F1 and F2, which are located on thebottom rear panel and are a part of the power entry module (see below).
Main power entry module
Fuse F1 and F2 are located in
this drawer. Use a
screwdriver and pull the
drawer out to see the fuses.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
8/50
8
Section 3 Product Description
3.1 CCD Camera and Lenses
A CCD camera is placed on top of the Universal Hood II for capturing images. The camera comes with
a motorized zoom lens (MZL) that allows a remote adjustment of the lens control functions (zoom
focus, and iris).
For ChemiDocXRS systems optional lenses with low f-numbers are available that offer extreme light
collection efficiency for low-light applications like chemiluminescence.
Lens f/0.95, 25 mm, wide-angle catalog #170-8072
Lens f/0.95, 17 mm, wide-angle catalog #170-8073
3.2 Darkroom Enclosure, Emission Filters, and Illumination Sources
The Universal Hood II is designed to capture fluorescent and chemiluminescent images without using a
photographic darkroom. The enclosure has built-in white light epi-illumination and UV
transillumination. An optional white light transilluminator (catalog #170-7950) or a white light
conversion screen (catalog #170-8001) may also be used with the system.
For easy sample loading, the UV transilluminator is located in the drawer of the Universal Hood II and
can be accessed from the front of the enclosure. For your convenience, the lights in the darkroom
enclosure turn off automatically after about 15 minutes. This time period can be extended indefinitely
by pressing theHoldbutton.
The Universal Hood II has a three-position filter slider that offers the flexibility of using two different
emission filters for fluorescent applications as well as the ability to image chemiluminescent samples
(no filter).
Bio-Rad Laboratories, Inc. offers a selection of filters and illumination sources.
Standard Filter Comes with each system. Catalog #580AF120 62 mm Ethidium Bromide filter 170-8081
Optional Filters Can be purchased for optimal performance.520DF30 62 mm Optional filter for SYBR Green/SYBR Gold/GFP/
fluorescein 170-8074
560DF50 62 mm Optional filter for Cy3/rhodamine 170-8075
630BP30 62 mm Optional filter for SYPRO Ruby 170-8076480BP70 62 mm Optional filter for Hoechst/coumarin 170-8077
Optional Illumination SourcesWhite light transilluminator 170-7950
White light conversion screen 170-8001
302 nm UV lamps, 6 (shipped standard with each system) 170-8097
254 nm UV lamps, 6 170-8098
365 nm UV lamps, 6 170-6887
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
9/50
9
3.3 USB and FireWire Connections
The Universal Hood, common to both Gel Doc XR and ChemiDoc XRS systems, requires a USB
connection in order to communicate with the PC. The Gel Doc XR camera has a USB connection that
connects to the Hood. However, the ChemiDoc XRS camera is directly connected to the PC and will
need a FireWire port to communicate with the PC, which is where the user interface software is
installed. In case of PCs that do not offer a FireWire (IEEE1394) interface, in order to connect the
ChemiDoc XRS system, a FireWire card (included as an accessory) needs to be installed in the hostcomputer to communicate with the camera.
Note: Read and follow the instructions for installing the USB cable (Gel Doc XR) or FireWire
cable (ChemiDoc XRS), card, drivers, and software in the following sections.
3.4 Quantity One
Software
Each system ships with a full version of Quantity One1-D Analysis Software and unlimited copies of
Quantity One Basic software for data sharing. The software can be used to annotate and document
images, analyze molecular weights, print, and perform a host of other applications. See the software
manual for detailed instructions on how to install and operate the software.
Note: Read and follow the instructions in the following pages prior to installing Quantity One
software.
3.5 Printer (Optional)
For your convenience, Bio-Rad offers an optional USB printer for use with the Gel Doc XR and
ChemiDoc XRS systems:
Mitsubishi P93DW Thermal Printer catalog #170-8089
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
10/50
10
3.6 Unpacking the System
Gel Doc XR ComponentsThe Gel Doc XR and ChemiDoc XRS systems consist of multiple components. Each of them may be in
a separate box or in one box with a label describing the catalog numbers and its contents. Please make
sure that all the system components are in your shipment. Please unpack each component carefully and
verify that you have the correct one. Each system includes the following:
Catalog # Descriptions
170-8170 Gel Doc XR System PC/Mac170-8162 Universal Hood II with TLUM 100/240 V
170-8169 Gel Doc XR Camera w/MZL(motorized zoom lens)
170-8080 Kit Accessory, Universal Hood II
170-9600 Quantity One Software PC/Mac
Universal Hood II with TLUM 100/240V
Gel Doc XR camera with MZL & power supply
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
11/50
11
6 ft USB cable
10 ft USB cable
580 AF 120 Band Pass Filter (EtBr)
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
12/50
12
Quantity One software
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
13/50
13
ChemiDoc XRS Components
Catalog # Descriptions
170-8070 / 1708071 ChemiDoc XRS System PC/Mac170-8162 Universal Hood II with TLUM 100/240 V
170-8099 ChemiDoc XRS camera with MZL (including
camera controller cable and power supply)
170-8173 FireWire card (PC system only)170-8080 Kit accessory, Universal Hood II
170-8008 Fluorescent reference plate
170-9600 Quantity One software PC/Mac
170-8177 FireWire cable
ChemiDoc XRS Camera with MZL and Power Supply
Firewire card
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
14/50
14
Section 4 Getting Started
4.1 Selecting a Location
The Gel Doc
XR and ChemiDoc
XRS systems can be placed on any benchtop. The best position is
near the computer where Quantity One is operating. Remember to leave space to easily access the
power switch on the Universal Hood II, which is located at the rear bottom-left side of the enclosure.
