In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*,...
-
date post
21-Dec-2015 -
Category
Documents
-
view
213 -
download
1
Transcript of In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*,...
![Page 1: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/1.jpg)
In-Vivo Pan/Tilt Endoscope with Integrated Light Source
Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.***Department of Computer Science
**Department of SurgeryColumbia University, New York
![Page 2: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/2.jpg)
Outline Motivation Introduction Proposed Solution Prototype Device I and II Animal Tests Results and Conclusion Contributions and Future Work Acknowledgement
![Page 3: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/3.jpg)
Minimally Invasive Surgery
![Page 4: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/4.jpg)
Limitations of Standard Endoscope Paradigm of pushing
long sticks into small openings. Narrow angle imaging. Limited workspace. Counter intuitive
motion. Assistants needed to
control the camera. Additional incisions
for other laparoscopic instruments.
![Page 5: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/5.jpg)
Robotic Surgery Robotic surgery is future of surgery.
Increase the dexterity and fine motion capabilities of surgeon.
Decrease the tremor of surgeon’s hand. Enable remote operation.
Current surgical robot. Enormous size. Extremely high cost. Multiple assistants needed. Multiple incisions needed as non-robotic MIS.
Compact and inexpensive surgical robot is needed.
![Page 6: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/6.jpg)
In-Vivo Imaging Devices Rod-lens by Hopkins and cold light source of
fiber optics by Karl Storz. Flexible endoscope using fiber optics to
delivery light and transmit image. Pill camera without locomotion. Endoscope with rotating mirror. Endoscope positioned by multilink arm with
piezoelectric actuators.
Gao et al., 1998
Pill camera
Karl Storz Endoscope
Flexible Endoscope
Ikuta et. al., 2002
![Page 7: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/7.jpg)
Proposed Solution An insertable surgical robot platform with sensors and
effectors in the body cavity where they can perform surgical & imaging tasks unfettered by traditional endoscopic instrument design.
A prototype of a novel insertable Pan/Tilt endoscope with integrated light source.
The incision port is left open access, allowing for single port surgery.
![Page 8: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/8.jpg)
Prototype Device I
Design of 5-DOF insertable camera device
Stereo camera with 6 DOF are desirable – full mobility
Difficult to achieve in small actuated package
Compromise – 3 DOF per camera Cameras share tilt axis
(1 DOF) Independent translation
(2 DOF) Independent pan (2
DOF)
![Page 9: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/9.jpg)
Single Camera Prototype IDiameter: 22 mm; Length: 190 mmCamera opening: 58 mm3 DOF: Pan: 120°; Tilt: 130°; Translation: 50 mm
Video
![Page 10: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/10.jpg)
Improvements Needed for Prototype II
Much reduced size: Smaller diameter Reduce length
Needs light source Make imaging head modular design Tradeoff: Degrees-of-freedom for
compactness
![Page 11: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/11.jpg)
Prototype II 110 mm in length and 11
mm in diameter. 120 degrees Pan/ 90 degrees
Tilt. Integrated 8 LED light
source and miniature camera module.
Package to protect the delicate electronics and fragile wires from body fluid and moisture. Fully sealed camera
module. Joint sealed by rubber boot.
Joystick control.
![Page 12: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/12.jpg)
Prototype II (Cont.)
![Page 13: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/13.jpg)
LED Light Source Light-emitting diode (LED) as
a light source in laparoscopy: Lower power Higher efficiency Compact package Longer lifespan Lower cost
Luxeon portable PWT white LED(LXCL_PWT1) 2.0 X1.6 X 0.7 mm 26 lumens of light at 350 mA
8 PWT LED in a printed circuit board with 9mm diameter. 208 lumens light at 8.4 w
![Page 14: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/14.jpg)
Camera Module
Pin hole lens (PTS 5.0 from Universe Kogaku America) Focal length 5.0 mm. F number 4. Angle of view D-H-V(85.4-68.3-50.9 ).
6.5 mm CCD camera sensor. NET USA Inc, CSH-1.4-V4-END-R1. 450 TV lines in horizontal resolution and 420 TV lines in vertical resolution.
Protective window by sapphire.
![Page 15: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/15.jpg)
Pan/Tilt Mechanism
Miniature Brushless DC motor (0513G, Faulhaber Group). 25mNm output torque. 5.8 mm in diameter.
Miniature worm gear (Kleiss Gear Inc.) Gear ratio 16:1. Compact size. Increased torque.
Sleeve bearing to reduce the friction of tilt motion.
![Page 16: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/16.jpg)
Working Device in the Animal Test
![Page 17: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/17.jpg)
Laparoscopic Procedures
Running (measuring) the bowel Appendectomy
Suturing Nephrectomy
Video
![Page 18: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/18.jpg)
Timing of Each Procedures for Laparoscope and our device
Procedure Laparoscope(minutes)
Robot(minutes)
Running Bowel 4:20 3:30
Appendectomy 2:20 2:20
Suturing 5:00 4:00
![Page 19: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/19.jpg)
Conclusion Easier and more intuitive to use than a
standard laparoscope. Joystick operation requires no specialized
operator training. Field of view and access to relevant regions
of the body were superior to a standard laparoscope using a single port.
Time to perform procedures was better or equivalent to a standard laparoscope.
![Page 20: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/20.jpg)
Contributions and Future Work Contributions
Built an insertable camera with Pan/Tilt and integrated LED light source.
Accomplished a series of standard laparoscopic procedures( appendectomy, running (measuring) the bowel, suturing, and nephrectomy) by laparoscopic surgeon using the device.
Future Work Camera with zoom and auto-focusing capabilities. Stereo camera in one package. Function with tool or organ tracking.
![Page 21: In-Vivo Pan/Tilt Endoscope with Integrated Light Source Tie Hu, Ph. D*, Peter K. Allen, Ph. D*, Dennis L. Fowler, M.D.** *Department of Computer Science.](https://reader035.fdocuments.us/reader035/viewer/2022062714/56649d565503460f94a33cbb/html5/thumbnails/21.jpg)
Acknowledgement
We thank Nancy Hogle for her help in project development and lab support.
This work was supported by NIH grant1R21EB004999-01A1.