Part Number: RTC-SS Rev 9
INSTALLATION/OPERATIONMANUAL
Notes:
Part Number: RTC-SS Rev 9
Part Number: RTC-SS Rev 9
TABLEOFCONTENTS:
PageIntroduction 1SafetyInstructions 2‐3BlockDiagram 4CheckingMaterials 5‐6RoomLayoutandInstallationoverview 7HowtoConnectTransducerPowercables 8‐9RunningMRLaserLinkCable 10ComponentsInterconnection 11‐13ControlRoomSetup 14PatientSetup 15OperationandTroubleshooting 15‐16
Part Number: RTC-SS Rev 9
Congratulations on your purchase of the Serene SoundDigitalMRI‐compatible system. This devicerepresentsmorethan24yearsofdevelopmentandstate‐of‐the‐artengineering.WeareconfidentthisproductwillgiveyouthetoolsyouneedforcomfortingthepatientwhileundergoingMRIscans.Thisuser/installationmanualoutlineshowtoproperlycareforandoperatethissystem.ThankyouforchoosingtopurchasethissystemfromResonanceTechnology,Inc.theleaderinfMRIandMRIpatientcomfortsystems.Suggestionsonhowtoimprovethissystemarealwayswelcome.
Sincerely,MokhtarZiaratiResonanceTechnology,Inc.PresidentandCEO
Page1
Introduction
Part Number: RTC-SS Rev 9
ResonanceTechnology,Inc.,patientsafetyisourtoppriority.Iftheheadsetbecomesdamaged,stopusingtheheadsetimmediatelyandnotifyResonanceTechnology,Inc.CustomerServiceforassistanceat(818)882‐1997or(800)428‐6788.Useofbrokencomponentscancauseinjurytotheclinicianorthepatient.
Caution:Seizures.Ifyouhaveahistoryoftemporaryspasms,unconsciousness,orepilepticseizures,consultyourdoctorbeforeusingthissystem.Useofthisproductbysuchindividualsmaycausespasms,unconsciousnessorseizures.Ifyouexperiencethesesymptoms,stoptheusingthissystemimmediatelyandconsultyourdoctor.System:WarrantyandShelflifeThesystemcomeswithaoneyearwarrantyonallpartsandlabor,andashelflifeoftwoyearsstartingfromthedayofitsoriginalinstallation.ResonanceTechnology,Inc.suggestamaintenancecalleverysixmonthsaftertheoneyearoriginalwarrantyperiodtocheckuponfrequentlyuseditemssuchasthepatientaudioheadsetandtheTechnologistRemoteControl.Thiswayyouwillmakethebestuseofthissystemforyourfacilityandyourpatients.
Caution:SoundVolume.Avoidusingtheaudioheadsetonpatientsathighvolume.Hearingexpert’sadviceagainstcontinuousloudandextendedaudioplay.Ifyourpatientsexperienceringingintheirears,reducetheaudiovolumeordiscontinueusethesystemuseuntilthepatientfeelscomfortable.
Thisdocumentcontainsproprietaryinformationthatisprotectedbycopyright.Allrightsarereserved.NopartofthisdocumentmaybephotocopiedorreproducedwithoutpriorwrittenconsentofResonanceTechnologyInc.InformationcontainedinthisdocumentiscoveredbyoneormoreofthefollowingU.S.patents5627902,5877732,5432544,5412419
Page2
Part Number: RTC-SS Rev 9
WARNING:Inspectsystemcomponentsthatcomeincontactwiththepatientpriortouse.Discontinueproductusageifanydamageisvisibleorpresentsotherpotentialhazards.Useofbrokencomponentscancauseinjurytotheclinicianorthepatient.
Caution:MRIEnvironmentHazards.InstallationofmaterialsinsidetheMRIsuitemustbedonewithextremecaution.Takecarethatferromagneticmaterialsbekeptatleastthreemetersawayfromthemagnetandthatnoinstallationshallbedonenearthefilterpanelifascanisinprogress.Inaddition,nopersonswithferromagneticprostheticdevices,suchaspacemakersorjointreplacements,shouldentertheMRIsuiteatanytime.Extremehighmagneticfieldsinsidethemagnetroomhavethepotentialtodislodgeitemsathighvelocitiesandcanresultinseriousinjury,ordeath.Forquestionsregardinginstallationproceduresortechnicalsupport,callResonanceTechnology,Inc.viatelephoneMondaysthroughFridays,8a.m.to5p.m.PacificStandardTimeat(818)882‐[email protected],Inc.willnotbeliableforanyinjuryorpropertydamagethatoccursasaresultofimproperinstallationproceduresorimproperuseofthissystem.Byagreeingtothisnotice,userscertifythattheyarefamiliarwithbasicMRIsafetyproceduresandthattheyhavereadandunderstandthisnotice.OnlysystemcomponentsexplicitlydesignatedforuseinsidetheMRIsuiteshouldbeplacedinsidethemagnetroom.ComponentsnotdesignatedforMRIusemaypresentaprojectilehazardandcanbecomeairborne,causingseriousinjury,damageordeath.
