In-vitro measurement of Intra-Ocular Pressure
Alex LuceUndergraduate Research Symposium 2008
![Page 2: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/2.jpg)
Motivation: Treat Glaucoma
Glaucoma: a degenerative
disease of the eye
•Affects 3 million Americans
•Leading cause of adult
blindness in the world
•No available cure
•Entirely preventable if
caught early!
Normal range of vision
Advanced vision loss due to glaucoma
04/20/23 Undergraduate Research Symposium - Alex Luce
2
![Page 3: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/3.jpg)
Intraocular Pressure (IOP)The major risk factor for glaucoma
Normal eye pressure : 10-21 mm Hg
10X increased risk to develop glaucoma: >23mm Hg
Measurement of the intraocular pressure (IOP) with an applanation tonometer
04/20/23 3Undergraduate Research Symposium - Alex Luce
![Page 4: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/4.jpg)
Methods to Measure IOP
Indirect Methods:•Applanation Tonometry (ex. Goldman Applanation Tonometry)
•MacKay-Marg tonometry
•Digital Palpation Tonometry
Measurement of IOP with Dynamic Contour Tonometer. Measurements must be performed in clinical setting.
Problems:•Diurnal pressure variation (24 hr cycle)•Laser eye surgery•Requires anesthesia •Potential for infection•Cannot be completed at home
Photo credit: http://en.wikipedia.org/wiki/Image:PASCAL_tonometer.jpg
04/20/23 4Undergraduate Research Symposium - Alex Luce
![Page 5: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/5.jpg)
Experiment Goal
Utilize concept of digital palpation tonometry for experimental IOP measurements
Determine required sensor specifications
Digital palpation tonometry: fingertips push on eye- ‘feel’ pressure Measure F vs X vs IOP
Disp.F1(x) F2(x)
04/20/23 5Undergraduate Research Symposium - Alex Luce
![Page 6: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/6.jpg)
Force application
Applied
Force F
04/20/23 6Undergraduate Research Symposium - Alex Luce
![Page 7: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/7.jpg)
Force application
04/20/23 7Undergraduate Research Symposium - Alex Luce
![Page 8: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/8.jpg)
Force measurement
Bending-beam load cell•deflection of cantilever induces strain
04/20/23 Undergraduate Research Symposium - Alex Luce
8
Strain Gage•Deformation of metal foil causes change in electrical resistance
![Page 9: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/9.jpg)
Block Diagram of Experiment
04/20/23 Undergraduate Research Symposium - Alex Luce
9
![Page 10: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/10.jpg)
Experimental setup
04/20/23 Undergraduate Research Symposium - Alex Luce
10
![Page 11: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/11.jpg)
04/20/23 11Undergraduate Research Symposium - Alex Luce
![Page 12: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/12.jpg)
04/20/23 12Undergraduate Research Symposium - Alex Luce
Force sensor•Measure F
Micrometer•Measure x
Syringe•Adjust IOP
![Page 13: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/13.jpg)
Force Sensor Calibration
C alibration C urve F S 2
0
0.1
0.2
0.3
0.4
0.5
0 5 10 15
Applied We ig ht (g )
Mea
sure
d v
olt
age
(Vo
lt)
Weights of known value measured with each force sensor
Voltage output measured
Standard deviation 0.3% of measured force
04/20/23 13Undergraduate Research Symposium - Alex Luce
![Page 14: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/14.jpg)
Experimental procedure Adjust IOP: 10-35
mmHG FS1: Increment
micrometer x - apply force F1 to eyeball
FS2: measure force F2
F1 F2
Pressure adjustment
Applied force
04/20/23 14Undergraduate Research Symposium - Alex Luce
Disp. X
![Page 15: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/15.jpg)
Experimental Results
•Increased Pressure•Increased risk of glaucoma
04/20/23 15Undergraduate Research Symposium - Alex Luce
![Page 16: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/16.jpg)
04/20/23 Undergraduate Research Symposium - Alex Luce
16
Disp. (mm)
![Page 17: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/17.jpg)
04/20/23 17Undergraduate Research Symposium - Alex Luce
![Page 18: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/18.jpg)
Conclusions Digital palpation tonometry: a viable IOP
measurement technique
To screen for glaucoma, required sensor resolution is.: F ≤ 0.5N, x ≤ 0.5mm.
Force measurements must be done below ~2.5 N, to avoid errors introduced by motion of eye in model “socket”
04/20/23 18Undergraduate Research Symposium - Alex Luce
![Page 19: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/19.jpg)
Future Work Compare experimental
results of experiment to Finite Element simulation
Develop motorized experiment to mimic human palpation exam institute
Explore means to measure IOP independently of displacement
04/20/23 Undergraduate Research Symposium - Alex Luce
19
![Page 20: In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.](https://reader030.fdocuments.us/reader030/viewer/2022032612/56649f055503460f94c1a615/html5/thumbnails/20.jpg)
Acknowledgements Dr. Eniko Enikov, Dr. Gholam Peyman, Vasco
Polyzoev, Balint Magyar, Ed White Advanced Microsystems Lab Funding from NSF Grant CBET-0603198
04/20/23 20Undergraduate Research Symposium - Alex Luce