Inhalation and Humidification
Transcript of Inhalation and Humidification
-
7/27/2019 Inhalation and Humidification
1/10
632
Indian Journal of Anaesthesia, October 2008(P.G.Issue)
Inhalation therapy and Humidification
V Saraswat
Summary
Inhalation therapy is a preferred mode of drug delivery as it provides rapid onset of action, increased bioavailability,
low gastrointestinal side effects and user-friendly environment with better patient compliance. A variety of drugs are
in use and more drugs formulations are being investigated, consequently leading to development of compatible inhal-
ers, which necessitates good understanding of drug formulations and functioning of inhalers. There is no significant
difference between various inhaler devices in their efficacy outcome, however most are drug specific depending on
drug formulations. Multi-drug inhalers are in developmental phase and may be available in future. Oxygen therapy
being the mainstay of inhalation therapy needs special emphasis especially with regards to delivery devices. Climatisation
of inspired gases, both in terms of humidification and temperature maintenance, is the integral part of inhalation
therapy especially so if upper airway is bypassed by intubation or tracheostomy. The problem is compounded incritically ill patients on long-term ventilation. Both active and passive humidifiers are in use, heated humidifiers (HH)
and Heat and moisture exchanger (HME), being most common in use at present. Both, HH and HME, have proved
their efficacy in varying patient requirements, however their use may be guided more by clinical and cost consider-
ations. Use of microbial filters in humidifiers, in some studies, have shown a significant reduction in ventilator-
associated pneumonia (VAP).
Key words Nebulizer, Metered dose inhaler (MDI), Dry powder inhaler (DPI), Oxygen delivery
devices, Heated humidifier (HH), Heat and moisture exchanger (HME),
ventilator-associated pneumonia (VAP)
1.DPMO (A), HQ, WAC, Correspondence to: V Saraswat Subroto Park, New Delhi 110010,
Email: [email protected], [email protected]
Introduction
Inhalation therapy is a term used for a variety of
treatment techniques, including drugs administered via
inhalation. It aims at targeting lung tissue, airway secre-
tion and microorganisms in upper, central and/or pe-
ripheral airways. However, drugs targeting systemic
effects are used aiming at deposition in the alveoli where
it can be rapidly absorbed and distributed.1
Benefits of inhalation therapy
1. A principal benefit of inhalation therapy is the
rapid onset of action, compared to peroral medication
from large absorption area of the lung. Locally acting
drugs have immediate onset of action whereas, sys-
temically active reach the blood stream within seconds.
It is important in rescue medications, pain medication
and time-sensitive therapies, such as insulin.
2. Inhaled drugs avoid GI tract and are not sub-jected to the first-pass metabolism that reduces
bioavailability, avoid undesirable GIT side effects and
are better tolerated.
3. Injected drugs also avoid the problems asso-
ciated with the GI tract, but needles are invasive and
socially uncomfortable to effect self injection. Inhaled
drugs are more user-friendly and result in better patient
compliance.
Drugs used in inhalation therapy
1. Oxygen therapy - The most common form of
inhalation therapy is the provision of oxygen. There are
a multitude of reasons for administering oxygen, rang-
ing from minor to very serious breathing problems like
decreased lung function and systemic organ failures.
The treatments range from at-home oxygen therapy for
Indian Journal ofAnaesthesia 2008;52:Suppl (5):632-641
-
7/27/2019 Inhalation and Humidification
2/10
633
patients with chronic obstructive pulmonary disease
(COPD) to mechanical ventilation for patients with res-
piratory failure.
2. Anaesthetic agents - Anaesthetic agents have
been used by inhalation since the beginning of practice
of anaesthesia. The anaesthetic gases from nitrous ox-
ide to inert gases like helium have been used. Other
agents used by this route in ICU are nitric oxide and
surfactant. In addition, in absence of venous access,
most of the life saving drugs required for immediate
effect may be given endotracheally in ICU.
3. Beta2 agonists - Beta2 agonists arebronchodilators, extensively used in treatment of asthma.
Albuterol, is short-acting and used for acute exacerba-
tions. It relaxes smooth muscle and increases airflow
within 5 minutes and last 3-6 hours. Salmeterol is a
long-acting with slower onset (15-30 minutes), but
longer duration of action (over 12 hours). Inhalers are
colour coded and deliver fixed dose at actuation.
