Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok...
Transcript of Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok...
![Page 1: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/1.jpg)
B 1-2. practicals – Introduction to invasive monitoring
B 3-4. practicals – Monitoring of the cardiovascular system
Cannulation of jugular vein → CVP measurement;Cannulation of femoral a. → Arterial pressuremeasurement;Thermodilution cardiac output measurement;Positioning of Swan Ganz catheter to the pulmonalisartery;Disection of carotis artery → Blood flow measurement;
B 5-6. practicals – Complex monitoring
Institute of Surgical ResearchBasic Medical Skills
![Page 2: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/2.jpg)
INVASIVE HEMODYNAMIC
MONITORING
„When a patient is critically ill, even the most thorough examination by the most experiencedphysician is not enough; then we have tomeasure, and measure particularly accurately.”
Gábor Petri
![Page 3: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/3.jpg)
Dissection of veins - Cannulation ofjugular vein
Venous cutdown, that is surgical exposure of a peripheral vein, is necessary if it is impossible toinsert a cannula into a satisfactory vein - or thepercutaneous insertion of a vena cava catheter is contraindicated.
Indications for catheterization of a vein
1. Replacement of fluids (infusion, transfusion)2. Iv. administration of drugs3. Obtaining serial blood samples
![Page 4: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/4.jpg)
Venous catheter can be introducedinto the following veins
1. Arm veins (median cubital vein, basilic vein, sometimes cephalic vein)
2. External jugular vein3. Superficial veins of the foot and the leg (saphenous
vein)
![Page 5: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/5.jpg)
Venous cutdown, cannulation (video)
![Page 6: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/6.jpg)
Removal of venous catheter (video)
![Page 7: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/7.jpg)
Arterial/venous catheterization with percutaneous punction
![Page 8: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/8.jpg)
Central venous catheter
1. Measurement of central venous pressure
Pressure transducer(sensor)
![Page 9: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/9.jpg)
1. Central Venous Catheterization
Surgical Team 1.
• Surgical exposure of the left jugular vein:1. Stretching the skin on both sides of the planned incision;2. Skin incision (5-7 cm) with a skin scalpel;3. Dissection of subcutaneous tissues with diathermy pencil; 4. Handling bleeding if necessary;5. Localisation of the vein (in approx. 2 cm deptht);6. Blunt dissection of the vein;7. Opening the vessel and introduction of the CV catheter
• Fix the cannula securely and connect it to the pressure measurement system;
• CVP measurement
![Page 10: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/10.jpg)
Arterial thermodilution catheter
Thermosensor for cold saline bolus
Pressure transducer
2. Measurement of arterialpressure in the femoral artery
![Page 11: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/11.jpg)
2. Catheterization of femoral arterySurgical Team 2.
• Skin incision on the inguinal area;• Cutting the connective tissue by diathermy;• Abdominal wall is retracted;• Careful, blunt disection of the femoral artery
by using dressing forceps only (!);• Catheterization with a thermistor-pressure
supplied catheter; • Fix the catheter securely and connect it to the
pressure sensor;• Arterial pressure measurements
![Page 12: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/12.jpg)
PiCCO Monitor
Arterial thermodilution catheter
Injectate temperature sensor
Pressure transducer
Central venous catheterInjectate salinetemperaturesensor housing
3. Cardiac Output measurement with a transpulmonary thermodilution (TDa) method
![Page 13: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/13.jpg)
Cold salinebolusinjection
Thermistorcatheter inaorta
Left HeartRight heart
RA PBVEVLW
LA LV
EVLW
RV
Scheme for the TranspulmonaryThermodilution Method
![Page 14: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/14.jpg)
A known volume of cold saline (5-10-20 ml) is injected iv, as fast as possible. Saline temperature is at least 10°C lower than blood temperature.
The passage of the heat bolus injected into the central vein is registered by a thermistor catheter positioned in the femoral artery.
The temperature change recorded downstream is depending on the flow and on the volume through which the cold indicator has passed. As a result, a thermodilution curve can be obtained.
The cardiac output is calculated from the area underthe thermodilution curve.
Measurement of Cardiac Output (CO)
![Page 15: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/15.jpg)
Time components of the thermodilutioncurve
MTt: Mean transit time ≈ half of the indicator passed the point of detection.DSt: Downslope time ≈ exponential downslope time of TD curve.
![Page 16: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/16.jpg)
ITTV = IntraThoracic Total Volume
ITTV = MTt x CO( V = t x Q )
ITTV = volume from the point of injection to the point of detection.Flow measurement between two points, based on theV = t x Q relationship;
![Page 17: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/17.jpg)
PTV = Pulmonary Total Volume
PTV = DSt x CO( V = t x Q )
Between the two detection points, the lung is thelargest volume space, where the flow slowes downand the rate of thermodilution is the highest → thisis shown by the DSt time component.