Caution:ElectricShock.Failuretoobservealloperatingandmaintenanceinstructionsmaycausedamagetoyourproduct,propertydamage,orinjuryordeathfromelectricshock,fireorothercauses.Caution:DoNotDisassemble.Onlytrained,authorizedpersonnelshouldperformanyrequiredserviceonthisproduct.Failuretocomplywiththiswarningmayresultinpropertydamage,injuryordeathfromelectricshock,fireorothercauses.Caution:AvoidExtremeHeat,Wet,Humid,DustyandSmokyEnvironments.Thisproductmaybedeformedbyhightemperature,directsunlightanddroppingorothermechanicalshock.Donotexposetheunittorainormoisture.Caution:UnplugtheProductWhenNotinUse.Alwaysunplugthisproductwhennotinuseforextendedperiods,suchasduringvacationsandMRImaintenance.
Page3
Part Number: RTC-SS Rev 9
BlockDiagram
SereneSoundSystemComponents
Note:Theantennashown in theSereneSoundTransducerunit isanoption for futuresystemfeatures thatmaynotbeincludedwith your system. If the antenna ismissingproceedusing the system sincenoneof the system functions areaffectedinanyway.
Page4
Patient Audio Headset
Serene Sound Transducer
AM/FM-CD-Ipod Iphone Sound System
Serene Sound Controller
Tech Remote Unit Magnet Room Components (MRI SAFE)
Technologist Room Components (NOT MRI SAFE)
Part Number: RTC-SS Rev 9
CheckingMaterials
YourSereneSoundsystemcomescompletewithallthenecessarycomponentsforfullinstallationatyoursite.Thefollowingchecklistisprovidedforverification:
PartNumber Quantity Description
‐‐MainItems‐‐RTC‐660‐170‐370‐000 1 SereneSoundControllerRTC‐660‐050‐535‐000 1 SereneSoundController PowerSupplyRTC‐660‐050‐540‐000 1 SereneSoundTransducerRTC‐650‐050‐005‐000 1 PowerSupply(ForTransducer)RTC‐661‐050‐005‐000 1 SwitchingPowerSupplyRTC‐650‐050‐260W‐000 1 LinearPowerSupply(optional)RTC‐LPS 1 LinearPowerSupplyInstallation/OperationManual(optional)RTC‐650‐020‐000‐000 1 TechnologistRemoteControlRTC‐650‐062‐000‐000 1 AudioHeadsetRTC‐651‐060‐621‐000 1 HeadsetHeadbandRTC‐ALS‐SFA 1 StereoSystemRTC‐SS 1 Installation/OperationManual(Thismanual)PartNumbervaries 1 FilterPlateRTC‐650‐300‐605‐000 1 9pinD‐SubFilter ‐‐Accessories‐‐RTC‐ALS‐HEC 50 HeadsetEarpiece CoversRTC‐ALS‐TWR 20 TieWrapsRTC‐ALS‐TWH 20 TieWrapHoldersRTC‐650‐060‐052‐000 10 HeadsetMicrophoneCushionsRTC‐651‐000‐453‐000 1 PatientHeadrestRTV‐108 1 SiliconeGlueTubeRTC‐ALS‐HDH 1 HeadsetHook17304 1 CableBundler ‐‐Cables‐‐RTC‐101‐102‐001‐101 1 MRLaserLink Cable
RTC‐101‐950‐003‐000 1 RG‐59AntennaCable‐ 35feet
RTC‐330‐170‐404‐000 1 MiniStereotoMiniStereoCable(male)ATS300SP 1 “T”FM Transformer forradioantennaRTC‐101‐210‐001‐101
1 Transducer DCPowerCords(3m)
RTC‐101‐245‐001‐001 1 Transducer DCPowerCord(10m)
RTC‐101‐306‐003‐000 2 HospitalGradePowerCords (USA)
RTC‐ALS‐HGC 2 HospitalGradePowerCords(Europe,etc)
Page5
Part Number: RTC-SS Rev 9
CheckingMaterials(continued)
Thefollowinglistdetailsplacementofallequipmentcomponents:
PartNumber Quantity Description
‐‐ControlRoom–
RTC‐660‐170‐370‐000 1 SereneSoundController
