Agilent Private Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and Microscopy 21 June 2012 1

Transcript of Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the...

Page 1: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Agilent Private

Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples

Lou Tisinger

FTIR/FTIR Imaging and Microscopy

21 June 2012 1

Page 2: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

POM

QA/QC, Geoscience,

Art Conservation,

Materials & Surfaces

Analyzer

Fuel BioDiesel QA/QC

Routine

QA/QC Pharma & Chemical, Academia

Mid

Chemicals, Petrochem, Food/Ag,

Forensics, Bio Research

High

Materials & Polymers,

Bio Research, Forensics

FTIR Product Line

Page 3: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Agilent 4500/5500 DialPath/Tumbler makes measurement easy

Agilent 4500/5500 DialPath/Tumbler

• Easy to clean liquid cell • Reproducible path length

Page 4: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Agilent Private

DialPath Sampling Technology

•Agilent-exclusive DialPath/Tumbler technology features an “open cell” transmission design for qualitative and quantitative FTIR analyses of

liquids. •As easy for liquid measurement as using ATR, and with the added

benefit of variable and extended pathlength

4

Page 5: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Three simple steps to measuring liquids!

Pathlength (μm) Typical Conc. Range 30 Neat - 0.1% 50 50% - 500 ppm 100 20% - 100 ppm 200 10 % - 50 ppm

Page 6: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Agilent Private

Dialpath Geometry 3 µ L Fill Volume

250 µ L Fill Volume

µ L Fill Volume

250 µ L Fill Volume

The combination of class-leading performance and rapid scanning capability means that most liquid samples can be measured in < 30 secs. Volatile liquids require > 120 secs. to show any appreciable change in concentration, once placed in the DialPath cell that is in the closed position.

Page 7: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Advantages of DialPath Technology for analysis of liquids • Non-viscous and/or volatile liquids

can easily fill the DialPath, even when in the “close” position, by simple “capillary action” as touching the liquid between the windows, will draw the liquid in.

• Viscous samples (oils, syrups, etc) can easily fill DialPath in open position

• Effective for volatile solutes and/or solvents since detectable evaporation takes much longer than typical measurement times.

• Post analysis cleaning is a simple process; both windows are exposed and need only be wiped off

3 µ L Fill Volume

250 µ L Fill Volume

µ L Fill Volume

250 µ L Fill Volume

7

Page 8: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Agilent Private

• Standards of benzene in hexane were prepared

• Evaluation samples of 0.5% benzene in hexane were prepared and measured. 18 sec scan time was employed for both calibration and evaluation samples

• Reproducibility of 0.5% sample was measured with increasing time delays of 0 to 60 sec.

• Red line is mean benzene concentration measured at zero delay time; purple lines are +/- 2% error bars

Benzene in Hexane Method Prediction Variations of a Nominally 0.5vol% Standard, Testing the Possible Effects of Solvent and Solute Evaporation, Using

the TumblIR Transmission Cell

0.51

0.515

0.52

0.525

0.53

0.535

0.54

0.545

0.55

0.555

0.56

0 0 45 45 45 60 60 60Time Delay (seconds) After Adding Sample to TumblIR Cell

Be

nze

ne

(V

ol%

)

Benzene Vol%

Concentration of a volatile benzene analyte measured in volatile hexane solvent shows no significant change up 60 secs after analysis initiates.

Solvent evaporation with DialPath – Benzene in hexane

Page 9: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Agilent Private

Summary

• Dialpath/Tumbler provides for liquid transmission as easy as ATR

• Dialpath provides for a variable pathlength, from 30 to 250 microns to greater range of concentrations measurements

• Suitable for viscous and even highly volatile liquids, with little or no detectable evaporation effects up to 1 minute measurement time

Page 10: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Environment – Oil in Water by Extraction

Oil in water a key environmental measurement

• FTIR measurement by Extraction

• Historically extracted by Freon 113 (ASTM D3921) • Currently extracted by S 316 (ASTM 7066-04) • Also extracted by cyclohexane in ASTM D7678 • Oil measured directly in extracted fluid using Transmission

Agilent 4500/5500 DialPath makes measurement easy

• Easy to clean liquid cell • Reproducible path length • Easy to use software

Red or Green results will tell the user if there is critical warning and recommend the next steps

Page 11: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Environment – Oil in Water by Cyclohexane Extraction (ASTM D7678)

Previous Freon FTIR extraction method is being phased out

• Freon and other halogenated solvents have been banned by the Montreal protocol due to their ozone depleting activity

Cyclohexane is a CFC free non-ozone depleting solvent

Agilent’s FTIR version of the ASTM D7678 features a limit of quantification (LOQ) of .75mg/L (0.75ppm) oil in water

Standard procedures for liquid-liquid solvent extraction remain unchanged from previous ASTM methods.