![Page 18: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/18.jpg)
GEDV = Global End-Diastolic Volume
GEDV = ITTV - PTV
GEDV = (MTt - DSt) x CO
GEDV= End-diastolic volume of the 4 heart chambersGEDV = Preload
![Page 19: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/19.jpg)
ITBV* = IntraThoracal Blood Volume
ITBV* = 1,25 x GEDV
r = 0.96
ITBV
TD (m
l)
GEDVST (ml)
Sakka et al, Intensive Care Med 26: 180-187, 2000
![Page 20: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/20.jpg)
EVLW* = ExtraVascular Lung Water
EVLW* = ITTV - ITBV*
![Page 21: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/21.jpg)
Arterial Pulse Contour AnalysisPulse Contour Cardiac Output (PCCO)
t [s]
P [mm Hg]
![Page 22: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/22.jpg)
1) Thermodilution parameters
CO → Cardiac output (L/min)
CI → Cardiac index = CO/body surface (ml/min/m2)
GEDV→ Global End-diastolic volume = PRELOAD (ml)
ITBVI* → Intrathracic blood volume (estimated) (ml)
EVLWI* → Extravascular lung water (estimated) (ml)
2) Pulse Contour parameters
PCCO→ Pulse contour cardiac output (continouos) (L/min)
SV → stroke volume = CO/heart rate (ml)
AP → arterial pressure (systolic, diastolic, mean) (mmHg)
SVR → Systemic vascular resistance = (AP-CVP)/CO
dP/dtmax→ heart contractility index (mmHg/sec)
Measured and calculated parameters
![Page 23: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/23.jpg)
Positioning of Swan-Ganz catheter into the pulmonary artery
4. Monitoring of pulmonary circulation
![Page 24: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/24.jpg)
Swan-Ganz catheter, a device of monitoring of pulmonary circulation: Pressure and CO
Yellow: PA pressureBlue: for CVP and saline injectionWhite: thermistor cabelRed: catheter balloon
![Page 25: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/25.jpg)
4. Monitoring of pulmonary circulation
1. Exposure of the right jugular vein;
2. Catheterisation with a Swan-Ganz catheter;
3. Introduce the catheter into the arteria pulmonaliswitth the help of continous monitoring of thepressure signal and the inflatable balloon at the tipof catheter;
4. Pulmonary artery pressure, pulmonary capillarywedge pressure and cardiac output can be measuredsimultaneously
Introduce a Swan-Ganz catheter into the a. pulmonalisfrom the jugular vein, through the right heart
Surgical Team 2.
![Page 26: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/26.jpg)
Pulmonary artery and wedge pressures
Right ventricle A. pulmonalisRight atrium PCWP
![Page 27: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/27.jpg)
Thermodilution cardiac output measurementwith Swan-Ganz catheter in clinical practice
Injector branch of catheter
Thermal bolus sensingthermistor branch
Pressure transducer
Internal reference thermistor sensorfor cold saline injecting
Inflatable branch of catheter
Computermonitor
Tb injection
t
![Page 28: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/28.jpg)
The transonic research flowmeter consists of a bench-top electronic flow detection unit with enhanced frequency resolution and volume flow sensing probes.
5. Blood flow measurement on carotid artery
![Page 29: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/29.jpg)
5. Disecting of carotis artery
• Surgical exposure of the right carotis arteryfrom a previous incision
• Localisation of the artery betwwen thetrachea and jugular vein, under a musclebundle, together with nervus vagus.
• Fix the artery by a double thread securely.
• Take around of artery a perivascular flowprobe.
![Page 30: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/30.jpg)
The execution of the practical
Scrub preparation, gowning, gloving: approx. 8 min;
6-8 students/operating tables
Two surgical teams /operating table:as surgeon, first and second assistants and nurse
Surgical Team 1:Dissection and cannulation of left jugular vein and
femoral artery;
Surgical Team 2:Dissection and cannulation of right jugular vein and
carotid artery;
![Page 31: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/31.jpg)
Blood pressureand
thermosensor
Swan-Ganzcatheter
Tb
CVP catheter
1. Dissecting of left jugular vein
2. Dissecting of left femoral artery
3. Transpulmonary(TDa) cardiac
output measurement
4. Dissecting of right jugular vein
5. Dissecting of rightcarotid artery
Summary of the Practical
![Page 32: Institute of Surgical Researchweb.szote.u-szeged.hu/expsur/rop/aktualis_gyakorlatok PDF-ben/B3-4...Dissection of veins - Cannulation of jugular vein Venous cutdown, that is surgical](https://reader031.fdocuments.us/reader031/viewer/2022021811/5cd5965e88c9937d508c0d29/html5/thumbnails/32.jpg)