Influence of implant number on the movement of …...RESEARCH AND EDUCATION Influence of implant...

6
RESEARCH AND EDUCATION Inuence of implant number on the movement of mandibular implant overdentures Ken Oda, DDS, a Manabu Kanazawa, DDS, PhD, b Shin Takeshita, DDS, PhD, c and Shunsuke Minakuchi, DDS, PhD d The percentage of elderly in- dividuals in the population has been increasing in a number of countries. 1,2 The number of people with edentulism is expected to grow with the in- crease in the older population, 3,4 indicating that the demand for treatment for edentulism will also increase. The use of an implant overdenture (IOD) has been regarded as an effective treat- ment option for edentulous mandibles 5,6 and has been widely used in clinical prac- tice. 7-9 However, the rotational movement of the IOD has a negative effect on the perceived masticatory ability of denture wearers. 10 Thus, rotational movement must be prevented to enhance the quality of life of patients with edentulism. The number of implants should be reduced because of the cost of treatment and the invasiveness of the operation. Consensus statements have Supported by the Grant-in-Aid for Scientic Research from the Japan Society for the Promotion of Science, grant 15K20429. a Graduate student, Gerodontology and Oral Rehabilitation, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, Yushima, Bunkyo, Tokyo, Japan. b Assistant Professor, Gerodontology and Oral Rehabilitation, Tokyo Medical and Dental University, Tokyo, Japan. c Resident, Gerodontology and Oral Rehabilitation, Tokyo Medical and Dental University, Tokyo, Japan. d Professor, Gerodontology and Oral Rehabilitation, Tokyo Medical and Dental University, Yushima, Bunkyo, Tokyo, Japan. ABSTRACT Statement of problem. The rotational movement of an implant overdenture (IOD) has a negative effect on the perceived masticatory ability of the denture wearer. However, the inuence of implant number on the movement of IODs has not been investigated. Purpose. The purpose of this in vitro study was to evaluate the denture movement of mandibular IODs anchored by different numbers of implants. Material and methods. An edentulous mandibular test model with articial mucosa and 5 experi- mental overdentures (N=5) was fabricated. The locator attachment system with blue nylon inserts was chosen for this study. Three implant positions were prepared: anterior midline (1-IOD), bilateral lateral incisor regions (2-IOD), and anterior midline and bilateral canine regions (3-IOD). Vertical loads of 50 N were applied to the mid-anterior region, the left canine region, the left premolar region, and the left rst molar region. The vertical and horizontal displacements at the right distal edge and the vertical displacements at the loading point were measured. The displacement values were statistically analyzed using a 1-way analysis of variance and the post hoc Tukey honest signicant difference test with the implant number as a factor. In addition, the values of the vertical and horizontal displacements at the distal edge of the overdenture were statistically compared using a paired t test, and the values of the vertical displacement at the distal edge of the overdenture were statistically analyzed by a repeated measures analysis of variance and the post hoc Tukey honest signicance difference test with the loading point as a factor (a=.05). Results. Upon anterior loading, the 2-IOD showed signicantly larger vertical displacements at the right distal edge than the 1-IOD or 3-IOD (P<.01). The horizontal displacements at the right distal edge were small compared with the vertical displacements at the same point, although the displacement of the 2-IOD was signicantly larger than that of the 1-IOD upon anterior loading (P=.03). The magnitude of the vertical displacement at the loading point of the 2-IOD was signicantly larger than that of the 3-IOD upon anterior loading (P<.01). No statistically signicant differences were shown under the other conditions for each loading (P>.05). Conclusions. Within the limitations of this in vitro study, the following conclusions were drawn: During mastication with the anterior teeth, the use of 2 implants for anchoring an IOD increased the rotation of the denture base more than the use of 1 or 3 implants. The horizontal movements of the IODs were small compared with the vertical movements. Denture movement under the occlusal force in the molar region was smaller than that in the anterior region. (J Prosthet Dent 2017;117:380-385) 380 THE JOURNAL OF PROSTHETIC DENTISTRY

Transcript of Influence of implant number on the movement of …...RESEARCH AND EDUCATION Influence of implant...

