Infant Resuscitation | Technical Manual
|
N E O P U F F ™ t E c h N i c a l m a N U a l 1
TABLE OF CONTENTS
1. Specifications .....................................................................................................................................................................2
1.1 Warnings ..........................................................................................................................................................................2
1.2 PerformanceSpecifications ..............................................................................................................................................2
1.3 TechnicalSpecifications ...................................................................................................................................................2
1.4 SymbolDefinitions ...........................................................................................................................................................3
1.5 LabelIdentification ..........................................................................................................................................................3
2. Cleaning .............................................................................................................................................................................4
2.1 CleaningProductstoAvoid ...............................................................................................................................................4
2.2 CleaningoftheTestLung ..................................................................................................................................................4
2.2.1 Latex(black)TestLung(discontinuedMay2007) .................................................................................................4
2.2.2 Silicone(blue)TestLungRD020-01 .......................................................................................................................5
2.3 CleaningReusableT-Piece(discontinuedSeptember2008) ..............................................................................................5
2.4 Sterilization ......................................................................................................................................................................5
2.5 AfterCleaning ...................................................................................................................................................................5
3. Service Information .............................................................................................................................................................6
3.1 FunctionalSchematic .......................................................................................................................................................6
3.2 InstallationChecksandPreventativeMaintenance ...........................................................................................................6
3.2.1 InstallationChecks ................................................................................................................................................6
3.2.2PreventativeMaintenance ......................................................................................................................................6
3.3 TestingoftheF&PNeopuff™/Perivent™Performance(ManometerandValveSystem) .....................................................6
3.3.1 TestingoftheManometer ......................................................................................................................................7
3.3.2 TestingoftheValveSystem ...................................................................................................................................7
3.3.3 SettingMaxPressureReliefto40 cmh2O ...............................................................................................................7
3.3.4 ResettingtheManometertoZero ...........................................................................................................................7
3.3.5 ManometerReplacement ......................................................................................................................................7
3.3.6 ValveSystemReplacement ....................................................................................................................................7
3.4 Set-upandMaintenanceChecklist ...................................................................................................................................8
4. Assembly Diagrams .............................................................................................................................................................9
4.1 F&PNeopuff™/Perivent™InfantResuscitator(PostJune2010) ........................................................................................9
4.1.1 assembly Diagram ................................................................................................................................................9
4.1.2 PartsList ...............................................................................................................................................................9
4.1.3 SparesKit ...........................................................................................................................................................10
5. Mounting Options .............................................................................................................................................................11
5.1 900RD301Side-MountingBlock .....................................................................................................................................11
5.2 900MR088RailMountBracket .......................................................................................................................................11
5.3 900RD302 WallMountBracket .......................................................................................................................................11
5.4 RD050-01PoleandRailCentralMount ...........................................................................................................................12
N E O P U F F ™ t E c h N i c a l m a N U a l 2
1.1 Warnings
1.3 Technical Specifications
Peak Inspiratory Pressure (PIP) Range
@ 5 L/min approx. 2 to 70 cmh2O [mbar]
@ 8 L/min approx.3 to 72 cmh2O [mbar]
@ 10 L/min approx.4 to 73 cmh2O [mbar]
@ 15 L/min approx.8 to 75 cmh2O [mbar]
Positive End Expiratory Pressure (PEEP) Range
@ 5 L/min approx.1 to 6 cmh2O [mbar]
@ 8 L/min approx.1 to 10 cmh2O [mbar]
@ 10 L/min approx. 2 to 15 cmh2O [mbar]
@ 15 L/min approx.4 to 17 cmh2O [mbar]
Input Gas Flow Range
minimum 5L/min
Maximum 15L/min
Operating Time (400 L Cylinder)
5 L/min 80minutes
10 L/min 40minutes
15 L/min 26minutes
NOTE:Allperformancefigureslistedabovearerepresentativeonly. PEEPvaluesstatedarebasedontypicalclinicalPIPsettings. HigherPEEPvaluescanbeachievedifhigherPIPvaluesareset.