4.2 Assembling the System
4.2.1 Assembling the Universal Hood II
Important: It is recommended that the Universal Hood II be carried by at least two people
holding the instrument from the bottom side. Be sure that the door and the drawer are
completely closed.
Wichtig: Es wird empfohlen, dass mindestens zwei Personen den Hood transportieren und dabei
das Geraet am Geraeteboden halten.
1. Lift the top lid off the main box.2. Open the accessory box and review the unpacking instructions.3. Lift off the top foam.4. Lift the cardboard enclosure off the pallet.5. Remove the plastic wrapping and carefully lift the enclosure from the bottom foam, and place it
in a suitable location.
Note: Please do not connect the power cable to a power source until all connections are made.
The power source has to be grounded and protected by a circuit breaker.
6. Locate the band pass filter in the accessory kit and thread it onto one of the two 62 mm rings ofthe Filter Slider (inside of the hood). The filter slider facilitates the selection between two
filters and an open position for chemiluminescence.
7. The three selectable positions are indicated by NO FILTER, FILTER 1, or FILTER 2.on the right side of the camera support. Note the position of the 580 AF 120 Ethidium Bromide
filter for future use. See picture below.
Three positions sliding filter
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
15/50
15
4.3 Gel Doc XR SystemIf you have purchased the ChemiDoc XRS System, proceed to page 22 for installation insructions
4.3.1 Installing the Gel Doc XR Camera
1. Place the camera on the bench sideways as shown. Take out the lens cap and put it a safe place.Take out the thumb screw from the box and assemble it to the camera bracket exactly as shown.
2. Place the thumb screw as shown.
Gel Doc XR camera with bracket, MZL, and thumb screw
3. Place the Camera over the Universal Hood II by inserting the lens into the hole in gasket andseating the lens on the gasket while the thumb screw is positioned on the camera bracket as
shown. The thumb screw slides through the slot in the vertical camera bracket on the Hood.
4. Secure the thumb screw (see pictures below). Make sure that the camera/lens is fully seated onthe gasket and is upright.
5. The camera position can be fine adjusted in all directions; loosen the two Phillips head screwsand move the entire assembly, along with the bracket.
Closeup of Gel Doc XR camera mounted with thumb screw
Phillips head
screws
Thumb screw
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
16/50
16
4.3.2 Installation of Quantity One Software for Gel Doc XRQuantity One ships with the CD and the HASP key.
HASP key
For a PC computer Quantity One installation
1. Insert CD for Quantity One.2. When the Installer launches, select Install Quantity One.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
17/50
17
3. The Select Features screen has Quantity One and HSK Support selected as default.4. In addition, select Gel Doc XR drivers.
5. Click on the pull down arrow, next to X on Gel Doc XR Drivers.6. Select Will be installed on local hard drive.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
18/50
18
7. Finish Installation.8. Insert the HASP key in the USB port.9. Restart the computer.
For MAC computer Quantity One installation
1. Insert CD. It appears as an icon on the desktop.2. Open the CD by clicking on the icon.3. Select Install Quantity One.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
19/50
19
4. Select Custom Install.
5. Select Quantity One and Gel Doc XR drivers, users guide is optional.
6. Click on Install.7. Insert the HASP key in the USB port.8. Shut down and restart the computer.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
20/50
20
4.3.3 Gel Doc XR System Cable Configuration
Camera connections
Connecting the cabling harnesses
1. Locate the 6 ft USB cable. Connect one end of the USB cable connector to the USB port on thecamera. Connect the other end to the corresponding port on the rear panel on the Universal
Hood that says TO CAMERA.
2. The camera power supply cable should be plugged into the CCD camera power connector.3. Connect the MZL (Motorized Zoom Lens, page 13) cable connector to the connector labeled
LENS on the rear panel of the Universal Hood.
4. Connect the 10 ft USB cable from the Universal Hood to the PC.5. The main power cable that connects the Universal Hood to the power outlet.
The following table describes the cable/connectors and their designations.
Label Instrument/Connection
10 ft USB connector/cable From hood to PC
6 ft USB connector/cable From camera to hood
Camera power supply cord/power supply Camera/PWR connector
MZL (motorized zoom lens) cable Hood/LENS Din connector
Power cable Connects the Hood to the power outlet on
the wall
See the cabling diagram on the following page.