4. Glucocorticoids - Inhaled glucocorticoids are
the most potent inhaled anti-inflammatory drugs cur-
rently available. Early intervention improves asthma
control, normalizes lung function and prevents irrevers-
ible airway damage. Improvement can occur within 24
hours, although maximum benefit may take 1 to 2
weeks. The risks of systemic side effects, such as sup-
pression of the hypothalamic-pituitary axis, are less than
with oral glucocorticoids. Inhaled glucocorticoids in-
clude fluticasone, beclomethasone, budesonide, and
triamcinolone.
5. Anticholinergic drugs - Ipratropium bromide(Atrovent) is a quaternary derivative of atropine that
lacks its adverse side effects. It is used as an additional
reliever medication when short-acting beta2agonists do
not give enough relief. Its anticholinergic action also
decreases mucous secretions.
6. Other drugs Antibiotics, insulin, contracep-
tives and analgesics have been tried and some are in
developmental stage.
DEVICES FOR INHALATION THERAPY
1. Oxygen delivery devices
A wide variety of oxygen delivery systems are
available. (Fig-1) The mask and valve design and oxy-
gen flow rate allows delivery of inspired oxygen of 24-
90% (FiO20.26-0.90). The concentration of oxygen
depends on the ventilatory minute volume (MV) and
the flow rate of oxygen, greater the ventilation, the lower
the FiO2for a given flow rate of oxygen. It is impos-
sible to provide a fixed FiO2to a patient with a varying
ventilatory requirement unless the total MV is provided
at the required FiO2.2 There are two basic types of
oxygen mask which deliver either the entire (high flow
mask) or a proportion (low flow mask) of the ventila-
tory requirement.
Fig 1 Oxygen mask and nasal canula
Nasal prongs are simple, convenient to use and
can provide 24-40% oxygen with flow rates up to 6 L/
min in adults. The FiO2depends on the flow rate of
oxygen (1-6L/min) and varies according to MV. At anoxygen flow rate of 2L/min the oxygen concentration
in the hypopharynx is 25-30%. Nasal prongs prevent
rebreathing, are comfortable for long periods, and al-
low oxygen to be continued during talking and eating.
Local irritation and dermatitis may occur with high flow
rates. Nasal prongs are less prone to complications,
and oxygenation in children is equally effective.3 Neo-
nates and small babies can be given oxygen by a tent
or oxygen hood.
V Saraswat. Inhalation therapy and humidification
-
7/27/2019 Inhalation and Humidification
3/10
634
Indian Journal of Anaesthesia, October 2008(P.G.Issue)
Low flow masks Simple oxygen masks can
provide 35-50% FiO2, depending on fit, at flow rates
from 5-10 L/min. Flow rates should be maintained at 5L/min or more in order to avoid rebreathing exhaled
CO2that can be retained in the mask. A concentration
of up to 60% can be achieved with addition of reser-
voir bag, and are used mainly in type I respiratory fail-
ure. These masks are generally not suitable for patients
with type II respiratory failure.
High flow, jet mixing masksdeliver about 40L/min of gas through the mask, sufficient to meet the total
respiratory demand unaffected by the pattern of venti-
lation and avoid rebreathing as mask is flushed by thehigh flow rates. The masks contain venturi valves, which
use the principle of jet mixing (Bernoulli effect). When
oxygen passes through a narrow orifice it produces a
high velocity stream that draws a constant proportion
of room air through the base of the venturi valve. Air
entrainment depends on the velocity of the jet (the size
of orifice and oxygen flow rate) and the size of the valve
ports. It can be controlled to give inspired oxygen lev-
els of 24-60%. In COPD and type II respiratory fail-
ure patients, these masks reduce the risk of carbon di-oxide retention while improving hypoxemia. Jet-mixing
masks rated at 35% or higher usually do not deliver
flow rates adequate to meet the inspiratory flow rates
of adults in respiratory distress. They are loose fitting
and comfortable to wear. The OxyMask, a new face
mask for oxygen delivery that uses a small diffuser to
concentrate and direct oxygen toward the mouth and
nose delivers oxygen more efficiently than a Venturi
mask.4 Aerosol masks, tracheostomy collars, T-tube
adapters, and face tents can be used with high-flow
supplemental oxygen systems.