Height 250mm(9.8”)
Width 200mm(7.9”)
Depth 104mm(4.1”)
Weight 1.9kg(4.2lb)
Manometer Range -10to80 cmh2O [mbar]
Manometer Accuracy +/-2.0%ofFullScaleDeflection
Maximum Pressure Setting 65 to 80 cmh2O [mbar]
(dependentonflowrate)
900RD010 / 900RD110 Dead Space 6 ml
RD1300-10 Dead Space 3.3 ml
Storage Temperature Range -10to50°C(+14to+122°F),upto95%humidity
Operating Temperature Ranges
Humidified Circuit
Non-Humidified Circuit
+18to26°C(+64to+78°F),30 - 75%humidity
-18to50°C(-0.4to+122°F),upto95%humidity
Recommended Patient Body Weight
0 to 10kg(22lb)
Delivered Oxygen Concentration Upto100% depending on gas supply
Inputgasflowrate5 to15L/min.Recommended operating gas
flow rate 8 L/min.Inputflowrangesarecircuitspecific,referto
circuitUserInstructions.
Do not attempt to use a higher gas flow rate than 15 L/min.
Factorysettingofpressurelimitvalveisat40 cmh2O.Thissetting
isuseradjustableupto80 cmh2O. Do not attempt to set the
pressure relief valve above 80 cmH2O. Reset pressure relief valve
to 40 cmH2O after use (refer section 3.3.3 of this manual).
UseonlyrecommendedF&PNeopuff™/Perivent™Infant
Resuscitatoraccessories.
Use only a F&P gas supply line or approved equivalent.
Donotuseoil,greaseorothersubstancesthatareincompatible
withoxygenonanypartoftheF&PNeopuff/Perivent
T-PieceCircuit.
Incorrectusecanbehazardous.
Thedevicemustnotbeusedonunattendedpatients.
Tobeusedforresuscitationonly.
The operation of the F&P Neopuff / Perivent must be checked prior
to first use.
Refertosection 3.3“TestingoftheF&PNeopuff/Perivent
Performance”.
1.2 Performance Specifications
1. SPECiFiCATiONS
N E O P U F F ™ t E c h N i c a l m a N U a l 3
SYMBOL DEFiNiTiON
Attention:ConsulttheOperatingInstructions
MAX-P SetstheMax Pressure Relief thatmaybedeliveredtothepatient (factorysetat40 cmh
2O)
PiP ControlsthePeak Inspiratory Pressure deliveredtothepatient
5-15 L/min Gas inletconnectionfromgassupply(5 to 15litresperminute)
Gas outletconnectiontopatient
1.4 Symbol Definitions
1. SPECiFiCATiONS continued
1.5 Label identification
REF - Thisdetailstheproductcodewhichidentifiestheoperating
instructionlanguageandfasciatype
LOT - Theproductiondate:(10) = 2010, (06) =June,
(15) =fifteenthday
SN- TheserialnumberisspecifictoaparticularF&PNeopuff/
PeriventInfantResuscitator.ThefirstsixdigitsmatchtheLOT
numberandlastsixdigitsidentifythespecificproductserial
number.
Thebarcodeconsistsofthefollowingpredefinedapplicationidentifiers:
01 -Notused
0-Sequencestart
94-Countryofmanufacture-NewZealand
200124-CompanyIdentifier-Fisher&PaykelHealthcareLtd
1093-ProductCode
1-CheckSum
REF - RD900XXX
RD900 XXX Operating Instruction Language Fascia Type
900 Series
T-Piece
Resuscitator
aEU English EnglishText
aDU Danish Symbols
OnlyaFU French
alU Italian
amU Finnish
aNU Dutch
aSU Spanish
atU Portuguese
AVU Norwegian
aWU Swedish
AZU chinese
aGU German(Perivent)
ThelabelpicturedistypicaloftheinformationcontainedonaF&PNeopuff™/Perivent™InfantResuscitator.