Power supply
connectorUSB connector
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
21/50
21
Figure 1
This completes the installation of the Gel Doc Camera and Quantity One
software. Proceed to Section 4.7 on page 32 for a description and initial test
for the Universal Hood II.
Power cable
from power
supply
MZL cable
USB cable
USB cable
To main power outlet
110/240 VAC
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
22/50
22
4.4 ChemiDoc XRS System
Unpacking the system was covered in Section 3.6.
1. Place the camera on the bench so that the locking hole of the tripod mount is facing you.2. Thread the thumb screw on the locking tripod hole. Turn twice.3. Remove the lens cap and put it in a safe place.4. Place the filter assembly in the NO FILTER position.5. Place the ChemiDoc XRS camera/lens (MZL: motorized zoom lens) on the mounting plate, on
top of the enclosure, by inserting the lens through the hole in the gasket while the screw slides
through the camera bracket slot. Make sure that the lens is fully seated on the gasket.
6. Tighten the screw against the camera bracket. See the following pictures.
ChemiDoc XRS camera with thumb screw
7. Fix the camera bracket. Tighten the screw to lock the camera against the camera bracket.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
23/50
23
4.4.1 Installation of Quantity One Software for ChemiDoc XRS
Installation of Quantity One on a PC
1. Insert the Installation CD.2. The Select features screen has Quantity One and HSK Support selected as a default.3. In addition to these defaults, select the ChemiDoc XRS driver.
4. Finish the installation wizard.5. Insert the HASP key in the USB port.6. Shut down the computer.
When you restart and click on Quantity One, you will see a Welcome screen.
You can either click on the registration form button to send your registration online, or call Bio-Rad Tech
Support with your system ID to register.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
24/50
24
To register the software, you need the software serial number and the system ID.
The Quantity One serial number is on the outside of the box and on the CD jacket.
Format of serial number: BRQ1AXXXXXX
To get the system ID :
Click on Enter Password in the Welcome page.
The system ID is seen in the Enter Password window.
Once you register online or by phone, you get a password. Enter the password in the Quantity One field.
The Password is not case sensitive and can be entered without spaces. If the password is correct, a Green
OK is seen. Click Done to finish.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
25/50
25
Installation of Quantity One On MAC
1. Insert the CD. It appears as an icon on the desktop.2. Click on the CD icon.3. Select Install Quantity One.
4. Select Custom Install.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
26/50
26
5. Select Quantity One and ChemiDoc XRS drivers.
6. Click on Install.7. Insert the HASP key in the USB port.8. Shut down and restart the computer.
When you click on newly-installed Quantity One software, you will see a Welcome screen.
You can click on the Registration Form button to register online, or call with the system ID to register.
To register the software, you need the software serial number and the system ID. The serial number of
Quantity One is on the outside of the box and on the CD jacket.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
27/50
27
Format of serial number: BRQ1AXXXXXX
To get the system ID :
Click on Enter password in the Welcome page.
The system ID is seen in the Enter Password window that opens.
Once you register online or by phone, you get a password. Enter the password in the Quantity One field.
Password is not case sensitive and can be entered without spaces. If the password is correct, a Green OK is
seen. Click Done to finish.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
28/50
28
4.4.2 ChemiDoc
XRS System Cable Configuration
Connecting the cabling harnesses.
1. Locate the FireWire cable. Connect one end of the FireWire cable connector to the FireWireconnector on the camera.
2. Connect the other end to the FireWire connector on the PC/Mac.3. The camera power supply cable should be plugged into the CCD camera power connector.4. Connect one end of the USB cable to the connector labeled TO PC on the hood, and the other
end to the USB port on the PC.
5. Connect the MZL cable connector the extension cable. Connect the other end of the extensioncable to the connector labeled LENS on the rear panel of the universal hood.
The following table describes the cable connectors and their designations.
Label Instrument/Connection
6-pin FireWire cable/connector FireWire connector from camera to PC
Camera power supply cord/power supply Camera/PWR connector10 ft USB cable From Hood to PC
MZL (motorized zoom lens) cable LENS Din connector to extension cable din connector
MZL extension cable Extension cable din connector to Hood connector
Power cable Connects the Hood to the power outlet on the wall
Power cable
from power
supply
MZL cables
USB cable
FireWire cable
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
29/50
29
4.4.3 Firewire Card InstallationPlease follow the instructions below to install the FireWire card into your PC or Mac.
ChemiDoc XRS PC:1. Install the FireWire (IEEE 1394) adaptor card into the PC if necessary. Make sure that the computer
is powered off and the power cord is disconnected from the PC while installing.
2. Turn on the computer.3. The Windows XP or 2000 operating systems will recognize the card and automatically install the
appropriate driver for it.
ChemiDoc XRS Mac:Current Mac systems generally have onboard FireWire ports (connectors). The ChemiDoc XRS FireWire
cameras can be plugged directly into any one of these connectors, and the Bio-Rad software interface will
be able to communicate with the camera via this port.