Page 1: Influence of implant number on the movement of …...RESEARCH AND EDUCATION Influence of implant number on the movement of mandibular implant overdentures Ken Oda, DDS,a Manabu Kanazawa,

RESEARCH AND EDUCATION

Supported byaGraduate stuBunkyo, TokbAssistant PrcResident, GdProfessor, G

380

Influence of implant number on the movement of mandibularimplant overdentures

Ken Oda, DDS,a Manabu Kanazawa, DDS, PhD,b Shin Takeshita, DDS, PhD,c andShunsuke Minakuchi, DDS, PhDd

ABSTRACTStatement of problem. The rotational movement of an implant overdenture (IOD) has a negativeeffect on the perceived masticatory ability of the denture wearer. However, the influence of implantnumber on the movement of IODs has not been investigated.

Purpose. The purpose of this in vitro study was to evaluate the denture movement of mandibularIODs anchored by different numbers of implants.

Material and methods. An edentulous mandibular test model with artificial mucosa and 5 experi-mental overdentures (N=5) was fabricated. The locator attachment systemwith blue nylon inserts waschosen for this study. Three implant positions were prepared: anterior midline (1-IOD), bilateral lateralincisor regions (2-IOD), and anterior midline and bilateral canine regions (3-IOD). Vertical loads of 50 Nwere applied to the mid-anterior region, the left canine region, the left premolar region, and the leftfirst molar region. The vertical and horizontal displacements at the right distal edge and the verticaldisplacements at the loading point were measured. The displacement values were statisticallyanalyzed using a 1-way analysis of variance and the post hoc Tukey honest significant differencetest with the implant number as a factor. In addition, the values of the vertical and horizontaldisplacements at the distal edge of the overdenture were statistically compared using a paired ttest, and the values of the vertical displacement at the distal edge of the overdenture werestatistically analyzed by a repeated measures analysis of variance and the post hoc Tukey honestsignificance difference test with the loading point as a factor (a=.05).

Results. Upon anterior loading, the 2-IOD showed significantly larger vertical displacements at theright distal edge than the 1-IOD or 3-IOD (P<.01). The horizontal displacements at the right distaledge were small compared with the vertical displacements at the same point, although thedisplacement of the 2-IOD was significantly larger than that of the 1-IOD upon anterior loading(P=.03). The magnitude of the vertical displacement at the loading point of the 2-IOD wassignificantly larger than that of the 3-IOD upon anterior loading (P<.01). No statisticallysignificant differences were shown under the other conditions for each loading (P>.05).

Conclusions.Within the limitationsof this in vitro study, the following conclusionsweredrawn:Duringmasticationwith the anterior teeth, the useof 2 implants for anchoring an IOD increased the rotation ofthe denture base more than the use of 1 or 3 implants. The horizontal movements of the IODs weresmall compared with the vertical movements. Denture movement under the occlusal force in themolar region was smaller than that in the anterior region. (J Prosthet Dent 2017;117:380-385)

The percentage of elderly in-dividuals in the population hasbeen increasing in a number ofcountries.1,2 The number ofpeople with edentulism isexpected to grow with the in-crease in the older population,3,4

indicating that the demand fortreatment for edentulism willalso increase.

The use of an implantoverdenture (IOD) has beenregarded as an effective treat-ment option for edentulousmandibles5,6 and has beenwidely used in clinical prac-tice.7-9 However, the rotationalmovement of the IOD has anegative effect on the perceivedmasticatory ability of denturewearers.10 Thus, rotationalmovement must be preventedto enhance the quality of life ofpatients with edentulism.

The number of implantsshould be reduced because ofthe cost of treatment and theinvasiveness of the operation.Consensus statements have

the Grant-in-Aid for Scientific Research from the Japan Society for the Promotion of Science, grant 15K20429.dent, Gerodontology and Oral Rehabilitation, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, Yushima,yo, Japan.ofessor, Gerodontology and Oral Rehabilitation, Tokyo Medical and Dental University, Tokyo, Japan.erodontology and Oral Rehabilitation, Tokyo Medical and Dental University, Tokyo, Japan.erodontology and Oral Rehabilitation, Tokyo Medical and Dental University, Yushima, Bunkyo, Tokyo, Japan.

THE JOURNAL OF PROSTHETIC DENTISTRY

Page 2: Influence of implant number on the movement of …...RESEARCH AND EDUCATION Influence of implant number on the movement of mandibular implant overdentures Ken Oda, DDS,a Manabu Kanazawa,

Figure 1. Edentulous mandibular model with silicone artificial mucosa.Five implant analogs placed in mid-anterior region, bilateral lateralincisor regions, and bilateral canine regions.

