MasterManualofProceduresfor
HumanTestingCenter
InstituteofAppliedLifeSciences
LifeScienceLaboratoriesS360240ThatcherRoad,
Amherst,MA01003
atthe
UniversityofMassachusetts,Amherst
Updated:16October2017 ii
HumanTestingCenter
Chapter1:CoreAccess,Training,Scheduling,PricingandBilling
ExerciseInterventionandOutcomesCore
Chapter1:ParvoMedicsTrueOne2400MetabolicCart
Chapter2:VelotronBicycleErgometer
Chapter3:Monark828EBicycleErgometer
Chapter4:Monark874EBicycleErgometer
Chapter5:BiodexSystem4ProDynomometer
Chapter6:QuintonQ-Stressw/TM55Treadmill
Chapter7:GECarescapePatientMonitors
Chapter8:GELunariDXAa
LivingScienceCore
Chapter1:OxyconMobileMetabolicSystem
Chapter2:ActigraphLinkActivityMonitor
Chapter3:ActigraphGT3X+BTLEActivityMonitor
Chapter4:Actilife&ActilifeLiteSoftware
HumanMotionCore
Chapter1:QualysisOQUS700+/500+w/AMTIForcePlatformIntegration
Chapter1.1–CreatinganewprojectinQTM
Chapter1.2–QTMDataCollection
Chapter1.3-QTMDataProcessing
Chapter1.4–ExampleMarkersets
Chapter1.5–Visual3DTutorial
Chapter2:DelsysTrignoEMG/IMUunits–MotionCaptureIntegration
RoomCalorimeterCore
Chapter1:QuickGuide forValidation/Calibration
Updated:16October2017 HumanTestingCenter 1
MANUALOFPROCEDURES
HumanTestingCenter
InstituteofAppliedLifeSciences
LifeScienceLaboratoriesS360240ThatcherRoad,
Amherst,MA01003
atthe
UniversityofMassachusetts,Amherst
Updated:16October2017 HumanTestingCenter 2
UNIVERSITYOFMASSACHUSETTSAMHERST
INSTITUTEFORAPPLIEDLIFESCIENCES
HumanTestingCenter
CoreAccess,Training,Scheduling,Pricing,andBilling
MostrelevantinformationforgainingaccesstotheCoreswithintheHumanTestingCentercanbefoundontheCorewebsites.Accesswillonlybegiventousersonceallstepshavebeencompleted.
a) CoreAccess:
InternalUser.
IfyouareaUMassAmherst,UMasssystem,orFiveCollegebasedresearcherwhoisdoingresearchthatwouldreceive‘internal’pricing.ThefollowingstepsoutlinehowyouwillgainaccesstotheCoreFacilities.
i) ContactMichaelBusatheCoreFacilityDirectoranddiscussyourprojectsneedsandtimelines([email protected],413-755-0574).
ii) IfyouarenewtoIALSaMOUandpaymentauthorizationformmustbecompletedforeachcorefacilityyouwouldliketouse.ThePI’sStartUppaperworkcanbefoundintheFormstabofthewebpage.
ExternalUser
Ifyouareanexternaluserfromanacademicorindustrypartner,pleasecontactMichaelBusa,theCoreDirector,todiscussyourneeds([email protected],413-577-0574).
b) Training:IndividualswhowouldliketouseanypieceofequipmentintheHumanTestingCentermustcompleteatrainingsessionwithCorestaff(CoreRatesApply).ThesetrainingsserveasbothatechnicaltutorialforequipmentuseandanorientationforuserstotheHumanTestingCenter.Foruserswhoalreadyhaveexpertiseintheuseofaparticularpieceofequipment,theycanrequestthattheyjustdemonstrateproficiencyinuseandskipthetutorial.PleasenotethateventhoughausermaybeproficientinthetechnicalaspectsofacertainpieceofequipmentthattheintegrationofthebillinganddatastoragesystemsintheIALScoresdiffersfromstandalonepiecesofequipment.Beingtrainedontheseaspectswillensurethatdataisproperlystoredandbillingrecordsareaccurate.
Aspartofthetraining,theprospectiveusermustalsoprovidetheCoreStaffwithproofoftherequireEHStrainings.AllrequiretrainingsarelocatedatthebeginningofeachinstrumentsStandardOperatingProcedure.
c) Scheduling:AllschedulingofequipmenttimeistobedonethroughtheFacilitiesOnline
Managementsystem(FOM).i) Supervisors(i.e.PIs)musthavesupervisoraccountscreatedpriortohavingtheirRA’screate
profiles,theCoreDirectorwilldothisforthemfollowinganinitialemailorphonecallindicatingtheirintentiontouseacore.
ii) AllusersmustcreatetheirownFOMaccountandlinkittotheirsupervisor.
Updated:16October2017 HumanTestingCenter 3
(1) Theymustthenclickontheindividualcapacitythattheywanttogainaccessto,andrequestatrainingtimewiththeCoreStaff.
(2) Duringregistration,theymustusethespeedtypethatmatchesthePaymentAuthorizationFormthatissubmittedaspartofthePI’sStartUppaperwork.Additionalspeedtypescanbeaddedbyusersinthecasewhentheyworkonmultipleprojects.
d) Pricing/Usage:PricingforfacilityusearepostedontheCorewebsites.i) Ratesaresetonanannualbasisandaresubjecttochange.ii) PricesaresetinFOM.ChargesforusearebasedonFOMactivetimeroundeduptothe
nextcompleteschedulingperiod.
e) Billing:BillingiscarriedoutbytheIALSaccountingstaff.i) Statementsaresentouttosupervisorsandtheirdesignatedrepresentative.ii) Disputesorchangesinspeedtypetotheinvoicemustbemadewithin10businessdays,
otherwisetheselectedspeedtypewillbeautomaticallychanged.
Updated:16October2017 ExerciseIntervention&OutcomesCore 1
ManualofProcedures
ExerciseIntervention&OutcomesCore
InstituteofAppliedLifeSciences
LifeScienceLaboratoriesS360240ThatcherRoad,
Amherst,MA01003
atthe
UniversityofMassachusetts,Amherst
Updated:16October2017 ExerciseIntervention&OutcomesCore 2
UNIVERSITYOFMASSACHUSETTSAMHERST
INSTITUTEFORAPPLIEDLIFESCIENCE
ExerciseInterventionandOutcomesCore
Chapter1:ParvoMedicsTrueOne2400
1.1Introduction:TheTrueOne®2400MetabolicMeasurementSystem(ParvoMedics,INC.,Sandy,UT,USA)sanintegratedmetabolicmeasurementsystemformaximalO2consumptiontestingandindirectcalorimetryassessment.TheTrueOne®2400'sAnalyzerModule,withparamagneticoxygenandinfra-redcarbondioxideanalyzers,isdesignedtobeaccurate,reliable,andeasytouse.Theuniqueflowmetercalibrationalgorithm,utilizinganimagereconstructiontechnique,correctsthenon-linearityofthepneumotachandprovideshighlyaccurateflowmeasurement.
Features:
•HighPrecisionFlowmeterCalibration
•CompactAnalyzerModule
•HighlyReliableMixingChamberPrincipleRecommendedApplications
•AthleticTesting
•ExercisePhysiologyTeachingandResearch
•SportsMedicineResearch
•EnergyExpenditureAssessment
•DifferentialDiagnosisinCardiopulmonaryDiseases
Updated:16October2017 ExerciseIntervention&OutcomesCore 3
1.2HowtoCalibrate:Belowisastep-by-stepoutlineforcalibratingtheunit.
PreliminaryCalibrationSteps1. LocatethepowerswitchontherearoftheAnalyzerModuleandturniton.
a. Allowit30minutestowarmup
2. TurnonthecomputerandmonitorandlaunchTrueOnesoftware
a. Theydonotrequireawarmuptimeofmorethan2-minutes3. Assemblethewater-trapconnection
a. ConnectthefiltertotheT-connectorusingthegreyconnectionpiece.Gentlypushthetwotogetherinacircularmotion.
Updated:16October2017 ExerciseIntervention&OutcomesCore 4
4. Attachthisassemblytothepneumotach.Attachbypushingthewatertraptothepneumotachwithslightpressure.Ensurethatthemixingchamber(tintedrectangularbox)doesnotfalloffthecart.
5. Ontheotherend(labeledtube)attachedoneendofthehose.
Updated:16October2017 ExerciseIntervention&OutcomesCore 5
6. Next,assembletheT-mouthpiece3-Lsyringecalibrationassembly.Makesurethesmallwhitemembranesarefacingthesamedirection.Lookingatimage#3below,whenholdingtheunitwithupright,thesideofthebodywiththeair-flowarrowfacingup(smallarrowonthecenterbodypiece)themembraneonthissideshouldhavesurfaceparallelwiththelongwaterchamber.Themembraneontheoppositesideshouldbefacingthesameway.Ifyouarehavingtroubleusethisresource:https://www.youtube.com/watch?v=YSg65afxMCM
FINALPRODUCTforT-mouthpiece3-Lsyringecalibrationassembly
Updated:16October2017 ExerciseIntervention&OutcomesCore 6
7. Taketheotherendofthehose(endnotattachedtothewatertrap)andconnectittotheT-mouthpiece3-Lsyringecalibrationassembly.Makesurethehoseisattachedtotheoutputvalve(endwith3metalrings).
8. Next,grabthe3-Lsyringe,ensuringithasthegreyconnectorattached
Updated:16October2017 ExerciseIntervention&OutcomesCore 7
9. InterfacetheT-mouthpiece3-Lsyringecalibrationassemblywiththegreyendofthe3-Lsyringe.Makesurethattheconnectionistight,howeverdonotforcefullypushthemtogether.Rather,gentlymakesmallcircularmotionstoconnectthem.
10. Next,prepareforgascalibrationbyopeningthegascylinder.Turnthemetalknobonthetopofthe
cylindercounterclockwise1and½turns.TheCalGascalibrationiskeytoaccuratetestingresults.Youcanrunthegascalibrationasmanytimesasneededduringcalibration.Itisrecommendedtorunthecalgascalibrationonelasttimejustbefore(getsonthetreadmillorergometer)testingthesubject.WeneedtoensuretheCO2levelstheParvocartisreadingforroomairhavenotfloatedabove3-4%.Ifthishappens,theRERvalueinthetestwindowwithbeinflatedwhentheteststarts.
Updated:16October2017 ExerciseIntervention&OutcomesCore 8
11. Nowyoucanattachthecalibrationgashose/whiteconnectorfromthecylindertotheCalGasPortonthemetaboliccart.Thisislocatedonthebackoftheunit.ItistotheleftoftheRoomAirportandjustabovetheO2Rangeswitch.Youmustholdthefrontoftheunit,soitdoesnotfalltothefloor,whilebringingthecalibrationgashose/whiteconnectortotheCalGasport.Stayparalleltothefloorandpushtheconnectorstraightintheportconnector.Toremove,pushdownlightlyonthemetalclaspandpullthecalibrationgashose/whiteconnectoraway.
12. Now,touchpneumotach,itshouldbewarm,indicatingitisshouldbewarmedup.Ensurethat30-
minuteshaspassed,andcontinewithcalibration
13. Thesystemisnowreadyforthethreestepcalibration.Oncethesystemhasbeensetup,thecalibrationprocessisveryquickandeasy.Thesystemshouldbecalibratedagaininbetweenmultiplesubjectsorifithasbeensittingturnedonbutnotusedforalongperiodoftime.Theonlycalibrationsteptheoreticallyneededafterinitialcalibrationistoapplythecalibrationgasagain.However,theenemyofthepneumotachismoisture.Excessmoisturewillelectricallydistortwhatthepneumotachissensing;thiswillbereflectedinerroneousreadingofthe3litersyringeas9,12,15litersetc...Thescreeninsidetheroundovenneedstobeasdryaspossible,soyoumayhavetousethesyringetoinputroomairacrossthesensorandintothemixingchamberifyouaretestingmultiplesubjects.ThepneumotachisverysensitivetosubtlechangesinCO2.Thisisthereasonwhyweneedtoletthepneumotachheatuptooperatingtemperature.DoingatestwithacoldpneumotachwillhavedistortedCO2results–theRERandVO2kgreadingswillbewayoff.
Updated:16October2017 ExerciseIntervention&OutcomesCore 9
1.3FLOWMETERCALIBRATION
1. OpenTrueOnesoftwareandclick“FlowmeterCalibration”2. Enterroomtemperature,relativehumidity,andbarometricpressurereadingsfromtheVantage
VUEintodialogueboxonTrueOne
3. ClickandopenSampleBaselineintheupperrightcornerofthedialoguebox4. TheFlowmeterCalibrationProcessingwindowwillopen.Allinputstrokeswillbedonewiththe
3-Lsyringe.
5. Detectionstroke-oneinputstrokeindicatesairisflowingintothesystem(mouthpieceandallassembliesareproperlyconfigured).*NoticetheFlowindicatorwillshowtherealtimeinput
Updated:16October2017 ExerciseIntervention&OutcomesCore 10
velocity,StrokePeakwithshowthepeakinrealtimeandintheblankbluefieldbesidetheCancelbutton,thevolumeofthesyringewillbedisplayedattheendofanystroke.Thisvalueshouldbearound3liters.Ifitreadshigh–9,12,15litersetc..thepneumotachhasamoistureissue(iftestingmultiplesubjectsoverashortperiodoftime)andmustbeallowedtodry.YoucanusethesyringesetuptodotheFlowmetercalibration,butdon'topenthesoftware.Youarejustphysicallydrivingairacrossthepneumotach,throughthemixingchamberandintothesystem
6. FlushStroke–fourinputstrokestopushfreshairintothesystem7. Thepneumotachcansense800lpm.Ofthefivecalibrationstrokesinputnext,twomustbeless
than80lpm.ThenumberwillfluctuatebesidetheFlownumericalindicator(totheleftofStrokePeak.ForExerciseStrokevelocityisveryveryslow,veryslow,slow,fast,faster.
8. Stroke1→approximately50-80lpm(very,veryslow)9. Stroke2→approximately100+lpm(veryslow)10. Stroke3→approximately200+lpm(slow)11. Stroke4→approximately300+lpm(fast)12. Stroke5→approximately400+lpm(faster)13. ForRestingStrokevelocitycanstayat50-80lpm14. Theresultissomethinglikethiscurve
Updated:16October2017 ExerciseIntervention&OutcomesCore 11
1.4GASCALIBRATION
1. Onthehomescreen,click“GasCalibration”2. TheGasCalibrationwindowwillopen,likewithFlowmetercalibration,correctlyinput
temperature,humidity,andpressureintothedialoguebox.ClickOK.3. Theunitwillperformgascalibrationandafterabout20seconds.4. Apop-upwillappearpromptingyoutoturnoffthecalibrationgas.5. Asummarywindow(similartotheonepicturedbelow)willarise.Itisrecommendedtodothe
CalGascalibrationuntilyougettheexactnumbersonthecylinder.Also,youmaydothisstepjustbeforeyoursubjectisreadyfortesting
6. Click“Save”.7. Whendone,turnoffthecalibrationgascylinderbyturningclockwisehandtight.Thenbleedof
theexcessleftintheregulatorandthecalibrationgashose/whiteconnectorbygentlypushingontheplasticconnector/openingattheend.Bleedoffuntilbothindicatorsreadzero
Updated:16October2017 ExerciseIntervention&OutcomesCore 12
1.5TESTINGwiththeParvo2400:1. Fitparticipantwiththeheartratemonitor,theheartratetransmitterislocatedonthereticulated
arm.Besuretowetthestraporapplyelectrodegelbeforethemonitorisused.2. Ifyouareusingareusingtheheadsupportsystem,attachedsweatguardtothefrontofthesystem.
Updated:16October2017 ExerciseIntervention&OutcomesCore 13
3. Whiletheparticipantisseated,fittheheadsupportsothefrontportionisjustabovetheeyebrows.
Ensureasnugfit.Tightentheentirebandaroundtheheadusingtheknobontheback.Youwillbeusingamouthpieceandnoseclipwiththissetup.Attacharubbermouthadaptertothesmallwhitecylinderofthemouthpieceyoumadeearlier.Youwanttoavoidthesubjectbreathingoutofthecorneroftheirmouth(whenatmaximalexertion).Wewantatightsealregardlessofwhatdeviceisused-headsupportorfullmask.
4. Givethesubjectanosecliptoputon5. Theyarereadyforthetreadmillorergometer.Attachthehosetotheoutputendofthe
mouthpiece.Donotletthesubjectinserttheadaptoryet.Whenitistimetostartthetest,thesubjectcanjustspintheadaptortowardtheirface,placetheadaptorintheirmouthandbegin
Updated:16October2017 ExerciseIntervention&OutcomesCore 14
mouthbreathingintothevalves.Wemustwaituntiltheappropriatetestwindowopensforthesubjecttodothis.Theyarenottobreatheintothehoseatthistime.
6. Thefullfacemaskdoesnotrequireanosecliptobeused.Tousethispiece,switchoutthesmall
whitecylinderforthesmallerclearadapter(picturedbelow).Threadtheendofthesmallclearadapterintothefrontholeofthebluefacemask.Facialhairmayinterferewithatightseal.Donothookupthehosetothefullfacemaskuntiltheteststarts.Thestrapswraparoundtheheadandthentheblackconnectorslockintotheclearbracketsofthemask.ThestrapsareadjustablebyslidingthestrapthroughtheblackconnectortheusingVelcrotokeepinplace.Themaskcoversthemouthandnose–sothestrapsneedtobetighttoensureagoodseal
Updated:16October2017 ExerciseIntervention&OutcomesCore 15
1.6METABOLICTESTING1. Onthehomescreen,clickonVO2/MetabolicTesting.2. Inputparticipantinformation.Updatetemperature,humidityandbarometricpressureifithas
changed.Choosewhichdeviceyouareusing.
3. Ifyouusethetreadmillyouneedtochooseaprotocol.Youcancreateanewoneandnameitwhateveryouwant.Or,modifyanexistingoneandsaveit.Usewhatistherealreadyfromthepulldownmenu.OrmodifyaprotocolandnotsaveitbyjustclickingOK.
4. Treadmillprotocolscanhavearesting,warmup,exerciseandcooldownphase.Ifyoudonotwant
restingorwarmupphases,maketheduration=0.
Updated:16October2017 ExerciseIntervention&OutcomesCore 16
5. ClickOK6. Ifyouusetheergometeryoualsoneedtochooseaprotocol.Youcancreateanewoneandnameit
whateveryouwant.Modifyanexistingoneandsaveit.Usewhatistherealreadyfromthepulldownmenu.OrmodifyaprotocolandnotsaveitbyjustclickingOK.
7. YouhavesimilaroptionswhenusingabikechooseaprotocolandpressOK.8. Nowthetestingwindowswillappear.Ifusingtheheadsupport,youcanattachthehosefromthe
reticulatedarm/T-connectortotheoutputconnectoronthemouthpiece.Thenoseclipcanbeworn.Donotletthesubjectbreathintothemouthpieceyet.Ifusingthefullfacemask,donotattachthehosefromthereticulatedarm/T-connectortothemaskyet.
9. Aseriesofsoftwarewindowswillopenandclose.10. Thenthiswindowwillopen.Youwillseetheblanktestresultswindowopenaswell.Thefieldswill
beblankbecausethesubjectisnotbreathingintothesystemyet.WaituntiltheCO2levelsexceeds0.03%,thenclickOK.
11. Ifusingtheheadgear,thesubjectcanspinthemouthpiecetowardtheirface.Theadaptorcanbe
placedinthemouthwhilethenoseclipison.Ifusingthefullfacemask,thehosefromthereticulatedarm/Tcannowbeattachedtotheinputconnectoronthefullfacemask.