RTC‐660‐050‐535‐000 1 SereneSoundController PowerSupply
RTC‐101‐306‐003‐000 1 HospitalGradePowerCord
RTC‐650‐020‐000‐000 1 TechnologistRemoteControl
RTC‐ALS‐SFA 1 StereoSystem
RTC‐660‐000‐072‐101 1 Installation/OperationManual(Thismanual)
RTC‐101‐950‐003‐000 1 RG‐59AntennaCable‐ 35footARX‐Mini315 1 3.5mmMiniStereocable
ATS300SP 1 “T”FMTransformerforradioantenna
‐‐InsidetheMagnetRoom‐‐
RTC‐660‐050‐540‐000 1 SereneSoundTransducer
RTC‐101‐245‐001‐001 1 Transducer DCPowerCord (10mlong)
RTC‐650‐062‐000‐000 1 AudioHeadset
RTC‐101‐102‐001‐001 1 MRLaserLinkCable
RTC‐ALS‐HDH 1 HeadsetHook
‐‐OnthepenetrationpanelOutsidetheMagnetRoom‐‐
RTC‐101‐102‐001‐101 1 MRLaserLinkCable
RTC‐101‐210‐001‐101
1 Transducer DCPowerCords(3m)
RTC‐650‐300‐605‐000 1 DB‐9D‐SubFilter
PartNumbervaries 1 FilterPlate
Page6
Part Number: RTC-SS Rev 9
RoomLayoutandInstallationOverview
BelowisthestandardsetupfortheSereneSoundsystem.Yourindividualinstallationmayvary,butwillgenerallybedistributedinthreeareas:1)ControlRoom,2)ComputerRoomand3)MagnetRoom.
TheControlRoomsetupconsistsofplacementandconnectionoftheSereneSoundController,TechnologistRemoteControlandAM/FM/CD/I‐pod/I‐phoneStereosystem.AsingleLaserLinkFiberOpticscablebundlewillrunfromtheSereneSoundControllerthroughthepenetrationpanelwave‐guidetotheSereneSoundTransducer.Withtheexceptionoftheheadset,allMagnetRoomSereneSoundcomponentsmustbeinstalledtothesideofthemagnetshroudalsothetransducerpowercableshouldneverinparallelwithmagneticcoil. Additionally,thesecomponentsshouldbeplacedinalightlytraffickedarea.
Page7
Serene Sound Controller
Headset
9pin D-Sub Filter Connection
Part Number: RTC-SS Rev 9
HowToConnectTheTransducerPowerCables
InordertoconnecttheTransducerpowercabletothepowersupplyoutsidethemagnetroom,youneedtouseaDB‐9filterprovidedwiththesystem.Makesuretoconnectthetwometerlongcabletoboththepowersupplyandpenetrationpanelandconnectthe10meterlongcabletoboththepenetrationpanelandtheTransducer.GEFilterPanelPlate:PlugtheFemale9Pin“D”endofthecablethathasbeenrunthroughwiththefiberopticsfromtheControlroomtotheEquipmentroom.Connectthecable’sFemaleendtotheJ‐51plugfoundintheGEFilterPanel
SiemensFilterPanelPlate:RemoveanyoftheblankplatesonthefilterpanelwithametricAllen(hex)wrenchandinstalltheRTCsuppliedfilterpanelplatefortheSiemensMagnet.
or
Page8
10 meters long power cable connected inside the magnet room
2 meters long power cable connected outside the magnet room into the DB-9 filter and power supply.
DB-9 Filter installed on the penetration panel used to connect the 2-meter power cable to the 10-meter
Facility Penetration Panel
Part Number: RTC-SS Rev 9
RunningtheMRLaserLinkCable
Phillips Filter Panel Plate: It is advised to drop the female end of the 2ft 9pin D cable through the filter panel into the MRI room side and to attach it to the male end of the filter before putting back the removed plate.
Page9
RTC recommends the Transducer Power Supply be installed outside the magnet room.