N E O P U F F ™ t E c h N i c a l m a N U a l 4
2. CLEANiNg
• Complywithhospital,localandnationalguidelinesforproduct
cleaning frequencies.
• Ensurealloxygenandairsuppliesareturnedoffanddisconnected
fromtheF&PNeopuff™/Perivent™InfantResuscitatorbefore
performingcleaningprocedures.Explosionandfirehazardscan
existwhenperformingcleaningproceduresinanoxygen-enriched
environment.
• Cleaningshallbeperformedatambientconditions.
• Beforecleaning,removeanddiscardalluseddisposableproducts
usingtherecommendedmethodofdisposal.
• Dustallsurfaceswithacleandampsoftcloth.
• Cleanallplasticsurfaceswithdetergent-basedsolution(maximum
2%inwater)ensuringthemanufacturer’sdirectionsforuseofthe
cleaningagentarefollowed.
• Thefollowingproprietarychemicalcleaningwipesare
recommendediftheF&PNeopuff/Periventfasciarequirescleaning
forinfectioncontrolpurposes.
Caviwipes(MetrexResearchCorporation)
Sani-clothHB(ProfessionalDisposables,Inc.)
Asepti-WipesII(Ecolab,Inc.)
Spartan’sTB-CideQuatWipes
VernacareTuffieWipesAlcoholFree
• Dryallsurfacesaftercleaningwithacleansoftclothor
papertowel.
CAUTION: EnsurenopartoftheF&PNeopuff/PeriventInfant
Resuscitatororrelatedaccessoriesisimmersedinanycleaningliquid
orcleaningsolution.ForTestLungcleaninginstructions,seesection
2.2.2.
CAUTION: DonotcleantheF&PNeopuff/PeriventInfantResuscitator
fasciawithproprietarycleaningproductscontainingeither
hydroxides,hypochlorites,peroxides,gluteraldehydeorcleaning
productswithagreaterthan30% alcohol base.
NOTE:Therecommendedchemicalcleaningwipeslistedabovehave
beencheckedforlong-termcompatibilitywiththeIW900SeriesInfant
WarmersandtheF&PNeopuff/PeriventInfantResuscitator.
2.1 Cleaning Products to Avoid
CAUTION: Thechemicalsusedintheseproprietarycleaningproducts
mayleadtodiscoloration,crazingandeventualcrackingofthefascia.
Examplesofproprietarycleaningproductswhichcontainsuch
cleaningchemicalsincludebutarenotlimitedto:
Asepti-Wipes(Ecolab,Inc)
Clorox(TheCloroxCompany)
Endbac 256(JohnsonWaxProfessional)
Quat(3M)
Sporiciden(LibertyIndustries,Inc)
SporoxII(SultanHealthcare)
Steris(SterisCorporation)
Terralin(Schülke&Mayr)
Virtek(AVirtekCompany)
Virox(ViroxTechnologies)
CAUTION: Donotuseabrasivecleaningsolutions.
CAUTION: EnsureF&PNeopuff/PeriventInfantResuscitatorand
accessoriesarecheckedbeforereturningtheInfantResuscitatorto
service.
2.2 Cleaning of the Test Lung
Ifrequired,theTestLungcanbedisinfected.Therearetwodifferent
modelsofTestLungandeachhasspecificcleaningmethods.
• TheTestLungisaconsumableitem.Irrespectiveofmodel,the
TestLungshouldregularlybeinspectedforsignsofdamagesuch
asdiscoloration,perishingorcracking.ReplacetheTestLungif
damageisobserved.
2.2.1 Latex (black) Test Lung (discontinued May 2007)
WARNINGTheTestLungcontainsnaturalrubberlatexthatmay
causeallergicreactions.
CAUTION:Latexissusceptibletoattackbysolvents.Ensurenosolvents
areusedtocleantheTestLung.
ThisTestLungconsistsoftwoparts,theflexiblelatextestlungand
arigidblackplasticconnector.TheTestLungcanbedisassembled
fordisinfection.Itcanbedisinfectedusingethyleneoxidegasonly.