4.5 Installing the White Light Transilluminator (Optional)
1. To install the white light transilluminator, open the enclosure door and pull the drawer toward you.2. Locate the power cable positioned on the inside left side of the enclosure, behind the drawer slide.3. Remove the black rubber that covers the banana plug and insert it into the white light
transilluminator outlet. Make sure that the main power switch on the transilluminator is in the ON
position. The transilluminator will not illuminate, because power to the white light transilluminator
is controlled from the membrane touch pad.
4. To use the white light transilluminator, place it horizontally on the UV transilluminator and press theTransWhite button on the membrane touch pad.
When the white light transilluminator is not in use, store it in the two plastic holders located at the rear of
the enclosure, as it stands against the back wall.
Storage of the white light transilluminator
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
30/50
30
4.6 Installing the Optional 17 mm or 25 mm Wide-Angle Lenses
(ChemiDoc XRS System only)
This installation guide covers installation of the optional lenses for the ChemiDoc XRS system. The
following catalog numbers are covered under this category:
a.Lens f 0.95, 25 mm, wide-angleb.Lens f 0.95, 17 mm, wide-angle
The lens kit includes a lens and the light seal adaptor ring. See pictures below:
Setting Up the Lens:
1. Turn off the power to the enclosure and camera.2. Disconnect the camera and lens cables.3. Remove the ChemiDoc XRS camera from the bracket.4. Remove the existing motorized zoom lens from the camera by turning the lens counterclockwise.
Attach the 25 mm or 17 mm lens to the camera as shown below:
Remove existing motorized zoom lens Install the wide angle lens
17 mm wide-angle lens with
light seal adaptor
25 mm wide-angle lens with
light seal adaptor
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
31/50
31
5. The wide-angle lenses come with an adapter. See picture below.
6. Place the adapter ring on the mounting plate.7. Position the camera lens assembly on the adapter ring while sliding the thumb screw through the
slot on the camera bracket. Before tightening the screw, make sure that the wide-angle lens is
seated on the gasket of the adapter ring, and that the back of the camera closely faces the camera
bracket.
8. Tighten thumb screw. See pictures below:
Make sure that the lens is seated properly on the gasket in the adapter.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
32/50
32
The wide angle lenses have an iris and focus ring that allows users to adjust for optimum light collection
and focal plane. See image below:
For imaging of chemiluminescent samples, make sure that the iris is opened all the way.
4.7 Description of System Function and System Initialization
4.7.1 Control Panel
The front membrane touch pad control panel of the Universal Hood II allows full control of the motorized
zoom lens, UV transillumination, white epi illumination and white light transillumination. Please read the
following section to become familiar with each function.
Membrane touch pad
Epi White LED
and button
Trans White
LED and
button
Hold LED and
button
Trans UV LED
and button
Prep UV LED and
button
Power indicator
LED
Imaging indicator
LED
Focusing ring Iris control ring
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
33/50
33
ThePower light tells you whether the system is turned on. The power on/off switch is located at the rear
bottom-left side of the enclosure.
TheImaging LED blinks when you are acquiring an image with the camera indicating integration by the
Camera. It also blinks if the computer is turned off.
Light Sources
The TransUV button controls the UV transilluminator with full intensity. The transilluminator will
automatically turn off after 15 minutes unless the Holdbutton is activated. For safety purposes, this switch
is subject to the following automatic controls:
It is automatically turned off when the front door is open It is automatically turned off when the drawer is open
If the door or drawer is opened while the transilluminator is ON, the blinking LED will warn you that the
transilluminator has been turned OFF. After the door or the drawer is closed, you have to press the UV
Trans switch again to turn it ON.
Important:For the purposes of band cut applications, it is possible to turn ON the UV
light with the drawer fully open. You must wear all possible UV protections, especially for your eyes,
when the transilluminator switch is turned ON with the drawer fully open. The UV shield has to be
used but glass or mask and gloves are recommended to block the UV radiation. This option is not
applicable to the door.
Wichtig: Wer die Absicht hat, Banden aus dem Gel zu schneiden, kann bei eingeschaltetem UV-Licht
die untere Schublade oeffnen. Sie muessen Schutzkleidung und die notwendigen UV-
Schutzmassnahmen tragen, besonders Augenschutz, wenn der transilluminator bei offener Schubladeeingeschaltet ist. Das Plexiglasschutzschild muss benutzt werden, ebenso Schutzbrille oder
Schutzmaske und Einmalhandschuhe, um sich vor der UV-Strahlung zu schuetzen. Diese
Moeglichkeit ist von der Geraetetuer aus nicht anwendbar.
The Prep UVswitch is used to decrease the light output of the UV Transilluminator. The UV Preparative
function is a lower intensity light designed to minimize the effects of UV exposure on DNA. The
Preparative mode is recommended for applications that require longer UV exposure times. In order to
activate the UV Prep switch, the UV Trans button must be on.