Part Number: RTC-SS Rev 9
TheMRLaserLinkcableisfedthroughthewave‐guideinthePenetrationPanelandthenroutedtotheTransducerunitinsidethemagnetroom.CaringFortheFiberOpticsCable:Note:Takecaretoensurethecableisnotbentinradiusoflessthan15cm.
Page10
RunningtheMRLaserLinkCable
r ≥ 90° or Right angle
r
r ≤ 90° or Right angle
r
MR Laser Link Cable MR Laser Link Cable
Part Number: RTC-SS Rev 9
ComponentsInterconnection
A.MagnetRoom
PlacetheSereneSoundTransducerinasecureareaawayfromtrafficasillustratedbelow.InstalltheaudioheadsetfirmlyintotheTransducerusingthebracketprovided.MakesurethattheaudioheadsetconnectorisconnectedallthewayintotheTransduceraudiosocketforproperinstallation.
Page11
SERENE SOUND TRANSDUCER PREFERRED LOCATION
Press down firmly on the audio headset connector into the Transducer.
Part Number: RTC-SS Rev 9
ComponentsInterconnection(continued)
MagnetRoom(cont’d)
AttachtheMRLaserLinkconnectorstotheTransducerasillustratedbelow.Matchthecablecolorstothoseontheunit’slabels.
Note:Remembernottobendtheindividualcablessmallerthana15cmradius.
Page12
Press down firmly on the audio headset connector into the Transducer.
Fiber Optic Cable
Transducer Power (DC) Cable
r≥ 90° or Right angle
r
r< 90° or Right angle
r
Individual Cable Individual Cable
Part Number: RTC-SS Rev 9
ComponentsInterconnection(continued)
B.ControlRoom
ConnectallappropriatecablestotheControllerasillustratedbelow.
Connectthe5‐pinRoundCablefromtheControllerPowerSupplytotheControllerconnectorlabeledPOWER.
ConnecttheHospitalGradePowerCordtotheControllerPowerSupplyandintoanavailableactiveACpoweroutlet.
ConnecttheTechnologistRemoteControltotheControllerconnectorlabeledREMOTE.
ConnecttheministereojackcabletotheControllerconnectorController*labeledControlRoomAudioIN.
ConnecttheMRLaserLinkcabletotheControllerconnectorlabeledOpticLink.
Note:MakesuretomatchthecoloroftheMRLaserLinkcablestotheController’slabels.
Connecttheministereojackcablefromrecord‐outtocomputeraudioinput.
Thisfeaturecanbeusedtorecordthepatients’headsetmicrophoneresponseintoacomputermicrophoneinput.Thisisanoptionalfeature.
ApplypowertotheControllerfirstandthenturnontheTransducerPowerSupply.Waituntilthesystemisfullysynchronized(logolightsrotatingontheTransducerinsidetheMagnetRoom).
*TheControllerisIPODandIphoneready.
Page13
Part Number: RTC-SS Rev 9
ControlRoomSetup
TECHNOLOGISTREMOTECONTROLOPERATION:!
Wheel#1:PatientAudioVolume(scrollupanddowntoadjust)Scrollingthewheelupincreasesthepatient’sheadsetvolume.Scrollingthewheeldowndecreasesthepatient’sheadsetvolume.
TechnologistMicrophone–Pressandholdthetalkbutton.ThenscrolltheadjustmentwheelThisadjustmentistochangethevolumethatthepatientcanhearduringcommunication.
Wheel#2:PatientMicrophoneVolume(scrollupanddowntoadjust)Scrollingthewheelupincreasesthevolumeofthepatient’smicrophonevolumeheardinthecontrolroom.Scrollingthewheeldowndecreasesthevolumeofthepatient’smicrophone.
Wheel#3:MainVolume(scrollupanddowntoadjust)Scrollingthewheelupincreasesthecontrolroomvolume(youneedtoconnectexternalspeakerstotheController).Scrollingthewheeldowndecreasesthecontrolroomvolume.
Wheel#4:MainMenu(scrollandpresstoselectsub‐menuitems)Pressingdownonthiswheelactivatesthesystemmenu.Scrollingthewheelupordownrotatesthroughthesystemfunctions.TosetthepatientmicrophoneinAUTOorMANUALmodeusethemenuoptionstoselectthedesiredmode.WhenintheMANUALmodeyoumustpressthetalkbuttontoactivatemicrophone.Toadjustyourtechnologistmicrophonevolumebuiltintothetechnologistremoteunit,simplypressandholdtheTALKbuttonandatthesametimescrollthewheel#1tosetthePTTvolumetoabout60to70.Ifyourvoicesoundstooloudinthepatient’saudioheadsetlowerthePTTvolume.