12. Therewillbeadelayofapproximately15–20secondsbeforetheMetabolicMeasurementTesting
windowupdateswithdata.Thewhitespacebelowwillbefilledwithdataevery5seconds(thisresolutionlevelcanbechanged)
Updated:16October2017 ExerciseIntervention&OutcomesCore 17
13. TheEventsbuttontotherightcanbeusedtoenterRPEvalues,increasedwattage,etc...apop-upwindowwillappear.SimplyentertheRPEvaluetotheemptydatafieldinthepop-up.Thisneweventwillbeaddedtothedatalinesinthewhitespacebelow
14. Welldocumentedreportsareavailableaftertesting.TotherightoftheCancelbuttonisasetoftimers.Thiswillshowoveralltesttime,restphasetimeleft,warm-upphasetime,andexercisephasetimeleftasyougothroughthestagesoftheprotocolyouareusing.Afterthetestyoucanviewdataunderthe“Reports”tabonthehomescreen.
Updated:16October2017 ExerciseIntervention&OutcomesCore 18
1.7CLEANINGEQUIPTMENT1.) Whilewearinggloves,removeheadgearorthefullfacemaskfromthesubject.Rememberto
keepthesubjectattachedtothecartforabout20secondsafterthetesthasfinished.Thisallowssoftwaredelaytofillintheremainingdata
2.) Takeallcomponentsoffandbringtogreycart/sinkarea3.) Detachmouthpiecefromheadgear4.) Rinsethesweatguardunderwater.Theheartratestrapandtransmittercanberinsedaswell.
Letthesepartsairdryonthepeggedrack5.) DisassemblethemouthpieceandrinseallcomponentsinsinkwithJohnsonbabysoap
6.) Rinsein______________solution
1.8Contacts1. ParvoMedics
8152South1715EastSandy,UT84093,USA
1800-942-72552. MichaelBusa–CoreDirector
Updated:16October2017 ExerciseIntervention&OutcomesCore 19
Chapter3:Monark828E,weightergometer3.1Introduction
MonarkErgometermodel828Eistheworld’smostcommonlyusedexerciseergometer.TheadjustablebrakesystemandtheforcecanbesetandreadKiloponds(kp)andNewton(N).Theresultoftheresistanceandcadenceisdisplayedinpower(Watts)ontheelectronicmeter.Thecycleisequippedwithanelectronicmetershowingpedalrevolutionsperminute(RPM),heartrateinbpm(HR),exercisetimeinminutesandseconds(TIME),speedinkmperhourormilesperhour(SPEED),covereddistanceinkmormile(DISTANCE),burnedCalories(CAL)andthepoweronthecycle(WATT).Thewattleveldependsonpedallingspeed,itcanbefinetunedbyincreasingordecreasingthespeedorpedalrpm.Each828Eiscalibratedatthefactory.ThismeansthatyoucanbegintousetheErgometerdirectlyafterassembly.However,iftheuserwishestoverifythescale,pleasereadtheinstructionfor“Calibration”inthismanual.
Quickfactsaboutthe874E:
•Large,well-balancedflywheel20kg(44lbs)
•Pendulumscale,easytocalibrate
•Adjustableseatheight
•Adjustablehandlebarwithquickreleaselever
•Stableframe,solidsteeltube
•Electronicdisplaywithheartrate
Corecompetencies:
1) Understandingthebrakedeviceandbasicsofthetensioncenter2) BefamiliarwiththeFitnesscomputerandwhatmetricsitgenerates3) Reviewserviceactionsandunderstandwhatactionsyoumightneedtotakebefore/afterusing
thisequipment.3.2UserInstructions
CycleadjustmentsSeatheightshouldbeadjustedtoacomfortableposition.Theappropriateheightistohavethekneeslightlybentwhenthesoleofthefootiscenteredoverthepedalaxlewiththepedalinthebottomposition.Toadjusttheseatheight,loosenthelever(1)ontheseattube.SeeFig:Adjustments.Alsonotethatthehandlebarstemshouldbeinsertedintotheframetubeatleast3inches.Thismeasureismarkedwith“MAX”onthestem(3).
Updated:16October2017 ExerciseIntervention&OutcomesCore 20
FitnesscomputerTheMonark874EisequippedwithaFitnesscomputerthatdisplaysthefollowingmetrics:
1. Revolutionsperminute(RPM)2. Heartrateinbpm(HR)
a. Atthedisplayforheartrate(HR)aheartsymbolislitwhichmeansthatthemeteristryingtofindapulsesignalfromanexternalsource,chestbeltwithelectrodes,IfthemetercannotfindsuchasignaltheHRfunctionisautomaticallyturnedoffafter30seconds.Whenthefunctionisturnedofftheheartsymbolisnotlitanymore.Theheartratefunctioncanbeturnedonagainbypressingakey.
3. Exercisebouttimeinminutesandseconds(TIME)a. Thetimerstartsautomaticallywhenpedalsaremoved.MetervaluesforTime,Distanceand
CaloriescanbesettozerobypressingtheRESETbutton(seeimagebelowforRESTbuttonlocation)formorethantwoseconds.
4. Cyclingspeedinkmperhourormilesperhour(SPEED)5. Covereddistanceinkmormiles(DISTANCE)6. Caloriesburned(CAL)andpower(WATT)arealsodisplayedonthecomputer.Togetcorrect
readingsforcaloriesandwatts,thekp-valueontheelectronicmeter(Fitnesscomputer)hastobesettothesamevalueasthependulum,orthekpwindowshowntotheleftoftheelectronicmeter.
a. Forexample,thependulumandthekp-windowisshowing2kp.Pressthekp-keytotheleftonthemeter.Thelowerdisplaywindowisnowshowingfiguresinkp.Increaseordecreaseinstepsof0.1kpbypressingthekp-key-(arrowup)ortheRESET-key-(arrowdown)untilthereadingcorrespondswiththeactualordesiredkp-valuesonthependulumscaleorinthekp-window.AfterthatpresstheCAL/WATTbuttontoeithershowtheCAL-orWATT-figures.Thewattreadinginthedisplayisdependingonthepedallingspeed.Thewattreadingcanthenbefinetunedthroughincreaseordecreaseofthepedallingspeed.Caloriesarecalculatedallthetime.
Comment[MB1]:Makeavideo?
Comment[GP2R1]:Made.WaitingonYouTubeaccount.
Updated:16October2017 ExerciseIntervention&OutcomesCore 21
Toactivateanyofthefunctionsdescribedabove,pressanykeyormovepedals.
Toreset,holdtheRESETkeyfortwoseconds.
BelowaresampleandimagesoftheFitnesscomputerandatableoutliningmetricsthecomputerdisplayswithcorrespondingunitsandranges.
Updated:16October2017 ExerciseIntervention&OutcomesCore 22
Scale-zeroadjustmentLoosentensiondevicesothatthebrakebeltfeelsloose.Checkthatthependulumwillhanginverticalposition.
Scaleboard:Checkthattheindexonthependulum(2)weighisalignedwiththeindexatthe0-positiononthescaleboard.Ifadjustmentisnecessary,firstloosenthelocknut(1)andthenchangethepositionoftheboard.Tightenthelock-nutaftertheadjustment.Seefig:Scaleadjustments.
Kp-scale:Atthesametime,checkthatthekilopond-scale(5)totheleftoftheFitnesscomputer,shows0inlinewiththeindex.Toadjustthekp-scaleloosenthelockscrewforscaleindicator(9).Tightenthescrewfirmlyaftertheadjustment.Seefig:Calibration,Adjustmentkp-scalewindow.
Updated:16October2017 ExerciseIntervention&OutcomesCore 23
CalibrationAlthoughallErgometersarecalibratedatthefactory,theusermaywishtoverifythisbyperformingamechanicalscalecalibration.Tocalibratefollowthestepsbelow:
1. Removethebluebrakebeltfromthespring2. Leanthebikeforwardsothatthecalibrationweight,4kg(4)ishangingindependentlyfromthe
framecover3. Besurethatthescaleboardissettozerobeforetheweightisputon,seesection”Scale-zero
adjustment”4. Putthe4kgweightonthespring(6).Whencorrectlyset,itshouldbepossibletoreadthisweight
fromthecorrespondingplaceonthescaleboard(7).Seefig:Calibration
Shouldtherebeadeviationfromthecalibrationweight,adjustthependulumtothecorrectweightonthescalebymeansoftheadjustingweightinsidethependulum(3).Seefig:Scaleadjustment.Tochangethepositionoftheadjustingweight,loosenthelockscrew(8)onthebackofthependulumweight.Shouldtheindexofthependulumweightbetoolow,movetheadjustingweightupwardsintheweight.Iftheindexshouldbetoohigh,theadjustingweightismovedsomewhatdownwardsandlockedinthenewposition.Repeatuntilthecorrectreading(4kg)isachieved.
Checkthecalibrationofthependulumweightonceayearorwhenneeded
Updated:16October2017 ExerciseIntervention&OutcomesCore 24
SampleDataSheetforGradedExerciseTest
TIME DISTANCE SPEED HR RPM kP/meter/minàWatts(1kP/m/min=0.1634watts)
1:00
2:00
3:00
4:00
5:00
6:00
7:00
Updated:16October2017 ExerciseIntervention&OutcomesCore 25
3.3Generaltroubleshootingguide
MaintenanceSchedule:
Serviceaction:
• Manufacturersuggestusingisopropylalcohol.Useadampbut,notwetclothtocleanthesurfaceyouwishtodisinfect.
• Periodicallywipethesurfacewitharustpreventative,especiallywhenithasbeencleanedandthesurfaceisdry.Thisisdonetoprotectthechromeandzincpartsaswellasthepaintedparts(4timesperyear).
• Checknowandthenthatbothpedalsarefirmlytightened.Ifnot,thethreadinginthepedalarmswillbedamaged.Also,checkthatpedalarmsarefirmlytightenedonthecrankaxle,tightenifnecessary.WhentheErgometerisnewitisimportanttotightenthepedalsafter5hoursofpedalling(4timesperyear).
• Checkthatthepedalcrankissecuretothecrankaxle(4timesperyear).• Besurethatthepedalsaremovingsmoothly,andthatpedalaxleisclearofdirtandfibres(4
timesperyear).• Whencleaningandlubricatingbesuretocheckthatallscrewsandnutsareproperlytightened
(2timesperyear).• Checkthatthechainissnugandthereisnoplayinthepedalcrank(2timesperyear).• Checkthatpedals,chainandfreewheelsprocketarelubricated(2timesperyear).• Besurethatthebrakebeltdoesnotshowsignificantsignsofwear(2timesperyear).• Checkthatthehandlebarsandseatadjustmentscrewsarelubricated(2timesperyear).• Besurethatallmovingpartsascrankandflywheelareworkingnormalandthatnoabnormal
playorsoundexists.(I.e.playinbearingscausesfastwearingandwiththatfollowsahighlyreducedlifetime.)
• Checkthattheflywheelisplacedinthecenterandwithplanerotation.• Thecrankandflywheelbearingsarelongtermgreasedanddonotrequiresupplementary
lubrication.Ifproblemariseswithcrankorflywheelbearings,contactMonarkdealer.• Replacebatteriesifthecomputerisnoton/seemstobeproducingincorrectmeasurements
Updated:16October2017 ExerciseIntervention&OutcomesCore 26
3.4Contacts
1.)MonarkExerciseAB(Vabsbro,Sweeden)+46(0)281594940
2.) MichaelBusa–[email protected]
Updated:16October2017 ExerciseIntervention&OutcomesCore 27
Chapter4:Monark874E,weightergometer
4.1Introduction:MonarkErgometermodel874Eisasafe,easy-tousebikeforfitnesstestingandworktests.Ithasabrakesystemwheretheworkloadisdeterminedbyweightsaddedintheweightbasket.Thepatentedweightbasketsystemdoesnotrequirecalibration,theprecisionoftheweightsensuresthattheworkloadiscorrect.
Quickfactsaboutthe874E:
•Large,well-balancedflywheel20kg(44lbs)
•Adjustablesaddleandhandlebar
•Stableframe,solidsteeltube
•Wheelsforeasytransport
•Electronicmeterwithheartrate
Corecompetencies:
1) Understandingthebrakedeviceandbasicsofthetensioncenter2) BefamiliarwiththeFitnesscomputerandwhatmetricsitgenerates3) Reviewserviceactionsandunderstandwhatactionsyoumightneedtotakebefore/afterusing
thisequipment
Updated:16October2017 ExerciseIntervention&OutcomesCore 28
4.2Userinstructions
CycleadjustmentsSeatheightshouldbeadjustedtoacomfortableposition.Theappropriateheightistohavethekneeslightlybentwhenthesoleofthefootiscenteredoverthepedalaxlewiththepedalinthebottomposition.Toadjusttheseatheight,loosenthelever(1)ontheseattube.SeeFig:Adjustments.Alsonotethatthehandlebarstemshouldbeinsertedintotheframetubeatleast3inches.Thismeasureismarkedwith“MAX”onthestem(3).
FitnesscomputerTheMonark874EisequippedwithaFitnesscomputerthatdisplaysthefollowingmetrics:
1. Revolutionsperminute(RPM)2. Heartrateinbpm(HR)
a. Atthedisplayforheartrate(HR)aheartsymbolislitwhichmeansthatthemeteristryingtofindapulsesignalfromanexternalsource,chestbeltwithelectrodes,IfthemetercannotfindsuchasignaltheHRfunctionisautomaticallyturnedoffafter30seconds.Whenthefunctionisturnedofftheheartsymbolisnotlitanymore.Theheartratefunctioncanbeturnedonagainbypressingakey.
3. Exercisebouttimeinminutesandseconds(TIME)a. Thetimerstartsautomaticallywhenpedalsaremoved.MetervaluesforTime,Distanceand
CaloriescanbesettozerobypressingtheRESETbutton(seeimagebelowforRESTbuttonlocation)formorethantwoseconds.
4. Cyclingspeedinkmperhourormilesperhour(SPEED)5. Covereddistanceinkmormiles(DISTANCE)6. Workload(kp=weightbasket+weightinkg)canbesetwhichgivesareadingofcaloriesburned
(CAL)andpower(WATT).EnergyisexpressedinkJ(kilojoule)orcal(kilocalories,kcal).Conversion:1
Updated:16October2017 ExerciseIntervention&OutcomesCore 29
kCal=4,2kJ.Powerisdependentoncadence,orpedalingspeed,whichmakesitpossibletoadjusttheworkload/powerbyincreasingordecreasingcadence.
a. Togetcorrectreadingsforcaloriesandwattsthekpvalueontheelectronicmeterhastobesettothesamevalueastheworkloadthatistheweightofthebasketincludingtheweightsinit.Therubberplatesareincludedinthecalibratedweightofthe1kgweightbasket.
i. Forexample,theworkloadis3kg(weightbasket0,1kg+2x1kgweight).Pressthekpkeytotheleftonthemeter.Thelowerdisplaywindowisnowflashingandshowingfiguresinkp.Increaseordecreaseinstepsof0.1kpbypressingthekpbutton(arrowup)ortheRESETbutton(arrowdown)untilthereadingiscorrespondingwiththeactualordesiredkpvalues(workload)fromtheweightbasket.Afterthat,presstheCAL/WATTbuttontoeithershowtheCALorWATTfigures.Next,presstheCAL/WATTbuttontoeithershowtheCALorWATTfigures.Refertosampleimagesbelow.
ii. Thewattreadingsinthedisplayaredependentoncadence.Thewattscanaccordinglybeadjustedbyincreasingordecreasingcadence.Caloriesarecalculatedinreal-time.
Toactivateanyofthefunctionsdescribedabove,pressanykeyormovepedals.
BelowaresampleandimagesoftheFitnesscomputerandatableoutliningmetricsthecomputerdisplayswithcorrespondingunitsandranges.
Updated:16October2017 ExerciseIntervention&OutcomesCore 30
HeartrateUsingtheMonarkchest-strapheartratebeltresearcherscanmonitorparticipantheartrate.Thistechnologysenseselectricaloutputofthehearttocomputeheartrate.Properlyfittingthechest-strapisintegraltoprovidingaccuratereadings.
Beforeharnessingthechest-strap,cleanskinwherethebeltistobeplaced.Additionally,electrodesonthebacksideofthestrapshoulidebemoistenedwithwaterorspecializedgel.Thechest-strapshouldbesecuredatacomfortabletensionaroundtheparticipant’ssternum,seeFig:Placementofchestbeltforvisitation.Thebeltshouldbeplacedsuchthattheword“Monark”isoutwardfacing,upright,andcenteredontheparticipant’schest.InordertoproperlyconnectwiththeHRreceiveronthebike,thestrapshouldnotexceed100cmindistancefromthereceiver.Itisespeciallyimportantwhenfirstpairingthechestbeltwiththereceiver,thetwounitsarewithin60cmofeachother.
Figure:Placementofchestbelt
WingatetestingWhenusingthispieceforwingatetesting,resistanceintheweightbasketshouldbesetat0.075kgperkilogramoftheparticipant’sbodyweight.Whentestinghighly-trainedathletesconsiderusingahigherresistance,forexample1.0-1.3kgperkilogramofparticipantbodyweight.
Aftertheparticipanthascompleteda3-5minutewarmup,thetestshouldbegin.Afterthreesecondsofalloutpedaling,thebrakecordistobereleased,dosobylettinggoofcord.Thiswillapplyresistancetothesystem.The30-secondstestallowsforcalculatingthefollowingvariables:
1.) Peakpower(PP),inwatts2.) Relativepeakpower(RPP),inW/kg3.) Anaerobicfatigue(AF)4.) Anaerobiccapacity,inkilogram-Joules
Updated:16October2017 ExerciseIntervention&OutcomesCore 31
Adjustingthebrakecord/belttensionImportantnote:Toreceivecorrectworkloadmetrics,itisimportanttoplacetheweightbasketaccordingtothedescriptionthatfollows.Ifthebaskethangstoolow,itmaytouchtheflywheel.Ifthebasketistoohigh,wrongworkloadinformationwillbedisplayed.
First,checkthatthebrakebeltislyingcorrectlyontheflywheelbrakesurface,seeimage1belowforproperalignment.Next,put4kgintheweightbasket(4),seeFig:Brakedevice,below.Whenrotatingtheflywheelbyhand,orwhenpulling/releasingthebrakecord,thebasketshouldlift/dropbetween40mmand60mm.Ifthisisnotthecase,thebrakebelthastobeloosenedortightenedalittleatthetensioncenter.
Loosenthecordbracket(#2inFigureBrakedevice)sothatthecordlengthcanbeadjusted.Ifthebasketistoolow,shortenbytensioningthecord.Alternatively,ifthebasketistoohigh,lengthenthecordbyeasingtension.
Toshortenorlengthentension,turnthetensioncenter(3)approximately45degrees,andmakeappropriatemodificationstocordlength.Afterthattightenthebracketagain,andcheckthatthedistancebetweentheweightbasketandflywheelisbetween40and60mmwhenyoupullonbrakecord.
Image1:Brakebeltpositioningontheflywheel
Updated:16October2017 ExerciseIntervention&OutcomesCore 32
ReplacingthebrakebeltIfthebrakebeltneedstobereplaced,loosenthebrakecordandsetthebaskettoitsupperposition.Loosenthelockwasherthatisholdingthecord(#2)andremoveitfromthetensioncenter.Loosenorcutofftheknotontheotherendofthecordandthenremovethewholecordfromthebike.Whenassemblinganewbrakecord,firstenteroneendintotheholeinthetensioncenter,andtieaknotandlettheknotfallintothebiggerpartofthehole.Locktheendofthecordwiththelockwasher.