These ASTM D7678 based FTIR results will correlate to other methods

• ASTM D3921, D7066, ISO 9377-2, EPA 413.2, and 418.1 methods

Page 12: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Simple 5 step process 1. Add Cyclohexane Add 20mL of cyclohexane to 900mL of process water and vigorously shake for 2minutes.

2. Remove Top Layer Allow layers to separate, and remove the top layer for analysis

3. Add Cleaning Agents Add 2g drying agent (sodium sulfate) and 2g Florisil™ to the cyclohexane extract, shake

vigorously, and allow to settle for 2minutes.

4. Filter Cyclohexane Extract Filter the “cleaned” cyclohexane with a 0.45um nylon syringe filter (17mm).

5. Measure FTIR Spectrum Add the cyclohexane to the Dialpath™ or TumblIR™ cell and initiate the FTIR scanning.

The result will be displayed after 30seconds of scanning.

Environment – Oil in Water by Cyclohexane Extraction (ASTM D7678) Simplified Procedure

Page 13: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

The overlaid FT-IR spectra of crude oil (Blue) and mineral oil (Red). The zoomed region indicates the local baseline (dashed line) used for the 1378cm-1 peak area measurement by FTIR.

811.

1

1604

.8

2956

.729

25.8

2855

.0

1459

.913

77.8

868.

5

722.

0

3500 3000 2500 2000 1500 1000

1.2

1.0

0.8

0.6

0.4

0.2

0.0

Wavenumber

Abs

orba

nce

1378

1461

1600 1550 1500 1450 1400 1350

0.6

0.5

0.4

0.3

0.2

0.1

0.0

-0.1

Wavenumber

Abs

orba

nce

Mineral oil (abs area) = 2.661

Page 14: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

The FTIR spectrum of cyclohexane, Dialpath attachment. The zoom box indicates the overlaid spectral region of the hydrocarbons using the methyl absorbance at 1378cm-1.

4333

.5

4173

.141

10.4

4083

.340

07.8

3961

.3

3797

.237

56.0

3715

.536

36.5

3565

.8

3443

.8

3318

.6

3155

.330

94.1

2553

.5 2477

.8

2367

.223

04.1 22

48.5

2177

.321

38.2

2060

.1 1989

.719

29.4

1871

.9

1777

.116

83.6

1644

.2

1349

.6

1109

.3

4500 4000 3500 3000 2500 2000 1500 1000

3.0

2.8

2.6

2.4

2.2

2.0

1.8

1.6

1.4

1.2

1.0

0.8

0.6

0.4

0.2

0.0

-0.2

Wavenumber

Abs

orba

nce

Hydrocarbons, Methyl Bend Absorbance

Page 15: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

The mineral oil in cyclohexane calibration plot of actual (x-axis) vs. predicted (y-axis) values. The values displayed are the final concentrations of oil in water based on the ASTM D7678 parameters (900mL water, 20mL cyclohexane).

Oil Calibration SetStandard Name Oil (mg/L)OIW Soln A 0.00OIW Soln B 0.05OIW Soln C 0.15OIW Soln D 0.26OIW Soln E 0.84OIW Soln F 1.70OIW Soln G 2.54OIW Soln H 4.20OIW Soln I 8.30OIW Soln J 16.54OIW Soln K 24.90OIW Soln L 32.55

Page 16: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

OIW Validation Results – Samples are 9.3mg/L (solution A) and 1.4mg/L (solution B) mineral oil respectively

Solution A Solution B9.70 1.349.43 1.549.12 1.449.60 1.46

Mean Value 9.46 1.45Std. Deviation 0.25 0.08Relative Std. Dev. (%) 2.69 5.69% Recovery 101.7 103.2

TPH

The samples are measured in quadruplicate with the Dialpath™ accessory.

Page 17: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

The automatically obtained data results from the MicroLab software for the Agilent Cary 630, 4500series, and 5500series FTIR spectrometers.

Page 18: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Using Mobile FTIR to Measure Gasoline in Diesel

The intensity of an infrared absorbance band is proportional to the concentration of that component in a mixture, as stated in Beer’s Law. This relationship allows the Agilent 5500t and 4500t FTIR spectrometers to accurately measure gasoline in diesel fuels.

September 6, 2012 Confidentiality Label

18

4500t FTIR Series

5500t FTIR Series

Page 19: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Chemical Differences Between Gasoline and Diesel

Figure 1. The IR spectral overlay of gasoline (Blue) and diesel fuel (Red) using the Agilent 5500t FTIR spectrometer, 100 μm pathlength

The chemical differences between gasoline and diesel can be seen in the infrared spectrum of each. Below is a comparison between the infrared spectra of gasoline and diesel highlighting the ethanol and aromatics in gasoline. Neither ethanol or light aromatics present in gasoline are observed in the diesel fuel spectrum; in fact, the spectrum of diesel fuel is similar to mineral oil.