Somecarriergasescancausestresscrackingandarenotsuitable.If
indoubt,checkwiththechemicalsupplier.
N E O P U F F ™ t E c h N i c a l m a N U a l 5
2.2.2 Silicone (blue) Test Lung RD020-01
TheTestLungmustbecleanedbeforeeveryuseorin
accordancewithhospitalcleaningprotocols.Thistestlungconsistsof
twoparts,theflexiblesiliconetestlungandarigidplasticconnector.
ThetwopartsoftheTestLungmaybedisassembled
fordisinfection.Itcanbecleanedusingdisinfectantscontaining
eitherperaceticacid(e.g.Perasafe®)ororthopthaldehyde
(e.g.Cidex®OPA).Otherdisinfectionmethodscancausedamage
totheTestLungandarenotrecommended.
Pleasefollowthemanufacturer’sinstructionsforstorage,handling
andusageofthesechemicals.
WARNING EnsurethesiliconeTestLungandplasticconnectorare
bothremovedfromtheT-Piecebeforeuseonapatient.
2.3 Cleaning Reusable T-Piece (discontinued September 2008)
ThereusableT-Piece500RD104consistsoftwopartsandcanbe
disassembledfordisinfection(referdiagram).TheT-Pieceshould
bedisinfectedbyautoclavingatupto136°C,220kPafor4
minutes.Followingreassembly,theT-Pieceshouldbetestedprior
tousetoensurethatitisfunctioningcorrectly.RefertotheF&P
Neopuff™/Perivent™OperatingManual(Fisher&PaykelHealthcare
PartNo.185041726)forset-upinstructions.
UnscrewPEEPcapfordisinfection
2.4 Sterilization
CAUTION:
• Donotautoclaveorgas-sterilizeanypartofthe
F&PNeopuff™/Perivent™InfantResuscitator.
• DonotautoclaveanyoftheNeopuff/Periventaccessories,
e.g.TestLungandGasSupplyLine.
• Forcoldsterilization:Ensurethecoldsterilizationagentsaresafe
forusewiththerelevantsurfaces.
2.5 After Cleaning
CAUTION:
• EnsureallF&PNeopuff/PeriventInfantResuscitatorpartsand
accessoriesarecheckedbeforereturningthedevicetoservice.
PlasticConnectorSiliconeTestLung
2. CLEANiNg continued
N E O P U F F ™ t E c h N i c a l m a N U a l 6
3.1 Functional Schematic
3.2 installation Checks and Preventative Maintenance
WARNINGDroppingtheF&PNeopuff™/Perivent™Infant
Resuscitatororothersimilarformsofimpactmaycausedamage
resultinginincorrectoperationoftheunit.Ifyoususpectdamage
tohaveoccurred,pleaseperformchecksasoutlinedinsection3.3
beforeconnectiontoapatient.
3.2.1 installation Checks
Priortofirstuse:
–Removemanometercover
–CompletePerformanceTesting(section 3.3)
3.2.2 Preventative Maintenance
• Theintegrityofthesystemandmanometershouldbecheckedprior
tofirstuse,annuallyandafterservicing,usingthe
“TestingoftheF&PNeopuff/PeriventPerformance”procedure
(section3.3).
• Theuseofamountingbrackettohelppreventthe
F&PNeopuff/Periventbeingdamagedordroppedisrecommended.
Mountingsolutionsavailablecanbefoundinsection5ofthis
manual.
• Allmaintenanceandserviceproceduresmustbeperformedby
qualifiedpersonnelusingonlyFisher&PaykelHealthcareparts.
• Alwaysensuregaspassagesarefreefromcontaminants,especially
hydrocarbons,oilsandgrease,priortoreassembly.
• PleasecontactanauthorizedFisher&PaykelHealthcare
representativeforfurtherassistancewithanyservicingor
maintenancerequirements.
• TheTestLungisaconsumableitem;itshouldbemonitoredfor
signsofwearandmaterialdegradationandreplacedasrequired.