Rebreathing and anaesthetic type oxygen
masksPartial rebreathing masks incorporating non-
rebreathing valves and reservoir bags can provide con-
centrations 40-70% at flow rates of 6-10 L/min. Oxy-
gen flow should always be supplied to maintain the res-
ervoir bag at least one third to one half full on inspira-
tion. In cardiac or respiratory arrest, tight fitting anaes-
thetic-type masks can achieve 100% oxygen. The non-
rebreathing mask is similar to the partial rebreathing
mask except it has a series of one-way valves. One
valve is placed between the bag and the mask to pre-vent exhaled air from returning to the bag. There should
be a minimum flow of 10 L/min to deliver FiO2of 60-
80%.
Non-invasive assisted ventilationSupplemental
oxygen may be provided through tight fitting nasal or
full-face masks during nasal intermittent positive pres-
sure ventilation and continuous positive airways pres-
sure. (Fig-2) It is used to support ventilation in sleep-
associated hypoventilation, during weaning from me-
chanical ventilation, and in respiratory failure associ-ated with COPD. The technique improves oxygenation
by recruiting under ventilated alveoli. As intubation is
avoided the risks of nosocomial pneumonia are re-
duced. The CPAP mask often becomes uncomfortable
and gastric distension may occur, hence patients must
be cooperative, able to protect their airway, have the
strength to breathe spontaneously and cough effectively,
failing which mechanical ventilation can be instituted.
The type of interface affects the NIMV outcome more
than the ventilatory mode.5
Fig 2 Mask for noninvasive ventilation (NIV)
Hyperbaric oxygen therapy (HBO), delivers pure
oxygen under pressure equal to two to three times nor-
mal atmospheric pressure. At a pressure of 300 kPa
the small quantity of oxygen in solution in the blood can
be increased by up to 300% and diffusion through tis-
sues may be improved.2 This treatment has been espe-
cially effective in high altitude pulmonary edema and
-
7/27/2019 Inhalation and Humidification
4/10
635
scuba divers. The patient enters the chamber and oxy-
gen is delivered under forced pressure. The therapy
usually lasts one hour, although it can take up to fivehours in serious decompression cases. Before exiting
the chamber, the pressure will eventually be lowered to
normal atmospheric level.
2.Devices for inhaled medications
There are four different kinds of inhalation de-
vices:
(a)Nebulizer systems - Nebulizer changes liq-
uid medicine into fine droplets (aerosol or mist) thatare inhaled through a mouthpiece or mask. (Fig-3) It
can be used to deliver saline, distilled water or medi-
cines in liquid formulation. It is based on one of three
different techniques - jet, ultrasonic and vibrating mem-
brane. The device may be heated or unheated. Choos-
ing the optimal device for a patient is essential to en-
sure the effectiveness of the therapy. However, the
choice is dependent on several different factors and
there is no single device that is best for all patients in all
situations. Inhalation technique and handling of the de-vice is essential for the effective treatment.
age group. Spacers decrease the amount of drug de-
posited in the oropharynx (up to 80% of the actuated
dose with the MDI alone), thereby reducing systemicand other side effects. The drugs currently available in
MDI formulations include beta agonists, glucocorticoids,
ipratropium bromide, cromolyn sodium, and nedo-
cromil. Each product delivers a set amount of drug-
per actuation (puff). In the United States, MDIs are
labeled according to the amount of drug delivered at
the mouthpiece, whereas in Canada and the European
Union, they are labeled according to the amount of drug
delivered from the valve.
MDIs use chlorofluorocarbons (CFCs) as thepropellant. New devices, propelled by Hydrof-
luoroalkane (HFA) have been developed and are cur-
rently available with different medicines. The Clean Air
Act and an international environmental treaty have led
to the phase out of CFC-propelled inhalers by De-
cember 31, 2008 as they deplete the ozone layer.
(c) Dry powder inhaler (DPI) - Special inhaler
are in use to deliver the dry powder medicines. (Fig-4)
When inhaled fast enough the medicine is released. DPIwhen compared with MDI are as efficient and may be
superior delivery device for some drugs.6
Fig 3 Nebulizer
(b) Metered dose inhaler (MDI) - It consists
of a pressurized canister of medication in a plastic case
with a mouthpiece. Pressing the MDI releases a mist of
medication. It is portable, efficient and convenient and
are designed for actuation during a slow, deep inhala-
tion for optimal lung delivery. The addition of a spacer
improves drug delivery and can be used in paediatric
Fig 4 Dry powder inhaler (DPI)
(d) Monodispersed aerosols with a very nar-
row range of smaller particle sizes target drug delivery
to specific areas of the lung where it is most effective.