Page 20: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

The Sensitivity of Mobile FTIR

The sensitivity of the 5500t FTIR allows gasoline to be measured down to 0.025 % in diesel. To demonstrate this, several concentrations of gasoline (87 octane) are carefully prepared in ultra low sulfur diesel (US, Danbury CT). The samples were prepared with 0 %, 0.0269 %, 0.2669 %, and 1.0586 % gasoline in diesel. The FTIR spectra were measured on the 5500t spectrometer and the gasoline absorbance results are plotted against their concentrations. This gasoline absorbance plot indicates a very good linear correlation with concentration.

Page 21: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Figure 2. The IR absorbance vs. concentration plot for gasoline in diesel, Agilent 5500t FTIR 100 um pathlength

Gasoline absorbance results are plotted against their concentrations

Page 22: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

3 Minute Mobile FTIR Analysis

Linear correlation is common in spectroscopy and can be easily added to 5500 or 4500 FTIR methods. Multiple components can be reported from a single 3 minute analysis, such as gasoline in diesel, biodiesel in diesel, oxidation, and water.

The instrument can be operated from a laptop (5500t) or the Agilent 4500t FTIR which is a fully field portable version with an onboard battery and operated from a hand held computer (PDA). The instrument software is simple to use with little to no sample preparation.

September 6, 2012 Confidentiality Label

22

Red or Green results will tell the user if there is critical warning and recommend the next steps

Page 23: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Agilent Private

Advancement in Mobile Solutions – Taking On-Site Measurements is Now Easier Than Ever

21 June 2012 23

Page 24: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Value of Moving Measurements From the Lab to the Field Fast answers Define sampling strategy based on current

results Triage - Send more relevant samples back to

lab. Decide which areas need more investigation.

Non-destructive analysis of large objects Too big for lab Too expensive to disassemble

Screen incoming materials before they enter the production process

On-site adulteration or contamination determination

September 6, 2012

Confidentiality Label 24

Page 25: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Instrument Requirements Needed for Field and Routine Use Effective performance Achieve required measurement limits “Spectroscopic performance still matters”

Form Factor – Compact, rugged, portable, versatile System needs to go where the sample is Frequent travel subjects it to ↑ shock and

vibration Correct sampling interfaces Little or no sample prep Multiple uses and matrices

Ease of use and provide answers Easy to understand, meaningful results

25

Page 26: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

What approach does Agilent Technologies take when designing systems for field use?

26

Develop high performance optical engine

Integrate engine into appropriate form factors

4100 Exoscan / 4200 Flexscan – handheld Cary 630 – routine lab

Lab systems benefit -> compact size and robustness

Field systems benefit -> spectroscopic performance

Easy method transfer from development lab to field

Cary 630 routine benchtop 4100 handheld

4100 in lab configuration

4200 FlexScan

Page 27: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Dual Nature of the Agilent 4100 ExoScan

Method development for field applications is often accomplished in the lab

Exoscan was designed to specifically meet this need Docking Station Convert handheld system into bench-top system PC connection Sample stages

Interchangeable sample interfaces

Increased flexibility

27

Page 28: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Instrument Evolution

28

Laboratory (1970’s)

Spectroscopy Measurements Continue to Move Directly to the Sample

Current (Today)

Page 29: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Field FTIR Form Factor

4 lbs optical head Optical Enclosure wire connected to Electronics PDA Controlled Dedicated sample interfaces

29

4100 ExoScan FTIR 4200 FlexScan FTIR

7 lbs 7” x 4.7” x 9” inches Single Enclosure (optics & electronics) PDA Controlled Interchangeable sample interfaces

Page 30: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Field Sample Interfaces ATR

Solids or liquids Easy sample identification Diamond and Germanium available

External Reflectance (45⁰)

Metal and composite surfaces Specular reflectance or reflection absorption Surface oxidation, coating thickness and chemistry

Grazing Angle (~82⁰)

Smooth metal surfaces Enhanced absorption Trace contaminant identification

Diffuse Reflectance

Normal incidence Large collection angle Low reflective samples

30

Page 31: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Identifying applications that can be migrated to the field Does the measurement involves chemical differences? Change in concentration, thickness, oxidation, chemical damage, UV

degradation, presence or absence of chemical, etc?

Is the measurement made by FTIR in the lab?

Do you have access to the sample? Not behind glass or plastic

Is there an advantage to field measurement? Speed or non-destructive

Actionable limit ATR ~ 1 % Diffuse Reflectance ~ 0.01 – 0.1% External Reflectance ~ 0.05 mils or ~ 2 μg/cm2

Grazing Angle ~ 0.2 μg/cm2

31

Page 32: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Considerations for Handheld Measurements

Sample Interface Substrate Required limit Ability to measure handheld

Diffuse or external reflectance easier than ATR Diffuse most forgiving WRT focus

Measurement Time ↑ Time = ↓ Noise…. however How long can the user keep the system steady

Especially true for ATR Sometimes shorter = better measurement.