TheEpi White button controls the epi white illumination. Press the button to turn ON epi-white; press the
button again to turn it OFF. Epi illumination will automatically turn off after 15 minutes, unless the Holdbutton is activated. When the UV transilluminator switch is activated,Epi White is automatically turned off.
The Trans White Light button controls the white light transilluminator when installed (optional). It will
automatically turn off after 15 minutes, unless theHoldbutton is activated.
If the Conversion Screen option is selected, place the Conversion Screen on the sample area, select Trans
White Light. It turns on the Trans UV.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
34/50
34
The Hold key disables the automatic shut-off of the UV and White light transilluminators and the epi
illumination; the lights will remain active until the hold status is cancelled. When the Holdkey is pressed
the orange LED will light up, indicating that theHoldfunction is active.
If opening the door or drawer (activating the UV interlock, which turns off the UV), you will need to turn
off theHoldbutton before pressing any other switch. The control panel has a second section that includes
the buttons to run all the MZL functions (see the Lens Control Section below).
Lens Control
The membrane pad has a Fast and a Slow button to obtain rapid or fine adjustments of the lens. An image
can be optimized with the following lens adjustments:
Aperture is adjusted using the Iris buttons on the membrane touch pad. The size of the aperture opening
allows the CCD sensor to take in more or less light. The Iris has two buttons. One button, Close (-), reduces
the size of the aperture and decreases the amount of light coming into the sensor, thereby making the image
darker. The other button, Open (+), opens the aperture and increases the amount of light coming into the
sensor, resulting in a brighter image.
Focus is adjusted using the Focus buttons on the keypad. Pushing theNear (+) button or theFar (-) button
changes the focal point of the lens and allows the user to focus on the sample.
Zoom is adjusted using the Zoom buttons on the keypad. These buttons allow you to change the size of the
image on the screen. Pushing the Wide (-) button zooms out (far field). Pushing the Tele (+) button zooms
in (close up).
Membrane touch pad Lens control section
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
35/50
35
Lens and Light Source Control Through the Computer
The functions described in the previous paragraphs can also be performed through the PC. The attached
screen shot as applied to Gel Doc XRS is self-explanatory.
Filters: A +1 diopter is factory installed on the lens. This diopter should always remain on the lens
assembly. A 580AF120 Ethidium Bromide filter is included in the accessory kit. Place this filter in
position 1 or 2 of the filter slider (see 4.2.1). The Ethidium Bromide filter should be used for fluorescence
(UV) and colorimetric (white light) applications.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
36/50
36
Please follow the procedure in the table to ensure that the Universal Hood II is functioning properly.
STEP 1. Initial test for the Universal Hood II:
Procedure Control Panel
Make sure the door and drawer are closed and the computer is switched on.
Turn on the system Power LED turns on after short blinking.
Wait 30 minutes for camera to warm up Nothing will change except the camera will warm up.
Press the Epi White key Epi-white LED turns on.
Open the door Epi-lights are on, Trans UV LED blinks.
Press the Epi White key again
Close the door
Epi-lights and LED are turned off.
Trans UV LED turns off.
Press the Trans UV key Trans UV LED turns on.
Open the door Trans UV LED blinks.
Close the door Trans UV LED turns off.
Press the Trans UV key Trans UV LED turns on.
Open the drawer half way Trans UV LED blinks.Close the drawer Trans UV LED turns off.
Press the Trans UV key Trans UV LED turns on.
Press the Prep UV key Prep UV LED turns on.
Press the Hold key Hold LED turns on.
Open the drawer Trans UV LED blinks, Prep UV LED and
Hold LED turns off.
Close the drawer Trans UV LED turns off.
Note: The light source buttons on the control panel are disabled when the Imaging LED blinks
(indicating that the computer is switched off).
STEP 2. Procedure for initialization of the camera and lens assembly both (for GelDoc XR/
ChemiDoc XRS systems)1. Press the epi iIllumination button.2. Open the door and determine if lights are on.3. Place the focusing target on the UV transilluminator.4. Double click on Quantity One software icon on the desktop. From File Menu, select Gel Doc XR
or ChemiDoc XRS.
Note: Dark Reference comes up on first use for both GelDoc XR system. Please refer to the step 3
for details about reference file image acquisition.
5. Select Live/Focus. This applies to Chemi-Doc XRS System only, Gel Doc XR does thisautomatically.
6. An image of the target will appear on the computer screen. Using the lens control buttons on themembrane touch pad or the control panel that appears in the software, find the best Iris, Focus, and
Zoom conditions.
Note: Working with a widely open Iris and changing the Zoom may require you to adjust the Focus.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
37/50
37
STEP 3. Reference file acquisition:
Please note that the steps described below must be performed in either of the following scenarios:
When installing the system with a new computer or camera When software is first installed or reinstalled for any reason
1. Make sure that all lights in the Universal Hood II are off.2.
Make sure that the lens cap is on the lens. This is critical!3. Power ON the camera and the Universal Hood II.