However, as smaller particles are more easily absorbed
into the pulmonary circulation they may be associated
with a higher incidence of systemic side effects. The
optimal particle size required to maximize the thera-
peutic ratio of a molecule may be different for a 2-
agonist and inhaled corticosteroid.
V Saraswat. Inhalation therapy and humidification
-
7/27/2019 Inhalation and Humidification
5/10
636
Indian Journal of Anaesthesia, October 2008(P.G.Issue)
Effect of particle size
The destination of aerosol particles is critical to
the efficacy of inhalation therapy. For locally-acting
drugs, the particles need to be deposited in the area of
the respiratory tract requiring treatment. For systemi-
cally-acting drugs, particles need to reach the alveoli
for absorption.
Large particles (5-10 m) do not follow changes
in the direction of airflow and are deposited by inertial
impact in the upper airways. Particles deposited in the
mouth and throat can be swallowed and lead to local
or systemic side effects. Intermediate-sized particles(3-5 m) can be carried further, into smaller airways of
the bronchi and bronchioles. Small particles (
-
7/27/2019 Inhalation and Humidification
6/10
637
America, oxygen is not humidified as long as flow is
less than 4L/min, as per guidelines for oxygen therapy
announced by the ACCP-NHLBI in 1984 and revisedby AARC in 2002. The oxygen received through a
nasal cannula at 0.5-4 L/min or through a Venturi mask
at 24-40% constitutes only a small percentage of the
patients inspiratory tidal volume compared with the
water content delivered from the airway. Increasing the
relative humidity of the room air easily compensates
this humidity deficit, hence routine humidification of low-
flow oxygen or low-concentration oxygen is not justifi-
able.15
Noninvasive positive-pressure ventilation deliversair at a high flow, with a low relative humidity, at high
inspiratory pressure. Addition of a heated humidifier
increases the relative and absolute humidity to levels
acceptable for nonintubated patients, with minimal ef-
fect on delivered pressure.16 Heated humidifiers have
been found to be more effective than helmet17 and oily
nose drops while delivering CPAP18 and improves ab-
solute humidity in nasal CPAP therapy with nasal mask
leak making it more acceptable.19
Ventilation with endotracheal intubation bypasses
the upper airway and the normal heat and moisture
exchanging process of inspired gases. A continuous loss
of moisture and heat occurs and predisposes patients
to serious airway damage. Ventilation with dry and cold
gases leads to epithelial cell disorders, increased mu-
cus viscosity and restriction of the mucociliary function
resulting in hypothermia, atelectasis and hypoxaemia.
On the other hand, over-humidification or ventilation
with hyperthermic gases may lead to tracheal burning,
alteration in surfactant and epithelial cell disorders lead-
ing to hyperthermia, hyponatraemia, atelectasis and
hypoxaemia. Thus, it is of crucial importance to moni-
tor ventilatory gas conditioning closely, especially in ICU
patients subjected to prolonged mechanical ventilation.
Currently heated humidifying systems (HH) (ac-
tive) (Fig-5) and heat and moisture exchange filter
(HMEF) (passive) (Fig-6) are in common use in ana-
esthesia and intensive care units (ICU).
