In vivo dosimetry Eirik Malinen Eva Stabell Bergstrand Dag Rune Olsen.
-
Upload
ginger-thornton -
Category
Documents
-
view
234 -
download
1
Transcript of In vivo dosimetry Eirik Malinen Eva Stabell Bergstrand Dag Rune Olsen.
In vivo dosimetry
Eirik MalinenEva Stabell Bergstrand
Dag Rune Olsen
In vivo dosimetry
• In vivo: In the living• Dosimetry: Estimates of radiation dose by theory and
measurement• Verification of delivered
dose to individual patients• Radiotherapy requires
accurate dose delivery
error
Prescribed dose
Pro
bali
lity
Errors in patient dose
• Patient contour / planning basis (CT images)• Patient motion • Organ motion• Dose calculations (inhomogeneities, scatter)• Patient positioning• Transfer of treatment data from simulator to linac• Linac settings (energy, monitor units, field size) and
calibration• Beam modifiers (blocks, wedges)
Dose characteristics
Dose measurements
Patient curvature
beam
wedgeOutput, SSD
Wedge, curvature
Thickness, density
Entrance dose:
Exit dose:
Point detector
2D detector array
Desired in vivo dosimeter characteristics
• Accurate and precise
• Multiple readouts• Reusability
• No cables
• Non-destructive readout
High accuracyLow precision
Low accuracyHigh precision
In vivo dosimetry principles
• Point detector:– Semiconductors (diodes)– Thermoluminescent crystals– EPR (electron paramagnetic resonance) sensitive
materials– ….
• 2D detector, (electronic) portal imaging device; EPID:– Film– Arrays (ion chambers, semiconductors)
Dosimeter reading → absorbed dose
• Absorbed dose, D:
R: dosimeter reading
ND: calibration factor
Ci: correction factor
ii
D CRND
Calibration
Rcal
Dcal
beam
dmax
water phantom
ion chamber
dosimeter
cal
calD R
DN
• Under reference conditions:
Example – diodes
spherical
droplet
Buildup cap
Correction factors
• Dosimeter reading may depend on:– Temperature– (Accumulated) Dose– Dose rate– Beam energy– Field size– ...
• Accuracy may be reduced if dependence is not corrected
Temperature and sensitivity, diodes
Detector temperature after placing on patient
Sensitivity dependence
• Regular calibration must be performed
Accumulated dose and sensitivity, diodes
Field size and sensitivity, diodes
8 or 18 MV photonsEntrance (in) or exit (out)
Supralinearity, TLD
Energy dependence, TLD
Correction factor forEPR/
alanineTLD Diode
Dose rate 1 1 <1
Linearity 1 <1 1
Beam inclination 1 1> 1
Temperature ≈1 1 <1
Energy ≈1 ≈1 ≈1
Stability ≈1 ≈1 Immediate readout
Total uncertainty(following corrections)
3-4 %(~1 Gy)
2-3 % 2 %
Comparison
Action level
• Relative dose difference:
• At what dose difference level should the treatment be revised? 1% ? 2.5 % ? 5 %?
• Depends on:– dosimetric accuracy and precision– non-systematic errors– …
prescribed
measured
D
Dr 1
Clinical example
Methods
Portal image profile
Measured dose / prescribed dose
Action level: 2.5%
measured dose
dose after correction
%2.1
008.1r
Frequency distribution of relative dose
2D dose maps
Treatment planning algorithm Portal image
Collapsed cone algorithm Location of normalization point
Novel methods – ”dose guided radiotherapy”
dose image
Backprojection of filtered dose image into patient image
→OK
→correction
target
prescribed isodose
Novel methods – ”dose guided radiotherapy”
Corrections
bladder
prosta
te
rectum