4. Wait 30 minutes for the system to warm up and stabilize.5. Double click on Quantity One software icon on the desktop.6. From File Menu, select Gel Doc XR, OR ChemiDoc XRS.
TDS Quantity One Window Select GelDoc XR
OR ChemiDoc XRS from file
menu
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
38/50
38
The following window will open with a notification that the system will be acquiring reference files,
depending on the imager:
a. Gel Doc XR
Note: No lights should be turned ON. Please allow approximately 30 seconds for the acquisition
of the reference file.
b. ChemiDoc XRS
This process takes a fraction of a second.
After the reference file acquisition is over the system will be ready to operate.
1. Turn ON the lights using the Epi-white touch pad button.2. Place the focusing target on the transilluminator platen.3. Click on Live/Focus.4. An image of the target will be seen on the imaging screen.5. Using the lens control buttons on the membrane touch pad or the control panel that appears in the
Quantity One acquisition window, adjust the iris, focus, and zoom.
Note: For additional details on Reference File Acquisition, refer to the Quantity One software
manual.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
39/50
39
Section 5 Operation of the Universal Hood II
5.1 Operating the Unit
The Gel Doc XR and ChemiDoc XRS systems are easy-to-use instruments. In the Quantity One
acquisition window on your computer, select Live/Focus mode and adjust your image position, size, focus,
and intensity using the lens controls. After the image is optimized, capture the image. A typical procedure is
described below.
5.1.1 Switch on the Universal Hood II
1. Turn on the Universal Hood II main switch (on the rear bottom-left side of the enclosure).
2. Turn on the computer and start the software.
3. Select the acquisition mode from the File menu.
NOTE: Turn the camera power on, if you have not already done so.
5.1.2 Position Your Gel
1. Open the drawer of the Universal Hood II.2. Press the Epi White button to turn on the epi white lights.3. Center your gel on the transilluminator platen and close the drawer.4. Adjust the lens iris, zoom, and focus while looking at the computer screen.5. Open the door and reposition the gel, if necessary.6. When using the white light conversion screen or the white light transilluminator, focusing is easily
achieved if the iris is slightly closed.
5.1.3 Acquiring an Image
1. Press the appropriate light source for your sample.2. Select exposure time (see software manual for details) or select auto exposure..3. When a satisfactory image is seen, clickFreeze.
5.1.4 Acquiring an Image With Flat Fielding
Note: This applies to ChemiDoc XRS systems only where flat fielding is made available for acquiring
images with UV or white light transillumination.
1. Center your sample on the platen.2. Turn ON the UV or white light transilluminator.3. Adjust zoom, focus, and iris to get the best possible image in Live/Focus mode.4. Make sure that the box labeled Flat Fielding is checked to enable flat fielding.5. Select Auto Expose or Manual Expose.6. Following the exposure, a window will open asking if you would like to use a saved reference
image for flat fielding.
7. Select NO unless you have recently acquired a flat field reference image and are still using thesame illumination and lens settings.
8. Select YES if you are using the same lens and illumination settings that were used to previouslygenerate a flat field reference image file. See image on the following page.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
40/50
40
9. Next you will be prompted to remove your sample from the transilluminator and place thefluorescent reference plate on the platen for UV-illuminated images.For whitelight illumination
the white light transilluminator or the White Light conversion screen is used to collect the
reference flat fielding image. In either case, you must remove the sample from the platen.
10. Remove the sample from the transilluminator. Remove any liquid or residue remaining from thesample platen. Place the fluorescent reference plate on the platen. Make sure that the plate covers
the entire glass surface. Close the door and then click on Continue in the window shown above.
11. The next prompt will direct you to turn ON the UV transilluminator. Press the UV transilluminatoror white light transilluminator button on the hood, as the case may be, and click on Continue.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
41/50
41
Note: When using the white light, you must turn ON the UV transilluminator after removing the
sample.
12. The system will automatically acquire a reference image, and a flat-fielded corrected image
of the sample will be generated for you to save and analyze.
13.For additional details on flat fielding please refer to the Quantity One Software Manual. You mayalso click on the Quantity One Help button for a detailed description.
5.2 Cutting Gels
1. Open the drawer until it is fully extended and the Trans UV LED stops blinking.2. Place the clear acrylic UV protection screen in the horizontal slot in the inside of the drawer facing
the transilluminator.
!! CAUTION!! !! VORSICHT !!
WARNING: EXPOSURE TO UV RADIATION IS HAZARDOUS TO HEALTH. PLEASE
WEAR PROPER UV PROTECTIVE CLOTHING AND FACE AND EYE SHIELDS WHEN
CUTTING GELS. THE UV SHIELD PROVIDED WITH THE SYSTEM IS NOT ADEQUATE
PROTECTION AGAINST UV GENERATED BY THIS SYSTEM.
3. Press the Trans UV button to turn on the UV illumination.4. Cut the gel.5.
Press the Trans UV button and the UV transilluminator will turn off. Remove the acrylic UVprotection shield before closing the drawer.