Fig 5 Heated humidifier (HH)
Fig 6 Heat and moisture exchanger filter (HMEF)
During general anaesthesia gas climate is signifi-
cantly improved by low flow techniques with fresh gas
flows equal or less than 1 L/min. With higher FGF ratesthe humidity decreases markedly while high gas tem-
peratures are maintained. Gas climatisation is also in-
fluenced by the temperature loss and water condensa-
tion within the hoses, factors that are related to the tech-
nical design and material of the patient hose system. In
routine clinical practice conventional and coaxial hose
systems are similar in respect to the climatisation of
breathing gases and acceptable, but not optimal. The
coaxial hose system lead to improved climatisation in
long lasting procedures only. Whereas, heated breath-
ing hoses perform markedly better and sufficient mois-
ture and temperature values can be achieved.20 The
HMEs improve the inspiratory AH, RH and tempera-
ture of the anaesthetic gases during different FGFs but
do not prevent drop in body temperature.21Unheated
wire circuits perform better than heated circuits in terms
of delivered relative humidity.22
Heated humidifier (HH) Vs heat and
moisture exchanger (HME)
V Saraswat. Inhalation therapy and humidification
-
7/27/2019 Inhalation and Humidification
7/10
638
Indian Journal of Anaesthesia, October 2008(P.G.Issue)
HH are considered better in terms of preserva-
tion of relative humidity and temperature than HME,23
however, modern HMEs are able to maintain physi-ological air-conditioning even in long-term ventilated
patients with less labour and costs per day.24 In spon-
taneously breathing, tracheostomized ICU patients also
both, HH and HMEs perform satisfactorily in preser-
vation of heat and humidity of inspired gases.25 HMEs
of different manufacturers may differ in their perfor-
mance,26 but different composition of condensation
surface or filter membrane does not seem to have much
effect on their performance.27 The use of HMEs in neo-
nates can achieve an absolute humidity of 28 mgH2
O/L
and atemperature of 30 C or more, but higher values
are achieved using aHH.28
There does not seem to have a consensus on how
frequently HME filters should be changed. Since using
the HME for a longer time may have effect on its per-
formance as well as colonization of bacteria, changing
the HME every 24 hours has been advocated. Earlier
studies suggested that changing HME filters after 48
hours did not affect its performance and maintained heat
and humidity.29,30
Other studies reported satisfactoryperformance after three days to one week31-33 How-
ever, all these studies reported some indirect evidence
of very little, if any, change in HME resistance.
In a recently published study, four electrostatic
HMEF were assessed on a bench-test in conformity
with the ISO 9360: 2000 standard. They showed dif-
ferences between the HMEFs for both humidification
and resistance parameters. When compared to the new
version of the standards, HMEFs demonstrated their
reliability, however, evolution of humidification and flow
resistance characteristics over 24h showed a structural
degradation, limiting their use over a longer period.34
The type of humidification device may negatively
influence the mechanical efficacy of ventilation. HMEs
have been noted to have small increase in airway resis-
tance and dead space25,35 that may alter the alveolar
ventilation and may play a crucial role in difficult to wean
patients and unless the pressure support level is con-
siderably increased, the use of a HME should not be
recommended in difficult or potentially difficult to wean
patients with chronic respiratory failure.36
It may alsoprovide some PEEP effect during spontaneous ventila-
tion.35 Some studies have reported endotracheal tube
occlusion caused by inadequate airway humidification
with use of HMEs in ICU patients.11,37-39
There is major concern regarding the type of
humidification system on the rate of ventilator-associ-
ated pneumonia (VAP). Various studies have failed to
record the colonization of HMEs with microbial after
varying period of ventilation nor different kind of HMEs
have shown any change in tracheal colonization orVAP.27,30-32 Two meta-analyses have suggested an as-
sociation between the use of HME and a decreased
VAP rate40,41
There has been equivocal evidence in respect to
choice of humidifiers in relation to reducing the inci-
dence of VAP, some favoring HME42,43 while others
reporting lower incidence from HH in patients requir-
ing long-term respirator therapy.44 Both HHand HMEs
can be used with no
significant impact on the incidenceof VAP and other criteria may justify their choice.45This
was summarized well by Kola et al in a meta-analysis,
who found a significant reduction in the incidence of
VAP in patients humidified with HMEs during MV, par-
ticularly in patients ventilated for 7days or longer. How-
ever, this finding is limited by the exclusion of patients
at high risk for airway occlusion from some of the stud-
ies. Moreover, contraindications (tenacious secretions,
airway obstructive disease, hypothermia) and techni-
cal issues of HMEs must be considered.41
Two mechanisms have been postulated in devel-
opment of VAPs; the inclusion of microbiological gas
filter in the HME, which protects the patient from ex-
ogenous VAP; and reduced contaminated condensate
in the HME circuit. Previous studies evaluating the ef-
fect of gas filtration in anaesthesia machines46,47 and in
ventilators48were unable to demonstrate differences in
the incidence of VAP between the patient groups with
and without filters. The contaminated condensate can
-
7/27/2019 Inhalation and Humidification
8/10
639
be reduced by changing HMEs at an interval of 48 hours
or less and improvement of design of HH (servo-con-
trolled with dual-heated circuit and an autofeed cham-ber). Lower incidence of tube occlusion,11,37,38,49 thick
bronchial secretions and atelectasis 11,23,37,50have been
reported with HH than patients with HMEs.
References
1. Lannefors L, Inhalation therapy: practical consider-
ations for nebulisation therapy. Physical Therapy Re-
views 2006 ;11: 21-27.