Meaningful Results Must be useable No interpretation required

32

Page 33: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Considerations for Identifying “Good” Data Reflectance techniques can produce negative or totally absorbing bands Often due to strongly absorbing samples No sample preparation

Useful data can often be found outside these regions Example: Sulfate concentration in soil High percentage of sand = strongly absorbing Si-O bands Overlap sulfate absorbance Overtone of sulfate at 2200 cm-1 correlates well with sulfate concentration

33

Page 34: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

ATR’s ease of use SHOULD make it ideal for hand held use No sample preparation High signal to noise Consistent path length High acceptance in lab

BUT one feature makes hand held use difficult -> CONTACT

Crystal/sample contact is the key - > Flatness not pressure In Lab, press brings crystal and sample in line In Field, contact is made with hand movement – hard to keep steady

Crystal Geometry is the ANSWER

Spherical surface makes good contact at many angles Strong absorbance is observed even with a light touch

34

Sample Interface Considerations - ATR

Page 35: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

How are methods developed and what types of methods are useful?

Useable results require intelligent output Methods developed for specific applications Most methods are universal Only developed once Usable on all instruments of the same type

Qualitative Methods Standard library search techniques Smaller libraries = focused results

Quantitative methods Meaningful units to the end user Threshold limited and color coded Often Multivariate methods

35

Page 36: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Implementing methods into end user software

Software guides the user through the

selected method and gives users

instructions to prevent mistakes.

Agilent Confidential

Page 37: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Displaying Meaningful Results

Library results are shown with the hit list

Quant results can be programmed for action levels with color codes

(red, green, yellow)

Page 38: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Exoscan can identify all 10 groups of seals Fluorosilicone, Silicone, Viton, EPR/EPDM, Neoprene, Butyl, Kalrez, NBR, Polyurethane, Natural Rubber

Easy Identification by spectral search Similarity = correlation Highest correlation = correct ID

Demonstration 15 samples Exoscan correctly identified all samples Only one sample had a close second match

38

Hand Held FTIR Material Identification O-ring and Seal Identification

Page 39: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Hand Held FTIR Coating Analysis

Sensitivity of FTIR is ideal for thin coatings - Primers, adhesives, protective coats

Ideal thickness required - Cost and performance

Many types of surfaces

Two examples: - Epoxy primer - Ultra thin anodization

39

Page 40: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Epoxy paint primer on aluminum

Ideal thickness ~ 0.2 mils

Range from 0.5 to 5.0 mils

Linear correlation to epoxy band at 1610 cm-1

Upper Limit of Quant = 0.6 mils R2 = 0.99

40

Epoxy Primer Thickness External Reflectance

Page 41: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Ultra Thin Anodization Coating Thickness Grazing Angle

41 1

18

7.7

92

5.2

90

8.7

front 00_2009-05-08t17-12-10(1)

12

64

.0

11

10

.9

1500 1450 1400 1350 1300 1250 1200 1150 1100 1050 1000 950 900 850 800 750 700

0.16

0.14

0.12

0.10

0.08

0.06

0.04

0.02

0.00

-0.02

-0.04

-0.06

-0.08

W b

Abs

orba

nce

Metal Oxide Peak 1Metal OxidePeak 2

11

87

.7

92

5.2

90

8.7

front 00_2009-05-08t17-12-10(1)

12

64

.0

11

10

.9

1500 1450 1400 1350 1300 1250 1200 1150 1100 1050 1000 950 900 850 800 750 700

0.16

0.14

0.12

0.10

0.08

0.06

0.04

0.02

0.00

-0.02

-0.04

-0.06

-0.08

W b

Abs

orba

nce

Metal Oxide Peak 1Metal OxidePeak 21 µm

0.5 µm

50 nm

blank

Anodization on aluminum Thickness crucial for further coating Exoscan – Grazing angle ~10 second measurement time LOD ~ 25 nm thickness Action Level = 0.5 μm

Page 42: Innovations in FTIR Technology for the Rapid Analysis … · Innovations in FTIR Technology for the Rapid Analysis of Liquid Petrochemical Samples Lou Tisinger FTIR/FTIR Imaging and

Conclusions

Instruments are moving out of the lab to give users cost and measurement advantages

As they move out, design philosophies must accommodate new requirements

Size, weight, durability and usability Given the right design, instruments can perform as well as many lab systems

The realities of handheld measurement require method developers to consider the entire measurement process.

4100 ExoScan development supports both field and method development Well developed methods can provide easy data acquisition and actionable

answers even on difficult samples.