Updated:16October2017 ExerciseIntervention&OutcomesCore 33
SampleDatasheetforWingateTest
Time RPM
0:05
0:10
0:15
0:20
0:25
0:30
4.3Generaltroubleshootingguide
Updated:16October2017 ExerciseIntervention&OutcomesCore 34
4.4MaintenanceSchedule:
Serviceaction:
• Manufacturersuggestusingisopropylalcohol.Useadampbut,notwetclothtocleanthesurfaceyouwishtodisinfect.
• Periodicallywipethesurfacewitharustpreventative,especiallywhenithasbeencleanedandthesurfaceisdry.Thisisdonetoprotectthechromeandzincpartsaswellasthepaintedparts(4timesperyear).
• Checknowandthenthatbothpedalsarefirmlytightened.Ifnot,thethreadinginthepedalarmswillbedamaged.Also,checkthatpedalarmsarefirmlytightenedonthecrankaxle,tightenifnecessary.WhentheErgometerisnewitisimportanttotightenthepedalsafter5hoursofpedaling(4timesperyear).
• Checkthatthepedalcrankissecuretothecrankaxle(4timesperyear).• Besurethatthepedalsaremovingsmoothly,andthatpedalaxleisclearofdirtandfibers(4
timesperyear).• Whencleaningandlubricatingbesuretocheckthatallscrewsandnutsareproperlytightened
(2timesperyear).• Checkthatthechainissnugandthereisnoplayinthepedalcrank(2timesperyear).• Checkthatpedals,chainandfreewheelsprocketarelubricated(2timesperyear).• Besurethatthebrakebeltdoesnotshowsignificantsignsofwear(2timesperyear).• Checkthatthehandlebarsandseatadjustmentscrewsarelubricated(2timesperyear).• Besurethatallmovingpartsascrankandflywheelareworkingnormalandthatnoabnormal
playorsoundexists.(I.e.playinbearingscausesfastwearingandwiththatfollowsahighlyreducedlifetime.)
• Checkthattheflywheelisplacedinthecenterandwithplanerotation.• Thecrankandflywheelbearingsarelongtermgreasedanddonotrequiresupplementary
lubrication.Ifproblemariseswithcrankorflywheelbearings,contactMonarkdealer.• Replacebatteriesifthecomputerisnoton/seemstobeproducingincorrectmeasurements
4.5Contacts
1.)MonarkExerciseAB(Vabsbro,Sweeden)+46(0)2815949402.)MichaelBusa–[email protected]
Updated:16October2017 ExerciseIntervention&OutcomesCore 35
Chapter5:VeltonCycleErgometerbyRacermate
5.1 Introduction-TheVelotronisalaboratory-gradecycleergometerthatdelivershighaccuracyandrepeatability.BuiltwithRacerMate’ssignaturedurability,enjoythemostrealistic,sciencegradetrainingexperienceontheworld’smostadvancedcycleergometer.Velotronisusedworldwidebythefinestuniversities,sportssciencelabs,andcoachingcenters,ofteninconjunctionwithpowerfulWingatesoftware.TheonlybicycleergometereverapprovedforuseintheUSACyclingNationalTalentSearch,Velotronisalsotheultimatecyclingtrainer.
TheVelotronisequippedwithaprecisionloadgenerator,providingsmoothvirtualshifting-pedalthroughthegearshift,withnowaitorfriction.Secondly,theradicalflywheeldesignallowsforendlesscoastingforatrueroadfeel,withacurrentdrivepowerfulenoughtodeliverthewidestspectrumofloadresistance.
POWERGENERATOR
FLYWHEEL
Updated:16October2017 ExerciseIntervention&OutcomesCore 36
5.2CoreCompetenciesGettingstartedwithVelotronCoachingSoftware
1.) ConnectEXTPCcabletotheappropriatespotonthebackofthepowergenerator
2.) Attachoppositeend(stereotoUSBadaptor)toUSBonExercisetestinglaptop
3.) OpenVelotronCoachingSoftware(CS)2008fromthestartmenu
4.) GotoSourcemenuandselectReal-time5.) Ifthisisyourfirstuse,select“New”onsub-directoryofReal-timeinSourcemenu
a. Createauserdatafile,enteruserinformationandsaveàOK.OnsubsequentusesselectOpentochoosethisuserfile.Ifthereismorethanoneuser,createaUserDataFileforeachpersonwhowillusetheVelotron.
Updated:16October2017 ExerciseIntervention&OutcomesCore 37
6.) GotoUtilitiesmenu,select“CommPorts”todeterminewhatserialportVelotronisconnected
to.Withthisinformation,selectSource-Real-time-CommPortandverifythecorrectportfoundinthetesthasbeenset.
7.) SelecttheOptionsmenutosetupyourManualErgosettings,ScreenCaptureOptions,Drag
Factor™settings,Reports,anddecidewhetheryouwanttorunthetestinMilesPerHourorKilometersPerHour.
8.) SelectStart,thenCharts,ManualErgo,Mode,orSpinScan.ByopeningChartsaselectionwindowforthecoursesortestsyouwanttouseforyourtestwillappear.Selectacoursefromthe“Courses”folderfromtheVelotronCS2008directory.ThenclickonOpen,theChartscreenwillappearwiththisselection.
9.) ByselectingacoursethiswillopenthedesiredcourseintheChartsScreen,clicktheredorgreenbuttonsontheleft-handsideofthechartwindowtoenableordisablethevariouschannelsofdata.
10.) HitF1ontheVelotronHandlebarControllertostartthetimer.Thecourseortestwillstart.11.) Belowisageneraloverviewofthechartfeatures:
Updated:16October2017 ExerciseIntervention&OutcomesCore 38
a. FullarrayofdatatakendirectlyfromaVelotronsavedperformancefile.b. Current,PeakandAveragedisplayofdata.HRzone%andPulsePower™,Gearand
Gearing.c. Userselectablecolorgraphsthatcorrespondtothedatasection.d. TimeIndicatorBartopinpointexacttimecoordinates.e. Course/programprofileforRealTimeusetoindicatecurrentposition,orworkload.f. MultipleZoomLevelsinFileModeoperation.g. RPMzonepointertohelpmaintainafixedRPMduringanytest.Itisidealforfixed
wattagetests.Concludingatest
1.) Whenyouhavecompletedyourtest,youcaneitherresetthetimerandrunanewtest,orsavetheperformancedataasapermanentrecord.
a. Attheconclusionofaknownlengthtestadialogwillappearaskingyouto“ResetorSave.”
b. Ifthetestisopen-endedyouwouldpresstheF5(reset)buttontoprompta“Save,Reset,orCancel”(CanceladdedincaseyoupressedF5bymistake).ThesequestionscanbeansweredusingtheHandlebarControllerkeysshowninthedialogtoeliminatehavingtodismountthebike
Saving/Viewing/Exportingatest
PerformancefilesinCSaresavedas.velfiles.VelfileextensionsarecreatedwhetherthetestwasanERGfile,CRSfile,orsavedSpinScansession.TheperformancefileswillbesavedwithaUserName/Courseortesttype/Date/Time.velformat.1.) PressRESETafterthecompletionofatesttodisplaythe“Doyouwanttosave”dialog.
AnswerYes.2.) UsingtheHandlebarKeys,pressF1toSaveorF2nottosave.3.) Exportingthedata.SavedRaceFilesfromanyVelotronPCprogram(availableonyourhard
drive)canbeexportedasASCIItextfiles.Thiscanbedoneautomaticallyormanually.Ifyou
Updated:16October2017 ExerciseIntervention&OutcomesCore 39
alwayswantanexportperformancefilesafteratest(forusein3rdpartyapplications)checktheAutoExportoptionintheExportOptionsmenu.FormanualexportchooseFile,thenselecteithertheExportfunctionorchangetheExportOptionsasneeded(check-markboth”shortformat”&“commadelimited”intheoptionsifyouplantousetheseexportfilesinaspreadsheet).Youwillbeaskedtoselecttheoriginalsavedperformancefile.Youcanthenbrowsetoanyfolderwherethefilemayreside.Onceexported,thesefileshavethesamenameastheoriginalfile,butwithanew.txtfileextension.
4.) Toopenaperformancereport,selectFile,OpenReport,andnavigatetothedesiredfileandselect.
FileMode,viewingsavedracefiles
1.) Select“SavedFile”fromtheSourcemenu.2.) SelectOptionsandcheckoruncheckMetricasdesired.3.) SelectStart,thenCharts,andyouwillbepresentedwithawindowwhereyoucanbrowseto
anyfolderfromthe“LookIn”dropdownwindowtoselectasavedperformancefile.Thiswillopeninthebackground,thereforeminimizeCS2008andopenthisdialoguebox.Oncethefileisfound,highlightitandselectOpen.Thechartscreenwillappearwiththeracedatashowinginit’sentirety.Toopenanotherfile,pressCtrl-O.
4.) Clicktheredorgreenbuttonsontheleft-handsideofthechartwindowtoenableordisablethevariouschannelsofdata.
Updated:16October2017 ExerciseIntervention&OutcomesCore 40
5.) Youcannowmovethemouseleftandrightoverthedataandzoominandoutwiththeleftandrightmousebuttons.Right-clickyourmouseforFullZoomIn,andscrollforwardandbackthroughthedatabymovingtheTimeIndicatorBarfromtheleft-handorright-handsideoftheChartWindow,orestablishaZoomWindowusingtheleft/rightandup/downarrowkeysofyourPCkeyboardandhitEntertoactivatethislevelofzoom
Creatinganewcourse/programBecausecoursefilesmustretaintheformatasshownintheimageonthepreviouspage,itisrecommendedtoeditoneofthepreexistingcourseswhenmakinganewcourse.Thiswillalmostguaranteeausablecourseeverytime.
1.) TocreateanewcoursefromanexistingcoursewhileinNotepad,firstopenapreexistingcourse.
a. Toopenapreexistingcourse,openNotepadfromtheStartmenub. InNotepad,clickFile,thenOpenc. NavigatetoVelotronCS2008folderontheCdrive,thenopentheCoursessubdirectoryd. BydefaultNotepadlooksonlyfor“textfiles”,butifyouleft-clickonthedownarrowin
the“Filesoftype:”sectionandchoose“AllFiles(*.*)”,thenallthefilesinthisfolderwillappear.SelectthecoursetheyoudesiretoamendandclickOpen
Updated:16October2017 ExerciseIntervention&OutcomesCore 41
e. WhenthecourseyouwanttoeditisopeninNotepad,renameitbyclickingFile,Save
As.TypeanewnameforthiscourseandchooseAllFiles(*.*)extensionwhensavingnewcourse.
f. LeftclickSavetomakea“carboncopy”of,andclose,thefirstopenedcourse.g. Nowyoucanproceedwithchangestomaketoyournewcourse:
i. Generally,toamendacoursechangeparametersfromtheoriginalcourse.Followtheformattinginstructionsoutlinedbelow:
h. Whenfinishedediting,clickFile/Save.
Updated:16October2017 ExerciseIntervention&OutcomesCore 42
5.3GettingstartedwithVelotronWingateSoftware
1.) GototheStartmenuandopenVelotronWingate
2.) GotoClientmenuandselectNewDatabase
Updated:16October2017 ExerciseIntervention&OutcomesCore 43
3.) Next,gotoClientmenuandselectNewClienta. Enterthedatabasethatyouwantdatatobesavedto,entersubjectantropometric
measuresandapplyoneofthefollowing:1.)torquefactor,2.)Torquevalue,3.)Droppedweight
b. OKtosave
4.) Now,gotoWingatemenu,andselectCreatenewprotocol
a. Createanameforprotocolandedittestingparametersb. Oktosavetodesiredpathway.Suggestedtosaveinclientdatafolder
Updated:16October2017 ExerciseIntervention&OutcomesCore 44
5.) SelectWingatemenuandclickGraphScaling.EditWattsandHR/CadenceScaling6.) CheckthatVelotronispluggedintocomputer(seeabove).SelectSetupmenuandclickrunTest
CommPorts.NotewhichComminconnectedtoVelotron.ClickOK.Next,clickSetupmenu,andselectCommPort,verifythatthissettingmatchesTestCommPortsresults.Seeabovesectionforimages.
7.) GotoClientmenu,selectLoadClientandnavigatetoclientdata.a. GotoC-drive,VeloTronWingate2008folder,datafolder,selectyourdesireddatabase
folder,andselectandopenclientfolder
Updated:16October2017 ExerciseIntervention&OutcomesCore 45
8.) Now,selectLoadProtocolfromWingatemenu.Navigatetotheprotocolyoumadeinstep4.
9.) Next,gotoWingatemenuandclickRunWingatea. TheVelotronisnowis“Warm-upmode.”Theclientcanwarm-upusingthepreselected
load-or-byusingtheUp/DownbuttonsontheWingateScreenorthoseontheVelotronHandlebarControlleradjusttheworkloadtosuitthewarm-uprequirements.Duetox-axisscalinglimitationsthereisonlyabout20secondsofwarm-upshown,butthewarm-upperiodpriortostartingtheclockisopen-ended.
Updated:16October2017 ExerciseIntervention&OutcomesCore 46
10.) Whenyouarereadytobeginthetest,selecttheStartWingatebutton,thefirsticononthetoolbar
STARTa. Thewarm-upphasethatyouselectedwillbegin,thenaredcountdowntimerwill
appearintheupperrighthandcorner,indicatingwhentheloadwillbeappliedb. Attheendofthetestaplotoftheclient’sperformancewillappear.ClicktheReset
button,fourthbuttonfromtheleftonthetoolbar.Selectthepathwaywhereyouliketosavetheperformance.Suggestedsavinglocationistheoriginalclientdatafolder
RESETFilemode,viewingandexportingsaveddata
1.) ReturntotheWingatesoftwarehomepage2.) Toviewdatainfilemode,intheTree-viewdiagramontherightsideofthepageselectyour
database,thenclient,andfinallytheclientperformancefilea. Rightclickontheperformancefileandchooseplot.APlotwillappearintherightpanel.
Updated:16October2017 ExerciseIntervention&OutcomesCore 47
3.) Toexportdata,returntothetree-viewdiagramandrightclick,butselectexport.a. Adialogueboxwillappearthatnoteswherethefilehasbeensaved
b. Theprogramwillexporttheperformancefileincommaseparatedtextformatintothesame
databasefoldertheoriginalperformancefilewasfound.Thefilewillhavethesamefilename,butwithaTXTfileextension.ThesefilescanbeusedMicrosoftexcel.
5.4Contacts
1.) Velotronforums:https://www.racermateinc.com/help-center-faq/2.) MichaelBusa–CoreDirector
413-577-0574 [email protected]
Updated:16October2017 ExerciseIntervention&OutcomesCore 48
Chapter6:Quinton-55EKGStressSystem
6.1.Introduction-ImportanceofElectrocardiogramsAstheheartisamuscularpumpcomprisedoffourchambers,itrequireselectricalstimulationtocontrolthemuscularcontractionsandflowofbloodthroughoutthebody.Anelectrocardiogramisamethodtypicallyusedtoassesselectricalconductionintheheart.Thissimplenon-invasivetestallowstocheckfornormalrhythms,ratesandproperconductionofelectricalimpulsethroughoutthecardiactissue.ThismanualwillallowaresearchertounderstandandutilizetheEKGsysteminordertoobtainEKG’sformedicalinterpretation.Forallsubjects>70yearsofageorifasubjecthas≥2cardiovascularriskfactors,anEKGmustbecompletedpriortobeginningexerciseprotocols.
6.2EKGSettingsUnlessspecificallyrequestedotherwise,EKGsshouldberunwiththefollowingstandardsettings:PaperSpeed:25mm/sec
CalibrationStandard:1mV=10mmRhythmLead:DefaultLeadII
6.2.1SkinPrepPatientswithexcessivechesthairshouldbeshavedattheelectrodesite.Abradetheseareaswithfinesandpaperoranabrasivepad,andthencleanwithalcohol-saturatedgauze.Allowtheskintoairdrybeforeplacingelectrodes.Besurethesubjectdoesnothaveanyjewelrythatcouldinterferewiththetracings(necklaces,piercingsetc.).
6.3LimbLeadPlacementRA(White)–RightArmelectrodeisplacedonadistalportionoftherightlateralsideoftheupperarmbelowtheshoulder.LA(Black)–LeftArmelectrodeisplacedonadistalportionoftheleftlateralsideoftheupperarmbelowtheshoulder.RL(Green)–RightLegelectrodeisplacedontheinsidecalf,midwaybetweenkneeandankle.LL(Red)–LeftLegelectrodeisplacedontheinsidecalf,midwaybetweenkneeandankle.
6.4PrecordialLeadPlacementV1(COLOR)–4thintercostalplaceattherightmarginofthesternum.V2(COLOR)–4thintercostalplaceattheleftmarginofthesternum.V3(COLOR)–MidwaybetweenV2andV4(ontopofthe5thrib).V4(COLOR)–5thintercostalplaceattheleftmid-clavicularline.V5(COLOR)–AtthehorizontallevelofV4,attheleftanteriorline.V6(COLOR)–AtthehorizontallevelofV4,atthemid-axillaryline.
Updated:16October2017 ExerciseIntervention&OutcomesCore 49
NOTE:LeadplacementdoesaffecttheEKGwaveform.Whenthelimbleadsareplacedonthetorso,waveformchangesmightbeseenintheQRSamplitude,axisshiftoccurs,Qwavescanbeseen,andTwavesmightappearflippedorflattened.Thesechangesareclinicallysignificantinthattheyareassociatedwithcardiacischemia.Ifanon-standardleadplacementisused,notethevariationintheEKGcommentfield.
6.5 InputtingaNewPatient
a. ToinitializetheQ-stressprogram,clickontheicononthedesktop
Updated:16October2017 ExerciseIntervention&OutcomesCore 50
b. Tofirstsearchforpreviouspatients,clickthepatienticon
c. SearchforthepatientnameorMRN(MedicalRecordNumber)
Updated:16October2017 ExerciseIntervention&OutcomesCore 51
d. Ifpatientinformationisnotfound,Select“New”andinputpatientdata
e. Thefollowingscreenwillallowyoutoenterallpossibledemographicinformation.ClickOKwhencompletedtocompletethedataentry.Alsonotethereshouldbenochangesmadetoprotocol,procedure,orreportunlessinstructed.
Updated:16October2017 ExerciseIntervention&OutcomesCore 52
6.6 CheckLeadPlacementandArtifact
a. Afterpatientdataiscomplete,clickonthelightblueicontocheckleadplacementandconnections.
b. Click“Start”andwaitfortheiconstochangecolor
NOTE:Leadsshouldbeinthe“good”range
c. IfLeadsarenotwithinacceptable“good”range,removeleadsandpreptheelectrodesitesagain
Updated:16October2017 ExerciseIntervention&OutcomesCore 53
d. Clickexitafterallleadsareproperlyplaced.