Clear acrylic UV
protection screen
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
42/50
42
Section 6 Troubleshooting
Problem Possible Cause Solution
Image is not visible on the
monitor Aperture is closed Incorrect monitor settings Wrong cable connections
Open the apertureSee your computer manualSee Appendix A for correct
cabling
Image is not bright enough Wrong aperture setting Insufficient exposure time Open the apertureSelect longer exposure time
Impossible to image whole
sample area Lens is zoomed-in too close Zoom-out the lens
Printout does not look the
same as the image on the
monitor
Monitor settings are wrong Thermal printer settings are
wrong
Read the manual for the printer toensure proper settings
Light leakage into the
darkroom The lens body is not properly
centered and against the light
seal gasket on the hood
he rear hole on the MZL notclosed with the screw
Loosen the bracket and move thecamera around, to properly seal
the hood opening or adaptor plate,as the case may be
Close the hole with the blackplastic screw
Hot (very bright) pixels are
seen in the image
(ChemiDoc XRS only)
Hot pixels are moresusceptible to dark current
than surrounding pixels.
Select reference by clicking onthe appropriate box in the Options
window
Hot (very bright) pixels are
seen in the image
(Gel Doc XR only)
Corrupted reference darkimage
Click on Options button and thenclick on Reset Reference Dark
ChemiDoc XRS camera willnot respond/camera not
found
Camera power may be turnedoffController cables may not be
seated properly
Software driver for thecamera is missing
Computer power-savingmodes may be interfering
with the camera driver
Multiple FireWire devicesare installed on the same
computer
Camera may be defective
Turn ON the power to the cameraMake sure that all cables areconnected properly
Install Quantity One softwareagain if the camera driver is not
present
Disable power-saving modes onthe computer
Exit out of the Quantity Onesoftware and turn off power to the
ChemiDoc XRS camera
Power the camera back on andrestart the software
Replace cameraChemiDoc XRS image will
not autoexposeThe sample is too bright and
saturates the image at the
minimum possible exposure
Close the Iris control untilautoexpose is possible
Unable to focus on the
sample using white light
transilluminator or
conversion screen
Aperture is open too widecausing a shallow depth of
field
Close the iris slightly and thenfocus again. Repeat until good
focus is achieved
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
43/50
43
Section 7 Accessories and Replacement Parts
Accessories and Replacement Parts
Catalog
numberDescription
170-8001 UV/White light conversion screen
170-7950 White light transilluminator upgrade for Universal Hood II
170-8089 Printer, digital, Mitsubishi P93DW, 100/240 V, USB
170-7581 Mitsubishi thermal printer paper, 4 rolls
170-8074 Filter, 520DR30
170-8075 Filter, 560DF50
170-8076 Filter, 630BP30
170-8077 Filter, 440BP70
170-8081 Kit, filter, 580AF120, 62 mm
170-8072 Lens, f/0.95, 25 mm, wide-angle lens (ChemiDoc XRS system only)
170-8073 Lens, f/0.95, 17 mm, wide-angle lens (ChemiDoc XRS system only)
931-0198 Lens, motorized, 1275 mm, f/1.2 (ChemiDoc XRS system)
921-0265 Lens, motorized, 8.551 mm, f/1.2 (Gel Doc XR system)
170-8068 Shield, UV, Universal Hood II
170-3759 Bio-Rad fluorescent ruler
170-3760 Gel cutter ruler
170-7813 Sample holders for gels
170-8008 Fluorescent reference plate (ChemiDoc XRS system only)
931-0071 Cable, USB, Type A to B, 10' long
901-0064 Cable, USB, Type A to B, 6' long
10007733 Cable, Din, 5 pin, 3 ft, blk.
100-2827 Lamp, epi illumination, 5 W
100-1361 Lamp, UV B, 302 nm, 1
170-8097 Kit, 6 lamp, 302 nm
170-8098 Kit, 6 lamp, 254 nm
170-6887 Kit, 6 lamp, 365 nm
170-8080 Kit, accessory, ChemiDocXRS system
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
44/50
44
170-8173 Card, FireWire, ChemiDoc XRS systems
170-8177 Cable, FireWire, 10' for ChemiDoc XRS system
170-8099 ChemiDocXRS camera with MZL
100-2784 UV transilluminator lid (includes filter glass)
100-2785 Universal Hood II right shield epi-illumination (lid complete)
100-1948 Universal Hood II opal filter epi-illumination (without metal frame)
100-2786 Universal Hood II left shield epi-illumination (lid complete)
100-2787 Universal Hood II screw feet, 4
100-3951 -20 screw with black knob
100-1951 Fuse T 2 A, 250 V, 10
100-1952 Fuse T 4 A, 250 V, 10
100-1370 UV TR starter, 3
170-9600 Quantity One software, PC/Mac
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
45/50
45
Section 8 Maintenance and Parts Replacement
This section covers the replacement of parts in the Universal Hood II.
8.1 Epi Illumination Lamp Replacement
The lamps are located behind the two panels on each side (left and right) of the internal side of the hood.