Figure 2. Experimental overdenture and 5 metal caps with blue nyloninserts (without framework).

Clinical ImplicationsMandibular 2-implant overdentures move easilyand rotate around the fulcrum line formed by 2implants during mastication with the anterior teeth.The use of 1 or 3 implants may prevent thismovement, improving denture stability and themasticatory ability of patients with edentulousmandibles.

March 2017 381

cited strong evidence for the use of 2 implants as anadequate number to retain a mandibular IOD.5,6 Recently,an overdenture retained by a single implant placed in themid-anterior region (1-IOD) has been used in a clinicalapplication, and high implant survival rates and highpatient satisfaction have been reported.11-14 Furthermore,Misch15 proposed that an overdenture retained by 3implants (3-IOD) produced less force and less prosthesismovement than that of an overdenture retained by 2implants (2-IOD). Therefore, the 3-IOD has also beenused as a treatment option for edentulous mandibles.16,17

IOD movement under an occlusal load has beeninvestigated using an in vitro model that simulates theintraoral environment. Studies by Manju and Sreelal18 andTokuhisa et al19 compared the movements of 2-IODs withball, magnet, and bar attachments and reported that the2-IOD with magnetic attachments showed the mostmovement. Haruta et al20 measured the movement of2-IODs with ball and magnet attachments and reportedthat the movement was larger with the ball attachmentthan with the magnetic attachment. Maeda et al21

measured the movements of 1-IODs and 2-IODs withball and magnet attachments and reported that themovements of the 1-IODs were larger than those of the2-IODswith themagnetic attachment. Liu et al22measuredthe denture base movement and lateral force on the im-plants with IODs retained by 1 to 4 implants, using a finiteelement model. They concluded that the 2-IOD showedthe largest denture movement and the largest lateral force.

These in vitro studies have focused on the differencesamong the attachment systems.18-21 Only the studies byMaeda et al21 and Liu et al22 compared denture movementamong IODs with different numbers of implants. Thesestudies reported inconsistent findings. We are unaware ofany study that measures the denture movement of IODsretained by 1, 2, and 3 implants by using an edentulousmandibular model. Therefore, whether the number ofimplants influences the movement of IODs is not clear.

The purpose of this in vitro study was to evaluate themovement of mandibular IODs anchored by differentnumbers of implants. The null hypothesis was that theimplant number would have no influence on the denturebase movement of mandibular IODs.

Oda et al

MATERIAL AND METHODS

An edentulous mandibular test model was fabricated froman acrylic resin (Acron Clear; GC Corp). A 3-mm-thick layer of the model surface was replaced with siliconeimpression material (Exahiflex regular; GC Corp) tosimulate the mucosa. The elastic modulus of the siliconematerials was reported to be in the same range as that ofthe oral mucosa.23 Five implant replicas (Implant ReplicaInternal tri-channel connection RP; Nobel Biocare) wereplaced parallel to each other in the anterior midline,bilateral lateral incisor regions, and bilateral canineregions (Fig. 1). All tests were performed on this model.

Five experimental overdentures without a metal frame-work were fabricated from a multipurpose pourable resin(Procast DSP clear; GC Corp). Locator attachment systemswere chosen for this study. Fivemetal caps (metal cap;NobelBiocare) and blue nylon inserts (Extra Light RetentionReplacementMale [blue]; Nobel Biocare) were incorporatedinto the intaglio surface in the anterior midline, bilateral

THE JOURNAL OF PROSTHETIC DENTISTRY

Page 3: Influence of implant number on the movement of …...RESEARCH AND EDUCATION Influence of implant number on the movement of mandibular implant overdentures Ken Oda, DDS,a Manabu Kanazawa,

Figure 3. Implant positions. A, Single implant in mid-anterior region(1-IOD). B, Two implants in bilateral lateral incisor regions (2-IOD). C, Threeimplants in mid-anterior region and bilateral canine regions (3-IOD).

Figure 4. Loading points and denture base movement. Loading pointsin mid-anterior region (Anterior), left canine region (Canine), leftpremolar region (Premolar), and left molar region (Molar). Two types ofdenture base movement were measured: horizontal (X) and vertical (Y)displacement of right distal edge.

382 Volume 117 Issue 3

lateral incisor regions, and bilateral canine regions of eachoverdenture (Fig. 2). Three implant positions were preparedat the anteriormidline (1-IOD), the lateral incisor regions (2-IOD), and the midline and canine regions (3-IOD) (Fig. 3).These implant positionswere reproduced on themandibularmodel by changing the position of the locator abutments.