6.7 BaselineWaveforma. Aftertheleadsareplacedandthepatientdataisinput,clickontheresticontobeginrestingwaveforms.
Updated:16October2017 ExerciseIntervention&OutcomesCore 54
b. ThescreenwillbegintodisplayheartrateandEKGwaveforms.Allowsabout10secondstopassbeforecollectingarestingEKG.
c. PressF5toprintoutafull12-leadrestingEKGd. PressF6toprintthecurrent3-leadscreene. Tobeginexerciseprotocolsclickonthe“startbelt”iconfollowedbythestartexercisebutton
Updated:16October2017 ExerciseIntervention&OutcomesCore 55
f. Toendexerciseprotocolsclickthestartrecoverybuttonfollowedbythe“stopbelt”button
g. ToreviewthecollectEKGsclickthereviewtabandclick“Yes”
Updated:16October2017 ExerciseIntervention&OutcomesCore 56
h. Clickyeswhenenteringreviewmode.Thiswillendthetest.
i. Whenthetesthasended,ifareportisneedclick“PartialReport”
Updated:16October2017 ExerciseIntervention&OutcomesCore 57
Thiswillprintasummaryreportoftheprocedure
j. Whenfinishedreviewing,click“Exit”tosavethetest.Whenprompted,besurePDFisclickedyes,andpress“OK”
k. Toexitcurrentpatient,press“Exit”again
Updated:16October2017 ExerciseIntervention&OutcomesCore 58
Afterobtainingabaselinewaveform,theEKGshouldbeinterpretedbyamedicalprofessional.Theinterpretationwillbecompletedbyeitheracertifiedclinicalexercisephysiologist(CCEP)ormedicaldoctor(MD)withinthedepartment.Anexercisephysiologistmayinterpretthebaselinewaveformswithsomerestriction.IftheEKGisdeterminedtobesatisfactoryandwithoutcontraindicationstoexercise,theexerciseportionwillresumewithoutreferralfromthemedicaldoctor.However,iftheexercisephysiologistdeterminesthatthebaselinewaveformisquestionableinanyway(basedoncurrentACSMrecommendationstoexercise),themedicaldoctorshouldbeconsultedtoverifytheEKG.Thedoctorwillthendeterminewhetherornottoproceedwithtesting.QuestionablebaselineEKG’scanincludebutarenotlimitedto:frequentectopy,abnormalheartrates(<40bpmor>100bpmatrest),STsegmentabnormalities<2mmdepressionorconductiondelays.Subjectswithsignificantabnormalitieswillbeexcludedfromthestudy.SignificantEKGabnormalitiesmayincludebutarenotlimitedto:dysrhythmias,electronicallypacedrhythms,conductionblocks,>1mmSTsegmentelevationsor>2mmSTsegmentdepressions.
6.8 ExerciseProtocols
AsthereisnocontinuousEKGmonitoringduringCADENCE-Adults,protocolisnotsignificantwhencollectingEKG’s.EKGwillbean“asneeded”proceduretoensuresafety.
6.9 Follow-Up
Ifatanytimeduringtestingprotocols,asubjectbecomesdizzy,lightheadedorcomplainsofchestpain,afollow-upEKGshouldbeperformedalongwithbloodpressures.InterpretationoftheEKGshouldbeimmediatelyreferredtotheExercisePhysiologistorMDonsite.
Updated:16October2017 ExerciseIntervention&OutcomesCore 59
6.10 Cleaning
Allwiresandconnectionsshouldbewipeswithdisinfectingwipes.Usedelectrodesshouldbethrownaway.Thetreadmillandanyotherequipmentassociatedwiththetestshouldbethoroughlycleanedbetweenparticipantswithalcoholwipes.
Updated:16October2017 ExerciseIntervention&OutcomesCore 60
ChapterX:
BiodexSystem4ProDynamometer
Turningthingson:
1.MakesuretheBiodexispluggedin.
2.Turnonthemainpowerswitch(locatedonthebottomleftofthebackoftheBiodexcart)byflippingitfromOtoI.
3.Turnonthepowerforthecomputeranddynamometer(greenswitchesonthebottomrightofthebackoftheBiodexcart)byflippingthemfromOtoI.
4.Turnonthecomputertower.
BasicsofsettinguptheBiodex
ThedynamometerandtheBiodexchaircanbetranslatedinspacesidetosideandverticallyandrotated.Thefollowingimagespointouthowtomovethechairanddynamometer.
Updated:16October2017 ExerciseIntervention&OutcomesCore 61
Tiltthedynamometerheadbylooseningthisknob.
Comfortstopbutton.Pushingthisbuttonwillimmediatelystopatestandexitoutoftestingmode.Useasakillswitchonly.
Updated:16October2017 ExerciseIntervention&OutcomesCore 62
Thisknobisusedtotilttheseatback
Thisleverisusedtorotatetheseat
Whenputtingattachmentsondynamometer,thesedotsmustalign.Youcanrotatethedynamometerdotusingthebluebuttonsonthedynamometerhead
Watchout,fingers/handsarecrushable!
Attachmentswithoutanutonthemwillusetheattachmenthandlestoredonthedynamometerhead(seeabove)Lettercorrespondstoside
(left/right)you’retesting
Updated:16October2017 ExerciseIntervention&OutcomesCore 63
RunningaTestinBiodexAdvantage:
TheseinstructionswilltakeyouthroughrunningatestontheBiodex.IfyouwishtousetheExerciseorBiofeedbackmodes,pleaseseetheinstructionsintheBiodexusermanualorhelpdocumentation.
1.Whenthecomputerbootsup,itwillaskyoutoinitializethedynamometer.Removeanyattachments(theattachmentpostwillrotate)andclickOK.Thedynamometerwilltheninitialize.
2.BiodexAdvantageusuallybootsuptothelasttestthatwasrun.
A.Ifyouneedtoseehowtosetupthebiodexattachmentsforagiventest,clickthe“Anatomical”button.Inthewindowthatpopsup,youcanselectthejointandjointmotionyou’retestingtoseeavideoofhowtosetupthemachine.
Updated:16October2017 ExerciseIntervention&OutcomesCore 64
B.Beforestartinganewtestyouneedtoloada“patient”andselectaprotocol.Follow thebuttonsontheleftfrom1-4.Asyougothroughthesteps,theactivewindowwillbe yellow.
1.Selectajointandmotion2.Selectamethodtoviewtheinstructions
3.Clickview
Updated:16October2017 ExerciseIntervention&OutcomesCore 65
3.ThePatientwindowgivesyouoptionstoopenanexistingpatient,addapatient,oreditanopenpatient.
Tostartatest,clickherefirst
Toviewinstructionsforhowtosetupagiventest,clickhere
Notreadybecauseyouhavetogothroughthetestsetupfirst
Updated:16October2017 ExerciseIntervention&OutcomesCore 66
A.Ifyouclick“Open,”youcanseealistofexistingentriesandselectonetocreateanewtestingsessionfor.Ifyouselectapatientfromthislist,youcanbeginanewtestbyclickingthe“NEW”buttonorEDITthepatient’sinformation.Ifapatienthasexistingtests,youcanREPEAToneofthese,printareportforanexistingtest,orDELETEanexistingtest.
B.Ifyouclick“AddPatient,”thewindowwillchangetotheviewbelow.Atminimum, youmustfilloutanyinformationboxesthatareinred.NOTE:Personalidentifying informationshouldneverbeenteredintotheBiodexsoftware.Onlyenterstudy participantIDsintonamefields.Neverenternames,birthdates,age,address,phone number,orotherinformationthatcouldidentifyanindividual.Weightandgenderare requiredbyBiodexandcanbeentered.
Updated:16October2017 ExerciseIntervention&OutcomesCore 67
C.Afterenteringallnecessarypatientinformation,click“Save,”andthenclickon “Protocol.”
Updated:16October2017 ExerciseIntervention&OutcomesCore 68
4.Hereyoucanloadaprotocol,addaprotocol,oreditaselectedprotocol.
Optionbuttons
Updated:16October2017 ExerciseIntervention&OutcomesCore 69
A.Ifyouclickthe button,thefollowingwindowwillpopup.Hereyoucanviewbuilt-inorpreviouslydesignedtestingprotocols.
Updated:16October2017 ExerciseIntervention&OutcomesCore 70
B.Ifyouwanttocreateanewprotocol,clickthe button.Thiswillactivatethe editableportionsoftheprotocolpane.
i.Dependingonthe“StudyType”and“Mode”optionsyouchoose,theotheroptionsintheblueboxabovewillchange.Forexample,ifyouchangefromIsometricmodetoIsokineticmode,insteadof“#Positions,”thegreenboxwillbelabeled“#Sets.”
Ifyouclickthis,theseoptionsbecomeavailable
Updated:16October2017 ExerciseIntervention&OutcomesCore 71
5.Afteryouselectaprotocol,clickonthe“SetROM”button.Atthispoint,thechairanddynamometershouldbeinposition,thecorrectattachmentshouldbeonthedynamometer,andtheparticipantshouldbeseatedandstrappedin.
A.Belowisthe“SetROM”window.Followtheblinkingredtextthrougheachstep.
1.First,selectthesidebeingtested
Usingthisoptionisnotrecommended(generallyonlyavailablewhenswitchingfromonesidetotheotherinbilateralprotocols)
2.ClickCLEARLIMITS.SetAWAYandTOWARDlimitbymovingparticipant’slimbtothesepositions,holdingthere(seebelow),andclickingtheSETbutton.Notethatthissetstherangethattheparticipantwillbetestedin.
3.MovelimbtoAnatomicalReferenceposition,holdthere,andclicktheCalibratePositionbutton.AnatomicalReferencecanbechangedbyclickingonthebuttonthatreads“90”here.
4.MovelimbtoAWAYlimit,holdthere(usingholdbutton),haveparticipantcompletelyrelaxlimb,andclicktheLimbWeightbutton.Ifyougetanerroronthis,seetextbelowini.
5.ClickContinueViewSetupbuttonwillshowaschematicofwhat“AWAY”and“TOWARD”are,aswellashowanglesarereferenced
Updated:16October2017 ExerciseIntervention&OutcomesCore 72
i.Clicking“LimbWeight”maygivethefollowingerror:
ii.IfthisoccursandthelimbwasweighedattheAWAYlimit,youwillneedtoincreasetheAWAYlimit,movethelimbtothisposition,andweighagain.IftheAWAYlimitisalreadyattheparticipant’slimitforthegiventest.
6.Afterhitting“Continue”intheSetROMwindow,theprogramwillmoveontotesting.TheprogramwillswitchtothefollowingwindowandwaitthereuntilyouhitSTART.IfStartisnotanoption,itmeansyoumissedsomethingbetweenstep3andthisstep.Whenyouarereadytobeginyourtest,clickSTART.
A.Ifyouwanttochangethevisualizationforatest(e.g.,bargraphinsteadofline),clickonthe
options button.
Holdbuttoncanholdparticipant’slimbinaposition.
Updated:16October2017 ExerciseIntervention&OutcomesCore 73
B.First,atrialrepswindowwillappear.Here,theparticipantcan practicethemotionandspeedofthetestbeforetheprogramstarts collectingdata.Whenyou’rereadytostarttherealreps/sets,have theparticipantholdtheirlimbinthestartposition(usually TOWARDlimit).
C.Oncetheparticipantholdsatthestartpositionforasufficient time,thetestwillbegin.Youwill seethiswindow:
Startbutton
Updated:16October2017 ExerciseIntervention&OutcomesCore 74
D.Oncethetestbegins,youwillseereal-timedataofthetorquetheparticipantis generating,aswellasprogressthroughthecurrentsetviathePieChart.
E.Duringrestperiodsbetweensetsorsectionsofaprotocol,the windowtotherightwillcountdowntothenextset.
7.Whenatestisdone,youcanclickontheReportbuttonortheCurvebuttontoviewresults.
A.Thereportwindowgivesoptionsforgeneratingstandardreports.Thisisaneasywaytogiveparticipantsfeedbackontheirperformance.Hereyoucanpreviewareportorprintareport.Ifyouwillnotbeprintingareportandalltestsarecomplete,thetestingsessionisover.
Updated:16October2017 ExerciseIntervention&OutcomesCore 75
B.ToanalyzeresultsinBiodexAdvantageorexportdatatoanASCIIformat,clickontheCurvebutton.
i.Therearemanyoptionsforreal-timeanalysisinthecurvewindow.Thesearepointedoutbelow.
ii.DataexportedfromtheCurvewindowcanbetakenoffthecomputeronaUSBdriveandanalyzedusingothersoftware(e.g.,MatLab,Excel).
Clickheretoexportdata
Enablecursors(bluelinesingraphabove)andusebuttonsheretoselectrangesofdata
Optionsforadjustinggraphdisplayandperformingsomereal-timeanalysis.Wedonotrecommendusingthe“Filtered”optionexceptfordemonstrationpurposes
Changebetweentestshere
Updated:16October2017 RoomCalorimeterCore 76
MANUALOFPROCEDURES
HumanMotionCore
InstituteofAppliedLifeSciences
LifeScienceLaboratoriesS360240ThatcherRoad,
Amherst,MA01003
atthe
UniversityofMassachusetts,Amherst
Updated:16October2017 RoomCalorimeterCore 77
Chapter1QualisysOqus700+/500+w/AMTIForcePlatformintegration
Chapter1.1-CreatingaNewProjectinQTM
WhenbeginninganewstudyyouwillneedtocreateanewprojectinQTM.Thisprocessensuresthatyoudon’tinadvertentlyalteranotherstudy’sprojectsettingsandalsocreatesadesignateddirectoryforyourstudy’sdata.
1.First,openQTM
2.AwindowdisplayingcurrentprojectswillpopupwhenyouopenQTM.Whencreatinganewproject,click“Newproject…”
3.Thefollowingwindowwillpopup.In“Projectname:”typethenameofyourstudyorthenameyouwouldliketobeappliedtoyourdirectory.Notethatwhenyouchangethisname,theendofthe“Projectfolderpath:”willchange.Ifyouwouldliketheprojectfoldertobelocatedsomewhereotherthanwhereisindicatedbythe“Projectfolderpath:”youcanbrowsetothatlocationusingthe“Browse…”button.Whenyournameandpathareset,select“Settingsimportedfromanotherproject”andhit“OK.”
Updated:16October2017 RoomCalorimeterCore 78
4.Anotherdialogwindowwillopenaskingyoutoselecta“Settings”filetobaseyournewprojecton.NavigatetoC:\Users\Qsys02\Documents\Project_Starterandselectthe“Settings”filefoundthere.Theseprojectsettingshavebeencreatedspecificallytoactasastartingpointfornewprojects.
Typeyourdesiredprojectnamehere
Checkthispath
Usethisoption
Updated:16October2017 RoomCalorimeterCore 79
5.Afterclicking“Open,”QTMshouldstartup.EnsurethatyouareinyournewprojectbycheckingthetopbarontheQTMwindow.Itshouldnowread[Project:Yourprojectname]
Additionally,thereshouldnowbeafolderatthepathyoupreviouslydesignatedthathasthefollowingfolders/filesinit:
6.Afterverifyingthatyournewprojecthasbeencreatedcorrectly,youcanbeginchangingsettings.Todo
this,firstclickontheprojectoptions button.Thereareseveralprojectsettingsthatyoumaywishtochange(NOTE:afteranychangeismade,youmusthit“Apply”).Thissectioninthismanualisnotall-encompassing;forfurtherdetailsseetheonlineQualisysusermanualsection“Projectoptions”byclickingHelp>Helptopics.
InputDevices.Hereyoucanchangewhichdevicesyouwillbecollectingwith.Oquscamerascannotbedisabled,butinthissettingwindowyoucanselectordisableforceplatecollection(USB-2533)andEMGinputs(MEGA,Noraxon).
Updated:16October2017 RoomCalorimeterCore 80
Markercapturefrequency.Changethisnumbertoadjusthowmanyframesofmarkerdataarecollectedpersecond.Capturefrequencyneededdependsonspeedofmovementbeingcaptured.
Timingoptions.Usethesesettingstoactivateanexternaltrigger.Thismaybedesiredifyouaresynchronizingmultipledataacquisitionsoftwares(e.g.,QTMandEMGsoftware)orifyouwishtotriggeradatacapturefrominthecollectionspaceinsteadoffromthecomputer.
Updated:16October2017 RoomCalorimeterCore 81
Camerasettings.Inthistabyoucanadjustcapturesettingsforindividualcameras.NotethatyoucanadjustsettingsformultiplecamerasatatimebyholdingCtrlwhileclickingoncameranumbersoryoucanadjusteachcameraindividually.Ingeneral,camerasettingsshouldnotneedtobechanged.Instanceswhentheymayneedtobeadjustedincludetheuseofadeviceinthecapturespacethatisveryshinyorverysmallorverylargemarkers.IfyouwishtochangecamerasettingsitisrecommendedthatyoureadtheQualisyshelptopics.
Updated:16October2017 RoomCalorimeterCore 82
Analogsettings.Hereyoucanchangethecapturerateforanyanalogdevicesyouincludeinyourproject.Sampleratesmustbeamultipleofthecameraframerateandcanbechangedintheboxoutlinedbelow.Therearealsosettingshereforremovingoffsetfromanalogdata(yellowbox).Usecautionifactivatingtheseoptionsasthesoftwareoffsetremovalassumesthatthereisnothingontheforceplateswhenitremovesanybaselineoffset,afeaturethatmaybeproblematicifsomeonehappenstocrossoveraforceplateduringtheoffsetremoval.
Updated:16October2017 RoomCalorimeterCore 83
3Dtracking.Hereyoucanadjustsettingsforerrortolerancein3Dmarkertracking.
“Predictionerror”isthemaximumdistancebetweenamarker’spredicted locationandthecapturedlocationofthemarker.Ifthecapturedlocationis outsideofthispredictedlocation,thesoftwarewillnotreconstructit.Increasethis valuetoincreasethetoleranceinreconstruction.
“Maximumresidual”isthemaximumdeviationincamera2Dviewsofamarkerthatcanbeconvergedintoasinglemarkertrajectory.Qualisysrecommendsthatthisvaluebe2-5timestheaverageresidualvaluesfromthecalibrationresults.
“Minimumtrajectorylength”istheminimumnumberofframesofdatato createatrajectory.Dataseenforfewerframeswillnotbeassignedtoatrajectory. Theminimumvalueforthisoptionis2.
“Autojoin”automaticallyjoinstrajectoriesthatappeartobelongtothesame markerin3Dspace.Thismayreducethenumberofpartsinatrajectoryandmay maketrackingeasier.Thisdoesnotautomaticallyfillgaps.
Trajectories.Hereyoucanadjustsettingsforautomaticallyfillinggaps.
Updated:16October2017 RoomCalorimeterCore 84
ForceData.Inthesesettingsyoucanselectwhichforceplatesyou’dliketocollectwithandcheckadvancedforceplateoptions.
GUI.InGUI,2DViewSettings,and3DViewSettingsyoucanadjustsettingssuchasthereal-timescreenrefresh,andtheappearanceofthescreenin2Dand3Dmodes.
Miscellaneous.HereyoucanadjustfoldersettingsandQTMstartupsettings.Pleasedo notadjustthe“Globaldirectories”options(inFolderOptions)orthe“Automaticallyloadaprojectatstartup”options(inStartup)asthesechangedefaultsforallusers.
ProjectoptionsareautomaticallysavedtoyourprojectSettingsfilewhenyouclick“Apply”andthen“OK.”Itisrecommendedthatyoualwayspilotdatacollection,processing,andanalysisusingyournewprojectsettings.Thiswillensurethatyoursettingsaresufficienttocapturetheinformationyouneed.
Updated:16October2017 RoomCalorimeterCore 85
Chapter1.2
QTMDataCollection
Ifcollectingdataonaparticipant,turningthesystemonandcalibrationstepsshouldbeperformedbeforetheparticipantarrives.Camerasandforceplatesshouldbeallowedtowarmupforatleast20minutesbeforecalibrating.Turningthingson:
1.Turnondatacollectioncomputer.
2.Turnonpowerstripforcameras(locatedontowertoleftofdatacollectioncomputer)
3.Makesureforceplatesareon(3boxesdirectlytotheleftofthecomputermonitors).
4.Find/addnewparticipantfoldertoprojectfolder(C:\Users\Qsys02\Documents\yourprojectfolder\Data).
5.OpenQTM
Powerswitch
Manualzeroingbutton
Updated:16October2017 RoomCalorimeterCore 86
6.Loadyourproject
7.Ifyouneedtocreateanewproject,followinstructionsinChapterX.X
Updated:16October2017 RoomCalorimeterCore 87
Calibration:
1.Hitthe“new”button (topleftoftoolbar)
2.Clickon2Dview,checkcameraviewstomakesureeachcameraisseeing0markers.