Page14
PleaseNote:Ifthetechremoteunitisnotrespondingyoucanresetittoitsoriginalfactorydefaultsettingsbyfollowingthese5simplesteps:1st:SetPatientAudio(AdjustmentWheel#1)to12nd:SetPatientMicrophone(AdjustmentWheel#2)to23rd:SettheMainVolume (AdjustmentWheel#3)to34th:Pressandholdtalkbuttonfor15seconds
5th:Releasethetalkbutton.Allthesettingshavebeenresettotheoriginalfactorydefaults.Younowhavetore‐adjustallthewheels settings.
Notice:Alwaysadjusttheaudioinputsourcetoalevelthatwillnotcausedistortion.Otherwiseacracklingnoisewillbeheardonthepatientaudioheadset!
#1 #2 #3 #4
Part Number: RTC-SS Rev 9
PatientSetup
MakesureallcableshavebeenconnectedsecurelyintotheTransducerandthatthecomponentsdonotimpedewalkways.EnsurethattheTransducerisnotplaceddirectlyinfrontofthemagnetboreordirectlyunderthebed.Beforeplacingtheheadsetonthepatient,makesurethatthetechnologistmicrophonevolumehasbeensettoacomfortablelevelforthepatient.Placetheaudioheadsetoverthepatient’searsandadjusttheheadband.Ifnecessary,positiontheheadsetmicrophoneaboutonecm.awayfromthepatient’smouth.
Operation
Whenoperatingnormally,thelogolightsontheTransducerarerotatingandthesidesoftheControllerareblue.However,aftertwotothreehoursofnon‐operation,thesystemgoesintoaSLEEPmodeandthesystemappearstobeturnedoff.Towakeupthesystem,justpressdownonceontheTALKbuttonoftheTechnologistRemoteControlandwaitforthesystemtosynchronizecompletely.Remembertoalwayscommunicatewiththepatienttoadjusttheheadsetaudiovolume.Alwaysstartatalowsettingprogressivelytoahighersettinguntilthepatientcanhearyouclearly.Readthestereosystemusermanualforitsoperation.Belowisanillustrationhowtoconnectthecontrolroomcomponents:
Page15
Technologist Remote Control
Controller
AM/FM/CD/Ipod/ Iphone Stereo
FM radio antenna
Stereo system power cord
Audio ouput cable connects into the Controller
Controller DC power cord attached to the Controller Power Supply
Fiber optics cable that connects to the Transducer in the Magnet Room
Audio input cable from the stereo system
Part Number: RTC-SS Rev 9
SystemTroubleshooting
Note:WhendisconnectingpowertotheController,alwaysdisconnecttheACpowercordthatconnectsintotheControllerPowerSupply.NeverdisconnecttheDCcableconnectorfromtheController.Tore‐connectpowertotheTransducer,justre‐connecttheACpowercord.
Symptom Verify
NoSynchronization ThefiberopticsareintherightpositionontheControllerandTransducer.
NoAudio Checkthemini‐stereocableproperlyconnectedontheControllerinput.VerifytheheadsetisproperlyseatedintheTransducer.
NoMicrophone OntheTechnologistRemoteControl,checkcommmodeisautoormanual.Checkmicrophonevolumesetting.
SystemReboots UnplugthepowersupplyfromtheA/Coutlet.UnplugtheController;turnoffthepowersupplyfortheTransducer.
Ifthesystemhangsup*orfailstorespondtotheTechnologistRemoteControlcommands.
PowerdowntheControllerforapproximately10secondsandthenpowerupagain.
Ifthesystemstilldoesnotrespond. Powerdownthecompletesystemforapproximately10seconds,thenpoweruptheControllerfirstintheControlRoom,thenpoweruptheTransducerintheMagnetRoom.WaitforthelogolightstorotateontheTransducerandthenusethesystemnormally.
*SystemHangupiswhenthesystemrefusestofunctionnormally.
ControllerUnitPowerSupply
Ifthesystemstillfailstorespond,contactResonanceTechnologyfortechnicalassistanceat(818)882‐1997.
Page16
AC power cable connects into an AC outlet.
DC power cable connects into the Controller.
Part Number: RTC-SS Rev 9