THE JOURNAL OF PROSTHETIC DENTISTRY

Loading tests were performed using a universal testingmachine (InstronModel 5544; Instron Corp). Vertical loadsof 50 N, reported to be the average occlusal force in pa-tients with edentulism,24 were applied on the occlusalsurface of the experimental overdentures. The loadingpoints were the mid-anterior region; the mid-anteriorpoint of the denture base, in the left canine region; 14mm apart from the mid-anterior point, in the left premolarregion; 10 mm apart from the canine point and left firstmolar region; and 10 mm apart from the premolar point.The load was applied with a 5-mm-diameter acrylic ball.The vertical and horizontal displacements at the right distaledge of the overdenture and the vertical displacement atthe loading point were measured under each loadingcondition (Fig. 4). The vertical and horizontal displace-ments at the right distal edge were measured with move-ment sensors that apply strain gauges (DTH-A-5; KyowaElectronic Instruments Co Ltd) (Fig. 5). The movementsensors were placed at the right distal edge and connectedto a computer with sensor interface boards (PCD300A;Kyowa Electronic Instruments Co Ltd). The vertical dis-placements at the loading point were recorded by theuniversal testing machine at the same time as loading.

The displacement values were statistically analyzedusing multiple comparison 1-way analysis of variance(ANOVA) and post hoc Tukey honest significant differ-ence (HSD) test with the number of implants as a factor.In addition, the vertical and horizontal displacementvalues at the distal edge of the overdenture werestatistically compared using the paired t test, and thevalues of the vertical displacement at the distal edge of theoverdenture were statistically analyzed by repeated mea-sures ANOVA and post hoc Tukey HSD test, using theloading point as a factor (SPSS Statistics v17.0; SPSS Inc)(a=.05).

Oda et al

Page 4: Influence of implant number on the movement of …...RESEARCH AND EDUCATION Influence of implant number on the movement of mandibular implant overdentures Ken Oda, DDS,a Manabu Kanazawa,

Figure 5. Positions of movement sensors. A, Horizontal direction. B, Vertical direction.

Table 1.Mean ±SD of vertical displacement at right distal edge ofoverdenture (mm)

LoadingPoint 1-IOD 2-IOD 3-IOD P

Anterior 0.65 ±0.09 1.06 ±0.08 0.64 ±0.08 1-IOD versus 2-IOD: <.01*

2-IOD versus 3-IOD: <.01*

1-IOD versus 3-IOD: .99

Canine 0.87 ±0.25 0.97 ±0.34 0.75 ±0.11 1-IOD versus 2-IOD: .78

2-IOD versus 3-IOD: .36

1-IOD versus 3-IOD: .74

Premolar 0.22 ±0.08 0.27 ±0.15 0.15 ±0.07 1-IOD versus 2-IOD: .68

2-IOD versus 3-IOD: .20

1-IOD versus 3-IOD: .59

Molar 0.05 ±0.02 0.03 ±0.02 0.03 ±0.03 1-IOD versus 2-IOD: .26

2-IOD versus 3-IOD: .99

1-IOD versus 3-IOD: .28

IOD, implant overdenture. *P<.05 Tukey honest significant difference test.

Table 2.Mean ±SD of horizontal displacement at right distal edge ofoverdenture (mm)

Loadingpoint 1-IOD 2-IOD 3-IOD P

Anterior -0.01 ±0.04 0.07 ±0.05 0.04 ±0.04 1-IOD versus 2-IOD: .03*

2-IOD versus 3-IOD: .58

1-IOD versus 3-IOD: .18

Canine -0.07 ±0.02 -0.04 ±0.03 -0.05 ±0.03 1-IOD versus 2-IOD: .17

2-IOD versus 3-IOD: .56

1-IOD versus 3-IOD: .65

Premolar -0.06 ±0.0 -0.06 ±0.04 -0.06 ±0.02 1-IOD versus 2-IOD: .96

2-IOD versus 3-IOD: >.99

1-IOD versus 3-IOD: .96

Molar -0.05 ±0.04 -0.09 ±0.03 -0.07 ±0.02 1-IOD versus 2-IOD: .13

2-IOD versus 3-IOD: .41

1-IOD versus 3-IOD: .72

IOD, implant overdenture. *P<.05 Tukey honest significant difference test.