3.PlaceLframeonforceplate1(endsmallFP)withthelongarmpointingtowardsthebigforceplate.Makesureitfitstightlytotheforceplatecorner(usedrop-downlegs),alignswiththeforceplateedges,andislevel.IfthelevelsareoffthetiltoftheLcanbeadjustedusingtheroundscrewsonthearms.AdjustthemuntiltheLislevelinbothplanes.
Updated:16October2017 RoomCalorimeterCore 88
4.Putthecalibrationwandtogether.Clickthecalibrationbutton .
5.Adjustcalibrationsettingsaspreferred.Forcalibratingavolumeforgait(all3forceplates+space)60secondsmaybenecessary.Usingacalibrationdelaywillgiveyoutimetohit“OK”andgetsetinthevolumebeforethestartofcalibrationcapture.Whenyouarereadytocalibrate,hit“OK.”
6.Calibratethevolume.It’sgenerallyrecommendedtodothisinasystematicmethod,coveringtheentiretyofthedesiredcapturevolumewiththecalibrationwandinawaythatallcamerasviewthewandoverthefullvolume(i.e.,don’thaveyourbacktoonewallofcamerasthroughouttheentirecalibration).
Updated:16October2017 RoomCalorimeterCore 89
7.Calibrationwillendwhenthesettimehaselapsed.Thiswindowwillpopup:
Standarddeviationofwandlengthshouldbelessthan1mm.Ifthiscriteriaismet,hit“OK”andputawaycalibrationequipment.
8.Calibrationwilloccasionallyfail.Ifcalibrationfails,checkcamera2DviewstomakesurenomarkersotherthantheLandthecalibrationwandarevisible.Ifnoothermarkersorreflectionsarevisible,startcalibrationoveragain.
Note:cameraswillpickupanythingreflective,includingrunningshoesandreflectivetagsonclothing.Makesureyoudonothaveanythingreflectiveonwhencalibrating.
Updated:16October2017 RoomCalorimeterCore 90
DataCollection:
1.Withnothingontheforceplates,zeroforceplates(orangebuttonsonforceplateboxes).Whenplatesarezeroed,theorangelightsshouldbeoff.
2.Clickthe“capture”button. Thiswindowwillpopup:
3.AdjusttheCapturePeriodtothedesiredlengthoftime.Changingthetimeboxwillautomaticallychangethenumberofmarkerframesbox.
4.Checkthe“Savecapturedandprocessedmeasurementautomatically”box.Makesurefolderthatisdirectedtoiswhereyouwouldlikeyourdatatobestored.
5.Changethe“Name”boxtowhateveryouwouldliketheprefixofyourfilestobe.Ifyouwouldlikefilestobeautomaticallynumberedasyoucollectthem,check“Addcounterstarting…”
6.Click“Options…”tocheckCameraSystemSettingstomakesureframerateandtriggersettingsarecorrect.
Updated:16October2017 RoomCalorimeterCore 91
7.Whenyouarereadytocollectatrial,hit“Start.”
8.Attheendofthecaptureperiod,thetrialwillopenin3Dview.Playthroughthetrialusingthebuttonsinthe
toolbar:
orbydraggingthetoparrowalongthebottomtimebar:
9.Labelingdata(tocreateanAIMmodel,useamotiontrial,notastatictrial):
Scanthroughthetrialtomakesureallnecessarymarkerandforcedatawerecaptured.Movedatatoatimepointwhereallmarkersarevisibleinthe3Dview.
Tolabeldata,firstopentrajectorywindowsbyclickingView>TrajectoryInfo>All.
Right-clickinthe“Labeledtrajectories”windowandselect“Loadlabellist…”
Selectyourmarkerlist.
Ifyouhavenotpreviouslycreatedalabellist,right-clickontheword“Trajectory”andselect“Addnewlabel.”Thiswillcreatea“New####”label.
Addasmanylabelsasmarkersinyourmarkersetandnamethemappropriately.
Whenalllabelsarecreated,right-clickinthe“Labeledtrajectories”windowandselect“Savelabellist…”
II.Tobeginlabelingmarkers,selectthefirstmarkerinthelabellistandclickthe
button(leftverticaltoolbar).
III.Inthe3Dview,clickonthemarkercorrespondingtothefirstlabel.InID mode,theprogramwillprogresstothenextlabelinthelabellistwhenamarkerisclickedon.
IV.Continueclickingonthemarkersinthe3Dviewthatcorrespondtothelabel highlightedinthelabellistuntilyoureachtheendofthelabellist.
V.Clickthe buttontoturnoffIDmode.
C.Trimtrialsothatonlylabeleddataareincludedusingthesidearrowsofthetimebar.Onlylabeleddataareincludedwhenallmarkerfilllevelsareat100%(inlabeledtrajectorywindow).
D.Click“GenerateAIMmodel”button
Updated:16October2017 RoomCalorimeterCore 92
E.Select“Createnewmodel”inthewindowthatpopsupandthenclick“Next.”
F.Inthenextwindow,verifythatthemodellookscorrect,andthenclick“Next.”
G.EnterafilenameforyourAIMmodelandclick“OK.”Inthefinalwindow,click“Finish.”
10.TohaveAIMappliedasyoucollectdata,inProjectOptions,gotoAIMandaddthemodelyoujustcreatedtothe“Appliedmodels:”window.Hit“Apply,”then“OK.”
11.Continuecollectingtrialsuntilfinished.
12.Ifyouwishtooverwriteatrial,settheNameandcountertomatchthetrialyouwishtooverwrite.Whenyouhit“Start,”awindowwillpopupaskingifyouwishtooverwritethetrial.Datawillbegincollectingwhenyouhit“Yes.”
13.Whendatacollectioniscompleted,turnoffcamerasandcopydatatoappropriatelocation(e.g.,externalharddrive,networkfolder,etc.).
Updated:16October2017 RoomCalorimeterCore 93
Chapter1.3-QTMDataProcessing
DataProcessinginQTM
Afterdataarecollected,theymustbetrackedandexportedfromQTMforfurtheranalysis.
Trackingmarkerdata
1.Navigatetolocationofmotioncapturedataonthecomputer.QTMfilesendin*.qtmandcanbeopenedonacomputerwithQTMsoftwareeitherbydouble-clickingonthefilefromWindowsExplorerorusingthe
“Openfile” buttoninQTM.
2.Beforetrackingyourdata,youmaywishtotrimthetrialtoonlytherangeofframesofinterest.Forexample,youmaywishtodothisifyouwantasinglestrideofdatafromatrialwhichincludesdataofyourparticipantwalkingallthewayacrossthedatacapturespace.
A.Toshortenatrial,clickanddragthearrowbuttonsoneitherendofthetimebar.
B.Forexample,hereIhavetrimmedoffthebeginningandendofatrialofdata.
C.Onceyoutrimatrial,markerinformationwillonlybedisplayedfortheframesinside thetimebararrows.Notethatthisrangecanalwaysbechangedlateronifyoudecideyou wanttotrackmoreofyourtrial.
3.Ifmarkersarealreadylabeled,skiptostep5.
Updated:16October2017 RoomCalorimeterCore 94
4.Ifmarkersarenotyetlabeled,alabellistwillneedtobecreated.Un-labeledmarkersappearredandalltrajectorieswillbelistedinthe“unidentifiedtrajectories”box.Ifyoudon’tseethetrajectoryboxes,clickView>TrajectoryInfo>Alltoopentheseboxes.
A.Right-clickinthe“Labeledtrajectories”windowandselect“Loadlabellist…” Selectyourmarkerlist.
I.Ifyouhavenotpreviouslycreatedalabellist,right-clickandselect “Addnewlabel.”Thiswillcreatea“New####”label.
II.Addasmanylabelsasmarkersinyourmarkersetandnamethem appropriately.
III.Whenalllabelsarecreated,right-clickinthe“Labeledtrajectories” windowandselect“Savelabellist…”
Updated:16October2017 RoomCalorimeterCore 95
B.Tobeginlabelingmarkers,selectthefirstmarkerinthelabellistandclickthe
button.
C.Inthe3Dview,clickonthemarkercorrespondingtothefirstlabel.InID mode,theprogramwillprogresstothenextlabelinthelabellistwhenamarkeris clickedon.
D.Continueclickingonthemarkersinthe3Dviewthatcorrespondtothelabel highlightedinthelabellistuntilyoureachtheendofthelabellist.
E.Clickthe buttontoturnoffIDmode.
F.Trimtrialsothatonlylabeleddataareincludedusingthebottomarrowofthetimebar:
G.Click“GenerateAIMmodel”button
E.Select“Createnewmodel”inthewindowthatpopsupandthenclick“Next.”
F.Inthenextwindow,verifythatthemodellookscorrect,andthenclick“Next.”
G.EnterafilenameforyourAIMmodelandclick“OK.”Inthefinalwindow,click “Finish.”
H.AftercreatinganAIMmodel,themodelcanbeappliedtonew,unlabeledfilesofthesame
participant/datacollectionsessionusingthe button.
Updated:16October2017 RoomCalorimeterCore 96
5.Ifmarkersarealreadylabeled,thenextstepistoclean-upthedata.Generally,thegoalofcleaning-upthedataistomakesuretherearenogapsinmarkerdata.
Note:itisrarethattherearenomarkersin“Unidentifiedtrajectories,”evenwhenallmarkersarelabeled.
Updated:16October2017 RoomCalorimeterCore 97
6.First,checkthatallmarkersarecorrectlylabeled(e.g.,theRightASISisactuallylabeledasRASIratherthanasRSHForsomethingelse).Tocheckthis,youcanclickonamarkerinthe“Labeledtrajectories”boxanditwillbehighlightedinthe3Dview(andviceversa).Alsonotethatlabelcolorsinthe“Labeledtrajectories”boxmatchthecolorofthemarkersinthe3Dview.
A.The“FillLevel”columnforeachmarkertellsyouwhatpercentofthetrialcontainsdataandalabelforthismarker.The“Range”columntellsyouduringwhichframesthelabeledmarkerappears.Attheendofdatatrackingmarkerdatayouwantallmarkersofinteresttobeat100%fortherangeofframesofinterest.
B.The+signnexttoamarkerindicatesthatthismarkertrajectoryismadeupofmultiple parts.Ifyouexpandbyclickingonthe+sign,youcanseehowmanypartsareinagiventrajectoryandwhatframestheyappearduring.
Updated:16October2017 RoomCalorimeterCore 98
7.Whenyouhaveconfirmedthatallmarkersarelabeledcorrectly,youcanbeginfixinggapsindata.Thereare2generalmethodsforfixinggapsindatawhicharedetailedbelow.
A.Option1:fillingagapinamarkertrajectorywhenthereisanunlabeledtrajectory available(bestoptionifavailable)
I.SometimesQTMdoesnotdoaperfectjobofidentifyingwhichmarkeris which,evenafteryoulabeldata.Thiswillresultinsectionsofdatawherethereis amarkerinthe3Dviewbutitisunlabeled(orinsomecases,islabeled incorrectly).Below,weseethattheLeftPSIS(LPSI)islabeledfor90%ofthe trial,exceptfortheframerange1960-2559.Thereisanunidentified trajectorythatcoincideswithmostofthisframerange.
II.ToaddthisunidentifiedtrajectorytotheLPSI,clickandholdoneitherthe unidentifiedmarkerinthe3DviewortheUnidentifiedtrajectoryinthe “Unidentifiedtrajectories”box,anddragittoLPSIinthe“Labeledtrajectories” box.
Updated:16October2017 RoomCalorimeterCore 99
III.NowtheunidentifiedtrajectoryhasbeenaddedtoLPSI.
IV.Sometimestheunidentifiedtrajectoryyouaretryingtoaddtofillagapina labeledmarkeroverlapswithexistingdataandyouwillgetthefollowing message:
a.Ifyouhit“OK,”aboxshowing3Dtrajectorieswillpopupinwhich youcanselecttherangeoftheunidentifiedmarkeryou’dliketokeep. Thistoolcanbeunwieldyanddifficulttouseattimes.
b.Ifyouhit“Cancel,”youcaninsteadmanuallyidentifytheoverlapping data.
Updated:16October2017 RoomCalorimeterCore 100
c.Intheexamplebelow,youcanseethattheunidentifiedtrajectorycontains2partsofdatawhichcovertherange1961-2558.LPSIalreadyhasdataforframes2553-2559.Todeterminewhichdatatokeep(thedataalreadyinLPSIfortheseframesorthedatafromtheunidentifiedtrajectoryfortheseframes),navigatetothisframerangeinthe3Dviewusingthetimebaratthebottomofthescreen.
d.Duringtheoverlappingframes,youshouldbeabletoseetwomarkers. Thesemayoverlapverycloselyinspace(asshownbelow)ortheymay not.
e.Atthispointyouneedtochoosewhichdatayouwanttokeep.Ifyouwishtokeepsectionsofonemarkerwhichgobeforeandafteranother(inthiscase,saywewanttotaketheunidentifiedmarkerfortherangebeforeandaftertheoverlappingLPSIdata),youmayneedtosplitonetrajectory.Todothis,navigatetothelasttimepointyou’dliketokeepofthetrajectoryofinterest,right-clickonthetrajectoryinthe“Labeledtrajectories”or“Unidentifiedtrajectories”box,andclick“Splitpartaftercurrentframe.”Thiswillcuttheselectedtrajectoryinto2parts.Youmayneedtodothisagainattheendoftherangeofinterest.Thenthesplitpartorpartscanbedraggedandaddedtothelabeledtrajectory.
B.Option2:gap-fillingagapinamarkerwhenthereisnounidentifiedtrajectorytoswapin
I.Below,weseeanexampleofaframeofdatawherethereisnoLPSI,andnounlabeledtrajectorytoswapintoitsplace.Incaseslikethis,youhavetoperformagapfill.
Updated:16October2017 RoomCalorimeterCore 101
II.Togapfilldata,selectthePartsoneithersideofthegap(below,therearedata from1-1959and1961-2552,butnodataforframe1960),right-click,andselect “Gapfilltrajectorywithpreview”
Updated:16October2017 RoomCalorimeterCore 102
III.Awindowdisplayingthe3Dtrajectoryoftheselectedmarkerwillpopupandshowhowthegapistobefilled.Followingtheinstructionsinthebottomofthiswindow,youcanzoominonthegapandadjusttherangeofframesyouwantthegapfilltospan.Notethatgapfillsmustbeginandendatlabeledmarkerdataasthefillisconstructedusingapolynomialfitbasedonexistingdata.
Gaprangeintrajectories(X,Y,andZ)
Rangeofgapbeingshown
Updated:16October2017 RoomCalorimeterCore 103
IV.Generalguidelinesforgapfillingdata:
a.Usecautionwhenfillinglargergaps(rangedependsonfrequencyof datacapture,speedofmovement,etc.).Gapfilleddataisnotrealdataand largegapfillsmaybeadistortedrepresentationofthetruemotionofa marker.
b.Adjustgapfillframessothattherearenotabruptshifts/discontinuitiesinthe3Dtrajectories.Oftentimesbeforeamarkerdropsoutandcreatesagap,thecamerasarehavingdifficultycalculatingitspositionandsothelast/firstframeortwoofexistingdatawillbeshifted.
c.Foranygapfilllargerthanafewframes,alwaysviewthe3Drepresentationofthegapfillbyplayingthroughthecorrespondingframeswhilewatchingthe3Dviewofyourdatatocheckforvisibleerrorsinthegapfilleddata.
8.Whenallmarkersarelabeledat100%,datatrackingiscomplete.
9.Afterdatatracking,datamustbeexportedfromQTMforfurtherprocessing.Filescanbeexportedinthefollowingformats:
-C3D:Visual3Dfileformat
-TSV:tab-delimiteddatathatcanbeviewedinexcelandreadintoavarietyof programs
-MAT:MatLabfileformat
-AVI:videofileofmarkerdata
A.Filescanbeexportedone-at-a-timeorinbulk.
I.Toexportfilesinbulk,clickFile>BatchProcess
a.WhenusingBatchProcess,thefirstwindowtopopupasksyouto selectthedesiredfiles.Afterselectingthedesiredfilesandclicking “Open,”awindowsimilartotheonebelowwillpopup.
Updated:16October2017 RoomCalorimeterCore 104
b.Inthe“Processing”tab(shownbelow),youcanselectwhichfileformatyouwanttoexporttoandalsosetadditionalprocessingoptions.Ifyouwanttoexportthedataintheformthatyoucurrentlyhave(e.g.,markerstracked,trialstrimmed,etc.),DONOTselectanyoptionsexceptforyourexportingfileformat.Otheroptionsmayundoworkthatyouhavealreadydone(e.g.,selecting“Trackthemeasurement”willre-trackallfilesinraw,unlabeledformat).
c.InthesecondaryBatchProcessingtab(above,thiswouldbethe“C3DExport”tabonthetopleft),youcanselectoptionsspecifictothatfileformat.
II.Toexportfilesone-at-a-time,clickFile>Export>[desiredfiletype].Using thisdialog,onlythesecondarywindowoptionsdescribedaboveinCwillpopup.
10.Afterexportingdata,processinginQTMiscomplete
Updated:16October2017 RoomCalorimeterCore 105
Chapter1.4-ExampleMarkerSets
TheHumanMotionCorehassomestandardmarkersetsthatmaybeusedformotioncaptureof
certaintasks.Thesemarkersetsincludefull-body,rightlowerextremity,leftlowerextremity,andheadwithFrankfortplane.Userscanfeelfreetocopythesemarkersetsandmodifyasneeded.Ingeneral,thefullbodyandlowerextremitymarkersetsareoptimizedforcalculationoflowerextremityjointmechanics.Amoredetailedmarkersetthanthefullbodymarkersetwouldbeneededtocalculateupperextremitymechanics,butthecurrentmarkersetcanbeusedtoaccountforsegmentsizes/massesforcenter-of-masscalculations.
Belowareimagesofthestandardmarkersetswithmarkerslabeled.Attheendofthisdocumentisadefinitionofthemarkerlabels.Notethatinthestandardmarkersets,markerlabelsareconsistentfrommarkersettomarkerset.
Updated:16October2017 RoomCalorimeterCore 106
UnilateralLowerExtremityThismarkersetexistsinarightsideandleftsideversion.Therightsideisshownbelow.Therearesomedifferencesinthemarkersonthelowerextremitybetweenthismodelandthefull-bodymodel.Thismodelhasafootmarkersetthatcanbeusedtocalculatewholefootandrearfootangles.Additionally,thismodelhasdifferentpelvicmarkerscomparedtothewholebodymodelwhichwillleadtodifferencepelvicsegmentdefinitionsbutnosubstantialdifferencesinsegmentmotioncalculations.
RASI LASI
LGRT RGRT
LPSI RPSI
RMCO RLCO
RMMA RLMA
R1MH R5MH
RTOE
RTHF RTHU
RTHL RTHR
RSHF RSHU
RSHL RSHR
RHLA
RHPR RTOE
RPSI RASI
RGRT
RHPR
RHLA RHMD
RMMA RLMA
R5MH
RTOE
Frontalviewofunilateralmarkerset
Closeupofrearfootofunilateralmarkerset
Posteriorviewofunilateralmarkerset
Updated:16October2017 RoomCalorimeterCore 107
RASI LASI LILC RILC
SLJ
RACJ LACJ
LLCO LMCO
RMCO RLCO
RMMA RLMA LLMA LMMA
L5MH L1MH LTOE RTOE
R1MH R5MH
FLH FRH
BRH BLH
TOPH
LELB RELB
RDUL
RDRA LDRA
LDUL LBTH
LHAN
LFIN RFIN
RHAN
RBTH
FullBody
Thismarkersetincludesthesamegenerallowerextremitymodelasthelowerextremitymarkersets,butaddsatrunk,arms,andahead.Notethatthismarkersetdoesnotincludethe3-markerrearfoot.Thiscanbeeasilyaddedasshowninthelower-extremityimage.