3.3 Testing of the F&P Neopuff™ /Perivent™ Performance (Manometer and Valve System)
TheintegrityoftheF&PNeopuff/Periventmanometerandvalve
systemcanbetestedusingthefollowingguidelines.Theinletport
mustbeconnectedtoagassupplycapableofgeneratingconstant
flowat5,10 and 15L/min,andacertifiedgaugeandbleedvalve
shouldbeavailabletocheckthemanometeraccuracy.The
F&PNeopuff/PeriventresuscitationcircuitandT-Piececan
beusedinplaceofableedvalveasshownbelow.
60 7080
-20100 -10
20
30
4050
cmH2O+–
60 7080
100 -10
20
30
4050
cmH2O+–
Manometer
GasOutletPortGasInletPort
CircuitPressureManometer
Maximum Pressure ReliefValve
Peak Inspiratory PressureValve
PeakInspiratoryPressure Control
OutletPort
MaximumPressure ReliefCover
InletPort
Connectpressuregauge betweenF&PNeopuff/Periventoutletandresuscitationcircuit
UseT-Piecewith TestLungattached asableedvalve
CertifiedGauge
3. SERViCE iNFORMATiON
N E O P U F F ™ t E c h N i c a l m a N U a l 7
3.3.1 Testing of the Manometer
1. LiftthecoverofftheMaxPressureReliefValveslightlyandturnoutoftheway.
2. DisconnectalldevicesfromtheF&PNeopuff™/Perivent™outletport.Checkthatthemanometerneedleiswithin+/-2 cmh
2O of
zeroonthemanometergauge.Ifthemanometerdoesnotreadzero,theresettingofthemanometertozeroprocedure(section3.3.4)shouldbefollowed.
3. ConnecttheoutletoftheF&PNeopuff/PeriventInfantResuscitatortoableedvalveandacertifiedgauge(e.g.MensorDigitalPressureGauge Series 2400).Setthegassupplyto10L/min.Completelyclosethemaximumpressurelimitvalvebyturningtheleft-handknobcompletelyclockwise.Withthebleedvalveclosed,adjustthePeakInspiratoryPressureknobtosetthepressuresothatthecertifiedgaugereads10, 20 and 40 cmh
2O.Checkthatthe
manometerreadswithin+/-2 cmh2Oofthesevaluesateachset
point.
4. Withthepressuresetto40 cmh2O,openandclosethebleedvalve
threetimesandcheckthemanometerneedlerisesandfallssmoothly.If the F&P Neopuff / Perivent Infant Resuscitator fails any of these tests, the manometer should be regarded as inaccurate and replaced with a new manometer (Part No. 043040841).
Follow the manometer replacement guidelines in section 3.3.5 of this manual, or contact your Fisher & Paykel Healthcare service representative for further guidance.
3.3.2 Testing of the Valve System
1. Setthegassupplyto5L/min.CompletelyclosethePeakInspiratoryPressurecontrolandtheMaxPressurecontrolbyturningbothknobscompletelyclockwise.Closethebleedvalveandcheckthatthegaugereadsatleast60 cmh
2O.
2. Setthegassupplyto15L/min.Closethebleedvalveandcheckthatthegaugereadsnohigherthan80cmH
2O.
3. Setthegassupplyto10L/min.ClosethebleedvalveandturntheMaxPressureReliefvalveuntilthemanometerreads40 cmh
2O.
Checkthatthemanometerneedlerisesandfallssmoothly.GentlyrotatetheMaxPressureReliefcoverovertheMaxPressureReliefknob.
4. Resetthepeakinspiratorypressureto20 cmh2Oandturnoffthe
gasflow.Testingisnowcomplete.
If the F&P Neopuff / Perivent Infant Resuscitator fails any of these tests, the valve assembly should be regarded as faulty and replaced with a new valve assembly (see the parts list (section 4.1.2) for part numbers). Follow the valve replacement guidelines in section 3.3.6 of this manual, or call your Fisher & Paykel Healthcare service representative for further information.