Table 3. Comparison of mean ±SD vertical versus horizontal displace-ment on all loadings (mm)

Type of IOD

Displacement Direction

PVertical Horizontal

1-IOD 0.44 ±0.36 -0.05 ±0.04 <.01*

2-IOD 0.58 ±0.49 -0.03 ±0.07 <.01*

3-IOD 0.39 ±0.33 -0.04 ±0.05 <.01*

IOD, implant overdenture. *P<.05 paired t test.

Table 4.Mean ±SD of vertical displacement at loading point ofoverdenture (mm)

Loadingpoint 1-IOD 2-IOD 3-IOD P

Anterior -0.80 ±0.09 -0.99 ±0.13 -0.66 ±0.13 1-IOD versus 2-IOD: .05

2-IOD versus 3-IOD: <.01a

1-IOD versus 3-IOD: .16

Canine -0.77 ±0.09 -0.65 ±0.14 -0.64 ±0.10 1-IOD versus 2-IOD: .23

2-IOD versus 3-IOD: .99

1-IOD versus 3-IOD: .21

Premolar -0.52 ±0.13 -0.49 ±0.08 -0.39 ±0.03 1-IOD versus 2-IOD: .83

2-IOD versus 3-IOD: .20

1-IOD versus 3-IOD: .07

Molar -0.61 ±0.10 -0.54 ±0.13 -0.48 ±0.07 1-IOD versus 2-IOD: .53

2-IOD versus 3-IOD: .57

1-IOD versus 3-IOD: .12

IOD, implant overdenture. aP<.05 Tukey honest significant difference test.

March 2017 383

RESULTS

Table 1 shows the average vertical displacement at the rightdistal edge of the overdenture. The 2-IOD showedsignificantly larger displacements than the 1-IOD or the 3-IOD on anterior loading (P<.01). The displacementresulting from anterior and canine loading was slightlylarger than that resulting frompremolar andmolar loading,but the differences were not statistically significant (P>.05).

Table 2 shows the average horizontal displacement atthe right distal edge of the overdenture. The horizontaldisplacement of the 2-IOD was significantly larger than

Oda et al

that of the 1-IOD upon anterior loading (P=.03).Furthermore, statistically significant differences were notdetected under the other conditions. However, regardlessof the loading point, the horizontal displacements were

THE JOURNAL OF PROSTHETIC DENTISTRY

Page 5: Influence of implant number on the movement of …...RESEARCH AND EDUCATION Influence of implant number on the movement of mandibular implant overdentures Ken Oda, DDS,a Manabu Kanazawa,

Table 5.Mean ±SD of vertical displacement at right distal edge ofoverdenture upon each loading point (mm)

Loading Point P

Anterior 0.78 ±0.22 A versus C: .59

Canine 0.86 ±0.25 A versus P: <.01*

Premolar 0.21 ±0.11 A versus M: <.01*

Molar 0.03 ±0.02 C versus P: <.01*

C versus M: <.01*

P versus M: <.04*

A, anterior loading; C, canine; IOD, implant overdenture; M, molar; P, premolar. *P<.05Tukey honest significant difference test.

384 Volume 117 Issue 3

small compared with the vertical displacements at allimplant positions (P<.01) (Table 3).

Table 4 indicates the vertical displacement of the over-denture at the loading point. The magnitude of thedisplacement of the 2-IOD at the loading pointwas significantly larger than that of the 3-IODuponanteriorloading (P< .01), while no statistically significant differenceswere observed under the other conditions (P>.05).

Table 5 shows the average of all vertical displace-ments at the distal edge of the overdenture on eachloading point. Statistically significant differences existedamong all loading points (P<.05), except between ante-rior and canine loading. The value of the displacementson molar loading was the smallest, followed by premolar,canine, and anterior loading.

DISCUSSION

The results of this study support the hypothesis that theimplant number affects the denture base movement of amandibular IOD upon anterior loading. The null hy-pothesis of the study was rejected for anterior loading butnot for canine, premolar, and molar loading.

Figure 6. Denture base movement in vertical direction. A, Overdenture rotatwith molar loading.