Frontalviewoffullbodymarkerset
Updated:16October2017 RoomCalorimeterCore 108
Viewofright/frontsideoffull-bodymarkerset
LTHF
LTHL
LTHU
LTHR
RTHF
RTHL
RTHU
RTHR
LSHF
LSHL
LSHU
LSHR RSHF
RSHL
RSHU
RSHR
Updated:16October2017 RoomCalorimeterCore 109
PosteriorViewoffull-bodymarkerset
RARF RARU
RARL RARR
RFAF
RFAL RFAU
RFAR LFAR LFAL LFAF
LARU
LFAU
LARF
LARR LARL
BLH
RACJ
SLJ
RHEEL LHEEL
LACJ
BRH
TOPH
Updated:16October2017 RoomCalorimeterCore 110
HeadMarkersetforGazeTracking
FLH FRH
BRH BLH
TOPH
LAUM
RZYA LZYA
RAUM LAUM RZYA
LZYA
RAUM
TOPH
BLH FLH
FRH
Updated:16October2017 RoomCalorimeterCore 111
MarkernamedefinitionsMarkerlabelsarelistedbysegment
Head:Crown/helmetTOPH–topofheadFRH–frontrightofheadBRH–backrightofheadFLH–frontleftofheadBLH–backleftofheadFrankfortplaneR/LAUM–right/leftauditorymeatusR/LZYA–right/leftzygomaticarchTrunk:R/LACJ–right/leftacromioclavicularjointUpperBackTrunkClusterLowerBackCluster(ifyouwantbothlumbar&thoracicspine)Arm:UpperarmrigidclusterR/LARU–right/leftarmupperR/LARF–right/leftarmfrontR/LARL–right/leftarmlowerF/LARR–right/leftarmrearForearmrigidclusterR/LFAU–right/leftforearmupperR/LFAF–right/leftforearmfrontR/LFAL–right/leftforearmlowerF/LFAR–right/leftforearmrearR/LDRA–right/leftdistalradius(radialstyloid)R/LDUL–right/leftdistalulna(ulnarstyloid)R/LHAN–right/lefthandR/LBTH–right/leftbaseofthumbR/LFIN–right/leftsecondfingerPelvis:R/LASI–right/leftanteriorsuperioriliacspine(ASIS)R/LILC–right/leftiliaccrestR/LPSI–right/leftposteriorsuperioriliacspine(PSIS)SLJ–sacrolumbarjointR/LGRT–right/leftgreatertrochanterLowerextremity:Thighrigidcluster
Updated:16October2017 RoomCalorimeterCore 112
R/LTHU–right/leftthighupperR/LTHF–right/leftthighfrontR/LTHL–right/leftthighlowerR/LTHR–right/leftthighrearShankrigidclusterR/LSHU–right/leftshankupperR/LSHF–right/leftshankfrontR/LSHL–right/leftshanklowerR/LSHR–right/leftshankrearR/LLCO–right/leftlateralfemoralepicondyleR/LMCO–right/leftmedialfemoralepicondyleR/LLMA–right/leftlateralmalleolusR/LMMA–right/leftmedialmalleolusR/L1MH–right/leftfirstmetatarsalphalangealjointR/L5MH–right/leftfifthmetatarsalphalangealjointR/LTOE–right/lefttoeR/LHEEL–right/leftheelR/LHPR–right/leftheelproximalR/LHMD–right/leftheelmedialR/LHLA–right/leftheellateral
CalibrationMarkersBothCalibration&TrackingMarkersTrackingMarkers
Updated:16October2017 RoomCalorimeterCore 113
CHAPTER1.5-Visual3DTutorial
Visual3D(V3D)isamotioncaptureanalysissoftwarepackagethatcanmakedataanalysisrelativelystraightforwardandstreamlined.C-Motion,thedevelopersofV3D,havecreatedahigh-qualitysetofdocumentationforthissoftware(https://www.c-motion.com/v3dwiki/index.php/Visual3D_Overview).Ifyouareafirst-timeuserofV3DitishighlyrecommendedthatyougothroughtheV3Dtutorials(https://www.c-motion.com/v3dwiki/index.php/Visual3D_Tutorials)tofullyappreciatethetoolsincludedinthisprogram,aswellastheimplicationsofincorrectlymodeling/computingyourresults.
I.V3DBasics
1.V3Dworkswithmotionfilesthatarein.c3dformat.MotioncapturedatacanbeexportedfromQTMinthisformatwhichwillincludeanymarker,analog,andforcedatacapturedinatrial.NOTE:V3Dcannotworkwithfiles/commands/tagsthathavespacesinthename.Neverusespacesinfilenames,markernames,etc.
2.ToloadrawdatafilesinV3D,clicktheopenbutton andselectthec3dfilesyouwishtoview.Whenfilesareloaded,theywillappearinthe“Workspace”tabinV3D.
Updated:16October2017 RoomCalorimeterCore 114
3.Toviewthesedatafiles,switchtothe“SignalsandEvents”tab.Hereyoucanselectwhichfiletoviewusingthedrop-downmenu,playorscrollthroughatrialusingthetoolsatthebottom,orviewrawTARGET(marker),ANALOG,orFORCEdata.
Drop-downboxtoviewtrials
Playbackcontrols
ORIGINAL(raw)targetdata
3Ddataview
Updated:16October2017 RoomCalorimeterCore 115
4.Ifyouleft-clickonanydataitem,aDataViewwindowwillpopuptoshowyoutrajectoriesandvaluesforthatdataitem.Youcanalsoright-clickonadataitemtoplotit.
5.Viewingrawdatahaslimitedutility.Generallyyouwillwanttobuildamodelandapplythismodelyourdatatrials.
II.ModelsinV3D
ModelsinV3Darethebasefilesfromwhichskeleton/segmentcharacteristicsareappliedtomotionfiles.Inordertocreateamodeltoapplytoyourmotionfiles,youmusthaveastaticcalibrationtrial.Forhumanmotion,especiallygait,thestatictrialisgenerallycapturedwhileanindividualstandsstillinanatomicposition.Theseinstructionswilltakeyouthroughsomegeneralstepsforcreatingamodelfromastatictrial.Thisisnottheonlywaytocreateamodel;furtherinstructionscanbefoundontheV3Dwiki.
TherearedemoV3Dfilesthatyoucanusetoexplorethefollowinginstructions.AcompletedanalysiscanbeviewedbyopeningR_LE_walk.cmo.Thefilesusedtobuildthisanalysisincludeonestatictrial(R_LE_static.c3d),twotreadmillwalkingtrials(R_LE_walk01.c3dandR_LE_walk02.c3d)amodelfile(R_LE_model.mdh),andapipeline(R_LE_walk_pipeline.v3s).ThesefilesarefordemonstrationonlyandcanbeusedtoexploresomecapabilitiesofV3DbutarenotinclusiveofalltasksthatV3Discapableof.
Therearesomepre-createdmodelsavailableforgeneraluse.Theseincludeunilaterallowerextremitymodelsforboththerightandleftlimbsandafull-bodymodel.Thesemodelsarebasedonthemarkersetsthatarealsoavailableforgeneraluseandcanbeusedandmodifiedasneeded.Thesemodelsarenotmeanttobeall-inclusivebutratherprovideexamplesandstartingpointsfordevelopmentofproject-specificmodels.NoteespeciallythattheupperextremitiesinthefullbodymodeldonotfollowISB
X,Y,Ztrajectory
Dataresidual
Dataidentity
Updated:16October2017 RoomCalorimeterCore 116
recommendationsfortheshoulder,elbow,orforearm.UsersshouldalwaysreviewtheV3Dwikirecommendationsformodelingofsegmentsandjointsabovethepelvis.
Generalfactsaboutmodels
ModelsinV3Dconsistofdefinitionsofsegmentsandlandmarksusedtocreatethesesegments.Thesedefinitionsareneededsothatthe3Dorientationofsegmentscanbetrackedduringamovement.Ingeneral,segmentsaredefinedusingrealorvirtually-definedmarkersattheproximalanddistalends,aswellasradiioftheproximalanddistalends.Atleastthreemarkersareneededtodefinetheorientationofasegment.Whenasegmentiscreated,itisassignedacoordinatesystem(seered,green,andblueorthogonalaxesinimagebelow)whichisrigidlyassociatedwiththesegment’sproximaljoint(e.g.,thethighcoordinatesystemwillbeanchoredatthehipjoint).Modelsmustalsoincludeassignmentoftrackingmarkers,ormarkersthatareregisteredtothesegmentdefinitioninacalibrationtrialandarethenusedtotrackthemovementofthesegmentduringmovementtrials.Thesegmentcoordinatesystemismodeledasrigidlyrelatedtotheorientationofthetrackingmarkersandforthisreasontrackingmarkersareoftenplacedonplasticshells(e.g.,thighandshankclusters)oronanatomicallocationswithlittleskinmotionartifact(e.g.,acromio-clavicularjointsduringlowerextremitytasks).Inadditiontosegment-specificdefinitions,modelsrequireanindividual’sheightandbodymassforcalculationofsegmentinertialpropertiesandjointkinetics.
Theorientationofadjacentsegmentcoordinatesystemswithrespecttoeachotherprovidesthejointangleorjointtranslationalposition.Forthisreason,itisimportanttopayattentiontotheorientationofsegmentcoordinatesystemsastheyarecreated.Intheimagebelow,youcanseethatallvisiblesegmentshavethered(X)axisorientedtothemodel’sleft.Thismatchingorientationprovideseaseofunderstandingforjointdefinitions.Forexample,rotationsaboutXbetweentheshankandthigharesagittalplanejointangles,thoseaboutYarefrontalplanejointangles,andthoseaboutZaretransverseplanejointangles.Inordertoobtainaccurateandmeaningfuldata,itiscriticalthatyoucorrectlydefineandconstructyourmodel.RecommendationsformodelingsegmentsandjointsofinterestcanbefoundontheV3Dwiki,aswellasinpublishedrecommendationsfromtheInternationalSocietyofBiomechanics.
Applyinganexistingmodeltomovementtrials
InanewV3Dworkspaceorwindow,onthemenu,click“Model,”then“Create(AddStaticCalibrationTrial),”and“HybridModelfromC3DFile.”
A.Inthewindowthatpopsup,selectyourstaticcalibrationtrial.
B.V3Dwillautomaticallyswitchtothe“Models”tab.
C.ClicktheLoadModelTemplatebutton
i.Selectthemodelfile(*.mdh)youwouldliketouse.
ii.Themodelshouldnowbeappliedtoyourstatictrial.
Updated:16October2017 RoomCalorimeterCore 117
iii.Insomecases,therewillbeerrorsinbuildingthemodel.Thesewillbedetailedina“BuildModelResults”window.
Updated:16October2017 RoomCalorimeterCore 118
iv.Ifthisoccurs,youcanscrollthroughtheerrorsand/orwarningstodetermine whatissuesarepresent.Fixingtheseissuesmayinvolvefindingacorrect, completestatictrial,correctingthenamesofmarkers,ormodifyingthemodelto correspondtodifferentsegment/landmarkdefinitions.
D.Definitionsofhowmodelsegmentsandlandmarksarecreatedcanbefoundbydouble-clickingonasegmentorlandmarkintheleft-handwindowintheModelstab.
Updated:16October2017 RoomCalorimeterCore 119
i.Forexample,ifyoulookattherightthighinthismodel,youwillseethefollowingdefinitions.
ii.Notethatsomeofthemarkersusedfordefiningtheproximalanddistaljointsarerealmarkersthatwereplacedonanindividual(e.g.,RLCO,RMCO)whileothermarkersare“landmarks,”orvirtualmarkerscreatedinV3Dbasedonthepositionofothermarkers.Here,theF_RHIPisalandmark.
Updated:16October2017 RoomCalorimeterCore 120
iii.IfwelookatF_RHIPinthe“Landmarks”tab,wecanseehowthislandmarkwascreated.Thislandmarkisahipjointcenterdefinitionbasedonliteratureregressionequations.
E.Afterapplyingyourmodel,theworkspaceshouldlooksomethinglikethis:
Updated:16October2017 RoomCalorimeterCore 121
F.Nowyoumustaddmotionfiles.Todothis,clicktheopenbutton andselect yourmotiontrials.Awarningwindowwillpopup,select“Insertnewfilesintoyour currentlyopenworkspace.”
i.Afterselectingyourmotiontrials,yourworkspacewilllooksomethinglikethis:
G.Youcannowapplyyourmodeltoyourmotionfiles.Todothis,clickModel>Assign ModeltoMotionFiles.Selectthetrialswhichyouwouldliketoapplythemodelto
Updated:16October2017 RoomCalorimeterCore 122
i.Nowyourworkspaceshouldshowyourmotionfileswithintheirassigned model.
ii.Ifyouviewyourmotiontrialsinthe“SignalsandEvents”tab,youwillseethe modelappliedtothetrials(assumingyourdataweretrackedcorrectlywith markernamesthemodelexpected.)Notethatthemodelisnotnecessarilytoscale orcompletelycorrectanatomically(e.g.,modelfemoralheadsareoftennotinthe acetabulum),themodelismeantasavisualaid.
iii.Ifasegmentismissingorincorrectlyorientedinamotiontrialitisdueto incorrectlylabeleddata.Checkthatalltrackingmarkersarepresentinthemotion trialandthatthemarkernamesassignedtothemotiontrialmatchthoseinthe model.
Updated:16October2017 RoomCalorimeterCore 123
III.DataAnalysisandPipelines
OnceyouhavedataloadedintoV3Dandamodelassignedtoyourmotionfiles,youcanprocessyoursignals,createmotionevents,andcomputeoutcomes(e.g.,jointkinematicsandkinetics).Thiscanbedoneusingsomepoint-and-clickstepsbutismoreefficientlyimplementedusingpipelines.Pipelinesconsistofaseriesofcommandswhichcanbesavedandusedformultipleparticipants/conditions/iterationsinastudyandthusassistwithbatchprocessingandcalculationconsistency.Thefollowinginstructionswillgothroughsomebasicpipelinecommandsthatmaybeusedduringanyanalysis.Asyouworkthroughpipelinecommandsitwillbecomeclearthattherearemanytoolsavailableinpipelines.FurtherinformationabouteachcommandcanbefoundintheV3Dwikidocumentation.Foranexamplepipeline,intheCPHMdemonstrationdataprovided,theanalysisoftheR_LE_walk.cmofilewasperformedusingtheR_LE_walk_pipeline.v3sfile.
1.Tousethepipelinetool,clickonthepipelinebutton. Thefollowingwindowwillpopup.
2.Inthelefthandpaneofthiswindowareallavailablecommands,organizedintosubgroups.Thecenterpaneshowscommandsinthecurrentpipeline,andtherighthandpaneshowsapreviewofacommandinthecurrentpipelinewhenitisselected.Forexample,ifyouweretostartcreatingapipelinewithacommandtocreateamodelfromastatictrial,thewindowwouldlooklikethis:
Updated:16October2017 RoomCalorimeterCore 124
3.Whenyouhaveacommandinacurrentpipelinehighlighted,theinitialtextofthatcommandaswellasadescriptionofthecommandwillappearintherighthandpane.Toeditthecommand,double-clickonthecommandinthemiddlepane.Forthe“Create_Hybrid_Model”command,double-clickingonthecommandwillbringupawindowoftextthatcanbeedited.Inthesetypesofwindows,“!”indicatesthatthepipelineshouldskipthislineofcodeorusethedefaultvalue.Inthewindowbelow,thismeansthatthereisnospecificfilethatthepipelineispulling,itwillinsteadopenadialogwindowaskingtheuserto“Openstandingfile.”Ifthereisaspecificfileyou’dliketoloadusingthiscommand,thepathtothisfilecanbeenteredinthe“/CALIBRATION_FILE=”lineandthe“!”canberemovedfromthisline.
Textofcommand
Descriptionofcommand
Updated:16October2017 RoomCalorimeterCore 125
4.Insomecases,therearetwodifferentformatsinwhichtoeditacommand.Forexamplewhenyoudouble-clickonthe“Event_Copy”command,itopensaGUI-typewindow.Ifyouprefertoeditthecommandintext
format,youcanclickonthecommandinthemiddlepaneofthepipelinewindowandclickthe button.
Updated:16October2017 RoomCalorimeterCore 126
5.Whenapipelineiscreatedandsaved,the.v3sfilecanbeopenedasatextdocumentforeaseofbatchediting(e.g.,changingexportfilenamesonasubject-by-subjectbasis).
6.Pipelinewindowbuttons:
A.“OpenPipeline…”allowsyoutoopenapreviously-createdpipeline.Ifyouclickthis buttonwhencommandsarealreadyinthemiddlepaneofthepipelinewindow,youwill begiventheoptiontoreplacethecurrentcommandsorappendthepipelineyouare openingtotheendofthecurrentcommands.
B.“SavePipelineAs…”allowsyoutosaveyourpipeline.Pipelinesdonotautomatically save.Ifyouclearthepipelineorreplacethepipelinebyopeninganotherpipeline,the pipelinewillnotbesavedandyouwillnotbepromptedtosaveit.
C.“ClearPipeline”clearsallcommandsfromthemiddlepaneofthepipelinewindow.
Updated:16October2017 RoomCalorimeterCore 127
D.“ClearParameters”willclearanyparametersenteredinpipelinecommands.
E.“Step”willexecutethecommandcurrentlyhighlightedinthemiddlepane.
F.“ExecutePipeline”willexecutetheentiretyofthecommandsinthemiddlepaneofthe pipelinewindowfromthefirstcommandtothelast.
G.Notethatthereisno“undo”buttonanywhereinV3D.
7.Keypipelinecommands.Userscanfollowalonginthe“R_LE_walk_pipeline”forexamplesoftheimplementationofeachofthesecommands.
A.Create_Hybrid_Model+Apply_Model_Template:whilemodelscanbecreatedandappliedusingthepointandclickstepsdetailedabove,itisoftenusefultoautomatethesestepsonceamodelhasbeencreatedandpiloted.The“Create_Hybrid_Model”commandwillpromptausertoopenastandingcalibrationtrialandloadthistrialintotheV3Dworkspace.The“Apply_Model_Template”commandwillprompttheusertoselectamodel(.mdh)fileandthenapplythismodeltothestandingcalibrationtrial.
B.File_Open:thiscommandpromptstheusertoopenfileswhichwillbeloadedintothe V3Dworkspace.Inthe“R_LE_walk_pipeline,”thiscommandisusedtoloadmovement trialsafterthecreationofthemodel.
C.Assign_Model_File:thiscommandwillprompttheusertoassignthemodelfileto movementtrials.
D.Lowpass_Filter:target(marker)andanalogdataoftenneedtobefilteredtoremove noise.TherearemanyfilteroptionsinV3Dpipelines(seethe“SignalFilter”sectioninthelefthandpaneofthepipelinewindow).Ingeneral,rawdatashouldbefilteredbeforeperformingcalculations.Whichfiltersshouldbeusedandatwhatstepoftheanalysistheyshouldbeimplementeddependsonthetaskbeingperformedandtheanalysisrequired.Inthe“R_LE_walk_pipeline”example,alowpassButterworthfilterisappliedtothemarkerdatatoremovesignalabove8Hz.The“NUM_REFLECTED”lineinthecommandprovides100reflectedpointsoneithersideofthedatatominimizetimeshiftsinthedataduetothefilter.Notethatwhenthetargetdataarefiltered,thefiltereddataareplacedinanewfolder:TARGET>PROCESSED,whiletheoriginal,unfiltereddataremainintheTARGET>ORIGINALfolder.