2. Bateman N T and Leach R M. ABC of Oxygen: Acute
oxygen therapy. Can Respir J 2006 ;13:247-52.3. Weber MW, Palmer A, Oparaugo A, Mulholland EK.
Comparison of nasal prongs and nasopharyngeal cath-
eter for the delivery of oxygen in children with hypox-
emia because of a lower respiratory tract infection. J
Pediatr 1995 ;127:378-83.
4. Beecroft JM, Hanly PJ. Comparison of the OxyMask
and Venturi mask in the delivery of supplemental oxy-
gen: pilot study in oxygen-dependent patients. BMJ
1998;317:798801.
5. Navalesi P, Fanfulla F, Frigerio P, Gregoretti C, Nava S.
Physiologic evaluation of noninvasive mechanical ven-
tilation delivered with three types of masks in patientswith chronic hypercapnic respiratory failure. Crit Care
Med 2000 ;28:1785-90.
6. Lotvall J. Inhalation therapy of the future - How will it
change the way we treat asthma? Journal of Aerosol
Medicine. 2001;14(supplement 1): 45-50.
7. Dolovich MB, Richard C. Ahrens, et al. Device selec-
tion and outcomes of aerosol therapy: Evidence-Based
Guidelines. Chest 2005;127:335-371.
8. Cole P. Respiratory mucosal vascular responses, air
conditioning and thermoregulation. J Laryngeal Tool
1954;68:613622.9. Shelly MP. Editorial: The upper airway - the forgotten
organ. Crit Care 2001; 5: 12.
10. Shelly MP. Conditioning of inspired gases in physiologi-
cal basis of ventilatory support. In Physiological Basis
of Ventilatory Support. Edited by Marini JJ, Slutsky AS
New York: Marcel Decker Inc; 1998. pp. 575599.
11. Cohen IL, Weinberg PF, Fein IA, Rowinski GS. Endotra-
cheal tube occlusion associated with the use of heat
and moisture exchangers in the intensive care unit. Crit
Care Med 1988;16:277279.
12. Dery R, Pelletier J , Jacques A. Humidity in
anaesthesiology. Heat and moisture patterns in the res-
piratory tract during anaesthesia with the semi-closed
system. Can Anaesth Soc J 1967;14:287294.
13. McFadden ER, Pichurko BM, Bowman HF. Thermal
mapping of the airways in humans. J Appl Physiol
1985;58:564570.
14. Weber MW, Palmer A, Jaffar S, Mulholland EK. Humidi-
fication of oxygen with unheated humidifiers in tropical
climates. Pediatr Pulmonol 1996 ;22:125-8.
15. Miyamoto K. Is it necessary to humidify inhaled low-
flow oxygen or low-concentration oxygen? Nihon
Kokyuki Gakkai Zasshi 2004 ;42:138-44.
16. Holland AE, Denehy L, Buchan CA, Wilson JW. Effi-
cacy of a heated passover humidifier duringnoninvasive ventilation: a bench study. Respir Care
2007;52:38-44.
17. Chiumello D, Chierichetti M, Tallarini F, Cozzi P, et al.
Effect of a heated humidifier during continuous posi-
tive airway pressure delivered by a helmet. Crit Care
2008;21;12:R55 [Epub ahead of print].
18. Wiest GH, Lehnert G, Brck WM, Meyer M, Hahn EG,
Ficker JH. A heated humidifier reduces upper airway
dryness during continuous positive airway pressure
therapy. Respir Med 1999;93:21-6.
19. Fischer Y, Keck T, Leiacker R, Rozsasi A, Rettinger G,
Gruen PM. Effects of nasal mask leak and heated hu-
midification on nasal mucosa in the therapy with nasal
continuous positive airway pressure (nCPAP). Sleep
Breath 2008 Mar 1[Epub ahead of print].
20. Baum J, Zchner K, Hlscher U, Sievert B, Stanke HG,
Gruchmann T, Rathgeber J. Climatization of anesthetic
gases using different breathing hose systems.
Anaesthesist 2000;49:402-11.
21. Johansson A, Lundberg D, Luttropp HH. The effect of
heat and moisture exchanger on humidity and body tem-
perature in a low-flow anaesthesia system. Acta
Anaesthesiol Scand 2003;47:564-8.22. Gilmour IJ, Boyle MJ, Rozenberg A, Palahniuk RJ. The
effect of heated wire circuits on humidification of in-
spired gases. Anesth Analg 1994;79:160-4.