THE JOURNAL OF PROSTHETIC DENTISTRY

The 2-IOD showed significantly larger vertical dis-placements at the right distal edge than the 1-IOD uponanterior loading, which is in agreement with previousresearch using a finite element model.22 This result maybe because the 2-IOD rotated easily around the fulcrumline formed between the 2 implants. By contrast, the1-IOD rotated little because of the proper fit between thedenture base and the edentulous mandibular model usedin this study and because the 1-IOD showed smallervertical displacements. Under actual oral conditions, the1-IOD may allow complex denture movements because asingle implant cannot limit the direction of movement.The experimental overdentures had an appropriatefit with the alveolar ridge in this in vitro study, preventingdenture rotation under occlusal loading. The 1-IOD maybe an effective treatment option for the edentulousmandible, with a low cost and reduced invasiveness.

The vertical displacement of the 3-IOD at the rightdistal edge upon anterior loading was significantlysmaller than that of the 2-IOD, supporting the proposalof Mericske-Stern et al25 that the use of 3 or moreimplants prevents denture rotation. Denture baserotation has a negative effect on perceived masticatoryability and patient satisfaction10 and must be minimized.The 3-IOD treatment may be more effective than that of2-IOD for patients with edentulism to prevent denturerotation if the cost and invasiveness required for 3implants are acceptable.

Regardless of the implant number, upon anterior andcanine loading, the vertical displacements at the rightdistal edge and the loading point of the overdenturesuggest that the IODwas pressed into the mandible at theloading point and raised above it at the distal edge. Thisresulted in rotation around the implant. In contrast, upon

ed around implants on anterior loading. B, Minimal overdenture rotation

Oda et al

Page 6: Influence of implant number on the movement of …...RESEARCH AND EDUCATION Influence of implant number on the movement of mandibular implant overdentures Ken Oda, DDS,a Manabu Kanazawa,

March 2017 385

premolar and molar loading, although the IOD waspressed into the mandible at the loading point, the IODhardly rotated and was slightly raised at the distal edge(Fig. 6). The denture-bearing area of the molar region islarger than that of the anterior region and resisted occlusalforcemore strongly. Therefore, the IODhardly rotated andwas slightly raised in the distal region. The horizontaldisplacements at the right distal edge measured in thisstudy were small; therefore, they hardly affected denturestability. This study suggests that the IOD moved mainlyin a vertical direction on anterior loading, simulating theincising of food with the anterior teeth. To prevent thismovement, an IOD should not be given definitive occlusalcontacts in the anterior area.

The actual denture movement in a patient’s oral cavityis not same as that observed in in vitro mandibularmodels, and denture movements should be measured inthe actual oral cavity. However, measuring denturemovement intraorally is difficult as the denture move-ment is smaller than the movement of the mandibleduring mastication. Therefore, denture movement wasmeasured using a simple model which could notcompletely simulate clinical situations such as behavior ofthe occlusal load. In addition, the success of the IODtreatment depends on complex factors such as forceapplied to the implants, retention of dentures, andprosthodontic complications. Further studies investi-gating these factors are required.

CONCLUSIONS

Within the limitations of this in vitro study, the followingconclusions were drawn:

1. The use of 2 implants for anchoring an IOD resultsin easier rotation of the denture base duringmastication with the anterior teeth than the use of1 or 3 implants.

2. The horizontal movements of IODs are smallcompared with the vertical movements.

3. Denture movement under occlusal force in themolar region is small compared with that in theanterior region.

REFERENCES

1. Anderson GF, Hussey PS. Population aging: a comparison among industri-alized countries. Health Aff (Millwood) 2000;19:191-203.

2. Petersen PE, Yamamoto T. Improving the oral health of older people: theapproach of the WHO Global Oral Health Programme. Community DentOral Epidemiol 2005;33:81-92.

3. Douglass CW, Shih A, Ostry L. Will there be a need for complete dentures inthe United States in 2020? J Prosthet Dent 2002;87:5-8.

4. Allen PF, McMillan AS. A longitudinal study of quality of life outcomes inolder adults requesting implant prostheses and complete removable den-tures. Clin Oral Implants Res 2003;14:173-9.

Oda et al

5. Feine JS, Carlsson GE, Awad MA, Chehade A, Duncan WJ, Gizani S, et al.The McGill consensus statement on overdentures. Mandibular two-implantoverdentures as first choice standard of care for edentulous patients.Gerodontology 2002;19:3-4.

6. Thomason JM, Feine J, Exley C, Moynihan P, Müller F, Naert I, et al.Mandibular two implant-supported overdentures as the first choice standardof care for edentulous patientsdthe York consensus statement. Br Dent J2009;207:185-6.