E.Interpolate:thiscommandwillfilldatagapsofaspecifiedsize.Thisshouldnotbe usedinlieuoftrackingdatainQTMbutratherasafinalsweeptocorrectanysmallgaps thatweremissedintracking.
F.Select_Active_File:thiscommandcanbeusedtoselectallfilesorgroupsoffilesfor furtherpipelinecommands.Whenpairedwithanearlier“Assign_Tags_To_Files,”the Select_Active_Filecommandcanbeusedtoperformpipelinecommandsononlyasubsetoffiles.
G.Eventcommands(e.g.,Event_Minimum,Event_Maximum,Event_Threshold):canbeusedtoautomaticallycreateeventssuchasheel-strikes,toe-offs,maximumflexion,etc. Thesecommandscanbefoundinthe“EventCreation”and“EventManagement” sectionsofthelefthandpaneofthepipelinewindow.Notethatitisoftenmostusefulto editthesecommandsusingthe“Text”optionratherthanthedefaultGUIwindow.
Updated:16October2017 RoomCalorimeterCore 128
H.Forloops:the“For_Each”and“End_For_Each”commandscanbeusedtobeginand endforloops.Inthe“R_LE_walk_pipeline”example,theseareusedtorecursively renamegaitcycleevents.
I.Signalmathcommands:thesecommandscanbeusedtoperformsimplearithmeticonsignals.Forexample,inthe“R_LE_walk_pipeline,”thenumberofframesbetweenconsecutiveheelstrikesandbetweenheelstrikesandtoe-offsareusedtodeterminegaitcycletimes(usingSubtract_Signals)and,subsequently,thepercentofthegaitcyclespentinthestancephase(usingDivid_Signals).
J.Metriccommands:thesecommandscreatenewsignalsinthe“METRIC”folder.
K.Compute_Model_Based_Data:thiscommandisperhapsthemostimportantcommandinV3Dpipelines.Withit,ausercancomputeahugerangeofvariables.Wheneditingthiscommandthefollowingwindowappears.Init,younamethedatavariableyouarecreating,identifywhattypeofvariableyouarecalculating,anddefinethedatabeing usedtocalculatethisnewvariable.TherearemanydifferenttypesofvariablesthatcanbecreatedandthespecificsofthealgorithmsusedorcomputationalordercanbefoundintheV3Donlinedocumentation.The“UseNegative”optionsallowyoutomultiplyasignalby-1asyoucalculateit,whichiscommonlyusedtoaligndatadirectionswithcommonreportingpractices(e.g.,theright-handrulewillgenerallyresultinkneeextensionbeingapositivevalue,whilereportingpracticesgenerallyfollowthatkneeflexionshouldbepositive).CardanSequenceshouldgenerallyfollowfromaxisoflargestexpectedexcursiontoaxisofsmallestexpectedexcursionforthemovementofinterest.Inthebelowexamplewindow,fortherighthipduringwalkingwewouldexpectthelargestexcursiontobeinthesagittal,orX,axis,followedbyfrontal(Yaxis)andtransverse(Z)motion.
Datavariableyou’recreating(musttypeinname)
Whattypeofvariableyou’recreating
Definitionsforvariablecalculation
Updated:16October2017 RoomCalorimeterCore 129
L.Export_Data_To_Ascii_File:thiscommandisusedtoexportdatatoatab-delimitedtextfilewhichcanbereadinexcelorimportedintoafurtherdataanalysisprogram,suchasMatLab.IntheGUIwindowwhichpopsupwhenyoueditthiscommand,youcandesignatethepathforyourexportedAsciifile,selectoptionssuchasexportingthemeanandstandarddeviationofyourdata(ifyouhavemultipletrials),normalizeyourdata(tomakealltrialsthesamerelativelength),amongotheroptions.Whenyouselectthesignaltype,folder,name,andcomponentyou’dliketoexport,youmustclick“AddaNewSignal”tosavetheseselectionsintothecommand.Likewise,ifyouwishtoeditanexistingcommand,youmustclickthe“ModifySelected”buttonaftermakingyoureditstosavethem.Intheexamplefrom“R_LE_walk_pipeline,”notethateachstrideofdatamusthaveaseparatelydesignatedeventsequenceandthusaseparatelineintheGUI.Ifyourunthiscommand(withanappropriatelyrenamedfilenamepath),youcanopentheexportedfileinexceltoviewtheformatofV3DAsciifiles.If“ExportMeanandStandardDeviation”isselected,themeanandstandarddeviationofalllinesintheGUIwillbecomputed,andthesewillappearascolumns2-7intheAsciiexport.Notethatifyoure-runthe“Export_Data_To_Ascii_File”commandwithoutchangingthe“ExportedAsciiFile”name,youwilloverwriteanyexistingfileatthatfileaddress.V3Dwillnotgivea“areyousureyouwishtooverwrite…”warningdialogbeforeoverwritingexistingdata.
Updated:16October2017 RoomCalorimeterCore 130
IV.Reports
V3Dcancreatereportstoassistwithdatavisualization.Thesereportscanplotvariablesthathavebeencalculatedandcangenerallybeusefulforcheckingcalculations.Areportisincludedinthe“Reports”tabofthe“R_LE_walk.cmo”fileandthecorrespondingreporttemplatefileisincludedindemonstrationfilesas“R_LE_report_template.rgt.”Thisreportconsistsentirelyof2Dgraphsofjointkinematics.Reportsarealsoabletogeneratetables,bargraphs,andothergraphics.Belowaresomeguidelinesforcreatingareport.
1.Whencreatinganewreportfromscratch,firstdesignatethelocationoftheitem(graph/graphic)youwishtoadd,selectthetypeofitemyouwishtoadd,andclickthe“Add”button.
Locationofitemtoadd
Itemtoadd
Updated:16October2017 RoomCalorimeterCore 131
2.An“Add/ModifyGraph”windowwillpopup.Hereyoucandesignatewhatdatatographinwhichposition.Belowisanexampleofthiswindowpopulatedforplottingonestrideofsagittalhipangledata.
3.Inareportwithmultiplemovementcyclesandmultiplesvariablesofinterest,therewillbemanylinesinthereportgenerationdialog.Alongwithmultiplecolumnsandrowsofplotsonasinglepage,reportscancontainmultiplepages,allowingforeasyvisualizationoflotsofdata.Ifyouareinterestedingeneratingreports,itisrecommendedthatyouexplorethedocumentationonreportsintheV3Dwiki(https://www.c-motion.com/v3dwiki/index.php/Tutorial:_Creating_a_Report)forfurtherinformationandexamples.
Updated:16October2017 RoomCalorimeterCore 132
UNIVERSITYOFMASSACHUSETTSAMHERST
INSTITUTEFORAPPLIEDLIFESCIENCE
HumanTestingCenter
HumanMotionCore
Chapter2:DelsysTrignoEMG
1. GeneralInformationaboutTrignoSensors1.1. Sensors1-8haveEMG+TriaxialAccelerometersensors.
1.1.1. WhenusedinconjunctionwithQTMthesensorshavecollectionfrequenciesofEMG1925.25Hz(upsampledto2000Hz)and148.48Hz(upsampledto??).
1.2. Sensors9-16haveEMG+TriaxialAccelerometer,GyroscopesandMagnetometers.1.2.1. WhenusedinconjunctionwithQTMthesensorshavecollectionfrequenciesofEMG1925.25Hz
(upsampledto2000Hz)and148.48Hz(upsampledto??).
2. ConfiguringtheTrignoSensors
2.1. Openthe‘DelsysTrignoContorlUtility’2.2. Inthelowerrightcornerofthescreenpress‘Configure’
Addtherestoftheconfigurationscreens–whattheyarehowtheywork.
Updated:16October2017 RoomCalorimeterCore 133
3. PairingSensors.
3.1. Pairthesensorsyouwouldliketouse3.1.1. Pushthepairbuttonin“DelsysTrignoControlUtility’software.
3.1.2. Pushthebuttononthesensoruntilyouseeaflashinggreenlight.Youshouldalsoreceivea
confirmationonthecomputerscreenthatthesensorhassuccessfullypaired
4. Placementandorientationofsensors5. EnablingSensorsinQTM
5.1. InputDevice
6. LabelingSensorsinQTM
Updated:16October2017 RoomCalorimeterCore 134
7. AccessingSensordatainQTM
7.1. Sensordatacanbeviewedinrealtime
Updated:16October2017 LivingScienceCore 1
MANUALOFPROCEDURES
LivingScienceCore
InstituteofAppliedLifeSciences
LifeScienceLaboratoriesS360240ThatcherRoad,
Amherst,MA01003
atthe
UniversityofMassachusetts,Amherst
Updated:16October2017 RoomCalorimeterCore 2
UNIVERSITYOFMASSACHUSETTSAMHERST
INSTITUTEFORAPPLIEDLIFESCIENCE
HumanTestingCenter
LivingScienceCore
Chapter1:OxyconMobile
1.1 Introduction:TheOxyconmobileisaportableindirectcalorimetryunitthatmeasuresmetabolicresponsestoexercise,workorotheractivitiesthatstationarymetabolicunitscannotmeasure.Thewirelesscapabilitiesofthisunitoperateviatelemetry,upto1,000minlineofsightrange.TheOxyconMobileaffordsuserstheopportunityforon-the-spotrecordingofcardiopulmonaryanddeducedparameterssuchas:VO2,VCO2,RER,VEO2,VECO2,PETO2.VE,VT,BF,BR,andSpO2,etc.Researcherscanchoosetomonitorparametersineither,breath-by-breathorintrabreathepochs.Additionally,researcherscanmonitorthesedatainreal-time(viatelemetry),orposthoc,asdataarestoredtoaninternalSDmemorycard.
CoreCompetencies:
1. Beabletofollowcalibrationinstructions2. Understandhowtooutfitparticipantwithequipotent3. Reviewstepsforsettingup,collecting,andreviewingsessiondatathroughascreenreport4. Familiarizeyourselfwithhowtogeneratea.CSVreport
1.2 HowtoCalibrate:Belowisastep-by-stepoutlineforcalibratingtheunit.
PreliminaryCalibrationSteps1.) RetrieveComputerfromOxyconcase
a. Username:labsup Password:lab2.) Collectreceiverunit(picturedbelow)
Updated:16October2017 RoomCalorimeterCore 3
3.) PluginpowerandconnectcomputertoreceiverunitusinggreyUSBcable
Updated:16October2017 RoomCalorimeterCore 4
4.) Grabthetwopacks(SBX,DEX)andamemorycarda. SBX:Sensorbox(O2&CO2sensor)b. DEX:Dataexchangeunit
Powercord
Updated:16October2017 RoomCalorimeterCore 5
5.) WarmupSBXbyconnectingittothereceiverwithyellowEthernetcable
6.) Openlabmanager
a. Allowwarm-upfor15minutes
Updated:16October2017 RoomCalorimeterCore 6
7.) Assemblenecessarycomponentsofmaska. Attachturbinetoturbineholder(metal(silver)sideinfirst)
b. Makesureturbineisnotwetc. Connectinfraredsensortoturbine
d. Besuresamplingtubeisattachedtotheturbineholder(oppositeendofinfrared)
Updated:16October2017 RoomCalorimeterCore 7
8.) AttachmaskcomponentsandtightensamplingtubeandinfraredsensortoSBX
Updated:16October2017 RoomCalorimeterCore 9
9.) PlaceappropriatestudymemorycardinDEXunitandaddbattery(makesureitischarged)
a. IfDEXdoesn’tturnoffrightaway,pressoffbutton
10.) After15minutes,labmanagershouldopenuptoahomescreen
Updated:16October2017 RoomCalorimeterCore 10
11.) ConnectSBXtoDEXusingyellowEthernetcable
a. AfterafewsecondsofSBXvibrating,alightshouldflashonSBXb. Ifthebatteryislow,SBXwillbeepandanotewillpopuponthescreenc. BesuregreenLED’sonbothunitsarelitpriortocalibration
1.3OxyconMobileCalibration
1. ClickAmbientconditionsonthelabmanager
Updated:16October2017 RoomCalorimeterCore 11
a. DEXsensesi. Usuallyaccuratetemperatureandbarometricpressureii. Manuallyenterhumidity
b. CheckgreyboxinthelaboratoryontheParvometaboliccartforenvironmentalconditions
i. PressF1tochangetemperature,barometricpressureandhumidityii. Leavebarometricpressureasis
c. PressF12tosave
2. ConnectSBXbacktoreceiverusingyellowEthernetcablea. WaitforSBXtogetpower(willvibrateandLEDlightsupgreenwhenreadytocontinue)
3. SetupforAutoVolumeCalibration
a. ClickAutovolumecalibrationsfromcalibrationsgroup
Updated:16October2017 RoomCalorimeterCore 12
b. Waitforgraphtocomeup
4. Connectwhitesamplingtube,intoreceiverandSBX
Updated:16October2017 RoomCalorimeterCore 13
5. ConnectmouthpiecewithturbineintoPCA(blackhole)withsampletubeholecovered
6. PressF1tobeginAutoVolumeCalibration
Updated:16October2017 RoomCalorimeterCore 14
a. If%differenceislessthan3%pressF12tosaveb. If%differenceisgreaterthan3%,checklinesandconnectionsandretry
7. GasAnalyzerCalibration
a. Connectgascylinderintounit,pushclipuntilitislocked
Updated:16October2017 RoomCalorimeterCore 15
b. Opengascylinderbytwistingthevalvefromthecalibrationgastwocompleteturns
i. Cylindercontains15.98%O2and3.9%CO2concentrations
c. ClickonGasAnalyzerCalibrationinthelabmanager
Updated:16October2017 RoomCalorimeterCore 16
d. Clickonthesettingstabatthetopandenterinanalysisresultsfromthe“CertificateofAnalysis”providedbyAirgas,thissheetshouldbewiththegastanks
e. Then,pressF1tobegin,Graphwillappearandcalibrationwillbeginf. Calibrationwillsampleroomairalongwithcylinderair
g. O2peaksat21andvalley;bluelineh. CO2willvalleyat0andthenpeak;redlinei. Ifpercentdifferencelessthan3%pressF12andsavej. Turnofaircylinderk. Ifpercenterrorgreaterthan3%,repeatprocess
8. ReassembleSBXandDEXwithyellowEthernet
Updated:16October2017 RoomCalorimeterCore 17
9. Reassemblemouthpiecewithsamplingtubea. PlacenewbatteryinDEX
Updated:16October2017 RoomCalorimeterCore 18
1.4TestingwiththeOxyconMobile
1. Inputpatientdataa. PressF2tocreatenewsubjectb. BesuretofilloutLastName,FirstName,Identification,DOB,Gender,HeightandWeight
(pressentertonavigatethroughpatientinputs)
c. PressF12tosavecurrentsubjectinformation
2. Sizesubjectforheartratebelta. Wetsensorsb. Attachsensorstobeltc. Subjectputsonheartratemonitorensuringthesensorisatthesternum
3. Sizesubjectformask
a. Havesubjectputmaskonandcoverbreathingholeb. Havethesubjectblowtomakesuretherearenoleaksc. Attachthefourclipstothemaskd. Havethesubjectholdthemaskinplacewhiletheresearcherclipstheblackheadgear
nettothemask(taggoesinfront)
4. Sizesubjectforbackpacka. AttachtheSBXandDEXtothebackpack(wiresfacingup)b. Subjectputsonbackpackc. Attachmaskwithwhitesamplingtube(whitesamplingtubeshouldbeclosesttothe
subjectsface)
Updated:16October2017 RoomCalorimeterCore 19
5. Tobegindatacollection,clickbreathbybreath
a. MakesureDEXandSBXhavegreenlights
i. Ifyellow,flipSBXandDEX,checkHRbeltposition,checkbattery
6. EnsureallPatientDataiscorrect
7. CheckTestsettings
Updated:16October2017 RoomCalorimeterCore 20
a. Ergometer-Modalitymustbebicycleb. DeadSpace-clickappropriatesizecorrespondingtosubject’smasksizec. Averaging-Method,Breaths
8. PressF1orplaytoanalyzebaselineinformation(backgroundzeroing,notedinbottomrightcorner)
9. PressF1orplayagaintostartrecordingdata10. Afterdatacollectioniscompletebesuretheupperleftbutton(play)isnowa“stop”buttonto
endrecording11. Press12toexitandsave
1.5ReviewingtheData
1. ClickonScreenReporttoviewcollecteddata
2. Fromthereportlist,choosethereportviewyouwouldlike(studyspecific,formoreinformationon
creatingnewdatavisualizationtemplatesseebelow)
Updated:16October2017 RoomCalorimeterCore 21
3. PressOKtogeneratethereport
4. ToexportusabledatapressF9andsaveonalocaldrive
5. ToimportfromstoragedeviceselectFlashdiskreader
Updated:16October2017 RoomCalorimeterCore 22
6. Importthefileyouwouldliketoreview7. MakesuretheCurrentpatientmatchesthefileyouaretryingtoimport8. PressF1toreadfile9. AfterfileisloadedpressF12tosave10. RepeatSteps1-411. ToviewsavedfileinExcel,openExcelandgotoFile-Open12. Findthesavedfile13. WhenpromptedbytheImportWizard,MakesureImportRowisat1andFixedWidthisselected
14. ClickNext
Updated:16October2017 RoomCalorimeterCore 23
15. Whenpromptedwithtextimport,alsoclicknext
16. Onthefinalstep,Clickadvancedandmakesureseparatorsarecorrect
17. Afterimportiscomplete,selectSaveasandbesuretocheckthecorrectfileextension
Updated:16October2017 RoomCalorimeterCore 24
1.6CleaningEquipment1. Disassembleallpiecestobecleaned2. Wipeturbineholderandinfraredsensorwithalcoholwipes.3. Headgearstrapsandheartratemonitorsshouldberinsedandhunguptodry4. Masksshouldbesoakedina25:1cleaningsolutionfor30secondsandrinsedthoroughly5. Theturbinemustbesoakedinaspecialsolution
a. Inthemarkedcontainer,fillthetothebrimwithSterilizingandDisinfectingsolutionb. Intheclearplasticcuplocatednearby,fillthecuptothefirstindent(blackline)withthe
Activatorsolutionc. PourtheActivatorsolutioninthesterilizingsolutionandmixthoroughlyd. Placetheplasticgrateintothesolutionwiththeturbineinsidee. Allowtheturbinetositfor15minutesf. Removefromsolutionandrinsewithwaterg. Letairdry
1.7CreatingaNewDataVisualizationTemplate1. Clickontheutilitiestab2. Enterpassword:abcd3. Clickonreportdesigner4. Namenewreport,outputunit:defaultprinter(JLABpdfprinter)5. Clickok6. Clickonthetableattheright(hasF3onbutton)
Comment[MB3]:Isthisactuallythepassword?
Updated:16October2017 RoomCalorimeterCore 25
7. ChooseOxycontable,thenclicksummary8. Hereyoucanchoosemeasurementprogram(breath-by-breath,endtidal,cardiacoutput,intrabreath,
etc.)9. Dependingonthemeasurementprogramselectedyoucanaddparameterstoyoursummary,dothis
byhighlightingaparameteryouareinterestedinandclick“Insert”.Thiswilladdittoyourscreenreportsummary
10. Byfurthernavigatingthroughthetabsatthetopofthedialoguebox,youcanselectaveragingtype(intervals),layout,marker,andsetupsettings
11. PressF12tosavethereport,andnamethereport.Note:thenameyougivethereportwillbehowit
ispresentedinthescreenreportsection(seeReviewingtheDatatore-familiarizeyourselfwiththisstep)
Updated:16October2017 RoomCalorimeterCore 26
1.8Contacts
1. KimatCarefusion:18002312466Operatorwillpromptyouthroughascreener,click1atfirstquestion,click2atsecondquestion,click3atthirdquestion.