1. Turn system power off.2. Remove the power cord.3. Open the enclosure door.4. Locate the epi-lamp housing.5. Locate the socket head screw that holds each lamp housing to the internal side of the hood.6. Remove this screw. The cover screen with a plastic piece will come loose.7. Pull on the cover to remove it from the box and the lamp will become visible.8. To remove the lamp, hold it from the receptacle and then pull it from the plastic holder.9. Insert the new lamp into the lamp holder and then push it into the receptacle.10. Reassemble the cover.
Epi-white light 5 W assembly (left side)
8.2 Fuse Replacement
Always unplug the instrument before changing or checking the fuses.
Power entry moduleThis unit is protected by 2 fuses (5 x 20 mm, 2A Slo-Blo). The fuses are located in the rear left side in a fuse
holder. See picture above.
Power cable
ON/OFF main
switch
Fuse holder
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
46/50
46
1. Unplug the main power cable from the power entry module outlet.2. Use a screwdriver lever on the fuse holder to extract it.3. Remove the blown fuses and replace them with two new ones (catalog #100-1951).4. Slide the fuse holder into the power entry module until it snaps in place.8.3 UV Transilluminator
UV transilluminator without cover
8.4 Lamp Replacement
Attention: The UV filter surface should always be kept clean from the chemical agents used as gel
dyes. Please use protective gloves when touching the UV transilluminator cover.
The lamps must be replaced after every 500 hours of use.
1. Turn off the power.2. Disconnect the power cord from the Universal Hood II.3. Remove the four screws located on the left/right sides of the trans cover.4. Remove the cover with the UV filter by sliding it forward, then lifting it up.5. Place it down on its backside on a nonabrasive surface so that the glass does not get scratched. 6. Caution: Do not put the UV filter directly on the bench. Wear gloves when touching the lamps
Remove the lamp by rotating until it becomes loose and the pins come to a vertical position.
7. Remove the lamp. Install the new lamp by rotating so that the pins are horizontal and the lamp istight.
8. Reassemble the cover and retighten the screws on both sides.
Lamp replacement Starter replacement
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
47/50
47
Starter Replacement (Catalog #100-1370)
1. Turn off the power.2. Disconnect the power cord from the Universal Hood II.3. Remove the four screws located on the left/right sides of the trans cover.4. Remove the cover with the UV filter by sliding it forward, then lifting it up.5. Place it down on its backside on a non-abrasive surface so the glass does not get
scratched.6. Caution: Do not put the UV filter directly on the bench.7. To remove the starter, rotate it counterclockwise and pull out.8. To replace the starter, insert it into the holder and rotate clockwise.9. Reassemble the cover and retighten the screws on both sides.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
48/50
48
Appendix A: Technical Specifications
Model Gel Doc XR ChemiDoc XRS
Application
Chemiluminescence No Yes
Fluorescence Yes Yes
Chemifluorescence No No
Colorimetry/densitometry Yes YesGel documentation Yes Yes
Isotopic imaging No No
Hardware specifications
Maximum sample size (cm) 28 x 36 28 x 36
Maximum image area (cm) 25 x 26 25 x 26
Excitation source
Trans-UV 302, (254, 365 nm
and white, optional),
epi white light
Trans-UV 302, (254, 365 nm
and white, optional)
epi white light
Illumination modes
3
Trans UV and white,
epi white
3
Trans UV and white,
epi white
Detector CCDSupercooled
CCD
CCD resolution (H x V) 1,392 x 1,040 1,392 x 1,040
Pixel size (H x V in microns) 4.65 x 4.65 6.7 x 6.7
Cooling system No Peltier cooled
Camera cooling temp. NA 50C
Filter holder
3 positions
2-emission filter
1-chemi
3 positions
2-emission filter
1-chemi
Emission filters1 included (amber),
4 optional
1 included (amber),
4 optional
Dynamic range >3.0 orders of magnitude 4.0 orders of magnitude
Pixel density (gray levels) 4096 65,536
Dynamic flat fielding - Yes,
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc
49/50
49
Legal Notices
Coomassie is a trademark of BASF Aktiengesellschaft. Cy is a trademark of GE Healthcare Group
Companies. FireWire, Mac, and Mac OS are trademarks of Apple, Inc. Hoechst is a trademark of Hoechst
Gmbw. Mitsubishi is a trademark of Mitsubishi Companies. Slo-Blo is a trademark of Littelfuse, Inc. Sony
is a trademark of Sony Corporation. SYBR, SYPRO, and Texas Red are trademarks of Invitrogen
Corporation. Bio-Rad Laboratories, Inc. is licensed by Molecular Probes, Inc. to sell SYPRO products for
research use only, under US patent 5,616,502. Windows, Windows XP, and Windows Vista are trademarks
of Microsoft Corporation. Practice of the polymerase chain reaction (PCR) may require a license.
-
8/14/2019 Instruction Manual Gel Doc and Chemi Doc