23. Luchetti M, Stuani A, Castelli G, Marraro G. Comparison
of three different humidification systems during pro-
longed mechanical ventilation. Minerva Anestesiol 1998
; 64:75-81.
24. Rathgeber J, Henze D, Zchner K. Air conditioning with
a high-performance HME (heat and moisture ex-
changer)an effective and economical alternative to
V Saraswat. Inhalation therapy and humidification
-
7/27/2019 Inhalation and Humidification
9/10
640
Indian Journal of Anaesthesia, October 2008(P.G.Issue)
active humidifiers in ventilated patients. A prospective
and randomized clinical study. Anaesthesist 1996 ;
45:518-25.
25. Thomachot L, Viviand X, Arnaud S, Vialet R, , Martin C.
Preservation of humidity and heat of respiratory gases
in spontaneously breathing, tracheostomized patients.
Acta Anaesthesiol Scand 1998;42:841-4.
26. Martin C, Papazian L, Perrin G, Saux P, Gouin F. Preser-
vation of humidity and heat of respiratory gases in pa-
tients with a minute ventilation greater than 10 L/min.
Crit Care Med 1994;22:1871-6.
27. Thomachot L, Vialet R, Arnaud S, Barberon B, Michel-
Nguyen A, Martin C. Do the components of heat and
moisture exchanger filters affect their humidifying effi-
cacy and the incidence of nosocomial pneumonia? CritCare Med 1999 ;27:923-8.
28. Fassassi M, Michel F, Thomachot L, Nicaise C, Vialet R,
Jammes Y, Lagier P, Martin C. Airway humidification with
a heat and moisture exchanger in mechanically venti-
lated neonates : a preliminary evaluation. Intensive Care
Med 2007;33:336343.
29. Thomachot L, Vialet R, Viguier JM, Sidier B, Roulier P,
Martin C. Efficacy of heat and moisture exchangers af-
ter changing every 48 hours rather than 24 hours. Crit
Care Med 1998;26:477-81.
30. Boisson C, Viviand X, Arnaud S, Thomachot L, Miliani
Y, Martin C. Changing a hydrophobic heat and moistureexchanger after 48 hours rather than 24 hours: a clinical
and microbiological evaluation. Intensive Care Med
1999;25:1237-43.
31. Thomachot L, Boisson C, Arnaud S, Michelet P, Cambon
S, Martin C. Changing heat and moisture exchangers af-
ter 96 hours rather than after 24 hours: a clinical and mi-
crobiological evaluation. Crit Care Med 2000 ; 28:714-20.
32. Davis K Jr, Evans SL, Campbell RS, et al. Prolonged use
of heat and moisture exchangers does not affect device
efficiency or frequency rate of nosocomial pneumonia.
Crit Care Med 2000;28:1412-8.
33. Ricard JD, Mire EL, Markowicz P, et al. Efficiency and
safety of mechanical ventilation with a heat and mois-
ture exchanger changed only once a week. J Respir Crit
Care Med 2000;161:104-109
34. Lannoy D, Dcaudin B, Resibois JP, et al. Humidifica-
tion assessment of four heat and moisture exchanger
filters according to ISO 9360: 2000 standard. Ann Fr
Anesth Reanim 2008;27:148-53. Epub 2007 Dec 31.
35. Le Bourdells G, Mier L, Fiquet B, Djedani K, Saumon G,
Coste F, Dreyfuss D. Comparison of the effects of heat
and moisture exchangers and heated humidifiers on
ventilation and gas exchange during weaning trials from
mechanical ventilation. Chest 1996 ; 110:1294-8.
36. Girault C, Breton L, Richard JC, et al. Mechanical effects
of airway humidification devices in difficult to wean pa-
tients. Critical Care Medicine 2003;31:1306-1311.
37. Martin C, Perrin G, Gevaudan M, Saux P, Gouin F. Heat
and moisture exchangers and vaporizing humidifiers in
the intensive care unit. Chest 1990;97:144149.
38. Roustan JP, Kienlen J, Aubas P, Aubas S, Cailar JD.
Comparison of hydrophobic heat and moisture exchang-
ers with heated humidifier during prolonged mechani-
cal ventilation. Intensive Care Med 1992;18:97100.
39. Jaber S, Pigeot J, Fodil R, et al. Long-term effects of
different humidification systems on endotracheal tube
patency: Evaluation by the Acoustic Reflection Method.Anesthesiology 2004;100:782-788.