Note:Havereceiverunitonhandwhenyoucall,ithasserialnumbertheywillneedwhenassistingyou.
2. MichaelBusa–CoreDirector413-577-0574
Updated:16October2017 RoomCalorimeterCore 27
UNIVERSITYOFMASSACHUSETTSAMHERST
INSTITUTEFORAPPLIEDLIFESCIENCE
HumanTestingCenter
LivingScienceCore
Chapter2:ActigraphLink(GT9X)
2.1Supplies
1) ActiGraphGTXaccelerometerchargingcable2) ActiGraphdock3) ActiLife6software4) ComputertorunActiLife6software
2.2DeviceCharging
TheActiGraphGT9XaccelerometercontainsarechargeableLithiumPolymerbatterywhichiscapableofprovidingpowerforcontinuousdatacollectionperiodsof10ormoredayswithoutrecharging.Ifthebatteryhasbeenfullydepletedatthetimeofcharging,itwilltakeapproximately3-4hoursfortheActiGraphGT9Xaccelerometertoreachfullcharge.ChargingProcedures
1) ConnectActiGraphdocktocomputerorwalloutletusingminiUSBcable2) PlugtheActiGraphLinkintothedockwiththeActiGraphlogofacingup.Once
connected,theredLEDlightontherightsideofthedockwillturnyellow,thedevicescreenwilldisplaytheserialnumber,andthebattericonwillblinktoindicatecharging.
3) Oncethedeviceisfullycharged,theyellowlightwillturngreenandthebatteryicononthedevicewillshowasfullandstayonsteady
2.3DeviceInitializationTheActiGraphGT9XaccelerometersareinitializedwiththeActiLife6softwarepackage.TheActiGraphGT9XaccelerometerwillnotcollectdataunlessitisinitializedusingtheActiLifesoftware.
Updated:16October2017 RoomCalorimeterCore 28
InitializationProcedures
1) OpentheActiLife6softwarebydouble-clickingonthedesktopiconorbysingle-clickingonthestartmenushortcutfortheprogram.
2) PlugthechargingcableintoastandardUSBportonthecomputerrunningtheActiLife6
software.3) Selecttheaccelerometerthatwillbeinitializedbycheckingtheboxbesidethe
accelerometerdevicenameandserialnumber.
4) Fromthe“Initialize”buttonchoose“RegularInitialization”.Thiswillopentheprimary
initializationwindow.
1)
3)
Updated:16October2017 RoomCalorimeterCore 29
5) Selectasamplingrate6) Selectidlesleepmode.7) Note:SelectWirelessOption,heartrate,sleepmode,andIMU8) Clickthe“EnterSubjectInfo”button.Thiswillopenthesecondaryinitializationwindow.
6)
7)
8)
4)
5)
Updated:16October2017 RoomCalorimeterCore 30
9) Selectastartdateandtimeatthetimesubjectorientationisscheduledtobegin10) Selectastopdateandtimetostopdatacollectionataspecifiedtime(canbeleftblank)
11) Inputanappropriateidentifier(e.g.,“S#”)inthe“SubjectName”field.12) Inputtheappropriate“Limb”and“Side”information(e.g.righthiporleftwrist)13) Clickthe“Initialize1Device”button.Thiswillclosethewindowandreturntothemain
softwarewindow.14) Ifproperlyinitialized,thewords“finishedinitializing”shouldappearunderneaththe
“Status”column.
15) TheActiGraphGT9Xaccelerometerisnowreadyfordatacollection.Unplugaccelerometerfromthechargingcable.
9)
11)
12)
13)
Updated:16October2017 RoomCalorimeterCore 31
2.4AccelerometerPlacement
TheActiGraphGT9Xcanbewornatthehip(aroundthewaist)andonthewrist.Elasticbeltsandwriststrapswillbeusedtosecuretheaccelerometerstotheparticipant’sbody(seepicturesfordetails).
Securedaroundthewaistattherighthipandattachedtoanelasticbelt.MakesurethattheGT9Xispositioneddirectlyonthebody’sside(mid-axillaryline)
2.5DataRetrievalandDownload
ThepreviouslydescribedActiLife6softwareutilityisrequiredtodownloaddatafromtheActiGraphGT9Xaccelerometer.
Updated:16October2017 RoomCalorimeterCore 32
DownloadProcedures1) OpentheActiLife6softwarebydouble-clickingonthedesktopiconorbysingle-clickingonthestart
menushortcutfortheprogram.
2) PlugtheActiGraphLinkintothedockwiththeActiGraphgoldplatescontactingthegoldpinsofthedock.3) PlugthechargingcableintoastandardUSBportonthecomputerrunningtheActiLife6software.4) Selecttheaccelerometerthatwillbeinitializedbycheckingtheboxbesidetheaccelerometerdevice
nameandserialnumber.
5) Clickonthe“Download”button.Thiswillopentheprimarydownloadwindow.Click“ChangeLocation”toselectdestinationfile
6) Withinthe“GT9XDownloadOptions”box,makesurethe“CreateAGDFile”boxischeckedandspecifythe“Epoch”.
7) Checktheappropriateboxesforyourprotocol.
1)
4)
5)
Updated:16October2017 RoomCalorimeterCore 33
8) Clickthe“DownloadAllDevices”button.Theprimarydownloadwindowwillcloseandreturntothemain
softwarewindowanddialogboxwillappearaskingyoutonamethefile.Thedownloadprocessmaytakeseveralminutes.
5)
5))
8)
7)
6)
9)
Updated:16October2017 RoomCalorimeterCore 34
9) Ifproperlydatawasproperlydownloaded,thewords“finisheddownloading”shouldappearunderneath
the“Status”column.
10) Nowitisnecessarytoprocessthedefaultfilter.agdfile.Select“Scoring”>Checkallitem>click“RemoveSelected”>Click“AddDataset”>SelectthepreviouslydownloadedAGDFile
11) Click“Calculate”12) Goto“Details”onthecalculatedfileandclick“ExportAllEpochs”13) Goto“OpenContainingFolder”,renamethefile(S#_ActiGraph_Location.csv)andputitinthe
appropriatefolder
10)
Updated:16October2017 RoomCalorimeterCore 1
MANUALOFPROCEDURES
RoomCalorimeterCore
InstituteofAppliedLifeSciences
LifeScienceLaboratoriesS360240ThatcherRoad,
Amherst,MA01003
atthe
UniversityofMassachusetts,Amherst
Updated:16October2017 RoomCalorimeterCore 2
UNIVERSITYOFMASSACHUSETTSAMHERST
INSTITUTEFORAPPLIEDLIFESCIENCE
HumanTestingCenter
RoomCalorimeterCore
Chapter1:QuickGuideforCalibration/Validation
Supplies
QuickGuide:CalorimeterRoomOperation
GasTankCheckListPriortocalibrationtestsandvalidationstudies,checktankpressures.
Tank MinimumPressure Tests
CO2 100psi Infusion,ShortCircuit,BlenderCalibration
O2 800psiforblendercal,NAforinfusion
Blendercalibration
StartCalRQ Login
SelectMode
HumanorValidationStudy Calibration
CalCheck
Complet
NullTest
No
Yes
RunStudy NullTest
Figure 1. Study Setup Flow
Updated:16October2017 RoomCalorimeterCore 3
N2 1500forblendercal,800psiforInfusion,500forShortCircuit
Infusion,ShortCircuit,BlenderCalibration
ZeroTank 400psi TankCalibration,CalCheck
SpanTank 400psi TankCalibration,CalCheck
RefTank 400psi Alltests–alwaysopen
Panelinstrumentsettingsforalltests StdSetting
1 Chamber1O2Rotameter 1.25L/min
2 Chamber1CO2Rotameter 1.25L/min
3 Chamber2O2Rotameter 1.25L/min
4 Chamber2CO2Rotameter 1.25L/min
5 InflowO2Rotameter 1.25L/min
6 InflowCO2Rotameter 1.25L/min
7 Chamber1SamplePressureGauge 2psi
8 Chamber1SamplePressureRegulator 2psi
9 Chamber2SamplePressureGauge 2psi
10 Chamber2SamplePressureRegulator 2psi
11 InflowSamplePressureGauge 2psi
12 InflowSamplePressureRegulator 2psi
Updated:16October2017 RoomCalorimeterCore 5
AnalyzerCalibration
CalCheck:flowzeroandspangasestocheckvalues
AnalyzerGain:Subtracttheapparentconcentrationonthecomputerofthespanvaluefromtheapparentconcentrationonthecomputerofthezerovalue.Thisdeltashouldmatchthedeltaofthespanfromtheblenderortanks(Span(20.95)–Zero(20.05)=0.9delta).
AnalyzerCalibration:Flowzeroandspangases,changeslopeandoffsetvaluesintheanalyzer.Ifanalyzervaluesareoutofrange(<19.8%or>21.2%forO2;>1.1%forCO2)performcalibrationonanalyzers,whenanalyzersarestable,performcalibrationonsoftware,ifneeded.
SoftwareCalibration:Flowzeroandspangases,changeslopeandoffsetvaluesinthesoftware
Delta:Thedifferencebetweenthespanvalueandthezerovalue.Subtracttheapparentconcentrationofthespanvaluedisplayedonthecomputerfromtheapparentconcentrationonthecomputerofthezerovalue.
𝐷𝑒𝑙𝑡𝑎 = 𝑧𝑒𝑟𝑜 − 𝑠𝑝𝑎𝑛 eq(1)0.9 = 20.95 − 20.05 eq(2)
Datafileisrecordingwhentheloggingbutton(J)displays“LoggingOn”andEnterNotebox(K)appears.Filerecordingwillstopwhenloggingbuttondisplays“LoggingOff”or“Done”(L)ispressed.“Done”alsoclosesthewindow.
A
B
C
ED
C
ED
F
IHG
KJ
LFigure2.CalibrationScreen
Updated:16October2017 RoomCalorimeterCore 6
BlenderCalibration:
1. OpenN2,CO2andO2tanks2. FlipInstrumentationPanelvalvestomatchthoseshowninFigure3.Flipblendervalvesfor
calibration.3. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer4. Oncethegasisflowing,watchtheO2(A)andCO2(B)graphsuntilthevaluesaresteady.Turn
Averagingon,andletblenderrunforatleastanhour.5. Afterhouriscomplete,pressthe“SetZero”button(D).6. Afterthevaluesarestored,enterthecalibrationtank’sgasconcentrationsintheirrespective
entryboxesinthe“CalGas”column(C)(i.e.ifacalibrationgasconsistingofablendof21.02%O2and0%CO2isflowingintotheanalyzers,21.02wouldbeenteredinthe“CalGas”columnfortheO2analyzersand0wouldbeenteredfortheCO2analyzers).DefaulttankvaluescanbechangedonthePreferencesscreen.
7. Repeatsteps3-6usingtheSpanBlend.8. Selectradiobuttons(F)foreachanalyzeryouwouldliketocalibrate.Buttonswillturnblueif
calculationswillbeperformedforthatanalyzer.9. Afterzeroandspanvalueshavebeenstored,press“Calculate”(G).Newslopesandoffsets
arecalculated,butnotappliedtoanalyzergasvalues.10. Ifslopesandoffsetsareacceptable,press“Apply”(H).Newslopesandoffsetsaresavedto
theconfigurationfile.11. Recordthenewslopeandoffsetvaluesinlabnotebook.12. FlowZerogasblendthroughtheanalyzerstoconfirmthecalibrationvalues.Repeatwiththe
Spangasblend.Ifeitherblenddoesnotmatchthesetvalues,repeatthecalibration.
Figure3.PanelLayoutforBlenderCalibration
Comment[MB4]:Whatareacceptablevalues?
Updated:16October2017 RoomCalorimeterCore 7
GasTankCalibration
1. Openzeroandspantanks2. FlipvalvesasthoseshowninFigure43. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer4. Oncethegasisflowing,watchtheO2(A)andCO2(B)graphsuntilthevaluesaresteady.Once
steady,pressthe“SetZero”button(D).5. Afterthevaluesarestored,enterthecalibrationtank’sgasconcentrationsintheirrespective
entryboxesinthe“CalGas”column(C)(i.e.ifacalibrationgasconsistingofablendof21.02%O2and0%CO2isflowingintotheanalyzers,21.02wouldbeenteredinthe“CalGas”columnfortheO2analyzersand0wouldbeenteredfortheCO2analyzers).DefaulttankvaluescanbechangedonthePreferencesscreen.
6. Repeatsteps3-6usingtheSpanCalibrationGas.7. Selectradiobuttons(F)foreachanalyzeryouwouldliketocalibrate.Buttonswillturnblueif
calculationswillbeperformedforthatanalyzer.8. Afterzeroandspanvalueshavebeenstored,press“Calculate”(G).Newslopesandoffsets
arecalculated,butnotappliedtoanalyzergasvalues.9. Ifslopesandoffsetsareacceptable,press“Apply”(H).Newslopesandoffsetsaresavedto
theconfigurationfile.10. Recordthenewslopeandoffsetvaluesinlabnotebook.11. FlowZerogasblendthroughtheanalyzerstoconfirmthecalibrationvalues.Repeatwiththe
Spangasblend.Ifeitherblenddoesnotmatchthesetvalues,repeatthecalibration.
Figure4.PanellayoutforSpanGasCalibration(flipthehighlightedCalibrationSourcevalvefromSpanGastoZeroGasforspancalibration)
Updated:16October2017 RoomCalorimeterCore 8
TankCalCheck
Purpose:Checkanalyzercalibrationvaluesbeforeastudy(humanandvalidation).
1. OpenZeroandSpancalibrationtanks2. SetpanelupasshowninFigure4.3. Flowzerogastotheanalyzersbyturningthe“CalibrationSource”valveto“ZeroGas”.4. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer5. Oncethegasisflowing,watchtheO2(A)andCO2(B)graphsuntilthevaluesaresteady.Once
steady,recordthezerovalue.6. Repeatsteps3-4usingthespangas.7. Calculatethegasdeltabysubtractingthezerogasmeasuredontheanalyzerfromthespan
valuemeasuredonthevalue.8.
𝐷𝑒𝑙𝑡𝑎 = 𝑍𝑒𝑟𝑜𝑇𝑎𝑛𝑘𝐺𝑎𝑠𝑃𝑒𝑟𝑐𝑒𝑛𝑡𝑎𝑔𝑒 − 𝑆𝑝𝑎𝑛𝑇𝑎𝑛𝑘𝐺𝑎𝑠𝑃𝑒𝑟𝑐𝑒𝑛𝑡𝑎𝑔𝑒 ex.O2:0.9 = 20.95 − 20.05ex.CO2:0.99 = 0 − 0.99
9. Verifythedeltaiswithin+/-3%oftheexpecteddelta.10. Ifdeltaisnotwithin+/-3%oftheexpecteddelta,performaTanksoftwarecalibration.
Table1.DeltaExample
ExpectedDelta
AllowableDifference
0.9 0.0271 0.03
Updated:16October2017 RoomCalorimeterCore 9
NullTest
Anulltestverifiestheoffsetofthegasanalyzers.Thetestisperformedbyflowingtheinflowair(medair)toboththeinflowandoutflowanalyzers.AnoffsetcorrectionismadeontheCalibrateAnalyzersscreeninCalRQTMorcanbemadeduringpostprocessing.
1. Openafiletologdata(HumanStudy,ValidationStudy,orCalibration)2. TurntheSampleSourceValvetoNull(Figure5).3. Letthesystemrununtilthehumidityreading,plusVO2andVCO2onthesoftwarestabilize.This
indicatesthatthedryersareconditioned.4. Oncethereadingsarestable,recordthenulldataforatleast10min.5. Ifapplyingthenullcorrectioninrealtime,calculateoffsetcorrectionbysubtractingInflowO2
concentrationfromOutflowO2concentration.RepeatforCO2.ApplyoffsetcorrectiontotheinflowanalyzeroffsetforbothO2andCO2.
Figure5.Panellayoutfornulltest
Updated:16October2017 RoomCalorimeterCore 10
ValidationTestingShortCircuit
Figure6.ShortCircuit(SC)RackPanelConfiguration.Redboxeshighlightthevalvesthatmustbeturnedforthecorrectgasstreamtoreachtheanalyzers.**ShadedredboxhighlightsthevalvethatdetermineswheretheSCinfusionissent.Turnthevalvetoselectthedesiredchamber**
Updated:16October2017 RoomCalorimeterCore 11
Figure7.ShortCircuitMFCPanelConfiguration
1. SetuprackpanelasshowninFigure4.ChangeMFCpanelasshowninFigure7
2. Figure9.3. StartaValidationStudy(File->NewValidationStudy)4. ChooseInfusionRadioButton5. EnterValidationID–FilenamewillbeValidationID–Chamber#-Infusion-Date6. SetdesiredInflowandOutflowrates(Edit->ManualMFCSet)
a. InflowRate=60lpmb. OutflowRate=0lpmc. UsethehighflowMFCforchamber2.Setto60lpm
7. EnterflowrateinInflowSettingpress“ApplyInflow”.CheckCurrentInflowRateisasexpected.
Updated:16October2017 RoomCalorimeterCore 12
8. Runanull–samegasshouldbegoingtoboththeinflowandoutflowanalyzers(Valvesonthepanelshouldbepointingtonull)
9. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer10. Setchambervolumeto0onthePreferencesScreen.11. OpenBlenderScreen(Edit->AdjustBlender).ChooseFlowRateradiobutton12. SelectAppropriateMFCandsetflowrateorselectdesiredprofile.Press“Apply”tostartflow
throughblender.13. FloweachN2/CO2blendforatleast10minafterVO2andVCO2havestabilized
a. Quickcheck:VCO2shouldbeclosetoCO2flowrate;VO2shouldbecloseto
(N2flow+CO2flow)*[O2inflow]
Updated:16October2017 RoomCalorimeterCore 13
Infusion
Figure8.InfusionPanelConfiguration-Redboxeshighlightthevalvesthatmustbeturnedforthecorrectgasstreamtoreachtheanalyzers.**Shadedredboxhighlightsthevalvethatdetermineswheretheinfusionissent.Turnthevalvetoselectthedesiredchamber**
Updated:16October2017 RoomCalorimeterCore 14
Figure9.ChamberStudyMFCPanelSetup
1. SetupRackpanelasshowninFigure6.Selectthecorrectchambertosendgasinfusion.ChangeMFCpanelasshownin
2. Figure93. StartaValidationStudy(File->NewValidationStudy)4. ChooseInfusionRadioButton5. EnterValidationID–FilenamewillbeValidationID–Chamber#-Infusion-Date6. SetdesiredInflowandOutflowrates(Edit->ManualMFCSet)
a. InflowRate=60lpmb. OutflowRate=0lpmc. UsethehighflowMFCforchamber2.Setto60lpm
7. EnterflowrateinInflowSettingpress“ApplyInflow”.CheckCurrentInflowRateisasexpected.
Updated:16October2017 RoomCalorimeterCore 15
8. Runanull(~15mins)–samegasshouldbegoingtoboththeinflowandoutflowanalyzers(Valvesonthepanelshouldbepointingtonull)
9. Checkflowrates(1.25lpm)andpressures(~2psi)foreachanalyzer10. SetChamberVolumetoappropriatevolume(Chambers=32500liters,Flex=5000liters)11. Closechamberdoor.12. Addanotetodatafilewithchambersetup(whatfurnitureistheroom,platform,bikeor
emptyinthesmallroom)13. SwitchnullvalvetoStudy.Letrunforabout15minutes14. OpenBlenderScreen(Edit->AdjustBlender).ChooseFlowRateradiobutton15. SelectAppropriateMFCandsetflowrateorselectdesiredprofile.Press“Apply”tostartflow
throughblender.16. Setflowratesandselect“Apply”17. VerifyOutflowO2andOutflowCO2valuesarechangingasexpected.