7. Sato D, Kanazawa M, Kim YK, Yokoyama S, Omura Y, Ozeki M, et al.Immediate loading of two freestanding implants placed by computer-guided flapless surgery supporting a mandibular overdenture with magneticattachments. J Prosthodont Res 2016;60:54-62.

8. Matsuda K, Kurushima Y, Enoki K, Ikebe K, Maeda Y. Replacement of amandibular implant-fixed prosthesis with an implant-supported overdentureto improve maintenance and care. J Prosthodont Res 2014;58:132-6.

9. Elsyad MA, Khirallah AS. Circumferential bone loss around splinted andnonsplinted immediately loaded implants retaining mandibularoverdentures: a randomized controlled clinical trial using cone beamcomputed tomography. J Prosthet Dent 2016;116:741-8.

10. Kimoto S, Pan S, Drolet N, Feine JS. Rotational movements of mandibulartwo-implant overdentures. Clin Oral Implants Res 2009;20:838-43.

11. Cordioli G, Majzoub Z, Castagna S. Mandibular overdentures anchored tosingle implants: a five-year prospective study. J Prosthet Dent 1997;78:159-65.

12. Wolfart S, Braasch K, Brunzel S, Kern M. The central single implant in theedentulous mandible: improvement of function and quality of life. A report of2 cases. Quintessence Int 2008;39:541-8.

13. Walton JN, Glick N, Macentee MI. A randomized clinical trial comparingpatient satisfaction and prosthetic outcomes with mandibular over-dentures retained by one or two implants. Int J Prosthodont 2009;22:331-9.

14. Kronstrom M, Davis B, Loney R, Gerrow J, Hollender L. A prospectiverandomized study on the immediate loading of mandibular overdenturessupported by one or two implants: a 3 year follow-up report. Clin ImplantDent Relat Res 2014;16:323-9.

15. Misch CE. Contemporary implant dentistry. 3rd ed. St. Louis: Mosby; 2007.p. 293-313.

16. Mericske-Stern R. Clinical evaluation of overdenture restorations supportedby osseointegrated titanium implants: a retrospective study. Int J OralMaxillofac Implants 1990;5:375-83.

17. El-Sheikh AM, Shihabuddin OF, Ghoraba SM. Two versus threenarrow-diameter implants with locator attachments supportingmandibular overdentures: a two-year prospective study. Int J Dent2012;2012:285684.

18. Manju V, Sreelal T. Mandibular implant-supported overdenture: an in vitrocomparison of ball, bar, and magnetic attachments. J Oral Implantol 2013;39:302-7.

19. Tokuhisa M, Matsushita Y, Koyano K. In vitro study of a mandibular implantoverdenture retained with ball, magnet, or bar attachments: comparison ofload transfer and denture stability. Int J Prosthodont 2003;16:128-34.

20. Haruta A, Matsushita Y, Tsukiyama Y, Sawae Y, Sakai N, Koyano K. Effects ofmucosal thickness on the stress distribution and denture stability ofmandibular implant-supported overdentures with unsplinted attachmentsin vitro. J Dent Biomech 2011;2:894395.

21. Maeda Y, Horisaka M, Yagi K. Biomechanical rationale for a single implant-retained mandibular overdenture: an in vitro study. Clin Oral Implants Res2008;19:271-5.

22. Liu J, Pan S, Dong J, Mo Z, Fan Y, Feng H. Influence of implant number on thebiomechanical behaviour of mandibular implant-retained/supported over-dentures: a three-dimensional finite element analysis. J Dent 2013;41:241-9.

23. Inoue K, Arikawa H, Fujii K, Shinohara N, Kawahata N. Viscoelastic prop-erties of oral soft tissue. 1. A method of determining elastic modulus of oralsoft tissue. Dent Mater J 1985;4:47-53.

24. Garrett NR, Kaurich M, Perez P, Kapur KK. Masseter muscle activity indenture wearers with superior and poor masticatory performance. J ProsthetDent 1995;74:628-36.

25. Mericske-Stern RD, Taylor TD, Belser U. Management of the edentulouspatient. Clin Oral Implants Res 2000;11(suppl 1):108-25.

Corresponding author:Dr Manabu Kanazawa1-5-45, Yushima, BunkyoTokyo 113-8549JAPANEmail: [email protected]

Copyright © 2016 by the Editorial Council for The Journal of Prosthetic Dentistry.

THE JOURNAL OF PROSTHETIC DENTISTRY