40. Cook D, De Jonghe B, Brochard L, Brun-Buisson C.
Influence of airway management on ventilator-associ-
ated pneumonia: evidence from randomized trials. JAMA
1998;279:781787.
41. Kola A, Eckmanns T, Gastmeier P. Efficacy of heat and
moisture exchangers in preventing ventilator-associated
pneumonia: meta-analysis of randomized controlled tri-
als. Intensive Care Med 2005;31:511.
42. Kirton OC, De Haven B, Morgan J, Morejon O, Civetta
J. A prospective, randomized comparison of an in-line
heat moisture exchange filter and heated wire humidifi-
ers. Rates of ventilator-associated early-onset (commu-
nity-acquired) or late-onset (hospital-acquired) pneu-
monia and incidence of endotracheal tube occlusion.
Chest 1997;112:10551059.
43. Kranabetter R, Leier M, Kammermeier D, Just HM,
Heuser D. The effects of active and passive humidifica-
tion on ventilation-associated nosocomial pneumonia.
Anaesthetist 2004;53:29-5.
44. Lorente L, Lecuona M, Jimnez A, Mora ML, and Sierra
A. Ventilator-associated pneumonia using a heated hu-
midifier or a heat and moisture exchanger: a randomizedcontrolled trial. Crit Care 2006; 10: R116.
45. Lacherade JC, Auburtin M, Cerf C, et al. Impact of hu-
midification systems on ventilator-associated pneumo-
nia: a randomized multicentre trial. Am J Respir Crit Care
Med 2005;172:1276-1282.
46. Garibaldi RA, Britt MR, Webster C, Pace NL. Failure of
bacterial filters to reduce the incidence of pneumonia
after inhalation anesthesia. Anesthesiology
1981;54:364368.
47. Feeley TW, Hamilton WK, Xavier B, Moyers J, Eger EI.
Sterile anesthesia breathing circuits do not prevent post-
-
7/27/2019 Inhalation and Humidification
10/10
641
operative pulmonary infection. Anesthesiology
1981;54:369372.
48. Lorente L, Lecuona M, Mlaga J, Revert C, Mora ML,
Sierra A. Bacterial filters in respiratory circuits: an un-
necessary cost? Crit Care Med 2003;31:21262130.
49. Hess DR. And now for the rest of the store. Respir Care
2002;47:696699.
50. Misset B, Escudier B, Rivara D, Leclercq B, Nitenberg G.
Heat and moisture exchanger vs heated humidifier dur-
ing long-term mechanical ventilation. A prospective ran-
domized study. Chest 1991;100:160163.
V Saraswat. Inhalation therapy and humidification
Election Notification 2008
The following office bearers of ISA HQ are retiring on 30th December 2008:
1. Dr. K. R. N. Tagore - President
2. Dr. B.V.R. Sastry - Vice President
3. Dr. Anjan Datta - Hon. Secretary
4. Dr. Bijendra Kumar Roy - Treasurer5. Dr. Rajib Bhattacharya - Governing Council Member
6. Dr. Rajnish Kumar Jain - Governing Council Member
Therefore Election to the Governing Council ISA 2008 will be held at conference venue
of ISACON 2008 at S N Medical College, Jodhpur. Vacancies open for the year 2008 are as
follows and nominations are hereby invited for:
1. The President
2. The Vice President
3. Hon. Secretary
4. Treasurer
5. Governing Council Members three (3) posts
Rules and regulations regarding eligibility and the election are as par Article VII of the
Memorandum of Association of ISA.
Last Date for receiving nomination at ISAHQ: 28th November08
Last date for withdrawal of nomination: 26th December08 by 12 noon
Publication of Valid Nominations: 27th December08 at 1 PM
Polling: 28th December08 from 9AM till 4 PM
Specimen nomination form is available at our web site. One representative of each candi-
date will be allowed to be present at the designated area at the polling booth for observation. All
candidates seeking election may inform the name of their representative to the Chief ElectionOfficer in writing by 27th December08. Any sort of canvassing inside the polling arena is prohib-
ited. MEMBERS WITH VALID PHOTO ID CARD WITH BARCODE WILL BE ALLOWED
TO CAST THEIR VOTE.
Requests for bar code will be entertain till 15th November 2008 at HQ.
Dr. Anjan Datta
31st August08 Hon. Secretary, ISA (HQ)