3.3.3 Setting Max Pressure Relief to 40 cmH2O
ThisisrequirediftheMaxPressureReliefhasbeenchanged.ThefactorysettingfortheMaxPressureReliefis40 cmh
2O.
AlternativesettingsfortheMaxPressureReliefshouldbemadeinaccordancewithhospitalprotocol.
1. Adjustgasflowto10L/min.
2. ClosethePeakInspiratoryPressurevalvebyturningtheknobfullyclockwise.
3. AdjusttheMaxPressureReliefknobclockwiseorcounterclockwiseuntilthemanometerreads40 cmh
2O.
4. TurnthePeakInspiratoryPressureknobcounterclockwisesothemanometerreads20 cmh
2Oandshutoffthegasflow.
3.3.4 Resetting the Manometer to Zero
Tosetthemanometertozero:
1. DisconnecttheF&PNeopuff™/Perivent™InfantResuscitatorfromanyotherequipment.
2. Removetheopaqueplasticpluginthelensofthemanometer.
3. Usingasuitableslotscrewdriver,carefullyadjustthescrewinthemanometerfaceclockwiseorcounterclockwisetoresetthemanometertozero.Caremustbetakenwhendoingthis,asover-rotationofthescrewcandamagethemanometerinternals.
4. Replacetheplasticpluginthelensofmanometer.
5. Verifythatthemanometerneedleisnowwithin+/-2.0 cmh2O of
zero.Ifnot,repeattheaboveprocedure.
3.3.5 Manometer Replacement
ThemanometerisnotaserviceableitemandmustbereplacedbyManometerKit043040841.
1. Removethebackcover,fixedbyfourscrews.
2. Disconnectthetubefromthemanometer.
3. Removethemanometerbyunscrewingthetworetainingnuts.
4. Fitthenewmanometerintothefrontpanel,tightentheretainingnutsandreconnectthemanometertube.
5. Refitthefrontpaneltothebackcoverwiththefourscrews.
6. Carryoutthemanometerperformancetestaspersection3.3.1.Itisrecommendedtorecordthelotnumberfromtheboxlabelofthereplacementmanometeronthemaintenancechecklist.
3.3.6 Valve System Replacement
NOTE:Thevalvesareanintegralpartofthevalve,panelandmanifoldassemblyandarenotabletobeserviced.Pleasespecifythemodelnumberfromthepartslist(section4.1.2)whenorderingareplacementvalveassembly.
1. Removethebackcover,fixedbyfourscrews.
2. Disconnectthetubefrommanometer.
3. Removethemanometerbyunscrewingthetworetainingnuts.
4. Fittheexistingmanometerintothenewvalveassemblypanel,tightentheretainingnutsandreconnectthemanometertube.
5.Refitthefrontpaneltothebackcoverwiththefourscrewsfromstep1.
6.Carryoutthevalvesystemperformancetestaspersection3.2.2. ItisrecommendedtorecordthelotnumberofthenewvalveassemblyontotheSet-upandMaintenanceChecklist.
3. SERViCE iNFORMATiON continued
N E O P U F F ™ t E c h N i c a l m a N U a l 8
Cust
omer
Det
ails
Name:
address:
Country:
Phone Number:
Email:
F&P
Neo
puff
™/P
eriv
ent™
Perf
orm
ance
Tes
ting
Rec
ord TestDate:
Serial Number:
For 900iW130 only:InfantWarmerSerialNumber:
TestedBy:
NextTestDueDate:
Check Ref or x Comments
Manometer Checks:
ReplacementManometerSerialNumber:
accuracy @ 0 cm (+/-2 cmh2O) 3.3.1-2
accuracy @ 10 cm (+/-2 cmh2O) 3.3.1-3
accuracy @ 20 cm (+/-2 cmh2O) 3.3.1-3
accuracy @ 40 cm (+/-2 cmh2O) 3.3.1-3
Needlemovementsmooth? 3.3.1-4
Valve System Checks:
Pressuregreaterthan60 cmh2O? 3.3.2-1
Pressurelessthan80 cmh2O? 3.3.2-2
Maximumpressuresetto40 cmh2O 3.3.2-3
PIPresetto20 cmh2O 3.3.2-4
Additional Comments:
3.4 Set-up and Maintenance Checklist
Thefollowingtableisprovidedtorecordtheresultsoftheperformancetestsdescribedinsection3.3.Anycomponentsreplacedshouldbe
recordedalso,asappropriate.Thetablemaybephotocopiedorotherwisereproducedasrequired.
Upon receipt of the product and prior to use on a patient, please complete the tests and fill in the table below.
3. SERViCE iNFORMATiON continued
N E O P U F F ™ t E c h N i c a l m a N U a l 9
4.1 F&P Neopuff™ / Perivent™ infant Resuscitator (Post June 2010)
ThisF&PNeopuff/Periventmodulecontainsasealedvalveassembly
thatcannotbeservicedinthefield.Ifafaultyvalveoccurs,theentire
panelandvalveassemblymustbereplaced.
4.1.1 Assembly Diagram
4. ASSEMBLY DiAgRAMS
UNiTS MANUFACTURED POST JUNE 2010
# Description Part Number Reqd
1 handle Neopuff Spare 043043976 1
2 Plug(setoffour) 693040706 2
3 ScrewM8x20 614040309 8
4 Endcap(upper) 043042565 1
5 Endcap(lower) 043042564 1
6 Backcover 641040816 1
7model specific panel and valveassembly (seeSparesKitsection4.1.3)
(seetableinsection4.1.3)
1
8 Manometerkit 043040841 1
9 CapValveNeopuffSpare(Blue) 043043977 1
10 Cover,maximumpressurereliefvalve 043041057 1
11 Foot 693041436 4
12 Screw#8x1”Csk 616050011 4
13ScrewM4x8 Pan hd(handleattachment)
614040117 2
NOTE:Whitevalvecap(pre-June2010)canbeorderedusing043042345(1/bag).
4.1.2 Parts List
N E O P U F F ™ t E c h N i c a l m a N U a l 1 0
Model specific panel and valve assembly Blue (kit includes fascia)
043043595 Fascia&ValveAssyBlueEnglish
043043596 Fascia&ValveAssyBlueGerman
043043597 Fascia&ValveAssyBlueItalian
043043598 Fascia&ValveAssyBlueSpanish
043043599 Fascia&ValveAssyBlueFrench
043043600 Fascia&ValveAssyBlueNorwegian
043043601 Fascia&ValveAssyBlueDutch
043043602 Fascia&ValveAssyBlueSwedish
043043603 Fascia&ValveAssyBluePortuguese
043043604 Fascia&ValveAssyBlueDanish
043043605 Fascia&ValveAssyBlueFinnish
Model specific panel and valve assembly Purple (kit includes fascia)
043042347 Fascia&ValveAssyEnglish
043042348 Fascia&ValveAssyGerman
043042349 Fascia&ValveAssyItalian
043042350 Fascia&ValveAssySpanish
043042351 Fascia&ValveAssyFrench
043042353 Fascia&ValveAssyDutch
043042354 Fascia&ValveAssySwedish
043042355 Fascia&ValveAssyPortuguese
043042356 Fascia&ValveAssyDanish
043042357 Fascia&ValveAssyFinnish
043042358 Fascia&ValveAssyNorwegian
4. ASSEMBLY DiAgRAMS continued
NOtE: all RD900unitsmanufacturedfromMay1999 can
accommodatetheabovespareskit.
4.1.3 Spares Kit
N E O P U F F ™ t E c h N i c a l m a N U a l 1 1
ImpacttotheF&PNeopuff™/Perivent™InfantResuscitator,
causedbyroughhandlingortheunitbeingdropped,candamagethe
valvesystemandproduceirregularresuscitationpressures.
Tohelppreventanyimpacttothedevice,Fisher&PaykelHealthcare
recommendstheuseofoneofthemountingsystemsshownbelow.
5.1 900RD301 Side-Mounting Block, 900MR170 C-Clamp, 900MR030 Pole Bracket
the 900RD301SideMountingBlockfitsintothedovetailslotonthe
sideoftheInfantResuscitator.The900RD301maythenbeconnected
toa17 to 40mmpoleusingeitherthe900MR170C-Clampor
900MR030PoleBracket.
5-15 L/minGAS INLETGAS OUTLET
900RD301
900MR170 C-Clamp
900MR030PoleBracket
Mounting Option Parts to order
Quickdisconnect(C-Clamp) 900RD301 & 900MR170
PermanentPoleBracket 900RD301 & 900MR030
5.2 900MR088 Rail Bracket
FormountingtheF&PNeopuff/PeriventInfantResuscitatorcentrally
onstandardrails(2.5 to 5.5cmx1cm/0.98 to 2.17”x0.39”).
AlsorequiresRD050-01.
5. MOUNTiNg OPTiONS
5.3 900RD302 Wall Mount Bracket
the 900RD302WallMountBrackethastwotonguefittingsthat
maybescrewedtothebackoftheF&PNeopuff™/Perivent™Infant
Resuscitatorbyremovingthefourplasticplugsfromtherearpanel,
andaT-shapedbracketthatshouldbescrewedtothewall.
Thisallowstheinfantresuscitatortobequicklyandeasilyremoved
fromandreplacedonthebracketasrequired.
N E O P U F F ™ t E c h N i c a l m a N U a l 1 2
5.4 RD050-01 Pole and Rail Central Mount
FormountingtheF&PNeopuff™/Perivent™InfantResuscitator
centrallyonapoleoronstandardrails(2.5 to 5.5cmx1cm/
0.98to2.17”x0.39”),acentralmountingblockcanbeaffixedto
thebackoftheunitbyremovingthefourplasticplugsfromtherear
panel.TheunitwillthenfittheF&Pbracketandclamps.
5. MOUNTiNg OPTiONS continued
www.fphcare.com
For more information please contact your local Fisher & Paykel Healthcare representative
1850
4159
7 REVm
EN
© 2
011Fisher&PaykelHealthcareLimited
AustraliaTel: +61 3 9879 5022Fax: +61 3 9879 5232
AustriaTel: 0800 29 31 23Fax: 0800 29 31 22
BeneluxTel: +31 40 216 3555Fax: +31 40 216 3554
ChinaTel: +86 20 3205 3486Fax: +86 20 3205 2132
FranceTel: +33 1 6446 5201 Fax: +33 1 6446 5221
GermanyTel: +49 7181 98599 0 Fax: +49 7181 98599 66
IndiaTel: +91 80 4284 4000 Fax: +91 80 4123 6044
Irish RepublicTel: 1800 409 011
ItalyTel: +39 06 7839 2939 Fax: +39 06 7814 7709
JapanTel: +81 3 3661 7205 Fax: +81 3 3661 7206
Northern IrelandTel: 0800 132 189
SpainTel: +34 902 013 346Fax: +34 902 013 379
SwedenTel: +46 8 564 76 680Fax: +46 8 36 63 10
SwitzerlandTel: 0800 83 47 63Fax: 0800 83 47 54
TaiwanTel: +886 2 8751 1739Fax: +886 2 8751 5625
TurkeyTel: +90 312 354 34 12Fax: +90 312 354 31 01
UK (EU Authorised Representative)Fisher & Paykel Healthcare Ltd Unit 16, Cordwallis Park Clivemont Road, Maidenhead Berkshire SL6 7BU, UK
Tel: +44 1628 626 136 Fax: +44 1628 626 146
USA/CanadaTel: +1 800 446 3908 or +1 949 453 4000 Fax: +1 949 453 4001
Manufacturer Fisher & Paykel Healthcare Ltd 15 Maurice Paykel Place East Tamaki, Auckland 2013
PO Box 14 348 PanmureAuckland 1741 New Zealand
Tel: +64 9 574 0100 Fax: +64 9 574 0158 Email: [email protected] Web: www.fphcare.com